The changes of corneal biomechanical properties with long-term treatment of prostaglandin analogue measured by Corvis ST

verfasst von: Na Wu, Yuhong Chen, Yaping Yang, Xinghuai Sun

Erschienen in: BMC Ophthalmology | Ausgabe 1/2020

Abstract

Background

To investigate the corneal biomechanical changes in primary open angle glaucoma (POAG) patients treated with long-term prostaglandin analogue (PGA).

Methods

One hundred eleven newly diagnosed POAG patients, including 43 high tension glaucoma (HTG) and 68 normal tension glaucoma (NTG), were measured by Corvis ST to obtain intraocular pressure (IOP), central corneal thickness (CCT) and corneal biomechanical parameters at baseline and at each follow-up visit after initiation of PGA treatment. The follow-up measurements were analyzed by the generalized estimate equation model with an exchangeable correlation structure. Restricted cubic spline was employed to estimate the dose–response relation between follow-up time and corneal biomechanics.

Results

The mean follow-up time was 10.3 ± 7.02 months. Deformation amplitude (β = -0.0015, P = 0.016), the first applanation velocity (AV1, β = -0.0004, P = 0.00058) decreased and the first applanation time (AT1, β = 0.0089, P < 0.000001) increased statistically significantly with PGA therapy over time after adjusting for age, gender, axial length, corneal curvature, IOP and CCT. In addition, AT1 was lower (7.2950 ± 0.2707 in NTG and 7.5889 ± 0.2873 in HTG, P = 0.00011) and AV1 was greater (0.1478 ± 0.0187 in NTG and 0.1314 ± 0.0191 in HTG, P = 0.00002) in NTG than in HTG after adjusting for confounding factors.

Conclusions

Chronic use of PGA probably influences the corneal biomechanical properties directly, which is to make cornea less deformable. Besides, corneas in NTG tended to be more deformable compared to those in HTG with long-term treatment of PGA.
Hinweise

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
AT1
The first applanation time
AT2
The second applanation time
AV1
The first applanation velocity
AV2
The second applanation velocity
bIOP
Biomechanically corrected IOP
CCT
Central corneal thickness
CH
Corneal hysteresis
Corvis ST
Corneal Visualization Scheimpflug Technology
DA
Deformation amplitude
HC
Highest concavity
HC
Highest concavity
HTG
High tension glaucoma
IOP
Intraocular pressure
MMPs
Matrix metalloproteinases
NTG
Normal tension glaucoma
OCT
Optical coherence tomography
ORA
Ocular Response Analyzer
PD
Peak distance
PGA
Prostaglandin analogue
POAG
Primary open angle glaucoma
SD
Standard deviation

Background

Corneal biomechanical properties have recently attracted increasing attention for its involvement in glaucoma and in other ocular diseases, such as keratoconus [1]. The biomechanical properties of cornea may determine its deformation response to applanation, which influences the intraocular pressure (IOP) measurement [2, 3]. In addition, it has been speculated that corneal biomechanics may represent structural vulnerabilities of the whole eyeball to the development of glaucoma [46]. Thus, understanding corneal biomechanics is helpful to improve the diagnosis and management of glaucoma [2, 3, 7].
Primary open angle glaucoma (POAG) is the most common type of glaucoma worldwide. Prostaglandin analogue (PGA) is the first-line hypotensive medication for POAG currently. Long-term use of PGA has been reported to decrease the central corneal thickness (CCT) due to activation of corneal stromal matrix metalloproteinases (MMPs) [8, 9]. Experimental studies also indicate that PGA causes morphological and biochemical changes of corneal stroma [10]. These changes imply that corneal biomechanical properties could possibly be affected by PGA.
Most of the available studies about corneal biomechanics were conducted by using the Ocular Response Analyzer (ORA) [11]. ORA is the first developed device, which is able to provide an in vivo measurement of corneal viscoelastic parameter called corneal hysteresis (CH) [12]. CH is shown to be decreased in glaucomatous eyes compared to that in normal controls [13] and associated with visual field progression in glaucomatous eyes [7].
The Corneal Visualization Scheimpflug Technology (Corvis ST) is a newly available non-touch instrument that provides precise, repeatable, and reproducible measurements of corneal biomechanics in vivo [11, 14]. The Corvis ST uses an ultra-high-speed Scheimpflug camera to record corneal deformation in real time and provides direct information on corneal biomechanical behavior [15] and could potentially provide more information on corneal biomechanics than ORA [16].
Our previous study conducted by Corvis ST observed significant differences in biomechanical parameters between treatment naïve POAG patients and patients with long-term therapy of PGA, even after adjusting for other confounding factors that might influence biomechanics, such as IOP and CCT [17]. The results suggested that PGA might have a potential effect on corneal biomechanical properties in addition to its indirect effect owning to IOP decrease and CCT reduction [17]. However, due to the cross-sectional design, the influence of PGA on corneal biomechanics remains unanswered. In this study, we prospectively followed up newly diagnosed POAG patients under PGA treatment to observe the changes of corneal biomechanical properties over time.

