27.08.2019 | Ventricular Tachycardia

Accuracy of cardiac ablation catheter guidance by means of a single equivalent moving dipole inverse algorithm to identify sites of origin of cardiac electrical activation

verfasst von: Wener Lv, Kichang Lee, Tatsuya Arai, Conor D. Barrett, Maysun M. Hasan, Alison M. Hayward, Robert P. Marini, Maya E. Barley, Anna Galea, Gordon Hirschman, Antonis A. Armoundas, Richard J. Cohen

Erschienen in: Journal of Interventional Cardiac Electrophysiology | Ausgabe 3/2020

Einloggen, um Zugang zu erhalten

Abstract

We have developed a system that could potentially be used to identify the site of origin of ventricular tachycardia (VT) and to guide a catheter to that site to deliver radio-frequency ablation therapy. This system employs the Inverse Solution Guidance Algorithm based upon Single Equivalent Moving Dipole (SEMD) localization method. The system was evaluated in in vivo swine experiments. Arrays consisting of 9 or 16 bipolar epicardial electrodes and an additional mid-myocardial pacing lead were sutured to each ventricle. Focal tachycardia was simulated by applying pacing pulses to each epicardial electrode at multiple pacing rates during breath hold at the end-expiration phase. Surface potentials were recorded from 64 surface electrodes and then analyzed using the SEMD method to localize the position of the pacing electrodes. We found a close correlation between the locations of the pacing electrodes as measured in computational and real spaces. The reproducibility error of the SEMD estimation of electrode location was 0.21 ± 0.07 cm. The vectors between every pair of bipolar electrodes were computed in computational and real spaces. At 120 bpm, the lengths of the vectors in the computational and real space had a 95% correlation. Computational space vectors were used in catheter guidance simulations which showed that this method could reduce the distance between the real space locations of the emulated catheter tip and the emulated arrhythmia origin site by approximately 72% with each movement. We have demonstrated the feasibility of using our system to guide a catheter to the site of the emulated VT origin.
Literatur
1.
Zurück zum Zitat Stevenson WG, John RM. Ventricular arrhythmias in patients with implanted defibrillators. Circulation. 2011;124:e411–4.PubMed Stevenson WG, John RM. Ventricular arrhythmias in patients with implanted defibrillators. Circulation. 2011;124:e411–4.PubMed
2.
Zurück zum Zitat Yousuf O, Chrispin J, Tomaselli GF, Berger RD. Clinical management and prevention of sudden cardiac death. Circ Res. 2015;116:2020–40.CrossRef Yousuf O, Chrispin J, Tomaselli GF, Berger RD. Clinical management and prevention of sudden cardiac death. Circ Res. 2015;116:2020–40.CrossRef
3.
Zurück zum Zitat Connolly SJ, Dorian P, Roberts RS, Gent M, Bailin S, Fain ES, et al. Optimal pharmacological therapy in cardioverter defibrillator patients (OPTIC) investigators. Comparison of beta-blockers, amiodarone plus beta-blockers, or sotalol for prevention of shocks from implantable cardioverter defibrillators: the OPTIC Study: a randomized trial. JAMA. 2006;295:165–71.CrossRef Connolly SJ, Dorian P, Roberts RS, Gent M, Bailin S, Fain ES, et al. Optimal pharmacological therapy in cardioverter defibrillator patients (OPTIC) investigators. Comparison of beta-blockers, amiodarone plus beta-blockers, or sotalol for prevention of shocks from implantable cardioverter defibrillators: the OPTIC Study: a randomized trial. JAMA. 2006;295:165–71.CrossRef
4.
Zurück zum Zitat Daubert JP, Zareba W, Cannom DS, McNitt S, Rosero SZ, Wang P, et al. Inappropriate implantable cardioverter-defibrillator shocks in MADIT II: frequency, mechanisms, predictors, and survival impact. J Am Coll Cardiol. 2008;51:1357–65.CrossRef Daubert JP, Zareba W, Cannom DS, McNitt S, Rosero SZ, Wang P, et al. Inappropriate implantable cardioverter-defibrillator shocks in MADIT II: frequency, mechanisms, predictors, and survival impact. J Am Coll Cardiol. 2008;51:1357–65.CrossRef
5.
