Accuracy of minus power intraocular lens calculation using OKULIX ray tracing software

verfasst von: Karim Mahmoud Nabil

Erschienen in: International Ophthalmology | Ausgabe 8/2019

Einloggen, um Zugang zu erhalten

Abstract

Purpose

The purpose of this retrospective study was to assess the accuracy of minus power intraocular lens calculation using partial coherence interferometry and OKULIX ray tracing software.

Methods

We included 25 consecutive, myopic eyes with axial length ≥ 30 mm (25 patients, 13 males and 12 females, and 57.6 ± 10.3 years old), which underwent phacoemulsification and implantation of a minus power intraocular lens in the capsular bag. Axial length measurement and corneal topography were performed using the OA-1000 optical biometer and Topographic Modeling System TMS-5, respectively. The IOL power was calculated using SRK/T formula and OKULIX ray tracing software. The implanted IOL power was chosen based on OKULIX ray tracing software calculation aiming for − 2 diopters (D) of myopia.

Results

SRK/T calculated IOL power (− 6.3 ± 2.8 D) showed statistically significant difference compared to OKULIX calculated IOL power (− 4.7 ± 2.6 D), rs 0.994 p < 0.001. The expected refraction with implanted IOL was − 1.7 ± 0.9 D based on OKULIX ray tracing software calculation. A statistically significant difference was reported between implanted IOL and OKULIX calculated IOL power (2.7 ± 1.4 D), rs 0.981 p < 0.001. A statistically significant difference was reported between the expected refraction with implanted IOL and the achieved spherical refraction at 1 month postoperatively (1.4 ± 0.7 D), rs 0.77 p < 0.001. The achieved spherical refraction at 1 month postoperatively was 0.2 ± 0.2 D.

