Skip to main content
Erschienen in: BMC Musculoskeletal Disorders 1/2019

Open Access 01.12.2019 | Research article

Total knee arthroplasty using patient-specific instrumentation for osteoarthritis of the knee: a meta-analysis

verfasst von: Kazuha Kizaki, Ajaykumar Shanmugaraj, Fumiharu Yamashita, Nicole Simunovic, Andrew Duong, Vickas Khanna, Olufemi R. Ayeni

Erschienen in: BMC Musculoskeletal Disorders | Ausgabe 1/2019

Abstract

Background

Total knee arthroplasty using patient-specific instrumentation (TKA-PSI), which are disposable cutting block guides generated to fit each patient’s 3-dimensional knee anatomy, has been developed to treat patients with end-stage osteoarthritis of the knee. Surrogate markers such as radiographic malalignment have been well investigated, however, patient-important outcomes are not well examined to elucidate the efficacy of TKA-PSI. The aim of this review is to determine if TKA-PSI improves patient-reported outcome measures (PROM), surgery time, blood loss, transfusion and complications (e.g. surgical site infection, deep venous thrombosis, and revision TKA).

Methods

We searched the Cochrane Central Register of Controlled Trials (CENTRAL), MEDLINE, EMBASE, and ongoing clinical trials. For PROMs, surgery time, blood loss, and transfusion rate, we included randomized controlled trials (RCT) comparing TKA-PSI and standard TKA to treat osteoarthritis of the knee. For complications, we also included non-randomized comparative studies (non-RCT).

Results

This review includes 38 studies, 24 of which were RCT and 14 of which were non-RCT. These included a total of 3487 patients. The predominant population in the included studies highly reflected the general population, with 62% being female, aged over 60 and having end-stage osteoarthritis of the knee. TKA-PSI did not improve PROMs as compared to standard TKA for less than 1-year (mean difference 0.48, 95% confidence interval (CI) -1.92–0.97 in the Oxford knee score, mean 3-month follow-up) and for 1-year or more (mean difference 0.25, 95%CI − 4.39–4.89 in the WOMAC score, mean 29-month follow-up). TKA-PSI did not reduce surgery time (mean difference − 3.09 min, 95%CI -6.73–0.55). TKA-PSI decreased blood loss with a small effect size corresponding to a 0.4 g/dl hemoglobin decrease (95%CI 0.18–0.88), but did not decrease transfusion rate (risk difference − 0.04, 95%CI -0.09–0.01). TKA-PSI did not reduce complication rates (risk difference 0.00, 95%CI − 0.01–0.01 in the composite outcome).

Conclusions

TKA-PSI does not improve patient-reported outcome measures, surgery time, and complication rates as compared to standard TKA. TKA-PSI decreases blood loss with a small effect, which is not enough to reduce transfusion rate.
Begleitmaterial
Additional file 1: Appendix A. PICO format research question. Appendix B.1. searching in Embase (1974 to February 15th, 2019). Appendix B.2. searching in Medline (1946 to February 15th, 2019). Appendix B.3. searching in CENTRAL (The Cochrane Library 2019, Issue 2). Appendix C. Excluded studies with special reasons. Appendix D. Unpublished ongoing studies. Appendix E. MINORS scores in non-RCTs. Appendix F.1. Forest plot in surgical site infection (RCTs and non-RCTs). Appendix F.2. Forest plot in surgical site infection (RCTs). Appendix F.2. Forest plot in surgical site infection (RCTs). Appendix G.2. Forest plot in DVT (RCTs). Appendix G.2. Forest plot in DVT (RCTs). Appendix H.2. Forest plot in revision TKA (RCTs). Appendix I.1. Forest plot in HKA. Appendix I.2. Forest plot in mechanical axis. Appendix I.3. Forest plot in femoral coronal alignment. Appendix I.4. Forest plot in femoral sagittal alignment. Appendix I.5. Forest plot in femoral rotational alignment. Appendix I.6. Forest plot in tibial coronal alignment. Appendix I.7. Forest plot in tibial sagittal alignment. Appendix J. Funnel plots. Appendix K.1. Assessment of the quality of the evidence for Oxford (less than 1-year). Appendix K.2. Assessment of the quality of the evidence for WOMAC (1-year or more). Appendix K.3. Assessment of the quality of the evidence for Surgery time. Appendix K.4. Assessment of the quality of the evidence for Blood loss. Appendix K.5. Assessment of the quality of the evidence for transfusion rate. Appendix K.6. Assessment of the quality of the evidence for complications.
Hinweise

Supplementary information

Supplementary information accompanies this paper at https://​doi.​org/​10.​1186/​s12891-019-2940-2.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
ADL
Activities of daily living
CENTRAL
Cochrane Central Register of Controlled Trials
CI
Confidence interval
CT
Computed tomography
DVT
Deep venous thrombosis
EQ-5D VAS
European quality of life 5-dimensions using the visual analogue scale
GRADE
Grading of Recommendations Assessment, Development and Evaluation
HKA
Hip-knee-ankle angle
KOOS
Knee injury and Osteoarthritis Outcome Score
KSS
Knee Society Score
MCID
Minimally clinically important differences
MD
Mean difference
MINORS
Methodological index for non-randomized studies
MRI
Magnetic resonance imaging
OA
Osteoarthritis
Oxford
Oxford Knee Score
PRISMA
Preferred reporting items for systematic reviews and meta-analyses
PROM
Patient-reported outcome measure
PSI
Patient-specific instrumentation
QoL
Quality of life
RCT
Randomized controlled trial
RR
Risk ratio
SF-12 mental score
12-item short form health survey mental score
SF-12 physical score
12-item short form health survey physical score
SMD
Standardized mean difference
SR
Systematic review
SSI
Surgical site infection
TKA
Total knee arthroplasty
VAS
Visual analogue scale
WOMAC
Western Ontario and McMaster Universities Osteoarthritis Index

Background

Globally, the proportion of the aging population has been dramatically increasing [1]. Osteoarthritis (OA) of the knee is the most common joint disorder among people aged over 60. Among this population, 10% of males and 13% of females have symptomatic knee OA [2]. Total knee arthroplasty (TKA) is the most common surgical option for end-stage knee OA with knee deformity and persistent pain [3]. In the last 20 years (1993–2012), a total of 7.8 million primary TKA procedures were performed in the United States [4]. The number of TKA procedures continues to increase and it is expected to increase 69% between 2012 and 2050 [4]. Currently, TKA using intramedullary and extramedullary alignment systems with cutting guides is the standard of care [5]. During the procedure, numerous surgical devices are utilized to ensure success. However, this creates a complicated workflow and prolonged surgery time [6]. Furthermore, it requires positioning an intramedullary nail in the femoral canal that increases invasiveness.
A new innovative surgical technique using patient-specific instrumentation (PSI) for performing TKA has been developed to reduce the technical difficulties and invasiveness associated with standard TKA. PSI is also called “patient-matched instrumentation,” “custom-fit instrumentation,” or “custom-made instrumentation” [79]. For TKA using PSI, disposable cutting blocks are generated to fit each patient’s 3-dimensional anatomy of the knee in reference to the preoperative computed tomography (CT) or magnetic resonance imaging (MRI) images combined with radiographs of the lower extremity [10]. The cutting blocks are made individually for each patient and they enable the surgeon to develop a surgical plan specific to each patient. The cutting blocks fit on the distal femur and proximal tibia and guide surgeons to cut them accurately [10].
The TKA procedure using PSI decreases rates of lower limb malalignment [11, 12] and it is expected to improve functional outcomes and decrease revision rates [1315]. In the procedure, PSI does not require positioning intramedullary nails in the femur and is expected to reduce blood loss and transfusion rates. Also, the simpler workflow owing to the cutting blocks potentially reduces surgical time. Long surgical time is one of the important risk factors for postoperative surgical site infection (SSI) [16] and deep venous thrombosis (DVT) [17] related to TKA. Thus, PSI is expected to reduce postoperative SSI and DVT.
Previous systematic reviews (SR) reported that TKA using PSI reduces surgical time, blood loss, and rates of lower limb malalignment as compared to standard TKA [12, 1826]. However, the primary outcomes are surrogate markers and the SRs did not examine whether PSI would contribute to decreasing rates of transfusion, SSI, DVT, and revision TKA as compared to standard TKA. Also, the previous SRs addressed inconsistent results for patient-reported outcome measures (PROMs) with a limited number of RCTs. The aim of this review is to assess the effects of TKA using PSI for patients with osteoarthritis of the knee as compared to standard TKA. The complete PICO-format research questions is shown in Additional file 1: Appendix A.

