A New Tool for Breast Anthropometric Measurements: Presentation and Validation for Women and Men

verfasst von: Maksym Mikołajczyk, Anna Kasielska-Trojan, Bogusław Antoszewski

Erschienen in: Aesthetic Plastic Surgery | Ausgabe 5/2019

Abstract

Introduction

Anthropometric measurements of breasts are crucial for planning surgical procedures; however, there are no practical solutions for their quick, digital performance. The aim of the study was to present and validate a self-designed web application BreastIdea (BI) designed for indirect breast anthropometry.

Methods

Ten male and 10 female volunteers had their chests measured directly according to the routine clinical practice. Then their chests were photographed in non-standardised conditions. Corresponding measurements were performed using BI. Accuracy and both relative and absolute reliability of BI measurements were investigated.

Results

Breast assessments using BI yielded highly accurate results and presented near-perfect precision when compared to direct anthropometric measurements of the breast. Indirect anthropometry eliminates the necessity to trace the body’s curves, which usually introduces a bias to linear measurements.

Conclusion

BI web application is a reliable tool for indirect breast measurements in a clinical setting, providing accurate results regardless of chest pathology and photograph standardisation.

Level of Evidence IV

This journal requires that authors assign a level of evidence to each article. For a full description of these Evidence-Based Medicine ratings, please refer to the Table of Contents or the online Instructions to Authors www.​springer.​com/​00266.
Hinweise
Maksym Mikołajczyk and Anna Kasielska-Trojan equally contributed to this work.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

Planning a surgical procedure on the breast, be it reconstructive, reductive, oncological or aesthetic, breast measurements prove indispensable for state-of-the-art surgery [1]. Proportions, asymmetries and levels play a major role in planning and optimising treatment results. Preoperative chest evaluation is also crucial to accurate breast implant choices for aesthetic and reconstructive reasons [2].
Even though the consistency of outcome of manual measurements is highly dependent on the measurer’s skills and experience, as well as available precision of the tools used, those crucial measurements are mostly done by hand (direct anthropometry). The female breast is one of the most difficult organs to assess objectively [2]. Not only size, but also contour, volume and asymmetry need to be considered [3]. Some breast surgeons use 3D scanning, which would yield accurate results. The technology is not yet as accessible in clinical practice as a simple tape measure. What is more, it is time-consuming and requires expensive devices and a trained specialist to acquire and analyse the scan [4]. There have been attempts to perform breast measurements on photographs using graphics software packages [5]. However, the studies were based on photographs captured in standardised conditions, which are not possible to replicate in clinical settings. What is more, the software used was not clinically oriented.
Due to the lack of freely accessible and easy to use tools for indirect breast anthropometry, we decided to design such software. The tool was not created as an alternative to direct measurements, which remain a gold standard in breast surgeons’ clinical practice. Photograph-based breast analysis may, however, be helpful as an additional source of clinical data. It may help to extract metric data from the image, e.g. while consulting the patient with other surgeons and assessing asymmetry. Moreover, it may serve as a valuable tool for scientific research, like metric evaluation of breast asymmetry performed by independent observers without the need for personal contact with the patient.
Our aim is to present the self-designed web application—BreastIdea (BI)—a tool for linear breast measurements, allowing to substitute a physical pen and tape measure with their digital counterparts, thus increasing the objectivity of breast assessment. In the article, we show the process of its validation, as well as some pitfalls to be avoided by potential users.

Methods

The protocol of the study was approved by the Bioethics Committee of the Medical University (RNN/100/18/KE). Written informed consent was obtained from all subjects. The study consisted of three phases. The first was web application design and development, the second—direct breast measurements and the third—digital measurements using BI (indirect breast measurements). The measurements performed with BI were compared with direct measurements performed by a plastic surgeon.
The study groups consisted of 10 males, aged 18–37 (mean age 23.7 years, SD 5.04 years) and 10 females, aged 18–45 (mean age 29.6 years, SD 9.38). The volunteers were randomly selected from students of the Medical University and from patients at the Plastic, Reconstructive and Aesthetic Surgery Clinic, University Hospital, regardless of age (over 18 years old), breast size, breast ptosis and deformation (e.g. Poland syndrome). In the female group, 6 women (volunteers: students, patients) had natural breasts and 4 were clinic patients after breast surgery [breast reduction and lift (n = 2), breast enlargement (n = 1), breast symmetrisation due to Poland syndrome (n = 1)]. None of the male participants had chest surgery; however, two had mild lipomastia. The exclusion criterion for the female and male groups was obesity—body mass index above 30. The study process was identical for all patients.

