The value of 18F-FDG PET/CT for the diagnosis of device-related infections in patients with a left ventricular assist device: a systematic review and meta-analysis

verfasst von: D. ten Hove, G. Treglia, R. H. J. A. Slart, K. Damman, M. Wouthuyzen-Bakker, D. F. Postma, O. Gheysens, R. J. H. Borra, G. Mecozzi, P. P. van Geel, B. Sinha, A. W. J. M. Glaudemans

Erschienen in: European Journal of Nuclear Medicine and Molecular Imaging | Ausgabe 1/2021

Abstract

Background

Left ventricular assist devices (LVADs) are increasingly used for the treatment of advanced heart failure. LVADs improve quality of life and decrease mortality, but the driveline carries substantial risk for major infections. These device-related LVAD and driveline infections are difficult to diagnose with conventional imaging. We reviewed and analysed the current literature on the additive value of 18F-fluorodeoxyglucose positron emission tomography combined with computed tomography (FDG-PET/CT) imaging for the diagnosis of LVAD-related infections.”

Materials/methods

We performed a systematic literature review using several databases from their inception until the 31st of December, 2019. Studies investigating the diagnostic performance of FDG-PET/CT in patients with suspected LVAD infection were retrieved. After a bias risk assessment using QUADAS-2, a study-aggregate meta-analysis was performed on a per examination-based analysis.

Results

A total of 10 studies were included in the systematic review, eight of which were also eligible for study-aggregate meta-analysis. For the meta-analysis, a total of 256 FDG-PET/CT scans, examining pump/pocket and/or driveline infection, were acquired in 230 patients. Pooled sensitivity of FDG-PET/CT was 0.95 (95% confidence interval (CI) 0.89–0.97) and pooled specificity was 0.91 (95% CI 0.54–0.99) for the diagnosis of device-related infection. For pump/pocket infection, sensitivity and specificity of FDG-PET/CT were 0.97 (95%CI 0.69–1.00) and 0.93 (95%CI 0.64–0.99), respectively. For driveline infection, sensitivity and specificity were 0.96 (95%CI 0.88–0.99) and 0.99 (95%CI 0.13–1.00) respectively. Significant heterogeneity existed across studies for specificity, mostly caused by differences in scan procedures. Predefined criteria for suspicion of LVAD and/or driveline infection were lacking in all included studies.

Conclusions

FDG-PET/CT is a valuable tool for assessment of device-related infection in LVAD patients, with high sensitivity and high, albeit variable, specificity. Standardization of FDG-PET/CT procedures and criteria for suspected device-related LVAD infections are needed for consistent reporting of FDG-PET/CT scans.
Begleitmaterial
Hinweise

Electronic supplementary material

The online version of this article (https://​doi.​org/​10.​1007/​s00259-020-04930-8) contains supplementary material, which is available to authorized users.
This article is part of the Topical Collection on Infection and inflammation
B. Sinha and A. W. J. M. Glaudemans are shared authorship

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
BSAC
British Society of Antimicrobial Chemotherapy
CF-LVAD
Continuous-flow left ventricular assist device
CT
Computed tomography
DOR
Diagnostic odds ratio
EANM
European Association of Nuclear Medicine
EARL
EANM Research ltd
ESC
European Society of Cardiology
FDG
18F-fluorodeoxyglucose
ISHLT
International Society of Heart and Lung Transplantation
ICD
Implantable cardioverter defibrillator
LVAD
Left ventricular assist device
NLR
Negative likelihood ratio
NR
Not reported
PET
Positron emission tomography
PLR
Positive likelihood ratio
PRISMA-DTA
Preferred reporting items for systematic reviews and meta-analyses of diagnostic test accuracy studies
QUADAS-2
Quality Assessment of Diagnostic Accuracy Studies-version 2
SUVmax
Maximum standardized uptake value
TBR
Target-to-background ratio
VAD
(any) ventricular assist device

Introduction

Left ventricular assist devices (LVADs) are an established treatment option for end-stage heart failure, either as a bridge-to-transplantation, bridge to decision or as destination therapy. LVAD treatment is associated with improvement in quality of life and improved survival. Already in 2001, LVADs have been shown to improve 1-year survival from 25 to 50% compared with conservative medical treatment [1]. With subsequent LVAD generations, outcomes have further improved, with 4-year survival for LVAD recipients now approaching that of heart transplantation (60% and 70%, respectively) [2].
However, infection of either the driveline or the LVAD pocket or pump itself still remains an important clinical problem. The overall incidence has decreased over time, but infection still occurs in 18.1% of patients during the first year after implantation and in 11.9% the years thereafter [2]. LVAD infections are associated with significant morbidity and mortality [3], in particular when complicated by bloodstream infection, which has an overall mortality rate of up to 50% [4, 5]. Establishing the diagnosis accurately and at an early stage is essential for effective management and optimal patient outcome.
The diagnosis of device-related LVAD infections mainly relies on clinical findings and results from microbiology and imaging. Swabs taken at the driveline exit site and blood cultures are a mainstay for the diagnosis, but they provide no information about the extent of an infection. Surgical removal and subsequent culture of the device is the gold standard for diagnosis, but this is often not feasible because of the severe risks associated with exchanging these devices. Imaging techniques such as echocardiography and CT angiography (CTA) are commonly used, but their diagnostic accuracy is limited due to device-related scatter artefacts, while LVAD components themselves may mimic infectious complications e.g. appearance of partially obstructed flow on echocardiography or blood between outflow graft and surrounding Gore-Tex mimicking thrombus on contrast-enhanced CT [6, 7].
Molecular imaging, and specifically 18F-fluordeoxyglucose positron emission tomography (FDG-PET) combined with low-dose or contrast-enhanced CT (FDG-PET/CT), is increasingly used for assessment of device-related infections. For endocarditis and infections in patients with cardiac implantable electronic devices, i.e. pacemakers, implantable cardioverter defibrillators (ICDs), FDG-PET/CT has already been incorporated in ESC guidelines and its diagnostic value for this indication is supported by an extensive body of evidence [811]. Its value for the diagnosis of LVAD-related infections is still being investigated, but here supporting evidence is also emerging.
The aim of this systematic review and meta-analysis is to provide a detailed overview of all evidence so far to establish the role of FDG-PET/CT in diagnosing LVAD-related infections. For the analysis, a distinction was made between driveline infections and infections of the pump/pocket.

