Open Access 01.06.2022 | Original Article

Body Contouring Using a Combination of Pulsed Ultrasound and Unipolar Radio Frequency: A Prospective Pilot Study

verfasst von: Fernando Urdiales-Gálvez, Sandra Martín-Sánchez, Mónica Maíz-Jiménez, Esther Viruel-Ortega

Erschienen in: Aesthetic Plastic Surgery | Ausgabe 5/2022

Abstract

Objective

To assess the efficacy and safety of a new non-invasive body contouring device in patients with localized fat in abdomen or in abdomen and hips. Additionally, we also evaluated the patient satisfaction with the procedure.

Methods

Prospective and non-randomized open label study. The patients underwent four sessions, separated by 1 week each, with the Alma PrimeX, a non-invasive body contouring device that combines pulsed non-focus ultrasound and a Unipolar radiofrequency. The primary end point was the mean change in fat tissue thickness, assessed by diagnostic ultrasound, from baseline to 3-months after the last treatment-session.

Results

Fifteen subjects were evaluated. As compared to pre-treatment thickness, Hodges-Lehmann median difference (95% CI) was − 85.3 (− 107.5 to − 62.0) mm, p = 0.0001; − 70.3 (− 95.0 to − 48.5) mm, p = 0.0001; − 100.0 (− 140.5 to − 49.5) mm, p = 0.0039; and − 71.8 (− 132.5 to − 23.0) mm, p = 0.0078 in infraumbilical, supraumbilical, right hip, and left hip, respectively. Pretreatment fat volume was significantly reduced from 32.9% to 31.2%, p = 0.0006. The median (interquartile range) degree of patient satisfaction was 4.0 (1.0–5.0), with 13 (86.7%) patients being “Highly satisfied” or “Satisfied” with the treatment results. The most common adverse event was discomfort, followed by erythema. All the adverse events were mild and were successfully resolved without treatment.

Conclusions

Combine therapy of a Pulsed non-focus ultrasound and Unipolar radiofrequency using the non-invasive device Alma PrimeX was an effective and safe treatment for reducing fat tissue thickness in abdomen and hips in patients with localized fat. Patients’ satisfaction with the procedure was high.

Level of Evidence IV

This journal requires that authors assign a level of evidence to each article. For a full description of these Evidence-Based Medicine ratings, please refer to the Table of Contents or the online Instructions to Authors www.​springer.​com/​00266.
Hinweise

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

There is an increasing pressure for looking good and healthy in today’s society. This is not only due to the ubiquitous of idealized bodies in the media, but also to a better knowledge of the detrimental effects of obesity [14].
Most of the surgical body contouring methods have been associated with several complications, including pain, swelling, prolonged recovery, scarring, hematoma or infection [5, 6]. These issues have led to the development of non-invasive techniques. Non-invasive body contouring has experienced a significant and rapid increased in recent years, mainly due to their favorable safety profile, minimal recovery time and reduced cost, while maintained a good efficacy profile [15, 7].
The main goal of non-invasive contouring devices is to improve body’s appearance by removing the excess adipose tissue, particularly in areas in which fat persists despite optimal diet and exercise routine [15].
Four main non-invasive techniques, for reducing localized subcutaneous adipose tissue, have emerged as preferred options: low-level laser therapy, cryolipolysis, radio frequency (RF) and high-intensity ultrasound, non-focused or focus (HIFU) [15]. However, up to date, no procedure has been accepted as the “gold standard”.
Radiofrequency devices are currently the most popular noninvasive body contouring devices used in practice. Although initially used for treating periorbital wrinkles, rhytids, and skin laxity [8, 9]; RF is now widely used for body contouring, skin tightening, and cellulite reduction [10]. This method is based on the difference in water and impedance between skin (low impedance) and fat tissue (high impedance) [3]. RF induces thermal injury to target tissue layers by focusing thermal energy on high impedance tissues, which induces apoptosis of subcutaneous adipose tissue cells with minimal risk of damaging other tissues, such as epidermis or muscle [11, 12].
High-intensity focused and non-focused ultrasounds had been introduced as a non-invasive body contouring strategy that focus mainly on skin tightening and rejuvenation [3]. It uses acoustic energy to induce fat tissue apoptosis. HIFU devices can deliver energy to the deep dermis, subdermal connective tissue, and fibromuscular layers in precise zones without damage to the epidermis. Besides its thermal effects, HIFU also has a mechanical effect that causes an immediate disruption of cell membranes, which favor apoptosis and necrosis [3]. Regarding non-focused ultrasound, its effect on adipose tissue seems to be related to apoptosis and cavitation, which causes a mechanical disruption of subcutaneous adipocytes [13, 14].
HIFU effects are easily perceived since there is a clear border between targeted and untargeted tissue [15]. Moreover, HIFU is thought to cause gradual skin tightening through collagen contraction and remodeling [16].
Non-focused ultrasound and electrical stimulation are therapeutic modalities commonly used in physiotherapy practice. In fact, the combination of electrical stimulation and ultrasound can be more effective than each of them used separately [17].
Alma PrimeX (Alma Lasers, Ltd. Caesarea, Israel) is a non-invasive body contouring platform that combines a unique pulsed non-focus ultrasound and unipolar radiofrequency.
Literature review revealed no evidence of previous attempt of evaluating the effect of the combination of pulsed non-focus ultrasound and unipolar radiofrequency on fat tissue reduction.
The purpose of the current study is to assess the efficacy and safety of this new non-invasive body contouring platform in patients with localized fat in abdomen or in abdomen and flanks. Additionally, we also evaluated the patient satisfaction with the procedure.

