Open Access 01.12.2017 | Research article

Volatile organic compounds in ventilated critical care patients: a systematic evaluation of cofactors

Erschienen in: BMC Pulmonary Medicine | Ausgabe 1/2017

Abstract

Background

Expired gas (exhalome) analysis of ventilated critical ill patients can be used for drug monitoring and biomarker diagnostics. However, it remains unclear to what extent volatile organic compounds are present in gases from intensive care ventilators, gas cylinders, central hospital gas supplies, and ambient air. We therefore systematically evaluated background volatiles in inspired gas and their influence on the exhalome.

Methods

We used multi-capillary column ion-mobility spectrometry (MCC-IMS) breath analysis in five mechanically ventilated critical care patients, each over a period of 12 h. We also evaluated volatile organic compounds in inspired gas provided by intensive care ventilators, in compressed air and oxygen from the central gas supply and cylinders, and in the ambient air of an intensive care unit. Volatiles detectable in both inspired and exhaled gas with patient-to-inspired gas ratios < 5 were defined as contaminating compounds.

Results

A total of 76 unique MCC-IMS signals were detected, with 39 being identified volatile compounds: 73 signals were from the exhalome, 12 were identified in inspired gas from critical care ventilators, and 34 were from ambient air. Five volatile compounds were identified from the central gas supply, four from compressed air, and 17 from compressed oxygen. We observed seven contaminating volatiles with patient-to-inspired gas ratios < 5, thus representing exogenous signals of sufficient magnitude that might potentially be mistaken for exhaled biomarkers.

Conclusions

Volatile organic compounds can be present in gas from central hospital supplies, compressed gas tanks, and ventilators. Accurate assessment of the exhalome in critical care patients thus requires frequent profiling of inspired gases and appropriate normalisation of the expired signals.

Background

Multi-capillary column ion-mobility spectrometry (MCC-IMS) can be used for real-time clinical breath analysis [1]. Volatile organic compounds in expired gases (exhalome) are linked to physiological processes and various diseases [2, 3]. It may also be possible to estimate plasma drug concentrations from the exhalome [4]. For example, there is great potential to be able to diagnose lung cancer, chronic obstructive pulmonary disease (COPD), lung infections, and renal failure which all need to be confirmed in clinical trials [3]. Exhalome analysis may also facilitate early detection of inflammation and sepsis — although this application has so far only been evaluated in rats [5].
Accurate assessment of volatile organic compounds in expired gas requires either that none be present in inspired gas or that inspired concentrations are measured and subtracted from the raw expired signal. Potential sources of exogenous volatile organic compounds include gas supplies (central hospital, compressed cylinders, ambient air) and ventilators.
Our concern was prompted by a previous study in which there was substantial contamination of inspired gas [6], including inhaled volatile compounds that were subsequently exhaled unchanged and might thus have been mistaken for biomarkers. We therefore evaluated volatile organic compounds in ambient air in a critical care unit, in gases from the central hospital supply as well as in compressed gas cylinders and from critical care ventilators.

Methods

Patients

With approval by the responsible ethics committee (Ärztekammer Saarland, Saarbrücken, Germany Ref-No. 232/14), five sedated and mechanically ventilated adults were each evaluated for 12 h. Legal guardians or the patients themselves subsequently agreed in written form to participate in this study.
Patients were ventilated with an intensive care respirator (EVITA 4, Dräger, Lübeck, Germany) with ventilation parameters and oxygen concentrations adjusted to maintain physiological partial pressures of carbon dioxide and oxygen. They were ventilated with either pressure-supported or pressure-controlled modes. A MCC-IMS aspiration tube was connected to the endotracheal tube distal to a heat-and-moisture exchanging filter (Humid-Vent Filter Compact S, Teleflex Medical, Athlone, Ireland) by a polytetrafluoroethylene tube (Bohlender, Grünsfeld, Germany). Samples were aspirated from the breathing circuit at 30-min intervals.

Ventilators and gas sources

We evaluated three different isolated EVITA 4 ventilators (Dräger, Lübeck, Germany), each provided by gas (oxygen and compressed air) from central hospital supplies. Each ventilated an artificial lung at a minute volume of eight litres per minute for 12 h with an oxygen fraction of 21%. Gas was sampled at 30-min intervals from tubing connected to a heat-and-moisture exchanging filter which was connected to the outlet port of the ventilator.
We also evaluated oxygen and compressed air provided by our hospital’s central gas supply. Gases were passed through a pressure regulator into polytetrafluoroethylene tubing and then into a five-litre glass bottle at a rate of 1.0–1.2 l per minute. To avoid contamination of ambient air, the glass container was sealed towards the outside and flushed with oxygen or compressed air for 60 min before each set of measurements. Oxygen and air from the hospital central supply system were each evaluated from three different wall distribution stations. We also evaluated compressed oxygen from three different cylinders using the method mentioned above. In each case, samples of gas from the glass bottles were taken at 30-min intervals for 12 h.
And finally, we evaluated ambient air in an ICU on three distinct days, each for 24 h with a sampling interval of 30 min.

Analysis of volatile organic compounds

Volatile organic compounds in all gas samples were evaluated as described previously using a BreathDiscovery MCC-IMS (B&S Analytik, Dortmund, Germany) [6, 7]. Briefly, 10-ml gas samples were analysed in multicapillary columns which evaluated compound retention times (RT) and combined with ion-mobility spectrometry also evaluated drift times. IMS-Peaks with an intensity of more than 5 mV in at least three consecutive measurements were included. Volatile organic compounds were thus characterised by their retention times and drift times which identify specific compounds, and peak intensity which is a function of concentration.
Specific volatile compounds were identified using the software Visual Now 3.6 (B&S Analytik, Dortmund, Germany) by automated alignment software (MIMA, version 1.1.2) with an existing database (BS-MCC/IMS-analytes database, version 1209, B&S Analytik, Dortmund, Germany) [8]. Peak area overlapping of at least 10% with preexisting reference substance in chromatogram defined alignment. If overlapping areas of two eligible compounds differed less than 10% in extent, the alternative compound was designated as well. Unknown volatiles were designated only by unique peak numbers. We performed calibrations for the volatiles acetone, cyclohexanone, dimethyl disulphide, 3-hydroxy-2-butanone, 2-methylfuran, 2-methylpentane, and 3-pentanone using exponential dilution technique with a 5.9 l glass bottle as described previously [9].
Intensities of different volatile organic compounds were expressed as means (± 95% confidence interval) of the relevant sampling periods. As in a previous study, we classified volatiles as expired (detected only in expired gas or having a patient-to-inspired gas ratio > 1.5), unaffected (having a patient-to-inspired gas ratio between 0.5–1.5), and resorbed (having a patient-to-inspired gas ratio < 0.5) [6]. Gases with patient-to-inspired gas ratios < 5 were considered clinically important because they might be mistaken for de novo expired compounds.

Results

A total of 76 different signals were detected by MCC-IMS of which 48 were identified. Volatile compounds and their peak numbers in the IMS chromatogram, CAS number, class, and chemical identification (when known) are shown in Table 1. Seven peaks aligned with each two eligible reference substances. Table 2 shows intensities of all detected volatiles according to different sampling points. Concentrations in ppb of selected compounds are summarized in Table 3. Figure 1 displays the occurrence and intersecting sets of all signals.
Table 1
Peak number in the IMS chromatogram, volatile organic compound, CAS number (Chemical Abstracts Service), class and occurrence of chemical substance; (*) = alternative volatile organic compound; P49 – P76 are “unknown” signals and are not displayed
 
