Impedance mapping with constant contact force on 3D electroanatomic map to characterize tissues at pulmonary veno-atrial junction

verfasst von: Corrado Tomasi, Alessandro Dal Monte, Maria Selina Argnani, Cristiana Corsi, Federica Giannotti, Stefano Severi, Andrea Rubboli

Erschienen in: Journal of Interventional Cardiac Electrophysiology | Ausgabe 3/2021

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Generator impedance (Im) mapping with constant contact force (CF) by tip catheter at PV isolation (PVI) was assessed for a proposal of tissue characterization at PV-LA junction (PV-LAJ).

Methods

In this observational, prospective, single-center study, Im mapping at constant CF = 10 g (± 2 g) was performed before PVI at PV-LAJ. PV in-vein, PV ostium (PVos), and antrum (PVan) contours were manually traced based on the 3D electroanatomic map (3DEAM) integrating intracardiac echocardiography and computerized tomography. PVan contour-methods based on Im mapping was defined on 3DEAM as the atrial-like Im contour closest to PVos, and its distance from anatomical PVan contour > 5 mm was assumed as the non-concordance marker between contour and methods.

Results

Sixty-two patients (62 ± 9 years; 43 males) were enrolled, and 244 PV-LAJ were assessed. From in-vein PV to LA and, less prominently, from PVos to PVan and LA, Im showed a unidirectional decrease with highly variable individual-specific distribution and values. PVan non-concordance was found in 59/665 segments (8.8%), 18% of PV-LAJs, and 53% of pts; it prevailed in superior PV-LAJ and measured on average 7.2 ± 1.1 mm. Im decrease patterns and non-concordance were not associated with any clinical or anatomical feature, including PV dimensions and shape.