Methods

This research was approved by the Medical Ethics Committee of Eye and ENT Hospital of Fudan University, and adhered to the Declaration of Helsinki. Written informed consents were obtained from all of the participants.
Patients who were newly diagnosed with POAG were recruited consecutively at Eye and ENT Hospital of Fudan University. POAG was diagnosed with the criteria as follows: an open anterior chamber angle on gonioscopy examination, characteristic glaucomatous optic disc changes (thinning of the optic disc rim and/or enlargement of the optic disc cupping) with corresponding visual field defects: (1) glaucoma hemifield test values outside the normal limits; or (2) three or more abnormal points with a probability of being normal of P < 5%, of which at least one point has a pattern deviation of P < 1%; or (3) a pattern standard deviation of P < 5%. In addition, patients were diagnosed with high tension glaucoma (HTG) when they had at least one measurement of IOP > 21 mmHg with a Goldmann applanation tonometer. Patients diagnosed as normal tension glaucoma (NTG) had an IOP ≤ 21 mmHg at all time points on a 24-h IOP variation test. Patients who received continuous therapy of only one type of PGA medication were enrolled in the study. Exclusion criteria included any coexisted ocular conditions such as corneal diseases, previous history of ocular laser or surgical treatment, which might affect corneal biomechanics. Secondary glaucoma, such as trauma, uveitis and steroid induced glaucoma, was excluded as well.
All participants underwent a comprehensive ophthalmologic examination when they were first enrolled in the study, including best corrected visual acuity, slit-lamp biomicroscopy, fundus evaluation with a 90D lens, gonioscopy examination, automated perimetry (Humphrey Field Analyzer, Carl Zeiss, USA) and the retinal nerve fiber layer thickness on optical coherence tomography (OCT) (Optovue RTVue OCT, USA). Axial length and corneal curvature were measured by IOL Master (Zeiss, Germany).
The Corvis ST (Oculus, GmbH, Wetzlar, Germany) not only measures IOP, biomechanically corrected IOP (bIOP), CCT, but also provides parameters of corneal biomechanical properties by exerting an air impulse on the cornea. Under the pressure, the cornea bends inward from the resting state to the first applanation point and continues to move until reaches the maximum deformation state, namely highest concavity (HC). When the pressure of air puff decreases, the cornea moves outward and passes the second applanation point before arriving at its resting state [11, 14]. The biomechanical parameters used in this study were: 1) time from the resting state to the first and second applanations (AT1 and AT2, respectively); 2) corneal velocity during the first and second applantions (AV1 and AV2, respectively); 3) highest concavity time (HC-time) and maximum deformation amplitude (DA) from the resting state to the HC at the corneal apex; peak distance between corneal bending points (HC-PD) and radius of curvature (HC-radius) at HC [11, 14, 18].
Patients were examined by Corvis ST at baseline and at each follow-up visit. Both eyes were included if they were eligible. Single eyes were selected on 7 patients because the other ones could not obtain reliable measurements. It should be noted that only Corvis ST measurements of post-treatment of PGA were used for statistical analysis.
Statistical analyses were performed with R 3.3 (https://​www.​R-project.​org).
Descriptive statistical results were presented as the mean ± standard deviation (SD). The repeated follow-up measurements were analyzed by the generalized estimate equation model with an exchangeable correlation structure. Restricted cubic spline was employed to estimate the dose–response relation between follow-up time and IOP, CCT and corneal biomechanical parameters with adjustment. Independent samples t test and chi-square test were used for comparison of demographic (age and gender) and clinical (axial length and corneal curvature) measurements between NTG and HTG groups. It was considered statistically significant if the P value was less than 0.05.