Methods

Design

This was a prospective and non-randomized open label study. The study protocol was approved by the Ethics Committee of the Instituto Médico Miramar (Málaga, Spain). Written informed consent was provided by all the participants before treatment course. The study protocol adhered to the tenets of the Declaration of Helsinki and the Good Clinical Practice/International Council for Harmonization Guidelines.

Patients

Patients with localized fat in abdomen or in abdomen and flanks were candidates for undergoing treatment for fat reduction. Eligible participants were healthy women or men aged ≥ 30 years who had clearly visible fat deposits on their abdominal region or abdominal and hip regions with a body mass index (BMI) < 30 kg/m2; proportion of fat mass < 25% (women) and < 42% (men), visceral fat < 9, or fat tissue thickness in abdominal area higher than 0.5 cm measured using a composition analyzer.
Patients were excluded if they had undergone previous surgical or non-surgical fat reduction procedures in the area of treatment; have a BMI > 30 kg/m2; have implanted electronic devices (pacemakers, insulin pumps, etc.); have presence of metal prostheses in the area to be treated; any inflammatory dermatological diseases (psoriasis, lichen planus, etc.); or have any condition that prevents or recommends not doing this treatment.

Assessments

Before treatment, all the patients underwent a quantitative and qualitative evaluation of their fat distribution, according to their individual metameric distribution.
The amount and distribution of fat was analyzed macroscopically by means 2D photos with SonyDSC-HX400V (Sony Group Corporation Konan Minato-ku, Tokyo, 108-0075 Japan) and 3D photographs with Canfield Vectra H2 (Canfield Scientific Inc; Parsippany-Troy Hills, NJ 07054, USA). Bioelectrical Impedance Analysis was performed with Tanita BF-180 (Tanita Corporation; Itabashi-Ku, Tokyo, Japan 174-8630). Fat thickness was assessed by ultrasound technology with the Samsung HS 30 ultrasound machine (Samsung Healthcare Global, Gangwon; South Korea).

Treatment Procedure

The treatment has been performed with the Alma PrimeX (Alma Lasers, Ltd. Caesarea, Israel) non-invasive body contouring device. Its main technical characteristics are summarized in Table 1.
Table 1
Main technical characteristics of the non-invasive body contouring device.
Radiofrequency (UniBody applicator)a
Ultrasound (UlatraWave applicator)b
∙ Power: Up to 300 W
∙ Ultrasonic intensity: Up to 3 W/cm2
∙ Operating frequency: 40.68 MHz
∙ Operating frequency: 64 ± 5% kHz continuous/modulated output
∙ Cooling: thermoelectric cooling
∙ Vibration frequency: 0, 10, 20 Hz
∙ Vacuum pump: Up to 0.045 MPa
∙ Control: adaptative digital frequency control
 
∙ Cooling: water cooling, air cooling
Patients received a total of four sessions separated by 1 week each
aRadiofrequency was applied only in the abdominal region in a 15 min session with an accumulative energy of 70 kJ
bUltrasound was applied in both abdominal and hip regions with a power level 8 *max intensity) and operating frequencies of 10 Hz, with high vacuum, during a session of 15 min
Ultrasound was administered in both abdominal and hip regions with a power level that ranged from 5 to 8 W and operating frequency of 10 Hz, with high vacuum, during a session of 15 min. RF treatment was administered, only in the abdominal region, in a 15-minute session with an accumulative energy of 70 kJ and a power that ranged from 140 to 200 W in sub-deep or deep planes.
Patients received a total of four sessions separated by 1 week each.
Treatment technique with the ultrasound involved an in-motion circular fashion in which according to the size of the treated abdomen would have been treated as a whole or divided into two equal segments, each treated for 7.5 mins. Using the radiofrequency an in-motion circular or brushing linear strokes where performed.

Patient Satisfaction

A five-point Likert scale (1 = highly dissatisfied; 2 = dissatisfied; 3 = neutral; 4 = satisfied; and 5 = highly satisfied) was used to assess the degree of patient satisfaction with the treatment results by referring to the following question: [1] How satisfied are you with the treatment results?

Outcomes

The primary outcome was the mean fat thickness reduction in abdominal and hip regions, assessed by diagnostic ultrasound, from pretreatment values to 3 months after the last treatment session.
Secondary outcomes include results of patient satisfaction survey, assessed 3 months after the last treatment session and safety.

