Tidal volumes during delivery room stabilization of (near) term infants

verfasst von: Janine Thomann, Christoph M. Rüegger, Vincent D. Gaertner, Eoin O’Currain, Omar F. Kamlin, Peter G. Davis, Laila Springer

Erschienen in: BMC Pediatrics | Ausgabe 1/2022

Abstract

Background

We sought to assess tidal volumes in (near) term infants during delivery room stabilization.

Methods

Secondary analysis of a prospective study comparing two facemasks used for positive pressure ventilation (PPV) in newborn infants ≥ 34 weeks gestation. PPV was provided with a T-piece device with a PIP of 30 cmH2O and positive end-expiratory airway pressure of 5 cmH2O. Expired tidal volumes (Vt) were measured with a respiratory function monitor. Target range for Vt was defined to be 4 – 8 ml/kg.

Results

Twenty-three infants with a median (IQR) gestational age of 38.1 (36.4 – 39.0) weeks received 1828 inflations with a median Vt of 4.6 (3.3 – 6.2) ml/kg. Median Vt was in the target range in 12 infants (52%), lower in 9 (39%) and higher in 2 (9%). Thirty-six (25—27) % of the inflations were in the target rage over the duration of PPV while 42 (25 – 65) % and 10 (3 – 33) % were above and below target range.

Conclusions

Variability of expiratory tidal volume delivered to term and late preterm infants was wide. Reliance on standard pressures and clinical signs may be insufficient to provide safe and effective ventilation in the delivery room.

Trial registration

This is a secondary analysis of a prospectively registered randomized controlled trial (ACTRN12616000768493).
Hinweise

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
Vt
Tidal volume
PPV
Positive pressure ventilation
IQR
Inter quartile range
RFM
Respiratory function monitor
PIP
Peak inspiratory pressure

Background

Although most newborn infants initiate respiration spontaneously or after tactile stimulation within the first minutes after birth, approximately 5% require additional support [1]. Current guidelines recommend mask ventilation as the next step to achieve aeration of the lung [1, 2]. While volume-targeted ventilation has become standard of care in the neonatal intensive care unit due to its lung-protective effects [3], tidal volumes (Vt) are not routinely measured during non-invasive positive pressure ventilation (PPV) in the delivery room. Studies in preterm infants suggest an association between excessive Vt and intraventricular hemorrhage [4]. Others suggest that low Vt are associated with a higher intubation rate [5]. A clinical trial comparing the use of a RFM during PPV of preterm infants in the delivery room as a guidance for Vt delivery with no RFM showed that only 30% of inflations were within the target range [6]. In term and late preterm infants, less is known about the variability of measured Vt and changes of Vt over time. A recent single center study reported a median tidal volume at the lower end of the recommended range and substantial variation in tidal volumes measured in term infants [7].
In spontaneously breathing newborn term infants, expired Vt are known to be highly variable, ranging from 0.5 to more than 20 ml/kg [8, 9]. This may be due to changing breathing patterns during the transitional period [10]. During the provision of PPV, expired Vt may also vary chiefly due to interaction between spontaneous breaths and mask inflations.
Current European Resuscitation (ERC) guidelines recommend providing initial peak inflation pressures (PIP) of 30 cm H2O for term infants [1]. These recommendations are based on limited evidence [11]. The primary aim of this study was to assess Vt measured in term and late preterm infants during delivery room stabilization and changes in applied Vt over the duration of PPV.

Methods

Population and intervention

This is a secondary analysis of a previously published randomized controlled trial conducted at the Royal Women’s Hospital in Melbourne, Australia (ACTRN12616000768493) [12]. The original study and secondary analyses were approved by the local ethics committee. All parents provided written informed consent. The original trial compared the effect of two facemasks on mask leak in newborn infants ≥ 34 weeks gestation receiving PPV immediately after birth. PPV was started according to local neonatal resuscitation guidelines if the infant was gasping, not breathing or had a heart rate < 100 min [13]. The original study showed that the suction mask may have negative effects on mask ventilation [12, 14]. Therefore, the current secondary analysis only included infants resuscitated with the conventional mask (Laerdal Silicone mask, Laerdal, Stavanger, Norway).

