Abstract

Purpose

The aim of the study was to evaluate the effects of imageless computer-assisted surgery (CAS) on the accuracy of positioning of the femoral component and on the short-term clinical outcome in hip resurfacing (HR) using a randomised prospective design.

Methods

A total of 75 consecutive patients undergoing HR were randomly allocated to CAS and conventional implantation, respectively. Preoperatively and six months post-operatively standardised pelvic anteroposterior X-ray images, the total Western Ontario and McMaster Universities Osteoarthritis Index, the Harris Hip Score and the EQ-5D utility index were evaluated in a blinded manner. The primary end point of the study was a post-operative femoral component malpositioning in five degrees or more either varus or valgus absolute deviation from the planned stem shaft angle.

Results

Patient demographics and algofunctional scores did not differ between the CAS and conventional implantation samples. Using CAS fewer femoral components were positioned in five or more degrees absolute deviation (4/37 vs 12/38, Fisher’s exact p = 0.047; 95 % confidence interval for the primary end point’s incidence difference: +3 %; +39 %); the respective incidences of five or more degrees of varus deviation were 0/37 vs 5/38. One conversion to a stemmed prosthesis (CAS group) was performed for periprosthetic femoral neck fracture. Radiological signs of superolateral femoral neck/implant impingement were observed in two cases (one CAS-based and one conventional implantation).

Conclusions

The accuracy of femoral HR component positioning was significantly improved using CAS. However, one major complication necessitated early revision in the CAS group at six months of observation. Apart from that adverse event no inter-group differences were observed for the short-term clinical outcome. Future studies need to address the clinical long-term relevance of CAS in HR.
Literatur
1.
Zurück zum Zitat Prosser GH, Yates PJ, Wood DJ, Graves SE, de Steiger RN, Miller LN (2010) Outcome of primary resurfacing hip replacement: evaluation of risk factors for early revision. Acta Orthop 81(1):66–71PubMedCrossRef Prosser GH, Yates PJ, Wood DJ, Graves SE, de Steiger RN, Miller LN (2010) Outcome of primary resurfacing hip replacement: evaluation of risk factors for early revision. Acta Orthop 81(1):66–71PubMedCrossRef
2.
Zurück zum Zitat Corten K, MacDonald SJ (2010) Hip resurfacing data from national joint registries: what do they tell us? What do they not tell us? Clin Orthop Relat Res 468(2):351–357PubMedCrossRef Corten K, MacDonald SJ (2010) Hip resurfacing data from national joint registries: what do they tell us? What do they not tell us? Clin Orthop Relat Res 468(2):351–357PubMedCrossRef
3.
Zurück zum Zitat Amstutz HC, Le Duff MJ, Campbell PA, Gruen TA, Wisk LE (2010) Clinical and radiographic results of metal-on-metal hip resurfacing with a minimum ten-year follow-up. J Bone Joint Surg Am 92(16):2663–2671PubMedCrossRef Amstutz HC, Le Duff MJ, Campbell PA, Gruen TA, Wisk LE (2010) Clinical and radiographic results of metal-on-metal hip resurfacing with a minimum ten-year follow-up. J Bone Joint Surg Am 92(16):2663–2671PubMedCrossRef
4.
Zurück zum Zitat Amstutz HC, Campbell PA, Le Duff MJ (2004) Fracture of the neck of the femur after surface arthroplasty of the hip. J Bone Joint Surg Am 86-A(9):1874–1877PubMed Amstutz HC, Campbell PA, Le Duff MJ (2004) Fracture of the neck of the femur after surface arthroplasty of the hip. J Bone Joint Surg Am 86-A(9):1874–1877PubMed
5.
Zurück zum Zitat Kim PR, Beaulé PE, Laflamme GY, Dunbar M (2008) Causes of early failure in a multicenter clinical trial of hip resurfacing. J Arthroplasty 23(6 Suppl 1):44–49 Kim PR, Beaulé PE, Laflamme GY, Dunbar M (2008) Causes of early failure in a multicenter clinical trial of hip resurfacing. J Arthroplasty 23(6 Suppl 1):44–49
6.
Zurück zum Zitat Vail TP, Glisson RR, Dominguez DE, Kitaoka K, Ottaviano D (2008) Position of hip resurfacing component affects strain and resistance to fracture in the femoral neck. J Bone Joint Surg Am 90(9):1951–1960PubMedCrossRef Vail TP, Glisson RR, Dominguez DE, Kitaoka K, Ottaviano D (2008) Position of hip resurfacing component affects strain and resistance to fracture in the femoral neck. J Bone Joint Surg Am 90(9):1951–1960PubMedCrossRef
7.