Web Application Design and Development

Our initial assumption was to design an application that was easy to use and access (through a web browser). The user interface was written in Hypertext Markup Language (HTML) with Cascading Style Sheets (CSS), while the logic was supplied by JavaScript (JS)—a standard in web development [6]. Seven plastic surgeons specialising in breast surgery were asked to identify clinically relevant breast measurements. The following linear breast parameters were chosen: the difference in the levels of nipples, medial line–nipple distance (right and left), sternal notch–nipple distance (right and left), the difference in the levels of inframammary folds apices, medial line–inframammary fold apex distance (right and left), inframammary fold apex–nipple distance (right and left), the difference in the levels of both upper pole apices, and upper pole apex–nipple distance (right and left) (Fig. 1). The parameters were not gender specific, applying to both males and females.
For aesthetic assessment, we added proportions of the upper to the lower pole as well as medial to lateral part, derived from the measurements listed above. Following Mallucci et al., for vertical parameters we used the ratio of distance from upper pole apex to nipple and distance from inframammary fold apex to nipple (with the most appealing proportion of 45:55) [7]. For horizontal parameters, we used the ratio of distance from the breast’s lateral border to the nipple and distance from the breast’s medial border to the nipple (with most appealing proportion of 40:60), as suggested by Lewin et al. [8].
BI guides the user through the process of digital breast measurements. Each measurement consists of placing a number of digital markers on the photograph by clicking in locations suggested by dynamic instructions. If needed, the markers are automatically connected by lines. Most of the measurements are based upon straightforward vector length computation. Level difference measurements need an additional step of horizontal projection of the digital markers onto the midline and then calculating the distance between the resulting coordinates. To allow precise point marking, BI functions are enhanced with photograph transformation capabilities—translation, scaling and rotation. The user can also adjust its brightness and contrast (Fig. 2).
In many patients, the upper breast border is not evident, especially on photographical data. To determine the breast footprint, the breast has to be lifted. Taking this into account, an option to use a second photograph was added. This option was designed especially for patients with breast ptosis—the patient is photographed while lifting their breasts to visualise inframammary folds and upper breast borders. This function makes it possible to correctly assess parameters referring to the inframammary fold and better depict the upper pole of the breasts. The two photographs are spatially bound together by marking the patient’s medial line on both of them. After all the measurements are made, BI generates a report with all the markers and lines present on the photograph, as well as essential patient information (Fig. 3). The report can then be printed or saved as a PDF file, providing an easy way to enclose the results of the breast assessment process in both physical and electronic medical documentation.
The application uses the web browser only as an environment to run JS logic. The photographs used in BI are not uploaded to the web. No data were sent out or fetched from a database—data were only handled locally in the browser. BI is available free of charge for everyone at www.​breastidea.​com/​app.

Direct Breast Measurements

Direct breast measurements took place at different times of day and in random lighting conditions. The surgeon placed a reference unit (2 cm long) on the chest, either below a clavicle or on the sternum, below the sternal angle. The location of the reference unit did not affect measurement accuracy, so optimal reference unit placement was dependent on the patient’s body shape, as long as the surface was flat and perpendicular to the camera. The unit was used to translate virtual length to real-world length. A digital photograph of the subject’s chest was taken, making sure the camera (We consider that most potential users will take pictures with their smartphones. However, because of security restrictions imposed by the Ethics Committee, we were not allowed to take photographs with a mobile phone. We made sure the digital camera parameters were as close to a standard smartphone’s camera as possible.) was perpendicular to the ground so as not to skew the perspective. Based on breast size and ptosis, the need for an additional photograph was evaluated. In those situations, the subject was asked to elevate their breasts for the second photograph.
Next, to acquire the reference data, all previously listed measurements were performed directly (gold standard method). All essential lines were drawn on the subject’s chest using a marker. Then, using a tape measure, the linear parameters were measured. To measure level differences, the right and left inframammary folds apices, upper pole apices and nipples were projected as points to the midline and the differences between these points were measured. The duration of measurements was noted. All measurements were done by one person (a plastic surgeon).

Digital Indirect Breast Measurements

After transferring the photographs onto a computer, all measurements were made using BI (Fig. 4). The user is guided through the measurement process, which consists of placing digital markers by moving the cursor to indicated anthropometric landmarks and clicking the left mouse button. Measurements are automatically generated when necessary markers are placed. They are also updated in real time when the user adjusts the markers.

Data Quality

To guarantee the quality of the data and their suitability we conducted the following procedures. Subjects were chosen randomly, regardless of breast shape and size. Direct measurements were made by one person (a plastic surgeon). Breast measurements using BI were performed independently by two researchers of different experience in breast measurements (a plastic surgeon other than the one collecting direct measurements and a medical student). Digital measurements on the subjects’ photographs were repeated at weekly intervals during three sessions, and a set of photographs was ordered randomly each time.

Statistical Analysis

Accuracy (the degree of relation between the established ground-truth reference values measured physically and the corresponding digital measurements) was presented as mean absolute difference (MAD), i.e. expected absolute difference of two independent variables. Precision of measurement (variability of measurements performed by the same observer or relative reliability) for each variable was established using intra-class correlation coefficients (ICCs), i.e. a summary statistic describing resemblance of units between groups, and their 95% confidence intervals (95% CI). ICCs and coefficients of variation (CV), i.e. ratio of the standard deviation to the mean, were used to test absolute reliability (measurement independent of observer) [9]. Hopkins (2000) and Atkinson and Nevill (1998) strongly recommend providing more than one measure of reliability [10, 11]. Thus, ICCs were supplemented with standard error of measurement (SEM), i.e. the standard deviation of measurement errors, as this measure is more explicit in clinical context by giving results in the same units as the measurements [12]. The most common interpretation of ICC values follows Cicchetti’s guidelines: values in range 0.00–0.39 are considered poor correlation, 0.40–0.59—fair correlation, 0.60–0.75—good correlation and 0.75–1.00—excellent correlation [13]. A more recent clinically oriented interpretation suggests that values below 0.75 represent poor to moderate correlation, values between 0.75 and 0.87 are good correlation, and values above 0.87 represent correlation high enough to be regarded as clinical measures [14]. Accepted levels of measurement deviation for the thoracic region are below 1 cm [15]. Hence, such a range will be classified as good accuracy.
To show clinical reliability, the Bland–Altman plot was created, juxtaposing differences between values measured directly and in BI, and the mean of three measurements [16, 17]. Calculations were performed using Statistica v13 (Dell Inc., Round Rock, TX). P value ≤ 0.05 was considered as the threshold of statistical significance.