Measurements and data collection

A Neopuff Infant Resuscitator (Fisher & Paykel Healthcare, Auckland, New Zealand) was used to provide PPV with initial peak inflation pressure (PIP) of 30 cm H2O and positive end-expiratory airway pressure (PEEP) of 5 cm H2O [1]. Changes to pressure settings were at the discretion of the treating clinician. Respiratory function parameters were recorded continuously during PPV using the NewLifeBox (Advanced Life Diagnostics UG, Weener, Germany). Breath-by-breath analysis was performed by manually placing inspiratory and expiratory markers using Pulmochart (Advanced Life Diagnostics UG). As previously published [6], all PPV inflations, including those coinciding with spontaneous breaths, were included. The clinical team was blinded to the RFM data. All resuscitations were video recorded.

Statistical analysis

Data from all available breaths were analysed. Consistent with previous studies, we defined the optimal range for expiratory Vt to be 4 – 8 ml/kg [15]. Leak was calculated as the difference between inspiratory Vt and expiratory Vt, and expressed as a percentage of inspiratory V[16]. Obstruction was defined as Vt < 2 ml/kg with leak < 30% [16]. Variability of Vt was assessed by calculating the coefficient of variation (standard deviation divided by mean). The higher the coefficient of variation, the greater the level of dispersion around the mean. Skewed data are presented as median and interquartile range (IQR). Friedman test was used to compare Vt between different time points throughout episodes of PPV and Mann–Whitney U-test was used to compare parameters between groups. P-values < 0.05 were considered statistically significant. Analyses were performed using SPSS (IBM SPSS Statistics for Windows, Version 26.0. Armonk, NY: IBM Corp).

Results

Population

Overall, 23 infants with a median (IQR) gestational age of 38.1 (36.4 to 39.0) weeks received 1828 PPV inflations (see Table 1). A median of 110 inflations (67 to 181) were recorded per infant.
Table 1
Demographics and respiratory outcomes for n = 23 infants. Numbers are presented as median (IQR) or numbers (%)
Demographics
 Gestational age (weeks)
38.1 (36.4 – 39.0)
 Birth weight (g)
3210 (2810 – 3330)
 Male (%)
13 (57%)
 Cesarean delivery (%)
21 (91%)
 Maternal general anesthesia (%)
7 (30%)
 Umbilical cord pH
7.20 (7.11 – 7.24)
 Apgar 1 min
3 (2 – 4)
 Apgar 5 min
8 (6 – 9)
 Apgar 10 min
9 (8 – 9)
 Meconium stained amniotic fluid
5 (22%)
 Admission to NICU
10 (43%)
 Intubation
2 (9%)
Respiratory ouctomes
 Vt (ml/kg)
4.6 (3.3 – 6.2)
 PIP (mbar)
30.5 (29.9 – 31.8)
 PEEP (mbar)
5.2 (4.2 – 5.8)
 Respiratory rate (bpm)
51 (44 – 56)
 Ti (s)
0.58 (0.53 – 0.66)

Tidal volume

Median (IQR) Vt was 4.6 (3.3 – 6.2) ml/kg. Nine infants (39%) had a median Vt < 4 ml/kg, 12 infants (52%) had a median Vt between 4–8 and 2 infants (9%) a median Vt > 8 ml/kg. There was a high intra- and interindividual variability in measured Vt (Fig. 1). The median (IQR) coefficient of variation was 52 (45 – 77) %. Per infant, a median (IQR) of 36 (25—27) % of inflations were within target range during the duration of PPV, while 42 (25 – 65) % and 10 (3 – 33) % were above and below target range, respectively. Typical RFM waveforms for ventilations within (appropriate ventilations), above (excessive ventilations) and below (inadequate ventilations) the target range are shown in Fig. 2. There were no important changes in Vt over the duration of an episode of PPV (Fig. 3; p = 0.673).

Leak and obstruction

Median (IQR) leak was 30 (11 – 44) %. The median number of inflations with obstruction was 0 (0 – 13). At least one obstructive episode was noted in 10/23 (43%) of the infants. Infants with at least one episode of obstruction had a median (IQR) of 9 (1 – 30) obstructed inflations. The infant with the highest Vt [13.4 (11.7 – 15.5) ml/kg] had a median leak of 3% and no obstruction (patient 11, Fig. 1). The video recording showed one resuscitator applying the mask tightly with two hands while a second operator applied the inflations to an apneic, floppy infant.