Results

The examined group consisted of 111 newly diagnosed POAG patients, including 43 female and 68 male with a mean age of 48.04 ± 14.14 years (22—82 years). Together, 215 eyes were examined (both eyes from 104 subjects and 7 eyes from 7 subjects). The mean follow-up time was 10.3 ± 7.02 months (1.15—41.57 months). Subjects had an average of 2.53 ± 1.42 (1–7) times of follow-up visits in the study. The mean axial length and corneal curvature of the total subjects were 25.6 ± 1.77 mm and 7.75 ± 0.24 mm, respectively (Table 1). A slight but statistically significant decrease in IOP (β = -0.0596, P = 0.000047), bIOP (β = -0.0369, P = 0.0029) and CCT (β = -0.6639, P < 0.000001) was observed during the period of PGA usage (Table 2, Fig. 1).
Table 1
Demographic, clinical and diagnostic characteristics of 111 newly diagnosed primary open angle glaucoma patients participating in the study
Category
Age, years
48.04 ± 14.14 (22–82)
Gender
 Female
43 (38.7%)
 Male
68 (61.3%)
Follow-up period, months
10.3 ± 7.02 (1.15—41.57)
Follow-up times
2.53 ± 1.42 (1–7)
Diagnosis
 High tension glaucoma (HTG)
43 (38.7%)
 Normal tension glaucoma (NTG)
68 (61.3%)
Axial length, mm
25.6 ± 1.77
Corneal curvature, mm
7.75 ± 0.24
Table 2
The changes of IOP, bIOP and CCT in POAG patients under PGA treatment over time
 
β
95% CI
P value
IOP
-0.0596
-0.0883 to -0.0309
0.000047
bIOP
-0.0369
-0.0612 to -0.0126
0.0029
CCT
-0.6639
-0.8048 to -0.5230
 < 0.000001
IOP intraocular pressure, bIOP biomechanically corrected intraocular pressure, CCT central corneal thickness, POAG primary open angle glaucoma, PGA prostaglandin analogue
Generalized estimate equation model with an exchangeable correlation structure was used to observe the corneal biomechanical changes with PGA treatment over time. After adjusting for age, gender, axial length, corneal curvature, IOP and CCT, AT1 increased (β = 0.0089, P < 0.000001), whereas DA and AV1 decreased statistically significantly over time (β = -0.0015, P = 0.016 for DA; β = -0.0004, P = 0.00058 for AV1) (Table 3, Fig. 2). Other corneal biomechanical parameters showed no statistically significant changes over time (Table 3).
Table 3
The changes of corneal biomechanical parameters in POAG patients under PGA treatment over time with adjustment for age, gender, axial length, corneal curvature, IOP and CCT
Corneal biomechanical parameters
After adjustment
β
P value
DA
-0.0015
0.016
AT1
0.0089
 < 0.000001
AV1
-0.0004
0.00058
AT2
-0.0059
0.139
AV2
0.0003
0.364
HC-time
-0.0044
0.272
HC-PD
-0.0023
0.196
HC-radius
-0.0076
0.346
POAG primary open angle glaucoma, PGA prostaglandin analogue, IOP intraocular pressure, CCT central corneal thickness, DA deformation amplitude, AT1 the first applanation time, AV1 the first applanation velocity, AT2 the second applanation time, AV2 the second applanation velocity, HC the highest concavity, PD peak distance
In addition, we analyzed the differences of corneal biomechanical parameters between NTG and HTG groups under PGA treatment over time. Among the 111 POAG patients, 68 were NTG patients and the other 43 were HTG (Table 1). There were no statistically significant differences in age, gender, axial length and corneal curvature between these two groups (all P > 0.05, Table 4). IOP, bIOP and CCT of post-treatment of PGA in NTG group were all statistically significantly lower than those in HTG group (all P < 0.0001, Table 4). After correcting for age, gender, axial length, corneal curvature, IOP and CCT, AT1 was lower (7.2950 ± 0.2707 in NTG and 7.5889 ± 0.2873 in HTG, P = 0.00011, Fig. 3a) and AV1 was higher (0.1478 ± 0.0187 in NTG and 0.1314 ± 0.0191 in HTG, P = 0.00002, Fig. 3b) in NTG group compared to that in HTG group with statistically significant differences (Table 5).
Table 4
Differences in demographic and clinical characteristics between NTG and HTG groups
 
NTG (n = 68)
HTG (n = 43)
P value
Age (male)
49.15 ± 14.51
46.28 ± 13.50
0.292
Gender
41 (60.3%)
27 (62.8%)
0.843
Axial length
25.52 ± 1.81
25.74 ± 1.70
0.374
Corneal curvature
7.76 ± 0.22
7.75 ± 0.27
0.826
IOP
16.05 ± 2.36
18.47 ± 2.36
 < 0.0001
bIOP
15.37 ± 2.09
16.90 ± 2.10
 < 0.0001
CCT
537.44 ± 32.13
564.35 ± 27.96
 < 0.0001
NTG normal tension glaucoma, HTG high tension glaucoma, IOP intraocular pressure, bIOP biomechanically corrected intraocular pressure, CCT central corneal thickness
Table 5
Differences in corneal biomechanical parameters between NTG and HTG groups under PGA treatment over time after adjusting for age, gender, axial length, corneal curvature, IOP and CCT
 
