Accuracy, agreement, and reliability of DECT-derived vBMD measurements: an initial ex vivo study

verfasst von: Le Qin, Jingjing Huang, Pei Yu, Jing Yan, Yingqian Ge, Yong Lu, Fuhua Yan, Lei Wang, Lianjun Du

Erschienen in: European Radiology | Ausgabe 1/2021

Einloggen, um Zugang zu erhalten

Abstract

Objectives

To assess the agreement and reliability of DECT (dual-energy CT)-derived vBMD (volumetric bone mineral density) measurements from excised human femoral heads and to compare DECT-derived BMD with that measured by DXA (dual-energy X-ray absorptiometry) and QCT (quantitative CT) to determine its accuracy.

Methods

Twenty patients that underwent total hip arthroplasty were enrolled to this study. Femoral heads were excised to rectangles without cortical bones for scanning. A dual-source DECT scanner generated images under 80/Sn140 kVp and 100/Sn140 kVp scanning conditions. Specimens were subsequently scanned by QCT and DXA to produce QCT-derived vBMD (mg/cm3) and DXA-derived BMM (bone mineral mass, g). DECT images were loaded to a post-processing workstation to calculate DECT-derived vBMD and BMM.

Results

Higher DECT-derived vBMD and BMM were found under 80/Sn140 and 100/Sn140 kVp compared with those for QCT and DXA (p = 0.005). DECT-derived vBMD was highly correlated with QCT-derived vBMD (r = 0.961 ~ 0.993, p < 0.05). Similarly, DECT-derived BMM was strongly correlated with DXA-derived BMM (r = 0.927 ~ 0.943, p < 0.05). Agreement of the inter- and intra-observation of DECT-derived vBMD was excellent. Linear regression was carried out to calibrate DECT-derived vBMD of 80/Sn140 kVp (14 + 0.7 × DECT-derived vBMD) and 100/Sn140 kVp (74 + 0.4 × DECT-derived vBMD) with the reference of QCT-derived vBMD. After calibration, excellent agreement was found for vBMD and BMM within various imaging modalities.

Conclusions

Our study showed that DECT-derived vBMD exhibited high agreement and reliability features, and after calibration, it also displayed a high degree of accuracy. However, in vivo studies are needed to extend its potential utility in clinical settings.