Chemical Substance
IMS-Peak
Volatile Organic Compound
CAS number
Class of chemical substance
Occurrence
P1
Acetone monomer
67–64-1
Ketones
Synthesis with a raw material, solvents, adhesives
P2
Acetone dimer
67–64-1
Ketones
Synthesis with a raw material, solvents, adhesives
P3
Benzofuran
271–89-6
Aromatics
Tabacco smoke, synthesis chemicals
P4
Butanal monomer (1-Butanol*)
123–72-8
Aldehydes
Artificial resin, plasticizer
P5
Butanal dimer
123–72-8
Aldehydes
Artificial resin, plasticizer
P6
1,2-Butandiol
584–03-2
Alcohol
Solvents, epoxy resins
P7
2,3-Butandiol
513–85-9
Alcohols
Solvents, plasticizers, epoxy resins, toiletries
P8
2-Butanone
78–93-3
Ketones
Solvents, plastics, sterilization of medical products
P9
(+)Camphene
79–92-5
Terpenes
Ethereal oils
P10
Cyclohexanol monomer (3-Heptanon*)
108–93-0
Alcohols
Solvents
P11
Cyclohexanol dimer
108–93-0
Alcohols
Solvents
P12
Cyclohexanone monomer
108–94-1
Ketones
Solvents
P13
Cyclohexanone dimer
108–94-1
Ketones
Solvents
P14
p-Cymol
99–87-6
Terpenes
plants
P15
Dimethyl disulphide monomer
624–92-0
Disulphide
Flavouring
P16
Dimethyl disulphide dimer
624–92-0
Disulphide
Flavouring
P17
2,5-Dimethylpyrazin
123–32-0
Azine
Food, flavouring
P18
Ethanol
64–17-5
Alcohols
Fermentation, disinfectant, solvents
P19
Ethylbenzene
100–41-4
Aromatics
Solvents, plastics, lacquers
P20
2-Ethyl-1-hexanol
104–76-7
Alcohols
Solvents, intermediates
P21
Heptanal
111–71-7
Aldehydes
Intermediates, odor agents
P22
2-Heptanone
110–43-0
Ketones
High boiling solvents, coating material
P23
3-Heptanone (4-Heptanone*)
106–35-4
Ketones
Solvents
P24
Hexanal
66–25-1
Aldehydes
Lipid peroxidation of unsaturated fatty acids
P25
1-Hexanol
111–27-3
Alcohols
Solvents, plasticizer
P26
2-Hexanol
626–93-7
Alcohols
Solvents
P27
2-Hexanon (Hexanal*)
591–78-6
Ketones
Solvents
P28
3-Hydroxy-2-Butanone
513–86-0
Ketones
Bacteria, tobacco smoke
P29
Isoprene monomer
78–79-5
Terpenes
rubber
P30
Isoprene dimer
78–79-5
Terpenes
rubber
P31
Menthone
10,458–14-7
Ketones
Ethereal oils
P32
Methanol
67–56-1
Alcohols
Solvents, synthesis with a raw material
P33
3-Methylbutanal
590–86-3
Aldehydes
Drug substances, vitamins, solvents, plasticizers
P34
2-Methylbutylacetat (Hexanal*)
624–41-9
Acetic Esters
Solvents, flavouring
P35
2-Methylfuran
534–22-5
Furans
Tobacco smoke
P36
2-Methylpentane
107–83-5
Hexane
Solvents, cleaning agents
P37
n-Nonane
111–84-2
Alkanes
Fuels, Entrainer, detergent substances
P38
2,2,4,6,6-Pentamethylheptane
236–757-0
Alkanes
Solvents, cleaning agents
P39
1-Pentanol (Cyclohexanol*)
71–41-0
Alcohols
Solvents, cleaning agents, disinfectant
P40
2-Pentanone
107–87-9
Ketones
Solvents
P41
3-Pentanone monomer
96–22-0
Ketones
Solvents
P42
3-Pentanone dimer
96–22-0
Ketones
Solvents
P43
Phenylacetylene monomer (Dimethyl disulphide*)
536–74-3
Alkynes
Plastics
P44
Phenylacetylene dimer
536–74-3
Alkynes
Plastics
P45
1-Propanol
71–23-8
Alcohols
Solvents, disinfectant, cleaning agents
P46
2-Propanol monomer
67–63-0
Alcohols
Solvents, cleaning agents, disinfectant
P47
2-Propanol dimer
67–63-0
Alcohols
Solvents, cleaning agents, disinfectant
P48
Propofol
2078–54-8
Phenol
Anaesthetic
Table 2
Peak intensities of all detected volatile organic compounds detected in gas from designated sources
IMS-Peak
Volatile Organic Compound
Patient
Inspiration
O2
Compressed Air
O2 Cylinder
Room Air
P1 *
Acetone monomer
71
(60.1–81.9)
6
(4.8–7.3)
        
P2 *
Acetone dimer
137
(91–183)
      
5
(3.4–6.6)
12.1
(10.6–13.6)
P3 *
Benzofuran
4.9
(3.2–6.5)
      
2.7
(2–3.4)
  
P4 *
Butanal monomer
7.6
(6.2–9)
          
P5 *
Butanal dimer
4.6
(3.8–5.3)
          
P6 *
1.2-Butandiol
5
(4.2–5.8)
          
P7 *
2.3-Butandiol
6.9
(5.9–7.8)
          
P8 *
2-Butanone
15.2
(10.5–19.9)
      
25.8
(14–37.6)
3.2
(2.8–3.6)
P9 *
(+)Camphene
8.8
(7.2–10.4)
          
P10 *
Cyclohexanol monomer
17.7
(12.3–23.1)
        
10.3
(9.2–11.4)
P11 *
Cyclohexanol dimer
8.9
(7.8–10)
          
P12 *
Cyclohexanone monomer
175
(160–190)
        
9.7
(9.3–10.1)
P13 *
Cyclohexanone dimer
4
(3.4–4.6)
          
P14 *
p-Cymol
4
(3.1–4.8)
        
4
(3.8–4.3)
P15 *
Dimethyl disulphide monomer
88
(68.9–107)
10
(8.7–11.3)
34.2
(32.8–35.6)
8.8
(7.5–10)
18.3
(16.2–20.4)
44.4
(42.3–46.5)
P16 *
Dimethyl disulphide dimer
7.6
(6.3–8.9)
        
12.7
(12–13.4)
P17 *
2.5-Dimethylpyrazin
98.3
(58.9–138)
        
7.6
(7–8.2)
P18 *
Ethanol
5.5
(2.9–8.2)
      
9.1
(6.6–11.7)
26.7
26.7
P19 *
Ethylbenzene
35.8
(22–49.6)
      
6.2
(3.8–8.6)
12.2
(11.2–13.2)
P20
2-Ethyl-1-hexanol
          
4.1
(3.9–4.4)
P21 *
Heptanal
9.9
(9–10.9)
          
P22 *
2-Heptanone
13.5
(11.9–15.1)
          
P23 *
3-Heptanone
20.4
(18.2–22.6)
          
P24 *
Hexanal
7.4
(6.8–8)
          
P25 *
1-Hexanol
10.5
(8.8–12.2)
          
P26 *
2-Hexanol
16.2
(13.4–19)
          
P27 *
2-Hexanone
7.1
(6.3–7.9)
          
P28 *
3-Hydroxy-2-Butanone
66.8
(43.7–89.9)
      
1.9
(1.5–2.2)
86.6
(76.6–96.6)
P29 §
Isoprene monomer
  
8.1
(7.7–8.5)
    
54.2
(31.2–77.2)
10.6
(8.8–12.4)
P30 *
Isoprene dimer
4.5
(3.7–5.3)
        
3.6
(3.3–3.8)
P31 *
Menthone
14.5
(12.9–16.1)
          
P32 *
Methanol
107
(102–112)
45.9
(36.8–55)
98.2
(96.7–99.7)
87.4
(85.3–89.5)
101
(97.2–105)
98
(97–99)
P33 *
3-Methylbutanal
11.2
(9.3–13.1)
          
P34 *
2-Methylbutylacetat
7.1
(6.3–7.8)
          
P35 *
2-Methylfuran
26.2
(15.3–37.1)
3.9
(3.5–4.4)
    
3.4
(2.3–4.5)
13.8
(11.9–15.7)
P36 *
2-Methylpentane
15.3
(10.4–20.2)
6
(5.2–6.8)
    
51.3
(29.2–73.4)
27.8
(24.1–31.5)
P37 *
n-Nonane
83.4
(66.5–100)
          
P38 *
2.2.4.6.6-Pentamethylheptan
7.7
(6.3–9)
          
P39 *
1-Pentanol
17.9
(12.3–23.5)
          
P40 *
2-Pentanone
130
(86.8–173)
      
5.7
(4–7.5)
3.8
(3.5–4.1)
P41 *
3-Pentanone monomer
230
(167–294)
        
35.5
(31.7–39.3)
P42 *
3-Pentanone dimer
21.1
(20–22.2)
        