Statistical Analysis

Statistical analysis was performed with MedCalc Statistical Software version 20.015 (MedCalc Software Ltd, Ostend, Belgium; https://​www.​medcalc.​org; 2021).
Continuous variables were presented as median and 95% confidence interval (95% CI) or median (Interquartile range, IqR), as needed; while categorical variables were presented as number (percentage).
The comparisons between pretreatment and posttreatment procedure measurements were performed using the Wilcoxon test, including the Hodges-Lehmann median difference and its 95% CI.
The Mann–Whitney U test was used to compare pretreatment values between subjects who underwent treatment in abdomen versus those who underwent treatment in abdomen and hip.
Analysis of covariance (ANCOVA) was used to assess the changes in Infraumbilical and supraumbilical thickness between subjects who underwent treatment in abdomen and those who did it in abdomen and hip. The model included “treated area” as factor and age, weight, BMI, visceral fat, fat mass, pretreatment infraumbilical and supraumbilical thickness as covariates.

Results

Pretreatment Demographic and Clinical Characteristics

Twenty subjects, 16 (80.0%) women and 4 (20.0%), were included in the study. Five of these did not attend follow-up visits due to the government-enacted lockdowns as a consequence of coronavirus disease 2019 (COVID-19) outbreak.
A total of 15 patients, 13 (86.7%) women, of the 20 treated attended follow-up visits. Median power of RF (IqR) was 170.0 (155.0–170.0) W administered in sub-deep (5/15 patients) or deep (10/15 patients) planes, with no differences between them (Hodges-Lehmann median difference = 20.0 W; 95% CI − 20.0–30.0 W, p = 0.2046).
The mean age was 43.5 ± 9.0 years. Six (40.0%) patients underwent treatment in abdomen and 9 (60.0%) in abdomen and hips.
Table 2 summarizes the main pretreatment parameters in the overall study sample. There were no significant differences between those subjects who underwent treatment in the abdomen or in abdomen and hips in any of the study variables.
Table 2
Overview of the main clinical and demographic characteristics of the study sample.
 
Overall (n = 15)
Abdomen (n = 6)
Abdomen & hips (n = 9)
pa
Age, years
0.9059
 Mean (SD)
43.5 (9.0)
43.7 (10.5)
43.3 (8.6)
 Median (IqR)
41.0 (35.0–51.5)
44.0 (34.0–50.0)
41.0 (35.0–52.5)
Sex, n (%)
0.4857b
 Woman
13 (86.7)
6 (100.0)
7 (77.8)
 Man
2 (13.3)
0 (0.0)
2 (22.2)
Weight, kg
0.3458
 Mean (SD)
68.4 (12.5)
63.7 (9.1)
71.5 (14.0)
 Median (IqR)
66.7 (59.0–75.3)
62.9 (55.4–71.6)
69.3 (61.1–82.3)
BMI, kg/m2
0.4094
 Mean (SD)
24.8 (3.9)
23.9 (3.3)
25.4 (4.3)
 Median (IqR)
24.4 (22.1–27.1)
22.8 (21.9–27.6)
25.1 (22.7–26.7)
Visceral fat
0.4383
 Mean (SD)
5.5 (2.6)
4.7 (3.0)
6.0 (2.4)
 Median (IqR)
5.0 (3.3–7.8)
4.0 (2.0–8.0)
5.0 (4.8–7.3)
Fat mass, %
0.8132
 Mean (SD)
31.0 (6.9)
31.5 (6.7)
30.7 (7.4)
 Median (IqR)
32.9 (26.9–36.2)
32.7 (26.4–36.2)
32.9 (26.6–35.3)
FTIU, mm
0.4094
 Mean (SD)
216.4 (86.5)
228.3 (65.1)
208.4 (101.4)
 Median (IqR)
195.0 (154.3–275.0)
220.0 (183.0–256.0)
182.0 (128.0–280.5)
FTSU, mm
0.9062
 Mean (SD)
167.4 (80.5)
180.8 (88.9)
158.4 (78.8)
 Median (IqR)
174.0 (105.0–217.0)
168.5 (119.0–176.0)
179.0 (81.0–230.3)
FTRH
N.A.
 Mean (SD)
195.6 (100.9)
N.A.
195.6 (100.9)
 Median (IqR)
189.0 (112.8–289.3)
 
189.0 (112.8–289.3)
FTLHc
N.A.
 Mean (SD)
178.6 (95.7)
N.A.
178.6 (9.5.7)
 Median (IqR)
182.0 (90.5–264.5)
 
182.0 (90.5–264.5)
SD Standard deviation, IqR Interquartile range, BMI Body mass index, FTIU Fat thickness infraumbilical region, FTSU Fat thickness supraumbilical region, FTRH Fat thickness right hip, FTLH Fat thickness left hip, N.A. Not applicable.
aMann-Whitney U test
bFisher exact test
cEight subjects

Fat Thickness Reduction

As compared to pre-treatment values, fat thickness in the infraumbilical region was significantly reduced from 195.0 mm (154.3–275.0 mm) to 118.0 mm (6.0–155.0 mm), p = 0.0001. Pre-treatment fat thickness in the supraumbilical region was significantly reduced from 174.0 mm (95% CI 105.0–217.0 mm) to 81.0 mm (95% CI 56.8–123.5 mm), p = 0.0001. Similarly, pre-treatment fat thickness in the right hip was significantly reduced from 189.0 mm (95% CI 112.8–289.3 mm) to 69.0 mm (95% CI 51.0–115.3 mm), p = 0.0039; while pre-treatment fat thickness in the left hip reduced from 182.0 mm (95% CI 90.5–264.5 mm) to 69.0 mm (63.5–139.0 mm), p = 0.0078.
Figures 1 and 2 show the fat thickness reduction, assessed by diagnostic ultrasound, in a 47-year-old woman who underwent treatment in abdomen and a 34-year-old woman who underwent treatment in abdomen and hips, respectively.