Clinical outcomes

Ten of 23 (43%) infants included in the study were admitted to the neonatal intensive care unit (NICU) with the following diagnosis on admission: sepsis (n = 5), respiratory distress syndrome (n = 2), transient tachypnea of the newborn (n = 1), hypoglycemia (n = 1) and hydrocephalus (n = 1). Four of them received non-invasive respiratory support after admission to NICU and two were intubated. Vt and number of PPV inflations per infant in those admitted versus those not admitted were similar [Vt: 4.8 (2.9 – 7.1) ml/kg versus 4.2  (3.4 – 5.9) ml/kg, p = 0.95; number of PPV inflations: 58 (31 – 166) versus 58 (32 – 100), p = 0.88].
Two infants were intubated (patient 17 and 18). There was a non-significant trend towards a lower Vt and a higher number of PPV inflations in those infants who were intubated compared to those who were not [Vt: 2.6 (1.9 – 3.3) ml/kg versus 5.1 (3.4 – 6.4) ml/kg p = 0.07; number of PPV inflations 142 (82 – 201) versus 49 (31 – 103), p = 0.24)]. The two infants that were intubated had a higher proportion of inflations with obstruction [19 (7 – 31) %] in comparison with the infants that were not intubated [0 (0 – 1.5) %], albeit the difference did not reach statistical significance (p = 0.095).
One of the infants who were intubated had respiratory distress syndrome and the other transient tachypnea of the newborn.

Discussion

We found marked variability of Vt measured in term and late preterm infants when using currently recommended pressure levels for PPV during delivery room stabilization. These findings are in line with recently published data in term infants [7]. Our data, however, highlight that not only was there substantial variability between infants, but Vt was inconsistent within the same individual and leak and obstruction were common.
There are several potential causes of the high variability of Vt observed in this study: (1) The presence of leak and obstruction may contribute. In this study median facemask leak was 30% and obstruction occurred in 43% of infants. Leak was similar to that previously reported [7]. The incidence of obstruction in term infants has not been reported before. However, obstruction was reported in 25% of preterm infants [5]. (2) The presence of simultaneous spontaneous breaths either during inflation or during deflation may influence Vt variability [17]. Spontaneous breaths during PPV can be triggered by tactile stimulation during mask ventilation [18].
Only half of the infants had a median Vt within the currently recommended target range of 4–8 ml/kg and Vt was in the target rage in only one third of the inflations over the duration of PPV. The optimal Vt during mask ventilation at birth is unknown and the currently recommended Vt target range of 4–8 mL/kg is largely based on data from endotracheal ventilation [3]. Recent studies suggested that Vt in spontaneously breathing term infants during transition are between 2 – 6.5 ml/kg [8] and 2.5 – 8.5 ml/kg [9]. It is unclear whether the suggested reference range of 4–8 ml/kg is safe and effective for term newborns receiving PPV.
The recommended PIP for PPV in the delivery room is based on limited evidence [1] and it remains unclear whether it is sufficient to achieve adequate tidal volumes and lung aeration. Consistent with other reports, our data show that tidal volumes vary considerably despite a fixed PIP [7]. Vt tended to be lower in infants who were subsequently intubated. We speculate that higher PIP levels may be necessary in selected critically ill infants, however more data are required. On the other hand, one case showed very high Vt under optimal conditions in the absence of leak and obstruction. Based on our data, an individualised approach whereby pressures are adjusted to maintain a safe and effective Vt may be beneficial. A RFM in the delivery room may assist in this regard.
Data from term infants who received 20 PPV inflations with a self-inflating bag without a PEEP valve indicate that higher than recommended PIP levels (36 cmH2O) were necessary to achieve adequate Vt of 3–6 ml/kg [11]. Moreover, there seems to be a positive relationship between heart rate increase and measured Vt in depressed infants needing PPV [19].
Clinical assessment of Vt and identification of leak and obstructions is challenging, and the use of a RFM may improve the effectiveness of mask ventilation [20]. However, a multicenter randomized controlled trial showed that the use of a RFM compared to no RFM as guidance for Vt, did not increase the percentage of inflations in a predefined target range [6]. It is unclear whether improvements in monitor design or education in the use of RFM might lead to a different result.
This study has some limitations. It is an exploratory analysis and outcomes were not prospectively defined. It is a single center study with a small sample size and results may not be generalizable to other units using different equipment. Different definitions for target range for Vt as well as leak and obstruction used in the literature make comparisons with other studies difficult. It would have been interesting to evaluate the effectiveness of the resuscitation in terms of heart rate and oxygen saturations between the groups according to tidal volume, but we were unable to measure these data consistently in our study population.

