Accurate quantitative myocardial perfusion using single cycle T1 mapping

verfasst von: D Chen, B Sharif, R Dharmakumar, D Li

Erschienen in: Journal of Cardiovascular Magnetic Resonance | Sonderheft 1/2013

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Background

Quantitative myocardial perfusion MRI depends on accurate determination of the arterial input function (AIF). Unfortunately, typical doses of contrast agent (CA) cause saturation in the AIF signal intensity at peak CA concentrations. Methods using dual boluses [1] or dual sequences [2] have been proposed to correct for this effect; however, they require additional data acquisition which complicates the imaging procedure. CA concentrations can be directly found by calculating blood and myocardial T1 values [3, 4]. This method was performed in vivo and compared to the dual sequence method [2].

Methods

Radial AIF correction was performed using the method described in [4]. T1 relaxation following an SR magnetization preparation and FLASH acquisition can be modeled [3]. By acquiring multiple images after a single SR and using the acquisition parameters, pixelwise T1 can be found by solving a simple non-linear data fitting exercise. Gd concentration is found using the known relaxivity of Gd CA [6].
Ten healthy volunteers underwent perfusion MRI studies on a Siemens 3T Verio system with IRB approval and written consent. Two first pass perfusion scans were performed at rest using a radial acquisition for T1 mapping and a dual TI Cartesian acquisition. The Cartesian scan had a resolution of 3×2×8 mm3 and TI = 120 ms. The short TI scan has TI = 28 ms. The multi-shot radial scan had a resolution of 1.7×1.7×8 mm3; 3 images per SR with TI = 50, 94, and 138 ms respectively. Reconstruction of radial data was performed using a compressed sensing method with a composite reference constraint [3].

Results

Results: Raw Cartesian and radial perfusion signal intensity curves showed higher than expected rest MBF. Using the proposed method, the Gd concentration of the AIF was found. Fig. 1 shows signal curves comparing uncorrected radial and corrected. Using the proposed T1 mapping technique allows for correction of the radial AIF (‘SI Curve’). The MBF using the corrected AIF compared favorably with MBFs found using the AIF from the dual sequence, as presented in Table 1.
Table 1
Comparisons of MBF (ml/min/g) measurements using Cartesian and radial techniques. There is no significant difference between radially acquired and Cartesian MBFs.
 
Cartesian
Radial
P-Value
Uncorrected
1.96 ± 0.45
1.88 ± 0.41
P<0.33
Corrected
1.34 ± 0.41
1.35 ± 0.27
P<0.23

Conclusions

Myocardial perfusion measurement using radial T1-mapping eliminates the need to acquire two datasets to produce an unsaturated AIF with the dual sequence approach, making it more practical for clinical use. The major problem with the dual sequence method is it produces AIF of different signal intensity scale due to different T1 weighting. The AIF needs to be scaled properly with respect to myocardial tissue function for accurate MBF measurements. The proposed method allows the myocardial tissue signal and AIF to be extracted using the same set of the images, thus eliminating a potential source of error.

Funding

NIH Grant T32 EB51705
NIH Grant RO1 EB002623
AHA Postdoctoral Fellowship Award 1POST7390063
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://​creativecommons.​org/​licenses/​by/​2.​0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Literatur
1.
3.
Zurück zum Zitat DiBella Kholmovski: MRM. 2007, 821-7. DiBella Kholmovski: MRM. 2007, 821-7.
6.
Zurück zum Zitat Chien Hanicke: Med Phys. 1990, 1004-10. Chien Hanicke: Med Phys. 1990, 1004-10.
Metadaten
Titel
Accurate quantitative myocardial perfusion using single cycle T1 mapping
verfasst von
D Chen
B Sharif
R Dharmakumar
D Li
Publikationsdatum
01.01.2013
Verlag
BioMed Central
DOI
https://doi.org/10.1186/1532-429X-15-S1-W5

Weitere Artikel der Sonderheft 1/2013

Journal of Cardiovascular Magnetic Resonance 1/2013 Zur Ausgabe

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.