5.6
(5.5–5.7)
P43 *
Phenylacetylene monomer
96.9
(79.2–115)
        
5
(4.7–5.3)
P44 *
Phenylacetylene dimer
19.4
(15.3–23.5)
        
6.1
(5.8–6.5)
P45 *
1-Propanol
5.8
(4.8–6.8)
          
P46 *
2-Propanol monomer
22
(17.2–26.8)
      
11.4
(6.8–16)
44.9
(38.8–51)
P47 *
2-Propanol dimer
4.7
(4–5.5)
          
P48 *
Propofol
23.2
(19.6–26.8)
          
P49 *
 
36.2
(33.1–39.3)
    
4.5
(4.1–5)
  
24.3
(24–24.6)
P50 *
 
43.6
(34.6–52.6)
17.5
(15.4–19.7)
6.1
(4.6–7.6)
5.1
(3.8–6.5)
2.8
(1.7–3.8)
2.7
(2.1–3.2)
P51 *
 
20.9
(14.5–27.3)
12.2
(10.2–14.2)
4.9
(3.7–6)
4.3
(3.2–5.3)
0.9
(0.2–1.6)
2.1
(1.7–2.4)
P52 *
 
15.8
(12.3–19.4)
        
9.1
(8.3–10)
P53 *
 
13.8
(12.9–14.7)
          
P54 *
 
4.6
(3.9–5.3)
          
P55 *
 
43.5
(37.8–49.2)
        
3.1
(2.3–4)
P56 *
 
7.1
(5.7–8.5)
          
P57 *
 
9.9
(6.9–13)
          
P58 *
 
2.4
(2–2.7)
          
P59 *
 
3.5
(1.2–5.7)
        
3.7
(3.5–4)
P60 *
 
4.7
(4.1–5.4)
          
P61 *
 
5.3
(3.6–7)
          
P62 *
 
2.8
(2.3–3.3)
          
P63 *
 
3.7
(3–4.3)
          
P64 *
 
5.1
(3.9–6.4)
          
P65 *
 
5.9
(3.6–8.3)
        
8.4
(7.6–9.2)
P66 *
 
6.2
(4–8.3)
3.2
(2.8–3.7)
      
4.8
(4.2–5.3)
P67 #
 
3.7
(1.1–6.2)
3.5
(3.1–4)
    
4.9
(3.5–6.3)
5.8
(5.1–6.5)
P68 *
 
1.8
(1.3–2.2)
          
P69 *
 
4
(2.6–5.3)
          
P70 *
 
8
(3.4–12.6)
          
P71 *
 
2.1
(1.5–2.6)
          
P72 *
 
15.6
(13.9–17.3)
3.1
(2.5–3.6)
        
P73 *
 
4.6
(2.3–6.8)
      
3.7
(2.6–4.8)
4.5
(4.2–4.8)
P74 #
 
2.4
(2–2.8)
3.1
(2.6–3.6)
        
P75
           
8.7
(8.1–9.2)
P76 *
 
8.6
(6.8–10.5)
          
Results are shown as means (95% confidence interval) in millivolt [mV]; Peak 49–76 represents “unknown” compounds; * = expired compounds, # = unaffected compounds, § = resorbed compounds
Table 3
Peak intensities and corresponding concentration [ppb] of selected volatile organic compounds
Volatile Organic Compound
Patient
Inspiration
O2
Compressed Air
O2 Cylinder
Room Air
mV
ppb
mV
ppb
mV
ppb
mV
ppb
mV
ppb
mV
ppb
Acetone
345 (242–448)
9.9 (7.7–12.3)
6 (4.8–7.3)
3.6 (3.6–3.6)
    
10 (6.8–13.2)
3.7 (3.6–3.7)
24.2 (21.2–27.2)
3.9 (3.8–3.9)
Cyclohexanone
183 (167–199)
6.4 (5.6–7.3)
        
9.7 (9.3–10.1)
0.3 (0.2–0,3)
Dimethyl disulphide
103 (82–125)
69 (57–82)
10 (8.7–11.3)
17.5 (16.8–18.1)
34.2 (32.8–35.6)
30.4 (29.6–31.1)
8.8 (7.5–10)
16.9 (16.2–17.5)
18.3 (16.2–20.4)
21.9 (20.8–23)
69.8 (66.3–73.3)
49.9 (48–51.9)
3-Hydroxy-2-Butanone
66.8 (43.7–89.9)
< 0.01 ppb
      
1.9 (1.5–2.2)
< 0.01 ppb
86.6 (76.6–96.6)
< 0.01 ppb
2-Methylfuran
26.2 (15.3–37.1)
< 0.01 ppb
3.9 (3.5–4.4)
< 0.01 ppb
    
3.4 (2.3–4.5)
< 0.01 ppb
13.8 (11.9–15.7)
< 0.01 ppb
2-Methylpentane
15.3 (10.4–20.2)
0.9 (0.1–1.8)
6 (5.2–6.8)
< 0.01 ppb
    
51.3 (29.2–73.4)
7.5 (3.4–12.1)
27.8 (24.1–31.5)
3.1 (2.5–3.8)
3-Pentanone
272 (207–338)
< 0.01 ppb
        
46.7 (42.7–50.7)
< 0.01 ppb
Results are shown as means (95% confidence interval) in millivolt [mV] and [ppb]. Total intensity of volatile organic compound was calculated on the basis of monomer and double dimer intensity
Participating patients had a mean age of 61 [± 16 SD] years, weight of 80.6 [± 16 SD] kg, and height of 172 [± 13 SD] cm. A total of 73 peaks were detected from patients, whereas individual measurements showed 44, 45, 55, 58, and 59 signals, respectively. 36 peaks were identified in the exhalome of all patients, whereas 14 signals were seen in but a single patient (Table 4).
Table 4
Main diagnosis, anaesthetics, and volatile organic compounds in the exhalome of five critical care patients
 
Patient 1
59 VOCs
Patient 2
58 VOCs
Patient 3
55 VOCs
Patient 4
45 VOCs
Patient 5
44 VOCs
Diagnosis
Haemorrhagic shock, Peripartum atonic bleeding
Sepsis, Perforated sigmoid diverticulitis
Polytrauma, Brain injury
Sepsis, Mamma carcinoma
Femur fracture, Respiratory insufficiency
Sedation
Propofol and Remifentanil
Piritramide
Propofol and Remifentanil
Propofol and Remifentanil
Propofol and Remifentanil
VOCs in
P14, P30, P58, P62, P68
P13, P18, P67, P74
P69, P70, P71, P73
 
P61
1 patient
 
(n = 14)
 
VOCs in 2 patients (n = 9)
  