Methods

Criteria for including studies

We included studies which compared TKA using PSI (TKA-PSI) and standard TKA in this review. Standard TKA was defined as TKA using intramedullary alignment guiding nail for cutting femur and either intramedullary or extramedullary alignment guiding nail for cutting tibia. Studies were included if patients had primary TKA for knee OA classified into grade 3 and 4, according to Kellgren-Lawrence grading system [27]. If patients had rheumatic diseases or if less than 80% of the study’s population had OA, the studies were excluded. Since rheumatic diseases are systematic diseases, controlling for these conditions using medications directly influences clinical outcomes. Types of primary outcome measures are shown below:
  • PROMs such as Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC), Oxford knee score (Oxford), Knee Society Score (KSS), and the Knee injury and Osteoarthritis Outcome Score (KOOS), European quality of life 5-dimensions using the visual analogue scale (EQ-5D VAS), 12-item short form health survey physical score (SF-12 physical score), 12-item short form health survey mental score (SF-12 mental score)
  • Transfusion rate
  • Blood loss
  • Surgery time
  • Complications (i.e. SSI, DVT, and revision TKA)
As a secondary outcome measure, we investigated the percentage of alignment outliers from the planned alignment in the included studies in which PROMs were examined.
For patient-reported outcome measures (PROM), surgery time, blood loss, and transfusion rate, we included randomized controlled trials (RCT). For complications (i.e. SSI, DVT, revision TKA), we also included non-randomized comparative studies (non-RCT).
We conducted a comprehensive literature search for all relevant articles using four electronic databases: the Cochrane Central Register of Controlled Trials (CENTRAL) (The Cochrane Library 2019, Issue 2), MEDLINE (1946 to February 15th, 2019), and EMBASE (1974 to February 15th, 2019). We also examined ongoing trials using the database of clinical trials (https://​clinicaltrials.​gov). Each searching strategies are shown in the Additional file 1: Appendix B. In hand-searching, we screened the reference list of previous SRs and relevant studies for additional articles potentially not identified through electronic search. We chose the key search terms “knee*,” “arthroplasty OR replacement,” and “patient-specific OR patient-matched OR custom-fit OR custom-made OR custom*.” The search was limited to articles published since 2001, because PSI was initially used in 2001 in institutional studies only, and the first report using PSI was published in 2004 [28].
We adopted a 3-step screening process (title screening, abstract screening, and full-text screening) to select eligible articles. After duplicate articles were removed, two reviewers (KK AND AS) independently performed title and abstract screenings. If either of the reviewers included an article during title or abstract screening, it was moved to the next stage for screening. During full-text screening, discrepancies were resolved through discussion and consensus with the senior authors (FY AND OA). We did not register this protocol in time before data collection was performed. For each study, two reviewers independently extracted data into a spreadsheet for the outcomes designed a priori. Differences were resolved by discussion.
Two reviewers independently assessed risk of bias for RCTs using the Cochrane risk of bias tool [29]. The following domains were assessed: sequence generation; allocation concealment; blinding of participants and personnel (performance bias); blinding of outcome assessment (separately for PROM, transfusion, blood loss, surgery time, and complications); incomplete outcome data (attrition bias); and selective reporting (reporting bias). For non-RCTs included for complications, we used the methodological index for non-randomized studies (MINORS) appraisal tool [30].
Dichotomous outcomes (transfusion and complications) were expressed as a risk ratio (RR) with 95% confidence intervals (CI). Continuous outcomes (i.e. PROM, blood loss, and surgery time) were expressed as a mean difference (MD) between TKA-PSI and standard TKA groups. We preferred to calculate effect size measures (standardized mean difference (SMD)), when studies used different instruments to assess the same outcome (e.g. blood loss).
We analyzed outcomes according to the modified intention-to-treat method without imputation. When data was not expressed with mean and standard deviation (SD), but expressed with median, minimum-maximum range, or interquartile range, we estimated the mean and SD in reference to Wan et al. (2014) [31].
Heterogeneity between pooled studies were assessed using the chi-square test with statistical significance set at p < 0.10 and the I2 statistic [32]. The I2 value was assessed as follows: 0–40% might not be important, 30–60% moderate, 50–90% substantial, and 75–100% considerable. Also, we considered variance of the point estimate and overlap in the confidence intervals.
For assessing reporting biases, we constructed funnel plots for each outcome for which there were at least five trials. We pooled included studies using the generic inverse variance method for continuous outcomes and the Mantel-Haenszel method for dichotomous outcomes in Review Manager 5 (Revman). Either one of fixed-effect or random- effect model was used depending on the heterogeneity. We created a ‘Summary of findings’ for the main comparison. For PROM, we selected an outcome measure in which most patients were pooled each for less than 1-year and for 1-year or more. We pre-specified and carried out sensitivity analyses by excluding non-RCTs in complications. We assessed the quality of evidence related to each outcome measure using the Grading of Recommendations Assessment, Development and Evaluation (GRADE) approach [33].

Results

We searched from February 15th 2019 to March 15th 2019. We screened a total of 1386 articles (CENTRAL 128, EMBASE 651, MEDLINE 607). After the removal of duplicates, we checked 913 articles in title-screening and 320 articles in abstract-screening processes. We added 13 articles from the references in the pooled articles and implemented full-text screening process. From the database of clinical trials, we included 1 study. Overall, 38 articles were included in this systematic review: 38 studies were included in qualitative synthesis and 37 studies were included in quantitative synthesis (meta-analysis), as seen in Fig. 1. Details of the process of screening are illustrated in PRISMA (Preferred reporting items for systematic reviews and meta-analyses) flowchart in Fig. 1.

Included studies

We included 38 studies, with a total of 3487 patients (1753 in TKA-PSI, and 1734 in standard TKA) [7, 3471]. All studies except NCT02539992 were published in 2012–2018. We included 15 studies in PROM (14 studies in meta-analysis), 18 studies in surgery time (18 studies in meta-analysis), 15 studies in blood loss (15 studies in meta-analysis), 9 studies in transfusion rate (8 studies in meta-analysis), and 24 studies in complications (23 studies in meta-analysis). Boonen et al. (2013) and Boonen et al. (2016) were the same cohort and we counted them as the same study. Details of the included studies are provided in the characteristics of included studies table (Table 1).
Table 1
Characteristics of included studies table
Author
Year
Country
Study design
Patients
%females
Age
Follow-up months
PSI device
CT/MRI
Eligibility criteria
PSI
ST
PSI
ST
PSI
ST
Noble Jr. [7]
2012
USA
RCT
15
14
47
57
65.4(57–76)
68.0(56–80)
NR
Visionaire
MRI
arthritis
Pietsch [34]
2013
Austria
RCT
40
40
68
53
71.4(6.6)
69.2(9.4)
3 months
Materialise
MRI
OA
Vundelinckx [35]
2013
Belgium
RCT
31
31
52
65
64.7(8.2)
68.2(8.5)
6 months
Visionaire
MRI
OA
Boonen [36]
2013
Netherlands
RCT
90
90
62
56
69.0(8.0)
65.0(8.8)
44 months
Signature
MRI
OA
Chareancholvanich [37]
2013
Thailand
RCT
40
40
85
90
69.5(55–84)
70.3(53–85)
NR
Materialise
MRI
OA
Hamilton [38]
2013
USA
RCT
26
26
46
73
68.1(52–86)
67.6(51–88)
NR
TruMatch
CT
OA
Roh [39]
2013
Korea
RCT
42
48
93
90
70.0(7.2)
70.0(5.1)
NR
Signature
CT
OA
Ng [40]
2014
USA
pros
51
27
67
52
65.6(7.5)
79.3(8.1)
3 months
Materialise
MRI
OA
Renson [41]
2014
Belgium
pros
71
60
62
52
68.2(8.9)
70.3(8.7)
6 months
Materialise
MRI
OA
Dossett [42]
2014
USA
RCT
44
44
7
14
66.0(7.7)
66.0(8.6)
24 months
Vanguard
MRI
arthritis
Abdel [43]
2014
France
RCT
20
20
60
60
71.0(61–81)
71(55–83)
3 months
Materialise
MRI
arthritis
Chotanaphuti [44]
2014
Thailand
RCT
40
40
88
88
69.7(5.5)
69.3(5.5)
1.5 months
TruMatch
CT
arthritis
Pfitzner [45]
2014
USA
RCT
60
30
57
57
64.0(54–74)
64.0(54–74)
3 months
Visionaire
MRI
OA
Woolson [46]
2014
USA
RCT
22
26
0
0
NR
NR
9 months
TruMatch
CT
no
Nabavi [47]
2015
Australia
retro
82
84
52
52
64(44–85)
65(45–88)
12 months
MyKnee
CT
no
Thienpont [48]
2015
Belgium
retro
75
75
67
67
65.6(9.6)
67.8(11.0)
1.5 months
Visionaire
NR
OA
Yan [49]
2015
China
RCT
30
30
57
80
67.5(8.0)
69.5(8.4)
3 months
Materialise
MRI
OA*
Chen [50]
2015
Singapore
pros
29
30
69
83
65.0(8.0)
65.0(8.0)
24 months
Materialise
MRI
OA
Kotela [51]
2015
Poland
RCT
49
46
67
72
66.1(8.4)
68.6(9.9)
12 months
Signature
CT
OA
Rathod [52]
2015
USA
retro
15
14
60
57
57.0(4.5)
59.0(6.5)
3 months
Visionaire
MRI
OA
Abane [53]
2015
France
RCT
70
70
43
39
67.8(47–84)
70.4(54–83)
3 months
Visionaire
MRI
OA
Molicnik [54]
2015
Slovenia
RCT
19
19
89
74
67.1(7.1)
66.8(6.7)
NR
Materialise
MRI
OA
Ferrara [55]
2015
Italy
RCT
15
15
60
53
75.3(6.7)
74.5(7.2)
NR
Materialise
MRI
OA
Anderl [56]
2016
Austria
pros
114
108
64
58
68.7(8.2)
67.7(9.6)
24 months
MyKnee
CT
OA
Boonen [57]
2016
Netherlands
RCT
90
90
62
56
69.0(8.0)
65.0(8.8)
44 months
Signature
MRI
OA
Huijbregts [58]
2016
Australia
RCT
69
64
58
50
66.7 (9.1)
69.0(9.6)
12 months
Visionaire
MRI
arthritis
Pourgiezis [59]
2016
Australia
pros
45
45
80
56
69.5(1.5)
69.3(1.5)
3 months
Visionaire
MRI
OA
White [60]
2016
USA
pros
21
42
67
67
59.1(7.4)
59.8(6.7)
24 months
iTotal
CT
OA
Culler [61]
2017
USA
retro
126
122
62
64
69.7(8.4)
68.3(9.5)
3 months
NR
CT
arthritis
Kwon [62]
2017
Korea
retro
48
50
96
94
73.0(4.3)
72.0(6.5)
3 months
Signature
MRI
OA
Zhu [63]
2017
Singapore
pros
42
48
71
77
69.3(7.2)
66.8(5.9)
24 months
TruMatch
CT
OA
Vide [64]
2017
Portugal
RCT
47
48
68
69
67.8(8.4)
69.3(6.5)
NR
Visionaire
MRI
OA
Kosse [65]
2018
Netherlands
RCT
21
21
62
43
62.7(4.5)
63.4(4.2)
12 months
Visionaire
MRI
OA
Steimle [66]
2018
USA
retro
31
49
NR
NR
60.8(8.5)
61.0(9.9)
NR
Materialise
MRI
OA
Tammachote [67]
2018
Thailand
RCT
54
54
78
72
72.0(7.0)
72.0(8.0)
24 months
Visionaire
MRI
arthritis
Maus [68]
2018
Germany
RCT
59
66
56
65
68.1(8.5)
71.5(8.1)
3 months
Imprint
MRI
OA
Stolarczyk [69]
2018
Poland
RCT
30
30
73
60
70.2(5.9)
69.6(7.1)
3 months
Visionaire
MRI
OA
Leeuwen [70]
2018
Norway
RCT
44
50
68
64
67.0(8.8)
64.0(6.9)
24 months
Signature
MRI
OA
NCT02539992 [71]
NR
Italy
RCT
26
18
38
44
NR
NR
12 months
Triathlon
MRI
OA
Date was expressed with mean (standard deviation). The underlined numbers represent minimum-maximum range. Abbreviations: PSI Patient-specific instrumentation, ST Standard TKA, NR Not reported, RCT Randomized controlled trial, pros Prospective non-RCT, retro Retrospective non-RCT, OA Osteoarthritis, USA The United States of America. Yan et al. (2015) recruited 93% OA and 6.7% RA
In the included studies, 29 studies with pooled 2547 patients explicitly included patients with osteoarthritis, whereas 7 studies with pooled 726 patients recruited patients with knee deformity. Twenty-seven studies used MRI-based PSI, whereas 10 studies used CT-based PSI. Many varieties of PSI device were used and the most prominent device was Visionnaire (Smith & Nephew, Memphis, USA) in 15 studies followed by Materialise (Zimmer via Materialise, Belgium) in 10 studies. Thirteen studies with the pooled 1384 patients followed for 1-year or more.

