A comparison of three adjustable cortical button ACL fixation devices

verfasst von: Meghan Pasquali, Matthew J. Plante, Keith O. Monchik, David B. Spenciner

Erschienen in: Knee Surgery, Sports Traumatology, Arthroscopy | Ausgabe 5/2017

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Adjustable cortical fixation has gained popularity recently for ACL reconstruction; however, one concern with these types of devices is the potential for laxity occurring post-operatively. An indicator of clinical laxity is cyclic displacement during bench-top mechanical testing. The hypothesis was that the cyclic displacement and maximum strength of different adjustable buttons currently on the market would vary depending on their mechanism of fixation.

Methods

Three devices were studied: Biomet’s ToggleLoc with ZipLoop Technology (ZL), Arthrex’s TightRope RT (TR), and DePuy Mitek’s RIGIDLOOP Adjustable (RLA). Each was tested in isolation on a servohydraulic test machine. The implants were pre-conditioned from 5 to 67 N for 10 cycles and then tested from 50 to 250 N for 1000 cycles at 1 Hz. Following cyclic loading, a load to failure test was conducted at a rate of 20 mm/min. Total displacement was calculated from the 1st to the 1000th cycle, and ultimate strength and failure mode were recorded.

Results

Results showed that the RLA had significantly lower displacement versus both the TR (p = 0.012) and the ZL (p < 0.001). The TR also showed significantly lower displacement than the ZL (p < 0.001). The RLA and the ZL were both significantly stronger than the TR (p < 0.001).

