Comparison of human expert and computer-automated systems using magnitude-squared coherence (MSC) and bootstrap distribution statistics for the interpretation of pattern electroretinograms (PERGs) in infants with optic nerve hypoplasia (ONH)

verfasst von: Anthony C. Fisher, Daphne L. McCulloch, Mark S. Borchert, Pamela Garcia-Filion, Cassandra Fink, Antonio Eleuteri, David M. Simpson

Erschienen in: Documenta Ophthalmologica | Ausgabe 1/2015

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Pattern electroretinograms (PERGs) have inherently low signal-to-noise ratios and can be difficult to detect when degraded by pathology or noise. We compare an objective system for automated PERG analysis with expert human interpretation in children with optic nerve hypoplasia (ONH) with PERGs ranging from clear to undetectable.

Methods

PERGs were recorded uniocularly with chloral hydrate sedation in children with ONH (aged 3.5–35 months). Stimuli were reversing checks of four sizes focused using an optical system incorporating the cycloplegic refraction. Forty PERG records were analysed; 20 selected at random and 20 from eyes with good vision (fellow eyes or eyes with mild ONH) from over 300 records. Two experts identified P50 and N95 of the PERGs after manually deleting trials with movement artefact, slow-wave EEG (4–8 Hz) or other noise from raw data for 150 check reversals. The automated system first identified present/not-present responses using a magnitude-squared coherence criterion and then, for responses confirmed as present, estimated the P50 and N95 cardinal positions as the turning points in local third-order polynomials fitted in the −3 dB bandwidth [0.25 … 45] Hz. Confidence limits were estimated from bootstrap re-sampling with replacement. The automated system uses an interactive Internet-available webpage tool (see http://​clinengnhs.​liv.​ac.​uk/​esp_​perg_​1.​htm).

Results

The automated system detected 28 PERG signals above the noise level (p ≤ 0.05 for H0). Good subjective quality ratings were indicative of significant PERGs; however, poor subjective quality did not necessarily predict non-significant signals. P50 and N95 implicit times showed good agreement between the two experts and between experts and the automated system. For the N95 amplitude measured to P50, the experts differed by an average of 13 % consistent with differing interpretations of peaks within noise, while the automated amplitude measure was highly correlated with the expert measures but was proportionally larger. Trial-by-trial review of these data required approximately 6.5 h for each human expert, while automated data processing required <4 min, excluding overheads relating to data transfer.