Excluded studies

In full-text screening process, we excluded 113 studies. We described studies in which discrepancy for agreement was found or excluded with special reasons, being listed in Additional file 1: Appendix C. We found 18 ongoing trials using the database of clinicaltrials.gov, and we searched their results using corresponding author’s name and institution where trials were conducted. A list of ongoing trials was shown in Additional file 1: Appendix D.

Risk of bias in included studies

We assessed the risk of bias in the included studies. In sequence generation, 8 out of 25 RCTs properly generated unpredictable randomization lists, creating low risk. In allocation concealment, 9 studies appropriately distanced from recruiters notifying allocation, whereas two studies were exposed to high risk of selection bias by recruiting participants in turns between PSI and standard TKA groups [35], or using block randomization without blinding block size for recruiters [65]. PSI requires preoperative MRI or CT to create patient-specific cutting blocks, and we assumed patients were notified as to whether they were allocated to the TKA-PSI group. For that reason, in almost all RCTs except for two studies, there was a high risk of bias in the domain of blinding of outcome assessment for PROM. In the two studies, all patients had preoperative MRI or CT and blinding of patients was robustly maintained with low risk of bias [42, 70]. Also, due to the trait of surgical trials, surgeons and scrub nurses could not be blinded. The unblinding potentially caused performance bias for surgery time, blood loss, and transfusion rate with high risk of bias. In incomplete outcome data, the following outcomes (i.e. surgery time, blood loss, transfusion rate) were less likely influenced, since these were perioperatively recorded in medical charts during admission. Six studies in PROM had high risk of attrition bias [42, 51, 53, 58, 68, 71]. In the included studies for the percentage of outliers in the positioning of the prosthetic implants, five studies clearly performed blinding to the outcome assessors for the images, having low risk of outcome detection bias [45, 53, 57, 58, 67].
In non-RCTs, the mean MINORS scores were 16.8 (SD 3.9). Seven studies were prospective non-RCTs, whereas 6 studies were retrospective non-RCTs. In only one study, the main study purpose was measuring complications. We presented the MINORS scores in the included non-RCTs in Additional file 1: Appendix E.

Patient-reported outcome measures

The following outcome measures included two or more studies: KSS knee, KSS function, KSS total, Oxford, WOMAC, KOOS symptom, KOOS pain, KOOS ADL, KOOS sports, KOOS QoL, EQ-5D VAS, SF-12 physical score, and SF-12 mental score.
In patients followed for less than 1-year, 17 studies were included with the pooled 1609 patients. There were no significant differences in KSS knee, KSS function, KSS total, WOMAC, and Oxford scores between TKA-PSI and standard TKA (MD 0.24 (95%CI -2.25 – 3.65) in KSS knee at mean 3-months follow-up; MD 0.56 (95%CI -1.98 – 3.10) in KSS function at mean 3-months follow-up: MD -2.48 (95%CI -10.24 – 5.29) in KSS total at mean 3-months follow-up; MD 0.05 (95%CI -1.69 – 1.80) in WOMAC at mean 3-months follow-up, and MD -0.48 (95%CI -1.92 – 0.97) in Oxford score at mean 3-months follow-up). The MDs were less than minimally clinically important differences (MCID) in KSS knee (MCID 5.3), KSS function (MCID 6.1), WOMAC (MCID 10), and Oxford scores (MCID 4.3) [7274]. There were no significant differences in EQ-5D VAS, SF-12 physical and mental scores between groups: MD -0.56 (95%CI -6.57 – 5.45) in EQ-5D VAS at mean 3-months follow-up; MD 0.51 (95%CI -2.58 – 3.59) in SF-12 physical score at mean 3-months follow-up (MCID 4.5); MD 1.84 (95%CI -3.02 – 6.69) in SF-12 mental score at mean 3-months follow-up (MCID 3.3) [72].
In patients followed for 1-year or more, 13 studies were included with the pooled 1384 patients. There were no significant differences in Oxford, WOMAC, KSS knee and total, KOOS ADL and sports, and EQ-5D VAS between groups: MD 1.00 (95%CI -11.54 – 0.59) in KSS knee at 12-months follow-up, MD -5.22 (95%CI -10.70 – 0.06) in KSS function at 12-months follow-up, MD -2.51 (95%CI -14.18 – 9.15) in KSS total at mean 18.8-months follow-up, MD 2.66 (95%CI -1.34 – 6.67) in Oxford at mean 20.0-months follow-up, MD 0.25 (95%CI -4.39 – 4.89) in WOMAC at mean 21.6-months follow-up, MD 6.09 (95%CI 6.09–0.03 – 12.21) in KOOS ADL at mean 18.3-months follow-up, MD -2.39 (95%CI -19.99 – 15.20) in KOOS sports at mean 18.3-months follow-up, and MD -1.38 (95%CI -6.87 – 4.10) in EQ-5D VAS at mean 24-months follow-up). We found differences in KSS function, and KOOS symptom, pain, and QoL between groups, however, the pooled patients were small (less than 150) and the differences were less than the MCID: MD -5.34 (95CI -10.50 – − 0.18) in KSS function at mean 12-months follow-up (MCID 6.1), MD 5.23 (95%CI 0.11–10.35) in KOOS symptom at mean 18.3-months follow-up (MCID 10.7), MD 9.67 (95%CI 3.88–15.46) in KOOS pain at mean 18.3-months follow-up (MCID 16.7), MD 9.77 (95%CI 2.56–16.97) in KOOS QoL at mean 18.3-months follow-up (MCID 15.6) [73, 75]. We present more details in the forest plots in Figs. 2, 3, 4, 5, 6 and 7.
Two studies adopted unique outcome measures (i.e. Kujala [65] and Lysholm [35] scores), but they did not show differences in the outcome measures between TKA-PSI and standard TKA groups.

Surgery time

Eighteen studies were included with the pooled 1592 patients. There was high heterogeneity between studies with I2 = 94%. The pooled mean difference was MD − 3.09 min (95%CI -6.73 – 0.55) and there was no significant difference between groups, as shown in Fig. 8.

Blood loss

Fifteen studies were included with various measuring scales and different time points when blood loss was measured postoperatively. In measuring blood loss, 5 studies measured volume of drainage fluid from suction drain device [34, 36, 39, 51, 67], 5 studies measured perioperative hemoglobin reduction [35, 42, 54, 64, 70], 3 studies measured intraoperative bleeding [7, 55, 68], 1 study calculated blood loss using the Mercuriali & Inghilleri formula [53, 76], and 1 study lacked information about the measurement [44]. The time points were most prominently at intraoperative in 3 studies, postoperative 24 h in 3 studies, 48 h in 2 studies, lowest hemoglobin in 1 study, and not shown in 4 studies. TKA-PSI decreased blood loss as compared to standard TKA with SMD -0.36 (95%CI -0.57 – − 0.15, p = 0.001) as seen in Fig. 9, although the effect size was small equivalent to hemoglobin 0.4 g/dl (95%CI 0.18–0.88) reduction. The effect size was estimated using the calculated SMD (− 0.36, 95%CI -0.57 - -0.15) and the pooled SD (1.2 g/dl) in hemoglobin decrease from a previous review examined blood loss after TKA [77, 78]: 0.43 g/dl (95%CI 0.18–0.88) = 0.36 (95%CI 0.15–0.57) × 1.2 g/dl.

Transfusion rate

Nine studies were chosen in qualitative synthesis and 8 studies were meta-analyzed with the pooled 487 patients. The overall transfusion rate was 17.3% (98/567), whereas two studies did not have any cases with transfusion. There was not significant difference in transfusion rate between TKA-PSI and standard TKA group with risk difference − 0.14 (95%CI -0.33 – 0.05, p = 0.16), as seen in Fig. 10. Pietsch et al. (2013) did not specify transfusion rate, but they described that there was no difference in transfusion rate in the trial.

Complications

Overall, 24 studies were included with 11 RCTs and 13 non-RCTs. For SSI, 18 studies with 8 RCTs and 10 non-RCTs were included in meta-analysis. The incidence of SSI in the follow-up periods was 1.2% (24/2067). We measured a composite outcome consisting of SSI, DVT, and revision TKA. We did not find any difference in the composite outcome between TKA-PSI and standard TKA groups: risk difference 0.00 (95%CI -0.01 – 0.01, p = 0.73), as shown in Fig. 11. In sensitivity analysis excluding non-RCTs, the result was consistent: risk difference 0.01 (95%CI -0.02 – 0.03, p = 0.46).
There was no significant difference in SSI between TKA-PSI and standard TKA groups: 13 out of 1049 patient in TKA-PSI versus 11 out of 1018 patients in standard TKA with risk difference 0.00 (95%CI -0.01 – 0.01, p = 0.83) as seen in Additional file 1: Appendix F. In sensitivity analysis excluding non-RCTs, the result was consistent: risk difference 0.01 (95%CI -0.02 – 0.03, p = 0.59) in RCTs.
For DVT, 16 studies were included in meta-analysis with 7 RCTs and 9 non-RCTs. The incidence of DVT was 1.5% (12/1716) in the pooled trials. There was no significant difference between groups: 9 out of 881 patients in TKA-PSI versus 3 out of 835 in standard TKA with risk difference 0.01 (95%CI -0.01 – 0.02, p = 0.28) as seen in Additional file 1: Appendix G. In sensitivity analysis excluding non-RCTs, the result was consistent: risk difference 0.01 (95%CI -0.01 – 0.03, p = 0.44) in RCTs. Pulmonary emboli in Boonen et al. (2016) and NCT 02539992 were counted as DVT, considering the overlap between pulmonary embolism and DVT.
For revision TKA, 14 studies were included and 13 studies were meta-analyzed with 4 RCTs and 9 non-RCTs. The incidence of revision TKA was 0.9% (11/1227) and there was no significant difference between groups: 3 out of 601 patients in TKA-PSI versus 8 out of 626 in standard TKA with risk difference − 0.01 (95%CI -0.02 – 0.01, p = 0.83) as seen in Additional file 1: Appendix H. The specific reasons for revision TKA were infection (n = 1), tibia loosening (n = 1), and patella resurfacing (n = 1) in TKA-PSI group, whereas infection (n = 2), tibia loosening (n = 2), patella resurfacing (n = 2), and instability (n = 2) in standard TKA group. Pourgiezis et al. (2016) experience one case with revision TKA in TKA-PSI group, but the data was not synthesized in meta-analysis, because the number of events in standard TKA was not shown. Also, Rathod et al. (2015) experienced reoperation for hematoma in TKA-PSI group, but we did not the case as revision TKA. Findings of the primary outcomes are summarized in ‘Summary of findings’ (Fig. 12).

