Background

Intraoperative blood loss during hepatectomy worsens prognosis, and various tools have been used to improve perioperative safety and feasibility. We aimed to retrospectively evaluate the feasibility and safety of the BiClamp® device for open liver resection.

Methods

We included 84 patients undergoing liver resection from a single centre, with all patients operated by the same surgical group. All hepatectomies were performed using BiClamp® (Erbe Elektromedizin GmbH, Tubingen, Germany), an electrosurgical device that simultaneously transects liver parenchyma and seals vessels <7 mm in diameter. We collected data on intraoperative blood loss, resection time, and perioperative complications, comparing cirrhotic and non-cirrhotic patients.

Results

The 84 patients enrolled in this study included 56 cirrhotic and 28 non-cirrhotic patients. All patients underwent hepatectomy (30 major and 54 minor hepatectomies) using the BiClamp®, exclusively, and 54 patients required inflow occlusion (Pringle manoeuvre). Overall intraoperative blood loss (mean ± standard deviation) was 523.5 ± 558.6 ml, liver parenchymal transection time was 36.3 ± 16.5 min (range, 13-80 min), and the mean parenchymal transection speed was 3.0 ± 1.9 cm2/min. Twelve patients received perioperative blood transfusion. The cost of BiClamp® for each patient was 800 RMB (approximately 109€). There were no deaths, and the morbidity rate was 25%. The mean (standard deviation) hospital stay was 9.3 (2.3) days. Comparisons between cirrhotic and non-cirrhotic patients revealed no difference in blood loss (491.0 ± 535.7 ml vs 588.8 ± 617.5 ml, P = 0.598), liver parenchymal transection time (34.1 ± 14.8 min vs 40.9 ± 19.2 min, P = 0.208), mean parenchymal transection speed (3.3 ± 2.1 cm2/min vs 2.5 ± 1.3 cm2/min, P = 0.217), and operative morbidity (28.6% vs 14.3%, P = 0.147).

Conclusions

The reusable BiClamp® vessel-sealing device allows for safe and feasible major and minor hepatectomy, even in patients with cirrhotic liver.

Trial registration

This trial was retrospectively registered and the detail information was as followed. Registration number: ChiCTR-ORC-17011873 (Chinese Clinical Trial Registry). Registration Date: 2017-07-05.
Hinweise

Electronic supplementary material

The online version of this article (doi:10.​1186/​s12885-017-3513-0) contains supplementary material, which is available to authorized users.
Abkürzungen
F
Female
M
Male
SD
Standard deviation

Background

Liver resection is a demanding procedure that is often the only chance of a cure for many patients [1, 2]. Intraoperative blood loss during hepatectomy is a significant factor in determining postoperative morbidity and mortality [3], and several surgical techniques have been recommended to reduce blood loss. Inflow occlusion (Pringle manoeuvre) and low central venous pressure anaesthesia are proven effective methods to minimize haemorrhage during parenchymal transection [4, 5]. Since the first introduction of the clamp-crushing technique in the 1970s [6], surgical strategies to reduce blood loss in patients undergoing hepatectomy have focused mainly on technical innovations in liver parenchymal transection, including an ultrasonic dissector (CUSA®, Tyco Healthcare, Mansfield, MA, USA), the Waterjet® (Erbe Elektromedizin GmbH, Tubingen, Germany), a dissecting sealer (TissueLink®, Dover, NH, USA), and others. However, randomised controlled trials have shown that the clamp-crushing technique remains the reference technique for liver parenchymal transection compared with alternative transection techniques [7].
BiClamp® (Erbe Elektromedizin GmbH) was designed as a reusable bipolar vessel-sealing device that successfully ligates 2–7 mm arteries and veins [8]. The shape of BiClamp® is similar to a typical clamp; therefore, it can be used in liver transection similar to the clamp-crushing technique, and the device’s coagulation function may address limitations in the clamp-crushing technique. Uchiyama and colleagues used BiClamp® in laparoscopic partial hepatectomy [9]. The median tumour diameter of the nine patients in their study was 1.5 cm, and results confirmed the safety and efficacy of BiClamp® in laparoscopic partial hepatectomy. In another study, BiClamp® was added for liver parenchymal transection with the CUSA device [10]. The authors found lower intraoperative blood loss in the BiClamp® group compared with the group receiving CUSA combined with bipolar electrocautery.
To our knowledge, there are no previous publications evaluating the use of BiClamp® for open liver resection, especially in patients with cirrhosis. We performed this respective study to evaluate the feasibility and safety of this surgical device for open liver parenchymal transection [11].

