Biomechanical testing of a new plate system for the distal humerus compared to two well-established implants

verfasst von: Christine Voigt, Christina Rank, Klaus Waizner, Robert Wendlandt, Andreas Unger, Arndt P Schulz, Christian Jürgens, Felix Renken

Erschienen in: International Orthopaedics | Ausgabe 4/2013

Einloggen, um Zugang zu erhalten

Abstract

Purpose

A biomechanical study was performed to test the hypothesis that a new anatomically preformed, thinner, soft-tissue protecting plate system for distal humeral fractures (Tifix®-hybridplate [HP]) would show comparable results in the quasi-static and dynamic testings compared to two conventional implants: The 3.5-mm reconstruction plate (RP) providing primary stability with normal bone mineral density (BMD), and a multidirectional locking plate (Tifix®-plate [P]) which can be used with poor bone quality.

Methods

The Tifix®-HP was developed by the working group. The biomechanical testing was performed on a C2-fracture-model in 24 synthetic humeri. Three groups, each with eight bone-implant-constructs, were analysed in quasi-static and dynamic tests.

Results

The quasi-static measurements showed that under extension loading both locking plates (Tifix®-P, Tifix®-HP) were significantly stiffer than the reconstruction plate, and that the Tifix®-HP had a significantly lower stiffness than the two other implants under flexion loading. In the dynamic tests the Tifix®-P allowed significantly less fracture motion compared to the Tifix®-HP and the reconstruction plate. In an osteopaenic bone model locking plates failed only under much higher dynamic force than the reconstruction plate. The reconstruction plate and the Tifix®-P always failed through screw loosening, whereas the newly developed Tifix®-HP showed screw loosening in only one third of cases.