Lower-limb alignment (secondary outcome)

Eight studies were included with the pooled 930 patients [45, 49, 53, 57, 58, 67, 68, 70]. Lower-limb alignment was monitored using hip-knee-angle (HKA) [53, 58, 67, 70] and mechanical axis [45, 49, 57, 68]. The positioning of the femoral prosthetic implant was examined using coronal [45, 49, 53, 57, 58, 67, 70] and sagittal [49, 53, 57, 58, 70] radiographs or CT scans. Also, rotational alignment of the femoral prosthetic implant was examined using CT scans [45, 58, 70]. The positioning of the tibial prosthetic implant was examined using coronal [45, 49, 53, 57, 58, 67, 70] and sagittal [45, 49, 53, 57, 58, 70] radiographs or CT scans. In the included studies, rotational alignment of the tibial prosthetic implant was not examined. All studies defined outliers as three or more degrees deviation from the planned alignment.
There were no significant differences in HKA (95%CI -0.09 – 0.05, p = 0.58), mechanical axis (95%CI -0.21 – 0.16, p = 0.81), femoral coronal positioning (95%CI -0.05 – 0.05, p = 0.95), femoral sagittal positioning (95%CI -0.13 – 0.14, p = 0.91), femoral rotational positioning (95%CI -0.32 – 0.07, p = 0.21), tibial coronal positioning (95%CI -0.03 – 0.05, p = 0.60), and tibial sagittal positioning (95%CI -0.16 – 0.10, p = 0.63) when comparing between TKA-PSI and standard TKA, as shown in Additional file 1: Appendix I.

Discussion

This systematic review included 38 studies (26 RCTs and 12 non-RCTs) to evaluate the efficacy of TKA using PSI as compared to standard TKA for patients with end-stage knee OA. For PROM, TKA-PSI did not show superior outcomes among patients followed for less than 1-year. Also, among patients followed for 1-year or more, we could not find clinically important differences between TKA-PSI and standard TKA groups. Lower-limb alignment and prosthetic implant positioning did not differ between TKA-PSI and standard TKA groups. TKA-PSI decreased perioperative blood loss, though the effect size was small. TKA-PSI did not reduce transfusion rate and surgery time. The most striking of this review was that we investigated three prominent complications (i.e. SSI, DVT, and revision TKA) and overall complication rates in TKA-PSI were small: 1.3% in SSI, 1.0% in DVT, and 0.5% in revision TKA in the short-term follow-up periods (maximum 44-months). We did not find any differences in complication rates between TKA-PSI and standard TKA groups, but the pooled events were insufficient to draw a conclusion.
We found 12 other systematic reviews assessing the efficacy of PSI as compared to standard TKA. Radiographic, CT-, or MRI-identified alignments of the components (i.e. mechanical axis) were the most common main outcomes in 9 reviews. Two reviews showed the efficacy of PSI for favorable mechanical axis and femoral rotational alignment. However, the effects of the favorable alignments on PROM and reduction of revision TKA were not reported in their reviews. Goyal et al. (2016) described that a meta-analysis with 5 RCTs showed no significant differences in PROMs between TKA-PSI and standard TKA, but not conclusive with limited numbers of the pooled patients (379 patients). Mannan et al. (2017) demonstrated that TKA-PSI did not improve PROM compared to standard TKA, including non-RCTs. In this review, we limited to RCTs and meta-analyzed the pooled 1299 patients (666 in TKA-PSI versus 633 in standard TKA) in PROM, concluding that PSI did not improve PROM among patients followed both for less than 1-year and for 1-year or more. The conclusion was consistent with the previous systematic reviews. Thienpont et al. (2017) reviewed that TKA-PSI decreased blood loss and surgery time; however, the potential benefits, such as decreasing transfusion, SSI and DVT, were not reported. In our review, TKA-PSI decreased blood loss with a small effect which corresponded to hemoglobin 0.4 g/dl (95%CI 0.1–0.7) reduction, but did not shorten surgery time. TKA-PSI did not reduce transfusion, SSI, and DVT rates.
The most noteworthy strength of this review is robust methodology: this review followed guidance from the Cochrane handbook of Systematic Reviews [79]. In our search of potentially eligible studies, we screened ongoing clinical trials. Also, we selected the main outcomes avoiding surrogate or interim outcomes [79]. Another strength is generalizability to the target population of interest. Among the included studies, 29 studies recruited patients solely diagnosed with knee OA. The prevalence of females in the pooled population was 62%, which highly reflected the general population [80]. Also, their ages were almost within 60–79, which is the predominant population in need of TKA [81]. Overall, the characteristics of the pooled population in this review accurately reflected the prevalence of knee OA in clinical practice. Also, 1361 patients were pooled in PROM and the generalizability of this review is favorable.
The most remarkable issue in the certainty of evidence was inappropriate blinding. Patients could not be blinded in the trait of PSI with preoperative MRI or CT. Also, surgeons could not be blinded, which potentially created performance bias. For these reasons, we interpreted the results in PROM as having a moderate quality of evidence. For surgery time and blood loss, each point estimate in the included studies varied with high heterogeneity (I2 = 94%), and we interpreted the evidence as being moderate in quality. The results in transfusion and complication rates were imprecise because of the small number of the pooled events with wide confidence intervals. We did not downgrade, but we suspected publication bias in the funnel plot in transfusion rate in Additional file 1: Appendix J. We interpreted them as having a moderate quality of evidence. The assessment sheet of quality of evidence for each outcome is presented in Additional file 1: Appendix K.
In limitation, we identified 4 studies in non-English literatures in the searching process, but we could not include the data in this review for our inability to literate the articles (language bias), as listed in Additional file 1: Appendix C. Also, in two studies, SD were not presented and we estimated the SD referring to the other studies in this review: Roh et al. (2013) and Boonen et al. (2013) for surgery time and blood loss. In this review, the longest follow-up periods were 44-month and future systematic reviews including studies with longer follow-up periods would be needed for examining long-term efficacy in addition to cost effectiveness using PSI. Also, a newer surgical device and procedure to perform robotic-arm assisted TKA, has been recently introduced with potentially superior clinical outcomes as compared to standard jig-based TKA [82]. Further studies are required to elucidate the full benefit and any limitations.

Conclusions

TKA using PSI does not improve PROMs, surgery time and transfusion rate as compared to standard TKA among patients with end-stage OA of the knee followed for less than 1-year and for 1-year or more. TKA using PSI decreases blood loss with a small effect, but the effect is not enough to decrease transfusion rate. TKA using PSI does not reduce surgery time, and if it does, the degree of reduction is not clinically significant. TKA using PSI may not reduce SSI, DVT, and revision TKA, but they are inconclusive.

Supplementary information

Supplementary information accompanies this paper at https://​doi.​org/​10.​1186/​s12891-019-2940-2.

Acknowledgements

We deeply appreciate librarians at McMaster Health Sciences Library and teaching staffs in the department of health research methods, evidence and impact, McMaster University.
Not applicable.
Not applicable.

Competing interests

The authors declare that they have no competing interests.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Anhänge