Zurück zum Zitat Serber ER, Sears SF, Sotile RO, Burns JL, Schwartzman DS, Hoyt RH, et al. Sleep quality among patients treated with implantable atrial defibrillation therapy: effect of nocturnal shock delivery and psychological distress. J Cardiovasc Electrophysiol. 2003;14:960–4.CrossRef Serber ER, Sears SF, Sotile RO, Burns JL, Schwartzman DS, Hoyt RH, et al. Sleep quality among patients treated with implantable atrial defibrillation therapy: effect of nocturnal shock delivery and psychological distress. J Cardiovasc Electrophysiol. 2003;14:960–4.CrossRef
6.
Zurück zum Zitat van Rees JB, Borleffs CJ, de Bie MK, Stijnen T, van Erven L, Bax JJ, et al. Inappropriate implantable cardioverter-defibrillator shocks: incidence, predictors, and impact on mortality. J Am Coll Cardiol. 2011;57:556–62.CrossRef van Rees JB, Borleffs CJ, de Bie MK, Stijnen T, van Erven L, Bax JJ, et al. Inappropriate implantable cardioverter-defibrillator shocks: incidence, predictors, and impact on mortality. J Am Coll Cardiol. 2011;57:556–62.CrossRef
7.
Zurück zum Zitat Delacretaz E, Stevenson WG. Catheter ablation of ventricular tachycardia in patients with coronary heart disease: part I: mapping. Pacing Clin Electrophysiol. 2001;24:1261–77.CrossRef Delacretaz E, Stevenson WG. Catheter ablation of ventricular tachycardia in patients with coronary heart disease: part I: mapping. Pacing Clin Electrophysiol. 2001;24:1261–77.CrossRef
8.
Zurück zum Zitat Strickberger SA, Man KC, Daoud EG, Goyal R, Brinkman K, Hasse C, et al. A prospective evaluation of catheter ablation of ventricular tachycardia as adjuvant therapy in patients with coronary artery disease and an implantable cardioverter-defibrillator. Circulation. 1997;96:1525–31.CrossRef Strickberger SA, Man KC, Daoud EG, Goyal R, Brinkman K, Hasse C, et al. A prospective evaluation of catheter ablation of ventricular tachycardia as adjuvant therapy in patients with coronary artery disease and an implantable cardioverter-defibrillator. Circulation. 1997;96:1525–31.CrossRef
9.
Zurück zum Zitat Tanawuttiwat T, Nazarian S, Calkins H. The role of catheter ablation in the management of ventricular tachycardia. Eur Heart J. 2016;37:594–609.CrossRef Tanawuttiwat T, Nazarian S, Calkins H. The role of catheter ablation in the management of ventricular tachycardia. Eur Heart J. 2016;37:594–609.CrossRef
10.
Zurück zum Zitat Dukkipati SR, Choudry S, Koruth JS, Miller MA, Whang W, Reddy VY. Catheter ablation of ventricular tachycardia in structurally normal hearts: indications, strategies, and outcomes-part I. J Am Coll Cardiol. 2017;70:2909–23.CrossRef Dukkipati SR, Choudry S, Koruth JS, Miller MA, Whang W, Reddy VY. Catheter ablation of ventricular tachycardia in structurally normal hearts: indications, strategies, and outcomes-part I. J Am Coll Cardiol. 2017;70:2909–23.CrossRef
11.
Zurück zum Zitat Dukkipati SR, Koruth JS, Choudry S, Miller MA, Whang W, Reddy VY. Catheter ablation of ventricular tachycardia in structural heart disease: indications, strategies, and outcomes-part II. J Am Coll Cardiol. 2017;70:2924–41.CrossRef Dukkipati SR, Koruth JS, Choudry S, Miller MA, Whang W, Reddy VY. Catheter ablation of ventricular tachycardia in structural heart disease: indications, strategies, and outcomes-part II. J Am Coll Cardiol. 2017;70:2924–41.CrossRef
12.
Zurück zum Zitat Stevenson WG, Wilber DJ, Natale A, Jackman WM, Marchlinski FE, Talbert T, et al. Irrigated radiofrequency catheter ablation guided by electroanatomic mapping for recurrent ventricular tachycardia after myocardial infarction: The Multicenter Thermocool Ventricular Tachycardia Ablation Trial. Circulation. 2008;118:2773–82. Stevenson WG, Wilber DJ, Natale A, Jackman WM, Marchlinski FE, Talbert T, et al. Irrigated radiofrequency catheter ablation guided by electroanatomic mapping for recurrent ventricular tachycardia after myocardial infarction: The Multicenter Thermocool Ventricular Tachycardia Ablation Trial. Circulation. 2008;118:2773–82.
13.
Zurück zum Zitat Cuculich PS, Schill MR, Kashani R, Mutic S, Lang A, Cooper D, et al. Noninvasive cardiac radiation for ablation of ventricular tachycardia. N Engl J Med. 2017;377:2325–36.CrossRef Cuculich PS, Schill MR, Kashani R, Mutic S, Lang A, Cooper D, et al. Noninvasive cardiac radiation for ablation of ventricular tachycardia. N Engl J Med. 2017;377:2325–36.CrossRef