Methods

Previous experience

We have used the BiClamp® for liver resection since 2007, and more than 200 patients underwent hepatectomy in our institution before the current trial. For patient safety reasons, the clamp-crushing technique and inflow occlusion (Pringle manoeuvre) were necessary when first using the BiClamp® in our institution. However, after obtaining acceptable operative complication and blood transfusion rates, we realized that liver resection can be performed using the BiClamp® exclusively. The overall rate of inflow occlusion decreased during this study’s duration because inflow occlusion is unnecessary when the intrahepatic duct structure is clearly distinguishable, which decreases ischemia reperfusion injury caused by hepatic ischemia.

Experimental design

We enrolled 84 consecutive patients between July 2010 and July 2012, and all patients were operated by the same surgical group. Eligibility criteria included patients with benign and malignant tumours and acceptable liver function (Child-Pugh score A), and acceptable status to undergo the operation. Patients undergoing local partial hepatectomy for less than one segment were excluded from the study.

Surgical procedure

Intraoperative ultrasound was used to define the tumour location relative to the major vascular and biliary structures. The power supply for the BiClamp® was fixed at level 4, and the effect was at set at 100. Inflow occlusion was performed only when blood loss exceeded the expected amount.
The BiClamp® has two blades that can crush liver parenchyma like a clamp (Fig. 1), and the device’s coagulation function seals intrahepatic vessels simultaneously (Fig. 2). Larger intrahepatic structures can be isolated by the BiClamp®; vessels can then be divided by an electrotome or scissors followed by standard ligation or ligation with titanium clips (see Additional file 1: Video S1).
All resections in our study were performed with low central venous pressure (0–5 mmHg), and prophylactic postoperative drainage was performed routinely. Major hepatectomy was defined as liver resection involving ≥ three segments, and minor hepatectomy was defined as resection of < three segments.

Outcome measures

Each patient was followed for at least 3 months and the safety of the BiClamp® was evaluated based on postoperative morbidity and mortality. The feasibility of the BiClamp® was evaluated by the speed of liver parenchymal transection and by blood loss. Resection time was defined as the duration between the beginning and the end of parenchymal transection. Following transection, blood loss volume was estimated by the weight of the suction fluid and blood-soaked swabs after subtracting the lavage fluid volume and the weight of any dry swabs that were used during transection using the formula 1 g = 1 ml of blood.
A sheet of sterile paper was placed against the liver cut surface immediately following liver resection to duplicate the liver cut-surface area. The outline of the cut surface on the sterile paper was copied to a sheet of dry 120 g/m2 paper, postoperatively. The liver cut-surface area was then calculated using the 120 g/m2 density of the second paper. The transection area was expressed as cm2, and the speed of transection was expressed as the transection area divided by the transection time (cm2/min). Complications were recorded based on severity according to the classification system [12]. Mortality was defined as death within 30 days of hepatectomy. Bile leakage was defined as a total bilirubin level ≥ 86 μmol/L in the drained fluid [13], and the indications for blood transfusion included blood loss >1500 ml or a haemoglobin level < 70 g/L.

Statistical analysis

Results are shown as mean ± standard deviation (SD). Fischer’s exact, χ2, and independent samples t-tests were used when needed. The difference was considered significant if P < 0.05. All statistical calculations were performed using SPSS 17.0 (SPSS, Inc., Chicago, IL).

