Open Access 25.07.2016 | Symposium: Advances in PEEK Technology

The Biotribology of PEEK-on-HXLPE Bearings Is Comparable to Traditional Bearings on a Multidirectional Pin-on-disk Tester

verfasst von: Doruk Baykal, PhD, Ryan S. Siskey, MSc, Richard J. Underwood, PhD, Adam Briscoe, PhD, Steven M. Kurtz, PhD

Erschienen in: Clinical Orthopaedics and Related Research® | Ausgabe 11/2016

Einloggen, um Zugang zu erhalten

Abstract

Background

All-polymer bearings involving polyetheretherketone (PEEK) have been proposed for orthopaedic applications because they may reduce stress shielding, reduce weight of the implants, reduce wear and risk of osteolysis, and prevent release of metal ions by replacing the metal articulating components. Little is known about the biotribology of all-polymer PEEK bearings, including the effects of cross-shear, which are relevant for implant longevity, especially in the hip, and increased temperature that may affect lubricant proteins and, hence, lubrication in the joint.

Questions/purposes

Using pin-on-disk in vitro testing, we asked: (1) Can all-polymer bearing couples involving PEEK have a comparable or lower wear rate than highly crosslinked UHMWPE (HXLPE) on CoCr bearing couples? (2) Is the wear rate of PEEK bearing couples affected by the amount of cross-shear? (3) Is there a difference in wear mechanism and surface morphology for all-polymer bearing surfaces compared with UHMWPE (HXLPE) on CoCr?

Methods

We simultaneously tested a total of 100 pin-on-disk couples (n = 10 per bearing couple) consisting of three traditional metal-on-UHMWPE and seven polymer-on-polymer bearings for 2 million cycles under physiologically relevant conditions and in accordance with ASTM F732. Using analysis of variance, we analyzed the effect of bearing surface topography and cross-shear on wear rate. The changes in surface topography were evaluated using optical microscopy. Sample size was sufficient to provide 80% power to detect a difference of 1.4 mm3/MC in average wear rates of bearing couples.

Results

The combined wear rates of all-polymer bearing couples were not different than traditional bearing couples. With the numbers available, the PEEK and HXLPE bearing couple had a mean wear rate (WR: mean ± SD) of 0.9 ± 1.1 mm3/MC (95% confidence interval [CI], 0.2–1.5 mm3/MC), which was not different than the wear rate of the CoCr and HXLPE bearing couple (1.6 ± 2.0 mm3/MC; 95% CI, 0.4–2.8 mm3/MC; mean difference = 0.73 mm3/MC, p = 0.36). Bearing couples with PEEK reinforced with a carbon fiber (CFR-PEEK) counterface had higher wear rates (14.5 ± 15.1 mm3/MC; 95% CI, 9.1–20.0 mm3/MC) than bearing couples with a PEEK (5.1 ± 3.7 mm3/MC; 95% CI, 3.7–6.4 mm3/MC) or CoCr (4.1 ± 2.7 mm3/MC; 95% CI, 3.2–5.1 mm3/MC) counterface (mean difference = 9.5 mm3/MC, p < 0.001; and mean difference = 10.4 mm3/MC, p < 0.001, respectively). PEEK and HXLPE were insensitive to the cross-shear scenario in the contact mechanics (WR: 0.3 ± 0.1 mm3/MC for PEEK pins [95% CI, 0.2–0.3 mm3/MC] [representing full cross-shear condition] and 0.0 ± 1.0 mm3/MC for PEEK disks [95% CI, −0.5 to 0.5 mm3/MC] [representing limited cross-shear condition], mean difference = 0.3 mm3/MC, p = 0.23; WR: 1.3 ± 1.0 mm3/MC for HXLPE pins [95% CI, 0.7–1.9 mm3/MC] [full cross-shear] and 2.1 ± 2.2 mm3/MC for HXLPE disks [95% CI, 0.8–3.3 mm3/MC] [limited cross-shear], mean difference = 0.8 mm3/MC, p = 0.24). Qualitatively, the surface morphology of UHMWPE appeared similar with PEEK or CoCr as a counterface, although it had a rougher appearance when coupled with carbon fiber-reinforced PEEK. No transfer film was detected on the specimens.