14.
Zurück zum Zitat Stevenson WG, Friedman PL, Kocovic D, Sager PT, Saxon LA, Pavri B. Radiofrequency catheter ablation of ventricular tachycardia after myocardial infarction. Circulation. 1998;98:308–14.CrossRef Stevenson WG, Friedman PL, Kocovic D, Sager PT, Saxon LA, Pavri B. Radiofrequency catheter ablation of ventricular tachycardia after myocardial infarction. Circulation. 1998;98:308–14.CrossRef
15.
Zurück zum Zitat Miller MA, Dukkipati SR, Mittnacht AJ, Chinitz JS, Belliveau L, Koruth JS, et al. Activation and entrainment mapping of hemodynamically unstable ventricular tachycardia using a percutaneous left ventricular assist device. J Am Coll Cardiol. 2011;58:1363–71.CrossRef Miller MA, Dukkipati SR, Mittnacht AJ, Chinitz JS, Belliveau L, Koruth JS, et al. Activation and entrainment mapping of hemodynamically unstable ventricular tachycardia using a percutaneous left ventricular assist device. J Am Coll Cardiol. 2011;58:1363–71.CrossRef
16.
Zurück zum Zitat Josephson ME, Anter E. Substrate mapping for ventricular tachycardia: assumptions and misconceptions. JACC Clin Electrophysiol. 2015;1:341–52.CrossRef Josephson ME, Anter E. Substrate mapping for ventricular tachycardia: assumptions and misconceptions. JACC Clin Electrophysiol. 2015;1:341–52.CrossRef
17.
Zurück zum Zitat Sacher F, Lim HS, Derval N, Denis A, Berte B, Yamashita S, et al. Substrate mapping and ablation for ventricular tachycardia: the LAVA approach. J Cardiovasc Electrophysiol. 2015;26:464–71.CrossRef Sacher F, Lim HS, Derval N, Denis A, Berte B, Yamashita S, et al. Substrate mapping and ablation for ventricular tachycardia: the LAVA approach. J Cardiovasc Electrophysiol. 2015;26:464–71.CrossRef
18.
Zurück zum Zitat Komatsu Y. Substrate-based approach for ventricular tachycardia in structural heart disease: tips for mapping and ablation. J Arrhythm. 2014;30:272–82.CrossRef Komatsu Y. Substrate-based approach for ventricular tachycardia in structural heart disease: tips for mapping and ablation. J Arrhythm. 2014;30:272–82.CrossRef
19.
Zurück zum Zitat Furniss S, Anil-Kumar R, Bourke JP, Behulova R, Simeonidou E. Radiofrequency ablation of haemodynamically unstable ventricular tachycardia after myocardial infarction. Heart. 2000;84:648–52.CrossRef Furniss S, Anil-Kumar R, Bourke JP, Behulova R, Simeonidou E. Radiofrequency ablation of haemodynamically unstable ventricular tachycardia after myocardial infarction. Heart. 2000;84:648–52.CrossRef
20.
Zurück zum Zitat Armoundas AA, Feldman AB, Mukkamala R, Cohen RJ. A single equivalent moving dipole model: an efficient approach for localizing sites of origin of ventricular electrical activation. Ann Biomed Eng. 2003;31:564–76.CrossRef Armoundas AA, Feldman AB, Mukkamala R, Cohen RJ. A single equivalent moving dipole model: an efficient approach for localizing sites of origin of ventricular electrical activation. Ann Biomed Eng. 2003;31:564–76.CrossRef
21.
Zurück zum Zitat Armoundas AA, Feldman AB, Mukkamala R, He B, Mullen TJ, Belk PA, et al. Statistical accuracy of a moving equivalent dipole method to identify sites of origin of cardiac electrical activation. IEEE Trans Biomed Eng. 2003;50:1360–70.CrossRef Armoundas AA, Feldman AB, Mukkamala R, He B, Mullen TJ, Belk PA, et al. Statistical accuracy of a moving equivalent dipole method to identify sites of origin of cardiac electrical activation. IEEE Trans Biomed Eng. 2003;50:1360–70.CrossRef
22.
Zurück zum Zitat Barley ME, Armoundas AA, Cohen RJ. A method for guiding ablation catheters to arrhythmogenic sites using body surface electrocardiographic signals. IEEE Trans Biomed Eng. 2009;56:810–9.CrossRef Barley ME, Armoundas AA, Cohen RJ. A method for guiding ablation catheters to arrhythmogenic sites using body surface electrocardiographic signals. IEEE Trans Biomed Eng. 2009;56:810–9.CrossRef
23.
Zurück zum Zitat Fukuoka Y, Oostendorp TF, Sherman DA, Armoundas AA. Applicability of the single equivalent moving dipole model in an infinite homogeneous medium to identify cardiac electrical sources: a computer simulation study in a realistic anatomic geometry torso model. IEEE Trans Biomed Eng. 2006;53:2436–44.CrossRef Fukuoka Y, Oostendorp TF, Sherman DA, Armoundas AA. Applicability of the single equivalent moving dipole model in an infinite homogeneous medium to identify cardiac electrical sources: a computer simulation study in a realistic anatomic geometry torso model. IEEE Trans Biomed Eng. 2006;53:2436–44.CrossRef
