Abstract

A prospective cross-sectional study was conducted in a tertiary care center to determine the accuracy of transcutaneous bilirubin measurements (TcB) measured by the Konica Minolta JM-103™ meter compared to total serum bilirubin (TSB) in Asian infants aged 5–14 days. There were 405 late-preterm and term infants involved, and 455 paired samples were obtained by venepuncture and analyzed for bilirubin levels. TcB measurements were performed using the average of three measurements (TcB3) and a single measurement (TcB1) method. The overall correlation between TSB and the TcB was 0.80 (p ≤ 0.001) for TcB3 and 0.76 (p ≤ 0.001) for TcB1, respectively. The mean (SD) difference between TcB3 and TSB was −17.6 (29.5) μmol/L (median, −17.0; interquartile range (IQR), −39.1 to 1.7) and between TcB1 and TSB was −20.7 (32.3) μmol/L (median, −20.4; IQR, −42.5 to 1.7). The mean difference (SD) between the TcB3 and TSB in the low-risk (TSB < 170 μmol/L), intermediate-risk (TSB 170–254 μmol/L), and high-risk (TSB ≥ 255 μmol/L) groups was −2.8 (27.2), −13.4 (27.0), and −33.4 (29.1) μmol/L, respectively. To detect a TSB level of ≥255 μmol/L, using the TcB cutoff level of 204 μmol/L provides a sensitivity of 96 % with a specificity of 58 %. Conclusion: The TcB meter using a specific cutoff level can be reliably used as a screening tool for jaundice detection in older, postdischarge neonates, including the Asian population. Lower cutoff values can be set to capture all infants who merit closer surveillance, potential investigation, and treatment with higher accompanying screening costs.
Literatur
1.
Zurück zum Zitat American Academy of Pediatrics Clinical Practice Guideline (2004) Management of hyperbilirubinemia in the newborn infant 35 or more weeks of gestation. Pediatrics 114:297–316CrossRef American Academy of Pediatrics Clinical Practice Guideline (2004) Management of hyperbilirubinemia in the newborn infant 35 or more weeks of gestation. Pediatrics 114:297–316CrossRef
2.
Zurück zum Zitat Bhutani VK, Vilms RJ, Hamerman-Johnson L (2010) Universal bilirubin screening for severe neonatal hyperbilirubinemia. J Perinatol 30(Suppl):S6–S15PubMedCrossRef Bhutani VK, Vilms RJ, Hamerman-Johnson L (2010) Universal bilirubin screening for severe neonatal hyperbilirubinemia. J Perinatol 30(Suppl):S6–S15PubMedCrossRef
3.
Zurück zum Zitat Bosschaart N, Kok JH, Newsum AM et al (2012) Limitations and opportunities of transcutaneous bilirubin measurements. Pediatrics 129:689–694PubMedCrossRef Bosschaart N, Kok JH, Newsum AM et al (2012) Limitations and opportunities of transcutaneous bilirubin measurements. Pediatrics 129:689–694PubMedCrossRef
4.
Zurück zum Zitat Ebbesen F, Vandborg PK, Trydal T (2012) Comparison of the transcutaneous bilirubinometers BiliCheck and Minolta JM-103 in preterm neonates. Acta Paediatr 101:1128–1133PubMedCrossRef Ebbesen F, Vandborg PK, Trydal T (2012) Comparison of the transcutaneous bilirubinometers BiliCheck and Minolta JM-103 in preterm neonates. Acta Paediatr 101:1128–1133PubMedCrossRef
5.
Zurück zum Zitat Engle WD, Jackson GL, Stehel EK, Sendelbach DM, Manning MD (2005) Evaluation of a transcutaneous jaundice meter following hospital discharge in term and near-term neonates. J Perinatol 25:486–490PubMedCrossRef Engle WD, Jackson GL, Stehel EK, Sendelbach DM, Manning MD (2005) Evaluation of a transcutaneous jaundice meter following hospital discharge in term and near-term neonates. J Perinatol 25:486–490PubMedCrossRef