Conclusion

Tidal volumes delivered to term and late preterm infants in the delivery room vary widely both between and within infants, despite a consistent PIP. Airway obstruction and facemask leak are common, resulting in only one third of inflations having an expired Vt in the recommended range. Finding optimal pressure settings seems challenging and a respiratory function monitor may improve the safety and effectiveness of mask ventilation.

Acknowledgements

Not applicable.

Declarations

This is a secondary analysis of a previously published randomized controlled trial conducted at the Royal Women’s Hospital in Melbourne, Australia. The original study and secondary analyses were approved by the local ethics committee (The Human Research Ethics Committee at The Royal Women's Hospital, Melbourne). All parents provided written informed consent. All methods were performed in accordance with the NHMRC National Statement on Ethical Conduct in Human Research and other relevant federal and state legislation and regulations.
Not applicable.

Competing interests

The authors declare that they have no competing interests.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Madar J, Roehr CC, Ainsworth S, Ersdal H, Morley C, Rudiger M, et al. European resuscitation council guidelines 2021: newborn resuscitation and support of transition of infants at birth. Resuscitation. 2021;161:291–326.CrossRef Madar J, Roehr CC, Ainsworth S, Ersdal H, Morley C, Rudiger M, et al. European resuscitation council guidelines 2021: newborn resuscitation and support of transition of infants at birth. Resuscitation. 2021;161:291–326.CrossRef
2.
Zurück zum Zitat Perlman JM, Wyllie J, Kattwinkel J, Wyckoff MH, Aziz K, Guinsburg R, et al. Part 7: Neonatal resuscitation: 2015 international consensus on cardiopulmonary resuscitation and emergency cardiovascular care science with treatment recommendations. Circulation. 2015;132(16 Suppl 1):S204–41.PubMed Perlman JM, Wyllie J, Kattwinkel J, Wyckoff MH, Aziz K, Guinsburg R, et al. Part 7: Neonatal resuscitation: 2015 international consensus on cardiopulmonary resuscitation and emergency cardiovascular care science with treatment recommendations. Circulation. 2015;132(16 Suppl 1):S204–41.PubMed
3.
Zurück zum Zitat Klingenberg C, Wheeler KI, McCallion N, Morley CJ, Davis PG. Volume-targeted versus pressure-limited ventilation in neonates. Cochrane Database Syst Rev. 2017;10:CD003666.PubMed Klingenberg C, Wheeler KI, McCallion N, Morley CJ, Davis PG. Volume-targeted versus pressure-limited ventilation in neonates. Cochrane Database Syst Rev. 2017;10:CD003666.PubMed
4.
Zurück zum Zitat Mian Q, Cheung PY, O’Reilly M, Barton SK, Polglase GR, Schmölzer GM. Impact of delivered tidal volume on the occurrence of intraventricular haemorrhage in preterm infants during positive pressure ventilation in the delivery room. Arch Dis Child Fetal Neonatal Ed. 2019;104(1):F57–62.CrossRef Mian Q, Cheung PY, O’Reilly M, Barton SK, Polglase GR, Schmölzer GM. Impact of delivered tidal volume on the occurrence of intraventricular haemorrhage in preterm infants during positive pressure ventilation in the delivery room. Arch Dis Child Fetal Neonatal Ed. 2019;104(1):F57–62.CrossRef
5.
Zurück zum Zitat Schmolzer GM, Dawson JA, Kamlin CO, O’Donnell CP, Morley CJ, Davis PG. Airway obstruction and gas leak during mask ventilation of preterm infants in the delivery room. Arch Dis Child Fetal Neonatal Ed. 2011;96(4):F254–7.CrossRef Schmolzer GM, Dawson JA, Kamlin CO, O’Donnell CP, Morley CJ, Davis PG. Airway obstruction and gas leak during mask ventilation of preterm infants in the delivery room. Arch Dis Child Fetal Neonatal Ed. 2011;96(4):F254–7.CrossRef
6.