Conclusion

The hypothesis that the newly designed plate system showed comparable results in the quasi-static and dynamic tests compared to the conventional implants with a significantly lower implant volume and thickness was confirmed.
Literatur
1.
Zurück zum Zitat Palvanen M, Kannus P, Parkkari J, Pitkäjärvi T, Pasanen M, Vuori I, Järvinen M (2000) The injury mechanisms of osteoporotic upper extremity fractures among older adults: a controlled study of 287 consecutive patients and their 108 controls. Osteoporos Int 11:822–831PubMedCrossRef Palvanen M, Kannus P, Parkkari J, Pitkäjärvi T, Pasanen M, Vuori I, Järvinen M (2000) The injury mechanisms of osteoporotic upper extremity fractures among older adults: a controlled study of 287 consecutive patients and their 108 controls. Osteoporos Int 11:822–831PubMedCrossRef
2.
Zurück zum Zitat Popovic D, King GJW (2012) Fragility fractures of the distal humerus: what is the optimal treatment? J Bone Joint Surg Br 94(1):16–22PubMedCrossRef Popovic D, King GJW (2012) Fragility fractures of the distal humerus: what is the optimal treatment? J Bone Joint Surg Br 94(1):16–22PubMedCrossRef
3.
Zurück zum Zitat Jupiter JB, Morrey BF (1993) Fractures of the distal humerus in the adult. In: Morrey BF (ed) The elbow and its disorders, 2nd edn. Saunders, Philadelphia, London, pp 328–366 Jupiter JB, Morrey BF (1993) Fractures of the distal humerus in the adult. In: Morrey BF (ed) The elbow and its disorders, 2nd edn. Saunders, Philadelphia, London, pp 328–366
4.
Zurück zum Zitat Robinson CM, Hill RMF, Jacobs N, Dall G, Court-Brown CM (2003) Adult distal humeral metaphyseal fractures: epidemiology and results of treatment. J Orthop Trauma 17:38–47PubMedCrossRef Robinson CM, Hill RMF, Jacobs N, Dall G, Court-Brown CM (2003) Adult distal humeral metaphyseal fractures: epidemiology and results of treatment. J Orthop Trauma 17:38–47PubMedCrossRef
5.
Zurück zum Zitat Srinivasan K, Agarwal M, Matthews SJ, Giannoudis PV (2005) Fractures of the distal humerus in the elderly: is internal fixation the treatment of choice? Clin Orthop Relat Res 434:222–230PubMedCrossRef Srinivasan K, Agarwal M, Matthews SJ, Giannoudis PV (2005) Fractures of the distal humerus in the elderly: is internal fixation the treatment of choice? Clin Orthop Relat Res 434:222–230PubMedCrossRef
6.
Zurück zum Zitat Bonnaire F, Bula P (2010) Distale humeral fractures. Trauma Berufskrankh 12(2 Suppl):96–103CrossRef Bonnaire F, Bula P (2010) Distale humeral fractures. Trauma Berufskrankh 12(2 Suppl):96–103CrossRef
7.
Zurück zum Zitat Korner J, Diederichs G, Arzdorf M, Lill H, Josten C, Schneider E, Linke B (2004) A biomechanical evaluation of methods of distal humerus fracture fixation using locking compression plates versus conventional reconstruction plates. J Orthop Trauma 18:286–293PubMedCrossRef Korner J, Diederichs G, Arzdorf M, Lill H, Josten C, Schneider E, Linke B (2004) A biomechanical evaluation of methods of distal humerus fracture fixation using locking compression plates versus conventional reconstruction plates. J Orthop Trauma 18:286–293PubMedCrossRef
8.
Zurück zum Zitat Windolf M, Maza ER, Gueorguiev B, Braunstein V, Schwieger K (2010) Treatment of distal humeral fractures using conventional implants. Biomechanical evaluation of a new implant configuration. BMC Musculoskelet Disord 11:172PubMedCrossRef Windolf M, Maza ER, Gueorguiev B, Braunstein V, Schwieger K (2010) Treatment of distal humeral fractures using conventional implants. Biomechanical evaluation of a new implant configuration. BMC Musculoskelet Disord 11:172PubMedCrossRef
9.
Zurück zum Zitat Kollias CM, Darcy SP, Reed JG, Rosvold JM, Shrive NG, Hildebrand KA (2010) Distal humerus internal fixation: a biomechanical comparison of 90° and parallel constructs. Am J Orthop (Belle Mead NJ): 39(9):440–444 Kollias CM, Darcy SP, Reed JG, Rosvold JM, Shrive NG, Hildebrand KA (2010) Distal humerus internal fixation: a biomechanical comparison of 90° and parallel constructs. Am J Orthop (Belle Mead NJ): 39(9):440–444
10.
Zurück zum Zitat Penzkofer R, Hungerer S, Wipf F, von Oldenburg G, Augat P (2010) Anatomical plate configuration affects mechanical performance in distal humerus fractures. Clin Biomech (Bristol, Avon) 25(10):972–978CrossRef Penzkofer R, Hungerer S, Wipf F, von Oldenburg G, Augat P (2010) Anatomical plate configuration affects mechanical performance in distal humerus fractures. Clin Biomech (Bristol, Avon) 25(10):972–978CrossRef
11.
Zurück zum Zitat Schwartz A, Oka R, Odell T, Mahar A (2006) Biomechanical comparison of two different periarticular plating systems for stabilization of complex distal humerus fractures. Clin Biomech 21:950–955CrossRef Schwartz A, Oka R, Odell T, Mahar A (2006) Biomechanical comparison of two different periarticular plating systems for stabilization of complex distal humerus fractures. Clin Biomech 21:950–955CrossRef
12.