NTG (n = 68)
HTG (n = 43)
P value
DA
1.1070 ± 0.1115
1.0425 ± 0.1248
0.681
AT1
7.2950 ± 0.2707
7.5889 ± 0.2873
0.00011
AV1
0.1478 ± 0.0187
0.1314 ± 0.0191
0.00002
AT2
21.8646 ± 0.6164
21.6138 ± 0.5827
0.173
AV2
-0.2682 ± 0.0580
-0.2551 ± 0.0426
0.709
HC-time
16.8531 ± 0.5389
16.8520 ± 0.8962
0.477
HC-PD
5.0539 ± 0.2704
4.8712 ± 0.3023
0.241
HC-radius
6.5643 ± 0.9992
6.8861 ± 2.5276
0.428
NTG normal tension glaucoma, HTG high tension glaucoma, PGA prostaglandin analogue, IOP intraocular pressure, CCT central corneal thickness, DA deformation amplitude, AT1 the first applanation time, AV1 the first applanation velocity, AT2 the second applanation time, AV2 the second applanation velocity, HC the highest concavity, PD peak distance

Discussion

PGA is the most widely used first-line therapeutic medication for POAG due to its highest effectiveness in decreasing IOP [19] and reducing circadian IOP fluctuations, as well as its convenient dosing and tolerable side effects [20]. However, PGA was found to have some other effects, such as decreasing the thickness of central cornea. Our results showed CCT declined under treatment of PGA over time, which was in accordance with previous studies [21, 22], implying a possible time cumulative effect of PGA on cornea tissue. The mechanisms of CCT reduction are possibly due to the MMPs up-regulation and extracellular matrix degradation stimulated by PGA in the cornea [8, 9, 23].
The cornea is considered to have both viscous and elastic properties [1]. Studies conducted on the ORA showed that IOP decreased and CH increased significantly after topical PGA treatment [24, 25]. Increased CH indicates an increased corneal viscoelastic response [26]. IOP is a significant influencing factor of corneal deformation [1, 27] and should always be taken into account when analyzing corneal biomechanics. Hence, to better elucidate the corneal biomechanical changes with long-term treatment of PGA and to avoid the influences of IOP sudden decrease on it, baseline Corvis ST measurements, namely those before initiation of PGA treatment, were not analyzed.
In this study, we showed that DA, AT1 and AV1 changed with statistical significances under PGA therapy over time. Since the results were adjusted for age, gender, axial length, corneal curvature, IOP and CCT, it suggested long-term use of PGA might influence the corneal biomechanics directly. Our observation was consistent with an previous observational study, which found CH increased significantly after 6 months treatment of PGA on newly diagnosed POAG patients measured by ORA, and this increase was only significantly correlated with basal IOP but not with the drug induced IOP reduction, suggesting PGA could have a direct effect on cornea biomechanical properties [25].
A more deformable cornea is characterized by reaching the first applanation faster (with shorter AT1 and greater AV1), the second applanation slower (with longer AT2 and lower AV2) and the highest concavity with greater DA [14]. In our study, although not all of the biomechanical parameters reached statistical significances, DA (β = -0.0015), AV1 (β = -0.0004) and AT2 (β = -0.0059) decreased and AT1 (β = 0.0089) increased as PGA treatment time extended after adjusting for potential confounding factors. Based on the meanings of the parameters, our results indicated a less deformable cornea as PGA treatment prolonged.
Our observations were supported by a few previous studies as follows. Hussnain et al. observed a significant decrease in CH among medication, laser or surgery treated POAG patients in a retrospective study [28]. They hypothesized that CH decreased over time in POAG patients even though increased in a short term with therapies to reduce IOP [28]. However, the percentages and types of medical treatment were not mentioned in their publication. Another recent prospective study conducted by Meda et al. showed that POAG patients with long-term PGA therapy for at least 1 year had significantly lower CH values than those whose PGA were discontinued [12], which indicated that long-term use of PGA might cause CH decrease.
Pathogenesis differences between NTG and HTG have been investigated by many studies. Factors other than IOP probably contribute more to the onset and development of NTG. Interestingly, we found there were statistically significant differences in corneal biomechanical parameters under PGA treatment between NTG and HTG groups. After adjusting for age, gender, axial length, corneal curvature, IOP and CCT, AT1 was shorter and AV1 was greater in NTG than in HTG, which means that corneas in NTG tend to be more deformable compared to those in HTG with the extension of PGA treatment.
To our knowledge, this is the first prospective study investigating the long-term changes of corneal biomechanical properties with chronic use of PGA measured by Corvis ST. However, some of the limitations need to be considered. The patient profile may not necessarily represent the whole population data because of our clinic-based design. Furthermore, the patients had variant follow-up times and different intervals between visits, which could potentially introduce bias into current results.