Supplementary information

Additional file 1: Appendix A. PICO format research question. Appendix B.1. searching in Embase (1974 to February 15th, 2019). Appendix B.2. searching in Medline (1946 to February 15th, 2019). Appendix B.3. searching in CENTRAL (The Cochrane Library 2019, Issue 2). Appendix C. Excluded studies with special reasons. Appendix D. Unpublished ongoing studies. Appendix E. MINORS scores in non-RCTs. Appendix F.1. Forest plot in surgical site infection (RCTs and non-RCTs). Appendix F.2. Forest plot in surgical site infection (RCTs). Appendix F.2. Forest plot in surgical site infection (RCTs). Appendix G.2. Forest plot in DVT (RCTs). Appendix G.2. Forest plot in DVT (RCTs). Appendix H.2. Forest plot in revision TKA (RCTs). Appendix I.1. Forest plot in HKA. Appendix I.2. Forest plot in mechanical axis. Appendix I.3. Forest plot in femoral coronal alignment. Appendix I.4. Forest plot in femoral sagittal alignment. Appendix I.5. Forest plot in femoral rotational alignment. Appendix I.6. Forest plot in tibial coronal alignment. Appendix I.7. Forest plot in tibial sagittal alignment. Appendix J. Funnel plots. Appendix K.1. Assessment of the quality of the evidence for Oxford (less than 1-year). Appendix K.2. Assessment of the quality of the evidence for WOMAC (1-year or more). Appendix K.3. Assessment of the quality of the evidence for Surgery time. Appendix K.4. Assessment of the quality of the evidence for Blood loss. Appendix K.5. Assessment of the quality of the evidence for transfusion rate. Appendix K.6. Assessment of the quality of the evidence for complications.
Literatur
3.
Zurück zum Zitat Carr AJ, Robertsson O, Graves S, Price AJ, Arden NK, Judge A, Beard DJ. Knee replacement. Lancet. 2012;379(9823):1331–40.PubMedCrossRef Carr AJ, Robertsson O, Graves S, Price AJ, Arden NK, Judge A, Beard DJ. Knee replacement. Lancet. 2012;379(9823):1331–40.PubMedCrossRef
4.
Zurück zum Zitat Inacio MCS, Paxton EW, Graves SE, Namba RS, Nemes S. Projected increase in total knee arthroplasty in the United States–an alternative projection model. Osteoarthr Cartil. 2017;25(11):1797–803.CrossRef Inacio MCS, Paxton EW, Graves SE, Namba RS, Nemes S. Projected increase in total knee arthroplasty in the United States–an alternative projection model. Osteoarthr Cartil. 2017;25(11):1797–803.CrossRef
5.
Zurück zum Zitat Confalonieri N, Manzotti A, Pullen C, Ragone V. Computer assisted technique versus intramedullary and extramedullary alignment systems in total knee replacement: a radiological comparison. Acta Orthop Belg. 2005;71(6):703.PubMed Confalonieri N, Manzotti A, Pullen C, Ragone V. Computer assisted technique versus intramedullary and extramedullary alignment systems in total knee replacement: a radiological comparison. Acta Orthop Belg. 2005;71(6):703.PubMed
6.
Zurück zum Zitat Sharkey PF, Hozack WJ, Rothman RH, Shastri S, Jacoby SM. Why are total knee arthroplasties failing today? Clin Orthop Relat Res. 2002;404:7–13.CrossRef Sharkey PF, Hozack WJ, Rothman RH, Shastri S, Jacoby SM. Why are total knee arthroplasties failing today? Clin Orthop Relat Res. 2002;404:7–13.CrossRef
7.
Zurück zum Zitat Noble JW Jr, Moore CA, Liu N. The value of patient-matched instrumentation in total knee arthroplasty. J Arthroplast. 2012;27(1):153–5.CrossRef Noble JW Jr, Moore CA, Liu N. The value of patient-matched instrumentation in total knee arthroplasty. J Arthroplast. 2012;27(1):153–5.CrossRef
8.
Zurück zum Zitat Klatt BA, Goyal N, Austin MS, Hozack WJ. Custom-fit total knee arthroplasty (OtisKnee) results in malalignment. J Arthroplast. 2008;23(1):26–9.CrossRef Klatt BA, Goyal N, Austin MS, Hozack WJ. Custom-fit total knee arthroplasty (OtisKnee) results in malalignment. J Arthroplast. 2008;23(1):26–9.CrossRef
9.
Zurück zum Zitat Bali K, Walker P, Bruce W. Custom-fit total knee arthroplasty: our initial experience in 32 knees. J Arthroplast. 2012;27(6):1149–54.CrossRef Bali K, Walker P, Bruce W. Custom-fit total knee arthroplasty: our initial experience in 32 knees. J Arthroplast. 2012;27(6):1149–54.CrossRef
10.
Zurück zum Zitat Nam D, McArthur BA, Cross MB, Pearle AD, Mayman DJ, Haas SB. Patient-specific instrumentation in total knee arthroplasty: a review. J knee Surg. 2012;25(03):213–20.PubMedCrossRef Nam D, McArthur BA, Cross MB, Pearle AD, Mayman DJ, Haas SB. Patient-specific instrumentation in total knee arthroplasty: a review. J knee Surg. 2012;25(03):213–20.PubMedCrossRef
11.
Zurück zum Zitat Thienpont E, Schwab PE, Fennema P. Efficacy of patient-specific instruments in total knee arthroplasty: a systematic review and meta-analysis. JBJS. 2017;99(6):521–30.CrossRef Thienpont E, Schwab PE, Fennema P. Efficacy of patient-specific instruments in total knee arthroplasty: a systematic review and meta-analysis. JBJS. 2017;99(6):521–30.CrossRef
12.
Zurück zum Zitat Mannan A, Smith TO. Favourable rotational alignment outcomes in PSI knee arthroplasty: a level 1 systematic review and meta-analysis. Knee. 2016;23(2):186–90.PubMedCrossRef Mannan A, Smith TO. Favourable rotational alignment outcomes in PSI knee arthroplasty: a level 1 systematic review and meta-analysis. Knee. 2016;23(2):186–90.PubMedCrossRef
13.
Zurück zum Zitat Huang NF, Dowsey MM, Ee E, Stoney JD, Babazadeh S, Choong PF. Coronal alignment correlates with outcome after total knee arthroplasty: five-year follow-up of a randomized controlled trial. J Arthroplast. 2012;27(9):1737–41.CrossRef Huang NF, Dowsey MM, Ee E, Stoney JD, Babazadeh S, Choong PF. Coronal alignment correlates with outcome after total knee arthroplasty: five-year follow-up of a randomized controlled trial. J Arthroplast. 2012;27(9):1737–41.CrossRef
14.
Zurück zum Zitat Longstaff LM, Sloan K, Stamp N, Scaddan M, Beaver R. Good alignment after total knee arthroplasty leads to faster rehabilitation and better function. J Arthroplast. 2009;24(4):570–8.CrossRef Longstaff LM, Sloan K, Stamp N, Scaddan M, Beaver R. Good alignment after total knee arthroplasty leads to faster rehabilitation and better function. J Arthroplast. 2009;24(4):570–8.CrossRef
15.
Zurück zum Zitat Fang DM, Ritter MA, Davis KE. Coronal alignment in total knee arthroplasty: just how important is it? J Arthroplast. 2009;24(6):39–43.CrossRef Fang DM, Ritter MA, Davis KE. Coronal alignment in total knee arthroplasty: just how important is it? J Arthroplast. 2009;24(6):39–43.CrossRef
16.
Zurück zum Zitat Peersman G, Laskin R, Davis J, Peterson MGE, Richart T. Prolonged operative time correlates with increased infection rate after total knee arthroplasty. HSS J. 2006;2(1):70–2.PubMedPubMedCentralCrossRef Peersman G, Laskin R, Davis J, Peterson MGE, Richart T. Prolonged operative time correlates with increased infection rate after total knee arthroplasty. HSS J. 2006;2(1):70–2.PubMedPubMedCentralCrossRef
17.
Zurück zum Zitat Fujita S, Hirota S, Oda T, Kato Y, Tsukamoto Y, Fuji T. Deep venous thrombosis after total hip or total knee arthroplasty in patients in Japan. Clin Orthop Relat Res. 2000;375:168–74.CrossRef Fujita S, Hirota S, Oda T, Kato Y, Tsukamoto Y, Fuji T. Deep venous thrombosis after total hip or total knee arthroplasty in patients in Japan. Clin Orthop Relat Res. 2000;375:168–74.CrossRef
18.
Zurück zum Zitat Sassoon A, Nam D, Nunley R, Barrack R. Systematic review of patient-specific instrumentation in total knee arthroplasty: new but not improved. Clin Orthop Relat Res. 2015;473(1):151–8.PubMedCrossRef Sassoon A, Nam D, Nunley R, Barrack R. Systematic review of patient-specific instrumentation in total knee arthroplasty: new but not improved. Clin Orthop Relat Res. 2015;473(1):151–8.PubMedCrossRef
19.
Zurück zum Zitat Thienpont E, Schwab PE, Fennema P. A systematic review and meta-analysis of patient-specific instrumentation for improving alignment of the components in total knee replacement. Bone Joint J. 2014;96(8):1052–61.PubMedCrossRef Thienpont E, Schwab PE, Fennema P. A systematic review and meta-analysis of patient-specific instrumentation for improving alignment of the components in total knee replacement. Bone Joint J. 2014;96(8):1052–61.PubMedCrossRef
20.
Zurück zum Zitat Cavaignac E, Pailhe R, Laumond G, Murgier J, Reina N, Laffosse JM, Chiron P. Evaluation of the accuracy of patient-specific cutting blocks for total knee arthroplasty: a meta-analysis. Int Orthop. 2015;39(8):1541–52.PubMedCrossRef Cavaignac E, Pailhe R, Laumond G, Murgier J, Reina N, Laffosse JM, Chiron P. Evaluation of the accuracy of patient-specific cutting blocks for total knee arthroplasty: a meta-analysis. Int Orthop. 2015;39(8):1541–52.PubMedCrossRef
21.
Zurück zum Zitat Schotanus MG, Thijs E, Heijmans M, Vos R, Kort NP. Favourable alignment outcomes with MRI-based patient-specific instruments in total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc. 2018;26(9):2659–68.PubMedCrossRef Schotanus MG, Thijs E, Heijmans M, Vos R, Kort NP. Favourable alignment outcomes with MRI-based patient-specific instruments in total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc. 2018;26(9):2659–68.PubMedCrossRef
22.
Zurück zum Zitat Alcelik I, Blomfield M, Öztürk C, Soni A, Charity R, Acornley A. A comparison of short term radiological alignment outcomes of the patient specific and standard instrumentation for primary total knee arthroplasty: a systematic review and meta-analysis. Acta Orthop Traumatol Turc. 2017;51(3):215–22.PubMedPubMedCentralCrossRef Alcelik I, Blomfield M, Öztürk C, Soni A, Charity R, Acornley A. A comparison of short term radiological alignment outcomes of the patient specific and standard instrumentation for primary total knee arthroplasty: a systematic review and meta-analysis. Acta Orthop Traumatol Turc. 2017;51(3):215–22.PubMedPubMedCentralCrossRef
23.
Zurück zum Zitat Mannan A, Akinyooye D, Hossain F. A meta-analysis of functional outcomes in patient-specific instrumented knee arthroplasty. J knee Surg. 2017;30(07):668–74.PubMedCrossRef Mannan A, Akinyooye D, Hossain F. A meta-analysis of functional outcomes in patient-specific instrumented knee arthroplasty. J knee Surg. 2017;30(07):668–74.PubMedCrossRef
24.
Zurück zum Zitat Goyal T, Tripathy SK. Does patient-specific instrumentations improve short-term functional outcomes after total knee arthroplasty? A systematic review and meta-analysis. J Arthroplast. 2016;31(10):2173–80.CrossRef Goyal T, Tripathy SK. Does patient-specific instrumentations improve short-term functional outcomes after total knee arthroplasty? A systematic review and meta-analysis. J Arthroplast. 2016;31(10):2173–80.CrossRef
25.
Zurück zum Zitat Zhang QM, Chen JY, Li H, Chai W, Ni M, Zhang ZD, Yang F. No evidence of superiority in reducing outliers of component alignment for patient-specific instrumentation for Total knee Arthroplasty: a systematic review. Orthop Surg. 2015;7(1):19–25.PubMedPubMedCentralCrossRef Zhang QM, Chen JY, Li H, Chai W, Ni M, Zhang ZD, Yang F. No evidence of superiority in reducing outliers of component alignment for patient-specific instrumentation for Total knee Arthroplasty: a systematic review. Orthop Surg. 2015;7(1):19–25.PubMedPubMedCentralCrossRef
26.
Zurück zum Zitat Fu H, Wang J, Zhou S, Cheng T, Zhang W, Wang Q, Zhang X. No difference in mechanical alignment and femoral component placement between patient-specific instrumentation and conventional instrumentation in TKA. Knee Surg Sports Traumatol Arthrosc. 2015;23(11):3288–95.PubMedCrossRef Fu H, Wang J, Zhou S, Cheng T, Zhang W, Wang Q, Zhang X. No difference in mechanical alignment and femoral component placement between patient-specific instrumentation and conventional instrumentation in TKA. Knee Surg Sports Traumatol Arthrosc. 2015;23(11):3288–95.PubMedCrossRef
28.
Zurück zum Zitat Hafez MA, Chelule KL, Seedhom BB, Sherman KP. Computer-assisted total knee replacement: could a two-piece custom template replace the complex conventional instrumentations? Comput Aided Surg. 2004;9:93–4. Hafez MA, Chelule KL, Seedhom BB, Sherman KP. Computer-assisted total knee replacement: could a two-piece custom template replace the complex conventional instrumentations? Comput Aided Surg. 2004;9:93–4.
29.
Zurück zum Zitat Higgins JP, Altman DG, Gøtzsche PC, Jüni P, Moher D, Oxman AD, et al. The Cochrane Collaboration’s tool for assessing risk of bias in randomised trials. Bmj. 2011;343:d5928.PubMedPubMedCentralCrossRef Higgins JP, Altman DG, Gøtzsche PC, Jüni P, Moher D, Oxman AD, et al. The Cochrane Collaboration’s tool for assessing risk of bias in randomised trials. Bmj. 2011;343:d5928.PubMedPubMedCentralCrossRef
30.
Zurück zum Zitat Slim K, Nini E, Forestier D, Kwiatkowski F, Panis Y, Chipponi J. Methodological index for non-randomized studies (MINORS): development and validation of a new instrument. ANZ J Surg. 2003;73(9):712–6.PubMedCrossRef Slim K, Nini E, Forestier D, Kwiatkowski F, Panis Y, Chipponi J. Methodological index for non-randomized studies (MINORS): development and validation of a new instrument. ANZ J Surg. 2003;73(9):712–6.PubMedCrossRef