Results

The following neoplasms occurred in the 84 patients in this study: hepatocellular carcinoma (56 cases), cholangiocarcinoma (6 cases), colorectal liver metastases (6 cases), and haemangioma (16 cases). There were 74 first, 6 s, and 4 third hepatectomies. Patients’ clinical details are shown in Table 1.
Table 1
Patient demographics and tumour characteristics
Characteristic
Finding
No. of patients
84
Gender, No, (F/M)
18/66
Age, Mean ± standard deviation (SD), (Y)
51.8 ± 11.3
Number of patients with cirrhosis
56
Hepatocellular carcinoma
56
Cholangiocarcinoma
6
Colorectal Metastases
6
Haemangioma
16
Tumour Size, Mean ± SD, (mm)
72 ± 32
No. of tumours, Mean ± SD
1.6 ± 1.3
Thirty patients underwent major hepatectomies and 54 patients underwent minor hepatectomies. The intraoperative transection-related features and surgical outcomes are shown in Table 2. The overall operative time was 168.9 ± 44.7 min (range, 100–300 min), and the parenchymal transection time was 36.3 ± 16.5 min (range, 13–80 min). The liver cut-surface area measurement was 95.1 ± 43.9 cm2, parenchymal transection time per square centimetre was 0.43 ± 0.23 min, and the speed of parenchymal transection was 3.0 ± 1.9 cm2/min. The overall intraoperative blood loss was 523.5 ± 558.6 ml (range, 55.0–2474.1 ml), and the mean blood loss volume per square centimetre was 6.2 ± 7.6 ml (range, 0.6–39.8 ml). Fifty-four patients required hepatic inflow occlusion (Pringle manoeuvre) during liver resection, but only 12 patients required intraoperative blood transfusion. No patients developed grounding pad skin burn, myocardial infarction, or cardiac arrhythmia during or after the operation, and there were no deaths within 30 days postoperatively. The cost of BiClamp® for each patient was 800 RMB (approximately 109€).
Table 2
Intraoperative transection-related features and surgical outcomes (n = 84)
Features
Finding
Operative time (minutes)
168.9 ± 44.7 (168.9 ± 44.7)
Parenchymal transaction time (minutes)
36.3 ± 16.5 (13-80)
Liver cut suface area (cm2)
95.1 ± 43.9 (31.82 ~ 202.43)
Mean transection speed (cm2/min)
3.0 ± 1.9
Total blood loss (ml)
523.5 ± 558.6(55.0 ~ 2474.1)
Mean blood loss (mL/cm2)
6.2 ± 7.6 (0.6 ~ 39.8)
No. of transfused patients
12 (14.3%)
No. of patients with Pringle (%)
54 (64.3%)
Overall operative morbidity
20 (25%)
Operative mortality
0 (0%)
Inhospital (days)
9.3 ± 2.7
Expense of BiClamp for each patient
800 RMB (approximately 109€)
Fifteen patients suffered grade 1–2 postoperative complications according to the new classification [12] with 13 patients developing right pleural effusion, which was diagnosed on upper abdominal CT. All 13 patients recovered smoothly without additional treatment. One patient developed biliary leak that was diagnosed by the total bilirubin level in the abdominal drainage fluid [13]; this resolved after 2 weeks of postoperative drainage. One patient suffered transient liver dysfunction with pleural effusion and ascites that resolved with albumin therapy and diuresis. No patients developed postoperative bleeding and none required reoperation. The mean postoperative hospital stay was 9.3 ± 2.3 days (range, 5–18 days). No readmissions occurred within the 3-month follow-up.
The surgical characteristics for patients in the cirrhotic and non-cirrhotic groups are compared in Table 3 . Other than the significantly higher proportion of males in the cirrhotic group compared with the non-cirrhotic group, there were no differences between the two groups for: operation time (166.4 ± 39.2 min vs 174.0 ± 55.5 min, P = 0.611), parenchymal transection time (34.1 ± 14.8 min vs 40.9 ± 19.2 min, P = 0.208), liver cut-surface area (98.4 ± 48.1 cm2 vs 88.7 ± 35.0 cm2, P = 0.508), mean transection speed (3.3 ± 2.1 cm2/min vs 2.5 ± 1.3 cm2/min, P = 0.217), blood loss (491.0 ± 535.7 ml vs 588.8 ± 617.5 ml, P = 0.598), Mean blood loss (mL/cm2) (5.5 ± 6.1 VS 7.6 ± 10.1, P = 0.406), number of patients requiring transfusion (6/56 vs 6/28, P = 0.321), patients receiving the Pringle manoeuvre (34/56 vs 20/28, P = 0.469), complications (16/56 vs 4/28, P = 0.147), or length of hospital stay (9.5 ± 3.1 days vs 9.1 ± 2.1 days, P = 0.635).
Table 3
Comparison of surgical results between patients with and without cirrhosis
Features
Cirrhosis
(n = 56)
Non-cirrhosis
(n = 28)
P value
Gender, No, (F/M)
3/53
15/13
0.000a
Age, (Y)
50.4 ± 12.3
54.9 ± 8.7
0.228
Operative time (minutes)
166.4 ± 39.2
174.0 ± 55.5
0.611
Parenchymal transaction time (minutes)
34.1 ± 14.8
40.9 ± 19.2
0.208
Liver cut surface area (cm2)
98.4 ± 48.1
88.7 ± 35.0
0.508
Mean transection speed (cm2/min)
3.3 ± 2.1
2.5 ± 1.3
0.217
Total blood loss (ml)
491.0 ± 535.7
588.8 ± 617.5
0.598
Mean blood loss (mL/cm2)
5.5 ± 6.1
7.6 ± 10.1
0.406
No. of transfused patients (0%)
6 (10.7%)
6 (21.4%)
0.321b
No. of patients with Pringle (%)
34 (42.9%)
20 (71.4%)
0.469b
Operative morbidity (%)
16 (28.6%)
4 (14.3%)
0.147a
Inhospital (days)
9.5 ± 3.1
9.1 ± 2.1
0.635
Comparisons of mean ± SD used independent samples t-tests; aFisher’s exact; bχ2