Results

Accuracy and precision for both female and male subjects, as well as inter-rater variability, are shown in Tables 12 and 3. Absolute differences between digital measurements and reference measurements lie below the 1-cm demarcation line. There are, however, accuracy drops for specific measurements (medial line–nipple distance).
Table 1
Summary of three-session validation for female subjects
Measurement
Observer A
Observer B
Accuracy
Precision
Accuracy
Precision
MAD (cm)
SD (cm)
ICC
95% CI
SEM (cm)
MAD (cm)
SD (cm)
ICC
95% CI
SEM (cm)
NLD
0.30
0.29
0.944
0.847–0.984
0.07
0.41
0.31
0.988
0.965–0.997
0.03
MLN
1.42
1.06
0.935
0.824–0.982
0.27
1.71
1.08
0.970
0.915–0.992
0.19
MLN*
1.91
1.21
0.930
0.812–0.980
0.32
1.61
1.03
0.958
0.883–0.988
0.21
SNN
0.11
1.38
0.960
0.889–0.989
0.28
0.15
1.07
0.977
0.935–0.994
0.16
SNN*
0.12
1.30
0.952
0.868–0.987
0.28
0.14
0.94
0.969
0.913–0.991
0.17
IMFLD
0.34
0.30
0.951
0.865–0.986
0.07
0.29
0.24
0.962
0.894–0.990
0.05
IMFML
0.39
0.99
0.935
0.824–0.982
0.25
0.26
1.04
0.926
0.802–0.979
0.28
IMF*ML
0.14
1.03
0.936
0.827–0.982
0.26
0.50
1.31
0.985
0.957–0.996
0.16
IMFN
0.32
1.25
0.843
0.617–0.954
0.50
0.53
1.51
0.948
0.857–0.986
0.34
IMF*N*
0.26
1.06
0.826
0.584–0.949
0.44
0.05
1.08
0.955
0.875–0.988
0.23
UPLD
0.57
1.44
0.905
0.752–0.973
0.45
0.70
1.33
0.932
0.817–0.981
0.35
UPN
0.58
1.18
0.992
0.977–0.998
0.11
0.68
1.00
0.988
0.965–0.997
0.11
UPN*
0.70
1.22
0.993
0.980–0.998
0.10
1.00
1.04
0.983
0.951–0.995
0.14
MAD mean absolute difference, SD standard deviation, ICC intra-class correlation coefficient, CI confidence interval, SEM standard error of measurement, * left, NLD nipple levels difference, MLN medial line to nipple distance, SNN sternal notch to nipple distance, IMFLD inframammary fold apex levels difference, IMFML inframammary fold apex to medial line distance, IMFN inframammary fold to nipple distance, UPLD upper pole apex levels difference, UPN upper pole apex to nipple distance
Table 2
Summary of three-session validation for male subjects
Measurement
Observer A
Observer B
Accuracy
Precision
Accuracy
Precision
MAD (cm)
SD (cm)
ICC
95% CI
SEM (cm)
MAD (cm)
SD (cm)
ICC
95% CI
SEM (cm)
NLD
0.36
0.20
0.887
0.711–0.968
0.07
0.31
0.18
0.953
0.870–0.987
0.04
MLN
0.47
0.46
0.986
0.960–0.996
0.05
0.62
0.47
0.996
0.988–0.999
0.03
MLN*
0.21
0.40
0.984
0.954–0.996
0.05
0.33
0.39
0.994
0.983–0.998
0.03
SNN
0.89
0.84
0.973
0.924–0.993
0.14
1.12
0.75
0.994
0.983–0.998
0.06
SNN*
0.74
0.77
0.973
0.924–0.993
0.13
0.90
0.68
0.991
0.974–0.998
0.06
IMFLD
0.27
0.22
0.909
0.761–0.974
0.07
0.26
0.22
0.950
0.862–0.986
0.05
IMFML
0.32
0.50
0.987
0.963–0.996
0.06
0.27
0.43
0.993
0.980–0.998
0.04
IMF*ML
0.09
0.59
0.972
0.921–0.992
0.10
0.07
0.54
0.976
0.932–0.993
0.08
IMFN
0.15
0.34
0.988
0.965–0.997
0.04
0.23
0.34
0.990
0.971–0.997
0.03
IMF*N*
0.37
0.38
0.979
0.940–0.994
0.06
0.41
0.41
0.997
0.991–0.999
0.02
UPLD
0.16
0.35
0.963
0.897–0.990
0.07
0.27
0.29
0.959
0.886–0.989
0.06
UPN
0.11
0.34
0.991
0.974–0.998
0.03
0.20
0.28
0.998
0.994–0.999
0.01
UPN*
0.14
0.32
0.996
0.988–0.999
0.02
0.29
0.31
0.999
0.997–1.000
0.01
MAD mean absolute difference, SD standard deviation, ICC intra-class correlation coefficient, CI confidence interval, SEM standard error of measurement, * left, NLD nipple levels difference, MLN medial line to nipple distance, SNN sternal notch to nipple distance, IMFLD inframammary fold apex levels difference, IMFML inframammary fold apex to medial line distance, IMFN inframammary fold to nipple distance, UPLD upper pole apex levels difference, UPN upper pole apex to nipple distance
Table 3
Summary of intra-observer reliability between Observer A and Observer B, based on average values from three sessions
Measurement
Female subjects
Male subjects
CV (%)
ICC
95% CI
SEM (cm)
CV (%)
ICC
95% CI
SEM (cm)
NLD
5.7
0.968
0.910–0.991
0.01
6.3
0.943
0.844–0.984
0.01
MLN
1.2
0.968
0.910–0.991
0.02
0.7
0.996
0.988–0.999
0.00
MLN*
0.1
0.924
0.797–0.979
0.00
0.5
0.998
0.994–0.999
0.00
SNN
0.1
0.989
0.968–0.997
0.00
0.6
0.996
0.988–0.999
0.01
SNN*
0.0
0.986
0.960–0.996
0.00
0.4
0.997
0.991–0.999
0.00
IMFLD
5.8
0.909
0.761–0.974
0.01
0.4
0.957
0.881–0.988
0.00
IMFML
2.2
0.953
0.870–0.987
0.04
0.2
0.992
0.977–0.998
0.00
IMF*ML
3.0
0.858
0.648–0.959
0.11
0.1
0.995
0.985–0.999
0.00
IMFN
4.5
0.810
0.553–0.943
0.15
0.9
0.998
0.994–0.999
0.00
IMF*N*
1.2
0.898
0.736–0.971
0.04
0.9
0.995
0.985–0.999
0.01
UPLD
4.9
0.944
0.847–0.984
0.09
0.5
0.967
0.907–0.991
0.00
UPN
2.3
0.987
0.963–0.996
0.02
0.6
0.998
0.994–0.999
0.00
UPN*
3.8
0.952
0.868–0.987
0.08
1.0
0.999
0.997–1.000
0.00
CV coefficient of variation, ICC intra-class correlation coefficient, CI confidence interval, SEM standard error of measurement, * left, NLD nipple levels difference, MLN medial line to nipple distance, SNN sternal notch to nipple distance, IMFLD inframammary fold apex levels difference, IMFML inframammary fold apex to medial line distance, IMFN inframammary fold to nipple distance, UPLD upper pole apex levels difference, UPN upper pole apex to nipple distance
Precision (consistency of measurement) as well as inter-rater reliability, is at clinical measures level (or excellent according to Cichetti) for all but two measurements (female subjects' right inframammary fold apex to right nipple distance, which was 0.060 below the limit and left inframammary fold to medial line distance, 0.012 below the limit). Precision ICC values have a larger variance for female subjects (0.001415) than for male subjects (0.000705). Absolute reliability ICC values also have a larger variance for female subjects (0.00290) when compared to male subjects (0.0003465).
Standard error of measurement is consistent with ICC values. It is mostly below 0.1 cm for female subjects and close to zero for male subjects. Coefficient-of-variation values are the highest (around 5–6%) for level differences of nipples, inframammary fold apices and upper pole apices.
The Bland–Altman plots (Fig. 5a–d) show how the overall data map out when juxtaposed with the reference data (direct measurements).
The effectiveness (probability of satisfactory result) of BI equals 87.5% for females and 95.8% for males.
Breast assessment using BI during the first measurement session took around 1.5 min (with an eight-second difference when two photographs were used) for females and about a minute for males. The average times in the last session are roughly 30% shorter for female subjects and 20% shorter for male subjects (Table 4).
Table 4
Summary of measurements average duration
 