Research design and methods

Screening and selection of literature

This systematic review and meta-analysis was performed according to the Cochrane methodology and PRISMA-DTA statement [12]. A comprehensive literature search was performed by two authors (DtH and GT) on PubMed, the Cochrane Library database and Embase. The search included the following terms: ‘Left Ventricular Assist Device’, ‘Infection’, ‘Driveline Infection’, ‘Endocarditis’”, and ‘Positron Emission Tomography’ or variations of these search terms. For the exact search strings, we refer to the supplemental data. Studies published up to the 31st of December 2019 were used in our analyses. Original articles that evaluated the diagnostic performance of FDG-PET/CT for suspicion of LVAD infection were eligible for inclusion in the systematic review. References in selected studies were cross-checked to find other relevant articles. Both retrospective and prospective studies as well as blinded and non-blinded studies were included. We excluded case reports and case series with small patient numbers (n < 5), review articles without original data, editorials, letters, and conference papers. All studies included in the systematic review were eligible for the study-aggregate meta-analysis, with exception for those with unacceptable risk of bias (e.g. no valid reference test) and/or patient overlap. Two researchers (DtH and GT) independently reviewed titles and abstracts of the retrieved articles, applying the inclusion and exclusion criteria mentioned above. The full text of the remaining articles was examined to assess their eligibility for inclusion in the study-aggregate meta-analysis. Disagreements were resolved in a consensus meeting with a third reviewer (AG).

Data extraction and quality assessment

QUADAS-2 [13] was used to systematically assess the risk of bias and applicability concerns for all included studies. The criteria considered by QUADAS-2 are selection bias, index test bias, reference test bias, and flow-and-timing bias.
Selection bias risk was considered high if there were unexplained exclusions in the study and considered unknown when selection criteria were not (fully) described. FDG-PET/CT was considered to be the index test. Bias risk for the FDG-PET/CT scan was deemed low if the imaging specialists were blinded to the results of other diagnostic modalities and the final diagnosis of patients and if the scan was performed according to EANM/EARL procedural guidelines [1416] These entail patient preparation with a low-carbohydrate, fat allowed diet and a period of fasting before the scan of at least 6 h and analysis of both attenuation corrected and uncorrected PET images. If assessors were not blinded and EARL/EANM procedural guidelines were not followed, the risk of bias was considered high. All studies in between, with either EARL/EANM recommendations not followed or with non-blinded assessors, were considered intermediate/unknown risk. Because the multidisciplinary consensus criteria according to the International Society of Heart and Lung transplantation (ISHLT) [17] do not constitute a true gold standard, but are currently the best known alternative, bias risk for reference test was considered intermediate for all studies that adhered to these criteria for the diagnosis. Those that deviated from ISHLT criteria were considered high risk. For flow and timing, assessment of bias risk was complicated by the fact that the ‘adequate’ time interval between index test and reference test is unknown (e.g. optimal duration of follow-up). Additionally, and in particular in the situation where patients were already treated with antibiotics at the time of FDG-PET/CT, the duration of antibiotic use may influence the value of the scan for the diagnosis, but its exact impact is unknown. Therefore, all studies were considered ‘unknown risk’ for this domain.

Reference standard of diagnosis

For classification of the diagnosis of both driveline infections and infections of the central LVAD components (pump housing, outflow tract, and pump pocket for earlier LVAD generations, e.g. Heartmate II), we adhered to diagnostic criteria proposed in the 2011 consensus statement by the ISHLT [17] and the similar adverse event definition of device specific major infection of INTERMACS [18]. Accordingly, it was verified for all studies whether they included findings of all clinical investigations, including cultures/swabs, trans-oesophageal echocardiography, CTA if available, clinical course, and follow-up. Because of the diagnostic challenge LVAD infections may present, it was also checked whether the final diagnosis was made by a specialized multidisciplinary team, consisting of cardiologists, thoracic surgeons, infectious disease specialists, medical microbiologists, and imaging specialists, with access to all relevant clinical information in case there was any doubt about the clinical diagnosis.

Statistical analysis

Statistical analyses were performed using Open Meta-Analyst (BROWN School of Public Health, Providence, RI, USA). Pooled subgroup analyses were performed for all included studies that evaluated FDG-PET/CT for its diagnostic value in establishing or ruling out driveline infections and/or infections of LVAD pump/pocket. Since two of the included studies only focused on FDG-PET/CT assessment of the driveline [19, 20] apart from the overall analysis of FDG-PET/CT accuracy, additional analyses were performed for driveline and central device components separately. Bivariate analysis of sensitivity and specificity was performed using likelihood ratio estimates with 95% confidence intervals (CI). An I2 higher than 50% was considered indicative of significant study heterogeneity [21]. Negative and positive likelihood ratios, as well as diagnostic odds ratios (DOR), were calculated. Negative and positive predictive values and the diagnostic accuracy were not considered as accurate since the prevalence of LVAD and/or driveline infections in the patient population of interest is unknown, violating an assumption for NPV, PPV, and diagnostic accuracy calculations. The values of negative likelihood ratio (NLR) and positive likelihood ratio (PLR) indicate to what extent the probability of having a disease decreases given a negative test result and how much the probability of the disease increases, given a positive test, respectively. The diagnostic odds ratio indicates how the probability of a correct diagnosis changes after performing the test (with a higher value indicating better performance).

Results

Selection of literature

A total of 71 articles were identified through an electronic database search (Fig. 1). After removing one duplicate, the remaining 70 articles were screened based on title and/or abstract. Fifty-nine studies were excluded because they either had a different focus than the research question, presented no original data, or lacked a full text. Eleven studies were deemed eligible for full-text analysis. Cross-checking references for any additional publications yielded no extra results. One of the articles was excluded from further analysis because it contained insufficient data specific to our research question (only two patients in the study population had an LVAD) [22]. In total, 10 studies (n = 382 scans in 318 patients) were included in the systematic review. Two of the studies that were included in the systematic review were excluded from the meta-analysis: one because of suspected data overlap with a later study published by the same author [23, 24], the other because of methodology/applicability concerns based on full-text analysis [24]. The latter study included analyses of FDG-PET/CT accuracy, but microbiological diagnosis was based on driveline exit site swabs only, which cannot be used as a standalone reference test for any deeper infection of the driveline or central device components. Furthermore, this study included only patients with a relatively late stage of infection, leading to selection bias and applicability concerns. Therefore, for the meta-analysis, eight studies (n = 256 scans in 230 patients) were found eligible.

