Skip to main content
Erschienen in: Graefe's Archive for Clinical and Experimental Ophthalmology 11/2009

01.11.2009 | Refractive Surgery

Comparison of different excimer laser ablation frequencies (50, 200, and 500 Hz)

verfasst von: C. Winkler von Mohrenfels, R. Khoramnia, C.-P. Lohmann

Erschienen in: Graefe's Archive for Clinical and Experimental Ophthalmology | Ausgabe 11/2009

Einloggen, um Zugang zu erhalten

Abstract

Purpose

The aim of this study was to compare the influence of different excimer laser (193 nm) ablation frequencies (50, 200, and 500 Hz) on porcine cornea.

Methods

For this study, fresh porcine cadaver eyes were obtained from the butcher. PRK and LASIK were performed using a WaveLight Concept System 500 excimer laser. Various repetition rates (50, 200, and 500 Hz) and intended corrections (between -3 and -9 D) were used. Corneal endothelium was investigated pre- and post-operatively with the endothelial cell microscope EM 1000 Tomey™. Post-operatively, the corneas were embedded for light microscopic (HE, Picrosirius Red) and electron microscopic evaluation.

Results

The mean pre-operative cell count was 2,953 (SD 149) and the mean post-operative cell count was 2,974 (SD 156). There was no difference between the pre- and post-operative cell count and the different ablation frequencies. The distribution of collagen fibers and keratocytes was comparable in all different ablation frequencies. Also, the ultrastructural evaluation of keratocytes and fibers, assessed with transmission electron microscopy, was equal in all three different ablation frequencies.

Conclusions

The endothelial cell density and structural and ultra structural evaluation of the different ablation frequencies (50, 200, 500 Hz) showed no specific side-effects that could be associated with higher repetition rates. The quality of the ablation was comparable in all three different laser ablation frequencies. The 500-Hz excimer laser can provide the same ablation quality as slower platforms, but can also offer the advantage of requiring less treatment time than other laser platforms with lower repetition rates.
Literatur
1.
Zurück zum Zitat Trokel SL, Srinivasan R, Braren B (1983) Excimer laser surgery of the cornea. Am J Ophthalmol 96:710–715PubMed Trokel SL, Srinivasan R, Braren B (1983) Excimer laser surgery of the cornea. Am J Ophthalmol 96:710–715PubMed
2.
Zurück zum Zitat O'Brart DP, Corbett MC, Lohmann CP, Kerr Muir MG, Marshall J (1995) The effects of ablation diameter on the outcome of excimer laser photorefractive keratectomy. A prospective, randomized, double-blind study. Arch Ophthalmol 113:438–443PubMed O'Brart DP, Corbett MC, Lohmann CP, Kerr Muir MG, Marshall J (1995) The effects of ablation diameter on the outcome of excimer laser photorefractive keratectomy. A prospective, randomized, double-blind study. Arch Ophthalmol 113:438–443PubMed
3.
Zurück zum Zitat Buratto L, Ferrari M, Genisi C (1993) Myopic keratomileusis with the excimer laser: one-year follow up. Refract Corneal Surg 9:12–19PubMed Buratto L, Ferrari M, Genisi C (1993) Myopic keratomileusis with the excimer laser: one-year follow up. Refract Corneal Surg 9:12–19PubMed
4.
Zurück zum Zitat Gartry DS, Kerr Muir MG, Marshall J (1991) Photorefractive keratectomy with an argon fluoride excimer laser: a clinical study. Refract Corneal Surg 7:420–435PubMed Gartry DS, Kerr Muir MG, Marshall J (1991) Photorefractive keratectomy with an argon fluoride excimer laser: a clinical study. Refract Corneal Surg 7:420–435PubMed
5.
Zurück zum Zitat Seiler T (1995) Current evaluation of myopia correction with the excimer laser. Ophthalmologe 92:379–384PubMed Seiler T (1995) Current evaluation of myopia correction with the excimer laser. Ophthalmologe 92:379–384PubMed
7.
Zurück zum Zitat Durrie DS (2003) First 100 CustomCornea commercial eyes. J Refract Surg 19:S687–S690PubMed Durrie DS (2003) First 100 CustomCornea commercial eyes. J Refract Surg 19:S687–S690PubMed
9.
10.
Zurück zum Zitat Guell JL, Velasco F (2003) Topographically guided ablations for the correction of irregular astigmatism after corneal surgery. Int Ophthalmol Clin 43:111–128PubMed Guell JL, Velasco F (2003) Topographically guided ablations for the correction of irregular astigmatism after corneal surgery. Int Ophthalmol Clin 43:111–128PubMed
11.
Zurück zum Zitat Junqueira LC, Bignolas G, Brentani RR (1979) Picrosirius staining plus polarization microscopy, a specific method for collagen detection in tissue sections. Histochem J 11:447–455. doi:10.1007/BF01002772 PubMedCrossRef Junqueira LC, Bignolas G, Brentani RR (1979) Picrosirius staining plus polarization microscopy, a specific method for collagen detection in tissue sections. Histochem J 11:447–455. doi:10.​1007/​BF01002772 PubMedCrossRef
12.
Zurück zum Zitat Puchtler H, Waldrop FS, Valentine LS (1973) Polarization microscopic studies of connective tissue stained with Picro-sirius Red FBA. Beitr Pathol 150:174–187PubMed Puchtler H, Waldrop FS, Valentine LS (1973) Polarization microscopic studies of connective tissue stained with Picro-sirius Red FBA. Beitr Pathol 150:174–187PubMed
15.
16.
Zurück zum Zitat Iseli HP, Mrochen M, Hafezi F, Seller T (2004) Clinical photoablation with a 500-Hz scanning spot excimer laser. J Refract Surg 20:831–834PubMed Iseli HP, Mrochen M, Hafezi F, Seller T (2004) Clinical photoablation with a 500-Hz scanning spot excimer laser. J Refract Surg 20:831–834PubMed
Metadaten
Titel
Comparison of different excimer laser ablation frequencies (50, 200, and 500 Hz)
verfasst von
C. Winkler von Mohrenfels
R. Khoramnia
C.-P. Lohmann
Publikationsdatum
01.11.2009
Verlag
Springer-Verlag
Erschienen in
Graefe's Archive for Clinical and Experimental Ophthalmology / Ausgabe 11/2009
Print ISSN: 0721-832X
Elektronische ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-009-1102-x