P54
 
P45, P57
 
P45, P57
  
P3, P6, P47, P59, P60, P63
   
VOCs in 3 patients (n = 7)
P5
 
P5
P5
 
 
P11
 
P11
P11
 
P38
 
 
P65, P66
P65, P66
 
P65, P66
P51, P64
  
VOCs in 4 patients (n = 7)
P24
  
P48
 
P48
P48
P48
P4, P16
P4, P16
P4, P16
 
P4, P16
P17, P35, P76
 
VOCs in all patients (n = 36)
P1, P2, P7, P8, P9, P10, P12, P15, P19, P21, P22, P23, P25,
P26, P27, P28, P31, P32, P33, P34, P36, P37, P39, P40, P41,
P42, P43, P44, P46, P49, P50, P52, P53, P55, P56, P72
36 compounds are detectable in all patients, respectively. Other volatiles are merely seen in 1, 2, 3, or 4 patients
Inspiratory gas supplied by an intensive care respirator yielded 12 distinct signals without distinction amongst the three tested ventilators. There were 4 peaks detected in oxygen and 5 in air from the hospital’s central gas supply at each tested distribution point. Oxygen from cylinders revealed 17 signals. Ambient air from the intensive care unit yielded 34 unique signals.
All detectable signals in oxygen from the central gas supply (dimethyl disulphide monomer, methanol, and two unknown compounds) were found in inhaled and exhaled gas as well as in compressed and room air. There were also 31 out of 34 signals from ambient air detectable in the exhalome of patients. 2-ethyl-1-hexanol and P75 were seen only in room air. The only volatile organic compound identified in inspired but not in expired gas (resorbed compound) was a monomer of isoprene. In contrast, isoprene dimer was merely seen in exhaled gas and in only one patient.
Eleven of the twelve inhaled volatiles were exhaled as well. Two unknown compounds (P67 and P74) were expired in similar concentrations to inspired gas and therefore designated as unaffected compounds; that is, they had expired-to-inspired peak intensities of 1.1 for P67 and 0.8 for P74. On the other hand, nine signals were seen at greater intensities in expired than in inspired gas and therefore termed expired compounds (acetone monomer, dimethyl disulphide monomer, methanol, 2-methylfuran, 2-methylpentan and four unknown compounds). In these 11 peaks, we recorded intensities between 3.1 (P72 and P74) and 45 mV (Methanol) for inhaled and peak intensities between 2.4 (P74) and 107 mV (Methanol) for exhaled gas. The derived patient-to-inspired gas ratios ranged from 0.8 (unaffected compounds) to 11.8 (expired compounds) and are summarised in Table 5. We observed 7 contaminating volatiles with patient-to-inspired gas ratios < 5.
Table 5
Volatile organic compounds detectable in inhaled and exhaled gas, derived patient-to-inspired gas ratios and corresponding classification into unaffected volatiles (patient-to-inspired gas ratio 0.5–1.5), and expired volatiles (patient-to-inspired gas ratio > 1.5)
IMS-Peak
Volatile Organic Compound
Patient-to-Inspired Gas Ratio
Classification
Retention Time
Drift Time
P74
Unknown
0.8
unaffected
41.5
0.697
P67
Unknown
1.1
unaffected
17.8
0.506
P51
Unknown
1.7
expired
38.7
0.497
P66
Unknown
1.9
expired
21.5
0.506
P32
Methanol
2.3
expired
0.0
0.478
P50
Unknown
2.5
expired
20.3
0.496
P36
2-Methylpentane
2.6
expired
9.1
0.510
P72
Unknown
5.0
expired
42.8
0.701
P35
2-Methylfurane
6.7
expired
4.8
0.540
P15
Dimethyl disulphide monomer
8.8
expired
8.0
0.498
P1
Acetone monomer
11.8
expired
3.0
0.498
The retention time (RT) and drift time (1/K0) describe the position of the peaks in the IMS-chromatogram. 7 volatile organic compounds yielded patient-to-inspired gas ratios <5 and were therefore designated as contaminants in expired air
There were a total of 71 expired signals (Table 2). Examples of three-dimensional ion-mobility spectrometry chromatograms are shown in Fig. 2 for the exhalome from patients, gas from ventilators, oxygen from the hospital’s central gas supply, and ambient air of the intensive care unit. Figure 3 compares inspired and expired gas in a typical patient.