Conclusions

Although OKULIX ray tracing software yielded more accurate minus power intraocular lens calculation in extreme myopia, compared to SRK/T formula, yet it still shows tendency toward hyperopia.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Zhang Y, Liang XY, Liu S, Lee JW, Bhaskar S, Lam DS (2016) Accuracy of intraocular lens power calculation formulas for highly myopic eyes. J Ophthalmol 2016:1917268PubMedPubMedCentral Zhang Y, Liang XY, Liu S, Lee JW, Bhaskar S, Lam DS (2016) Accuracy of intraocular lens power calculation formulas for highly myopic eyes. J Ophthalmol 2016:1917268PubMedPubMedCentral
2.
Zurück zum Zitat Abulafia A, Barrett GD, Rotenberg M (2015) Intraocular lens power calculation for eyes with an axial length greater than 26.0 mm: comparison of formulas and methods. J Cataract Refract Surg 41:548–556CrossRefPubMed Abulafia A, Barrett GD, Rotenberg M (2015) Intraocular lens power calculation for eyes with an axial length greater than 26.0 mm: comparison of formulas and methods. J Cataract Refract Surg 41:548–556CrossRefPubMed
3.
Zurück zum Zitat Roessler GF, Dietlein TS, Plange N, Roepke AK, Dinslage S, Walter P, Mazinani BA (2012) Accuracy of intraocular lens power calculation using partial coherence interferometry in patients with high myopia. Ophthalmic Physiol Opt 32:228–233CrossRefPubMed Roessler GF, Dietlein TS, Plange N, Roepke AK, Dinslage S, Walter P, Mazinani BA (2012) Accuracy of intraocular lens power calculation using partial coherence interferometry in patients with high myopia. Ophthalmic Physiol Opt 32:228–233CrossRefPubMed
4.
5.
Zurück zum Zitat Garg A, Hoyos JE, Dementiev J (2005) Update on IOL power calculations formulas, in Mastering techniques of IOL power calculations. Jaypee Brothers, New Delhi, pp 16–22 Garg A, Hoyos JE, Dementiev J (2005) Update on IOL power calculations formulas, in Mastering techniques of IOL power calculations. Jaypee Brothers, New Delhi, pp 16–22
6.
Zurück zum Zitat Sanders DR, Retzlaff JA, Kraff MC, Gimbel HV, Raanan MG (1990) Comparison of the SRK/T formula and other theoretical and regression formulas. J Cataract Refract Surg 16:341–346CrossRefPubMed Sanders DR, Retzlaff JA, Kraff MC, Gimbel HV, Raanan MG (1990) Comparison of the SRK/T formula and other theoretical and regression formulas. J Cataract Refract Surg 16:341–346CrossRefPubMed
7.
Zurück zum Zitat Kohnen S, Brauweiler P (1996) First results of cataract surgery and implantation of negative power intraocular lenses in highly myopic eyes. J Cataract Refract Surg 22:416–420CrossRefPubMed Kohnen S, Brauweiler P (1996) First results of cataract surgery and implantation of negative power intraocular lenses in highly myopic eyes. J Cataract Refract Surg 22:416–420CrossRefPubMed
8.
Zurück zum Zitat MacLaren RE, Sagoo MS, Restori M, Allan BDS (2005) Biometry accuracy using zero- and negative-powered intraocular lenses. J Cataract Refract Surg 31:280–290CrossRefPubMed MacLaren RE, Sagoo MS, Restori M, Allan BDS (2005) Biometry accuracy using zero- and negative-powered intraocular lenses. J Cataract Refract Surg 31:280–290CrossRefPubMed
9.
Zurück zum Zitat Wang JK, Hu CY, Chang SW (2008) Intraocular lens power calculation using the IOLMaster and various formulas in eyes with long axial length. J Cataract Refract Surg 34:262–267CrossRefPubMed Wang JK, Hu CY, Chang SW (2008) Intraocular lens power calculation using the IOLMaster and various formulas in eyes with long axial length. J Cataract Refract Surg 34:262–267CrossRefPubMed
10.
Zurück zum Zitat Aristodemou P, Knox Cartwright NE, Sparrow JM, Johnston RL (2011) Formula choice: Hoffer Q, Holladay 1, or SRK/T and refractive outcomes in 8108 eyes after cataract surgery with biometry by partial coherence interferometry. J Cataract Refract Surg 37:63–71CrossRefPubMed Aristodemou P, Knox Cartwright NE, Sparrow JM, Johnston RL (2011) Formula choice: Hoffer Q, Holladay 1, or SRK/T and refractive outcomes in 8108 eyes after cataract surgery with biometry by partial coherence interferometry. J Cataract Refract Surg 37:63–71CrossRefPubMed
11.
Zurück zum Zitat Rabsilber T, Reuland A, Holzer M, Auffarth G (2007) Intraocular lens power calculation using ray tracing following excimer laser surgery. Eye 21:697–701CrossRefPubMed Rabsilber T, Reuland A, Holzer M, Auffarth G (2007) Intraocular lens power calculation using ray tracing following excimer laser surgery. Eye 21:697–701CrossRefPubMed
12.
Zurück zum Zitat Hoffmann PC, Wahl J, Preussner PR (2012) Accuracy of intraocular lens calculation with ray tracing. J Refract Surg 28:650–655CrossRefPubMed Hoffmann PC, Wahl J, Preussner PR (2012) Accuracy of intraocular lens calculation with ray tracing. J Refract Surg 28:650–655CrossRefPubMed
13.
Zurück zum Zitat Olsen T, Funding M (2012) Ray-tracing analysis of intraocular lens power in situ. J Cataract Refract Surg 38:641–647CrossRefPubMed Olsen T, Funding M (2012) Ray-tracing analysis of intraocular lens power in situ. J Cataract Refract Surg 38:641–647CrossRefPubMed
14.
Zurück zum Zitat Hoffmann PC, Lindemann CR (2013) Intraocular lens calculation for aspheric intraocular lenses. J Cataract Refract Surg 39:867–872CrossRefPubMed Hoffmann PC, Lindemann CR (2013) Intraocular lens calculation for aspheric intraocular lenses. J Cataract Refract Surg 39:867–872CrossRefPubMed
15.
Zurück zum Zitat Minami K, Kataoka Y, Matsunaga J, Ohtani S, Honbou M, Miyata K (2012) Ray-tracing intraocular lens power calculation using anterior segment optical coherence tomography measurements. J Cataract Refract Surg 38:1758–1763CrossRefPubMed Minami K, Kataoka Y, Matsunaga J, Ohtani S, Honbou M, Miyata K (2012) Ray-tracing intraocular lens power calculation using anterior segment optical coherence tomography measurements. J Cataract Refract Surg 38:1758–1763CrossRefPubMed
16.
Zurück zum Zitat Saiki M, Negishi K, Kato N, Torii H, Dogru M, Tsubota K (2014) Ray tracing software for intraocular lens power calculation after corneal excimer laser surgery. Jpn J Ophthalmol 58:276–281CrossRefPubMed Saiki M, Negishi K, Kato N, Torii H, Dogru M, Tsubota K (2014) Ray tracing software for intraocular lens power calculation after corneal excimer laser surgery. Jpn J Ophthalmol 58:276–281CrossRefPubMed
17.
Zurück zum Zitat Holladay JT, Prager TC (1989) Accurate ultrasonic biometry in pseudophakia. Am J Ophthalmol 107:189–190CrossRefPubMed Holladay JT, Prager TC (1989) Accurate ultrasonic biometry in pseudophakia. Am J Ophthalmol 107:189–190CrossRefPubMed
19.
Zurück zum Zitat Preussner PR (2012) Ray tracing for IOL power calculations. Cataract Refract Surg Today Eur 5:46–48 Preussner PR (2012) Ray tracing for IOL power calculations. Cataract Refract Surg Today Eur 5:46–48
20.
Zurück zum Zitat Wang L, Shirayama M, Ma XJ, Kohnen T, Koch DD (2011) Optimizing intraocular lens power calculations in eyes with axial lengths above 25.0 mm. J Cataract Refract Surg 37:2018–2027CrossRefPubMed Wang L, Shirayama M, Ma XJ, Kohnen T, Koch DD (2011) Optimizing intraocular lens power calculations in eyes with axial lengths above 25.0 mm. J Cataract Refract Surg 37:2018–2027CrossRefPubMed
21.
Zurück zum Zitat Haigis W (2009) Intraocular lens calculation in extreme myopia. J Cataract Refract Surg 35:906–911CrossRefPubMed Haigis W (2009) Intraocular lens calculation in extreme myopia. J Cataract Refract Surg 35:906–911CrossRefPubMed
22.
Zurück zum Zitat Petermeier K, Gekeler F, Messias A, Spitzer MS, Haigis W, Szurman P (2009) Intraocular lens power calculation and optimized constants for highly myopic eyes. J Cataract Refract Surg 35:1575–1581CrossRefPubMed Petermeier K, Gekeler F, Messias A, Spitzer MS, Haigis W, Szurman P (2009) Intraocular lens power calculation and optimized constants for highly myopic eyes. J Cataract Refract Surg 35:1575–1581CrossRefPubMed
Metadaten
Titel
Accuracy of minus power intraocular lens calculation using OKULIX ray tracing software
verfasst von
Karim Mahmoud Nabil
Publikationsdatum
23.08.2018
Verlag
Springer Netherlands
Erschienen in
International Ophthalmology / Ausgabe 8/2019
Print ISSN: 0165-5701
Elektronische ISSN: 1573-2630
DOI
https://doi.org/10.1007/s10792-018-1007-y