Discussion

BiClamp® is an innovative bipolar coagulation system with adjustable current modulation and intelligent self-control, which transforms electrical energy into heat, thus attaining an ideal energy-based vessel seal. Vessels with a diameter ≤ 7 mm can be successfully sealed by BiClamp alone [9]. BiClamp® is a reusable, cost-effective instrument.
BiClamp® has been widely used in thyroidectomies, hysterectomies, and pulmonary lobectomies, with proven efficacy and safety [1416]. However, experience using the BiClamp® in open liver resections is limited. BiClamp® has two blades that simultaneously crush liver parenchyma like a clamp and seal intrahepatic vessels via the coagulation function. Theoretically, BiClamp®‘s vessel-sealing function could be more effective in liver parenchymal transection than clamp crushing.
In Uchiyama and colleagues’ study [10], BiClamp® was applied to laparoscopic hepatectomies, and the estimated blood loss was as low as 417 ml with all patients recovering smoothly without complications. In another study [10] in which BiClamp® was used in addition to CUSA, the median blood loss was 345 mL in the BiClamp® group, which was less than that in the group receiving CUSA combined with bipolar electrocautery. In our patients, the mean tumour diameter was 7.2 cm, and 30 patients underwent major hepatectomies including 10 anatomical hemi-hepatectomies.
Our results showed that some complicated hepatectomies can be performed safely and efficiently using BiClamp® alone, even in some cirrhotic patients. We encountered no severe postoperative morbidity and mortality according to the classification of complications by severity [12].
We found that using BiClamp® was similar to the clamp-crushing technique, and that coagulation helped with hemostasis of the liver cut surface. Intrahepatic structures are coagulated during liver parenchymal transection without transecting the vessels, using BiClamp®. As seen in the supplemental video, titanium clips were used to ligate the small vessels before transection, and 4-0 Prolene (Ethicon, Somerville, NJ, USA) was used to ligate the major intrahepatic vasculature when necessary.
Another useful feature of BiClamp® in liver parenchymal transection is that the saline lavage following coagulation prevents liver tissue from adhering to the BiClamp® blades, which also prevents carbonization of the liver cut surface. Using BiClamp® is easy; however, liver parenchymal transection must proceed slowly while activating the BiClamp®, which is better for small vessel coagulation.
Liver parenchymal transection speed and blood loss were the most important parameters when we evaluated the feasibility of BiClamp® for hepatectomy. Previous reports indicated a clamp-crushing speed of 0.89–3.9 cm2/min, and a mean blood loss per square centimetre of 1.5–7.0 ml [1719]. Our results indicated a parenchymal transection speed of 3.0 ± 1.9 cm2/min and a mean blood loss per square centimetre of 6.2 ± 7.6 ml, similar results to previous studies.
Hepatic cirrhosis may be an adverse factor leading to increased bleeding [20]. Our results showed no difference in intraoperative blood loss, transection speed, proportion of hepatic inflow occlusion, or need for transfusion between the cirrhosis and non-cirrhosis groups using BiClamp®. The post-operative complication incidence in our cirrhosis group was higher than that in the non-cirrhosis group but with no significant difference, suggesting that BiClamp® may be more advantageous for hepatectomy in patients with cirrhosis.
Another advantage of BiClamp® was to reduce the cost for patients due to its reusable. Being calculated by deprecition and disinfection cost of instruments, each patient should only pay 800RMB (approximately 109€) for BiClamp®, while the cost of CUSA [17] and Ligasure [21] was reported 661€ and 447.2 ± 58.9€, respectively.
The limitations of this study include the retrospective, single-centre design, the small sample size, and the short-term follow-up. Prospective, randomised control trials are required to confirm our results; therefore, we designed a prospective randomised trial to compare BiClamp® with clamp-crushing methods in open liver resection (NCT02197481) [22].