Conclusions

Our in vitro pin-on-disk data suggest that all-polymer bearings, especially PEEK-on-HXLPE bearing couples, may represent a viable alternative to traditional bearings with respect to their wear performance. Our results warrant further testing of all-polymer bearing couples in physiologically relevant joint simulator tests.

Clinical Relevance

The in vitro pin-on-disk wear resistance of all-polymer bearings incorporating PEEK-on-HXLPE warrants further investigation using joint simulator testing for their validation as useful, metal-free alternatives to traditional CoCr-on-HXLPE bearings for use in orthopaedic applications.
Literatur
1.
Zurück zum Zitat ASTM F732-00. Standard Test Method for Wear Testing of Polymeric Materials Used in Total Joint Prostheses. West Conshohocken, PA, USA: ASTM International; 2011. ASTM F732-00. Standard Test Method for Wear Testing of Polymeric Materials Used in Total Joint Prostheses. West Conshohocken, PA, USA: ASTM International; 2011.
2.
Zurück zum Zitat ASTM F1714-96. Standard Guide for Gravimetric Wear Assessment of Prosthetic Hip Designs in Simulator Devices. West Conshohocken, PA, USA: ASTM International; 2013. ASTM F1714-96. Standard Guide for Gravimetric Wear Assessment of Prosthetic Hip Designs in Simulator Devices. West Conshohocken, PA, USA: ASTM International; 2013.
3.
Zurück zum Zitat Baykal D, Siskey RS, Haider H, Saikko V, Ahlroos T, Kurtz SM. Advances in tribological testing of artificial joint biomaterials using multidirectional pin-on-disk testers. J Mech Behav Biomed Mater. 2014;31:117–134.CrossRefPubMed Baykal D, Siskey RS, Haider H, Saikko V, Ahlroos T, Kurtz SM. Advances in tribological testing of artificial joint biomaterials using multidirectional pin-on-disk testers. J Mech Behav Biomed Mater. 2014;31:117–134.CrossRefPubMed
4.
Zurück zum Zitat Blunn G, Walker P, Joshi A, Hardinge K. The dominance of cyclic sliding in producing wear in total knee replacements. Clin Orthop Relat Res. 1991:273:253–260.PubMed Blunn G, Walker P, Joshi A, Hardinge K. The dominance of cyclic sliding in producing wear in total knee replacements. Clin Orthop Relat Res. 1991:273:253–260.PubMed
5.
Zurück zum Zitat Dumbleton J, Shen C. The wear behavior of ultrahigh molecular weight polyethylene. Wear. 1976;37:279–289.CrossRef Dumbleton J, Shen C. The wear behavior of ultrahigh molecular weight polyethylene. Wear. 1976;37:279–289.CrossRef
6.
Zurück zum Zitat East RH, Briscoe A, Unsworth A. Wear of PEEK-OPTIMA(R) and PEEK-OPTIMA(R)-Wear Performance articulating against highly cross-linked polyethylene. Proc Inst Mech Eng [H]. 2015;229:187–193.CrossRef East RH, Briscoe A, Unsworth A. Wear of PEEK-OPTIMA(R) and PEEK-OPTIMA(R)-Wear Performance articulating against highly cross-linked polyethylene. Proc Inst Mech Eng [H]. 2015;229:187–193.CrossRef
7.
Zurück zum Zitat Greenbaum ES, Burroughs BB, Harris WH, Muratoglu OK. Effect of lipid absorption on wear and compressive properties of unirradiated and highly crosslinked UHMWPE: an in vitro experimental model. Biomaterials. 2004;25:4479–4484.CrossRefPubMed Greenbaum ES, Burroughs BB, Harris WH, Muratoglu OK. Effect of lipid absorption on wear and compressive properties of unirradiated and highly crosslinked UHMWPE: an in vitro experimental model. Biomaterials. 2004;25:4479–4484.CrossRefPubMed