Weitere Artikel der Ausgabe 11/2009

Graefe's Archive for Clinical and Experimental Ophthalmology 11/2009 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Ophthalmika in der Schwangerschaft

Die Verwendung von Ophthalmika in der Schwangerschaft und Stillzeit stellt immer eine Off-label-Anwendung dar. Ein Einsatz von Arzneimitteln muss daher besonders sorgfältig auf sein Risiko-Nutzen-Verhältnis bewertet werden. In der vorliegenden …

Operative Therapie und Keimnachweis bei endogener Endophthalmitis

Vitrektomie Originalie

Die endogene Endophthalmitis ist eine hämatogen fortgeleitete, bakterielle oder fungale Infektion, die über choroidale oder retinale Gefäße in den Augapfel eingeschwemmt wird [ 1 – 3 ]. Von dort infiltrieren die Keime in die Netzhaut, den …

Bakterielle endogene Endophthalmitis

Vitrektomie Leitthema

Eine endogene Endophthalmitis stellt einen ophthalmologischen Notfall dar, der umgehender Diagnostik und Therapie bedarf. Es sollte mit geeigneten Methoden, wie beispielsweise dem Freiburger Endophthalmitis-Set, ein Keimnachweis erfolgen. Bei der …

So erreichen Sie eine bestmögliche Wundheilung der Kornea

Die bestmögliche Wundheilung der Kornea, insbesondere ohne die Ausbildung von lichtstreuenden Narben, ist oberstes Gebot, um einer dauerhaften Schädigung der Hornhaut frühzeitig entgegenzuwirken und die Funktion des Auges zu erhalten.   

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.