Skip to main content
Erschienen in: Journal of Cardiothoracic Surgery 1/2019

Open Access 01.12.2019 | Case report

An alternative solution for patient with high risk of coronary obstruction underwent TAVI procedure using a novel second-generation device – a case series

verfasst von: Hong Qian, Yucheng Chen, Zhaoyun Cheng, Yingqiang Guo

Erschienen in: Journal of Cardiothoracic Surgery | Ausgabe 1/2019

Abstract

Background

Obstruction of the left or right coronary artery is a rare but lethal complication during transcatheter aortic valve implantation (TAVI). The new J-Valve™ prosthesis is a new second generation TAVI device which has several features to avoid the coronary obstruction such as low profile design and clip fixation of the native leaflets. The aim of this study is to report our initial experience of using this valve in treating patient with high risk factors for coronary obstruction during TAVI procedure.

Case presentation

Three high surgical risk patients (All females with 77, 76, and 75 years old) with symptomatic aortic stenosis were enrolled. All patients have the common feature of low coronary ostium height (< 10 mm) with narrowed aortic sinus (< 30 mm) on CT angiogram and marked leaflet calcification. Three 25 mm J-Valve prostheses were successfully implanted through trans-apical approach. No coronary obstruction was noted for these patients. Effective aortic open area was significantly increased after valve implantation (Preoperative 0.7, 0.7 and 0.65 cm2 – Postoperative 1.8, 1.9 and 2.0 cm2). Only one patient was noted to have trivial degree paravalvular leakage.

Conclusion

The new J-Valve prosthesis is a new second generation TAVI device. This system may provide another safety treatment option for patient with high risk factor for coronary obstruction underwent TAVI procedure.
Abkürzungen
EOA
Effective aortic valve opening area
TAVI
Transcatheter aortic valve implantation

Background

Transcatheter aortic valve implantation (TAVI) has evolved as a routine procedure to treat selected high-risk patients with severe aortic stenosis. Although obstruction of the left or right coronary ostium is a relatively rare but lethal complication during TAVI procedure, it is generally accepted that a distance between the coronary ostium and the annulus less than 10 mm or sinus of Valsalva diameter below 30 mm, and existing of bulky calcified leaflet are risks for coronary occlusion as for several type of classic TAVI devices such as Edward SAPIEN and Medtronic CoreValve [1, 2]. The J-Valve™ system is a brand new second-generation TAVI device featured by three U-shape anatomically oriented devices “claspers” which could facilitate intuitive “self-positioning” valve implantation [3]. Due to the ability of embracing the native calcified leaflets during valve implantation and specially designed bear area on valve stent, this low-profile designed valve is especially suitable for patient who is at high risk for coronary obstruction. We then reported our initial experience of using J-Valve in treating patient with high risk factor for coronary obstruction during TAVI procedure.

Case presentation

Three high surgical risk patients with symptomatic aortic stenosis who are at high risk for coronary occlusion during TAVI procedure were enrolled (three female patients). The basic characters shown in Table 1. These three patients have the common feature of low coronary ostium height (< 10 mm) with narrowed aortic sinus (< 30 mm). The effective aortic valve opening area (EOA) was 0.7 cm2, 0.7 cm2 and 0.65 cm2 in each patient with marked leaflet calcification. Aortic valve annular size was 24 mm, 25 mm, 24 mm on CT image.
Table 1
Basic character of three patients
Patient
Log-Euroscore I(%)
Aortic Valve
Calcification
LCH
RCH
Aortic Sinus
Annular size
Valve Size
No1 77y Female
21%
Severe AS
Moderate
7.8 mm
7.8 mm
27 mm
24 mm
25 mm
No2 76y Female
20%
Severe AS
Severe
8.9 mm
9.5 mm
29 mm
25 mm
25 mm
No3 75y Female
22%
Severe AS
Bulky
8.2 mm
7.9 mm
27 mm
24 mm
25 mm
LCH Left coronary ostium height, RCH Right coronary ostium height
The TAVI procedure using J-Valve™ prosthesis was performed through trans-apical approach (Fig. 1). The procedure detail is described previously [3]. Briefly the apical puncture was done with a super-stiff guide-wire placed into the descending aorta. A 14F sheath was carefully introduced into the left ventricle followed by a balloon catheter. Pre-procedural balloon-valvuloplasty of the native valve was performed under rapid pacing. In parallel, a J-Valve™ prosthesis was crimped into the 27-F Ausper-AS delivery system (Three 25 mm prostheses were used). The delivery system was bluntly inserted into the left ventricle and advanced under fluoroscopic guidance into a supra-annular position. Three ‘U-shaped’ graspers were then completely released first and carefully placed into the corresponding aortic sinus by gently pulling back the delivery system, thereby embracing the native leaflets. The angiogram was then performed to confirm that all the graspers were positioned correctly into the each aortic sinus. The valve was retrieved back gently into the annular plan with the guidance of the graspers and deployed without rapid ventricular pacing. After wound closure, the patient was transferred to the intensive care unit. Further postoperative course was uneventful. At discharge, TTE revealed an excellent valve function with EOA (1.8 cm2, 1.9 cm2, 2.0 cm2 in each patient) without any relevant incompetence. One patient was noted to have trivial degree paravalvular leakage.