Zurück zum Zitat van Zanten HA, Kuypers K, van Zwet EW, van Vonderen JJ, Kamlin COF, Springer L, et al. A multi-centre randomised controlled trial of respiratory function monitoring during stabilisation of very preterm infants at birth. Resuscitation. 2021;167:317–25.CrossRef van Zanten HA, Kuypers K, van Zwet EW, van Vonderen JJ, Kamlin COF, Springer L, et al. A multi-centre randomised controlled trial of respiratory function monitoring during stabilisation of very preterm infants at birth. Resuscitation. 2021;167:317–25.CrossRef
7.
Zurück zum Zitat Bjorland PA, Ersdal HL, Haynes J, Ushakova A, Oymar K, Rettedal SI. Tidal volumes and pressures delivered by the NeoPuff T-piece resuscitator during resuscitation of term newborns. Resuscitation. 2022;170:222–9.CrossRef Bjorland PA, Ersdal HL, Haynes J, Ushakova A, Oymar K, Rettedal SI. Tidal volumes and pressures delivered by the NeoPuff T-piece resuscitator during resuscitation of term newborns. Resuscitation. 2022;170:222–9.CrossRef
8.
Zurück zum Zitat Baixauli-Alacreu S, Padilla-Sanchez C, Hervas-Marin D, Lara-Canton I, Solaz-Garcia A, Alemany-Anchel MJ, et al. Expired tidal volume and respiratory rate during postnatal stabilization of newborn infants born at term via cesarean delivery. J Pediatr. 2021;6:100063. Baixauli-Alacreu S, Padilla-Sanchez C, Hervas-Marin D, Lara-Canton I, Solaz-Garcia A, Alemany-Anchel MJ, et al. Expired tidal volume and respiratory rate during postnatal stabilization of newborn infants born at term via cesarean delivery. J Pediatr. 2021;6:100063.
9.
Zurück zum Zitat Blank DA, Gaertner VD, Kamlin COF, Nyland K, Eckard NO, Dawson JA, et al. Respiratory changes in term infants immediately after birth. Resuscitation. 2018;130:105–10.CrossRef Blank DA, Gaertner VD, Kamlin COF, Nyland K, Eckard NO, Dawson JA, et al. Respiratory changes in term infants immediately after birth. Resuscitation. 2018;130:105–10.CrossRef
10.
Zurück zum Zitat te Pas AB, Wong C, Kamlin CO, Dawson JA, Morley CJ, Davis PG. Breathing patterns in preterm and term infants immediately after birth. Pediatr Res. 2009;65(3):352–6.CrossRef te Pas AB, Wong C, Kamlin CO, Dawson JA, Morley CJ, Davis PG. Breathing patterns in preterm and term infants immediately after birth. Pediatr Res. 2009;65(3):352–6.CrossRef
11.
Zurück zum Zitat Ersdal HL, Eilevstjonn J, Perlman J, Gomo O, Moshiro R, Mdoe P, et al. Establishment of functional residual capacity at birth: observational study of 821 neonatal resuscitations. Resuscitation. 2020;153:71–8.CrossRef Ersdal HL, Eilevstjonn J, Perlman J, Gomo O, Moshiro R, Mdoe P, et al. Establishment of functional residual capacity at birth: observational study of 821 neonatal resuscitations. Resuscitation. 2020;153:71–8.CrossRef
12.
Zurück zum Zitat Lorenz L, Rüegger CM, O’Currain E, Dawson JA, Thio M, Owen LS, et al. Suction mask vs conventional mask ventilation in term and near-term infants in the delivery room: a randomized controlled trial. J Pediatr. 2018;198:181-6.e2.CrossRef Lorenz L, Rüegger CM, O’Currain E, Dawson JA, Thio M, Owen LS, et al. Suction mask vs conventional mask ventilation in term and near-term infants in the delivery room: a randomized controlled trial. J Pediatr. 2018;198:181-6.e2.CrossRef
13.
Zurück zum Zitat Liley HG, Mildenhall L, Morley P, Australian New Zealand Committee on R. Australian and New Zealand committee on resuscitation neonatal resuscitation guidelines 2016. J Paediatr Child Health. 2017;53(7):621–7.CrossRef Liley HG, Mildenhall L, Morley P, Australian New Zealand Committee on R. Australian and New Zealand committee on resuscitation neonatal resuscitation guidelines 2016. J Paediatr Child Health. 2017;53(7):621–7.CrossRef
14.
Zurück zum Zitat Ruegger CM, O’Currain E, Dawson JA, Davis PG, Kamlin COF, Lorenz L. Compromised pressure and flow during suction mask ventilation. Arch Dis Child Fetal Neonatal Ed. 2019;104(6):F662–3.CrossRef Ruegger CM, O’Currain E, Dawson JA, Davis PG, Kamlin COF, Lorenz L. Compromised pressure and flow during suction mask ventilation. Arch Dis Child Fetal Neonatal Ed. 2019;104(6):F662–3.CrossRef