Conclusions

Overall, using the reusable BiClamp® vessel-sealing device assures hepatectomy safety and effectiveness, even in patients with cirrhotic liver.

Acknowledgements

None.

Funding

This study was supported by the Key Project of Anhui Province on Hepatocellular Carcinoma (2010A009) in the design of the study and collection, analysis, and interpretation of data and in writing the manuscript. The data used in this study were collected from the database supported by the Youth’s Project of Anhui Province (2009B098).

Availability of data and materials

The data sets supporting the conclusions of this article are included within the article. Data are available from the corresponding author upon reasonable request.
The project has been approved by the Committee on Medical Ethics of The First Affiliated Hospital of Anhui Medical University. The reference number is “Quick-PJ 2017-05-13”. Informed consent was obtained from all individuals included in the study and all the informed consent was written.
Not applicable.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
1.
Zurück zum Zitat Jarnagin WR, Gonen M, Fong Y, DeMatteo RP, Ben-Porat L, Little S, Corvera C, Weber S, Blumgart LH. Improvement in perioperative outcome after hepatic resection: analysis of 1,803 consecutive cases over the past decade. Ann Surg. 2002;236(4):397–406. 406-407CrossRefPubMedPubMedCentral Jarnagin WR, Gonen M, Fong Y, DeMatteo RP, Ben-Porat L, Little S, Corvera C, Weber S, Blumgart LH. Improvement in perioperative outcome after hepatic resection: analysis of 1,803 consecutive cases over the past decade. Ann Surg. 2002;236(4):397–406. 406-407CrossRefPubMedPubMedCentral
2.
Zurück zum Zitat Poon RT, Fan ST, Lo CM, Liu CL, Lam CM, Yuen WK, Yeung C, Wong J. Improving perioperative outcome expands the role of hepatectomy in management of benign and malignant hepatobiliary diseases: analysis of 1222 consecutive patients from a prospective database. Ann Surg. 2004;240(4):698–708. 708-710PubMedPubMedCentral Poon RT, Fan ST, Lo CM, Liu CL, Lam CM, Yuen WK, Yeung C, Wong J. Improving perioperative outcome expands the role of hepatectomy in management of benign and malignant hepatobiliary diseases: analysis of 1222 consecutive patients from a prospective database. Ann Surg. 2004;240(4):698–708. 708-710PubMedPubMedCentral
3.
Zurück zum Zitat Kooby DA, Stockman J, Ben-Porat L, Gonen M, Jarnagin WR, Dematteo RP, Tuorto S, Wuest D, Blumgart LH, Fong Y. Influence of transfusions on perioperative and long-term outcome in patients following hepatic resection for colorectal metastases. Ann Surg. 2003;237(6):860–9. 869-870PubMedPubMedCentral Kooby DA, Stockman J, Ben-Porat L, Gonen M, Jarnagin WR, Dematteo RP, Tuorto S, Wuest D, Blumgart LH, Fong Y. Influence of transfusions on perioperative and long-term outcome in patients following hepatic resection for colorectal metastases. Ann Surg. 2003;237(6):860–9. 869-870PubMedPubMedCentral
4.
Zurück zum Zitat Kimura F, Miyazaki M, Suwa T, Sugiura T, Shinoda T, Itoh H, Nakagawa K, Ambiru S, Shimizu H, Yoshitome H. Evaluation of total hepatic vascular exclusion and pringle maneuver in liver resection. Hepato-Gastroenterology. 2002;49(43):225–30.PubMed Kimura F, Miyazaki M, Suwa T, Sugiura T, Shinoda T, Itoh H, Nakagawa K, Ambiru S, Shimizu H, Yoshitome H. Evaluation of total hepatic vascular exclusion and pringle maneuver in liver resection. Hepato-Gastroenterology. 2002;49(43):225–30.PubMed
5.
Zurück zum Zitat Melendez JA, Arslan V, Fischer ME, Wuest D, Jarnagin WR, Fong Y, Blumgart LH. Perioperative outcomes of major hepatic resections under low central venous pressure anesthesia: blood loss, blood transfusion, and the risk of postoperative renal dysfunction. J Am Coll Surg. 1998;187(6):620–5.CrossRefPubMed Melendez JA, Arslan V, Fischer ME, Wuest D, Jarnagin WR, Fong Y, Blumgart LH. Perioperative outcomes of major hepatic resections under low central venous pressure anesthesia: blood loss, blood transfusion, and the risk of postoperative renal dysfunction. J Am Coll Surg. 1998;187(6):620–5.CrossRefPubMed