Discussion

Although the incidence of coronary obstruction reveled in large TAVI series and registries are relatively low, it is still the most lethal complication of TAVI procedure [1]. Evidences have shown that a distance between the coronary ostium and the annulus less than 10 mm or sinus of Valsalva diameter below 30 mm, and existing of bulky calcified leaflet are risks for coronary occlusion as for classic TAVI device such as Edward SAPIEN and Medtronic CoreValve due to their design [2]. Contrast to western population, Asian patients always associated with low body weight, relatively small aortic root and low coronary orifice height. In some specific situation, choose of the classic TAVI device may warrant caution or even contraindicated.
The J-Valve™ system is a new second-generation self-expendable low-profile prosthesis that is featured by three U-shape anatomical orientated devices — “Clasper” encircling around the valve frame. The design concept of this device is to facilitate intuitive ‘self-positioning’ valve implantation through embracing the native aortic valve leaflet. In this case serious, successful TAVI procedure was done using this unique system in all three patients who are at high risk for coronary obstruction with low coronary height and narrowed aortic sinus. This valve have several unique features to minimized the risk coronary obstruction including 1) More accurate position of the device with anatomical orientated devices to avoid malposition; 2) Low-profile design with extra bare mental area in coronary orifice region (Fig. 2); 3) Unlike the traditional design of classic stent valve system, the unique designed U-shape claspers could embrace the native aortic valve leaflet (clip mechanism) then subsequently leaving the coronary ostia freely accessible from heavily calcified aortic leaflet; 4) Meanwhile, J-Valve™ system has the advantage of achieving better anchoring through embrace the leaflets, then diminish the necessity of oversizing and minimize the risk of coronary obstruction.

Conclusion

The new J-Valve™ prosthesis is a brand new second generation TAVI device. This system may provide a safety treatment option for patient with high risk factor for coronary obstruction underwent TAVI procedure.

Acknowledgements

Not applicable.

Funding

None.

Availability of data and materials

The dataset of this case report is available from the corresponding author on reasonable request.
The hospital ethics committee approval was granted of this case report.
Written informed consent was obtained from the patient for publication of this case report and any accompanying images. A copy of the written consent is available for review by the Editor-in-Chief of this journal.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
1.
Zurück zum Zitat Ribeiro HB, Webb JG, Makkar RR, et al. Predictive factors, management, and clinical outcomes of coronary obstruction following transcatheter aortic valve implantation: insights from a large multicenter registry. JACC Cardiovasc Interv. 2013;6:452–61.CrossRef Ribeiro HB, Webb JG, Makkar RR, et al. Predictive factors, management, and clinical outcomes of coronary obstruction following transcatheter aortic valve implantation: insights from a large multicenter registry. JACC Cardiovasc Interv. 2013;6:452–61.CrossRef
2.
Zurück zum Zitat Ribeiro HB, Nombela-Franco L, Urena M, et al. Coronary obstruction following transcatheter aortic valve implantation: a systematic review. JACC Cardiovasc Interv. 2013;6:452–61.CrossRef Ribeiro HB, Nombela-Franco L, Urena M, et al. Coronary obstruction following transcatheter aortic valve implantation: a systematic review. JACC Cardiovasc Interv. 2013;6:452–61.CrossRef
3.
Zurück zum Zitat Zhu D, Wei L, Cheung A, et al. Treatment of Pure Aortic Regurgitation Using a Second-Generation Transcatheter Aortic Valve Implantation System. J Am Coll Cardiol. 2016;67:2803–5.CrossRef Zhu D, Wei L, Cheung A, et al. Treatment of Pure Aortic Regurgitation Using a Second-Generation Transcatheter Aortic Valve Implantation System. J Am Coll Cardiol. 2016;67:2803–5.CrossRef
Metadaten
Titel
An alternative solution for patient with high risk of coronary obstruction underwent TAVI procedure using a novel second-generation device – a case series
verfasst von
Hong Qian
Yucheng Chen
Zhaoyun Cheng
Yingqiang Guo
Publikationsdatum
01.12.2019
Verlag
BioMed Central
Erschienen in
Journal of Cardiothoracic Surgery / Ausgabe 1/2019
Elektronische ISSN: 1749-8090
DOI
https://doi.org/10.1186/s13019-019-0859-1

Weitere Artikel der Ausgabe 1/2019

Journal of Cardiothoracic Surgery 1/2019 Zur Ausgabe

Echinokokkose medikamentös behandeln oder operieren?

06.05.2024 DCK 2024 Kongressbericht

Die Therapie von Echinokokkosen sollte immer in spezialisierten Zentren erfolgen. Eine symptomlose Echinokokkose kann – egal ob von Hunde- oder Fuchsbandwurm ausgelöst – konservativ erfolgen. Wenn eine Op. nötig ist, kann es sinnvoll sein, vorher Zysten zu leeren und zu desinfizieren. 

Wie sieht der OP der Zukunft aus?

04.05.2024 DCK 2024 Kongressbericht

Der OP in der Zukunft wird mit weniger Personal auskommen – nicht, weil die Technik das medizinische Fachpersonal verdrängt, sondern weil der Personalmangel es nötig macht.

Umsetzung der POMGAT-Leitlinie läuft

03.05.2024 DCK 2024 Kongressbericht

Seit November 2023 gibt es evidenzbasierte Empfehlungen zum perioperativen Management bei gastrointestinalen Tumoren (POMGAT) auf S3-Niveau. Vieles wird schon entsprechend der Empfehlungen durchgeführt. Wo es im Alltag noch hapert, zeigt eine Umfrage in einem Klinikverbund.

Recycling im OP – möglich, aber teuer

02.05.2024 DCK 2024 Kongressbericht

Auch wenn sich Krankenhäuser nachhaltig und grün geben – sie tragen aktuell erheblich zu den CO2-Emissionen bei und produzieren jede Menge Müll. Ein Pilotprojekt aus Bonn zeigt, dass viele Op.-Abfälle wiederverwertet werden können.

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.