Discussion

There was significant variation in volatiles detected under various sampling conditions, but all gas sources were contaminated to some degree; specifically, we identified 17 signals from compressed oxygen cylinders, 12 signals from mechanical ventilators, 4 signals in oxygen from the central gas supply, and 5 signals in compressed air from the central gas supply. The central gas supply accounted for 4 signals found in ventilator gas. It can be presumed that volatile organic compounds in inhaled air originate from air being used for manufacturing oxygen and compressed air or derive from piping, seals, or respirator. Interestingly, nine of the 12 peaks detected in ventilator gases were detectable in room air as well — although no ambient air is supposed to be drawn into our ventilators. How these compounds got into ventilator gases remains unclear. Thirty-one of 34 detected signals in room air were also detectable in patients’ exhalome. Most likely, extubated patients, visitors, and staff exhale these compounds into the ambient atmosphere. However, it should be noted that we detected monomers and dimers in our sample collection. Thus, aforementioned unknown signals might include monomers and dimers, reducing the reported amount of unknown volatile organic compounds.
Analysis of the exhalome in ventilated critically ill patients shows promising approaches in detecting biomarkers, especially in the field of lung infections, to some extent ventilator associated and a common healthcare problem. Fowler et al. showed that volatiles in expired air of intubated and ventilated patients were able to classify breath profiles of patients with and without significant pathogen load in the lower respiratory tract [10]. Schnabel and colleagues detected 12 volatiles in ventilated critically ill patients that correctly discriminated between ventilator-associated pneumonia and the control group with high sensitivity and specificity [11]. Nevertheless, both authors did not evaluate background contamination sufficiently to prevent confounding by exogenous volatiles. Filipiak et al. demonstrated that appearance and concentration profile of pathogen-derived metabolites in expired air of ventilated patients with confirmed ventilator-associated pneumonia correlates with the presence of a particular pathogen [12]. They stated that several hours of continuous ventilation prevents bias of confounding exogenous contaminants like plastic-derived substances from tubing and ventilator by decreasing the concentration to levels prior to ventilation [1214]. Risby et al. stated that continuous ventilation with pure gas mixture prior to breath sampling was an effective method for elimination of exogenous compounds from exhaled air [15]. In our opinion, both approaches exhibit methodological deficiencies: evaporation of volatiles from tubing and ventilator is impossible to anticipate. Furthermore, pure gas mixture from central gas supplies is not available as our results show. Gao et al. defined VOC profile being able to distinguish between lower respiratory tract infection, colonisation, and absence of acinetobacter baumannii pathogens in ventilated critically ill patients. However, they did not simultaneously detect inspired gas for background correction [16]. Regardless of the subject, normalisation of expired gas for inspired confounders is mandatory.
In expired air of ventilated critically ill patients, we reported several volatile organic compounds with notable intensities: acetone, 3-pentanone, cyclohexanone, and 3-hydroxy-2-butanone. Acetone monomer and dimer are among the main components of exhaled breath. Acetone dimer was also detected in ambient air which is consistent with Bessonneau and colleagues who report that air in hospitals contains more acetone than other public buildings or private homes [17]. Nevertheless, it should be noted that expired acetone might decrease during critical illness, as we demonstrated in a previous study [5]. For this reason, our findings are specific for critically ill patients due to inflammation and sepsis. Additionally, we detected expired 3-pentanone in high intensities. We also found monomer and dimer forms at lower intensities in room air, possibly the result of exhalation by staff and visitors. Considering dimerization of several volatiles and thus twofold number of molecules in dimer clusters, acetone is the most abundant volatile organic compound in expired gas. Cyclohexanone had one of the highest intensities in expired gas but was detectable in room air in lower intensities. This volatile compound is widely used as an adhesive solvent during manufacture of medical devices which may explain its presence in ambient air [18]. Still, it remains uncertain why cyclohexanone was not present in inspiratory gas that is passing through the ventilator circuit. It is possible that evaporation might depend on running time of ventilator and tubing system and that washout kinetics cause a decline in concentration. Kischkel et al. detected cyclohexanone in medical synthetic air and much more in expired air under mechanical ventilation that had passed through an endotracheal tube, but not in ambient air [14]. They stated that cyclohexanone originates from the material of the endotracheal tube, supporting our findings with high intensities in expired air but not in inspiratory gases. 3-hydroxy-2-butanone, also known as acetoin, revealed substantial intensities in expired air. Staphylococcus aureus, a common pulmonary pathogen in ventilated patients, is well known to produce a characteristic profile of 3-hydroxy-2-butanone and might be partly responsible for the evidence in exhaled breath [19, 20]. Furthermore, acetoin in expired air might be released as a result of cellular damage due to the reactive oxygen species [21]. However, we detected acetoin in room air in high intensities as well. 3-hydroxy-2-butanone is a known flavouring chemical, widely used for food, cigarettes, cosmetics, or detergents and detectable in the breath of healthy individuals as well, which might explain our findings [21]. Our results are generally consistent with Filipiak et al. who determined that expired gas composition is altered by exogenous exposure including smoking and exposure to air pollutants. They identified 86 organic compounds in expired gas to tobacco smoke, most unsaturated hydrocarbons. Exposure to indoor-air contaminations and diet were identified as further contributing factors [22].
Inspired gas in our institution is polluted by six volatiles, all detectable in exhaled air as well: acetone, dimethyl disulphide, isoprene, methanol, 2-methylfuran, and 2-methylpentane. Sturney and colleagues detected acetone in the inspiratory limb of the respirator of intubated and ventilated patients in the intensive care unit as well [23]. They stated a correlation between inspired and expired acetone concentrations, possibly related to contamination of inspiratory samples by exhaled acetone in the inspiratory part of the ventilator in a rebreathing system. Otherwise, components of the respirator and breathing circuit itself might be the source of acetone [23]. These findings would support our detection of acetone in inspired air, but the exact origin remains unknown. In expired air, acetone is the main volatile in human breath and is produced endogenously by hepatic decarboxylation, mainly during lipolysis [23]. The occurrence is related to fasting, diet, patients with diabetes mellitus, and well described in critically ill patients [6, 23], explaining the high intensities of monomer and dimer we present.
We determined dimethyl disulphide in inspired air, gas supplies, and room air. This volatile sulphur compound is released by muscle cells in rats [24] and likewise by cultures of pseudomonas aeruginosa from patients with cystic fibrosis [25]. Nonetheless, we can only speculate about the detection of dimethyl disulphide in inhaled air.
Isoprene is also one of the most abundant volatiles in human breath and a byproduct of cholesterol biosynthesis [26]. Isoprene might be related to oxidative damage to the fluid lining of the lung and the body [27, 28]. Patients with pulmonary fibrosis showed significant higher peak intensities of isoprene compared to healthy subjects [21]. Interestingly, the concentration of isoprene in expired human air is age dependent and shows a circadian rhythm with a maximum in the morning and lower concentrations in the evening [26]. Schubert and colleagues detected isoprene in inspired air of mechanically ventilated patients as well, confirming our findings [29]. Yet, it should be mentioned that we detected isoprene in exhaled breath solely in a single patient, and that in low intensity. These findings are hypothetic and very unlikely. However, in using ion-mobility spectrometry for breath analysis, intensities of different volatiles are substance specific in itself. It is know that isoprene shows a weak signal at the detector of the IMS leading to a poor response even in higher concentrations of isoprene in humid exhaled breath. Protonated isoprene does not form hydrates or cluster like other volatiles and therefore can pass through the drift tube more rapidly. Moreover, the ion lifetime is short, leading to weak ion detection. Finally, presence of interfering ions might be another reason for poor detection of isoprene using ion-mobility spectrometry [30]. Furthermore, these findings have been described only for a few volatile organic compounds. Therefore, statements concerning isoprene intensities in the exhalome must be treated with caution.
Methanol is frequently used as a solvent or detergent. This volatile has been described as a typical outdoor air volatile and might therefore contribute to the presence in gas supplies and inspired air from the ventilator [31]. Moreover, methanol might originate from piping or seals of the ventilator circuit as well. Notably, methanol intensities in exhaled air were twice as high as detected in inspired air. This compound is a major endogenous breath metabolite and also present in expired air of healthy people [32]. A fraction of exhaled methanol was shown to be inhaled from the ambient atmosphere [33]. Elevated levels of this compound has been associated with liver cirrhosis in humans, interestingly decreasing after transplantation [34]. However, concerning volatile isoprene, methanol only leads to a weak response at the IMS detector, as stated previously. Methanol reacts with small hydrated hydronium ions but fails to further react with large ones due to their dipole moments [30]. Thus, quantitative statements comparing methanol intensities with other volatiles are challenging.
We determined 2-methylfuran in ventilator gas, room air, and expired air, but not in the central gas supply. This volatile is an odour component in cigarette smoke and is present especially in the exhalome of smoking subjects [35]. It is observed in ambient air and detectable in rainwater in considerable amounts [36]. The presence in inhaled air but not in the central gas supply suggests the release from components of the respirator.
In a previous animal study, we assigned 2-methylpentane, a branched-chain alkane and structural isomer of hexane, to be a ‘respirator peak’. We now report low intensities in ventilator gas as well. However, this volatile was not detectable in oxygen and compressed air from central gas supply as stated for 2-methylfuran. Therefore, 2-methylpentane might also be released by the ventilator. Yoshida demonstrated the pulmonary absorption of 2-methylpentane by inhalation in a rat model based on the knowledge of well-known diffusion [37]. To what extent inhaled 2-methylpentane affects concentrations in expired air also remains unknown. Filipiak and colleagues observed higher concentrations of 2-methylpentane in the breath of lung cancer patients compared to healthy controls [38], confirming our observation of 2-methylpentane in expired air in human subjects.
Propofol, an intravenous anaesthetic and frequently used for sedation in critical ill patients, can be detected in patients’ exhalome using IMS [39]. We detected propofol in all four patients who were given this intravenous anaesthetic for sedation.
We divided volatile organic compounds into three different groups: expired volatiles which originate from metabolism, resorbed compounds which originate outside the body and are absorbed, and unaffected compounds which are inert and thus present in comparable concentrations in inhaled and expired gas. We previously recorded 22 expired, 12 unaffected, and 3 resorbed volatiles in rats [6]. However, we now report 71 expired compounds, most likely because humans have more environmental influences, illnesses, metabolic differences, medications, and habits than rats. Twelve of the 22 expired substances in rats were also expired in humans.
Fortunately, most compounds detected in inspired gas were also found in expired gas from patients. Most are probably inert compounds that are unaffected by metabolism and thus unimportant for breath analysis. We also observed compounds that had considerably lower concentrations in expired than inspired gas, suggesting that they were resorbed and thus probably irrelevant for breath analysis. However, there were several signals with sustained higher concentrations in expired than inspired gas. These expired volatiles are presumably endogenously derived and thus potentially reflect the patient’s metabolic state. Overall, we identified 11 volatiles that could cause uncertainties in interpretation of the patient’s exhalome: acetone monomer, dimethyl disulphide monomer, methanol, 2-methylfurane, 2-methylpentane, and an additional 6 unknown compounds. However, only 7 compounds showed patient-to-inspired gas ratios < 5 and were therefore considered clinically important for contamination.
The obvious conclusion from our results is that expired concentrations of volatile organic compounds should be normalized for multiplexed inspired concentrations which are technically easy to assess. This approach is consistent with the recommendation of Philips and colleagues who proposed subtracting inhaled concentrations of volatile compounds from their expired concentrations [40], an approach that appears valid at the relatively low concentrations we observed [29]. We note, though, that the effects of inspired substance concentrations on expired concentrations depends on the blood-to-alveolar gradient which is not necessarily linear and can be influenced by shunt perfusion and dead space ventilation, especially in mechanically ventilated patients [41].
Spanel et al. proposed that all exogenous compounds are partially retained in the exhaled breath according to a close linear relationship between exhaled and inhaled concentrations. They defined retention coefficients (α), with values between 0.1 and 1.0, specifically for each compound. When the retention coefficient is close to 1, such as for hydrocarbons and alkanes, inspired concentrations can simply be subtracted from exhaled concentrations by full amount as proposed by Phillips [40]. On the other hand, when the retention coefficient is close to 0.1, such as for water-soluble compounds, inspired concentrations can essentially be ignored in breath analysis [42]. Spanel and colleagues recorded retention coefficients for seven volatiles, ranging from 0.76 for pentane to 0.09 for deuterated water.
In some ways, our study has several limitations. First, ion-mobility spectrometry uses intensities (millivolt) as a unit of quantity rather than of concentrations. Therefore, comparing our results with other data might be difficult. Quantitative statements are partially speculative and not comparable to other studies because detector response of ion mobility spectrometer is substance specific. In any case, the existing recommendation line is unaffected by this: analysis of the exhalome in critical care patients requires normalisation of the expired to inspired signals. In addition, the data we present are specific for our intensive care unit (the ventilators we tested, our own hospital’s central gas supply) and probably not applicable in different settings. Secondly, as mentioned above, quantification and even correct detection of isoprene and methanol is impossible at high levels observed in expired air using ion-mobility spectrometry. Therefore, comparing intensities of these compounds to other volatiles is impossible and statements must be treated with great caution. Only abundances of the same particular compound can be compared between samples. Thirdly, we assume that volatile organic compounds will differ in other contexts, but our goal was not to exactly characterise the patterns in ambient air, gas cylinders, hospital central gas supply, or specific ventilators. Instead, it was to demonstrate that volatile organic compounds are ubiquitous and that any clinical measurement system will need to incorporate multiplexed measurements and compensate for inspired compounds.
While not all volatile compounds were detectable in every patient, nearly half were. In contrast, 14 volatile compounds were detected from single patients and nine others from just two patients. The extent to which the common or unusual compounds reflect normal or abnormal biology — or perhaps drug metabolites — remains largely unknown at this point. Much larger studies will be required to characterise patient-to-patient variability, not to mention how various diseases moderate the exhalome, neither of which was a goal of the current study. In addition, further studies have to focus on the relationship between volatiles’ peak intensity in chromatogram and their normally used units in gases.

