Skip to main content
Erschienen in: Pediatric Radiology 7/2020

03.03.2020 | Original Article

Clinical evaluation of reconstruction and acquisition time for pediatric 18F-FDG brain PET using digital PET/CT

Erschienen in: Pediatric Radiology | Ausgabe 7/2020

Einloggen, um Zugang zu erhalten

Abstract

Background

18F-2-fluoro-2-deoxyglucose (FDG) positron emission tomography (PET) plays an important role in the diagnosis, evaluation and treatment of childhood epilepsy. The selection of appropriate acquisition and reconstruction parameters, however, can be challenging with the introduction of advanced hardware and software functionalities.

Objective

To quantify the diagnostic performance of a block-sequential regularized expectation maximization (BSREM) tool and reduced effective counts in brain PET/CT for pediatric epilepsy patients on a digital silicon photomultiplier system.

Materials and methods

We included 400 sets of brain PET/CT images from 25 pediatric patients (0.5–16 years old) in this retrospective study. Patient images were reconstructed with conventional iterative techniques or BSREM with varied penalization factor (β), at varied acquisition time (45 s, 90 s, 180 s, 300 s) to simulate reduced count density. Two pediatric nuclear medicine physicians reviewed images in random order — blinded to patient, reconstruction method and imaging time — and scored technical quality (noise, spatial resolution, artifacts), clinical quality (image quality of the cortex, basal ganglia and thalamus) and overall diagnostic satisfaction on a 5-point scale.

Results

Reconstruction with BSREM improved quality and clinical scores across all count levels, with the greatest benefits in low-count conditions. Image quality scores were greatest at 300-s acquisition times with β=500 (overall; noise; artifacts; image quality of the cortex, basal ganglia and thalamus) or β=200 (spatial resolution). No statistically significant difference in the highest graded reconstruction was observed between imaging at 180 s and 300 s with an appropriately implemented penalization factor (β=350–500), indicating that a reduction in dose or acquisition time is feasible without reduction in diagnostic satisfaction.