Unadjusted Clinical Outcomes

Pretreatment fat mass was significantly reduced from 32.9% to 31.2% (Hodges-Lehmann median difference: − 1.7%; 95% CI − 3.4% to − 0.1%, p = 0.0006). There were no significant differences between pre- and posttreatment values in BMI, weight, and visceral fat (Table 3) yet significantly regarding fat mass % reduction and in thickness in the four areas evaluated.
Table 3
A comparison between pre- and post-treatment values.
 
Pre-treatment
Post-treatment
Hodges-Lehmann median difference
pa
Weight, kg
   
0.1354
 Median (IqR)
66.7 (59.0–75.3)
66.0 (56.6–74.3)
− 0.5 (− 1.2 to 0.3)
BMI, kg/m2
   
0.1531
 Meadin (IqR)
24.4 (22.1–27.1)
24.0 (22.0–27.2)
− 0.2 (− 0.4 to 0.1)
Visceral fat
   
0.0625
 Median (IqR)
5.0 (3.3–7.8)
5.0 (2.3–7.5)
− 0.5 (− 0.5 to 0.0)
Fat mass, %
   
0.0006
 Median (IqR)
32.9 (26.9–36.2)
31.2 (21.9–35.8)
− 1.7 (− 3.4 to − 0.1)
FTIU, mm
   
0.0001
 Median (IqR)
195.0 (154.3–275.0)
118.0 (6.0–155.0)
− 85.3 (− 107.5 to − 62.0)
FTSU, mm
   
0.0001
 Median (IqR)
174.0 (105.0–217.0)
81.0 (56.8–123.5)
− 70.3 (− 95.0 to − 48.5)
FTRHb
   
0.0039
 Median (IqR)
189.0 (112.8–289.3)
69.0 (51.0–115.3)
− 100.0 (− 140.5 to 49.5)
FTLHc
   
0.0078
 Median (IqR)
182.0 (90.5–264.5)
69.0 (63.5–139.0)
− 71.8 (− 132.5 to − 23.0)
The Wilcoxon test was used to assess the p values between pre- and post-treatment values
IqR Interquartile range, BMI Body mass index, FTIU Fat thickness infraumbilical region, FTSU Fat thickness supraumbilical region, FTRH Fat thickness right hip, FTLH Fat thickness left hip, N.A. Not applicable
aWilcoxon test
bNine subjects
cEight subjects

Adjusted Clinical Outcomes

Once adjusted by age, weight, BMI, fat mass, visceral fat, and pre-treatment thickness our study did not find significant differences in any of the study variables between subjects who underwent treatment in abdomen and those who underwent treatment in abdomen and hip regions (Table 4).
Table 4
Evolution of fat thickness in the different regions.
 
Abdomen
Abdomen and hips
Difference between treatment groups
Median (95% CI) difference from baseline
pa
Median (95% CI) difference from baseline
pa
Mean (95% CI)
pb
Weight, kg
− 0.6 (− 1.7 to 1.3)
0.4375
− 0.5 (− 1.9 to 0.6)
0.2500
0.4 (− 1.8 to 2.6)
0.6803
BMI, kg/m2
− 0.2 (− 0.6 to 0.4)
0.4375
− 0.2 (− 0.8 to 0.1)
0.2500
0.1 (− 0.6 to 0.8)
0.6497
Visceral fat
0.0 (− 1.0 to 0.0)
0.2353
− 0.5 (− 1.0 to 0.0)
0.1250
0.6 (− 0.5 to 1.7)
0.2624
Fat mass, %
− 1.6 (− 3.9 to 0.3)
0.0938
− 1.8 (− 6.0 to − 0.6)
0.0078
2.9 (− 1.5 to 7.3)
0.1664
FTIU, mmc
− 97.5 (− 127.0 to − 63.0)
0.0313
− 74.0 (− 119.5 to − 37.0)
0.0039
− 10.5 (− 53.5 to 32.5)
0.5821
FTSU, mmd
− 79.0 (− 172.0 to − 52.0)
0.0313
− 61.5 (− 98.5 to − 38.0)
0.0039
− 2.9 (− 24.8 to 18.9)
0.7602
FTRH
N.A.
N.A.
− 100.0 (− 140.5 to 49.5)
0.0039
N.A.
N.A.
FTLH
N.A.
N.A.
− 71.8 (− 132.5 to − 23.0)
0.0078
N.A.
N.A.
The Wilcoxon test was used to assess the intra-group statistical significance. The analysis of covariance (ANCOVA) was used for determining the p values between groups. “Treated area” as a factor and age, weight, BMI, visceral fat, and fat mass as covariates
FTIU Fat thickness infraumbilical region, FTSU Fat thickness supraumbilical region, FTRH Fat thickness right hip, FTLH Fat thickness left hip
aWilcoxon test
bANCOVA test with the Bonferroni correction
cAlso adjusted by pretreatment infraumbilical thickness
dAlso adjusted by pretreatment supraumbilical thickness
Clinical results, assessed by means of the 3D imaging system Canfield Vectra H2 or by 2D photographs, have shown a significant improvement in the aesthetic results (Figs. 3, 4, 5, 6, 7, 8).
The median (IqR) degree of patient satisfaction was 4.0 (4.0–5.0), with 13 (86.7%) patients being “Highly satisfied” or “Satisfied” with the treatment results, and only 2 (13.3%) ones who were “Neutral” with their esthetic results.
Regarding safety, the treatment was well tolerated. Adverse events were mild in severity and were successfully controlled without sequelae or need for medical or physical therapy. The most commonly reported adverse events were mild discomfort, immediately after treatment, which was successfully resolved without treatment; erythema (mild), which was successfully resolved without treatment within 48 h after treatment. No cases of burns or skin atrophy have been observed.