Conclusions

Ambient air in critical care units as well as gas from compressed cylinders and from central hospital supplies are all contaminated with various volatile organic compounds. Consequently, gases from mechanical ventilators are as well. Future studies of the exhalome in mechanically ventilated patients should consider and compensate for background contamination.

Acknowledgements

This study contains data taken from the thesis presented by Mario Wachowiak as part of the requirements for a “Doctor of Medicine” degree at Saarland University Medical Centre and Saarland University Faculty of Medicine. The authors would like to thank Karen Schneider for revising this work linguistically and Professor Jörg Ingo Baumbach (Reutlingen, Germany) for his encouragement and support.

Funding

Supported solely by internal funds.

Availability of data and materials

All data generated or analysed during this study are included in this published article.

Authors’information

Centre of Breath Research, Department of Anaesthesiology, Intensive Care and Pain Therapy, Saarland University Medical Centre, Homburg (Saar), Germany
The Centre of Breath Research is part of the OUTCOMES RESEARCH Consortium, Cleveland, Ohio, USA.
This project was approved by the responsible ethics committee (identification number 232/14; Ärztekammer Saarland, Saarbrücken, Germany). Written informed consent was obtained from each patient.
Not applicable.

Competing interests

None of the authors have financial or non-financial competing interests related to this report.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
1.
Zurück zum Zitat Baumbach JI. Ion mobility spectrometry coupled with multi-capillary columns for metabolic profiling of human breath. J Breath Res. 2009;3:34001.CrossRef Baumbach JI. Ion mobility spectrometry coupled with multi-capillary columns for metabolic profiling of human breath. J Breath Res. 2009;3:34001.CrossRef
2.
Zurück zum Zitat de Lacy CB, Amann A, Al-Kateb H, Flynn C, Filipiak W, Khalid T, et al. A review of the volatiles from the healthy human body. J Breath Res. 2014;8:14001.CrossRef de Lacy CB, Amann A, Al-Kateb H, Flynn C, Filipiak W, Khalid T, et al. A review of the volatiles from the healthy human body. J Breath Res. 2014;8:14001.CrossRef
3.
Zurück zum Zitat Fink T, Baumbach JI, Kreuer S. Ion mobility spectrometry in breath research. J Breath Res. 2014;8:27104.CrossRef Fink T, Baumbach JI, Kreuer S. Ion mobility spectrometry in breath research. J Breath Res. 2014;8:27104.CrossRef
4.
Zurück zum Zitat Sakai EM, Connolly LA, Klauck JA. Inhalation anesthesiology and volatile liquid anesthetics: focus on isoflurane, desflurane, and sevoflurane. Pharmacotherapy. 2005;25:1773–88.CrossRefPubMed Sakai EM, Connolly LA, Klauck JA. Inhalation anesthesiology and volatile liquid anesthetics: focus on isoflurane, desflurane, and sevoflurane. Pharmacotherapy. 2005;25:1773–88.CrossRefPubMed
5.
Zurück zum Zitat Fink T, Wolf A, Maurer F, Albrecht FW, Heim N, Wolf B, et al. Volatile organic compounds during inflammation and sepsis in rats: a potential breath test using ion-mobility spectrometry. Anesthesiology. 2015;122:117–26.CrossRefPubMed Fink T, Wolf A, Maurer F, Albrecht FW, Heim N, Wolf B, et al. Volatile organic compounds during inflammation and sepsis in rats: a potential breath test using ion-mobility spectrometry. Anesthesiology. 2015;122:117–26.CrossRefPubMed
6.
Zurück zum Zitat Albrecht FW, Hüppe T, Fink T, Maurer F, Wolf A, Wolf B, et al. Influence of the respirator on volatile organic compounds: an animal study in rats over 24 hours. J Breath Res. 2015;9:16007.CrossRef Albrecht FW, Hüppe T, Fink T, Maurer F, Wolf A, Wolf B, et al. Influence of the respirator on volatile organic compounds: an animal study in rats over 24 hours. J Breath Res. 2015;9:16007.CrossRef
7.
Zurück zum Zitat Wolf A, Baumbach JI, Kleber A, Maurer F, Maddula S, Favrod P, et al. Multi-capillary column-ion mobility spectrometer (MCC-IMS) breath analysis in ventilated rats: a model with the feasibility of long-term measurements. J Breath Res. 2014;8:16006.CrossRef Wolf A, Baumbach JI, Kleber A, Maurer F, Maddula S, Favrod P, et al. Multi-capillary column-ion mobility spectrometer (MCC-IMS) breath analysis in ventilated rats: a model with the feasibility of long-term measurements. J Breath Res. 2014;8:16006.CrossRef
8.
Zurück zum Zitat Maurer F, Hauschild AC, Eisinger K, Baumbach J, Mayor A, Baumbach JI. MIMA - a software for analyte identification in MCC/IMS chromatograms by mapping accompanying GC/MS measurements. Int J Ion Mobil Spectrom. 2014;17:95–101.CrossRef Maurer F, Hauschild AC, Eisinger K, Baumbach J, Mayor A, Baumbach JI. MIMA - a software for analyte identification in MCC/IMS chromatograms by mapping accompanying GC/MS measurements. Int J Ion Mobil Spectrom. 2014;17:95–101.CrossRef
9.
Zurück zum Zitat Ritter JJ, Adams NK. Exponential Dilution as a Calibration Technique. Anal Chem. 1976;48:612–9.CrossRef Ritter JJ, Adams NK. Exponential Dilution as a Calibration Technique. Anal Chem. 1976;48:612–9.CrossRef
10.
Zurück zum Zitat Fowler SJ, Basanta-Sanchez M, Xu Y, Goodacre R, Dark PM. Surveillance for lower airway pathogens in mechanically ventilated patients by metabolomic analysis of exhaled breath: a case-control study. Thorax. 2015;70:320–5.CrossRefPubMed Fowler SJ, Basanta-Sanchez M, Xu Y, Goodacre R, Dark PM. Surveillance for lower airway pathogens in mechanically ventilated patients by metabolomic analysis of exhaled breath: a case-control study. Thorax. 2015;70:320–5.CrossRefPubMed
11.
Zurück zum Zitat Schnabel R, Fijten R, Smolinska A, Dallinga J, Boumans M-L, Stobberingh E, et al. Analysis of volatile organic compounds in exhaled breath to diagnose ventilator-associated pneumonia. Sci Rep. 2015;5:17179.CrossRefPubMedPubMedCentral Schnabel R, Fijten R, Smolinska A, Dallinga J, Boumans M-L, Stobberingh E, et al. Analysis of volatile organic compounds in exhaled breath to diagnose ventilator-associated pneumonia. Sci Rep. 2015;5:17179.CrossRefPubMedPubMedCentral
12.
Zurück zum Zitat Filipiak W, Beer R, Sponring A, Filipiak A, Ager C, Schiefecker A, et al. Breath analysis for in vivo detection of pathogens related to ventilator-associated pneumonia in intensive care patients: a prospective pilot study. J Breath Res. 2015;9:16004.CrossRef Filipiak W, Beer R, Sponring A, Filipiak A, Ager C, Schiefecker A, et al. Breath analysis for in vivo detection of pathogens related to ventilator-associated pneumonia in intensive care patients: a prospective pilot study. J Breath Res. 2015;9:16004.CrossRef
13.
Zurück zum Zitat Lee HJ, Meinardi S, Pahl MV, Vaziri ND, Blake DR. Exposure to potentially toxic hydrocarbons and halocarbons released from the dialyzer and tubing set during Hemodialysis. Am J Kidney Dis. 2012;60:609–16.CrossRefPubMed Lee HJ, Meinardi S, Pahl MV, Vaziri ND, Blake DR. Exposure to potentially toxic hydrocarbons and halocarbons released from the dialyzer and tubing set during Hemodialysis. Am J Kidney Dis. 2012;60:609–16.CrossRefPubMed
14.
Zurück zum Zitat Kischkel S, Miekisch W, Fuchs P, Schubert JK. Breath analysis during one-lung ventilation in cancer patients. Eur Respir J. 2012;40:706–13.CrossRefPubMed Kischkel S, Miekisch W, Fuchs P, Schubert JK. Breath analysis during one-lung ventilation in cancer patients. Eur Respir J. 2012;40:706–13.CrossRefPubMed
15.
Zurück zum Zitat Risby TH, Sehnert SS. Clinical application of breath biomarkers of oxidative stress status. Free Radic Biol Med. 1999;27:1182–92.CrossRefPubMed Risby TH, Sehnert SS. Clinical application of breath biomarkers of oxidative stress status. Free Radic Biol Med. 1999;27:1182–92.CrossRefPubMed
16.
Zurück zum Zitat Gao J, Zou Y, Wang Y, Wang F, Lang L, Wang P, et al. Breath analysis for noninvasively differentiating Acinetobacter Baumannii ventilator-associated pneumonia from its respiratory tract colonization of ventilated patients. J Breath Res. 2016;10:27102.CrossRef Gao J, Zou Y, Wang Y, Wang F, Lang L, Wang P, et al. Breath analysis for noninvasively differentiating Acinetobacter Baumannii ventilator-associated pneumonia from its respiratory tract colonization of ventilated patients. J Breath Res. 2016;10:27102.CrossRef
17.
Zurück zum Zitat Bessonneau V, Mosqueron L, Berrubé A, Mukensturm G, Buffet-Bataillon S, Gangneux J-P, et al. VOC contamination in hospital, from stationary sampling of a large panel of compounds, in view of healthcare workers and patients exposure assessment. PLoS One. 2013;8:e55535.CrossRefPubMedPubMedCentral Bessonneau V, Mosqueron L, Berrubé A, Mukensturm G, Buffet-Bataillon S, Gangneux J-P, et al. VOC contamination in hospital, from stationary sampling of a large panel of compounds, in view of healthcare workers and patients exposure assessment. PLoS One. 2013;8:e55535.CrossRefPubMedPubMedCentral
18.
Zurück zum Zitat Wang Y, Han H, Shen C, Li J, Wang H, Chu Y. Control of solvent use in medical devices by proton transfer reaction mass spectrometry and ion molecule reaction mass spectrometry. J Pharm Biomed Anal. 2009;50:252–6.CrossRefPubMed Wang Y, Han H, Shen C, Li J, Wang H, Chu Y. Control of solvent use in medical devices by proton transfer reaction mass spectrometry and ion molecule reaction mass spectrometry. J Pharm Biomed Anal. 2009;50:252–6.CrossRefPubMed
19.
Zurück zum Zitat Zechman JM, Aldinger S, Labows JN. Characterization of pathogenic bacteria by automated headspace concentration-gas chromatography. J Chromatogr. 1986;377:49–57.CrossRefPubMed Zechman JM, Aldinger S, Labows JN. Characterization of pathogenic bacteria by automated headspace concentration-gas chromatography. J Chromatogr. 1986;377:49–57.CrossRefPubMed
20.