Conclusions

An automated computer system for PERG analysis, using a panel of signal processing and statistical techniques, provides objective present/not-present detection and cursor positioning with explicit confidence intervals. The system achieves, within an efficient and robust statistical framework, estimates of P50 and N95 amplitudes and implicit times similar to those of clinical experts.
Literatur
1.
Zurück zum Zitat Holder GE (2001) Pattern electroretinography (PERG) and an integrated approach to visual pathway diagnosis. Prog Retin Eye Res 20(4):531–561PubMedCrossRef Holder GE (2001) Pattern electroretinography (PERG) and an integrated approach to visual pathway diagnosis. Prog Retin Eye Res 20(4):531–561PubMedCrossRef
2.
3.
Zurück zum Zitat Holder GE et al (1998) Electrophysiological findings in dominant optic atrophy (DOA) linking to the OPA1 locus on chromosome 3q 28-qter. Doc Ophthalmol 95(3–4):217–228PubMedCrossRef Holder GE et al (1998) Electrophysiological findings in dominant optic atrophy (DOA) linking to the OPA1 locus on chromosome 3q 28-qter. Doc Ophthalmol 95(3–4):217–228PubMedCrossRef
4.
Zurück zum Zitat Harrison JM et al (1987) The pattern erg in man following surgical resection of the optic-nerve. Invest Ophthalmol Vis Sci 28(3):492–499PubMed Harrison JM et al (1987) The pattern erg in man following surgical resection of the optic-nerve. Invest Ophthalmol Vis Sci 28(3):492–499PubMed
5.
6.
Zurück zum Zitat Borchert M et al (1995) Clinical-assessment, optic disk measurements, and visual-evoked potential in optic-nerve hypoplasia. Am J Ophthalmol 120(5):605–612PubMedCrossRef Borchert M et al (1995) Clinical-assessment, optic disk measurements, and visual-evoked potential in optic-nerve hypoplasia. Am J Ophthalmol 120(5):605–612PubMedCrossRef
7.
Zurück zum Zitat Hoyt CS, Billson FA (1986) Optic-nerve hypoplasia—changing perspectives. Aust N Z J Ophthalmol 14(4):325–331PubMedCrossRef Hoyt CS, Billson FA (1986) Optic-nerve hypoplasia—changing perspectives. Aust N Z J Ophthalmol 14(4):325–331PubMedCrossRef
8.
Zurück zum Zitat Skarf B, Hoyt CS (1984) Optic-nerve hypoplasia in children—association with anomalies of the endocrine and CNS. Arch Ophthalmol 102(1):62–67PubMedCrossRef Skarf B, Hoyt CS (1984) Optic-nerve hypoplasia in children—association with anomalies of the endocrine and CNS. Arch Ophthalmol 102(1):62–67PubMedCrossRef
9.
Zurück zum Zitat Zeki SM, Dudgeon J, Dutton GN (1991) Reappraisal of the ratio of disk to macula disk diameter in optic-nerve hypoplasia. Br J Ophthalmol 75(9):538–541PubMedCentralPubMedCrossRef Zeki SM, Dudgeon J, Dutton GN (1991) Reappraisal of the ratio of disk to macula disk diameter in optic-nerve hypoplasia. Br J Ophthalmol 75(9):538–541PubMedCentralPubMedCrossRef
10.
Zurück zum Zitat Janaky M et al (1994) Electrophysiologic alterations in patients with optic-nerve hypoplasia. Doc Ophthalmol 86(3):247–257PubMedCrossRef Janaky M et al (1994) Electrophysiologic alterations in patients with optic-nerve hypoplasia. Doc Ophthalmol 86(3):247–257PubMedCrossRef
11.
Zurück zum Zitat Hoyt CS, Good WV (1992) Do we really understand the difference between optic-nerve hypoplasia and atrophy. Eye 6:201–204PubMedCrossRef Hoyt CS, Good WV (1992) Do we really understand the difference between optic-nerve hypoplasia and atrophy. Eye 6:201–204PubMedCrossRef
12.
13.
Zurück zum Zitat Kriss A, Russelleggitt I (1992) Electrophysiological assessment of visual pathway function in infants. Eye 6:145–153PubMedCrossRef Kriss A, Russelleggitt I (1992) Electrophysiological assessment of visual pathway function in infants. Eye 6:145–153PubMedCrossRef
14.
Zurück zum Zitat McCulloch DL et al (2007) Retinal function in infants with optic nerve hypoplasia: electroretinograms to large patterns and photopic flash. Eye 21(6):712–720PubMedCrossRef McCulloch DL et al (2007) Retinal function in infants with optic nerve hypoplasia: electroretinograms to large patterns and photopic flash. Eye 21(6):712–720PubMedCrossRef
15.
Zurück zum Zitat McCulloch DL et al (2010) Clinical electrophysiology and visual outcome in optic nerve hypoplasia. Br J Ophthalmol 94(8):1017–1023PubMedCrossRef McCulloch DL et al (2010) Clinical electrophysiology and visual outcome in optic nerve hypoplasia. Br J Ophthalmol 94(8):1017–1023PubMedCrossRef
16.
Zurück zum Zitat Garcia-Filion P et al (2008) Neuroradiographic, endocrinologic, and ophthalmic correlates of adverse developmental outcomes in children with optic nerve hypoplasia: a prospective study. Pediatrics 121(3):E653–E659PubMedCrossRef Garcia-Filion P et al (2008) Neuroradiographic, endocrinologic, and ophthalmic correlates of adverse developmental outcomes in children with optic nerve hypoplasia: a prospective study. Pediatrics 121(3):E653–E659PubMedCrossRef
17.
Zurück zum Zitat Chaplin C et al (2009) Light-adapted electroretinograms in optic nerve hypoplasia. Doc Ophthalmol 119(2):123–132PubMedCrossRef Chaplin C et al (2009) Light-adapted electroretinograms in optic nerve hypoplasia. Doc Ophthalmol 119(2):123–132PubMedCrossRef
18.