Discussion

The results of the current evaluation suggest that the treatment with a non-invasive device that combines pulsed non-focus ultrasound and Unipolar radiofrequency provided a significant fat thickness reduction in abdomen and hips that lasts at least 3 months, with an excellent safety profile.
After adjusting by different covariates, our study did not find significant differences in fat reduction among patients who underwent treatment in abdomen and those who did it in abdomen and hips. Moreover, with the exception of the fat mass, which was significantly lower after treatment, during the follow-up period, there were no significant changes in weight, BMI or visceral fat, which suggested clearly that reduction in fat thickness was due to local treatment rather than behavioral causes.
Over the past several years, there has been a paradigm shift in the treatment of fat. The clear trend is toward noninvasive technologies over more traditional forms of fat removal such as liposuction [15, 7].
Non-invasive body contouring technologies can be classified by the energy source deployed, or by their effect on adipose tissue. In a more didactic way, these non-invasive therapies may be classified, according to their effect on the adipocyte, in short-term (metabolic size reduction) and long-term (permanent adipocyte death) effect devices [25].
Among the different non-invasive techniques used currently for body contouring, RF and US have become increasingly popular [3, 10, 15]. Adipose tissue is highly influenced by thermal changes. This fact, in combination with the development of new non-invasive ultrasound technologies, has led to the use of ultrasound as a strategy for thermal fat tissue destruction [1316]. Although the effects of HIFU technology have been widely evaluated, the effects of non-focused ultrasound on the adipocyte have been less explored [13, 14].
From a theoretical point of view, the combination of RF with ultrasound targets the problem of localized fat accumulation from two fronts [3, 17, 18]. Moreover, a synergistic effect has been suggested when combining both techniques, amplifying effect of one therapy over the other [17].
Although it has not been fully elucidated, the reduction in adipose tissue after the treatment may be a result of lipolysis or mechanical disruption of subcutaneous adipocytes. Additionally, it has been hypothesized that the absorption of ultrasonic energy can lead to changes in the structure and cellular activity of adipocytes, which causes the liberation of lipids to the bloodstream and extracellular space immediately after the treatment [16, 19]. Fat clearance is performed by physiological pathways, namely lymphatic, venous, and immune systems. Triglycerides from the broken fat cells are released into the interstitial fluid and metabolized subsequently by Lipase enzyme into glycerol and free fatty acids [20]. Glycerol is phosphorylated and transported through the vascular system, while free fatty acids are metabolized and these fat metabolites are processed in the liver in the same manner as fat originating from digested food [20].
According to our results, it can be deduced that to treat the abdomen and hip in the same session is as effective and safety as that to treat only the abdomen. We have not observed synergies or antagonisms when we treat the two regions in the same session. Nevertheless, we did not observe any safety issues related to our technique.
It is extremely difficult to compare our results with the currently available scientific evidence. In fact, as far as we know, this is the first study evaluating the effectiveness of this non-invasive body contouring platform, incorporating its two technologies of ultrasound and RF.
The RF technology delivers a thermal stimulus to the skin and superficial adipose tissue causing a thickening of the dermis and enhancement of fat cell metabolism, resulting in a reduction in skin laxity and adipocyte volume [21, 22].
Different studies have shown a fat tissue reduction using RF, suggesting that RF is an effective and safe method for reducing localized fat, particularly in abdomen and thighs [2325]. Additionally, this technique is less time-consuming than other ones, which may be an advantage.
Another key point of the current study was the assessment of patient’s satisfaction with the treatment. The median degree of satisfaction with the treatment results was high (4 out of 5), with 7 (46.7%) patients who were “satisfied” and 6 (40.0%) ones who were “highly satisfied” with the treatment results.
Regarding safety, the all the adverse effects were mild in severity and were successfully resolved without treatment. No severe adverse events, such as burns or skin atrophy, have been observed in the current study. RF and US are, in general terms, safe technologies for these procedures [3, 5]. Moreover, the absence of complications, such as abdominal pain, erythema, or burns, have been reported [26].
Among others, the main limitations of the current study are a small sample, the limited follow-up, and the lack of a control group.