Conclusions

Our study indicated that long-term treatment of PGA might make cornea less deformable. It should be taken into consideration when analyzing corneal biomechanics in patients with PGA treatment. We speculate the corneal extracellular matrix remodeling may be involved, although the underlying mechanisms remain unknown. It would be interesting for further studies to validate these changes at a molecular level.

Acknowledgements

We are grateful to the glaucoma patients who participated in the study.
Not applicable.

Competing interests

The authors declare that they have no competing interest.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Ali NQ, Patel DV, McGhee CN. Biomechanical responses of healthy and keratoconic corneas measured using a noncontact scheimpflug-based tonometer. Invest Ophthalmol Vis Sci. 2014;55:3651–9.CrossRef Ali NQ, Patel DV, McGhee CN. Biomechanical responses of healthy and keratoconic corneas measured using a noncontact scheimpflug-based tonometer. Invest Ophthalmol Vis Sci. 2014;55:3651–9.CrossRef
2.
Zurück zum Zitat Liu J, Roberts CJ. Influence of corneal biomechanical properties on intraocular pressure measurement: quantitative analysis. J Cataract Refract Surg. 2005;31:146–55.CrossRef Liu J, Roberts CJ. Influence of corneal biomechanical properties on intraocular pressure measurement: quantitative analysis. J Cataract Refract Surg. 2005;31:146–55.CrossRef
3.
Zurück zum Zitat Medeiros FA, Weinreb RN. Evaluation of the influence of corneal biomechanical properties on intraocular pressure measurements using the ocular response analyzer. J Glaucoma. 2006;15:364–70. .CrossRef Medeiros FA, Weinreb RN. Evaluation of the influence of corneal biomechanical properties on intraocular pressure measurements using the ocular response analyzer. J Glaucoma. 2006;15:364–70. .CrossRef
4.
Zurück zum Zitat Sigal IA, Flanagan JG, Ethier CR. Factors influencing optic nerve head biomechanics. Invest Ophthalmol Vis Sci. 2005;46:4189–99.CrossRef Sigal IA, Flanagan JG, Ethier CR. Factors influencing optic nerve head biomechanics. Invest Ophthalmol Vis Sci. 2005;46:4189–99.CrossRef
5.
Zurück zum Zitat Johnson CS, Mian SI, Moroi S, Epstein D, Izatt J, Afshari NA. Role of corneal elasticity in damping of intraocular pressure. Invest Ophthalmol Vis Sci. 2007;48:2540–4.CrossRef Johnson CS, Mian SI, Moroi S, Epstein D, Izatt J, Afshari NA. Role of corneal elasticity in damping of intraocular pressure. Invest Ophthalmol Vis Sci. 2007;48:2540–4.CrossRef
6.
Zurück zum Zitat Wells AP, Garway-Heath DF, Poostchi A, Wong T, Chan KC, Sachdev N. Corneal hysteresis but not corneal thickness correlates with optic nerve surface compliance in glaucoma patients. Invest Ophthalmol Vis Sci. 2008;49:3262–8.CrossRef Wells AP, Garway-Heath DF, Poostchi A, Wong T, Chan KC, Sachdev N. Corneal hysteresis but not corneal thickness correlates with optic nerve surface compliance in glaucoma patients. Invest Ophthalmol Vis Sci. 2008;49:3262–8.CrossRef
7.
Zurück zum Zitat Congdon NG, Broman AT, Bandeen-Roche K, Grover D, Quigley HA. Central corneal thickness and corneal hysteresis associated with glaucoma damage. Am J Ophthalmol. 2006;141:868–75.CrossRef Congdon NG, Broman AT, Bandeen-Roche K, Grover D, Quigley HA. Central corneal thickness and corneal hysteresis associated with glaucoma damage. Am J Ophthalmol. 2006;141:868–75.CrossRef
8.
Zurück zum Zitat Sen E, Nalcacioglu P, Yazici A, Aksakal FN, Altinok A, Tuna T, et al. Comparison of the effects of latanoprost and bimatoprost on central corneal thickness. J Glaucoma. 2008;17:398–402.CrossRef Sen E, Nalcacioglu P, Yazici A, Aksakal FN, Altinok A, Tuna T, et al. Comparison of the effects of latanoprost and bimatoprost on central corneal thickness. J Glaucoma. 2008;17:398–402.CrossRef
9.
Zurück zum Zitat Kocabeyoglu S, Mocan MC, Irkec M. Decreased keratocyte density and central corneal thickness in primary open-angle glaucoma patients undergoing treatment with topical prostaglandin analogues. Indian J Ophthalmol. 2015;63:15–9.CrossRef Kocabeyoglu S, Mocan MC, Irkec M. Decreased keratocyte density and central corneal thickness in primary open-angle glaucoma patients undergoing treatment with topical prostaglandin analogues. Indian J Ophthalmol. 2015;63:15–9.CrossRef
10.