8.
Zurück zum Zitat Grupp TM, Utzschneider S, Schröder C, Schwiesau J, Fritz B, Maas A, Blömer W, Jansson V. Biotribology of alternative bearing materials for unicompartmental knee arthroplasty. Acta Biomater. 2010;6:3601–3610.CrossRefPubMed Grupp TM, Utzschneider S, Schröder C, Schwiesau J, Fritz B, Maas A, Blömer W, Jansson V. Biotribology of alternative bearing materials for unicompartmental knee arthroplasty. Acta Biomater. 2010;6:3601–3610.CrossRefPubMed
9.
Zurück zum Zitat Haider H, Baykal D. Wear assessment of UHMWPE with pin-on-disc testing. In: Kurtz SM, ed. UHMWPE Biomaterials Handbook. Waltham, MA, USA: Elsevier; 2016. Haider H, Baykal D. Wear assessment of UHMWPE with pin-on-disc testing. In: Kurtz SM, ed. UHMWPE Biomaterials Handbook. Waltham, MA, USA: Elsevier; 2016.
10.
Zurück zum Zitat Kurtz SM. PEEK Biomaterials Handbook. Waltham, MA, USA: Elsevier; 2011. Kurtz SM. PEEK Biomaterials Handbook. Waltham, MA, USA: Elsevier; 2011.
11.
Zurück zum Zitat Kurtz SM. Compendium of HXLPEs. In: Kurtz SM, ed. UHMWPE Biomaterials Handbook. Waltham, MA, USA: Elsevier; 2016. Kurtz SM. Compendium of HXLPEs. In: Kurtz SM, ed. UHMWPE Biomaterials Handbook. Waltham, MA, USA: Elsevier; 2016.
12.
Zurück zum Zitat Kurtz SM. UHMWPE Biomaterials Handbook: Ultra High Molecular Weight Polyethylene in Total Joint Replacement and Medical Devices. Waltham, MA, USA: Elsevier; 2016. Kurtz SM. UHMWPE Biomaterials Handbook: Ultra High Molecular Weight Polyethylene in Total Joint Replacement and Medical Devices. Waltham, MA, USA: Elsevier; 2016.
13.
Zurück zum Zitat Langohr G, Gawel H, Medley J. Wear performance of all-polymer PEEK articulations for a cervical total level arthroplasty system. Proc Inst Mech Eng [J]. 2011;225:499–513.CrossRef Langohr G, Gawel H, Medley J. Wear performance of all-polymer PEEK articulations for a cervical total level arthroplasty system. Proc Inst Mech Eng [J]. 2011;225:499–513.CrossRef
14.
Zurück zum Zitat Laux K, Schwartz C. Influence of linear reciprocating and multi-directional sliding on PEEK wear performance and transfer film formation. Wear. 2013;301:727–734.CrossRef Laux K, Schwartz C. Influence of linear reciprocating and multi-directional sliding on PEEK wear performance and transfer film formation. Wear. 2013;301:727–734.CrossRef
15.
Zurück zum Zitat Lu Z, McKellop H. Frictional heating of bearing materials tested in a hip joint wear simulator. Proc Inst Mech Eng [H]. 1997;211:101–108.CrossRef Lu Z, McKellop H. Frictional heating of bearing materials tested in a hip joint wear simulator. Proc Inst Mech Eng [H]. 1997;211:101–108.CrossRef
16.
Zurück zum Zitat Mazzucco D, Spector M. Effects of contact area and stress on the volumetric wear of ultrahigh molecular weight polyethylene. Wear. 2003;254:514–522.CrossRef Mazzucco D, Spector M. Effects of contact area and stress on the volumetric wear of ultrahigh molecular weight polyethylene. Wear. 2003;254:514–522.CrossRef
17.
Zurück zum Zitat Muratoglu OK, Bragdon CR, O’Connor DO, Jasty M, Harris WH, Gul R, McGarry F. Unified wear model for highly crosslinked ultra-high molecular weight polyethylenes (UHMWPE). Biomaterials. 1999;20:1463–1470.CrossRefPubMed Muratoglu OK, Bragdon CR, O’Connor DO, Jasty M, Harris WH, Gul R, McGarry F. Unified wear model for highly crosslinked ultra-high molecular weight polyethylenes (UHMWPE). Biomaterials. 1999;20:1463–1470.CrossRefPubMed