Zurück zum Zitat Ganapathi M, Evans S, Roberts P (2008) Strain pattern following surface replacement of the hip. Proc Inst Mech Eng H 222(1):13–18PubMed Ganapathi M, Evans S, Roberts P (2008) Strain pattern following surface replacement of the hip. Proc Inst Mech Eng H 222(1):13–18PubMed
8.
Zurück zum Zitat Shimmin AJ, Back D (2005) Femoral neck fractures following Birmingham hip resurfacing: a national review of 50 cases. J Bone Joint Surg Br 87(4):463–464PubMedCrossRef Shimmin AJ, Back D (2005) Femoral neck fractures following Birmingham hip resurfacing: a national review of 50 cases. J Bone Joint Surg Br 87(4):463–464PubMedCrossRef
9.
Zurück zum Zitat Freeman MA (1978) Some anatomical and mechanical considerations relevant to the surface replacement of the femoral head. Clin Orthop Relat Res (134):19–24 Freeman MA (1978) Some anatomical and mechanical considerations relevant to the surface replacement of the femoral head. Clin Orthop Relat Res (134):19–24
10.
Zurück zum Zitat Beaulé PE, Lee JL, Le Duff MJ, Amstutz HC, Ebramzadeh E (2004) Orientation of the femoral component in surface arthroplasty of the hip. A biomechanical and clinical analysis. J Bone Joint Surg Am 86-A(9):2015–2021PubMed Beaulé PE, Lee JL, Le Duff MJ, Amstutz HC, Ebramzadeh E (2004) Orientation of the femoral component in surface arthroplasty of the hip. A biomechanical and clinical analysis. J Bone Joint Surg Am 86-A(9):2015–2021PubMed
11.
Zurück zum Zitat Little CP, Ruiz AL, Harding IJ, McLardy-Smith P, Gundle R, Murray DW et al (2005) Osteonecrosis in retrieved femoral heads after failed resurfacing arthroplasty of the hip. J Bone Joint Surg Br 87(3):320–323PubMedCrossRef Little CP, Ruiz AL, Harding IJ, McLardy-Smith P, Gundle R, Murray DW et al (2005) Osteonecrosis in retrieved femoral heads after failed resurfacing arthroplasty of the hip. J Bone Joint Surg Br 87(3):320–323PubMedCrossRef
12.
Zurück zum Zitat Langton DJ, Joyce TJ, Jameson SS, Lord J, Van Orsouw M, Holland JP et al (2011) Adverse reaction to metal debris following hip resurfacing: the influence of component type, orientation and volumetric wear. J Bone Joint Surg Br 93(2):164–171PubMedCrossRef Langton DJ, Joyce TJ, Jameson SS, Lord J, Van Orsouw M, Holland JP et al (2011) Adverse reaction to metal debris following hip resurfacing: the influence of component type, orientation and volumetric wear. J Bone Joint Surg Br 93(2):164–171PubMedCrossRef
13.
Zurück zum Zitat Hart AJ, Skinner JA, Henckel J, Sampson B, Gordon F (2011) Insufficient acetabular version increases blood metal ion levels after metal-on-metal hip resurfacing. Clin Orthop Relat Res 469(9):2590–2597PubMedCrossRef Hart AJ, Skinner JA, Henckel J, Sampson B, Gordon F (2011) Insufficient acetabular version increases blood metal ion levels after metal-on-metal hip resurfacing. Clin Orthop Relat Res 469(9):2590–2597PubMedCrossRef
14.
Zurück zum Zitat Brin YS, Nikolaou VS, Joseph L, Zukor DJ, Antoniou J (2011) Imageless computer assisted versus conventional total knee replacement. A Bayesian meta-analysis of 23 comparative studies. Int Orthop 35(3):331–339PubMedCrossRef Brin YS, Nikolaou VS, Joseph L, Zukor DJ, Antoniou J (2011) Imageless computer assisted versus conventional total knee replacement. A Bayesian meta-analysis of 23 comparative studies. Int Orthop 35(3):331–339PubMedCrossRef
15.