Systematic review: study characteristics

In the ten articles included in the systematic review, a total of 382 FDG-PET/CT scans were acquired for 318 patients. A suspicion of LVAD-related infection was the reason for performing the FDG-PET/CT in 232 scans, 6 of which were for evaluation of treatment, while all others were considered separate episodes. The remaining 150 scans were either part of work-up for heart transplantation or assessment of pathology unrelated to LVAD (e.g. malignancy). In 78 scans, only the driveline was evaluated [19, 20]. One publication had a prospective study design [25], while all others used retrospective patient data. Median age of participants ranged from 52 to 64 years. The study population was predominantly male, proportions ranging from 77.8 to 90.5%. The characteristics of the ten included studies are summarized in Table 1.
Table 1
Study and patient characteristics
Authors
Year
Country
Study design
Type of patients evaluated
No. of FDG-PET/CT scans (patients)
Median age (years)
% male
Diagnosis of LVAD-specific infection1
Akin et al. [26]
2018
Netherlands
R
Suspected device-related infection
10 (9)
54
77.8%
8/10
Avramovic et al. [19]
2017
Germany
R
Suspected device-related infection (focus of study: driveline) or PET/CT as part of work-up for heart transplantation
48 (48)
57
83.3%
24/48
Bernhardt et al. [27]
2017
Germany
R
Suspected device-related infection
29 (21)
54
90.5%
16/29
Dell’ Aquila et al. [28]
2016
Germany
R
Suspected device-related infection
40 (31)
52
78.1%
30/40
Dell’ Aquila et al. [23]
2018
Germany
R
Suspected device-related infection
61 (47)
64
82.0%
40/61
De Vaugelade et al. [29]
2019
France
R
Suspected device-related infection
24 (22)
57
87.5%
21/24
Kanapinn et al. [20]
2019
Germany
R
Suspected device-related infection (all had baseline scan before: focus of study: driveline)
30 (30)
54
86.7%
23/30
Kim et al. [25]
2019
USA
P
Suspected device-related infection. Controls: baseline PET/CT
35 (35)
54
80.0%
28/35
Sommerlath Sohns et al. [24]
2019
Germany
R
Device-related infection, evaluation of extent of infection
85 (57)
56
86.0%
85/85
Tam et al. [30]
2019
USA
R
Suspected device-related infection
19 (18)
61
78.9%
17/19
FDG fluorine-18 fluorodeoxyglucose, LVAD left ventricular assist device, P prospective, PET/CT positron emission tomography/computed tomography, R retrospective

Technical aspects

In all studies, FDG-PET scans were performed on a hybrid PET/CT system, combining an FDG-PET scan with a low-dose CT for anatomical reference and attenuation correction. In one study, the FDG-PET scan was combined with diagnostic CTA [27]. Reporting of injected activity differed between studies: while some reported an injected activity per kilogram of body weight, others reported a mean total injected activity with lower and upper ranges. The injected activity was also highly variable for included studies, ranging from 215 to 474 MBq for the mean total activity and 2.3 to 5 MBq per kg body weight (EANM guidelines advice: 2.5–5.0 MBq/kg [15]). According to study protocols, all scans were performed approximately 60 min after injection of FDG. However, the actual time intervals in clinical practice were not reported.
Visual analysis of the scans was performed in all studies; in 5 studies, this was combined with semi-quantitative analyses, using SUVmax [19, 20, 24] and target-to-background ratios (TBR) [23, 24, 28], reference regions being lung parenchyma and deltoid muscle [23, 28], or thoracic aorta and liver [24]. In one study, metabolic volume was also used: this was defined as the measured volume of a target lesion showing more FDG uptake than the mean FDG uptake in a delineated region of interest in the liver plus 2.5 standard deviations, with a minimum volume of 9 cm3 [19]. The technical details of the included studies are summarized in Table 2.
Table 2
Technical aspects of 18F-FDG PET/CT studies included in systematic review
First author, year
Imaging modality
Mean injected activity per kg and total
Time interval FDG injection and image acquisition1
Image analysis
Comparison to other imaging modalities
Akin et al. 2018 [26]
PET/CT(low-dose CT)
2.3 MBq/kg
μtot = NR
60 min.
Visual analysis
None
Avramovic et al. 2017 [19]
PET/CT, (low-dose CT)
5 MBq/kg μtot = 338 MBq
60 min.
Visual + semi-quantitative analysis (SUVmax and MV)
None
Bernhardt et al. 2017 [27]
PET/CT(contrast-enhanced CT)
MBq/kg NR
μtot = 351 MBq
60 min.
Visual analysis
None
Dell’ Aquila et al. 2016 [28]
PET/CT(low-dose CT)
5 MBq/kg μtot = 308 MBq
60 min.
Visual + semi-quantitative analysis (SUVmax, SUVmean, TBR)
None
Dell’ Aquila et al. 2018 [23]
PET/CT(low-dose CT)
5 MBq/kg μtot = 344 MBq
60 min.
Visual + semi-quantitative analysis (SUVmax, SUVmean, TBR)
None
De Vaugelade et al. 2019 [29]
PET/CT(low-dose CT)
3.5 MBq/kg μtot = 310 MBq
60 min.
Visual analysis
WBC-SPECT
Kanapinn et al. 2019 [20]
PET/CT(low-dose CT)
MBq/kg NR
μtot = 215 MBq (1st scan)
μtot = 218 MBq (2nd scan)
60 min.
Visual + semi-quantitative analysis (SUVmax, SUVmean,)
None
Kim et al. 2019 [25]
PET/CT(low-dose CT)
MBq/kg NR
μtot = 474 MBq
60 min.
Visual analysis
None
Sommerlath Sohns et al. 2019 [24]
PET/CT(low-dose CT)
MBq/kg NR
μtot = NR, range 198–326 MBq
60 min.
Visual + semi-quantitative analysis (SUVmax, TBR)
None
Tam et al. 2019 [30]
PET/CT(low-dose CT)
MBq/kg NR
μtot = NR, range 333–370 MBq
60 min.
Visual analysis
None
FDG fluorine-18 fluorodeoxyglucose, INTERMACS Interagency Registry for Mechanical Circulatory Support, LVAD left ventricular assist device, MBq MegaBecquerel, min. minutes, μtot mean total injected activity, MV metabolic volume, NR not reported, PET/CT positron emission tomography/computed tomography, SUVmax maximal standardized uptake value, SUVmean mean standardized uptake value, TBR target-to-background ratio, WBC-SPECT white blood cell single photon emission computed tomography
1According to study protocol; actual values during study not reported

Methodological quality of included studies

The QUADAS-2 risk of bias of all studies evaluated for meta-analysis eligibility is summarized in Fig. 2. Two studies had a high risk of bias for patient selection, one due to unexplained patient exclusions [25], the other because of a case series of patients with late-stage infections [24]. All other studies described a suspicion of device-related infection as inclusion criterion, but this suspicion was not further elaborated or defined. Therefore, all other studies were considered to have an unknown risk for patient selection bias. Only one study had a low risk of bias for the index test, having assessors of the FDG-PET/CT blinded to findings of other clinical tests and final diagnosis for patients, while also performing the FDG-PET/CT scan according to EANM recommendations with a high-fat, low carbohydrate diet, a pre-scan fast of more than 6 h, and assessment of both attenuation-corrected and attenuation-uncorrected images [25]. In other studies, observers were either not blinded to clinical context of patients or assessment of non-attenuation-corrected images was not described. Two studies performed the reference test fully in accordance with ISHLT recommendations [27, 29]. Two studies had high applicability concerns for both index test and reference test, because they focused on the LVAD driveline only [19, 20].