Zurück zum Zitat Shin SJ, Sohn HS, Do NH (2010) A clinical comparison of two different double plating methods for intraarticular distal humerus fractures. J Should Elb Surg 19:2–9CrossRef Shin SJ, Sohn HS, Do NH (2010) A clinical comparison of two different double plating methods for intraarticular distal humerus fractures. J Should Elb Surg 19:2–9CrossRef
13.
Zurück zum Zitat Frattini M, Soncini G, Corradi M, Panno B, Tocco S, Pogliacomi F (2011) Mid-term results of complex distal humeral fractures. Musculoskelet Surg 95(3):205–213PubMedCrossRef Frattini M, Soncini G, Corradi M, Panno B, Tocco S, Pogliacomi F (2011) Mid-term results of complex distal humeral fractures. Musculoskelet Surg 95(3):205–213PubMedCrossRef
14.
Zurück zum Zitat Athwal GS, Hoxie SC, Rispoli DM, Steinmann SP (2009) Precontoured parallel plate fixation of AO/OTA type C distal humerus fractures. J Orthop Trauma 23:575–580PubMedCrossRef Athwal GS, Hoxie SC, Rispoli DM, Steinmann SP (2009) Precontoured parallel plate fixation of AO/OTA type C distal humerus fractures. J Orthop Trauma 23:575–580PubMedCrossRef
15.
Zurück zum Zitat Dunlap JT, Chong AC, Lucas GL, Cooke FW (2008) Structural properties of a novel design of composite analogue humeri models. Ann Biomed Eng 36(11):1922–1926PubMedCrossRef Dunlap JT, Chong AC, Lucas GL, Cooke FW (2008) Structural properties of a novel design of composite analogue humeri models. Ann Biomed Eng 36(11):1922–1926PubMedCrossRef
16.
Zurück zum Zitat Schuster I (2005) Development of a method to test osteosyntheses in intraarticular distal humeral fractures with a metaphyseal comminution zone in decreased bone mineral density. Dissertation, University of Freiburg, Germany Schuster I (2005) Development of a method to test osteosyntheses in intraarticular distal humeral fractures with a metaphyseal comminution zone in decreased bone mineral density. Dissertation, University of Freiburg, Germany
17.
Zurück zum Zitat Morrey BF (1985) The elbow and its disorders. WB Saunders Co, Philadelphia, PA Morrey BF (1985) The elbow and its disorders. WB Saunders Co, Philadelphia, PA
18.
Zurück zum Zitat Claes L, Wolf S, Augat P (2000) Mechanical modification of callus healing. Chirurg 71(9):989–994PubMedCrossRef Claes L, Wolf S, Augat P (2000) Mechanical modification of callus healing. Chirurg 71(9):989–994PubMedCrossRef
19.
Zurück zum Zitat Schuster I, Korner J, Arzdorf M, Schwieger K, Diederichs G, Linke B (2008) Mechanical comparison in cadaver specimens of three different 90-degree double-plate osteosyntheses for simulated C2-type distal humerus fractures with varying bone densities. J Orthop Trauma 22(2):113–120PubMedCrossRef Schuster I, Korner J, Arzdorf M, Schwieger K, Diederichs G, Linke B (2008) Mechanical comparison in cadaver specimens of three different 90-degree double-plate osteosyntheses for simulated C2-type distal humerus fractures with varying bone densities. J Orthop Trauma 22(2):113–120PubMedCrossRef
20.
Zurück zum Zitat Siebenlist S, Stöckle U, Lucke M (2009) Problems of osteoporotic fractures of the elbow. Obere Extremität 4:160–167CrossRef Siebenlist S, Stöckle U, Lucke M (2009) Problems of osteoporotic fractures of the elbow. Obere Extremität 4:160–167CrossRef
21.
Zurück zum Zitat Nicol A, Berme N, Paul J (1977) A biomechanical analysis of elbow joint function. Joint replacement in the upper limb. Institution of Mechanical Engineers, London, pp 45–51 Nicol A, Berme N, Paul J (1977) A biomechanical analysis of elbow joint function. Joint replacement in the upper limb. Institution of Mechanical Engineers, London, pp 45–51
22.
Zurück zum Zitat Nordin M (2001) Biomechanics of tendons and ligaments. In: Nordin M, Lorenz T (eds) Basic biomechanics of the musculoskeletal system, 3rd edn. Lippincott Williams & Wilkins, Philadelphia, pp 350–356 Nordin M (2001) Biomechanics of tendons and ligaments. In: Nordin M, Lorenz T (eds) Basic biomechanics of the musculoskeletal system, 3rd edn. Lippincott Williams & Wilkins, Philadelphia, pp 350–356
23.
Zurück zum Zitat Rübberdt A, Surke C, Fuchs T, Frerichmann U, Matuszewski L, Vieth V, Raschke MJ (2008) Preformed plate-fixation system for type AO 13C3 distal humerus fractures: clinical experiences and treatment results taking access into account. Unfallchirurg 111(5): 308–322 [Article in German]PubMedCrossRef Rübberdt A, Surke C, Fuchs T, Frerichmann U, Matuszewski L, Vieth V, Raschke MJ (2008) Preformed plate-fixation system for type AO 13C3 distal humerus fractures: clinical experiences and treatment results taking access into account. Unfallchirurg 111(5): 308–322 [Article in German]PubMedCrossRef
Metadaten
Titel
Biomechanical testing of a new plate system for the distal humerus compared to two well-established implants
verfasst von
Christine Voigt
Christina Rank
Klaus Waizner
Robert Wendlandt
Andreas Unger
Arndt P Schulz
Christian Jürgens
Felix Renken
Publikationsdatum
01.04.2013
Verlag
Springer-Verlag
Erschienen in
International Orthopaedics / Ausgabe 4/2013
Print ISSN: 0341-2695
Elektronische ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-013-1779-7

Weitere Artikel der Ausgabe 4/2013

International Orthopaedics 4/2013 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.