Key Points

• Measurements of DECT-derived vBMD had high intra- and inter-observer agreement and reliability.
• Measurements of DECT-derived vBMD and BMM had a high correlation with those derived from QCT and DXA.
• DECT-derived vBMD and BMM were accurate after calibration compared with QCT and DXA.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Chen P, Li Z, Hu Y (2016) Prevalence of osteoporosis in China: a meta-analysis and systematic review. BMC Public Health 16:1039CrossRef Chen P, Li Z, Hu Y (2016) Prevalence of osteoporosis in China: a meta-analysis and systematic review. BMC Public Health 16:1039CrossRef
2.
Zurück zum Zitat Lupsa BC, Insogna K (2015) Bone health and osteoporosis. Endocrinol Metab Clin North Am 44:517–530CrossRef Lupsa BC, Insogna K (2015) Bone health and osteoporosis. Endocrinol Metab Clin North Am 44:517–530CrossRef
3.
Zurück zum Zitat Bass MA, Sharma A, Nahar VK et al (2019) Bone mineral density among men and women aged 35 to 50 years. J Am Osteopath Assoc 119:357CrossRef Bass MA, Sharma A, Nahar VK et al (2019) Bone mineral density among men and women aged 35 to 50 years. J Am Osteopath Assoc 119:357CrossRef
4.
Zurück zum Zitat Oei L, Koromani F, Rivadeneira F, Zillikens MC, Oei EH (2016) Quantitative imaging methods in osteoporosis. Quant Imaging Med Surg 6:680–698CrossRef Oei L, Koromani F, Rivadeneira F, Zillikens MC, Oei EH (2016) Quantitative imaging methods in osteoporosis. Quant Imaging Med Surg 6:680–698CrossRef
5.
Zurück zum Zitat Black DM, Rosen CJ (2016) Clinical Practice. Postmenopausal Osteoporosis. N Engl J Med 374:254–262 Black DM, Rosen CJ (2016) Clinical Practice. Postmenopausal Osteoporosis. N Engl J Med 374:254–262
6.
Zurück zum Zitat Lewiecki EM, Kendler DL, Davison KS et al (2019) Western osteoporosis alliance clinical practice series: treat-to-target for osteoporosis. Am J Med 132:e771–e777CrossRef Lewiecki EM, Kendler DL, Davison KS et al (2019) Western osteoporosis alliance clinical practice series: treat-to-target for osteoporosis. Am J Med 132:e771–e777CrossRef
7.
Zurück zum Zitat Heiss C, Govindarajan P, Schlewitz G et al (2012) Induction of osteoporosis with its influence on osteoporotic determinants and their interrelationships in rats by DEXA. Med Sci Monit 18:BR199–BR207CrossRef Heiss C, Govindarajan P, Schlewitz G et al (2012) Induction of osteoporosis with its influence on osteoporotic determinants and their interrelationships in rats by DEXA. Med Sci Monit 18:BR199–BR207CrossRef
8.
Zurück zum Zitat Dall'Ara E, Pahr D, Varga P, Kainberger F, Zysset P (2012) QCT-based finite element models predict human vertebral strength in vitro significantly better than simulated DEXA. Osteoporos Int 23:563–572CrossRef Dall'Ara E, Pahr D, Varga P, Kainberger F, Zysset P (2012) QCT-based finite element models predict human vertebral strength in vitro significantly better than simulated DEXA. Osteoporos Int 23:563–572CrossRef
9.
Zurück zum Zitat Amstrup AK, Jakobsen NF, Moser E, Sikjaer T, Mosekilde L, Rejnmark L (2016) Association between bone indices assessed by DXA, HR-pQCT and QCT scans in post-menopausal women. J Bone Miner Metab 34:638–645 Amstrup AK, Jakobsen NF, Moser E, Sikjaer T, Mosekilde L, Rejnmark L (2016) Association between bone indices assessed by DXA, HR-pQCT and QCT scans in post-menopausal women. J Bone Miner Metab 34:638–645
10.
Zurück zum Zitat Laugerette A, Schwaiger BJ, Brown K et al (2019) DXA-equivalent quantification of bone mineral density using dual-layer spectral CT scout scans. Eur Radiol 29:4624–4634CrossRef Laugerette A, Schwaiger BJ, Brown K et al (2019) DXA-equivalent quantification of bone mineral density using dual-layer spectral CT scout scans. Eur Radiol 29:4624–4634CrossRef
11.
Zurück zum Zitat Engelke K, Libanati C, Fuerst T, Zysset P, Genant HK (2013) Advanced CT based in vivo methods for the assessment of bone density, structure, and strength. Curr Osteoporos Rep 11:246–255CrossRef Engelke K, Libanati C, Fuerst T, Zysset P, Genant HK (2013) Advanced CT based in vivo methods for the assessment of bone density, structure, and strength. Curr Osteoporos Rep 11:246–255CrossRef
12.
Zurück zum Zitat van Hamersvelt RW, Schilham AMR, Engelke K et al (2017) Accuracy of bone mineral density quantification using dual-layer spectral detector CT: a phantom study. Eur Radiol 27:4351–4359CrossRef van Hamersvelt RW, Schilham AMR, Engelke K et al (2017) Accuracy of bone mineral density quantification using dual-layer spectral detector CT: a phantom study. Eur Radiol 27:4351–4359CrossRef
13.
Zurück zum Zitat Mei K, Kopp FK, Bippus R et al (2017) Is multidetector CT-based bone mineral density and quantitative bone microstructure assessment at the spine still feasible using ultra-low tube current and sparse sampling? Eur Radiol 27:5261–5271CrossRef Mei K, Kopp FK, Bippus R et al (2017) Is multidetector CT-based bone mineral density and quantitative bone microstructure assessment at the spine still feasible using ultra-low tube current and sparse sampling? Eur Radiol 27:5261–5271CrossRef
14.
Zurück zum Zitat Yasaka K, Akai H, Kunimatsu A, Kiryu S, Abe O (2020) Prediction of bone mineral density from computed tomography: application of deep learning with a convolutional neural network. Eur Radiol 30:3549–3557CrossRef Yasaka K, Akai H, Kunimatsu A, Kiryu S, Abe O (2020) Prediction of bone mineral density from computed tomography: application of deep learning with a convolutional neural network. Eur Radiol 30:3549–3557CrossRef