Zurück zum Zitat Wu KY, Wang HZ, Hong SJ. Effect of latanoprost on cultured porcine corneal stromal cells. Curr Eye Res. 2005;30:871–9.CrossRef Wu KY, Wang HZ, Hong SJ. Effect of latanoprost on cultured porcine corneal stromal cells. Curr Eye Res. 2005;30:871–9.CrossRef
11.
Zurück zum Zitat Chansangpetch S, Panpruk R, Manassakorn A, Tantisevi V, Rojanapongpun P, Hurst CP, et al. Impact of Myopia on Corneal Biomechanics in Glaucoma and Nonglaucoma Patients. Invest Ophthalmol Vis Sci. 2017;58:4990–6.CrossRef Chansangpetch S, Panpruk R, Manassakorn A, Tantisevi V, Rojanapongpun P, Hurst CP, et al. Impact of Myopia on Corneal Biomechanics in Glaucoma and Nonglaucoma Patients. Invest Ophthalmol Vis Sci. 2017;58:4990–6.CrossRef
12.
Zurück zum Zitat Meda R, Wang Q, Paoloni D, Harasymowycz P, Brunette I. The impact of chronic use of prostaglandin analogues on the biomechanical properties of the cornea in patients with primary open-angle glaucoma. Br J Ophthalmol. 2017;101:120–5.CrossRef Meda R, Wang Q, Paoloni D, Harasymowycz P, Brunette I. The impact of chronic use of prostaglandin analogues on the biomechanical properties of the cornea in patients with primary open-angle glaucoma. Br J Ophthalmol. 2017;101:120–5.CrossRef
13.
Zurück zum Zitat Shin J, Lee JW, Kim EA, Caprioli J. The effect of corneal biomechanical properties on rebound tonometer in patients with normal-tension glaucoma. Am J Ophthalmol. 2015;159:144–54.CrossRef Shin J, Lee JW, Kim EA, Caprioli J. The effect of corneal biomechanical properties on rebound tonometer in patients with normal-tension glaucoma. Am J Ophthalmol. 2015;159:144–54.CrossRef
14.
Zurück zum Zitat Salvetat ML, Zeppieri M, Tosoni C, Felletti M, Grasso L, Brusini P. Corneal Deformation Parameters Provided by the Corvis-ST Pachy-Tonometer in Healthy Subjects and Glaucoma Patients. J Glaucoma. 2015;24:568–74.CrossRef Salvetat ML, Zeppieri M, Tosoni C, Felletti M, Grasso L, Brusini P. Corneal Deformation Parameters Provided by the Corvis-ST Pachy-Tonometer in Healthy Subjects and Glaucoma Patients. J Glaucoma. 2015;24:568–74.CrossRef
15.
Zurück zum Zitat Hon Y, Lam AKC. Corneal Deformation Measurement Using Scheimpflug Noncontact Tonometry. Optom Vis Sci. 2013;90:E1–8.CrossRef Hon Y, Lam AKC. Corneal Deformation Measurement Using Scheimpflug Noncontact Tonometry. Optom Vis Sci. 2013;90:E1–8.CrossRef
16.
Zurück zum Zitat Wang W, Du S, Zhang X. Corneal Deformation Response in Patients With Primary Open-Angle Glaucoma and in Healthy Subjects Analyzed by Corvis ST. Invest Ophthalmol Vis Sci. 2015;56:5557–65.CrossRef Wang W, Du S, Zhang X. Corneal Deformation Response in Patients With Primary Open-Angle Glaucoma and in Healthy Subjects Analyzed by Corvis ST. Invest Ophthalmol Vis Sci. 2015;56:5557–65.CrossRef
17.
Zurück zum Zitat Wu N, Chen Y, Yu X, Li M, Wen W, Sun X. Changes in Corneal Biomechanical Properties after Long-Term Topical Prostaglandin Therapy. Plos One. 2016;11:e0155527.CrossRef Wu N, Chen Y, Yu X, Li M, Wen W, Sun X. Changes in Corneal Biomechanical Properties after Long-Term Topical Prostaglandin Therapy. Plos One. 2016;11:e0155527.CrossRef
18.
Zurück zum Zitat Huseynova T, Waring GO, Roberts C, Krueger RR, Tomita M. Corneal Biomechanics as a Function of Intraocular Pressure and Pachymetry by Dynamic Infrared Signal and Scheimpflug Imaging Analysis in Normal Eyes. Am J Ophthalmol. 2014;157:885–93.CrossRef Huseynova T, Waring GO, Roberts C, Krueger RR, Tomita M. Corneal Biomechanics as a Function of Intraocular Pressure and Pachymetry by Dynamic Infrared Signal and Scheimpflug Imaging Analysis in Normal Eyes. Am J Ophthalmol. 2014;157:885–93.CrossRef
19.
Zurück zum Zitat Li F, Huang W, Zhang X. Efficacy and safety of different regimens for primary open-angle glaucoma or ocular hypertension: a systematic review and network meta-analysis. Acta Ophthalmol. 2018;96:e277–84.CrossRef Li F, Huang W, Zhang X. Efficacy and safety of different regimens for primary open-angle glaucoma or ocular hypertension: a systematic review and network meta-analysis. Acta Ophthalmol. 2018;96:e277–84.CrossRef
20.