Conclusions

The results of this study suggested that the combine therapy with pulsed non-focus ultrasound and Unipolar radiofrequency was effective for reducing fat tissue thickness in abdomen and hips in patients with localized fat, while maintained a good safety profile. However, due to the limited sample size and the characteristics of our study, appropriate caution is therefore recommended when extending the results to other patients. New clinical trials including a larger sample size and a longer follow-up using this device will be necessary to confirm these findings.

Acknowledgements

Medical writing and Editorial assistant services have been provided by Ciencia y Deporte S.L. Support for this assistance was funded by Alma Lasers at the request of the investigator.

Declarations

Conflict of interest

Dr Urdiales-Gálvez has received a Grant from Alma Laser for covering the medical writing services and the publication fees. All the coauthors declare that they have no conflicts of interest to disclose.

Ethical Approval

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.

Research Involving Human and Animal Rights

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.
All patients were fully informed about the details of the study protocol and patients provided written informed consent at the beginning of the study.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

Die Chirurgie

Print-Titel

Das Abo mit mehr Tiefe

Mit der Zeitschrift Die Chirurgie erhalten Sie zusätzlich Online-Zugriff auf weitere 43 chirurgische Fachzeitschriften, CME-Fortbildungen, Webinare, Vorbereitungskursen zur Facharztprüfung und die digitale Enzyklopädie e.Medpedia.

Bis 30. April 2024 bestellen und im ersten Jahr nur 199 € zahlen!

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Dent – Das Online-Abo der Zahnmedizin

Online-Abonnement

Mit e.Dent erhalten Sie Zugang zu allen zahnmedizinischen Fortbildungen und unseren zahnmedizinischen und ausgesuchten medizinischen Zeitschriften.