Impact on prognosis and patient management

The ability of FDG-PET/CT to predict outcome and help inform management of device infections was discussed in three of the articles included in the systematic review [24, 25, 27]. In one study, a positive FDG-PET/CT was associated with a 50% mortality during follow-up (median survival 87.5 days), which contrasted with the non-infected group, in which no patients died during follow-up (median follow-up duration of 165 days). Twelve out of the 14 (86%) patients who died had involvement of pump or pocket infection [25]. In another study, FDG-PET/CT helped clinicians change their medical strategy for 12 out of 21 patients (57%), including four patients that were listed for high urgency heart transplantation based on FDG-PET/CT results. In all these cases, infection of the LVAD device or the deep driveline was confirmed at transplantation. [27]. In the third study, an association was found between FDG uptake of thoracic lymph nodes and adverse outcome, although this was not found for increased FDG uptake along the driveline or around any central LVAD device component [24].

Meta-analysis: pooled diagnostic performance

In the eight articles included in the study-aggregate meta-analysis, a total of 256 FDG-PET/CT scans were acquired in 230 patients. A suspicion of device-related infection was the reason for performing FDG-PET/CT in 232 scans. In 78 scans, only the driveline was evaluated [19, 20].
For the assessment of overall device-related infections, pooled sensitivity and specificity of FDG-PET/CT were 0.95 (95% CI 0.89–0.97) and 0.91 (95% CI 0.54–0.99) respectively. NLR was 0.14 and positive likelihood ratio, PLR, was 3.54 with an overall DOR of 38.43. When only assessing the driveline, FDG-PET/CT pooled sensitivity, specificity, NLR, PLR, and DOR were respectively 0.97 (95% CI 0.88–0.99) and 0.99 (95% CI 0.13–1.0, 0.13, 3.93, and 92.46. When only assessing pump/pocket infections, FDG-PET/CT pooled sensitivity and specificity were 0.97 (95% CI 0.70–1.0) and 0.93 (95% CI 0.64–0.99) respectively. NLR was 0.12 and PLR was 5.56 with an overall DOR of 49.43.
The I2 test for heterogeneity was positive (> 50%) for PLR of FDG-PET/CT, for assessment of driveline only, pump/pocket only, and the combination of both. Results of the meta-analysis for LVAD-specific infections, in which findings for pump/pockets and driveline are combined, are summarized in Table 3 and Fig. 3. The split analyses of driveline and pump/pocket infections are shown as ROC curves in Fig. 4. The corresponding tables and forest plots can be found under supplemental data: Tables 4 and 5, Figs. 6 and 7. Plots for FDG-PET/CT diagnostic odds ratios are represented in supplemental data Figs. 8–10.
Table 3
Overall diagnostic performance of FDG-PET/CT in patients with suspected LVAD and/or driveline infection
Authors, year
Reference standard for diagnostic performance
assessment
True positive
False negative
False positive
True negative
Sensitivity
Specificity
PLR
NLR
Akin et al. 2018 [26]
Clinical course review by research group including medical history, comorbidities, cultures of blood and driveline (sternal wound if suspect), laboratory tests, imaging results, and outcome at end of recorded follow-up. Diagnosis according to INTERMACS definition of LVAD infection.
8
0
0
2
1.0
1.0
0.00
Avramovic et al. 2017 [19]
Clinical course review at the end of recorded follow-up or transplantation: clinical evidence of infection or recurrence of symptoms, swabs at driveline exit, along driveline, surgical samples if available, and laboratory tests. Diagnosis according to INTERMACS definition of LVAD infection.
Visual 21
MV 23
3
1
5
3
19
21
0.875
0.958
0.792
0.875
4.20
7.67
0.16
0.05
Bernhardt et al. 2017 [27]
ISHLT criteria at end of follow-up, based on clinical symptoms, cultures, and swabs of exit site, along driveline and during surgery if available, and imaging data. In case of missing data, consensus diagnosis made during multidisciplinary meeting.
14
2
0
13
0.875
1.0
0.12
Dell’ Aquila et al. 2016 [28]
Findings of MMB (cultures of skin and/or tissue surrounding driveline or central device components if available), surgery, clinical evidence of infection, and recurrence of symptoms at end of recorded follow-up, diagnosis according to INTERMACS definition of LVAD infection.
30
0
2
8
1.0
0.800
5.00
0.00
Dell’ Aquila et al. 2018 [23]
Clinical evidence of infection, cultures of skin and/or tissue surrounding driveline or central device components if available), surgery, and recurrence of symptoms at end of recorded follow-up. Diagnosis according to INTERMACS definition of LVAD infection.
36
4
6
15
0.900
0.714
3.15
0.14
De Vaugelade et al. 2019 [29]
ISHLT criteria at end of follow-up, based on clinical symptoms, microbiology, and imaging data. In case of missing data, consensus diagnosis made during multidisciplinary meeting.
20
1
1
2
0.952
0.667
2.86
0.07
Kanapinn et al. 2019 [20]
Consensus by 2 physicians with access to clinical criteria, findings of MMB (not further defined), and all diagnostic imaging (incl. FDG-PET/CT).
23
0
0
7
1.0
1.0
0.00
Kim et al. 2019 [25]
Findings of MMB, surgery, clinical evidence of infection, and recurrence of symptoms; it was not reported who performed the reference test.
28
0
0
7
1.0
1.0
0.00
Sommerlath Sohns et al. 2019 [24]
Clinician determined presence or absence of LVAD infection based on history, laboratory tests, imaging studies, and clinical outcome. Confirmation at 30 day follow-up.
11
0
6
2
1.0
0.250
1.33
0.00
FDG fluorine-18 fluorodeoxyglucose, INTERMACS Interagency Registry for Mechanical Circulatory Support, ISHLT International Society of Heart and Lung Transplantation, LVAD left ventricular assist device, MMB medical microbiology, MV metabolic volume, PET/CT positron emission tomography/computed tomography
While 5 studies mentioned the use of semi-quantitative analysis [19, 20, 23, 24, 28], only 3 of these described its findings in comparison with visual analysis. Visual analysis outperformed semi-quantitative analysis in 2 studies [23, 28] while in one study [19], both semi-quantitative analyses using SUVmax and especially metabolic volume with a cutoff of > 9 cm3 outperformed visual analysis, with 2/3 false negatives and 2/5 false positives correctly classified using metabolic volume.
In one study, all patients underwent two scans: a baseline scan without suspicion of infection, and a second one for assessment of driveline infection [20]. The baseline scan may have facilitated the interpretation of the second diagnostic scan, which might explain the absence of any false positives or false negatives in this study, although this warrants validation in further studies.
Analysis of false positive and false negative scans was performed in 4 studies [23, 27, 29, 30]. In one study, the cause of 2 false negatives could not be established [24]. In another, it is implied that the reason for their single false negative result might have been the 30-day period of antibiotic treatment at the time of the scan [25]. In the third study, out of 6 false positives, 4 patients had concurrent bacteraemia or other possible sources of infection, 1 patient had increased cardiac sarcoidosis activity, and 1 had a newly diagnosed chronic myeloid leukaemia. The exact effect of these comorbidities on FDG-PET/CT results in their study remained unclear. The most extensive analysis of false positives and false negatives was performed by Dell’ Aquila et al. [23]. They described prolonged antibiotic use, infection limited to the pump housing as the causes for false negatives, and the presence of chronic fistulas as main causes for a false positives in 3 cases, whereas 7 other cases remained unexplained.
Analyses of scans performed shortly after LVAD implantation showed robustness of the scan in this setting: in one study, a true negative was reported 3 weeks after LVAD implantation [26] and in another, 5 true positives and 5 true negatives were reported within 3 months after LVAD implantation [23].