Zurück zum Zitat Filipiak W, Sponring A, Baur MM, Filipiak A, Ager C, Wiesenhofer H, et al. Molecular analysis of volatile metabolites released specifically by Staphylococcus Aureus and Pseudomonas Aeruginosa. BMC Microbiol. 2012;12:113.CrossRefPubMedPubMedCentral Filipiak W, Sponring A, Baur MM, Filipiak A, Ager C, Wiesenhofer H, et al. Molecular analysis of volatile metabolites released specifically by Staphylococcus Aureus and Pseudomonas Aeruginosa. BMC Microbiol. 2012;12:113.CrossRefPubMedPubMedCentral
21.
Zurück zum Zitat Yamada Y, Yamada G, Otsuka M, Nishikiori H, Ikeda K, Umeda Y, et al. Volatile organic compounds in exhaled breath of idiopathic pulmonary fibrosis for discrimination from healthy subjects. Lung. 2017;195:247–54.CrossRefPubMed Yamada Y, Yamada G, Otsuka M, Nishikiori H, Ikeda K, Umeda Y, et al. Volatile organic compounds in exhaled breath of idiopathic pulmonary fibrosis for discrimination from healthy subjects. Lung. 2017;195:247–54.CrossRefPubMed
22.
Zurück zum Zitat Filipiak W, Ruzsanyi V, Mochalski P, Filipiak A, Bajtarevic A, Ager C, et al. Dependence of exhaled breath composition on exogenous factors, smoking habits and exposure to air pollutants. J Breath Res. 2012;6:36008.CrossRef Filipiak W, Ruzsanyi V, Mochalski P, Filipiak A, Bajtarevic A, Ager C, et al. Dependence of exhaled breath composition on exogenous factors, smoking habits and exposure to air pollutants. J Breath Res. 2012;6:36008.CrossRef
23.
Zurück zum Zitat Sturney SC, Storer MK, Shaw GM, Shaw DE, Epton MJ. Off-line breath acetone analysis in critical illness. J Breath Res. 2013;7:37102.CrossRef Sturney SC, Storer MK, Shaw GM, Shaw DE, Epton MJ. Off-line breath acetone analysis in critical illness. J Breath Res. 2013;7:37102.CrossRef
24.
Zurück zum Zitat Mochalski P, Al-Zoairy R, Niederwanger A, Unterkofler K, Amann A. Quantitative analysis of volatile organic compounds released and consumed by rat L6 skeletal muscle cells in vitro. J Breath Res. 2014;8:46003.CrossRef Mochalski P, Al-Zoairy R, Niederwanger A, Unterkofler K, Amann A. Quantitative analysis of volatile organic compounds released and consumed by rat L6 skeletal muscle cells in vitro. J Breath Res. 2014;8:46003.CrossRef
25.
Zurück zum Zitat Carroll W, Lenney W, Wang TS, Spanel P, Alcock A, Smith D. Detection of volatile compounds emitted by Pseudomonas Aeruginosa using selected ion flow tube mass spectrometry. Pediatr Pulmonol. 2005;39:452–6.CrossRefPubMed Carroll W, Lenney W, Wang TS, Spanel P, Alcock A, Smith D. Detection of volatile compounds emitted by Pseudomonas Aeruginosa using selected ion flow tube mass spectrometry. Pediatr Pulmonol. 2005;39:452–6.CrossRefPubMed
26.
Zurück zum Zitat Miekisch W, Schubert JK, Noeldge-Schomburg GF. Diagnostic potential of breath analysis—focus on volatile organic compounds. Clin Chim Acta. 2004;347:25–39.CrossRefPubMed Miekisch W, Schubert JK, Noeldge-Schomburg GF. Diagnostic potential of breath analysis—focus on volatile organic compounds. Clin Chim Acta. 2004;347:25–39.CrossRefPubMed
27.
Zurück zum Zitat Foster WM, Jiang L, Stetkiewicz PT, Risby TH. Breath isoprene: temporal changes in respiratory output after exposure to ozone. J Appl Physiol. 1996;80:706–10.PubMed Foster WM, Jiang L, Stetkiewicz PT, Risby TH. Breath isoprene: temporal changes in respiratory output after exposure to ozone. J Appl Physiol. 1996;80:706–10.PubMed
28.
Zurück zum Zitat Mendis S, Sobotka PA, Euler DE. Expired hydrocarbons in patients with acute myocardial infarction. Free Radic Res. 1995;23:117–22.CrossRefPubMed Mendis S, Sobotka PA, Euler DE. Expired hydrocarbons in patients with acute myocardial infarction. Free Radic Res. 1995;23:117–22.CrossRefPubMed
29.
Zurück zum Zitat Schubert JK, Miekisch W, Birken T, Geiger K, Noldge-Schomburg GF. Impact of inspired substance concentrations on the results of breath analysis in mechanically ventilated patients. Biomarkers. 2005;10:138–52.CrossRefPubMed Schubert JK, Miekisch W, Birken T, Geiger K, Noldge-Schomburg GF. Impact of inspired substance concentrations on the results of breath analysis in mechanically ventilated patients. Biomarkers. 2005;10:138–52.CrossRefPubMed
30.
Zurück zum Zitat Mochalski P, Rudnicka J, Agapiou A, Statheropoulos M, Amann A, Buszewski B. Near real-time VOCs analysis using an aspiration ion mobility spectrometer. J Breath Res. 2013;7(026002):11. Mochalski P, Rudnicka J, Agapiou A, Statheropoulos M, Amann A, Buszewski B. Near real-time VOCs analysis using an aspiration ion mobility spectrometer. J Breath Res. 2013;7(026002):11.
31.
Zurück zum Zitat Tang X, Misztal PK, Nazaroff WW, Goldstein AH. Volatile Organic Compound Emissions from Humans Indoors. Environ Sci Technol. 2016;acs.est.6b04415. Tang X, Misztal PK, Nazaroff WW, Goldstein AH. Volatile Organic Compound Emissions from Humans Indoors. Environ Sci Technol. 2016;acs.est.6b04415.
32.
Zurück zum Zitat Turner C, Španěl P, Smith D. A longitudinal study of methanol in the exhaled breath of 30 healthy volunteers using selected ion flow tube mass spectrometry. SIFT-MS Physiol Meas. 2006;27(7):637–48.CrossRefPubMed Turner C, Španěl P, Smith D. A longitudinal study of methanol in the exhaled breath of 30 healthy volunteers using selected ion flow tube mass spectrometry. SIFT-MS Physiol Meas. 2006;27(7):637–48.CrossRefPubMed
33.
Zurück zum Zitat Ernstgård L, Shibata E, Johanson G. Uptake and disposition of inhaled methanol vapor in humans. Toxicol Sci. 2005;88:30–8.CrossRefPubMed Ernstgård L, Shibata E, Johanson G. Uptake and disposition of inhaled methanol vapor in humans. Toxicol Sci. 2005;88:30–8.CrossRefPubMed
34.
Zurück zum Zitat Fernandez del Rio R, O’Hara ME, Holt A, Pemberton P, Shah T, Whitehouse T, et al. Volatile biomarkers in breath associated with liver cirrhosis - comparisons of pre- and post-liver transplant breath samples. EBioMedicine. 2015;(2):1243–50. Fernandez del Rio R, O’Hara ME, Holt A, Pemberton P, Shah T, Whitehouse T, et al. Volatile biomarkers in breath associated with liver cirrhosis - comparisons of pre- and post-liver transplant breath samples. EBioMedicine. 2015;(2):1243–50.
35.
Zurück zum Zitat Van Berkel JJBN, Dallinga JW, Möller GM, Godschalk RWL, Moonen E, Wouters EFM, et al. Development of accurate classification method based on the analysis of volatile organic compounds from human exhaled air. J Chromatogr B Anal Technol Biomed Life Sci. 2008;861:101–7.CrossRef Van Berkel JJBN, Dallinga JW, Möller GM, Godschalk RWL, Moonen E, Wouters EFM, et al. Development of accurate classification method based on the analysis of volatile organic compounds from human exhaled air. J Chromatogr B Anal Technol Biomed Life Sci. 2008;861:101–7.CrossRef
36.
Zurück zum Zitat Mullaugh KM, Hamilton JM, Avery GB, Felix JD, Mead RN, Willey JD, et al. Temporal and spatial variability of trace volatile organic compounds in rainwater. Chemosphere. 2015;134:203–9.CrossRefPubMed Mullaugh KM, Hamilton JM, Avery GB, Felix JD, Mead RN, Willey JD, et al. Temporal and spatial variability of trace volatile organic compounds in rainwater. Chemosphere. 2015;134:203–9.CrossRefPubMed
37.
Zurück zum Zitat Yoshida T. Estimation of absorption of aromatic hydrocarbons diffusing from interior materials in automobile cabins by inhalation toxicokinetic analysis in rats. J Appl Toxicol. 2010;30:525–35.CrossRefPubMed Yoshida T. Estimation of absorption of aromatic hydrocarbons diffusing from interior materials in automobile cabins by inhalation toxicokinetic analysis in rats. J Appl Toxicol. 2010;30:525–35.CrossRefPubMed
38.
Zurück zum Zitat Filipiak W, Filipiak A, Sponring A, Schmid T, Zelger B, Ager C, et al. Comparative analyses of volatile organic compounds (VOCs) from patients, tumors and transformed cell lines for the validation of lung cancer-derived breath markers. J Breath Res. 2014;8:27111.CrossRef Filipiak W, Filipiak A, Sponring A, Schmid T, Zelger B, Ager C, et al. Comparative analyses of volatile organic compounds (VOCs) from patients, tumors and transformed cell lines for the validation of lung cancer-derived breath markers. J Breath Res. 2014;8:27111.CrossRef
39.
Zurück zum Zitat Perl T, Carstens E, Hirn A, Quintel M, Vautz W, Nolte J, et al. Determination of serum propofol concentrations by breath analysis using ion mobility spectrometry. Br J Anaesth. 2009;103:822–7.CrossRefPubMed Perl T, Carstens E, Hirn A, Quintel M, Vautz W, Nolte J, et al. Determination of serum propofol concentrations by breath analysis using ion mobility spectrometry. Br J Anaesth. 2009;103:822–7.CrossRefPubMed
40.
Zurück zum Zitat Phillips M, Greenberg J, Sabas M. Alveolar gradient of pentane in normal human breath. Free Radic Res. 1994;20:333–7.CrossRefPubMed Phillips M, Greenberg J, Sabas M. Alveolar gradient of pentane in normal human breath. Free Radic Res. 1994;20:333–7.CrossRefPubMed
41.
Zurück zum Zitat Dembinski R, Max M, Bensberg R, Bickenbach J, Kuhlen R, Rossaint R. High-frequency oscillatory ventilation in experimental lung injury: effects on gas exchange. Intensive Care Med. 2002;28:768–74.CrossRefPubMed Dembinski R, Max M, Bensberg R, Bickenbach J, Kuhlen R, Rossaint R. High-frequency oscillatory ventilation in experimental lung injury: effects on gas exchange. Intensive Care Med. 2002;28:768–74.CrossRefPubMed
42.
Zurück zum Zitat Spaněl P, Dryahina K, Smith D. A quantitative study of the influence of inhaled compounds on their concentrations in exhaled breath. J Breath Res. 2013;7:17106.CrossRef Spaněl P, Dryahina K, Smith D. A quantitative study of the influence of inhaled compounds on their concentrations in exhaled breath. J Breath Res. 2013;7:17106.CrossRef
Metadaten
Titel
Volatile organic compounds in ventilated critical care patients: a systematic evaluation of cofactors
Publikationsdatum
01.12.2017
Erschienen in
BMC Pulmonary Medicine / Ausgabe 1/2017
Elektronische ISSN: 1471-2466
DOI
https://doi.org/10.1186/s12890-017-0460-0