18.
Zurück zum Zitat Pritchett J. Heat generated by hip resurfacing prostheses: an in vivo pilot study. J Long Term Eff Med Implants. 2011;21:55.CrossRefPubMed Pritchett J. Heat generated by hip resurfacing prostheses: an in vivo pilot study. J Long Term Eff Med Implants. 2011;21:55.CrossRefPubMed
19.
Zurück zum Zitat Saikko V. A multidirectional motion pin‐on‐disk wear test method for prosthetic joint materials. J Biomed Mater Res. 1998;41:58–64.CrossRefPubMed Saikko V. A multidirectional motion pin‐on‐disk wear test method for prosthetic joint materials. J Biomed Mater Res. 1998;41:58–64.CrossRefPubMed
20.
Zurück zum Zitat Scholes S, Unsworth A. The wear properties of CFR-PEEK-OPTIMA articulating against ceramic assessed on a multidirectional pin-on-plate machine. Proc Inst Mech Eng [H]. 2007;221:281–289.CrossRef Scholes S, Unsworth A. The wear properties of CFR-PEEK-OPTIMA articulating against ceramic assessed on a multidirectional pin-on-plate machine. Proc Inst Mech Eng [H]. 2007;221:281–289.CrossRef
21.
Zurück zum Zitat Scholes S, Unsworth A. Wear studies on the likely performance of CFR-PEEK/CoCrMo for use as artificial joint bearing materials. J Mater Sci Mater Med. 2009;20:163–170.CrossRefPubMed Scholes S, Unsworth A. Wear studies on the likely performance of CFR-PEEK/CoCrMo for use as artificial joint bearing materials. J Mater Sci Mater Med. 2009;20:163–170.CrossRefPubMed
22.
Zurück zum Zitat Scholes S, Unsworth A. The wear performance of PEEK-OPTIMA based self-mating couples. Wear. 2010;268:380–387.CrossRef Scholes S, Unsworth A. The wear performance of PEEK-OPTIMA based self-mating couples. Wear. 2010;268:380–387.CrossRef
23.
Zurück zum Zitat Wang A, Essner A, Polineni V, Stark C, Dumbleton J. Lubrication and wear of ultra-high molecular weight polyethylene in total joint replacements. Tribol Int. 1998;31:17–33.CrossRef Wang A, Essner A, Polineni V, Stark C, Dumbleton J. Lubrication and wear of ultra-high molecular weight polyethylene in total joint replacements. Tribol Int. 1998;31:17–33.CrossRef
24.
Zurück zum Zitat Wang A, Polineni V, Essner A, Sokol M, Sun D, Stark C, Dumbleton J. The significance of nonlinear motion in the wear screening of orthopaedic implant materials. J Test Eval. 1997;25:239–245.CrossRef Wang A, Polineni V, Essner A, Sokol M, Sun D, Stark C, Dumbleton J. The significance of nonlinear motion in the wear screening of orthopaedic implant materials. J Test Eval. 1997;25:239–245.CrossRef
Metadaten
Titel
The Biotribology of PEEK-on-HXLPE Bearings Is Comparable to Traditional Bearings on a Multidirectional Pin-on-disk Tester
verfasst von
Doruk Baykal, PhD
Ryan S. Siskey, MSc
Richard J. Underwood, PhD
Adam Briscoe, PhD
Steven M. Kurtz, PhD
Publikationsdatum
25.07.2016
Verlag
Springer US
Erschienen in
Clinical Orthopaedics and Related Research® / Ausgabe 11/2016
Print ISSN: 0009-921X
Elektronische ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-016-4989-7

Weitere Artikel der Ausgabe 11/2016

Clinical Orthopaedics and Related Research® 11/2016 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.