6.
7.
Zurück zum Zitat Rahbari NN, Koch M, Schmidt T, Motschall E, Bruckner T, Weidmann K, Mehrabi A, Büchler MW, Weitz J. Meta-analysis of the clamp-crushing technique for Transection of the parenchyma in elective hepatic resection: back to where we started? Ann Surg Oncol. 2009;16(3):630–9.CrossRefPubMed Rahbari NN, Koch M, Schmidt T, Motschall E, Bruckner T, Weidmann K, Mehrabi A, Büchler MW, Weitz J. Meta-analysis of the clamp-crushing technique for Transection of the parenchyma in elective hepatic resection: back to where we started? Ann Surg Oncol. 2009;16(3):630–9.CrossRefPubMed
8.
Zurück zum Zitat Richter S, Kollmar O, Neunhoeffer E, Schilling MK, Menger MD, Pistorius G. Differential response of arteries and veins to bipolar vessel sealing: evaluation of a novel reusable device. J Laparoendosc Adv Surg Tech A. 2006;16(2):149–55.CrossRefPubMed Richter S, Kollmar O, Neunhoeffer E, Schilling MK, Menger MD, Pistorius G. Differential response of arteries and veins to bipolar vessel sealing: evaluation of a novel reusable device. J Laparoendosc Adv Surg Tech A. 2006;16(2):149–55.CrossRefPubMed
9.
Zurück zum Zitat Uchiyama H, Itoh S, Higashi T, Korenaga D, Takenaka K. Pure laparoscopic partial hepatectomy using a newly developed vessel sealing device, BiClamp. Surg Laparosc Endosc Percutan Tech. 2013;23(3):e116–8.CrossRefPubMed Uchiyama H, Itoh S, Higashi T, Korenaga D, Takenaka K. Pure laparoscopic partial hepatectomy using a newly developed vessel sealing device, BiClamp. Surg Laparosc Endosc Percutan Tech. 2013;23(3):e116–8.CrossRefPubMed
10.
Zurück zum Zitat Itoh S, Fukuzawa K, Shitomi Y, Okamoto M, Kinoshita T, Taketomi A, Shirabe K, Wakasugi K, Maehara Y. Impact of the VIO system in hepatic resection for patients with hepatocellular carcinoma. Surg Today. 2012;42(12):1176–82.CrossRefPubMed Itoh S, Fukuzawa K, Shitomi Y, Okamoto M, Kinoshita T, Taketomi A, Shirabe K, Wakasugi K, Maehara Y. Impact of the VIO system in hepatic resection for patients with hepatocellular carcinoma. Surg Today. 2012;42(12):1176–82.CrossRefPubMed
11.
Zurück zum Zitat Zhao YJ, Geng XP. Evaluation of the BiClamp® vessel–sealingdevice for open hepatic resection: experience from a single institution. J Hepatobiliary Pancreat Sci. 2017; A322.P-58-6. Zhao YJ, Geng XP. Evaluation of the BiClamp® vessel–sealingdevice for open hepatic resection: experience from a single institution. J Hepatobiliary Pancreat Sci. 2017; A322.P-58-6.
12.
Zurück zum Zitat Clavien PA, Barkun J, de Oliveira ML, Vauthey JN, Dindo D, Schulick RD, de Santibanes E, Pekolj J, Slankamenac K, Bassi C, et al. The Clavien-Dindo classification of surgical complications: five-year experience. Ann Surg. 2009;250(2):187–96.CrossRefPubMed Clavien PA, Barkun J, de Oliveira ML, Vauthey JN, Dindo D, Schulick RD, de Santibanes E, Pekolj J, Slankamenac K, Bassi C, et al. The Clavien-Dindo classification of surgical complications: five-year experience. Ann Surg. 2009;250(2):187–96.CrossRefPubMed
13.
Zurück zum Zitat Ijichi M, Takayama T, Toyoda H, Sano K, Kubota K, Makuuchi M. Randomized trial of the usefulness of a bile leakage test during hepatic resection. Arch Surg. 2000;135(12):1395–400.CrossRefPubMed Ijichi M, Takayama T, Toyoda H, Sano K, Kubota K, Makuuchi M. Randomized trial of the usefulness of a bile leakage test during hepatic resection. Arch Surg. 2000;135(12):1395–400.CrossRefPubMed
14.
Zurück zum Zitat Oussoultzoglou E, Panaro F, Rosso E, Zeca I, Bachellier P, Pessaux P, Jaeck D. Use of BiClamp decreased the severity of Hypocalcemia after Total Thyroidectomy compared with LigaSure: a prospective study. World J Surg. 2008;32(9):1968–73.CrossRefPubMed Oussoultzoglou E, Panaro F, Rosso E, Zeca I, Bachellier P, Pessaux P, Jaeck D. Use of BiClamp decreased the severity of Hypocalcemia after Total Thyroidectomy compared with LigaSure: a prospective study. World J Surg. 2008;32(9):1968–73.CrossRefPubMed