6.
Zurück zum Zitat Holland L, Blick K (2009) Implementing and validating transcutaneous bilirubinometry for neonates. Am J Clin Pathol 132:555–561PubMedCrossRef Holland L, Blick K (2009) Implementing and validating transcutaneous bilirubinometry for neonates. Am J Clin Pathol 132:555–561PubMedCrossRef
7.
Zurück zum Zitat Keren R, Tremont K, Luan X, Cnaan A (2009) Visual assessment of jaundice in term and late preterm infants. Arch Dis Child Fetal Neonatal Ed 94:F317–F322PubMedCrossRef Keren R, Tremont K, Luan X, Cnaan A (2009) Visual assessment of jaundice in term and late preterm infants. Arch Dis Child Fetal Neonatal Ed 94:F317–F322PubMedCrossRef
8.
Zurück zum Zitat Knudsen A, Kruse C, Ebbesen F (1993) Detection of hyperbilirubinemia by skin color measurements in icteric newborn infants at 5 to 14 days of age. Acta Paediatr 82:510–513PubMedCrossRef Knudsen A, Kruse C, Ebbesen F (1993) Detection of hyperbilirubinemia by skin color measurements in icteric newborn infants at 5 to 14 days of age. Acta Paediatr 82:510–513PubMedCrossRef
9.
Zurück zum Zitat Lam TS, Tsui KL, Kam CW (2008) Evaluation of a point-of-care transcutaneous bilirubinometer in Chinese neonates at an accident and emergency department. Hong Kong Med J 14:356–360PubMed Lam TS, Tsui KL, Kam CW (2008) Evaluation of a point-of-care transcutaneous bilirubinometer in Chinese neonates at an accident and emergency department. Hong Kong Med J 14:356–360PubMed
10.
Zurück zum Zitat Maisels MJ, Bhutani VK, Bogen D, Newman TB, Stark AR, Watchko JF (2009) Hyperbilirubinemia in the newborn infant ≥35 weeks’ gestation: an update with clarifications. Pediatrics 124:1193–1198PubMedCrossRef Maisels MJ, Bhutani VK, Bogen D, Newman TB, Stark AR, Watchko JF (2009) Hyperbilirubinemia in the newborn infant ≥35 weeks’ gestation: an update with clarifications. Pediatrics 124:1193–1198PubMedCrossRef
11.
Zurück zum Zitat Maisels MJ, Engle WD, Wainer S, Jackson GL, McManus S, Artinian F (2011) Transcutaneous bilirubin levels in an outpatient and office population. J Perinatol 31:621–624PubMedCrossRef Maisels MJ, Engle WD, Wainer S, Jackson GL, McManus S, Artinian F (2011) Transcutaneous bilirubin levels in an outpatient and office population. J Perinatol 31:621–624PubMedCrossRef
12.
Zurück zum Zitat Maisels MJ, Ostrea EM Jr, Touch S et al (2004) Evaluation of a new transcutaneous bilirubinometer. Pediatrics 113:1628–1635PubMedCrossRef Maisels MJ, Ostrea EM Jr, Touch S et al (2004) Evaluation of a new transcutaneous bilirubinometer. Pediatrics 113:1628–1635PubMedCrossRef
13.
Zurück zum Zitat Moyer VA, Ahn C, Sneed S (2000) Accuracy of clinical judgment in neonatal jaundice. Arch Pediatr Adolesc Med 154:391–394PubMed Moyer VA, Ahn C, Sneed S (2000) Accuracy of clinical judgment in neonatal jaundice. Arch Pediatr Adolesc Med 154:391–394PubMed
14.
Zurück zum Zitat Panburana J, Boonkasidach S, Rearkyai S (2010) Accuracy of transcutaneous bilirubinometry compare to total serum bilirubin measurement. J Med Assoc Thai 93(Suppl 2):S81–S86PubMed Panburana J, Boonkasidach S, Rearkyai S (2010) Accuracy of transcutaneous bilirubinometry compare to total serum bilirubin measurement. J Med Assoc Thai 93(Suppl 2):S81–S86PubMed
15.
Zurück zum Zitat Sanpavat S, Nuchprayoon I (2005) Comparison of two transcutaneous bilirubinometers—Minolta Airshields jaundice meter JM 103 and Spectrx Bilicheck—in Thai neonates. Southeast Asian J Tropl Med Public Health 36:1533–1537 Sanpavat S, Nuchprayoon I (2005) Comparison of two transcutaneous bilirubinometers—Minolta Airshields jaundice meter JM 103 and Spectrx Bilicheck—in Thai neonates. Southeast Asian J Tropl Med Public Health 36:1533–1537