Literatur
1.
Zurück zum Zitat Mazzoni D, Lin MJ, Dubin DP, Khorasani H (2019) Review of non-invasive body contouring devices for fat reduction, skin tightening and muscle definition. Australas J Dermatol 60(4):278–283CrossRef Mazzoni D, Lin MJ, Dubin DP, Khorasani H (2019) Review of non-invasive body contouring devices for fat reduction, skin tightening and muscle definition. Australas J Dermatol 60(4):278–283CrossRef
2.
Zurück zum Zitat Lee NY, Robinson DM (2017) Noninvasive body contouring. Semin Cutan Med Surg 36(4):170–178CrossRef Lee NY, Robinson DM (2017) Noninvasive body contouring. Semin Cutan Med Surg 36(4):170–178CrossRef
3.
Zurück zum Zitat Rzepecki AK, Farberg AS, Hashim PW, Goldenberg G (2018) Update on noninvasive body contouring techniques. Cutis 101(4):285–288PubMed Rzepecki AK, Farberg AS, Hashim PW, Goldenberg G (2018) Update on noninvasive body contouring techniques. Cutis 101(4):285–288PubMed
5.
Zurück zum Zitat Kennedy J, Verne S, Griffith R, Falto-Aizpurua L, Nouri K (2015) Non-invasive subcutaneous fat reduction: a review. J Eur Acad Dermatol Venereol 29(9):1679–1688CrossRef Kennedy J, Verne S, Griffith R, Falto-Aizpurua L, Nouri K (2015) Non-invasive subcutaneous fat reduction: a review. J Eur Acad Dermatol Venereol 29(9):1679–1688CrossRef
6.
Zurück zum Zitat Shermak MA (2012) Body contouring. Plast Reconstr Surg 129(6):963e–978eCrossRef Shermak MA (2012) Body contouring. Plast Reconstr Surg 129(6):963e–978eCrossRef
8.
Zurück zum Zitat Narins RS, Tope WD, Pope K, Ross EV (2006) Overtreatment effects associated with a radiofrequency tissue-tightening device: rare, preventable, and correctable with subcision and autologous fat transfer. Dermatol Surg 32(1):115–124CrossRef Narins RS, Tope WD, Pope K, Ross EV (2006) Overtreatment effects associated with a radiofrequency tissue-tightening device: rare, preventable, and correctable with subcision and autologous fat transfer. Dermatol Surg 32(1):115–124CrossRef
9.
Zurück zum Zitat Araújo AR, Soares VP, Silva FS, Moreira Tda S (2015) Radiofrequency for the treatment of skin laxity: mith or truth. An Bras Dermatol 90(5):707–721CrossRef Araújo AR, Soares VP, Silva FS, Moreira Tda S (2015) Radiofrequency for the treatment of skin laxity: mith or truth. An Bras Dermatol 90(5):707–721CrossRef
10.
Zurück zum Zitat Weiss RA (2013) Noninvasive radio frequency for skin tightening and body contouring. Semin Cutan Med Surg 32(1):9–17PubMed Weiss RA (2013) Noninvasive radio frequency for skin tightening and body contouring. Semin Cutan Med Surg 32(1):9–17PubMed
11.
Zurück zum Zitat Weiss R, Weiss M, Beasley K, Vrba J, Bernardy J (2013) Operator independent focused high frequency ISM band for fat reduction: porcine model. Lasers Surg Med 45(4):235–239CrossRef Weiss R, Weiss M, Beasley K, Vrba J, Bernardy J (2013) Operator independent focused high frequency ISM band for fat reduction: porcine model. Lasers Surg Med 45(4):235–239CrossRef
12.
Zurück zum Zitat Chilukuri S, Mueller G (2016) “Hands-Free” noninvasive body contouring devices: review of effectiveness and patient satisfaction. J Drugs Dermatol 15(11):1402–1406PubMed Chilukuri S, Mueller G (2016) “Hands-Free” noninvasive body contouring devices: review of effectiveness and patient satisfaction. J Drugs Dermatol 15(11):1402–1406PubMed
13.
Zurück zum Zitat Garcia O Jr, Schafer M (2013) The effects of nonfocused external ultrasound on tissue temperature and adipocyte morphology. Aesthet Surg J 33(1):117–127CrossRef Garcia O Jr, Schafer M (2013) The effects of nonfocused external ultrasound on tissue temperature and adipocyte morphology. Aesthet Surg J 33(1):117–127CrossRef
14.
Zurück zum Zitat Levi A, Amitai DB, Lapidoth M (2017) A novel transcutaneous, non-focused ultrasound energy delivering device is able to induce subcutaneous adipose tissue destruction in an animal model. Lasers Surg Med 49(1):110–121CrossRef Levi A, Amitai DB, Lapidoth M (2017) A novel transcutaneous, non-focused ultrasound energy delivering device is able to induce subcutaneous adipose tissue destruction in an animal model. Lasers Surg Med 49(1):110–121CrossRef
15.
Zurück zum Zitat Luo W, Zhou X, Gong X, Zheng M, Zhang J, Guo X (2007) Study of sequential histopathologic changes, apoptosis, and cell proliferation in rabbit livers after high-intensity focused ultrasound ablation. J Ultrasound Med 26(4):477–485CrossRef Luo W, Zhou X, Gong X, Zheng M, Zhang J, Guo X (2007) Study of sequential histopathologic changes, apoptosis, and cell proliferation in rabbit livers after high-intensity focused ultrasound ablation. J Ultrasound Med 26(4):477–485CrossRef
16.
Zurück zum Zitat Minkis K, Alam M (2014) Ultrasound skin tightening. Dermatol Clin 32(1):71–77CrossRef Minkis K, Alam M (2014) Ultrasound skin tightening. Dermatol Clin 32(1):71–77CrossRef
17.
Zurück zum Zitat Canela VC, Crivelaro CN, Ferla LZ, Canela VC, Crivelaro CN, Ferla LZ, Pelozo GM, Azevedo J, Liebano RE et al (2018) Synergistic effects of combined therapy: nonfocused ultrasound plus Aussie current for noninvasive body contouring. Clin Cosmet Investig Dermatol 11:203–212CrossRef Canela VC, Crivelaro CN, Ferla LZ, Canela VC, Crivelaro CN, Ferla LZ, Pelozo GM, Azevedo J, Liebano RE et al (2018) Synergistic effects of combined therapy: nonfocused ultrasound plus Aussie current for noninvasive body contouring. Clin Cosmet Investig Dermatol 11:203–212CrossRef
18.
Zurück zum Zitat Alizadeh Z, Halabchi F, Mazaheri R, Abolhasani M, Tabesh M (2016) Review of the mechanisms and effects of noninvasive body contouring devices on cellulite and subcutaneous fat. Int J Endocrinol Metab 14(4):e36727CrossRef Alizadeh Z, Halabchi F, Mazaheri R, Abolhasani M, Tabesh M (2016) Review of the mechanisms and effects of noninvasive body contouring devices on cellulite and subcutaneous fat. Int J Endocrinol Metab 14(4):e36727CrossRef
19.
Zurück zum Zitat Gonçalves WL, Graceli JB, Santos RL, Cicilini MA, Bissoli NS, Abreu GR et al (2009) Ultrasound lipoclasia on subcutaneous adipose tissue to produce acute hyperglycemia and enhance acute inflammatory response in healthy female rats. Dermatol Surg 35(11):1741–1745CrossRef Gonçalves WL, Graceli JB, Santos RL, Cicilini MA, Bissoli NS, Abreu GR et al (2009) Ultrasound lipoclasia on subcutaneous adipose tissue to produce acute hyperglycemia and enhance acute inflammatory response in healthy female rats. Dermatol Surg 35(11):1741–1745CrossRef
21.
Zurück zum Zitat Zelickson BD, Kist D, Bernstein E, Brown DB, Ksenzenko S, Burns J et al (2004) Histological and ultrastructural evaluation of the effects of a radiofrequency-based nonablative dermal remodeling device: a pilot study. Arch Dermatol 140(2):204–209CrossRef Zelickson BD, Kist D, Bernstein E, Brown DB, Ksenzenko S, Burns J et al (2004) Histological and ultrastructural evaluation of the effects of a radiofrequency-based nonablative dermal remodeling device: a pilot study. Arch Dermatol 140(2):204–209CrossRef
22.
Zurück zum Zitat Elsaie ML (2009) Cutaneous remodeling and photorejuvenation using radiofrequency devices. Indian J Dermatol 54(3):201–205CrossRef Elsaie ML (2009) Cutaneous remodeling and photorejuvenation using radiofrequency devices. Indian J Dermatol 54(3):201–205CrossRef
23.
Zurück zum Zitat Manuskiatti W, Wachirakaphan C, Lektrakul N, Varothai S (2009) Circumference reduction and cellulite treatment with a TriPollar radiofrequency device: a pilot study. J Eur Acad Dermatol Venereol 23(7):820–827CrossRef Manuskiatti W, Wachirakaphan C, Lektrakul N, Varothai S (2009) Circumference reduction and cellulite treatment with a TriPollar radiofrequency device: a pilot study. J Eur Acad Dermatol Venereol 23(7):820–827CrossRef
24.
Zurück zum Zitat Fritz K, Salavastru C (2017) Long-term follow-up on patients treated for abdominal fat using a selective contactless radiofrequency device. J Cosmet Dermatol 16(4):471–475CrossRef Fritz K, Salavastru C (2017) Long-term follow-up on patients treated for abdominal fat using a selective contactless radiofrequency device. J Cosmet Dermatol 16(4):471–475CrossRef
25.
Zurück zum Zitat Boisnic S, Divaris M, Nelson AA, Gharavi NM, Lask GP (2014) A clinical and biological evaluation of a novel, noninvasive radiofrequency device for the long-term reduction of adipose tissue. Lasers Surg Med 46(2):94–103CrossRef Boisnic S, Divaris M, Nelson AA, Gharavi NM, Lask GP (2014) A clinical and biological evaluation of a novel, noninvasive radiofrequency device for the long-term reduction of adipose tissue. Lasers Surg Med 46(2):94–103CrossRef
26.
Zurück zum Zitat Harth Y (2015) Painless, safe, and efficacious noninvasive skin tightening, body contouring, and cellulite reduction using multisource 3DEEP radiofrequency. J Cosmet Dermatol 14(1):70–75CrossRef Harth Y (2015) Painless, safe, and efficacious noninvasive skin tightening, body contouring, and cellulite reduction using multisource 3DEEP radiofrequency. J Cosmet Dermatol 14(1):70–75CrossRef
Metadaten
Titel
Body Contouring Using a Combination of Pulsed Ultrasound and Unipolar Radio Frequency: A Prospective Pilot Study
verfasst von
Fernando Urdiales-Gálvez
Sandra Martín-Sánchez
Mónica Maíz-Jiménez
Esther Viruel-Ortega
Publikationsdatum
01.06.2022
Verlag
Springer US
Erschienen in
Aesthetic Plastic Surgery / Ausgabe 5/2022
Print ISSN: 0364-216X
Elektronische ISSN: 1432-5241
DOI
https://doi.org/10.1007/s00266-022-02919-2