Zurück zum Zitat Leenders T, Vandevelde D, Mahieu G, Nuyts R (2002) Reduction in variability of acetabular cup abduction using computer assisted surgery: a prospective and randomized study. Comput Aided Surg 7(2):99–106PubMedCrossRef Leenders T, Vandevelde D, Mahieu G, Nuyts R (2002) Reduction in variability of acetabular cup abduction using computer assisted surgery: a prospective and randomized study. Comput Aided Surg 7(2):99–106PubMedCrossRef
16.
Zurück zum Zitat Ulrich SD, Bonutti PM, Seyler TM, Marker DR, Jones LC, Mont MA (2007) Outcomes-based evaluations supporting computer-assisted surgery and minimally invasive surgery for total hip arthroplasty. Expert Rev Med Devices 4(6):873–883PubMedCrossRef Ulrich SD, Bonutti PM, Seyler TM, Marker DR, Jones LC, Mont MA (2007) Outcomes-based evaluations supporting computer-assisted surgery and minimally invasive surgery for total hip arthroplasty. Expert Rev Med Devices 4(6):873–883PubMedCrossRef
17.
Zurück zum Zitat Ganapathi M, Vendittoli PA, Lavigne M, Günther KP (2009) Femoral component positioning in hip resurfacing with and without navigation. Clin Orthop Relat Res 467(5):1341–1347PubMedCrossRef Ganapathi M, Vendittoli PA, Lavigne M, Günther KP (2009) Femoral component positioning in hip resurfacing with and without navigation. Clin Orthop Relat Res 467(5):1341–1347PubMedCrossRef
18.
Zurück zum Zitat Brooker AF, Bowerman JW, Robinson RA, Riley LH Jr (1973) Ectopic ossification following total hip replacement. Incidence and a method of classification. J Bone Joint Surg Am 55(8):1629–1632PubMed Brooker AF, Bowerman JW, Robinson RA, Riley LH Jr (1973) Ectopic ossification following total hip replacement. Incidence and a method of classification. J Bone Joint Surg Am 55(8):1629–1632PubMed
19.
Zurück zum Zitat Stucki G, Meier D, Stucki S, Michel BA, Tyndall AG, Dick W et al (1996) Evaluation of a German version of WOMAC (Western Ontario and McMaster Universities) Arthrosis Index. Z Rheumatol 55(1):40–49PubMed Stucki G, Meier D, Stucki S, Michel BA, Tyndall AG, Dick W et al (1996) Evaluation of a German version of WOMAC (Western Ontario and McMaster Universities) Arthrosis Index. Z Rheumatol 55(1):40–49PubMed
20.
Zurück zum Zitat Bellamy N, Buchanan WW, Goldsmith CH, Campbell J, Stitt LW (1988) Validation study of WOMAC: a health status instrument for measuring clinically important patient relevant outcomes to antirheumatic drug therapy in patients with osteoarthritis of the hip or knee. J Rheumatol 15(12):1833–1840PubMed Bellamy N, Buchanan WW, Goldsmith CH, Campbell J, Stitt LW (1988) Validation study of WOMAC: a health status instrument for measuring clinically important patient relevant outcomes to antirheumatic drug therapy in patients with osteoarthritis of the hip or knee. J Rheumatol 15(12):1833–1840PubMed
21.
Zurück zum Zitat Beaton DE, Schemitsch E (2003) Measures of health-related quality of life and physical function. Clin Orthop Relat Res (413):90–105 Beaton DE, Schemitsch E (2003) Measures of health-related quality of life and physical function. Clin Orthop Relat Res (413):90–105
22.
Zurück zum Zitat Ethgen O, Bruyère O, Richy F, Dardennes C, Reginster JY (2004) Health-related quality of life in total hip and total knee arthroplasty. A qualitative and systematic review of the literature. J Bone Joint Surg Am 86-A(5):963–974PubMed Ethgen O, Bruyère O, Richy F, Dardennes C, Reginster JY (2004) Health-related quality of life in total hip and total knee arthroplasty. A qualitative and systematic review of the literature. J Bone Joint Surg Am 86-A(5):963–974PubMed
23.
Zurück zum Zitat Sun Y, Stürmer T, Günther KP, Brenner H (1997) Reliability and validity of clinical outcome measurements of osteoarthritis of the hip and knee–a review of the literature. Clin Rheumatol 16(2):185–198PubMedCrossRef Sun Y, Stürmer T, Günther KP, Brenner H (1997) Reliability and validity of clinical outcome measurements of osteoarthritis of the hip and knee–a review of the literature. Clin Rheumatol 16(2):185–198PubMedCrossRef
24.