Conclusions

Im mapping of LA-PVJ at constant CF added to 3DEAM may consistently track the tissue transition from PV to LA. PVan identified by Im was often located more toward LA than the 3D anatomical PVan, particularly in LSPV, suggesting the potential advantage of avoiding ablation of venous-like tissue. Im mapping can deserve further investigation for target characterization at LA-PVJ.
Literatur
1.
Zurück zum Zitat Calkins H, Hindricks G, Cappato R, Kim YH, Saad EB, Aguinaga L, et al. 2017 HRS/EHRA/ECAS/APHRS/SOLAECE expert consensus statement on catheter and surgical ablation of atrial fibrillation. Europace. 2018;20:157–208.CrossRef Calkins H, Hindricks G, Cappato R, Kim YH, Saad EB, Aguinaga L, et al. 2017 HRS/EHRA/ECAS/APHRS/SOLAECE expert consensus statement on catheter and surgical ablation of atrial fibrillation. Europace. 2018;20:157–208.CrossRef
2.
Zurück zum Zitat Edriss H, Denega T, Test V, Nugent K. Pulmonary vein stenosis complicating radiofrequency catheter ablation for atrial fibrillation: a literature review. Respir Med. 2016;117:215–22.CrossRef Edriss H, Denega T, Test V, Nugent K. Pulmonary vein stenosis complicating radiofrequency catheter ablation for atrial fibrillation: a literature review. Respir Med. 2016;117:215–22.CrossRef
3.
Zurück zum Zitat Teunissen C, Velthuis BK, Hassink RJ, van der Heijden JF, Vonken EPA, Clappers N, et al. Incidence of pulmonary vein stenosis after radiofrequency catheter ablation of atrial fibrillation. JACC Clin Electrophysiol. 2017;3:589–98.CrossRef Teunissen C, Velthuis BK, Hassink RJ, van der Heijden JF, Vonken EPA, Clappers N, et al. Incidence of pulmonary vein stenosis after radiofrequency catheter ablation of atrial fibrillation. JACC Clin Electrophysiol. 2017;3:589–98.CrossRef
4.
Zurück zum Zitat Lang CC, Gugliotta F, Santinelli V, Mesas C, Tomita T, Vicedomini G, et al. Endocardial impedance mapping during circumferential pulmonary vein ablation of atrial fibrillation differentiates between atrial and venous tissue. Heart Rhythm. 2006;3:171–8.CrossRef Lang CC, Gugliotta F, Santinelli V, Mesas C, Tomita T, Vicedomini G, et al. Endocardial impedance mapping during circumferential pulmonary vein ablation of atrial fibrillation differentiates between atrial and venous tissue. Heart Rhythm. 2006;3:171–8.CrossRef
5.
Zurück zum Zitat Pedrote A, Arana-Rueda E, García-Riesco L, Jiménez-Velasco A, Sánchez-Brotons J, Arizón-Muñoz JM, et al. Three-dimensional impedance mapping as an aid to circumferential pulmonary vein isolation in paroxysmal atrial fibrillation. Rev Esp Cardiol. 2009;62:315–9.CrossRef Pedrote A, Arana-Rueda E, García-Riesco L, Jiménez-Velasco A, Sánchez-Brotons J, Arizón-Muñoz JM, et al. Three-dimensional impedance mapping as an aid to circumferential pulmonary vein isolation in paroxysmal atrial fibrillation. Rev Esp Cardiol. 2009;62:315–9.CrossRef
6.
Zurück zum Zitat Makimoto H, Lin T, Rillig A, Metzner A, Wohlmuth P, Arya A, et al. In vivo contact force analysis and correlation with tissue impedance during left atrial mapping and catheter ablation of atrial fibrillation. Circ Arrhythm Electrophysiol. 2014;7:46–54.CrossRef Makimoto H, Lin T, Rillig A, Metzner A, Wohlmuth P, Arya A, et al. In vivo contact force analysis and correlation with tissue impedance during left atrial mapping and catheter ablation of atrial fibrillation. Circ Arrhythm Electrophysiol. 2014;7:46–54.CrossRef
7.
Zurück zum Zitat Kumar S, Morton JB, Lee J, Halloran K, Spence SJ, Gorelik A, et al. Prospective characterization of catheter-tissue contact force at different anatomic sites during antral pulmonary vein isolation. Circ Arrhythm Electrophysiol. 2012;5:1124–9.CrossRef Kumar S, Morton JB, Lee J, Halloran K, Spence SJ, Gorelik A, et al. Prospective characterization of catheter-tissue contact force at different anatomic sites during antral pulmonary vein isolation. Circ Arrhythm Electrophysiol. 2012;5:1124–9.CrossRef
8.
Zurück zum Zitat Nakagawa H, Kautzner J, Natale A, Peichl P, Cihak R, Wichterle D, et al. Locations of high contact force during left atrial mapping in atrial fibrillation patients: electrogram amplitude and impedance are poor predictors of electrode-tissue contact force for ablation of atrial fibrillation. Circ Arrhythm Electrophysiol. 2013;6:746–53.CrossRef Nakagawa H, Kautzner J, Natale A, Peichl P, Cihak R, Wichterle D, et al. Locations of high contact force during left atrial mapping in atrial fibrillation patients: electrogram amplitude and impedance are poor predictors of electrode-tissue contact force for ablation of atrial fibrillation. Circ Arrhythm Electrophysiol. 2013;6:746–53.CrossRef
9.