Zurück zum Zitat Konstas AG, Quaranta L, Katsanos A, Riva I, Tsai JC, Giannopoulos T, et al. Twenty-four hour efficacy with preservative free tafluprost compared with latanoprost in patients with primary open angle glaucoma or ocular hypertension. Br J Ophthalmol. 2013;97:1510–5.CrossRef Konstas AG, Quaranta L, Katsanos A, Riva I, Tsai JC, Giannopoulos T, et al. Twenty-four hour efficacy with preservative free tafluprost compared with latanoprost in patients with primary open angle glaucoma or ocular hypertension. Br J Ophthalmol. 2013;97:1510–5.CrossRef
21.
Zurück zum Zitat Sawada A, Yamamoto T, Takatsuka N. Randomized crossover study of latanoprost and travoprost in eyes with open-angle glaucoma. Graefes Arch Clin Exp Ophthalmol. 2012;250:123–9.CrossRef Sawada A, Yamamoto T, Takatsuka N. Randomized crossover study of latanoprost and travoprost in eyes with open-angle glaucoma. Graefes Arch Clin Exp Ophthalmol. 2012;250:123–9.CrossRef
22.
Zurück zum Zitat Lee H, Cho BJ. Long-term effect of latanoprost on central corneal thickness in normal-tension glaucoma: five-year follow-up results. J Ocul Pharmacol Ther. 2015;31:152–5.CrossRef Lee H, Cho BJ. Long-term effect of latanoprost on central corneal thickness in normal-tension glaucoma: five-year follow-up results. J Ocul Pharmacol Ther. 2015;31:152–5.CrossRef
23.
Zurück zum Zitat Lopilly Park HY, Kim JH, Lee KM, Park CK. Effect of prostaglandin analogues on tear proteomics and expression of cytokines and matrix metalloproteinases in the conjunctiva and cornea. Exp Eye Res. 2012;94:13–21.CrossRef Lopilly Park HY, Kim JH, Lee KM, Park CK. Effect of prostaglandin analogues on tear proteomics and expression of cytokines and matrix metalloproteinases in the conjunctiva and cornea. Exp Eye Res. 2012;94:13–21.CrossRef
24.
Zurück zum Zitat Tsikripis P, Papaconstantinou D, Koutsandrea C, Apostolopoulos M, Georgalas I. The effect of prostaglandin analogs on the biomechanical properties and central thickness of the cornea of patients with open-angle glaucoma: a 3-year study on 108 eyes. Drug Des Devel Ther. 2013;7:1149–56.PubMedPubMedCentral Tsikripis P, Papaconstantinou D, Koutsandrea C, Apostolopoulos M, Georgalas I. The effect of prostaglandin analogs on the biomechanical properties and central thickness of the cornea of patients with open-angle glaucoma: a 3-year study on 108 eyes. Drug Des Devel Ther. 2013;7:1149–56.PubMedPubMedCentral
25.
Zurück zum Zitat Bolivar G, Sanchez-Barahona C, Teus M, Castejon MA, Paz-Moreno-Arrones J, Gutierrez-Ortiz C, et al. Effect of topical prostaglandin analogues on corneal hysteresis. Acta Ophthalmol. 2015;93:e495-498. .CrossRef Bolivar G, Sanchez-Barahona C, Teus M, Castejon MA, Paz-Moreno-Arrones J, Gutierrez-Ortiz C, et al. Effect of topical prostaglandin analogues on corneal hysteresis. Acta Ophthalmol. 2015;93:e495-498. .CrossRef
26.
Zurück zum Zitat Jung Y, Park HL, Yang HJ, Park CK. Characteristics of corneal biomechanical responses detected by a non-contact scheimpflug-based tonometer in eyes with glaucoma. Acta Ophthalmol. 2017;95:e556–63. .CrossRef Jung Y, Park HL, Yang HJ, Park CK. Characteristics of corneal biomechanical responses detected by a non-contact scheimpflug-based tonometer in eyes with glaucoma. Acta Ophthalmol. 2017;95:e556–63. .CrossRef
27.
Zurück zum Zitat Tian L, Wang D, Wu Y, Meng X, Chen B, Ge M, et al. Corneal biomechanical characteristics measured by the CorVis Scheimpflug technology in eyes with primary open-angle glaucoma and normal eyes. Acta Ophthalmol. 2016;94:e317-324.CrossRef Tian L, Wang D, Wu Y, Meng X, Chen B, Ge M, et al. Corneal biomechanical characteristics measured by the CorVis Scheimpflug technology in eyes with primary open-angle glaucoma and normal eyes. Acta Ophthalmol. 2016;94:e317-324.CrossRef
28.
Zurück zum Zitat Hussnain SA, Alsberge JB, Ehrlich JR, Shimmyo M, Radcliffe NM. Change in corneal hysteresis over time in normal, glaucomatous and diabetic eyes. Acta Ophthalmol. 2015;93:e627-630.CrossRef Hussnain SA, Alsberge JB, Ehrlich JR, Shimmyo M, Radcliffe NM. Change in corneal hysteresis over time in normal, glaucomatous and diabetic eyes. Acta Ophthalmol. 2015;93:e627-630.CrossRef
Metadaten
Titel
The changes of corneal biomechanical properties with long-term treatment of prostaglandin analogue measured by Corvis ST
verfasst von
Na Wu
Yuhong Chen
Yaping Yang
Xinghuai Sun
Publikationsdatum
01.12.2020
Verlag
BioMed Central
Erschienen in
BMC Ophthalmology / Ausgabe 1/2020
Elektronische ISSN: 1471-2415
DOI
https://doi.org/10.1186/s12886-020-01693-6