Zurück zum Zitat March L, Cross M, Tribe K, Lapsley H, Courtenay B, Brooks P (2002) Cost of joint replacement surgery for osteoarthritis: the patients’ perspective. J Rheumatol 29(5):1006–1014PubMed March L, Cross M, Tribe K, Lapsley H, Courtenay B, Brooks P (2002) Cost of joint replacement surgery for osteoarthritis: the patients’ perspective. J Rheumatol 29(5):1006–1014PubMed
25.
Zurück zum Zitat Bachmeier CJ, March LM, Cross MJ, Lapsley HM, Tribe KL, Courtenay BG et al (2001) A comparison of outcomes in osteoarthritis patients undergoing total hip and knee replacement surgery. Osteoarthritis Cartilage 9(2):137–146PubMedCrossRef Bachmeier CJ, March LM, Cross MJ, Lapsley HM, Tribe KL, Courtenay BG et al (2001) A comparison of outcomes in osteoarthritis patients undergoing total hip and knee replacement surgery. Osteoarthritis Cartilage 9(2):137–146PubMedCrossRef
26.
Zurück zum Zitat Harris WH (1969) Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg Am 51(4):737–755PubMed Harris WH (1969) Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg Am 51(4):737–755PubMed
27.
Zurück zum Zitat Saklad M (1941) Grading of patients for surgical procedures. Anesthesiology 2:281–284CrossRef Saklad M (1941) Grading of patients for surgical procedures. Anesthesiology 2:281–284CrossRef
28.
Zurück zum Zitat Amstutz HC, Beaulé PE, Dorey FJ, Le Duff MJ, Campbell PA, Gruen TA (2004) Metal-on-metal hybrid surface arthroplasty: two to six-year follow-up study. J Bone Joint Surg Am 86-A(1):28–39PubMed Amstutz HC, Beaulé PE, Dorey FJ, Le Duff MJ, Campbell PA, Gruen TA (2004) Metal-on-metal hybrid surface arthroplasty: two to six-year follow-up study. J Bone Joint Surg Am 86-A(1):28–39PubMed
29.
Zurück zum Zitat Beaulé PE, Dorey FJ, LeDuff M, Gruen T, Amstutz HC (2004) Risk factors affecting outcome of metal-on-metal surface arthroplasty of the hip. Clin Orthop Relat Res (418):87–93 Beaulé PE, Dorey FJ, LeDuff M, Gruen T, Amstutz HC (2004) Risk factors affecting outcome of metal-on-metal surface arthroplasty of the hip. Clin Orthop Relat Res (418):87–93
30.
Zurück zum Zitat Anglin C, Masri BA, Tonetti J, Hodgson AJ, Greidanus NV (2007) Hip resurfacing femoral neck fracture influenced by valgus placement. Clin Orthop Relat Res 465:71–79PubMed Anglin C, Masri BA, Tonetti J, Hodgson AJ, Greidanus NV (2007) Hip resurfacing femoral neck fracture influenced by valgus placement. Clin Orthop Relat Res 465:71–79PubMed
31.
Zurück zum Zitat Gautier E, Ganz K, Krügel N, Gill T, Ganz R (2000) Anatomy of the medial femoral circumflex artery and its surgical implications. J Bone Joint Surg Br 82(5):679–683PubMedCrossRef Gautier E, Ganz K, Krügel N, Gill T, Ganz R (2000) Anatomy of the medial femoral circumflex artery and its surgical implications. J Bone Joint Surg Br 82(5):679–683PubMedCrossRef
32.
Zurück zum Zitat Beaulé PE, Campbell PA, Hoke R, Dorey F (2006) Notching of the femoral neck during resurfacing arthroplasty of the hip: a vascular study. J Bone Joint Surg Br 88(1):35–39PubMed Beaulé PE, Campbell PA, Hoke R, Dorey F (2006) Notching of the femoral neck during resurfacing arthroplasty of the hip: a vascular study. J Bone Joint Surg Br 88(1):35–39PubMed
33.
Zurück zum Zitat Beaulé PE, Amstutz HC (2005) Orientation of the femoral component in surface arthroplasty of the hip. J Bone Joint Surg Am 87(5):1162PubMed Beaulé PE, Amstutz HC (2005) Orientation of the femoral component in surface arthroplasty of the hip. J Bone Joint Surg Am 87(5):1162PubMed
34.