15.
Zurück zum Zitat Zubke W, Hornung R, Wasserer S, Hucke J, Fullers U, Werner C, Schmitz P, Lobodasch K, Hammermuller U, Wojdat R, et al. Bipolar coagulation with the BiClamp forceps versus conventional suture ligation: a multicenter randomized controlled trial in 175 vaginal hysterectomy patients. Arch Gynecol Obstet. 2009;280(5):753–60.CrossRefPubMed Zubke W, Hornung R, Wasserer S, Hucke J, Fullers U, Werner C, Schmitz P, Lobodasch K, Hammermuller U, Wojdat R, et al. Bipolar coagulation with the BiClamp forceps versus conventional suture ligation: a multicenter randomized controlled trial in 175 vaginal hysterectomy patients. Arch Gynecol Obstet. 2009;280(5):753–60.CrossRefPubMed
16.
Zurück zum Zitat Sakuragi T, Okazaki Y, Mitsuoka M, Yamasaki F, Masuda M, Mori D, Satoh T, Itoh T. The utility of a reusable bipolar sealing instrument, BiClamp((R)), for pulmonary resection. Eur J Cardiothorac Surg. 2008;34(3):505–9.CrossRefPubMed Sakuragi T, Okazaki Y, Mitsuoka M, Yamasaki F, Masuda M, Mori D, Satoh T, Itoh T. The utility of a reusable bipolar sealing instrument, BiClamp((R)), for pulmonary resection. Eur J Cardiothorac Surg. 2008;34(3):505–9.CrossRefPubMed
17.
18.
Zurück zum Zitat Ikeda M, Hasegawa K, Sano K, Imamura H, Beck Y, Sugawara Y, Kokudo N, Makuuchi M. The vessel sealing system (LigaSure) in hepatic resection. Ann Surg. 2009;250(2):199–203.CrossRefPubMed Ikeda M, Hasegawa K, Sano K, Imamura H, Beck Y, Sugawara Y, Kokudo N, Makuuchi M. The vessel sealing system (LigaSure) in hepatic resection. Ann Surg. 2009;250(2):199–203.CrossRefPubMed
19.
Zurück zum Zitat Arita J, Hasegawa K, Kokudo N, Sano K, Sugawara Y. Randomized clinical trial of the effect of a saline-linked radiofrequency coagulator on blood loss during hepatic resection. Brit J Surg. 2005;92:954–9. Arita J, Hasegawa K, Kokudo N, Sano K, Sugawara Y. Randomized clinical trial of the effect of a saline-linked radiofrequency coagulator on blood loss during hepatic resection. Brit J Surg. 2005;92:954–9.
20.
Zurück zum Zitat Romano F, Franciosi C, Caprotti R, Uggeri F, Uggeri F. Hepatic surgery using the Ligasure vessel sealing system. World J Surg. 2005;29(1):110–2.CrossRefPubMed Romano F, Franciosi C, Caprotti R, Uggeri F, Uggeri F. Hepatic surgery using the Ligasure vessel sealing system. World J Surg. 2005;29(1):110–2.CrossRefPubMed
21.
Zurück zum Zitat Thompson IM, Kappa SF, Morgan TM, Barocas DA, Bischoff CJ, Keegan KA, Stratton KL, Clark PE, Resnick MJ, Smith JA, et al. Blood loss associated with radical cystectomy: a prospective, randomized study comparing impact LigaSure vs. stapling device. Urol Oncol: Semin Orig Investig. 2014;32(1):11–45.CrossRef Thompson IM, Kappa SF, Morgan TM, Barocas DA, Bischoff CJ, Keegan KA, Stratton KL, Clark PE, Resnick MJ, Smith JA, et al. Blood loss associated with radical cystectomy: a prospective, randomized study comparing impact LigaSure vs. stapling device. Urol Oncol: Semin Orig Investig. 2014;32(1):11–45.CrossRef
22.
Zurück zum Zitat Chen JM, Geng W, Liu FB, Zhao HC, Xie SX, Hou H, Zhao YJ, Wang GB, Geng XP. BiClamp(R) forcep liver transection versus clamp crushing technique for liver resection: study protocol for a randomized controlled trial. Trials. 2015;16:201.CrossRefPubMedPubMedCentral Chen JM, Geng W, Liu FB, Zhao HC, Xie SX, Hou H, Zhao YJ, Wang GB, Geng XP. BiClamp(R) forcep liver transection versus clamp crushing technique for liver resection: study protocol for a randomized controlled trial. Trials. 2015;16:201.CrossRefPubMedPubMedCentral
Metadaten
Titel
BiClamp® vessel-sealing device for open hepatic resection of malignant and benign liver tumours: a single-institution experience
verfasst von
Yi-jun Zhao
Da-chen Zhou
Fu-bao Liu
Hong-chuan Zhao
Guo-bin Wang
Xiao-ping Geng
Publikationsdatum
01.12.2017
Verlag
BioMed Central
Erschienen in
BMC Cancer / Ausgabe 1/2017
Elektronische ISSN: 1471-2407
DOI
https://doi.org/10.1186/s12885-017-3513-0