Weitere Artikel der Ausgabe 1/2020

BMC Ophthalmology 1/2020 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Ophthalmika in der Schwangerschaft

Die Verwendung von Ophthalmika in der Schwangerschaft und Stillzeit stellt immer eine Off-label-Anwendung dar. Ein Einsatz von Arzneimitteln muss daher besonders sorgfältig auf sein Risiko-Nutzen-Verhältnis bewertet werden. In der vorliegenden …

Operative Therapie und Keimnachweis bei endogener Endophthalmitis

Vitrektomie Originalie

Die endogene Endophthalmitis ist eine hämatogen fortgeleitete, bakterielle oder fungale Infektion, die über choroidale oder retinale Gefäße in den Augapfel eingeschwemmt wird [ 1 – 3 ]. Von dort infiltrieren die Keime in die Netzhaut, den …

Bakterielle endogene Endophthalmitis

Vitrektomie Leitthema

Eine endogene Endophthalmitis stellt einen ophthalmologischen Notfall dar, der umgehender Diagnostik und Therapie bedarf. Es sollte mit geeigneten Methoden, wie beispielsweise dem Freiburger Endophthalmitis-Set, ein Keimnachweis erfolgen. Bei der …

So erreichen Sie eine bestmögliche Wundheilung der Kornea

Die bestmögliche Wundheilung der Kornea, insbesondere ohne die Ausbildung von lichtstreuenden Narben, ist oberstes Gebot, um einer dauerhaften Schädigung der Hornhaut frühzeitig entgegenzuwirken und die Funktion des Auges zu erhalten.   

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.