Female subjects
Male subjects
Average (min)
Average (min)
1 photograph
2 photographs
Session 1
1.43 ± 0.28
1.56 ± 0.22
0.97 ± 0.09
Session 2
1.21 ± 0.25
1.41 ± 0.25
0.83 ± 0.10
Session 3
0.96 ± 0.18
1.15 ± 0.09
0.82 ± 0.09

Discussion

The study presents validation of BI—software dedicated to indirect breast anthropometry. It was designed for clinical purposes; however, it can also be helpful in scientific analyses. The process of validation included male and female participants, regardless of breast anatomy, which mimics the variety of plastic surgeons’ patients. The measurements performed by two examiners of different clinical experience using BI were compared to ‘gold standard’ direct measurements performed by a plastic surgeon. High consistency of the results suggests high repeatability of measurements independent of examiner’s experience.
As proved by Odo et al. [18], direct anthropometry is a valid method to depict differences between breasts when comparing asymmetry before and after surgery [18]. However, direct measurements have several drawbacks. While some body regions can be assessed in a mostly linear manner, the chest has prominent curvatures, making measurements less accurate [2, 19, 20]. Also, the region is not only affected by whole-body movements, but also ribcage expansion while breathing. Indirect anthropometry on the other hand presents a wide array of advantages. According to numerous studies, not only does it allow higher reliability, but it is also much less stressful for the patient [19, 2123]. Data are also usually relatively easy to record, already being in digital format. Nonetheless, indirect anthropometry of the breasts based on digital photographs using standard software also fails in some aspects. The programs offer satisfactory precision and accuracy only for a specific group of patients [5]. For example, for women with ptotic breast these programs fail to produce highly reliable results. Adapting programs like Autodesk AutoCAD® or Adobe Photoshop® for such use imposes the need to train medical staff and create a specific list of commands, resulting in a nonlinear workflow [5]. The programs are vast packages, offering a number of tools needed in the industry they were designed for, while in clinical practice their capabilities can prove daunting and distracting for the user [24, 25].
The middle ground between direct anthropometry and measurements using two-dimensional photographs is photogrammetry [26]. It is a valid approach; however, producing stable, good results requires a lot of experience, especially for patients with severe breast ptosis. Another drawback is the need for lengthy computation using a high-end computer or external cloud service (unadvised for sensitive data) [27].
In this paper, we presented the self-designed tool (and its validation), which does not require a lot of time from a user to become confident in usage, being as well capable of running on any modern device without specific hardware requirements. Because of the natural differences between male and female chests, it was crucial to validate the tool in female and male groups. We found that most accuracy values for female subjects vary by ± 1 cm, consistently between observers. The exceptions are nipple level differences and inframammary fold apex level differences. Their standard deviation was around 0.30 cm. Standard deviation of mean absolute difference above 1 cm was not detected in measurements in males. BI measurements, which were juxtaposed with measurements mostly affected by curves in direct anthropometry are least accurate, with a mean absolute difference between 1.42 and 1.91 cm (medial line–nipple distance). This may result from the fact that for distances like the one traced from the nipple to the body’s medial line, one should demarcate a line segment tangent to the natural breast curve, which is not a routine approach in clinical breast measurements [2]. It appeared that BI gives valid results when compared to a textbook procedure (measuring straight lines instead of following natural curves) [2]. The discussed problem is also depicted in the Bland–Altman graphs, where the majority of points fall between the two limits, proving suitability for clinical use, with a cluster of outlying data present for female subjects for both observers, corresponding to the measurements listed above.
To properly evaluate BI’s accuracy, we need to consider measurements which were truly linear both in the direct and indirect anthropometry using the web application. These measurements (sternal notch to nipple distance, medial line to inframammary fold apex distance, upper pole apex to nipple distance) mostly present satisfactory accuracy with excellent correspondence between observers. MAD of the distance from the inframammary fold apex to the medial line was initially above 2 cm, but decreased to about 0.30 cm within the next two sessions. What is more, the average accuracy of both observers grew with each session, accompanied by a significant decrease in procedure duration and time variation between different subjects. In the case of the distance from the inframammary fold apex to the medial line, we suppose the progress can be attributed to better visual extrapolation of anthropometric landmarks—applying experience from assessing subjects, whose specific points were more distinct, to female subjects with breasts that are more difficult to assess. This would also explain why there was no accuracy increase for male subjects—their chests were much easier to measure correctly.
For both male and female subjects, nearly all the ICC values fall above 0.87 and can be considered clinical measures—most reach at least 0.93. Reliability of direct anthropometry lies between 0.03 and 0.25 cm [28]. The SEM values both for inter and intra-rater reliability prove BI’s reliability is higher. Manual measurements usually have a coefficient of variation of 5% or more [28]. Although this percentage is still acceptable for the thoracic region, measurements with BI, on average, scored less than 2.7% for female subjects and 1.0% for male subjects.
When compared to graphics software adapted for breast measurements, BI’s accuracy is similar. However, mean absolute differences for female breasts obtained using BI generally deviate more when referred to standard deviation values presented by Quieregatto et al., what may be attributed to the fact, that their direct anthropometry measurements were made using callipers (truly linear measurement), while we used a tape measure and introduced curvature bias [5]. Analysing BI in comparison to 3D scanners, it should be highlighted that low- to medium-tier 3D scanners provide a reliability comparable to manual methods [2931]. When comparing BI to the 3D scanner used by Conkle et al., the accuracy results for linear measurements are similar, while reliability outcomes are somewhat better [24]. On the other hand, BI’s accuracy is hindered by pixel resolution and is ten times lower than accuracy of an Artec Eva high-precision laser 3D scanner (Artec, Luxembourg) used by Seminati et al., while achieving comparable reliability [32].
What is more, even though we designed the application with ease of use as our top priority, we did not manage to avoid a learning curve for the user. While precision did not show a major change during the three sessions, accuracy and measurement time improved. The progress was noticeable for the assessment of female breasts.
Last but not least, striving for aesthetic breast assessment, we have used ratios proposed by Mallucci et al. and Lewin et al. [7, 8]. When performing direct measurements, the clinical practitioner had to calculate those proportions themselves. In BI the ratios are automatically calculated, saving the observer some effort.
One of the main limitations of the study was the number of participants. We decided ten males and ten females were sufficient to validate the application, especially when the stated data quality procedures were taken into account. The problem was tackled by random choice of subjects, resulting in a wide array of different breast shapes and sizes. However, this may also have resulted in data bias, so further study with a larger sample of patients should be performed. It is possible that separate validations should have been conducted for different breast deformations. Further limitation is related to the specificity of using indirect anthropometry—the technique that cannot address skin extensibility. On the other hand, skin extensibility makes direct measurement dependent on the technique of taking the measurements by certain surgeons, which is overcome by standardised indirect technique.