Discussion

We have pooled the data on the diagnostic value of FDG-PET/CT in detecting pump/pocket and driveline infections in patients with a LVAD to obtain more robust estimates of diagnostic performance of FDG-PET/CT in this setting. FDG-PET/CT is already included in guidelines for endocarditis and cardiovascular implantable electronic devices. Supporting evidence is emerging for the use of FDG-PET/CT in device-related infection in patients with LVADs. However, most of the reported studies have limited power, due to relatively small patient numbers enrolled and different acquisition and interpretation criteria. The separate evaluation of FDG-PET/CT accuracy for infections of LVAD pump/pocket and the driveline, next to the analysis in which these were combined, allowed us to include a significant amount of studies and patients to the analyses. In addition, we performed a further in-depth analysis of the included studies’ methodology and a stratification for driveline versus central device components, with recommendations for future studies. We performed no separate analyses for white blood cell (WBC) scintigraphy. Its diagnostic value was evaluated by only one study [29], making any pooling of data impossible beforehand. Summarily, in this study, FDG-PET/CT was found to have higher sensitivity (95.2 vs 71.9, p = 0.01), while a difference in specificity favouring WBC scintigraphy was not found to be statistically significant (66.7 vs 100, p = 0.32), although the study was underpowered to detect such a difference with only 3 negative cases. To our opinion, while potentially useful as a high specificity test in situations where FDG-PET/CT results are unclear, evidence for WBC scintigraphy so far is insufficient to make any recommendations in that regard.

Clinical value of FDG-PET/CT in suspicion of LVAD-related infections

Considering the value of FDG-PET/CT in LVAD-related infections, we found a high overall sensitivity and specificity (both above 90%), underscoring its value in clinical practice. It was also found to have impact on prognosis and patient management. This is particularly important because of the severity of these infections and the difficulty of both their diagnosis and treatment. Accurate information about the presence and extent of an infectious process is of great importance for determining appropriate treatment e.g. duration of antibiotic treatment and/or extent of surgical debridement, while follow-up scans may be used to evaluate treatment response.

Heterogeneity and technical considerations of FDG-PET/CT

Although the overall accuracy of FDG-PET/CT for the diagnosis of device-related infection was excellent, we also found significant heterogeneity amongst studies. The current lack of a standardized FDG-PET/CT procedure, such as the wide variety of injected activity, the possibly variable intervals between injection of the FDG and the subsequent scan, the variable use of a low carbohydrate, fat allowed diet prior to FDG-PET/CT, and missing analysis of non-attenuation-corrected PET images along with the attenuation-corrected images, may well explain the wide confidence intervals that were found for specificity of the test and the corresponding heterogeneous positive likelihood ratio. If the low carbohydrate, fat allowed diet is not used, there is a substantial risk of physiological myocardial uptake [31]. This may render any assessment of the pump housing impossible. The use of non-attenuation-corrected FDG-PET/CT images to confirm increased uptake surrounding the device is important, because the attenuation correction for the FDG-PET is based on the CT images, which means beam hardening artefacts are incorporated in the calculated FDG uptake, leading to distortions [32]. Further standardization of FDG-PET/CT protocols using the EARL criteria, applying a strict protocol for patient preparation, and providing robust interpretation criteria could substantially reduce heterogeneity caused by such confounders and increase consistency of the high overall specificity. The findings of scans performed shortly after LVAD implantation suggest that reactive inflammation after LVAD implantation may be relatively short, making FDG-PET/CT feasible early after surgery, possibly as soon as 1 month after device implantation and almost certainly 3 months after LVAD implantation.
Studies comparing visual analysis with semi-quantitative analysis found conflicting results on which of these is the most accurate during assessment of pump/pocket infections or driveline infections. Using both is probably the best approach in clinical practice until more evidence is gathered for preferring one method over the other. A clear limitation of semi-quantitative analysis is that cutoff values are not necessarily interchangeable between different PET/CT systems. This can be circumvented by calibrating the PET/CT system according to EARL or using either metabolic volume or reference regions to determine increased FDG avidity surrounding the device or driveline.
Inclusion criteria for all studies investigated were a clinical suspicion of driveline infection or infection of central LVAD components with or without a control group. However, no clearly defined criteria exist to establish a suspicion of device-related infection, which introduces a risk of selection bias e.g. overestimation of FDG-PET/CT accuracy if only performed in late-stage infection and underestimation of FDG-PET/CT accuracy if performed for incidental findings on echocardiography in spite of absent clinical signs of infection. Detailed description of the clinical presentation for all included patients can partially mitigate this risk, but to eliminate it entirely, predefined criteria for suspicion are needed.

Proposal for structured approach in suspicion of LVAD-related infection, including FDG/PET-CT

To eliminate selection bias, we propose to distinguish between pump/pocket and driveline infection, as both their assessment and clinical significance differs. Central infections would include infections of the pump pocket, and outflow tract, and, for older LVAD generations such as Heartmate II, the pump pocket. For the driveline infections, signs like localized pain and erythema with or without purulent discharge at the driveline exit site would lead to a suspicion of infection. For infection of pump or pocket, criteria could be derived from the guidelines aimed at standardizing the suspicion for infective endocarditis, as published by the BSAC [33], with adjustments for this specific patient group. Therefore, we propose the following criteria for suspecting an infection of central LVAD components; see Fig. 5. If infection of LVAD pump/pocket is suspected, FDG-PET/CT would be indicated either for establishing the diagnosis, for determining the extent of infection, or for assessing dissemination to other organs.