Weitere Artikel der Ausgabe 1/2017

BMC Pulmonary Medicine 1/2017 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Echinokokkose medikamentös behandeln oder operieren?

06.05.2024 DCK 2024 Kongressbericht

Die Therapie von Echinokokkosen sollte immer in spezialisierten Zentren erfolgen. Eine symptomlose Echinokokkose kann – egal ob von Hunde- oder Fuchsbandwurm ausgelöst – konservativ erfolgen. Wenn eine Op. nötig ist, kann es sinnvoll sein, vorher Zysten zu leeren und zu desinfizieren. 

Umsetzung der POMGAT-Leitlinie läuft

03.05.2024 DCK 2024 Kongressbericht

Seit November 2023 gibt es evidenzbasierte Empfehlungen zum perioperativen Management bei gastrointestinalen Tumoren (POMGAT) auf S3-Niveau. Vieles wird schon entsprechend der Empfehlungen durchgeführt. Wo es im Alltag noch hapert, zeigt eine Umfrage in einem Klinikverbund.

Proximale Humerusfraktur: Auch 100-Jährige operieren?

01.05.2024 DCK 2024 Kongressbericht

Mit dem demographischen Wandel versorgt auch die Chirurgie immer mehr betagte Menschen. Von Entwicklungen wie Fast-Track können auch ältere Menschen profitieren und bei proximaler Humerusfraktur können selbst manche 100-Jährige noch sicher operiert werden.

Die „Zehn Gebote“ des Endokarditis-Managements

30.04.2024 Endokarditis Leitlinie kompakt

Worauf kommt es beim Management von Personen mit infektiöser Endokarditis an? Eine Kardiologin und ein Kardiologe fassen die zehn wichtigsten Punkte der neuen ESC-Leitlinie zusammen.

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.