Weitere Artikel der Ausgabe 1/2017

BMC Cancer 1/2017 Zur Ausgabe

Adjuvante Immuntherapie verlängert Leben bei RCC

25.04.2024 Nierenkarzinom Nachrichten

Nun gibt es auch Resultate zum Gesamtüberleben: Eine adjuvante Pembrolizumab-Therapie konnte in einer Phase-3-Studie das Leben von Menschen mit Nierenzellkarzinom deutlich verlängern. Die Sterberate war im Vergleich zu Placebo um 38% geringer.

Alectinib verbessert krankheitsfreies Überleben bei ALK-positivem NSCLC

25.04.2024 NSCLC Nachrichten

Das Risiko für Rezidiv oder Tod von Patienten und Patientinnen mit reseziertem ALK-positivem NSCLC ist unter einer adjuvanten Therapie mit dem Tyrosinkinase-Inhibitor Alectinib signifikant geringer als unter platinbasierter Chemotherapie.

Bei Senioren mit Prostatakarzinom auf Anämie achten!

24.04.2024 DGIM 2024 Nachrichten

Patienten, die zur Behandlung ihres Prostatakarzinoms eine Androgendeprivationstherapie erhalten, entwickeln nicht selten eine Anämie. Wer ältere Patienten internistisch mitbetreut, sollte auf diese Nebenwirkung achten.

ICI-Therapie in der Schwangerschaft wird gut toleriert

Müssen sich Schwangere einer Krebstherapie unterziehen, rufen Immuncheckpointinhibitoren offenbar nicht mehr unerwünschte Wirkungen hervor als andere Mittel gegen Krebs.

Update Onkologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.