Conclusion

Clinical evaluation of pediatric 18F-FDG brain PET image quality was shown to be diagnostic at reductions of count density by 40% using BSREM with a penalization factor of β=350–500. This can be accomplished while maintaining confidence of achieving a diagnostic-quality image.
Literatur
1.
Zurück zum Zitat Aaberg KM, Gunnes N, Bakken IJ et al (2017) Incidence and prevalence of childhood epilepsy: a nationwide cohort study. Pediatrics 139:e20163908CrossRef Aaberg KM, Gunnes N, Bakken IJ et al (2017) Incidence and prevalence of childhood epilepsy: a nationwide cohort study. Pediatrics 139:e20163908CrossRef
2.
Zurück zum Zitat Fitsiori A, Hiremath S, Boto J et al (2019) Morphological and advanced imaging of epilepsy: beyond the basics. Children 6:43CrossRef Fitsiori A, Hiremath S, Boto J et al (2019) Morphological and advanced imaging of epilepsy: beyond the basics. Children 6:43CrossRef
3.
Zurück zum Zitat Berg AT, Shinnar S, Levy SR et al (2001) Early development of intractable epilepsy in children: a prospective study. Neurology 56:1445–1452CrossRef Berg AT, Shinnar S, Levy SR et al (2001) Early development of intractable epilepsy in children: a prospective study. Neurology 56:1445–1452CrossRef
4.
Zurück zum Zitat Ollenberger GP, Byrne AJ, Berlangieri SU et al (2005) Assessment of the role of FDG PET in the diagnosis and management of children with refractory epilepsy. Eur J Nucl Med Mol Imaging 32:1311–1316CrossRef Ollenberger GP, Byrne AJ, Berlangieri SU et al (2005) Assessment of the role of FDG PET in the diagnosis and management of children with refractory epilepsy. Eur J Nucl Med Mol Imaging 32:1311–1316CrossRef
5.
Zurück zum Zitat Chugani HT, Conti JR (1996) Etiologic classification of infantile spasms in 140 cases: role of positron emission tomography. J Child Neurol 11:44–48CrossRef Chugani HT, Conti JR (1996) Etiologic classification of infantile spasms in 140 cases: role of positron emission tomography. J Child Neurol 11:44–48CrossRef
6.
Zurück zum Zitat Burneo JG, Poon R, Kellett S, Snead OC (2015) The utility of positron emission tomography in epilepsy. Can J Neurol Sci 42:360–371CrossRef Burneo JG, Poon R, Kellett S, Snead OC (2015) The utility of positron emission tomography in epilepsy. Can J Neurol Sci 42:360–371CrossRef
7.
Zurück zum Zitat la Fougère C, Rominger A, Förster S et al (2009) PET and SPECT in epilepsy: a critical review. Epilepsy Behav 15:50–55CrossRef la Fougère C, Rominger A, Förster S et al (2009) PET and SPECT in epilepsy: a critical review. Epilepsy Behav 15:50–55CrossRef
9.
Zurück zum Zitat Hsu DFC, Ilan E, Peterson WT et al (2017) Studies of a next-generation silicon-photomultiplier-based time-of-flight PET/CT system. J Nucl Med 58:1511–1518CrossRef Hsu DFC, Ilan E, Peterson WT et al (2017) Studies of a next-generation silicon-photomultiplier-based time-of-flight PET/CT system. J Nucl Med 58:1511–1518CrossRef
10.
Zurück zum Zitat Teoh EJ, McGowan DR, Macpherson RE et al (2015) Phantom and clinical evaluation of the Bayesian penalized likelihood reconstruction algorithm Q.Clear on an LYSO PET/CT system. J Nucl Med 56:1447–1452CrossRef Teoh EJ, McGowan DR, Macpherson RE et al (2015) Phantom and clinical evaluation of the Bayesian penalized likelihood reconstruction algorithm Q.Clear on an LYSO PET/CT system. J Nucl Med 56:1447–1452CrossRef
11.
Zurück zum Zitat Lantos J, Mittra ES, Levin CS, Iagaru A (2018) Standard OSEM vs. regularized PET image reconstruction: qualitative and quantitative comparison using phantom data and various clinical radiopharmaceuticals. Am J Nucl Med Mol Imaging 8:110–118PubMedPubMedCentral Lantos J, Mittra ES, Levin CS, Iagaru A (2018) Standard OSEM vs. regularized PET image reconstruction: qualitative and quantitative comparison using phantom data and various clinical radiopharmaceuticals. Am J Nucl Med Mol Imaging 8:110–118PubMedPubMedCentral
12.
Zurück zum Zitat Lantos J, Iagaru A, Levin CS (2016) Scanner dependent noise properties of the Q. Clear PET image reconstruction tool. 2015 IEEE Nucl Sci Symp Med Imaging Conf NSS/MIC 2015:1–3 Lantos J, Iagaru A, Levin CS (2016) Scanner dependent noise properties of the Q. Clear PET image reconstruction tool. 2015 IEEE Nucl Sci Symp Med Imaging Conf NSS/MIC 2015:1–3