31.
Zurück zum Zitat Wan X, Wang W, Liu J, Tong T. Estimating the sample mean and standard deviation from the sample size, median, range and/or interquartile range. BMC Med Res Methodol. 2014;14(1):135.PubMedPubMedCentralCrossRef Wan X, Wang W, Liu J, Tong T. Estimating the sample mean and standard deviation from the sample size, median, range and/or interquartile range. BMC Med Res Methodol. 2014;14(1):135.PubMedPubMedCentralCrossRef
33.
Zurück zum Zitat Guyatt GH, Oxman AD, Vist GE, Kunz R, Falck-Ytter Y, Alonso-Coello P, Schünemann HJ. GRADE: an emerging consensus on rating quality of evidence and strength of recommendations. Bmj. 2008;336(7650):924–6.PubMedPubMedCentralCrossRef Guyatt GH, Oxman AD, Vist GE, Kunz R, Falck-Ytter Y, Alonso-Coello P, Schünemann HJ. GRADE: an emerging consensus on rating quality of evidence and strength of recommendations. Bmj. 2008;336(7650):924–6.PubMedPubMedCentralCrossRef
34.
Zurück zum Zitat Pietsch M, Djahani O, Zweiger C, Plattner F, Radl R, Tschauner C, Hofmann S. Custom-fit minimally invasive total knee arthroplasty: effect on blood loss and early clinical outcomes. Knee Surg Sports Traumatol Arthrosc. 2013;21(10):2234–40.PubMedCrossRef Pietsch M, Djahani O, Zweiger C, Plattner F, Radl R, Tschauner C, Hofmann S. Custom-fit minimally invasive total knee arthroplasty: effect on blood loss and early clinical outcomes. Knee Surg Sports Traumatol Arthrosc. 2013;21(10):2234–40.PubMedCrossRef
35.
Zurück zum Zitat Vundelinckx BJ, Bruckers L, De Mulder K, De Schepper J, Van Esbroeck G. Functional and radiographic short-term outcome evaluation of the Visionaire system, a patient-matched instrumentation system for total knee arthroplasty. J Arthroplast. 2013;28(6):964–70.CrossRef Vundelinckx BJ, Bruckers L, De Mulder K, De Schepper J, Van Esbroeck G. Functional and radiographic short-term outcome evaluation of the Visionaire system, a patient-matched instrumentation system for total knee arthroplasty. J Arthroplast. 2013;28(6):964–70.CrossRef
36.
Zurück zum Zitat Boonen B, Schotanus MGM, Kerens B, Van der Weegen W, van Drumpt RAM, Kort NP. Intra-operative results and radiological outcome of conventional and patient-specific surgery in total knee arthroplasty: a multicentre, randomised controlled trial. Knee Surg Sports Traumatol Arthrosc. 2013;21(10):2206–12.PubMedCrossRef Boonen B, Schotanus MGM, Kerens B, Van der Weegen W, van Drumpt RAM, Kort NP. Intra-operative results and radiological outcome of conventional and patient-specific surgery in total knee arthroplasty: a multicentre, randomised controlled trial. Knee Surg Sports Traumatol Arthrosc. 2013;21(10):2206–12.PubMedCrossRef
37.
Zurück zum Zitat Chareancholvanich K, Narkbunnam R, Pornrattanamaneewong C. A prospective randomised controlled study of patient-specific cutting guides compared with conventional instrumentation in total knee replacement. Bone Joint J. 2013;95(3):354–9.PubMedCrossRef Chareancholvanich K, Narkbunnam R, Pornrattanamaneewong C. A prospective randomised controlled study of patient-specific cutting guides compared with conventional instrumentation in total knee replacement. Bone Joint J. 2013;95(3):354–9.PubMedCrossRef
38.
Zurück zum Zitat Hamilton WG, Parks NL, Saxena A. Patient-specific instrumentation does not shorten surgical time: a prospective, randomized trial. J Arthroplast. 2013;28(8):96–100.CrossRef Hamilton WG, Parks NL, Saxena A. Patient-specific instrumentation does not shorten surgical time: a prospective, randomized trial. J Arthroplast. 2013;28(8):96–100.CrossRef
39.
Zurück zum Zitat Roh YW, Kim TW, Lee S, Seong SC, Lee MC. Is TKA using patient-specific instruments comparable to conventional TKA? A randomized controlled study of one system. Clin Orthop Relat Res. 2013;471(12):3988–95.PubMedPubMedCentralCrossRef Roh YW, Kim TW, Lee S, Seong SC, Lee MC. Is TKA using patient-specific instruments comparable to conventional TKA? A randomized controlled study of one system. Clin Orthop Relat Res. 2013;471(12):3988–95.PubMedPubMedCentralCrossRef
40.
Zurück zum Zitat Ng VY, Arnott L, Li J, Hopkins R, Lewis J, Sutphen S, et al. Comparison of custom to standard TKA instrumentation with computed tomography. Knee Surg Sports Traumatol Arthrosc. 2014;22(8):1833–42.PubMedCrossRef Ng VY, Arnott L, Li J, Hopkins R, Lewis J, Sutphen S, et al. Comparison of custom to standard TKA instrumentation with computed tomography. Knee Surg Sports Traumatol Arthrosc. 2014;22(8):1833–42.PubMedCrossRef
41.
Zurück zum Zitat Renson L, Poilvache P, Van den Wyngaert H. Improved alignment and operating room efficiency with patient-specific instrumentation for TKA. Knee. 2014;21(6):1216–20.PubMedCrossRef Renson L, Poilvache P, Van den Wyngaert H. Improved alignment and operating room efficiency with patient-specific instrumentation for TKA. Knee. 2014;21(6):1216–20.PubMedCrossRef
42.
Zurück zum Zitat Dossett HG, Estrada NA, Swartz GJ, LeFevre GW, Kwasman BG. A randomised controlled trial of kinematically and mechanically aligned total knee replacements: two-year clinical results. Bone Joint J. 2014;96(7):907–13.PubMedCrossRef Dossett HG, Estrada NA, Swartz GJ, LeFevre GW, Kwasman BG. A randomised controlled trial of kinematically and mechanically aligned total knee replacements: two-year clinical results. Bone Joint J. 2014;96(7):907–13.PubMedCrossRef
43.
Zurück zum Zitat Abdel MP, Parratte S, Blanc G, Ollivier M, Pomero V, Viehweger E, Argenson JNA. No benefit of patient-specific instrumentation in TKA on functional and gait outcomes: a randomized clinical trial. Clin Orthop Relat Res. 2014;472(8):2468–76.PubMedPubMedCentralCrossRef Abdel MP, Parratte S, Blanc G, Ollivier M, Pomero V, Viehweger E, Argenson JNA. No benefit of patient-specific instrumentation in TKA on functional and gait outcomes: a randomized clinical trial. Clin Orthop Relat Res. 2014;472(8):2468–76.PubMedPubMedCentralCrossRef
44.
Zurück zum Zitat Chotanaphuti T, Wangwittayakul V, Khuangsirikul S, Foojareonyos T. The accuracy of component alignment in custom cutting blocks compared with conventional total knee arthroplasty instrumentation: prospective control trial. Knee. 2014;21(1):185–8.PubMedCrossRef Chotanaphuti T, Wangwittayakul V, Khuangsirikul S, Foojareonyos T. The accuracy of component alignment in custom cutting blocks compared with conventional total knee arthroplasty instrumentation: prospective control trial. Knee. 2014;21(1):185–8.PubMedCrossRef
45.
Zurück zum Zitat Pfitzner T, Abdel MP, von Roth P, Perka C, Hommel H. Small improvements in mechanical axis alignment achieved with MRI versus CT-based patient-specific instruments in TKA: a randomized clinical trial. Clin Orthop Relat Res. 2014;472(10):2913–22.PubMedPubMedCentralCrossRef Pfitzner T, Abdel MP, von Roth P, Perka C, Hommel H. Small improvements in mechanical axis alignment achieved with MRI versus CT-based patient-specific instruments in TKA: a randomized clinical trial. Clin Orthop Relat Res. 2014;472(10):2913–22.PubMedPubMedCentralCrossRef
46.
Zurück zum Zitat Woolson ST, Harris AH, Wagner DW, Giori NJ. Component alignment during total knee arthroplasty with use of standard or custom instrumentation: a randomized clinical trial using computed tomography for postoperative alignment measurement. JBJS. 2014;96(5):366–72.CrossRef Woolson ST, Harris AH, Wagner DW, Giori NJ. Component alignment during total knee arthroplasty with use of standard or custom instrumentation: a randomized clinical trial using computed tomography for postoperative alignment measurement. JBJS. 2014;96(5):366–72.CrossRef
47.
Zurück zum Zitat Nabavi A, Olwill CM. Early outcome after total knee replacement using computed tomography–based patient-specific cutting blocks versus standard instrumentation. J Orthop Surg. 2015;23(2):182–4.CrossRef Nabavi A, Olwill CM. Early outcome after total knee replacement using computed tomography–based patient-specific cutting blocks versus standard instrumentation. J Orthop Surg. 2015;23(2):182–4.CrossRef
48.
Zurück zum Zitat Thienpont E, Grosu I, Paternostre F, Schwab PE, Yombi JC. The use of patient-specific instruments does not reduce blood loss during minimally invasive total knee arthroplasty? Knee Surg Sports Traumatol Arthrosc. 2015;23(7):2055–60.PubMedCrossRef Thienpont E, Grosu I, Paternostre F, Schwab PE, Yombi JC. The use of patient-specific instruments does not reduce blood loss during minimally invasive total knee arthroplasty? Knee Surg Sports Traumatol Arthrosc. 2015;23(7):2055–60.PubMedCrossRef
49.
Zurück zum Zitat Yan CH, Chiu KY, Ng FY, Chan PK, Fang CX. Comparison between patient-specific instruments and conventional instruments and computer navigation in total knee arthroplasty: a randomized controlled trial. Knee Surg Sports Traumatol Arthrosc. 2015;23(12):3637–45.PubMedCrossRef Yan CH, Chiu KY, Ng FY, Chan PK, Fang CX. Comparison between patient-specific instruments and conventional instruments and computer navigation in total knee arthroplasty: a randomized controlled trial. Knee Surg Sports Traumatol Arthrosc. 2015;23(12):3637–45.PubMedCrossRef
50.
Zurück zum Zitat Chen JY, Chin PL, Tay DKJ, Chia SL, Lo NN, Yeo SJ. Functional outcome and quality of life after patient-specific instrumentation in total knee arthroplasty. J Arthroplast. 2015;30(10):1724–8.CrossRef Chen JY, Chin PL, Tay DKJ, Chia SL, Lo NN, Yeo SJ. Functional outcome and quality of life after patient-specific instrumentation in total knee arthroplasty. J Arthroplast. 2015;30(10):1724–8.CrossRef
51.
Zurück zum Zitat Kotela A, Lorkowski J, Kucharzewski M, Wilk-Frańczuk M, Śliwiński Z, Frańczuk B, et al. Patient-specific CT-based instrumentation versus conventional instrumentation in total knee arthroplasty: a prospective randomized controlled study on clinical outcomes and in-hospital data. Biomed Res Int. 2015;2015:1.CrossRef Kotela A, Lorkowski J, Kucharzewski M, Wilk-Frańczuk M, Śliwiński Z, Frańczuk B, et al. Patient-specific CT-based instrumentation versus conventional instrumentation in total knee arthroplasty: a prospective randomized controlled study on clinical outcomes and in-hospital data. Biomed Res Int. 2015;2015:1.CrossRef
52.
Zurück zum Zitat Rathod PA, Deshmukh AJ, Cushner FD. Reducing blood loss in bilateral total knee arthroplasty with patient-specific instrumentation. Orthopedic Clin. 2015;46(3):343–50. Rathod PA, Deshmukh AJ, Cushner FD. Reducing blood loss in bilateral total knee arthroplasty with patient-specific instrumentation. Orthopedic Clin. 2015;46(3):343–50.
53.
Zurück zum Zitat Abane L, Anract P, Boisgard S, Descamps S, Courpied JP, Hamadouche M. A comparison of patient-specific and conventional instrumentation for total knee arthroplasty: a multicentre randomised controlled trial. Bone Joint J. 2015;97(1):56–63.PubMedCrossRef Abane L, Anract P, Boisgard S, Descamps S, Courpied JP, Hamadouche M. A comparison of patient-specific and conventional instrumentation for total knee arthroplasty: a multicentre randomised controlled trial. Bone Joint J. 2015;97(1):56–63.PubMedCrossRef
54.
Zurück zum Zitat Molicnik A, Naranda J, Dolinar D. Patient-matched instruments versus standard instrumentation in total knee arthroplasty: a prospective randomized study. Wien Klin Wochenschr. 2015;127(5):235–40.PubMedCentralCrossRef Molicnik A, Naranda J, Dolinar D. Patient-matched instruments versus standard instrumentation in total knee arthroplasty: a prospective randomized study. Wien Klin Wochenschr. 2015;127(5):235–40.PubMedCentralCrossRef
55.
Zurück zum Zitat Ferrara F, Cipriani A, Magarelli N, Rapisarda S, De Santis V, Burrofato A, et al. Implant positioning in TKA: comparison between conventional and patient-specific instrumentation. Orthopedics. 2015;38(4):e271–80.PubMedCrossRef Ferrara F, Cipriani A, Magarelli N, Rapisarda S, De Santis V, Burrofato A, et al. Implant positioning in TKA: comparison between conventional and patient-specific instrumentation. Orthopedics. 2015;38(4):e271–80.PubMedCrossRef
56.
Zurück zum Zitat Anderl W, Pauzenberger L, Kölblinger R, Kiesselbach G, Brandl G, Laky B, et al. Patient-specific instrumentation improved mechanical alignment, while early clinical outcome was comparable to conventional instrumentation in TKA. Knee Surg Sports Traumatol Arthrosc. 2016;24(1):102–11.PubMedCrossRef Anderl W, Pauzenberger L, Kölblinger R, Kiesselbach G, Brandl G, Laky B, et al. Patient-specific instrumentation improved mechanical alignment, while early clinical outcome was comparable to conventional instrumentation in TKA. Knee Surg Sports Traumatol Arthrosc. 2016;24(1):102–11.PubMedCrossRef
57.