15.
Zurück zum Zitat Patino M, Prochowski A, Agrawal MD et al (2016) Material separation using dual-energy CT: current and emerging applications. Radiographics 36:1087–1105CrossRef Patino M, Prochowski A, Agrawal MD et al (2016) Material separation using dual-energy CT: current and emerging applications. Radiographics 36:1087–1105CrossRef
16.
Zurück zum Zitat Wong WD, Shah S, Murray N, Walstra F, Khosa F, Nicolaou S (2018) Advanced musculoskeletal applications of dual-energy computed tomography. Radiol Clin North Am 56:587–600 Wong WD, Shah S, Murray N, Walstra F, Khosa F, Nicolaou S (2018) Advanced musculoskeletal applications of dual-energy computed tomography. Radiol Clin North Am 56:587–600
17.
Zurück zum Zitat Mallinson PI, Coupal TM, McLaughlin PD, Nicolaou S, Munk PL, Ouellette HA (2016) Dual-energy CT for the musculoskeletal system. Radiology 281:690–707 Mallinson PI, Coupal TM, McLaughlin PD, Nicolaou S, Munk PL, Ouellette HA (2016) Dual-energy CT for the musculoskeletal system. Radiology 281:690–707
18.
Zurück zum Zitat Wichmann JL, Booz C, Wesarg S et al (2014) Dual-energy CT-based phantomless in vivo three-dimensional bone mineral density assessment of the lumbar spine. Radiology 271:778–784CrossRef Wichmann JL, Booz C, Wesarg S et al (2014) Dual-energy CT-based phantomless in vivo three-dimensional bone mineral density assessment of the lumbar spine. Radiology 271:778–784CrossRef
19.
Zurück zum Zitat Roski F, Hammel J, Mei K et al (2019) Bone mineral density measurements derived from dual-layer spectral CT enable opportunistic screening for osteoporosis. Eur Radiol 29:6355–6363CrossRef Roski F, Hammel J, Mei K et al (2019) Bone mineral density measurements derived from dual-layer spectral CT enable opportunistic screening for osteoporosis. Eur Radiol 29:6355–6363CrossRef
20.
Zurück zum Zitat Van Hedent S, Su KH, Jordan DW et al (2019) Improving bone mineral density assessment using spectral detector CT. J Clin Densitom 22:374–381CrossRef Van Hedent S, Su KH, Jordan DW et al (2019) Improving bone mineral density assessment using spectral detector CT. J Clin Densitom 22:374–381CrossRef
21.
Zurück zum Zitat Bredella MA, Daley SM, Kalra MK, Brown JK, Miller KK, Torriani M (2015) Marrow adipose tissue quantification of the lumbar spine by using dual-energy CT and single-voxel 1H MR spectroscopy a feasibility study. Radiology 277:230–235 Bredella MA, Daley SM, Kalra MK, Brown JK, Miller KK, Torriani M (2015) Marrow adipose tissue quantification of the lumbar spine by using dual-energy CT and single-voxel 1H MR spectroscopy a feasibility study. Radiology 277:230–235
22.
Zurück zum Zitat Booz C, Hofmann PC, Sedlmair M et al (2017) Evaluation of bone mineral density of the lumbar spine using a novel phantomless dual-energy CT post-processing algorithm in comparison with dual-energy X-ray absorptiometry. Eur Radiol Exp 1:11CrossRef Booz C, Hofmann PC, Sedlmair M et al (2017) Evaluation of bone mineral density of the lumbar spine using a novel phantomless dual-energy CT post-processing algorithm in comparison with dual-energy X-ray absorptiometry. Eur Radiol Exp 1:11CrossRef
23.
Zurück zum Zitat Woisetschlager M, Spangeus A (2018) Model for improved correlation of BMD values between abdominal routine dual energy CT data and DXA scans. Eur J Radiol 99:76–81CrossRef Woisetschlager M, Spangeus A (2018) Model for improved correlation of BMD values between abdominal routine dual energy CT data and DXA scans. Eur J Radiol 99:76–81CrossRef
24.
Zurück zum Zitat Wichmann JL, Booz C, Wesarg S et al (2014) Quantitative dual-energy CT for phantomless evaluation of cancellous bone mineral density of the vertebral pedicle: correlation with pedicle screw pull-out strength. Eur Radiol 25:1714–1720CrossRef Wichmann JL, Booz C, Wesarg S et al (2014) Quantitative dual-energy CT for phantomless evaluation of cancellous bone mineral density of the vertebral pedicle: correlation with pedicle screw pull-out strength. Eur Radiol 25:1714–1720CrossRef
25.
Zurück zum Zitat Wait JM, Cody D, Jones AK, Rong J, Baladandayuthapani V, Kappadath SC (2015) Performance evaluation of material decomposition with rapid-kilovoltage-switching dual-energy CT and implications for assessing bone mineral density. AJR Am J Roentgenol 204:1234–1241 Wait JM, Cody D, Jones AK, Rong J, Baladandayuthapani V, Kappadath SC (2015) Performance evaluation of material decomposition with rapid-kilovoltage-switching dual-energy CT and implications for assessing bone mineral density. AJR Am J Roentgenol 204:1234–1241
Metadaten
Titel
Accuracy, agreement, and reliability of DECT-derived vBMD measurements: an initial ex vivo study
verfasst von
Le Qin
Jingjing Huang
Pei Yu
Jing Yan
Yingqian Ge
Yong Lu
Fuhua Yan
Lei Wang
Lianjun Du
Publikationsdatum
05.08.2020
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 1/2021
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-020-07118-8

Weitere Artikel der Ausgabe 1/2021

European Radiology 1/2021 Zur Ausgabe

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Medizinstudium Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Endlich: Zi zeigt, mit welchen PVS Praxen zufrieden sind

IT für Ärzte Nachrichten

Darauf haben viele Praxen gewartet: Das Zi hat eine Liste von Praxisverwaltungssystemen veröffentlicht, die von Nutzern positiv bewertet werden. Eine gute Grundlage für wechselwillige Ärzte und Psychotherapeuten.

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.