13.
Zurück zum Zitat Gelfand MJ, Parisi MT, Treves ST (2011) Pediatric radiopharmaceutical administered doses: 2010 North American consensus guidelines. J Nucl Med 52:318–322CrossRef Gelfand MJ, Parisi MT, Treves ST (2011) Pediatric radiopharmaceutical administered doses: 2010 North American consensus guidelines. J Nucl Med 52:318–322CrossRef
14.
Zurück zum Zitat Viera AJ, Garrett JM (2005) Understanding interobserver agreement: the kappa statistic. Fam Med 37:360–363PubMed Viera AJ, Garrett JM (2005) Understanding interobserver agreement: the kappa statistic. Fam Med 37:360–363PubMed
15.
Zurück zum Zitat Lindström E, Sundin A, Trampal C et al (2018) Evaluation of penalized-likelihood estimation reconstruction on a digital time-of-flight PET/CT scanner for 18 F-FDG whole-body examinations. J Nucl Med 59:1152–1158CrossRef Lindström E, Sundin A, Trampal C et al (2018) Evaluation of penalized-likelihood estimation reconstruction on a digital time-of-flight PET/CT scanner for 18 F-FDG whole-body examinations. J Nucl Med 59:1152–1158CrossRef
16.
Zurück zum Zitat Messerli M, Stolzmann P, Egger-Sigg M et al (2018) Impact of a Bayesian penalized likelihood reconstruction algorithm on image quality in novel digital PET/CT: clinical implications for the assessment of lung tumors. EJNMMI Phys 5:27CrossRef Messerli M, Stolzmann P, Egger-Sigg M et al (2018) Impact of a Bayesian penalized likelihood reconstruction algorithm on image quality in novel digital PET/CT: clinical implications for the assessment of lung tumors. EJNMMI Phys 5:27CrossRef
17.
Zurück zum Zitat Sekine T, Delso G, Zeimpekis KG et al (2018) Reduction of 18F-FDG dose in clinical PET/MR imaging by using silicon photomultiplier detectors. Radiology 286:249–259CrossRef Sekine T, Delso G, Zeimpekis KG et al (2018) Reduction of 18F-FDG dose in clinical PET/MR imaging by using silicon photomultiplier detectors. Radiology 286:249–259CrossRef
18.
Zurück zum Zitat Trägårdh E, Minarik D, Almquist H et al (2019) Impact of acquisition time and penalizing factor in a block-sequential regularized expectation maximization reconstruction algorithm on a Si-photomultiplier-based PET-CT system for 18F-FDG. EJNMMI Res 9:64CrossRef Trägårdh E, Minarik D, Almquist H et al (2019) Impact of acquisition time and penalizing factor in a block-sequential regularized expectation maximization reconstruction algorithm on a Si-photomultiplier-based PET-CT system for 18F-FDG. EJNMMI Res 9:64CrossRef
19.
Zurück zum Zitat Otani T, Hosono M, Kanagaki M et al (2019) Evaluation and optimization of a new PET reconstruction algorithm, Bayesian penalized likelihood reconstruction, for lung cancer assessment according to lesion size. AJR Am J Roentgenol 213:W50–W56CrossRef Otani T, Hosono M, Kanagaki M et al (2019) Evaluation and optimization of a new PET reconstruction algorithm, Bayesian penalized likelihood reconstruction, for lung cancer assessment according to lesion size. AJR Am J Roentgenol 213:W50–W56CrossRef
20.
Zurück zum Zitat Reynés-Llompart G, Gámez-Cenzano C, Vercher-Conejero JL et al (2018) Phantom, clinical, and texture indices evaluation and optimization of a penalized-likelihood image reconstruction method (Q.Clear) on a BGO PET/CT scanner. Med Phys 45:3214–3222CrossRef Reynés-Llompart G, Gámez-Cenzano C, Vercher-Conejero JL et al (2018) Phantom, clinical, and texture indices evaluation and optimization of a penalized-likelihood image reconstruction method (Q.Clear) on a BGO PET/CT scanner. Med Phys 45:3214–3222CrossRef
Metadaten
Titel
Clinical evaluation of reconstruction and acquisition time for pediatric 18F-FDG brain PET using digital PET/CT
Publikationsdatum
03.03.2020
Erschienen in
Pediatric Radiology / Ausgabe 7/2020
Print ISSN: 0301-0449
Elektronische ISSN: 1432-1998
DOI
https://doi.org/10.1007/s00247-020-04640-1

Weitere Artikel der Ausgabe 7/2020

Pediatric Radiology 7/2020 Zur Ausgabe

Hermes

Hermes

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.