Zurück zum Zitat Lacomis JM, Nigginton W, Fuhrman, Schwartzman D, Armfield DR, Pealer KM. Multi-detector row CT of the left atrium and pulmonary veins before radiofrequency catheter ablation for atrial fibrillation. Radiographics. 2003:535–48. Lacomis JM, Nigginton W, Fuhrman, Schwartzman D, Armfield DR, Pealer KM. Multi-detector row CT of the left atrium and pulmonary veins before radiofrequency catheter ablation for atrial fibrillation. Radiographics. 2003:535–48.
10.
Zurück zum Zitat Marom EM, Herndon JE, Kim YH, McAdams HP. Variations in pulmonary venous drainage to the left atrium: implications for radiofrequency ablation. Radiology. 2004;230:824–9.CrossRef Marom EM, Herndon JE, Kim YH, McAdams HP. Variations in pulmonary venous drainage to the left atrium: implications for radiofrequency ablation. Radiology. 2004;230:824–9.CrossRef
11.
Zurück zum Zitat Arentz T, Weber R, Bürkle G, Herrera C, Blum T, Stockinger J, et al. Small or large isolation areas around the pulmonary veins for the treatment of atrial fibrillation? Results from a prospective randomized study. Circulation. 2007;115:3057–63.CrossRef Arentz T, Weber R, Bürkle G, Herrera C, Blum T, Stockinger J, et al. Small or large isolation areas around the pulmonary veins for the treatment of atrial fibrillation? Results from a prospective randomized study. Circulation. 2007;115:3057–63.CrossRef
12.
Zurück zum Zitat Proietti R, Santangeli P, Di Biase L, Joza J, Bernier ML, Wang Y, et al. Comparative effectiveness of wide antral versus ostial pulmonary vein isolation. A systematic review and meta-analysis. Circ Arrhythm Electrophysiol. 2014;7:39–45.CrossRef Proietti R, Santangeli P, Di Biase L, Joza J, Bernier ML, Wang Y, et al. Comparative effectiveness of wide antral versus ostial pulmonary vein isolation. A systematic review and meta-analysis. Circ Arrhythm Electrophysiol. 2014;7:39–45.CrossRef
13.
Zurück zum Zitat Reichlin T, Michaud GF. Our approach to maximizing the durability of pulmonary vein isolation during a paroxysmal atrial fibrillation ablation procedure. J Cardiovasc Electrophysiol. 2012;23:1272–6.CrossRef Reichlin T, Michaud GF. Our approach to maximizing the durability of pulmonary vein isolation during a paroxysmal atrial fibrillation ablation procedure. J Cardiovasc Electrophysiol. 2012;23:1272–6.CrossRef
14.
Zurück zum Zitat Vaseghi M, Cesario DA, Valderrabano M, Boyle NG, Ratib O, Finn JP, et al. Impedance monitoring during catheter ablation of atrial fibrillation. Heart Rhythm. 2005;2:914–20.CrossRef Vaseghi M, Cesario DA, Valderrabano M, Boyle NG, Ratib O, Finn JP, et al. Impedance monitoring during catheter ablation of atrial fibrillation. Heart Rhythm. 2005;2:914–20.CrossRef
15.
Zurück zum Zitat Scharf C, Sneider M, Case I, Chugh A, Lai SW, Pelosi F Jr, et al. Anatomy of the pulmonary veins in patients with atrial fibrillation and effects of segmental ostial ablation analyzed by computed tomography. J Cardiovasc Electrophysiol. 2003;14:150–5.CrossRef Scharf C, Sneider M, Case I, Chugh A, Lai SW, Pelosi F Jr, et al. Anatomy of the pulmonary veins in patients with atrial fibrillation and effects of segmental ostial ablation analyzed by computed tomography. J Cardiovasc Electrophysiol. 2003;14:150–5.CrossRef
16.
Zurück zum Zitat Knecht S, Reichlin T, Pavlovic N, Schaer B, Osswald S, Sticherling C, et al. Contact force and impedance decrease during ablation depends on catheter location and orientation: insights from pulmonary vein isolation using a contact force-sensing catheter. J Interv Card Electrophysiol. 2015;43:297–306.CrossRef Knecht S, Reichlin T, Pavlovic N, Schaer B, Osswald S, Sticherling C, et al. Contact force and impedance decrease during ablation depends on catheter location and orientation: insights from pulmonary vein isolation using a contact force-sensing catheter. J Interv Card Electrophysiol. 2015;43:297–306.CrossRef
17.
Zurück zum Zitat Kistler PM, Ho SY, Rajappan K, et al. Electrophysiologic and anatomic characterization of sites resistant to electrical isolation during circumferential pulmonary vein ablation for atrial fibrillation: a prospective study. J Cardiovasc Electrophysiol. 2007;18:1282–8.CrossRef Kistler PM, Ho SY, Rajappan K, et al. Electrophysiologic and anatomic characterization of sites resistant to electrical isolation during circumferential pulmonary vein ablation for atrial fibrillation: a prospective study. J Cardiovasc Electrophysiol. 2007;18:1282–8.CrossRef
18.
Zurück zum Zitat Ho SH, Cabrera JA, Sanchez-Quintana D. Left atrial anatomy revisited. Circ Arrhythm Electrophysiol. 2012;5:220–8.CrossRef Ho SH, Cabrera JA, Sanchez-Quintana D. Left atrial anatomy revisited. Circ Arrhythm Electrophysiol. 2012;5:220–8.CrossRef
19.