Weitere Artikel der Ausgabe 5/2022

Aesthetic Plastic Surgery 5/2022 Zur Ausgabe

Wie erfolgreich ist eine Re-Ablation nach Rezidiv?

23.04.2024 Ablationstherapie Nachrichten

Nach der Katheterablation von Vorhofflimmern kommt es bei etwa einem Drittel der Patienten zu Rezidiven, meist binnen eines Jahres. Wie sich spätere Rückfälle auf die Erfolgschancen einer erneuten Ablation auswirken, haben Schweizer Kardiologen erforscht.

Hinter dieser Appendizitis steckte ein Erreger

23.04.2024 Appendizitis Nachrichten

Schmerzen im Unterbauch, aber sonst nicht viel, was auf eine Appendizitis hindeutete: Ein junger Mann hatte Glück, dass trotzdem eine Laparoskopie mit Appendektomie durchgeführt und der Wurmfortsatz histologisch untersucht wurde.

Mehr Schaden als Nutzen durch präoperatives Aussetzen von GLP-1-Agonisten?

23.04.2024 Operationsvorbereitung Nachrichten

Derzeit wird empfohlen, eine Therapie mit GLP-1-Rezeptoragonisten präoperativ zu unterbrechen. Eine neue Studie nährt jedoch Zweifel an der Notwendigkeit der Maßnahme.

Ureterstriktur: Innovative OP-Technik bewährt sich

19.04.2024 EAU 2024 Kongressbericht

Die Ureterstriktur ist eine relativ seltene Komplikation, trotzdem bedarf sie einer differenzierten Versorgung. In komplexen Fällen wird dies durch die roboterassistierte OP-Technik gewährleistet. Erste Resultate ermutigen.

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.