The influence of ambient lighting on the detection of small contrast elements in digital dental radiographs

verfasst von: Till Schriewer, Ralf Schulze, Andreas Filippi, Irene Mischak, Michael Payer, Dorothea Dagassan-Berndt, Sebastian Kühl

Erschienen in: Clinical Oral Investigations | Ausgabe 7/2013

Einloggen, um Zugang zu erhalten

Abstract

Objectives

The aim of this study was to evaluate the effect of ambient light on the detection of contrast elements in digital dental radiographs.

Materials and methods

A high-contrast standardized digital radiograph of an aluminum step wedge containing 32 boreholes of different depth was cut into 40 isometric images. Images were presented at random on a 17-in cathode ray tube (CRT) monitor at different ambient background illuminations of 0, 50, 200, and 500 lx. Twenty observers stated twice their blinded decision whether or not they could perceive a dark spot on a five-point confidence scale. Areas (Az) under receiver operating characteristic curves were calculated and compared between the four different ambient illuminations using the Friedman test. Statistical significance was set at p < 0.05. Overall agreement was estimated determining the intraclass correlation coefficient.

Results

The Az values (0.735 for 0 lx, 0.728 for 50 lx, 0.735 for 200 lx, and 0.788 for 500 lx) did not significantly differ (p = 0.796) between the four ambient lighting levels.

Conclusions

The detection of small contrast features in digital dental radiographs on a CRT monitor seems to be comparable over a wide range of ambient background illumination.

Clinical relevance

The ambient light in dental offices of <500 lx may not negatively impact the diagnostic performance of digital dental radiographs.
Literatur
1.
Zurück zum Zitat Geissler O, Rother U (2007) The film-based printout using tetenal printing solution II. Int J Comput Dent 10:285–291PubMed Geissler O, Rother U (2007) The film-based printout using tetenal printing solution II. Int J Comput Dent 10:285–291PubMed
2.
Zurück zum Zitat Wenzel A (1999) Matters to consider when implementing direct digital radiography in the dental office. Int J Comput Dent 2:269–290PubMed Wenzel A (1999) Matters to consider when implementing direct digital radiography in the dental office. Int J Comput Dent 2:269–290PubMed
3.
Zurück zum Zitat Pollard B, Chawla AS, Delong DM, Hashimoto N, Samei E (2008) Object detectability at increased ambient lighting conditions. Med Phys 35:2204–2213PubMedCrossRef Pollard B, Chawla AS, Delong DM, Hashimoto N, Samei E (2008) Object detectability at increased ambient lighting conditions. Med Phys 35:2204–2213PubMedCrossRef
4.
Zurück zum Zitat Chawla AS, Samei E (2007) Ambient illumination revisited: a new adaptation-based approach for optimizing medical imaging reading environments. Med Phys 34:81–90PubMedCrossRef Chawla AS, Samei E (2007) Ambient illumination revisited: a new adaptation-based approach for optimizing medical imaging reading environments. Med Phys 34:81–90PubMedCrossRef
5.
Zurück zum Zitat Pollard BJ, Samei E, Chawla AS, Beam C, Heyneman LE, Hurwitz-Koweek LM et al (2012) The effect of ambient lighting in chest radiology reading rooms. J Digit Imaging. doi:10.1007/s10278-012-9459-5 Pollard BJ, Samei E, Chawla AS, Beam C, Heyneman LE, Hurwitz-Koweek LM et al (2012) The effect of ambient lighting in chest radiology reading rooms. J Digit Imaging. doi:10.​1007/​s10278-012-9459-5
6.
Zurück zum Zitat Siegel EL (1998) Economic and clinical impact of filmless operation in a multifacility environment. J Digit Imaging 11:42–47PubMedCrossRef Siegel EL (1998) Economic and clinical impact of filmless operation in a multifacility environment. J Digit Imaging 11:42–47PubMedCrossRef
7.
Zurück zum Zitat Grassl U, Schulze RKW (2007) In vitro perception of low-contrast features in digital, film, and digitized dental radiographs: a receiver operating characteristic analysis. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 103:694–701PubMedCrossRef Grassl U, Schulze RKW (2007) In vitro perception of low-contrast features in digital, film, and digitized dental radiographs: a receiver operating characteristic analysis. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 103:694–701PubMedCrossRef
8.
Zurück zum Zitat Kühl S, Krummenauer F, Dagassan-Berndt D, Lambrecht JT, d’Hoedt B, Schulze RKW (2011) Ink-jet printout of radiographs on transparent film and glossy paper versus monitor display: an ROC analysis. Clin Oral Investig 15:351–356PubMedCrossRef Kühl S, Krummenauer F, Dagassan-Berndt D, Lambrecht JT, d’Hoedt B, Schulze RKW (2011) Ink-jet printout of radiographs on transparent film and glossy paper versus monitor display: an ROC analysis. Clin Oral Investig 15:351–356PubMedCrossRef
9.
Zurück zum Zitat Schöfer H (2001) Monitorbefundung in Tageslicht beleuchteter Umgebung basierend auf DIN 6868–57. Bayerisches Landesamt für Arbeitsschutz, Arbeitsmedizin und Sicherheitstechnik Schöfer H (2001) Monitorbefundung in Tageslicht beleuchteter Umgebung basierend auf DIN 6868–57. Bayerisches Landesamt für Arbeitsschutz, Arbeitsmedizin und Sicherheitstechnik
10.
Zurück zum Zitat Scharitzer M, Prokop M, Weber M, Fuchsjager M, Oschatz E, Schaefer-Prokop E (2005) Detectability of catheters on bedside chest radiographs: comparison between liquid crystal display and high-resolution cathode-ray tube monitors. Radiology 234:611–616PubMedCrossRef Scharitzer M, Prokop M, Weber M, Fuchsjager M, Oschatz E, Schaefer-Prokop E (2005) Detectability of catheters on bedside chest radiographs: comparison between liquid crystal display and high-resolution cathode-ray tube monitors. Radiology 234:611–616PubMedCrossRef
11.
Zurück zum Zitat Goo JM, Choi JY, Im JG, Lee HJ, Chung MJ, Han D et al (2004) Effect of monitor luminance and ambient light on observer performance in soft-copy reading of digital chest radiographs. Radiology 232:762–766PubMedCrossRef Goo JM, Choi JY, Im JG, Lee HJ, Chung MJ, Han D et al (2004) Effect of monitor luminance and ambient light on observer performance in soft-copy reading of digital chest radiographs. Radiology 232:762–766PubMedCrossRef
12.
Zurück zum Zitat Hellén-Halme K, Lith A (2012) Effect of ambient light level at the monitor surface on digital radiographic evaluation of approximal carious lesions: an in vitro study. Dentomaxillofac Radiol 41:192–196PubMedCrossRef Hellén-Halme K, Lith A (2012) Effect of ambient light level at the monitor surface on digital radiographic evaluation of approximal carious lesions: an in vitro study. Dentomaxillofac Radiol 41:192–196PubMedCrossRef
13.
Zurück zum Zitat Buls N, Shabana W, Verbeek P, Pevenage P, De Mey J (2007) Influence of display quality on radiologist’ performance in the detection of lung nodules on radiographs. Br J Radiol 80:738–743PubMedCrossRef Buls N, Shabana W, Verbeek P, Pevenage P, De Mey J (2007) Influence of display quality on radiologist’ performance in the detection of lung nodules on radiographs. Br J Radiol 80:738–743PubMedCrossRef
14.
Zurück zum Zitat Balassy C, Prokop M, Weber M, Sailer J, Herold C, Schaefer-Prokop C (2005) Flat-panel display (LCD) versus high-resolution grays-scale display (CRT) for chest radiography: an observer preference study. Am J Roentgenol 184:752–756CrossRef Balassy C, Prokop M, Weber M, Sailer J, Herold C, Schaefer-Prokop C (2005) Flat-panel display (LCD) versus high-resolution grays-scale display (CRT) for chest radiography: an observer preference study. Am J Roentgenol 184:752–756CrossRef
15.
Zurück zum Zitat Otis L, Sherman RG (2005) Assessing the accuracy of caries diagnosis via radiograph. Film versus print. J Am Dent Assoc 136:323–330PubMed Otis L, Sherman RG (2005) Assessing the accuracy of caries diagnosis via radiograph. Film versus print. J Am Dent Assoc 136:323–330PubMed
Metadaten
Titel
The influence of ambient lighting on the detection of small contrast elements in digital dental radiographs
verfasst von
Till Schriewer
Ralf Schulze
Andreas Filippi
Irene Mischak
Michael Payer
Dorothea Dagassan-Berndt
Sebastian Kühl
Publikationsdatum
01.09.2013
Verlag
Springer Berlin Heidelberg
Erschienen in
Clinical Oral Investigations / Ausgabe 7/2013
Print ISSN: 1432-6981
Elektronische ISSN: 1436-3771
DOI
https://doi.org/10.1007/s00784-012-0858-2