Zurück zum Zitat McMinn D, Daniel J (2006) History and modern concepts in surface replacement. Proc Inst Mech Eng H 220(2):239–251PubMedCrossRef McMinn D, Daniel J (2006) History and modern concepts in surface replacement. Proc Inst Mech Eng H 220(2):239–251PubMedCrossRef
35.
Zurück zum Zitat Davis ET, Olsen M, Zdero R, Waddell JP, Schemitsch EH (2008) Femoral neck fracture following hip resurfacing: the effect of alignment of the femoral component. J Bone Joint Surg Br 90(11):1522–1527PubMedCrossRef Davis ET, Olsen M, Zdero R, Waddell JP, Schemitsch EH (2008) Femoral neck fracture following hip resurfacing: the effect of alignment of the femoral component. J Bone Joint Surg Br 90(11):1522–1527PubMedCrossRef
36.
Zurück zum Zitat Seyler TM, Lai LP, Sprinkle DI, Ward WG, Jinnah RH (2008) Does computer-assisted surgery improve accuracy and decrease the learning curve in hip resurfacing? A radiographic analysis. J Bone Joint Surg Am 90(Suppl 3):71–80PubMedCrossRef Seyler TM, Lai LP, Sprinkle DI, Ward WG, Jinnah RH (2008) Does computer-assisted surgery improve accuracy and decrease the learning curve in hip resurfacing? A radiographic analysis. J Bone Joint Surg Am 90(Suppl 3):71–80PubMedCrossRef
37.
Zurück zum Zitat Gallart X, Fernández-Valencia JA, Riba J, Bori G, García S, Carrillo S (2010) Alignment guided by computer navigation of the femoral component in hip resurfacing. Hip Int 20(Suppl 7 S7):43–47 Gallart X, Fernández-Valencia JA, Riba J, Bori G, García S, Carrillo S (2010) Alignment guided by computer navigation of the femoral component in hip resurfacing. Hip Int 20(Suppl 7 S7):43–47
38.
Zurück zum Zitat Hart R, Sváb P, Filan P (2008) Intraoperative navigation in hip surface arthroplasty: a radiographic comparative analysis study. Arch Orthop Trauma Surg 128(4):429–434PubMedCrossRef Hart R, Sváb P, Filan P (2008) Intraoperative navigation in hip surface arthroplasty: a radiographic comparative analysis study. Arch Orthop Trauma Surg 128(4):429–434PubMedCrossRef
39.
Zurück zum Zitat Schnurr C, Nessler J, Koebke J, Michael JW, Eysel P, König DP (2010) Imageless computer navigation of hip resurfacing arthroplasty. Oper Orthop Traumatol 22(3):307–316PubMedCrossRef Schnurr C, Nessler J, Koebke J, Michael JW, Eysel P, König DP (2010) Imageless computer navigation of hip resurfacing arthroplasty. Oper Orthop Traumatol 22(3):307–316PubMedCrossRef
40.
Zurück zum Zitat Krüger S, Zambelli PY, Leyvraz PF, Jolles BM (2009) Computer-assisted placement technique in hip resurfacing arthroplasty: improvement in accuracy? Int Orthop 33(1):27–33PubMedCrossRef Krüger S, Zambelli PY, Leyvraz PF, Jolles BM (2009) Computer-assisted placement technique in hip resurfacing arthroplasty: improvement in accuracy? Int Orthop 33(1):27–33PubMedCrossRef
41.
Zurück zum Zitat Langton DJ, Jameson SS, Joyce TJ, Hallab NJ, Natu S, Nargol AV (2010) Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: a consequence of excess wear. J Bone Joint Surg Br 92(1):38–46PubMedCrossRef Langton DJ, Jameson SS, Joyce TJ, Hallab NJ, Natu S, Nargol AV (2010) Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: a consequence of excess wear. J Bone Joint Surg Br 92(1):38–46PubMedCrossRef
42.
Zurück zum Zitat Hartmann A, Lützner J, Kirschner S, Witzleb WC, Günther KP (2012) Do survival rate and serum ion concentrations 10 years after metal-on-metal hip resurfacing provide evidence for continued use? Clin Orthop Relat Res 470:3118–3126 Hartmann A, Lützner J, Kirschner S, Witzleb WC, Günther KP (2012) Do survival rate and serum ion concentrations 10 years after metal-on-metal hip resurfacing provide evidence for continued use? Clin Orthop Relat Res 470:3118–3126
43.