Conclusion

To conclude, we found that BI may be a useful tool in breast surgeon’s clinical practice. It provides an objective method of breast measurement, offers high accuracy and precision without the need to process 3D data and omits the need for drawing dots and lines on the patient’s body. The user can accurately assess asymmetry, not being limited by chest deformations, severe ptosis or the necessity to take photographs in specific conditions. Finally, by using a digital medium, breast measurements no longer have to be written on physical paper or typed in a spreadsheet, while the report generated by BI may help analyse the treatment plan and consult it with another specialist regardless of the patient’s presence.

Compliance with Ethical Standards

Conflict of interest

All authors declare that there are no conflicts of interest in this work.

Ethical Approval

The protocol of the study was approved by the Bioethics Committee of the Medical University (RNN/100/18/KE).
Written informed consent was obtained from all subjects.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

Die Chirurgie

Print-Titel

Das Abo mit mehr Tiefe

Mit der Zeitschrift Die Chirurgie erhalten Sie zusätzlich Online-Zugriff auf weitere 43 chirurgische Fachzeitschriften, CME-Fortbildungen, Webinare, Vorbereitungskursen zur Facharztprüfung und die digitale Enzyklopädie e.Medpedia.

Bis 30. April 2024 bestellen und im ersten Jahr nur 199 € zahlen!

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Dent – Das Online-Abo der Zahnmedizin

Online-Abonnement

Mit e.Dent erhalten Sie Zugang zu allen zahnmedizinischen Fortbildungen und unseren zahnmedizinischen und ausgesuchten medizinischen Zeitschriften.