Weitere Artikel der Ausgabe 7/2013

Clinical Oral Investigations 7/2013 Zur Ausgabe

Parodontalbehandlung verbessert Prognose bei Katheterablation

19.04.2024 Vorhofflimmern Nachrichten

Werden Personen mit Vorhofflimmern in der Blanking-Periode nach einer Katheterablation gegen eine bestehende Parodontitis behandelt, verbessert dies die Erfolgsaussichten. Dafür sprechen die Resultate einer prospektiven Untersuchung.

Invasive Zahnbehandlung: Wann eine Antibiotikaprophylaxe vor infektiöser Endokarditis schützt

11.04.2024 Endokarditis Nachrichten

Bei welchen Personen eine Antibiotikaprophylaxe zur Prävention einer infektiösen Endokarditis nach invasiven zahnärztlichen Eingriffen sinnvoll ist, wird diskutiert. Neue Daten stehen im Einklang mit den europäischen Leitlinienempfehlungen.

Zell-Organisatoren unter Druck: Mechanismen des embryonalen Zahnwachstums aufgedeckt

08.04.2024 Zahnmedizin Nachrichten

Der Aufbau von Geweben und Organen während der Embryonalentwicklung wird von den Zellen bemerkenswert choreografiert. Für diesen Prozess braucht es spezielle sogenannte „Organisatoren“. In einer aktuellen Veröffentlichung im Fachjournal Nature Cell Biology berichten Forschende durch welchen Vorgang diese Organisatoren im Gewebe entstehen und wie sie dann die Bildung von Zähnen orchestrieren.

Die Oralprophylaxe & Kinderzahnheilkunde umbenannt

11.03.2024 Kinderzahnmedizin Nachrichten

Infolge der Umbenennung der Deutschen Gesellschaft für Kinderzahnheilkunde in Deutsche Gesellschaft für Kinderzahnmedizin (DGKiZ) wird deren Mitgliederzeitschrift Oralprophylaxe & Kinderzahnheilkunde in Oralprophylaxe & Kinderzahnmedizin umbenannt. Aus diesem Grunde trägt die erste Ausgabe in 2024 erstmalig den neuen Titel.

Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Zahnmedizin und bleiben Sie gut informiert – ganz bequem per eMail.