Zurück zum Zitat Leslie IJ, Williams S, Isaac G, Ingham E, Fisher J (2009) High cup angle and microseparation increase the wear of hip surface replacements. Clin Orthop Relat Res 467(9):2259–2265PubMedCrossRef Leslie IJ, Williams S, Isaac G, Ingham E, Fisher J (2009) High cup angle and microseparation increase the wear of hip surface replacements. Clin Orthop Relat Res 467(9):2259–2265PubMedCrossRef
44.
Zurück zum Zitat De Haan R, Pattyn C, Gill HS, Murray DW, Campbell PA, De Smet K (2008) Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement. J Bone Joint Surg Br 90(10):1291–1297PubMedCrossRef De Haan R, Pattyn C, Gill HS, Murray DW, Campbell PA, De Smet K (2008) Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement. J Bone Joint Surg Br 90(10):1291–1297PubMedCrossRef
45.
Zurück zum Zitat Ybinger T, Kumpan W, Hoffart HE, Muschalik B, Bullmann W, Zweymüller K (2007) Accuracy of navigation-assisted acetabular component positioning studied by computed tomography measurements: methods and results. J Arthroplasty 22(6):812–817PubMedCrossRef Ybinger T, Kumpan W, Hoffart HE, Muschalik B, Bullmann W, Zweymüller K (2007) Accuracy of navigation-assisted acetabular component positioning studied by computed tomography measurements: methods and results. J Arthroplasty 22(6):812–817PubMedCrossRef
46.
Zurück zum Zitat Marker DR, Seyler TM, Jinnah RH, Delanois RE, Ulrich SD, Mont MA (2007) Femoral neck fractures after metal-on-metal total hip resurfacing: a prospective cohort study. J Arthroplasty 22(7 Suppl 3):66–71PubMedCrossRef Marker DR, Seyler TM, Jinnah RH, Delanois RE, Ulrich SD, Mont MA (2007) Femoral neck fractures after metal-on-metal total hip resurfacing: a prospective cohort study. J Arthroplasty 22(7 Suppl 3):66–71PubMedCrossRef
47.
Zurück zum Zitat Schnurr C, Michael JW, Eysel P, König DP (2009) Imageless navigation of hip resurfacing arthroplasty increases the implant accuracy. Int Orthop 33(2):365–372PubMedCrossRef Schnurr C, Michael JW, Eysel P, König DP (2009) Imageless navigation of hip resurfacing arthroplasty increases the implant accuracy. Int Orthop 33(2):365–372PubMedCrossRef
48.
Zurück zum Zitat Schnurr C, Münnich U, Eysel P, König DP (2010) Computer-assisted joint replacement surgery. Versicherungsmedizin 62(1):16–19PubMed Schnurr C, Münnich U, Eysel P, König DP (2010) Computer-assisted joint replacement surgery. Versicherungsmedizin 62(1):16–19PubMed
49.
Zurück zum Zitat Olsen M, Schemitsch EH (2011) Avoiding short-term femoral neck fracture with imageless computer navigation for hip resurfacing. Clin Orthop Relat Res 469:1621–1626 Olsen M, Schemitsch EH (2011) Avoiding short-term femoral neck fracture with imageless computer navigation for hip resurfacing. Clin Orthop Relat Res 469:1621–1626
50.
Zurück zum Zitat Resubal JR, Morgan DA (2009) Computer-assisted vs conventional mechanical jig technique in hip resurfacing arthroplasty. J Arthroplasty 24(3):341–350PubMedCrossRef Resubal JR, Morgan DA (2009) Computer-assisted vs conventional mechanical jig technique in hip resurfacing arthroplasty. J Arthroplasty 24(3):341–350PubMedCrossRef
Metadaten
Titel
The First SICOT Oral Presentation Award 2011: Imageless computer-assisted femoral component positioning in hip resurfacing: a prospective randomised trial
verfasst von
Maik Stiehler
Jens Goronzy
Albrecht Hartmann
Frank Krummenauer
Klaus-Peter Günther
Publikationsdatum
01.04.2013
Verlag
Springer-Verlag
Erschienen in
International Orthopaedics / Ausgabe 4/2013
Print ISSN: 0341-2695
Elektronische ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-012-1762-8

Weitere Artikel der Ausgabe 4/2013

International Orthopaedics 4/2013 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.