Zurück zum Zitat Woźniak-Skowerska I, Skowerski M, Wnuk-Wojnar A, Hoffmann A, Nowak S, Gola A, et al. Comparison of pulmonary veins anatomy in patients with and without atrial fibrillation: analysis by multislice tomography. Int J Cardiol. 2011;146:181–5.CrossRef Woźniak-Skowerska I, Skowerski M, Wnuk-Wojnar A, Hoffmann A, Nowak S, Gola A, et al. Comparison of pulmonary veins anatomy in patients with and without atrial fibrillation: analysis by multislice tomography. Int J Cardiol. 2011;146:181–5.CrossRef
20.
Zurück zum Zitat Tsao HM, Hu WC, Wu MH, Tai CT, Chang SL, Lin YJ, et al. Characterization of the dynamic function of the pulmonary veins before and after atrial fibrillation ablation using multi-detector computed tomographic images. Int J Cardiovasc Imaging. 2011;27:1049–58.CrossRef Tsao HM, Hu WC, Wu MH, Tai CT, Chang SL, Lin YJ, et al. Characterization of the dynamic function of the pulmonary veins before and after atrial fibrillation ablation using multi-detector computed tomographic images. Int J Cardiovasc Imaging. 2011;27:1049–58.CrossRef
21.
Zurück zum Zitat Buist TJ, Gal P, Ottervanger JP, Smit JJ, Ramdat Misier AR, Delnoy PP, et al. Association between pulmonary vein orientation and ablation outcome in patients undergoing multi-electrode ablation for atrial fibrillation. Cardiovasc Comput Tomogr. 2016;10:251–7.CrossRef Buist TJ, Gal P, Ottervanger JP, Smit JJ, Ramdat Misier AR, Delnoy PP, et al. Association between pulmonary vein orientation and ablation outcome in patients undergoing multi-electrode ablation for atrial fibrillation. Cardiovasc Comput Tomogr. 2016;10:251–7.CrossRef
22.
Zurück zum Zitat Gebhard C, Krasniqi N, Stähli BE, Klaeser B, Fuchs TA, Ghadri JR, et al. Characterization of pulmonary vein dimensions using high-definition 64-slice computed tomography prior to radiofrequency catheter ablation for atrial fibrillation. Cardiol Res Pract. 2014;2014:179632 1–8.CrossRef Gebhard C, Krasniqi N, Stähli BE, Klaeser B, Fuchs TA, Ghadri JR, et al. Characterization of pulmonary vein dimensions using high-definition 64-slice computed tomography prior to radiofrequency catheter ablation for atrial fibrillation. Cardiol Res Pract. 2014;2014:179632 1–8.CrossRef
23.
Zurück zum Zitat Fourdrain A, De Dominicis F, Bensussan M, Iquille J, Lafitte S, Michel D, et al. Three-dimensional computed tomography angiography of the pulmonary veins and their anatomical variations: involvement in video-assisted thoracoscopic surgery-lobectomy for lung cancer. Folia Morphol (Warsz). 2017;76:388–93.CrossRef Fourdrain A, De Dominicis F, Bensussan M, Iquille J, Lafitte S, Michel D, et al. Three-dimensional computed tomography angiography of the pulmonary veins and their anatomical variations: involvement in video-assisted thoracoscopic surgery-lobectomy for lung cancer. Folia Morphol (Warsz). 2017;76:388–93.CrossRef
24.
Zurück zum Zitat Okumura Y, Watanabe I, Kofune M, Nagashima K, Sonoda K, Mano H, et al. Effect of catheter tip-tissue surface contact on three-dimensional left atrial and pulmonary vein geometries: potential anatomic distortion of 3D ultrasound, fast anatomical mapping, and merged 3D CT-derived images. J Cardiovasc Electrophysiol. 2013;24:259–66.CrossRef Okumura Y, Watanabe I, Kofune M, Nagashima K, Sonoda K, Mano H, et al. Effect of catheter tip-tissue surface contact on three-dimensional left atrial and pulmonary vein geometries: potential anatomic distortion of 3D ultrasound, fast anatomical mapping, and merged 3D CT-derived images. J Cardiovasc Electrophysiol. 2013;24:259–66.CrossRef
25.
Zurück zum Zitat Spies F, Kühne M, Reichlin T, Osswald S, Sticherling C, Knecht S. A quantitative comparison of the electrical and anatomical definition of the pulmonary vein ostium. Pacing Clin Electrophysiol. 2017;40:1213–7.CrossRef Spies F, Kühne M, Reichlin T, Osswald S, Sticherling C, Knecht S. A quantitative comparison of the electrical and anatomical definition of the pulmonary vein ostium. Pacing Clin Electrophysiol. 2017;40:1213–7.CrossRef
26.
Zurück zum Zitat Shah D. Electrophysiological evaluation of pulmonary vein isolation. Europace. 2009;11:1423–33.CrossRef Shah D. Electrophysiological evaluation of pulmonary vein isolation. Europace. 2009;11:1423–33.CrossRef
Metadaten
Titel
Impedance mapping with constant contact force on 3D electroanatomic map to characterize tissues at pulmonary veno-atrial junction
verfasst von
Corrado Tomasi
Alessandro Dal Monte
Maria Selina Argnani
Cristiana Corsi
Federica Giannotti
Stefano Severi
Andrea Rubboli
Publikationsdatum
04.08.2020
Verlag
Springer US
Schlagwort
Atrial Fibrillation
Erschienen in
Journal of Interventional Cardiac Electrophysiology / Ausgabe 3/2021
Print ISSN: 1383-875X
Elektronische ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-020-00845-4