Limitations

When assessing the diagnostic accuracy of any test for establishing device-specific infections in patients with an LVAD, a fundamental difficulty is the absence of a gold standard, due to associated risk of surgery and the inability of conventional investigations to accurately determine the extent of infection. Furthermore, the included studies were all relatively small and with significant differences in study protocols, leading to large heterogeneity. We took these factors into account to provide the most comprehensive review of the evidence so far.
The included studies focused almost exclusively on continuous flow LVAD systems. While this might limit the generalization of the results, these devices represent the vast majority of modern ventricular assist devices. Moreover, although the devices were almost exclusively LVADs, they were not all of the same type and/or generation, and it is certainly possible that LVADs made by different manufacturers and different materials may show different physiological uptake, impacting FDG-PET/CT accuracy. Furthermore, a difficult implantation of the device may cause a prolonged inflammatory response, impairing test accuracy, but there are currently no FDG-PET/CT data available on the impact of these factors for clinical practice.

Conclusion

Our systematic review and meta-analysis demonstrates that FDG-PET/CT is a valuable tool for establishing or excluding the diagnosis of device specific infection in patients with a left ventricular assist device, with a high sensitivity and a high, albeit variable, specificity. Future studies, in which criteria for suspecting device infection and scan procedures are standardized, are needed to confirm that this will lead to consistently high specificity and to further elucidate the role of semi-quantitative analyses that can be used across different PET/CT systems. Despite these limitations, the current evidence strongly supports implementation of FDG-PET/CT in the standard work-up of patients with suspected LVAD-related infections, in particular when initial clinical investigations are inconclusive.

Compliance with ethical standards

Conflict of interest

The authors declare that they have no conflict of interest.

Ethics approval

This article does not contain any study with human participants or animals performed by any of the authors.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Anhänge