Zurück zum Zitat Boonen B, Schotanus MGM, Kerens B, Van der Weegen W, Hoekstra HJ, Kort NP. No difference in clinical outcome between patient-matched positioning guides and conventional instrumented total knee arthroplasty two years post-operatively: a multicentre, double-blind, randomised controlled trial. Bone Joint J. 2016;98(7):939–44.PubMedCrossRef Boonen B, Schotanus MGM, Kerens B, Van der Weegen W, Hoekstra HJ, Kort NP. No difference in clinical outcome between patient-matched positioning guides and conventional instrumented total knee arthroplasty two years post-operatively: a multicentre, double-blind, randomised controlled trial. Bone Joint J. 2016;98(7):939–44.PubMedCrossRef
58.
Zurück zum Zitat Huijbregts HJTAM, Khan RJK, Fick DP, Hall MJ, Punwar SA, Sorensen E, et al. Component alignment and clinical outcome following total knee arthroplasty: a randomised controlled trial comparing an intramedullary alignment system with patient-specific instrumentation. Bone Joint J. 2016;98(8):1043–9.PubMedCrossRef Huijbregts HJTAM, Khan RJK, Fick DP, Hall MJ, Punwar SA, Sorensen E, et al. Component alignment and clinical outcome following total knee arthroplasty: a randomised controlled trial comparing an intramedullary alignment system with patient-specific instrumentation. Bone Joint J. 2016;98(8):1043–9.PubMedCrossRef
59.
Zurück zum Zitat Pourgiezis N, Reddy SP, Nankivell M, Morrison G, VanEssen J. Alignment and component position after patient-matched instrumentation versus conventional total knee arthroplasty. J Orthop Surg. 2016;24(2):170–4.CrossRef Pourgiezis N, Reddy SP, Nankivell M, Morrison G, VanEssen J. Alignment and component position after patient-matched instrumentation versus conventional total knee arthroplasty. J Orthop Surg. 2016;24(2):170–4.CrossRef
60.
Zurück zum Zitat White PB, Ranawat AS. Patient-specific total knees demonstrate a higher manipulation rate compared to “off-the-shelf implants”. J Arthroplast. 2016;31(1):107–11.CrossRef White PB, Ranawat AS. Patient-specific total knees demonstrate a higher manipulation rate compared to “off-the-shelf implants”. J Arthroplast. 2016;31(1):107–11.CrossRef
61.
Zurück zum Zitat Culler SD, Martin GM, Swearingen A. Comparison of adverse events rates and hospital cost between customized individually made implants and standard off-the-shelf implants for total knee arthroplasty. Arthroplasty Today. 2017;3(4):257–63.PubMedPubMedCentralCrossRef Culler SD, Martin GM, Swearingen A. Comparison of adverse events rates and hospital cost between customized individually made implants and standard off-the-shelf implants for total knee arthroplasty. Arthroplasty Today. 2017;3(4):257–63.PubMedPubMedCentralCrossRef
62.
Zurück zum Zitat Kwon OR, Kang KT, Son J, Suh DS, Heo DB, Koh YG. The effect of patient-specific instrumentation incorporating an extramedullary tibial guide on operative efficiency for total knee arthroplasty. Biomed Res Int. 2017;2017:1. Kwon OR, Kang KT, Son J, Suh DS, Heo DB, Koh YG. The effect of patient-specific instrumentation incorporating an extramedullary tibial guide on operative efficiency for total knee arthroplasty. Biomed Res Int. 2017;2017:1.
63.
Zurück zum Zitat Zhu M, Chen JY, Chong HC, Yew AKS, Foo LSS, Chia SL, et al. Outcomes following total knee arthroplasty with CT-based patient-specific instrumentation. Knee Surg Sports Traumatol Arthrosc. 2017;25(8):2567–72.PubMedCrossRef Zhu M, Chen JY, Chong HC, Yew AKS, Foo LSS, Chia SL, et al. Outcomes following total knee arthroplasty with CT-based patient-specific instrumentation. Knee Surg Sports Traumatol Arthrosc. 2017;25(8):2567–72.PubMedCrossRef
64.
Zurück zum Zitat Vide J, Freitas TP, Ramos A, Cruz H, Sousa JP. Patient-specific instrumentation in total knee arthroplasty: simpler, faster and more accurate than standard instrumentation—a randomized controlled trial. Knee Surg Sports Traumatol Arthrosc. 2017;25(8):2616–21.PubMedCrossRef Vide J, Freitas TP, Ramos A, Cruz H, Sousa JP. Patient-specific instrumentation in total knee arthroplasty: simpler, faster and more accurate than standard instrumentation—a randomized controlled trial. Knee Surg Sports Traumatol Arthrosc. 2017;25(8):2616–21.PubMedCrossRef
65.
Zurück zum Zitat Kosse NM, Heesterbeek PJ, Schimmel JJ, van Hellemondt GG, Wymenga AB, Defoort KC. Stability and alignment do not improve by using patient-specific instrumentation in total knee arthroplasty: a randomized controlled trial. Knee Surg Sports Traumatol Arthrosc. 2018;26(6):1792–9.PubMedCrossRef Kosse NM, Heesterbeek PJ, Schimmel JJ, van Hellemondt GG, Wymenga AB, Defoort KC. Stability and alignment do not improve by using patient-specific instrumentation in total knee arthroplasty: a randomized controlled trial. Knee Surg Sports Traumatol Arthrosc. 2018;26(6):1792–9.PubMedCrossRef
66.
Zurück zum Zitat Steimle JA, Groover MT, Webb BA, Ceccarelli BJ. Acute perioperative comparison of patient-specific instrumentation versus conventional instrumentation utilization during bilateral total knee arthroplasty. Surg Res Pract. 2018;2018:1.CrossRef Steimle JA, Groover MT, Webb BA, Ceccarelli BJ. Acute perioperative comparison of patient-specific instrumentation versus conventional instrumentation utilization during bilateral total knee arthroplasty. Surg Res Pract. 2018;2018:1.CrossRef
67.
Zurück zum Zitat Tammachote N, Panichkul P, Kanitnate S. Comparison of customized cutting block and conventional cutting instrument in total knee arthroplasty: a randomized controlled trial. J Arthroplast. 2018;33(3):746–51.CrossRef Tammachote N, Panichkul P, Kanitnate S. Comparison of customized cutting block and conventional cutting instrument in total knee arthroplasty: a randomized controlled trial. J Arthroplast. 2018;33(3):746–51.CrossRef
68.
Zurück zum Zitat Maus U, Marques CJ, Scheunemann D, Lampe F, Lazovic D, Hommel H, et al. No improvement in reducing outliers in coronal axis alignment with patient-specific instrumentation. Knee Surg Sports Traumatol Arthrosc. 2018;26(9):2788–96.PubMedCrossRef Maus U, Marques CJ, Scheunemann D, Lampe F, Lazovic D, Hommel H, et al. No improvement in reducing outliers in coronal axis alignment with patient-specific instrumentation. Knee Surg Sports Traumatol Arthrosc. 2018;26(9):2788–96.PubMedCrossRef
69.
Zurück zum Zitat Stolarczyk A, Nagraba L, Mitek T, Stolarczyk M, Deszczyński JM, Jakucinski M. Does patient-specific instrumentation improve femoral and tibial component alignment in total knee arthroplasty? A prospective randomized study. In: Rehabilitation Science in Context. Cham: Springer; 2018. p. 11–7.CrossRef Stolarczyk A, Nagraba L, Mitek T, Stolarczyk M, Deszczyński JM, Jakucinski M. Does patient-specific instrumentation improve femoral and tibial component alignment in total knee arthroplasty? A prospective randomized study. In: Rehabilitation Science in Context. Cham: Springer; 2018. p. 11–7.CrossRef
70.
Zurück zum Zitat Van Leeuwen JA, Snorrason F, Röhrl SM. No radiological and clinical advantages with patient-specific positioning guides in total knee replacement: a multicenter randomized controlled trial. Acta Orthop. 2018;89(1):89–94.PubMedCrossRef Van Leeuwen JA, Snorrason F, Röhrl SM. No radiological and clinical advantages with patient-specific positioning guides in total knee replacement: a multicenter randomized controlled trial. Acta Orthop. 2018;89(1):89–94.PubMedCrossRef
72.
Zurück zum Zitat Clement ND, MacDonald D, Simpson AHRW. The minimal clinically important difference in the Oxford knee score and short form 12 score after total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc. 2014;22(8):1933–9.PubMedCrossRef Clement ND, MacDonald D, Simpson AHRW. The minimal clinically important difference in the Oxford knee score and short form 12 score after total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc. 2014;22(8):1933–9.PubMedCrossRef
73.
Zurück zum Zitat Lee WC, Kwan YH, Chong HC, Yeo SJ. The minimal clinically important difference for knee society clinical rating system after total knee arthroplasty for primary osteoarthritis. Knee Surg Sports Traumatol Arthrosc. 2017;25(11):3354–9.PubMedCrossRef Lee WC, Kwan YH, Chong HC, Yeo SJ. The minimal clinically important difference for knee society clinical rating system after total knee arthroplasty for primary osteoarthritis. Knee Surg Sports Traumatol Arthrosc. 2017;25(11):3354–9.PubMedCrossRef
74.
Zurück zum Zitat Clement ND, Bardgett M, Weir D, Holland J, Gerrand C, Deehan DJ. What is the minimum clinically important difference for the WOMAC index after TKA? Clin Orthop Relat Res. 2018;476(10):2005–14.PubMedPubMedCentralCrossRef Clement ND, Bardgett M, Weir D, Holland J, Gerrand C, Deehan DJ. What is the minimum clinically important difference for the WOMAC index after TKA? Clin Orthop Relat Res. 2018;476(10):2005–14.PubMedPubMedCentralCrossRef
75.
Zurück zum Zitat Monticone M, Ferrante S, Salvaderi S, Motta L, Cerri C. Responsiveness and minimal important changes for the knee injury and osteoarthritis outcome score in subjects undergoing rehabilitation after total knee arthroplasty. Am J Phys Med Rehab. 2013;92(10):864–70.CrossRef Monticone M, Ferrante S, Salvaderi S, Motta L, Cerri C. Responsiveness and minimal important changes for the knee injury and osteoarthritis outcome score in subjects undergoing rehabilitation after total knee arthroplasty. Am J Phys Med Rehab. 2013;92(10):864–70.CrossRef
76.
Zurück zum Zitat Mercuriali F, Inghilleri G. Proposal of an algorithm to help the choice of the best transfusion strategy. Curr Med Res Opin. 1996;13(8):465–78.PubMedCrossRef Mercuriali F, Inghilleri G. Proposal of an algorithm to help the choice of the best transfusion strategy. Curr Med Res Opin. 1996;13(8):465–78.PubMedCrossRef
77.
Zurück zum Zitat Devji T, Johnston BC, Patrick DL, Bhandari M, Thabane L, Guyatt GH. Presentation approaches for enhancing interpretability of patient-reported outcomes (PROs) in meta-analysis: a protocol for a systematic survey of Cochrane reviews. BMJ Open. 2017;7(9):e017138.PubMedPubMedCentralCrossRef Devji T, Johnston BC, Patrick DL, Bhandari M, Thabane L, Guyatt GH. Presentation approaches for enhancing interpretability of patient-reported outcomes (PROs) in meta-analysis: a protocol for a systematic survey of Cochrane reviews. BMJ Open. 2017;7(9):e017138.PubMedPubMedCentralCrossRef
78.
Zurück zum Zitat Palmer A, Chen A, Matsumoto T, Murphy M, Price A. Blood management in total knee arthroplasty: state-of-the-art review. J ISAKOS. 2018;3(6):358–66.CrossRef Palmer A, Chen A, Matsumoto T, Murphy M, Price A. Blood management in total knee arthroplasty: state-of-the-art review. J ISAKOS. 2018;3(6):358–66.CrossRef
79.
Zurück zum Zitat Higgins, J. P., & Green, S. (Eds.). (2008). Cochrane handbook for systematic reviews of interventions. Higgins, J. P., & Green, S. (Eds.). (2008). Cochrane handbook for systematic reviews of interventions.
80.
Zurück zum Zitat Stundner O, Danninger T, Chiu YL, Sun X, Goodman SM, Russell LA, et al. Rheumatoid arthritis vs osteoarthritis in patients receiving total knee arthroplasty: perioperative outcomes. J Arthroplast. 2014;29(2):308–13.CrossRef Stundner O, Danninger T, Chiu YL, Sun X, Goodman SM, Russell LA, et al. Rheumatoid arthritis vs osteoarthritis in patients receiving total knee arthroplasty: perioperative outcomes. J Arthroplast. 2014;29(2):308–13.CrossRef
81.
Zurück zum Zitat Kremers HM, Larson DR, Crowson CS, Kremers WK, Washington RE, Steiner CA, Jiranek WA, Berry DJ. Prevalence of total hip and knee replacement in the United States. J Bone Joint Surg Am. 2015;97(17):1386–97.PubMedCrossRef Kremers HM, Larson DR, Crowson CS, Kremers WK, Washington RE, Steiner CA, Jiranek WA, Berry DJ. Prevalence of total hip and knee replacement in the United States. J Bone Joint Surg Am. 2015;97(17):1386–97.PubMedCrossRef
82.
Zurück zum Zitat Kayani B, Konan S, Tahmassebi J, Pietrzak JRT, Haddad FS. Robotic-arm assisted total knee arthroplasty is associated with improved early functional recovery and reduced time to hospital discharge compared with conventional jig-based total knee arthroplasty: a prospective cohort study. Bone Joint J. 2018;100(7):930–7.PubMedCrossRef Kayani B, Konan S, Tahmassebi J, Pietrzak JRT, Haddad FS. Robotic-arm assisted total knee arthroplasty is associated with improved early functional recovery and reduced time to hospital discharge compared with conventional jig-based total knee arthroplasty: a prospective cohort study. Bone Joint J. 2018;100(7):930–7.PubMedCrossRef
Metadaten
Titel
Total knee arthroplasty using patient-specific instrumentation for osteoarthritis of the knee: a meta-analysis
verfasst von
Kazuha Kizaki
Ajaykumar Shanmugaraj
Fumiharu Yamashita
Nicole Simunovic
Andrew Duong
Vickas Khanna
Olufemi R. Ayeni
Publikationsdatum
01.12.2019
Verlag
BioMed Central
Erschienen in
BMC Musculoskeletal Disorders / Ausgabe 1/2019
Elektronische ISSN: 1471-2474
DOI
https://doi.org/10.1186/s12891-019-2940-2