Literatur
1.
Zurück zum Zitat Choppin SB, Wheat JS, Gee M et al (2016) The accuracy of breast volume measurement methods: a systematic review. Breast 28:121–129CrossRefPubMed Choppin SB, Wheat JS, Gee M et al (2016) The accuracy of breast volume measurement methods: a systematic review. Breast 28:121–129CrossRefPubMed
2.
Zurück zum Zitat Westreich M (1997) Anthropomorphic breast measurement: protocol and results in 50 women with aesthetically perfect breasts and clinical application. Plast Reconstr Surg 100(2):468–479CrossRefPubMed Westreich M (1997) Anthropomorphic breast measurement: protocol and results in 50 women with aesthetically perfect breasts and clinical application. Plast Reconstr Surg 100(2):468–479CrossRefPubMed
3.
Zurück zum Zitat Watmough DJ (1982) Diaphanography: mechanism responsible for the images. Acta Radiol Oncol 21(1):11–15CrossRefPubMed Watmough DJ (1982) Diaphanography: mechanism responsible for the images. Acta Radiol Oncol 21(1):11–15CrossRefPubMed
4.
Zurück zum Zitat Daneshmand M, Helmi A, Avots E et al (2018) 3D scanning: a comprehensive survey. arXiv:801.08863v1 [cs.CV]. January 24, 2018. Available from: Cornell University, Ithaca, NY. Accessed 22 Aug 2018 Daneshmand M, Helmi A, Avots E et al (2018) 3D scanning: a comprehensive survey. arXiv:​801.​08863v1 [cs.CV]. January 24, 2018. Available from: Cornell University, Ithaca, NY. Accessed 22 Aug 2018
5.
Zurück zum Zitat Quieregatto PR, Hochman B, Furtado F et al (2014) Image analysis software versus direct anthropometry for breast measurements. Acta Cir Bras 29(10):688–695CrossRefPubMed Quieregatto PR, Hochman B, Furtado F et al (2014) Image analysis software versus direct anthropometry for breast measurements. Acta Cir Bras 29(10):688–695CrossRefPubMed
6.
Zurück zum Zitat Flanagan D (2011) JavaScript: the definitive guide, 6th edn. O’Reilly Media, Sebastopol Flanagan D (2011) JavaScript: the definitive guide, 6th edn. O’Reilly Media, Sebastopol
7.
Zurück zum Zitat Mallucci P, Branford OA (2014) Population analysis of the perfect breast: a morphometric analysis. Plast Reconstr Surg 134(3):436–447CrossRefPubMed Mallucci P, Branford OA (2014) Population analysis of the perfect breast: a morphometric analysis. Plast Reconstr Surg 134(3):436–447CrossRefPubMed
8.
Zurück zum Zitat Lewin R, Amoroso M, Plate N et al (2016) The aesthetically ideal position of the nipple–areola complex on the breast. Aesth Plast Surg 40:724–732CrossRef Lewin R, Amoroso M, Plate N et al (2016) The aesthetically ideal position of the nipple–areola complex on the breast. Aesth Plast Surg 40:724–732CrossRef
9.
Zurück zum Zitat Bruton A, Conway JH, Holgate ST (2000) Reliability: what is it, and how is it measured? Physiotherapy 86(2):94–99CrossRef Bruton A, Conway JH, Holgate ST (2000) Reliability: what is it, and how is it measured? Physiotherapy 86(2):94–99CrossRef
10.
Zurück zum Zitat Hopkins WG (2000) Measures of reliability in sports medicine and science. Sports Med 30(1):1–15CrossRefPubMed Hopkins WG (2000) Measures of reliability in sports medicine and science. Sports Med 30(1):1–15CrossRefPubMed
11.
Zurück zum Zitat Atkinson G, Nevill AM (1998) Statistical methods for assessing measurement error (reliability) in variables relevant to sports medicine. Sports Med 26(4):217–238CrossRefPubMed Atkinson G, Nevill AM (1998) Statistical methods for assessing measurement error (reliability) in variables relevant to sports medicine. Sports Med 26(4):217–238CrossRefPubMed
12.
Zurück zum Zitat Šerbetar I (2015) Establishing some measures of absolute and relative reliability of a motor tests. Croat J Educ 17(1):37–48 Šerbetar I (2015) Establishing some measures of absolute and relative reliability of a motor tests. Croat J Educ 17(1):37–48
13.
Zurück zum Zitat Cicchetti DV (1994) Guidelines, criteria, and rules of thumb for evaluating normed and standardized assessment instruments in psychology. Psychol Assess 6(4):284–290CrossRef Cicchetti DV (1994) Guidelines, criteria, and rules of thumb for evaluating normed and standardized assessment instruments in psychology. Psychol Assess 6(4):284–290CrossRef
14.
Zurück zum Zitat Portney LG, Watkins MP (2000) Foundations of clinical research: applications to practice. Prentice Hall Inc., New Jersey Portney LG, Watkins MP (2000) Foundations of clinical research: applications to practice. Prentice Hall Inc., New Jersey
15.
Zurück zum Zitat Mony PK, Swaminathan S, Gajendran JK et al (2016) Quality assurance for accuracy of anthropometric measurements in clinical and epidemiological studies: [Errare humanum est = to err is human]. Indian J Community Med 41(2):98–102CrossRefPubMedPubMedCentral Mony PK, Swaminathan S, Gajendran JK et al (2016) Quality assurance for accuracy of anthropometric measurements in clinical and epidemiological studies: [Errare humanum est = to err is human]. Indian J Community Med 41(2):98–102CrossRefPubMedPubMedCentral
16.
Zurück zum Zitat Altman DG, Bland JM (1983) Measurement in medicine: the analysis of method comparison studies. J R Stat Soc D 32:307–317 Altman DG, Bland JM (1983) Measurement in medicine: the analysis of method comparison studies. J R Stat Soc D 32:307–317
17.
Zurück zum Zitat Giavarina D (2015) Understanding bland altman analysis. Biochem Med 25(2):141–151CrossRef Giavarina D (2015) Understanding bland altman analysis. Biochem Med 25(2):141–151CrossRef
18.