Conclusions

Both the RLA and TR showed clinically acceptable amounts of cyclic displacement and maximum strength. The clinical relevance of this study is that cyclic displacement results from bench-top tests may correlate with post-operative laxity, which ideally should be kept to a minimum.
Literatur
1.
Zurück zum Zitat Barrow AE, Pilia M, Guda T, Kadrmas WR, Burns TC (2014) Femoral suspension devices for anterior cruciate ligament reconstruction: do adjustable loops lengthen? Am J Sports Med 42:343–349CrossRefPubMed Barrow AE, Pilia M, Guda T, Kadrmas WR, Burns TC (2014) Femoral suspension devices for anterior cruciate ligament reconstruction: do adjustable loops lengthen? Am J Sports Med 42:343–349CrossRefPubMed
2.
Zurück zum Zitat Benjaminse A, Gokeler A, van der Schans CP (2006) Clinical diagnosis of an anterior cruciate ligament rupture: a meta-analysis. J Orthop Sports Phys Ther 36:267–288CrossRefPubMed Benjaminse A, Gokeler A, van der Schans CP (2006) Clinical diagnosis of an anterior cruciate ligament rupture: a meta-analysis. J Orthop Sports Phys Ther 36:267–288CrossRefPubMed
3.
Zurück zum Zitat Boyle MJ, Vovos TJ, Walker CG, Stabile KJ, Roth JM, Garrett WE Jr (2015) Does adjustable-loop femoral cortical suspension loosen after anterior cruciate ligament reconstruction? A retrospective comparative study. Knee 22:304–308CrossRefPubMed Boyle MJ, Vovos TJ, Walker CG, Stabile KJ, Roth JM, Garrett WE Jr (2015) Does adjustable-loop femoral cortical suspension loosen after anterior cruciate ligament reconstruction? A retrospective comparative study. Knee 22:304–308CrossRefPubMed
4.
Zurück zum Zitat Chechik O, Amar E, Khashan M, Lador R, Eyal G, Gold A (2013) An international survey on anterior cruciate ligament reconstruction practices. Int Orthop 37:201–206CrossRefPubMed Chechik O, Amar E, Khashan M, Lador R, Eyal G, Gold A (2013) An international survey on anterior cruciate ligament reconstruction practices. Int Orthop 37:201–206CrossRefPubMed
5.
Zurück zum Zitat Daniel DM, Stone ML, Dobson BE, Fithian DC, Rossman DJ, Kaufman KR (1994) Fate of the ACL-injured patient. A prospective outcome study. Am J Sports Med 22:632–644CrossRefPubMed Daniel DM, Stone ML, Dobson BE, Fithian DC, Rossman DJ, Kaufman KR (1994) Fate of the ACL-injured patient. A prospective outcome study. Am J Sports Med 22:632–644CrossRefPubMed
6.
Zurück zum Zitat Eguchi A, Ochi M, Adachi N, Deie M, Nakamae A, Usman MA (2014) Mechanical properties of suspensory fixation devices for anterior cruciate ligament reconstruction: comparison of the fixed-length loop device versus the adjustable-length loop device. Knee 21:743–748CrossRefPubMed Eguchi A, Ochi M, Adachi N, Deie M, Nakamae A, Usman MA (2014) Mechanical properties of suspensory fixation devices for anterior cruciate ligament reconstruction: comparison of the fixed-length loop device versus the adjustable-length loop device. Knee 21:743–748CrossRefPubMed
7.
Zurück zum Zitat Gillquist J, Good L (1993) Load and length changes in an artificial ligament substitute. 10 cases of anterior cruciate ligament reconstruction. Acta Orthop Scand 64:575–579CrossRefPubMed Gillquist J, Good L (1993) Load and length changes in an artificial ligament substitute. 10 cases of anterior cruciate ligament reconstruction. Acta Orthop Scand 64:575–579CrossRefPubMed
8.
Zurück zum Zitat Johnson JS, Smith SD, LaPrade CM, Turnbull TL, LaPrade RF, Wijdicks CA (2014) A biomechanical comparison of femoral cortical suspension devices for soft tissue anterior cruciate ligament reconstruction under high loads. Am J Sports Med 43:154–160CrossRefPubMed Johnson JS, Smith SD, LaPrade CM, Turnbull TL, LaPrade RF, Wijdicks CA (2014) A biomechanical comparison of femoral cortical suspension devices for soft tissue anterior cruciate ligament reconstruction under high loads. Am J Sports Med 43:154–160CrossRefPubMed
9.
Zurück zum Zitat Kessler MA, Behrend H, Henz S, Stutz G, Rukavina A, Kuster MS (2008) Function, osteoarthritis and activity after ACL-rupture: 11 years follow-up results of conservative versus reconstructive treatment. Knee Surg Sports Traumatol Arthrosc 16:442–448CrossRefPubMed Kessler MA, Behrend H, Henz S, Stutz G, Rukavina A, Kuster MS (2008) Function, osteoarthritis and activity after ACL-rupture: 11 years follow-up results of conservative versus reconstructive treatment. Knee Surg Sports Traumatol Arthrosc 16:442–448CrossRefPubMed
10.
Zurück zum Zitat Kurosaka M, Yoshiya S, Andrish JT (1987) A biomechanical comparison of different surgical techniques of graft fixation in anterior cruciate ligament reconstruction. Am J Sports Med 15:225–229CrossRefPubMed Kurosaka M, Yoshiya S, Andrish JT (1987) A biomechanical comparison of different surgical techniques of graft fixation in anterior cruciate ligament reconstruction. Am J Sports Med 15:225–229CrossRefPubMed
11.
Zurück zum Zitat McGinty G, Irrgang JJ, Pezzullo D (2000) Biomechanical considerations for rehabilitation of the knee. Clin Biomech (Bristol Avon) 15:160–166CrossRef McGinty G, Irrgang JJ, Pezzullo D (2000) Biomechanical considerations for rehabilitation of the knee. Clin Biomech (Bristol Avon) 15:160–166CrossRef
12.
Zurück zum Zitat Morrison JB (1970) The mechanics of the knee joint in relation to normal walking. J Biomech 3:51–61CrossRefPubMed Morrison JB (1970) The mechanics of the knee joint in relation to normal walking. J Biomech 3:51–61CrossRefPubMed
13.
Zurück zum Zitat Petre BM, Smith SD, Jansson KS, de Meijer PP, Hackett TR, LaPrade RF, Wijdicks CA (2012) Femoral cortical suspension devices for soft tissue anterior cruciate ligament reconstruction: a comparative biomechanical study. Am J Sports Med 41:416–422CrossRefPubMed Petre BM, Smith SD, Jansson KS, de Meijer PP, Hackett TR, LaPrade RF, Wijdicks CA (2012) Femoral cortical suspension devices for soft tissue anterior cruciate ligament reconstruction: a comparative biomechanical study. Am J Sports Med 41:416–422CrossRefPubMed
14.
Zurück zum Zitat Tashman S, Collon D, Anderson K, Kolowich P, Anderst W (2004) Abnormal rotational knee motion during running after anterior cruciate ligament reconstruction. Am J Sports Med 32:975–983CrossRefPubMed Tashman S, Collon D, Anderson K, Kolowich P, Anderst W (2004) Abnormal rotational knee motion during running after anterior cruciate ligament reconstruction. Am J Sports Med 32:975–983CrossRefPubMed
Metadaten
Titel
A comparison of three adjustable cortical button ACL fixation devices
verfasst von
Meghan Pasquali
Matthew J. Plante
Keith O. Monchik
David B. Spenciner
Publikationsdatum
22.07.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
Knee Surgery, Sports Traumatology, Arthroscopy / Ausgabe 5/2017
Print ISSN: 0942-2056
Elektronische ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-015-3711-8

Weitere Artikel der Ausgabe 5/2017

Knee Surgery, Sports Traumatology, Arthroscopy 5/2017 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Proximale Humerusfraktur: Auch 100-Jährige operieren?

01.05.2024 DCK 2024 Kongressbericht

Mit dem demographischen Wandel versorgt auch die Chirurgie immer mehr betagte Menschen. Von Entwicklungen wie Fast-Track können auch ältere Menschen profitieren und bei proximaler Humerusfraktur können selbst manche 100-Jährige noch sicher operiert werden.

Sind Frauen die fähigeren Ärzte?

30.04.2024 Gendermedizin Nachrichten

Patienten, die von Ärztinnen behandelt werden, dürfen offenbar auf bessere Therapieergebnisse hoffen als Patienten von Ärzten. Besonders gilt das offenbar für weibliche Kranke, wie eine Studie zeigt.

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.