Electronic supplementary material

Literatur
1.
Zurück zum Zitat Rose EA, Gelijns AC, Moskowitz AJ, Heitjan DF, Stevenson LW, Dembitsky W, et al. Long-term use of a left ventricular assist device. N Engl J Med. 2001;345(20):1435–43.CrossRef Rose EA, Gelijns AC, Moskowitz AJ, Heitjan DF, Stevenson LW, Dembitsky W, et al. Long-term use of a left ventricular assist device. N Engl J Med. 2001;345(20):1435–43.CrossRef
2.
Zurück zum Zitat Goldstein DJ, Meyns B, Xie R, Cowger J, Pettit S, Nakatani T, et al. Third annual report from the ISHLT mechanically assisted circulatory support registry: a comparison of centrifugal and axial continuous-flow left ventricular assist devices. J Heart Lung Transplant. 2019;38(4):352–63.CrossRef Goldstein DJ, Meyns B, Xie R, Cowger J, Pettit S, Nakatani T, et al. Third annual report from the ISHLT mechanically assisted circulatory support registry: a comparison of centrifugal and axial continuous-flow left ventricular assist devices. J Heart Lung Transplant. 2019;38(4):352–63.CrossRef
3.
Zurück zum Zitat Kirklin JK, Naftel DC, Pagani FD, Kormos RL, Stevenson LW, Blume ED, et al. Seventh INTERMACS annual report: 15,000 patients and counting. J Heart Lung Transplant. 2015;34(12):1495–504.CrossRef Kirklin JK, Naftel DC, Pagani FD, Kormos RL, Stevenson LW, Blume ED, et al. Seventh INTERMACS annual report: 15,000 patients and counting. J Heart Lung Transplant. 2015;34(12):1495–504.CrossRef
4.
Zurück zum Zitat Aggarwal A, Gupta A, Kumar S, Baumblatt JA, Pauwaa S, Gallagher C, et al. Are blood stream infections associated with an increased risk of hemorrhagic stroke in patients with a left ventricular assist device? ASAIO J. 2012;58(5):509–13.CrossRef Aggarwal A, Gupta A, Kumar S, Baumblatt JA, Pauwaa S, Gallagher C, et al. Are blood stream infections associated with an increased risk of hemorrhagic stroke in patients with a left ventricular assist device? ASAIO J. 2012;58(5):509–13.CrossRef
5.
Zurück zum Zitat Trachtenberg BH, Cordero-Reyes AM, Aldeiri M, Alvarez P, Bhimaraj A, Ashrith G, et al. Persistent blood stream infection in patients supported with a continuous-flow left ventricular assist device is associated with an increased risk of cerebrovascular accidents. J Card Fail. 2015;21(2):119–25.CrossRef Trachtenberg BH, Cordero-Reyes AM, Aldeiri M, Alvarez P, Bhimaraj A, Ashrith G, et al. Persistent blood stream infection in patients supported with a continuous-flow left ventricular assist device is associated with an increased risk of cerebrovascular accidents. J Card Fail. 2015;21(2):119–25.CrossRef
6.
Zurück zum Zitat Lesicka A, Feinman JW, Thiele K, Andrawes MN. Echocardiographic artifact induced by HeartWare left ventricular assist device. Anesth Analg. 2015;120(6):1208–11.CrossRef Lesicka A, Feinman JW, Thiele K, Andrawes MN. Echocardiographic artifact induced by HeartWare left ventricular assist device. Anesth Analg. 2015;120(6):1208–11.CrossRef
7.
Zurück zum Zitat Shroff GS, Ocazionez D, Akkanti B, Vargas D, Garza A, Gupta P, et al. CT imaging of complications associated with continuous-flow left ventricular assist devices (LVADs). Semin Ultrasound CT MRI. 2017;38(6):616–28.CrossRef Shroff GS, Ocazionez D, Akkanti B, Vargas D, Garza A, Gupta P, et al. CT imaging of complications associated with continuous-flow left ventricular assist devices (LVADs). Semin Ultrasound CT MRI. 2017;38(6):616–28.CrossRef
8.
Zurück zum Zitat Habib G, Lancellotti P, Antunes MJ, Bongiorni MG, Casalta JP, Del Zotti F, et al. Guidelines for the management of infective endocarditis. Eur Heart J. 2015;36:3075–123.CrossRef Habib G, Lancellotti P, Antunes MJ, Bongiorni MG, Casalta JP, Del Zotti F, et al. Guidelines for the management of infective endocarditis. Eur Heart J. 2015;36:3075–123.CrossRef
9.
Zurück zum Zitat Sollini M, Lauri C, Boni R, Lazzeri E, Erba PA, Signore A. Current status of molecular imaging in infections. Curr Pharm Des. 2018;24(7):754–71.CrossRef Sollini M, Lauri C, Boni R, Lazzeri E, Erba PA, Signore A. Current status of molecular imaging in infections. Curr Pharm Des. 2018;24(7):754–71.CrossRef
10.
Zurück zum Zitat Erba PA, Lancellotti P, Vilacosta I, et al. Recommendations on nuclear and multimodality imaging in IE and CIED infections. Eur J Nucl Med Mol Imaging. 2018;45(10):1795–815.CrossRef Erba PA, Lancellotti P, Vilacosta I, et al. Recommendations on nuclear and multimodality imaging in IE and CIED infections. Eur J Nucl Med Mol Imaging. 2018;45(10):1795–815.CrossRef
11.
Zurück zum Zitat Blomström-Lundqvist C, Traykov V, Erba PA, Burri H, Nielsen JC, et al. ESC scientific document group. European Heart Rhythm Association (EHRA) international consensus document on how to prevent, diagnose, and treat cardiac implantable electronic device infections. Eur J Cardiothorac Surg. 2020;57(1):1–31.CrossRef Blomström-Lundqvist C, Traykov V, Erba PA, Burri H, Nielsen JC, et al. ESC scientific document group. European Heart Rhythm Association (EHRA) international consensus document on how to prevent, diagnose, and treat cardiac implantable electronic device infections. Eur J Cardiothorac Surg. 2020;57(1):1–31.CrossRef
12.
Zurück zum Zitat McInnes MDF, Moher D, Thombs BD, McGrath TA, Bossuyt PM, Clifford T, et al. Preferred reporting items for a systematic review and meta-analysis of diagnostic test accuracy studies the PRISMA-DTA statement. JAMA. 2018;319(4):388–96.CrossRef McInnes MDF, Moher D, Thombs BD, McGrath TA, Bossuyt PM, Clifford T, et al. Preferred reporting items for a systematic review and meta-analysis of diagnostic test accuracy studies the PRISMA-DTA statement. JAMA. 2018;319(4):388–96.CrossRef
13.
Zurück zum Zitat Whiting P, Rutjes AWS, Westwood M, Mallett S, Deeks JJ, Reitsma JB, et al. QUADAS-2: a revised tool for the quality assessment of diagnostic accuracy studies. Ann Intern Med. 2011;155:529–36.CrossRef Whiting P, Rutjes AWS, Westwood M, Mallett S, Deeks JJ, Reitsma JB, et al. QUADAS-2: a revised tool for the quality assessment of diagnostic accuracy studies. Ann Intern Med. 2011;155:529–36.CrossRef
14.
Zurück zum Zitat Boellaard R, Delgado-Bolton R, Oyen WJG, Giammarile F, Tatsch K, Eschner W, et al. FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0. Eur J Nucl Med Mol Imaging. 2015;42(2):328–54.CrossRef Boellaard R, Delgado-Bolton R, Oyen WJG, Giammarile F, Tatsch K, Eschner W, et al. FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0. Eur J Nucl Med Mol Imaging. 2015;42(2):328–54.CrossRef
15.
Zurück zum Zitat Jamar F, Buscombe J, Chiti A, Christian PE, Delbeke D, Donohoe KJ, et al. EANM/SNMMI guideline for 18F-FDG use in inflammation and infection. J Nucl Med. 2013;54(4):647–58.CrossRef Jamar F, Buscombe J, Chiti A, Christian PE, Delbeke D, Donohoe KJ, et al. EANM/SNMMI guideline for 18F-FDG use in inflammation and infection. J Nucl Med. 2013;54(4):647–58.CrossRef
17.
Zurück zum Zitat Hannan MM, Husain S, Mattner F, Danziger-Isakov L, Drew RJ, Corey GR, et al. Working formulation for the standardization of definitions of infections in patients using ventricular assist devices. J Heart Lung Transplant. 2011;30(4):375–84.CrossRef Hannan MM, Husain S, Mattner F, Danziger-Isakov L, Drew RJ, Corey GR, et al. Working formulation for the standardization of definitions of infections in patients using ventricular assist devices. J Heart Lung Transplant. 2011;30(4):375–84.CrossRef
19.
Zurück zum Zitat Avramovic N, Dell’Aquila AM, Weckesser M, Milankovic D, Vrachimis A, Sindermann JR, et al. Metabolic volume performs better than SUVmax in the detection of left ventricular assist device driveline infection. Eur J Nucl Med Mol Imaging. 2017;44(11):1870–7.CrossRef Avramovic N, Dell’Aquila AM, Weckesser M, Milankovic D, Vrachimis A, Sindermann JR, et al. Metabolic volume performs better than SUVmax in the detection of left ventricular assist device driveline infection. Eur J Nucl Med Mol Imaging. 2017;44(11):1870–7.CrossRef