Zurück zum Zitat Odo LM, Guimarães PA, Silva AL et al (2009) Assessing the outcome of surgical treatment of breast asymmetry by means of linear measures. Arq Catarin Med 38(Suppl. 1):43–45 Odo LM, Guimarães PA, Silva AL et al (2009) Assessing the outcome of surgical treatment of breast asymmetry by means of linear measures. Arq Catarin Med 38(Suppl. 1):43–45
19.
Zurück zum Zitat Nechala P, Mahoney J, Farkas LG (1999) Digital two-dimensional photogrammetry: a comparison of three techniques of obtaining digital photographs. Plast Reconstr Surg 103(7):1819–1825CrossRefPubMed Nechala P, Mahoney J, Farkas LG (1999) Digital two-dimensional photogrammetry: a comparison of three techniques of obtaining digital photographs. Plast Reconstr Surg 103(7):1819–1825CrossRefPubMed
20.
Zurück zum Zitat Pietruski P, Majak M, Dębski T et al (2017) A novel computer system for the evaluation of nasolabial morphology, symmetry and aesthetics after cleft lip and palate treatment. Part 1: general concept and validation. J Craniomaxillofac Surg 45(4):491–504CrossRefPubMed Pietruski P, Majak M, Dębski T et al (2017) A novel computer system for the evaluation of nasolabial morphology, symmetry and aesthetics after cleft lip and palate treatment. Part 1: general concept and validation. J Craniomaxillofac Surg 45(4):491–504CrossRefPubMed
21.
Zurück zum Zitat Zancanaro C, Milanese C, Lovato C et al (2015) Reliability of three-dimensional photonic scanner anthropometry performed by skilled and naïve operators. IJEG 5(1):1–11 Zancanaro C, Milanese C, Lovato C et al (2015) Reliability of three-dimensional photonic scanner anthropometry performed by skilled and naïve operators. IJEG 5(1):1–11
22.
Zurück zum Zitat Medina-Inojosa J, Somers VK, Ngwa T et al (2016) Reliability of a 3D body scanner for anthropometric measurements of central obesity. Obes Open Access 2(3):1–4 Medina-Inojosa J, Somers VK, Ngwa T et al (2016) Reliability of a 3D body scanner for anthropometric measurements of central obesity. Obes Open Access 2(3):1–4
23.
Zurück zum Zitat Conkle J, Suchdev PS, Alexander E et al (2018) Accuracy and reliability of a low-cost, handheld 3D imaging system for child anthropometry. PLoS One 13(10):e0205320CrossRefPubMedPubMedCentral Conkle J, Suchdev PS, Alexander E et al (2018) Accuracy and reliability of a low-cost, handheld 3D imaging system for child anthropometry. PLoS One 13(10):e0205320CrossRefPubMedPubMedCentral
26.
Zurück zum Zitat Schenk T (2005) Introduction to photogrammetry, 1st edn. The Ohio State University, Columbus Schenk T (2005) Introduction to photogrammetry, 1st edn. The Ohio State University, Columbus
27.
Zurück zum Zitat Hashizume K, Rosado DG, Fernández-Medina E et al (2013) An analysis of security issues for cloud computing. J Internet Serv Appl 4(5):1–1326 Hashizume K, Rosado DG, Fernández-Medina E et al (2013) An analysis of security issues for cloud computing. J Internet Serv Appl 4(5):1–1326
28.
Zurück zum Zitat Gordon CC, Blackwell CL, Bradtmiller B et al (2012) Anthropometric survey of U.S. Army Personnel: methods and summary statistics. U.S. Army Natick Soldier Research, Development and Engineering Center, Natick Gordon CC, Blackwell CL, Bradtmiller B et al (2012) Anthropometric survey of U.S. Army Personnel: methods and summary statistics. U.S. Army Natick Soldier Research, Development and Engineering Center, Natick
29.
Zurück zum Zitat Koepke N, Zwahlen M, Wells JC et al (2017) Comparison of 3D laser-based photonic scans and manual anthropometric measurements of body size and shape in a validation study of 123 young Swiss men. Peer J 5:e2980CrossRefPubMed Koepke N, Zwahlen M, Wells JC et al (2017) Comparison of 3D laser-based photonic scans and manual anthropometric measurements of body size and shape in a validation study of 123 young Swiss men. Peer J 5:e2980CrossRefPubMed
30.
Zurück zum Zitat Parker C, Gill S, Hayes S (2017) 3D body scanning has suitable reliability: an anthropometric investigation for garment construction. In: Proceedings of 3DBODY.TECH, pp 298–305 Parker C, Gill S, Hayes S (2017) 3D body scanning has suitable reliability: an anthropometric investigation for garment construction. In: Proceedings of 3DBODY.TECH, pp 298–305
31.
Zurück zum Zitat Bougourd JP, Dekker L, Grant Ross P et al (2000) A comparison of women’s sizing by 3D electronic scanning and traditional anthropometry. J Tex I 91(2):163–173 Bougourd JP, Dekker L, Grant Ross P et al (2000) A comparison of women’s sizing by 3D electronic scanning and traditional anthropometry. J Tex I 91(2):163–173
32.
Zurück zum Zitat Seminati E, Canepa Talamas D, Young M et al (2017) Validity and reliability of a novel 3D scanner for assessment of the shape and volume of amputees’ residual limb models. PLoS ONE 12(9):e0184498CrossRefPubMedPubMedCentral Seminati E, Canepa Talamas D, Young M et al (2017) Validity and reliability of a novel 3D scanner for assessment of the shape and volume of amputees’ residual limb models. PLoS ONE 12(9):e0184498CrossRefPubMedPubMedCentral
Metadaten
Titel
A New Tool for Breast Anthropometric Measurements: Presentation and Validation for Women and Men
verfasst von
Maksym Mikołajczyk
Anna Kasielska-Trojan
Bogusław Antoszewski
Publikationsdatum
01.08.2019
Verlag
Springer US
Erschienen in
Aesthetic Plastic Surgery / Ausgabe 5/2019
Print ISSN: 0364-216X
Elektronische ISSN: 1432-5241
DOI
https://doi.org/10.1007/s00266-019-01467-6