15.
Zurück zum Zitat Yang KC, Te Pas AB, Weinberg DD, Foglia EE. Corrective steps to enhance ventilation in the delivery room. Arch Dis Child Fetal Neonatal Ed. 2020;105(6):605–8.CrossRef Yang KC, Te Pas AB, Weinberg DD, Foglia EE. Corrective steps to enhance ventilation in the delivery room. Arch Dis Child Fetal Neonatal Ed. 2020;105(6):605–8.CrossRef
16.
Zurück zum Zitat Kamlin COF, Schmölzer GM, Dawson JA, McGrory L, O’Shea J, Donath SM, et al. A randomized trial of oropharyngeal airways to assist stabilization of preterm infants in the delivery room. Resuscitation. 2019;144:106–14.CrossRef Kamlin COF, Schmölzer GM, Dawson JA, McGrory L, O’Shea J, Donath SM, et al. A randomized trial of oropharyngeal airways to assist stabilization of preterm infants in the delivery room. Resuscitation. 2019;144:106–14.CrossRef
17.
Zurück zum Zitat Schilleman K, van der Pot CJ, Hooper SB, Lopriore E, Walther FJ, te Pas AB. Evaluating manual inflations and breathing during mask ventilation in preterm infants at birth. J Pediatr. 2013;162(3):457–63.CrossRef Schilleman K, van der Pot CJ, Hooper SB, Lopriore E, Walther FJ, te Pas AB. Evaluating manual inflations and breathing during mask ventilation in preterm infants at birth. J Pediatr. 2013;162(3):457–63.CrossRef
18.
Zurück zum Zitat Gaertner VD, Ruegger CM, Bassler D, O’Currain E, Kamlin COF, Hooper SB, et al. Effects of tactile stimulation on spontaneous breathing during face mask ventilation. Arch Dis Child Fetal Neonatal Ed. 2021;107(5):508–12.CrossRef Gaertner VD, Ruegger CM, Bassler D, O’Currain E, Kamlin COF, Hooper SB, et al. Effects of tactile stimulation on spontaneous breathing during face mask ventilation. Arch Dis Child Fetal Neonatal Ed. 2021;107(5):508–12.CrossRef
19.
Zurück zum Zitat Linde JE, Schulz J, Perlman JM, Oymar K, Blacy L, Kidanto H, et al. The relation between given volume and heart rate during newborn resuscitation. Resuscitation. 2017;117:80–6.CrossRef Linde JE, Schulz J, Perlman JM, Oymar K, Blacy L, Kidanto H, et al. The relation between given volume and heart rate during newborn resuscitation. Resuscitation. 2017;117:80–6.CrossRef
20.
Zurück zum Zitat O’Currain E, Thio M, Dawson JA, Donath SM, Davis PG. Respiratory monitors to teach newborn facemask ventilation: a randomised trial. Arch Dis Child Fetal Neonatal Ed. 2019;104(6):F582–6.CrossRef O’Currain E, Thio M, Dawson JA, Donath SM, Davis PG. Respiratory monitors to teach newborn facemask ventilation: a randomised trial. Arch Dis Child Fetal Neonatal Ed. 2019;104(6):F582–6.CrossRef
Metadaten
Titel
Tidal volumes during delivery room stabilization of (near) term infants
verfasst von
Janine Thomann
Christoph M. Rüegger
Vincent D. Gaertner
Eoin O’Currain
Omar F. Kamlin
Peter G. Davis
Laila Springer
Publikationsdatum
01.12.2022
Verlag
BioMed Central
Erschienen in
BMC Pediatrics / Ausgabe 1/2022
Elektronische ISSN: 1471-2431
DOI
https://doi.org/10.1186/s12887-022-03600-y

Weitere Artikel der Ausgabe 1/2022

BMC Pediatrics 1/2022 Zur Ausgabe

Embryotransfer erhöht womöglich Leukämierisiko der Kinder

13.05.2024 Assistierte Reproduktion Nachrichten

Reproduktionsmedizinische Techniken haben theoretisch das Potenzial, den epigenetischen Code zu verändern und somit das Krebsrisiko der Kinder zu erhöhen. Zwischen Embryotransfer und Leukämie scheint sich ein solcher Zusammenhang bestätigt zu haben.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Häufigste Gründe für Brustschmerzen bei Kindern

06.05.2024 Pädiatrische Diagnostik Nachrichten

Akute Brustschmerzen sind ein Alarmsymptom par exellence, schließlich sind manche Auslöser lebensbedrohlich. Auch Kinder klagen oft über Schmerzen in der Brust. Ein Studienteam ist den Ursachen nachgegangen.

Update Pädiatrie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.