Weitere Artikel der Ausgabe 1/2019

BMC Musculoskeletal Disorders 1/2019 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Aquatherapie bei Fibromyalgie wirksamer als Trockenübungen

03.05.2024 Fibromyalgiesyndrom Nachrichten

Bewegungs-, Dehnungs- und Entspannungsübungen im Wasser lindern die Beschwerden von Patientinnen mit Fibromyalgie besser als das Üben auf trockenem Land. Das geht aus einer spanisch-brasilianischen Vergleichsstudie hervor.

Endlich: Zi zeigt, mit welchen PVS Praxen zufrieden sind

IT für Ärzte Nachrichten

Darauf haben viele Praxen gewartet: Das Zi hat eine Liste von Praxisverwaltungssystemen veröffentlicht, die von Nutzern positiv bewertet werden. Eine gute Grundlage für wechselwillige Ärzte und Psychotherapeuten.

Proximale Humerusfraktur: Auch 100-Jährige operieren?

01.05.2024 DCK 2024 Kongressbericht

Mit dem demographischen Wandel versorgt auch die Chirurgie immer mehr betagte Menschen. Von Entwicklungen wie Fast-Track können auch ältere Menschen profitieren und bei proximaler Humerusfraktur können selbst manche 100-Jährige noch sicher operiert werden.