Weitere Artikel der Ausgabe 3/2021

Journal of Interventional Cardiac Electrophysiology 3/2021 Zur Ausgabe

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

„Jeder Fall von plötzlichem Tod muss obduziert werden!“

17.05.2024 Plötzlicher Herztod Nachrichten

Ein signifikanter Anteil der Fälle von plötzlichem Herztod ist genetisch bedingt. Um ihre Verwandten vor diesem Schicksal zu bewahren, sollten jüngere Personen, die plötzlich unerwartet versterben, ausnahmslos einer Autopsie unterzogen werden.

Hirnblutung unter DOAK und VKA ähnlich bedrohlich

17.05.2024 Direkte orale Antikoagulanzien Nachrichten

Kommt es zu einer nichttraumatischen Hirnblutung, spielt es keine große Rolle, ob die Betroffenen zuvor direkt wirksame orale Antikoagulanzien oder Marcumar bekommen haben: Die Prognose ist ähnlich schlecht.

Schlechtere Vorhofflimmern-Prognose bei kleinem linken Ventrikel

17.05.2024 Vorhofflimmern Nachrichten

Nicht nur ein vergrößerter, sondern auch ein kleiner linker Ventrikel ist bei Vorhofflimmern mit einer erhöhten Komplikationsrate assoziiert. Der Zusammenhang besteht nach Daten aus China unabhängig von anderen Risikofaktoren.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.