20.
Zurück zum Zitat Kanapinn P, Burchert W, Körperich H, Körfer J. 18F-FDG PET/CT-imaging of left ventricular assist device infection: a retrospective quantitative intrapatient analysis. J Nucl Cardiol. 2019;26(4):1212–21. Kanapinn P, Burchert W, Körperich H, Körfer J. 18F-FDG PET/CT-imaging of left ventricular assist device infection: a retrospective quantitative intrapatient analysis. J Nucl Cardiol. 2019;26(4):1212–21.
21.
Zurück zum Zitat Higgins JPT, Thompson SG, Deeks JJ, Altman DG. Measuring inconsistency in meta-analyses. BMJ. 2003;327(2):557–60.CrossRef Higgins JPT, Thompson SG, Deeks JJ, Altman DG. Measuring inconsistency in meta-analyses. BMJ. 2003;327(2):557–60.CrossRef
22.
Zurück zum Zitat Gomes A, van Geel PP, Santing M, Prakken NHJ, Ruis ML, van Assen, et al. Imaging infective endocarditis: Adherence to a diagnostic flowchart and direct comparison of imaging techniques. J Nucl Cardiol. 2020;27(2):592–608. Gomes A, van Geel PP, Santing M, Prakken NHJ, Ruis ML, van Assen, et al. Imaging infective endocarditis: Adherence to a diagnostic flowchart and direct comparison of imaging techniques. J Nucl Cardiol. 2020;27(2):592–608.
23.
Zurück zum Zitat Dell’aquila AM, Avramovic N, Mastrobuoni S, Motekallemi A, Wisniewski K, Scherer M, et al. Fluorine-18 fluorodeoxyglucose positron emission tomography/computed tomography for improving diagnosis of infection in patients on CF-LVAD: longing for more insights. Eur Heart J Cardiovasc Imaging. 2018;19(5):532–43.CrossRef Dell’aquila AM, Avramovic N, Mastrobuoni S, Motekallemi A, Wisniewski K, Scherer M, et al. Fluorine-18 fluorodeoxyglucose positron emission tomography/computed tomography for improving diagnosis of infection in patients on CF-LVAD: longing for more insights. Eur Heart J Cardiovasc Imaging. 2018;19(5):532–43.CrossRef
24.
Zurück zum Zitat Sommerlath Sohns J, Kroehn H, Schoede A, Derlin T, Haverich A, Schmitto J, et al. 18 F-fluorodeoxyglucose positron emission tomography / computed tomography in left-ventricular assist device infection: initial results supporting the usefulness of image-guided therapy. J Nucl Med. 2019:jnumed.119.237628. Sommerlath Sohns J, Kroehn H, Schoede A, Derlin T, Haverich A, Schmitto J, et al. 18 F-fluorodeoxyglucose positron emission tomography / computed tomography in left-ventricular assist device infection: initial results supporting the usefulness of image-guided therapy. J Nucl Med. 2019:jnumed.119.237628.
25.
Zurück zum Zitat Kim J, Feller ED, Chen W, Liang Y, Dilsizian V. FDG PET/CT for early detection and localization of left ventricular assist device infection: impact on patient management and outcome. JACC Cardiovasc Imaging. 2019;12(4):722–9.CrossRef Kim J, Feller ED, Chen W, Liang Y, Dilsizian V. FDG PET/CT for early detection and localization of left ventricular assist device infection: impact on patient management and outcome. JACC Cardiovasc Imaging. 2019;12(4):722–9.CrossRef
26.
Zurück zum Zitat Akin S, Muslem R, Constantinescu AA, Manintveld OC, Birim O, Brugts JJ, et al. 18F-FDG PET/CT in the diagnosis and management of continuous flow left ventricular assist device infections: a case series and review of the literature. ASAIO J. 2018;64(2):11–9. Akin S, Muslem R, Constantinescu AA, Manintveld OC, Birim O, Brugts JJ, et al. 18F-FDG PET/CT in the diagnosis and management of continuous flow left ventricular assist device infections: a case series and review of the literature. ASAIO J. 2018;64(2):11–9.
27.
Zurück zum Zitat Bernhardt AM, Pamirsad MA, Brand C, Reichart D, Tienken M, Barten MJ, et al. The value of fluorine-18 deoxyglucose positron emission tomography scans in patients with ventricular assist device specific infections. Eur J Cardiothorac Surg. 2017;51(6):1072–7.CrossRef Bernhardt AM, Pamirsad MA, Brand C, Reichart D, Tienken M, Barten MJ, et al. The value of fluorine-18 deoxyglucose positron emission tomography scans in patients with ventricular assist device specific infections. Eur J Cardiothorac Surg. 2017;51(6):1072–7.CrossRef
28.
Zurück zum Zitat Dell’Aquila AM, Mastrobuoni S, Alles S, Wenning C, Henryk W, Schneider SRB, et al. Contributory role of fluorine 18-fluorodeoxyglucose positron emission tomography/ computed tomography in the diagnosis and clinical management of infections in patients supported with a continuous-flow left ventricular assist device. Ann Thorac Surg. 2016;101(1):87–94.CrossRef Dell’Aquila AM, Mastrobuoni S, Alles S, Wenning C, Henryk W, Schneider SRB, et al. Contributory role of fluorine 18-fluorodeoxyglucose positron emission tomography/ computed tomography in the diagnosis and clinical management of infections in patients supported with a continuous-flow left ventricular assist device. Ann Thorac Surg. 2016;101(1):87–94.CrossRef
29.
Zurück zum Zitat De Vaugelade C, Mesguich C, Nubret K, Camou F, Greib C, Dournes G, et al. Infections in patients using ventricular-assist devices: comparison of the diagnostic performance of 18 F-FDG PET/CT scan and leucocyte-labeled scintigraphy. J Nucl Cardiol. 2019;26(1):42–55.CrossRef De Vaugelade C, Mesguich C, Nubret K, Camou F, Greib C, Dournes G, et al. Infections in patients using ventricular-assist devices: comparison of the diagnostic performance of 18 F-FDG PET/CT scan and leucocyte-labeled scintigraphy. J Nucl Cardiol. 2019;26(1):42–55.CrossRef
30.
Zurück zum Zitat Tam MC, Patel VN, Weinberg RL, et al. Diagnostic accuracy of FDG PET/CT in suspected LVAD infections: a case series, systematic review, and meta-analysis. JACC Cardiovasc Imaging. 2020;13(5):1191–202. Tam MC, Patel VN, Weinberg RL, et al. Diagnostic accuracy of FDG PET/CT in suspected LVAD infections: a case series, systematic review, and meta-analysis. JACC Cardiovasc Imaging. 2020;13(5):1191–202.
31.
Zurück zum Zitat Harisankar CNB, Mittal BR, Agrawal KL, Abrar ML, Bhattacharya A. Utility of high fat and low carbohydrate diet in suppressing myocardial FDG uptake. J Nucl Cardiol. 2011;18(5):926–36.CrossRef Harisankar CNB, Mittal BR, Agrawal KL, Abrar ML, Bhattacharya A. Utility of high fat and low carbohydrate diet in suppressing myocardial FDG uptake. J Nucl Cardiol. 2011;18(5):926–36.CrossRef
32.
Zurück zum Zitat Kinahan PE, Hasegawa BH, Beyer T. X-ray-based attenuation correction for positron emission tomography / computed tomography scanners. Semin Nucl Med. 2003;33(3):166–79.CrossRef Kinahan PE, Hasegawa BH, Beyer T. X-ray-based attenuation correction for positron emission tomography / computed tomography scanners. Semin Nucl Med. 2003;33(3):166–79.CrossRef
33.
Zurück zum Zitat Gould FK, Denning DW, Elliott TSJ, Foweraker J, Perry JD, Prendergast BD, et al. Guidelines for the diagnosis and antibiotic treatment of endocarditis in adults: a report of the working party of the British society for antimicrobial chemotherapy. J Antimicrob Chemother. 2012;67(2):269–89.CrossRef Gould FK, Denning DW, Elliott TSJ, Foweraker J, Perry JD, Prendergast BD, et al. Guidelines for the diagnosis and antibiotic treatment of endocarditis in adults: a report of the working party of the British society for antimicrobial chemotherapy. J Antimicrob Chemother. 2012;67(2):269–89.CrossRef
Metadaten
Titel
The value of 18F-FDG PET/CT for the diagnosis of device-related infections in patients with a left ventricular assist device: a systematic review and meta-analysis
verfasst von
D. ten Hove
G. Treglia
R. H. J. A. Slart
K. Damman
M. Wouthuyzen-Bakker
D. F. Postma
O. Gheysens
R. J. H. Borra
G. Mecozzi
P. P. van Geel
B. Sinha
A. W. J. M. Glaudemans
Publikationsdatum
27.06.2020
Verlag
Springer Berlin Heidelberg
Erschienen in
European Journal of Nuclear Medicine and Molecular Imaging / Ausgabe 1/2021
Print ISSN: 1619-7070
Elektronische ISSN: 1619-7089
DOI
https://doi.org/10.1007/s00259-020-04930-8

Weitere Artikel der Ausgabe 1/2021

European Journal of Nuclear Medicine and Molecular Imaging 1/2021 Zur Ausgabe