16.
Zurück zum Zitat Sanpavat S, Nuchprayoon I (2004) Noninvasive transcutaneous bilirubin as a screening test to identify the need for serum bilirubin assessment. J Med Assoc Thai 87:1193–1198PubMed Sanpavat S, Nuchprayoon I (2004) Noninvasive transcutaneous bilirubin as a screening test to identify the need for serum bilirubin assessment. J Med Assoc Thai 87:1193–1198PubMed
17.
Zurück zum Zitat Sanpavat S, Nuchprayoon I, Smathakanee C, Hansuebsai R (2005) Nomogram for prediction of the risk of neonatal hyperbilirubinemia, using transcutaneous bilirubin. J Med Assoc Thai 88:1187–1193PubMed Sanpavat S, Nuchprayoon I, Smathakanee C, Hansuebsai R (2005) Nomogram for prediction of the risk of neonatal hyperbilirubinemia, using transcutaneous bilirubin. J Med Assoc Thai 88:1187–1193PubMed
18.
Zurück zum Zitat Yasuda S, Yasuda S, Itoh S, Isobe K et al (2003) New transcutaneous jaundice device with two optical paths. J Perinat Med 31:81–88PubMed Yasuda S, Yasuda S, Itoh S, Isobe K et al (2003) New transcutaneous jaundice device with two optical paths. J Perinat Med 31:81–88PubMed
19.
Zurück zum Zitat Yamauchi Y, Yamanouchi I (1991) Transcutaneous bilirubinometry: effect of postnatal age. Acta Paediatr Jpn 33:663–667PubMedCrossRef Yamauchi Y, Yamanouchi I (1991) Transcutaneous bilirubinometry: effect of postnatal age. Acta Paediatr Jpn 33:663–667PubMedCrossRef
20.
Zurück zum Zitat Yu ZB, Dong XY, Han SP et al (2011) Transcutaneous bilirubin nomogram for predicting neonatal hyperbilirubinemia in healthy term and late-preterm Chinese infants. Eur J Pediatr 170:185–191PubMedCrossRef Yu ZB, Dong XY, Han SP et al (2011) Transcutaneous bilirubin nomogram for predicting neonatal hyperbilirubinemia in healthy term and late-preterm Chinese infants. Eur J Pediatr 170:185–191PubMedCrossRef
Metadaten
Titel
Accuracy and precision of transcutaneous bilirubinometry in postdischarge Asian neonates
verfasst von
Ratchada Kitsommart
Pornpat Pornladnun
Chulathida Chomchai
Pornvilai Urujchutchairut
Bosco Paes
Publikationsdatum
01.06.2013
Verlag
Springer-Verlag
Erschienen in
European Journal of Pediatrics / Ausgabe 6/2013
Print ISSN: 0340-6199
Elektronische ISSN: 1432-1076
DOI
https://doi.org/10.1007/s00431-013-1960-4

Weitere Artikel der Ausgabe 6/2013

European Journal of Pediatrics 6/2013 Zur Ausgabe

Neuer Typ-1-Diabetes bei Kindern am Wochenende eher übersehen

23.04.2024 Typ-1-Diabetes Nachrichten

Wenn Kinder an Werktagen zum Arzt gehen, werden neu auftretender Typ-1-Diabetes und diabetische Ketoazidosen häufiger erkannt als bei Arztbesuchen an Wochenenden oder Feiertagen.

Neue Studienergebnisse zur Myopiekontrolle mit Atropin

22.04.2024 Fehlsichtigkeit Nachrichten

Augentropfen mit niedrig dosiertem Atropin können helfen, das Fortschreiten einer Kurzsichtigkeit bei Kindern zumindest zu verlangsamen, wie die Ergebnisse einer aktuellen Studie mit verschiedenen Dosierungen zeigen.

Spinale Muskelatrophie: Neugeborenen-Screening lohnt sich

18.04.2024 Spinale Muskelatrophien Nachrichten

Seit 2021 ist die Untersuchung auf spinale Muskelatrophie Teil des Neugeborenen-Screenings in Deutschland. Eine Studie liefert weitere Evidenz für den Nutzen der Maßnahme.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Pädiatrie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.