24.
Zurück zum Zitat Lee K, Lv W, Ter-Ovanesyan E, Barley ME, Voysey GE, Galea AM, et al. Cardiac ablation catheter guidance by means of a single equivalent moving dipole inverse algorithm. Pacing Clin Electrophysiol. 2013;36:811–22.CrossRef Lee K, Lv W, Ter-Ovanesyan E, Barley ME, Voysey GE, Galea AM, et al. Cardiac ablation catheter guidance by means of a single equivalent moving dipole inverse algorithm. Pacing Clin Electrophysiol. 2013;36:811–22.CrossRef
25.
Zurück zum Zitat Sohn K, Lv W, Lee K, Galea A, Hirschman G, Barrett C, et al. A method to noninvasively identify cardiac bioelectrical sources. Pacing Clin Electrophysiol. 2014;37:1038–50.CrossRef Sohn K, Lv W, Lee K, Galea A, Hirschman G, Barrett C, et al. A method to noninvasively identify cardiac bioelectrical sources. Pacing Clin Electrophysiol. 2014;37:1038–50.CrossRef
26.
Zurück zum Zitat Barley M. Bioelectrical strategies for image-guided therapies. Harvard-MIT Division of Health Sciences and Technology, Massachusetts Institute of Technology; 2007 Barley M. Bioelectrical strategies for image-guided therapies. Harvard-MIT Division of Health Sciences and Technology, Massachusetts Institute of Technology; 2007
27.
Zurück zum Zitat Rosbury TS. Computer simulation of a novel technique for radio-frequency ablation of ventricular arrhythmias. Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology; 2006. Rosbury TS. Computer simulation of a novel technique for radio-frequency ablation of ventricular arrhythmias. Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology; 2006.
28.
Zurück zum Zitat Sapp JL, Bar-Tal M, Howes AJ, Toma JE, El-Damaty A, Warren JW, et al. Real-time localization of ventricular tachycardia origin from the 12-lead electrocardiogram. JACC Clin Electrophysiol. 2017;3:687–99.CrossRef Sapp JL, Bar-Tal M, Howes AJ, Toma JE, El-Damaty A, Warren JW, et al. Real-time localization of ventricular tachycardia origin from the 12-lead electrocardiogram. JACC Clin Electrophysiol. 2017;3:687–99.CrossRef
29.
Zurück zum Zitat Svehlikova J, Teplan M, Tysler M. Geometrical constraint of sources in noninvasive localization of premature ventricular contractions. J Electrocardiol. 2018;51:370–7.CrossRef Svehlikova J, Teplan M, Tysler M. Geometrical constraint of sources in noninvasive localization of premature ventricular contractions. J Electrocardiol. 2018;51:370–7.CrossRef
Metadaten
Titel
Accuracy of cardiac ablation catheter guidance by means of a single equivalent moving dipole inverse algorithm to identify sites of origin of cardiac electrical activation
verfasst von
Wener Lv
Kichang Lee
Tatsuya Arai
Conor D. Barrett
Maysun M. Hasan
Alison M. Hayward
Robert P. Marini
Maya E. Barley
Anna Galea
Gordon Hirschman
Antonis A. Armoundas
Richard J. Cohen
Publikationsdatum
27.08.2019
Verlag
Springer US
Erschienen in
Journal of Interventional Cardiac Electrophysiology / Ausgabe 3/2020
Print ISSN: 1383-875X
Elektronische ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-019-00605-z

Weitere Artikel der Ausgabe 3/2020

Journal of Interventional Cardiac Electrophysiology 3/2020 Zur Ausgabe

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

Niedriger diastolischer Blutdruck erhöht Risiko für schwere kardiovaskuläre Komplikationen

25.04.2024 Hypotonie Nachrichten

Wenn unter einer medikamentösen Hochdrucktherapie der diastolische Blutdruck in den Keller geht, steigt das Risiko für schwere kardiovaskuläre Ereignisse: Darauf deutet eine Sekundäranalyse der SPRINT-Studie hin.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Adipositas-Medikament auch gegen Schlafapnoe wirksam

24.04.2024 Adipositas Nachrichten

Der als Antidiabetikum sowie zum Gewichtsmanagement zugelassene Wirkstoff Tirzepatid hat in Studien bei adipösen Patienten auch schlafbezogene Atmungsstörungen deutlich reduziert, informiert der Hersteller in einer Vorab-Meldung zum Studienausgang.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.