Weitere Artikel der Ausgabe 5/2019

Aesthetic Plastic Surgery 5/2019 Zur Ausgabe

Wie erfolgreich ist eine Re-Ablation nach Rezidiv?

23.04.2024 Ablationstherapie Nachrichten

Nach der Katheterablation von Vorhofflimmern kommt es bei etwa einem Drittel der Patienten zu Rezidiven, meist binnen eines Jahres. Wie sich spätere Rückfälle auf die Erfolgschancen einer erneuten Ablation auswirken, haben Schweizer Kardiologen erforscht.

Hinter dieser Appendizitis steckte ein Erreger

23.04.2024 Appendizitis Nachrichten

Schmerzen im Unterbauch, aber sonst nicht viel, was auf eine Appendizitis hindeutete: Ein junger Mann hatte Glück, dass trotzdem eine Laparoskopie mit Appendektomie durchgeführt und der Wurmfortsatz histologisch untersucht wurde.

Mehr Schaden als Nutzen durch präoperatives Aussetzen von GLP-1-Agonisten?

23.04.2024 Operationsvorbereitung Nachrichten

Derzeit wird empfohlen, eine Therapie mit GLP-1-Rezeptoragonisten präoperativ zu unterbrechen. Eine neue Studie nährt jedoch Zweifel an der Notwendigkeit der Maßnahme.

Ureterstriktur: Innovative OP-Technik bewährt sich

19.04.2024 EAU 2024 Kongressbericht

Die Ureterstriktur ist eine relativ seltene Komplikation, trotzdem bedarf sie einer differenzierten Versorgung. In komplexen Fällen wird dies durch die roboterassistierte OP-Technik gewährleistet. Erste Resultate ermutigen.

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.