Weitere Artikel der Ausgabe 8/2019

International Ophthalmology 8/2019 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Ophthalmika in der Schwangerschaft

Die Verwendung von Ophthalmika in der Schwangerschaft und Stillzeit stellt immer eine Off-label-Anwendung dar. Ein Einsatz von Arzneimitteln muss daher besonders sorgfältig auf sein Risiko-Nutzen-Verhältnis bewertet werden. In der vorliegenden …

Operative Therapie und Keimnachweis bei endogener Endophthalmitis

Vitrektomie Originalie

Die endogene Endophthalmitis ist eine hämatogen fortgeleitete, bakterielle oder fungale Infektion, die über choroidale oder retinale Gefäße in den Augapfel eingeschwemmt wird [ 1 – 3 ]. Von dort infiltrieren die Keime in die Netzhaut, den …

Bakterielle endogene Endophthalmitis

Vitrektomie Leitthema

Eine endogene Endophthalmitis stellt einen ophthalmologischen Notfall dar, der umgehender Diagnostik und Therapie bedarf. Es sollte mit geeigneten Methoden, wie beispielsweise dem Freiburger Endophthalmitis-Set, ein Keimnachweis erfolgen. Bei der …

So erreichen Sie eine bestmögliche Wundheilung der Kornea

Die bestmögliche Wundheilung der Kornea, insbesondere ohne die Ausbildung von lichtstreuenden Narben, ist oberstes Gebot, um einer dauerhaften Schädigung der Hornhaut frühzeitig entgegenzuwirken und die Funktion des Auges zu erhalten.   

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.