Zurück zum Zitat Dawson WW, Trick GL, Litzkow CA (1979) Improved electrode for electroretinography. Invest Ophthalmol Vis Sci 18(9):988–991PubMed Dawson WW, Trick GL, Litzkow CA (1979) Improved electrode for electroretinography. Invest Ophthalmol Vis Sci 18(9):988–991PubMed
19.
Zurück zum Zitat McCulloch DL, Van Boemel GB, Borchert MS (1997) Comparisons of contact lens, foil, fiber and skin electrodes for patterns electroretinograms. Doc Ophthalmol 94(4):327–340PubMedCrossRef McCulloch DL, Van Boemel GB, Borchert MS (1997) Comparisons of contact lens, foil, fiber and skin electrodes for patterns electroretinograms. Doc Ophthalmol 94(4):327–340PubMedCrossRef
20.
Zurück zum Zitat Bach M et al (2013) ISCEV standard for clinical pattern electroretinography (PERG): 2012 update. Doc Ophthalmol 126(1):1–7PubMedCrossRef Bach M et al (2013) ISCEV standard for clinical pattern electroretinography (PERG): 2012 update. Doc Ophthalmol 126(1):1–7PubMedCrossRef
22.
Zurück zum Zitat Barnett V, Lewis T (1994) Outliers in statistical data. Wiley Series in probability and Mathematical Statistics. Wiley, Chichester Barnett V, Lewis T (1994) Outliers in statistical data. Wiley Series in probability and Mathematical Statistics. Wiley, Chichester
23.
Zurück zum Zitat Carter GC, Knapp CH, Nuttall AH (1973) Estimation of magnitude-squared coherence function via overlapped fast Fourier-transform processing. IEEE Trans Audio Electroacoust 21(4):337–344CrossRef Carter GC, Knapp CH, Nuttall AH (1973) Estimation of magnitude-squared coherence function via overlapped fast Fourier-transform processing. IEEE Trans Audio Electroacoust 21(4):337–344CrossRef
24.
Zurück zum Zitat Dobie RA, Wilson MJ (1989) Analysis of auditory evoked-potentials by magnitude-squared coherence. Ear Hear 10(1):2–13PubMedCrossRef Dobie RA, Wilson MJ (1989) Analysis of auditory evoked-potentials by magnitude-squared coherence. Ear Hear 10(1):2–13PubMedCrossRef
25.
Zurück zum Zitat Oppenheim AVS, Schafer RW (1975) Digital signal processing. Prentice-Hall International, Hemel Hempstead Oppenheim AVS, Schafer RW (1975) Digital signal processing. Prentice-Hall International, Hemel Hempstead
26.
Zurück zum Zitat Koopmans LH (1974) The spectral analysis of time series. Academic Press, San Diego Koopmans LH (1974) The spectral analysis of time series. Academic Press, San Diego
27.
Zurück zum Zitat Efron B, Tibshirani R (1993) An introduction to the bootstrap. Chapman & Hall/CRC, Boca RatonCrossRef Efron B, Tibshirani R (1993) An introduction to the bootstrap. Chapman & Hall/CRC, Boca RatonCrossRef
28.
Zurück zum Zitat Viswanathan S, Frishman LJ, Robson JG (2000) The uniform field and pattern ERG in macaques with experimental glaucoma: removal of spiking activity. Invest Ophthalmol Vis Sci 41(9):2797–2810PubMed Viswanathan S, Frishman LJ, Robson JG (2000) The uniform field and pattern ERG in macaques with experimental glaucoma: removal of spiking activity. Invest Ophthalmol Vis Sci 41(9):2797–2810PubMed
29.
Zurück zum Zitat Bland JM, Altman DG (2010) Statistical methods for assessing agreement between two methods of clinical measurement. Int J Nurs Stud 47(8):931–936CrossRef Bland JM, Altman DG (2010) Statistical methods for assessing agreement between two methods of clinical measurement. Int J Nurs Stud 47(8):931–936CrossRef
30.
Zurück zum Zitat Bach M (2001) Electrophysiological approaches for early detection of glaucoma. Eur J Ophthalmol 11:S41–S49PubMed Bach M (2001) Electrophysiological approaches for early detection of glaucoma. Eur J Ophthalmol 11:S41–S49PubMed
31.
Zurück zum Zitat Bach M, Gerling J, Geiger K (1992) Optic atrophy reduces the pattern-electroretinogram for both fine and coarse stimulus patterns. Clin Vis Sci 7(4):327–333 Bach M, Gerling J, Geiger K (1992) Optic atrophy reduces the pattern-electroretinogram for both fine and coarse stimulus patterns. Clin Vis Sci 7(4):327–333
32.
Zurück zum Zitat Mashima Y, Oguchi Y (1985) Clinical-study of the pattern electroretinogram in patients with optic-nerve damage. Doc Ophthalmol 61(1):91–96PubMedCrossRef Mashima Y, Oguchi Y (1985) Clinical-study of the pattern electroretinogram in patients with optic-nerve damage. Doc Ophthalmol 61(1):91–96PubMedCrossRef
33.
Zurück zum Zitat Thompson DA, Drasdo N (1994) The origins of luminance and pattern responses of the pattern electroretinogram. Int J Psychophysiol 16(2–3):219–227PubMedCrossRef Thompson DA, Drasdo N (1994) The origins of luminance and pattern responses of the pattern electroretinogram. Int J Psychophysiol 16(2–3):219–227PubMedCrossRef
34.
Zurück zum Zitat Ver Hoeve JN et al (1999) VEP and PERG acuity in anesthetized young adult rhesus monkeys. Vis Neurosci 16(4):607–617PubMed Ver Hoeve JN et al (1999) VEP and PERG acuity in anesthetized young adult rhesus monkeys. Vis Neurosci 16(4):607–617PubMed
35.
Zurück zum Zitat Bach M, Hoffmann MB (2008) Update on the pattern electroretinogram in glaucoma. Optom Vis Sci 85(6):386–395PubMedCrossRef Bach M, Hoffmann MB (2008) Update on the pattern electroretinogram in glaucoma. Optom Vis Sci 85(6):386–395PubMedCrossRef
36.
Zurück zum Zitat Meigen T, Bach M (1999) On the statistical significance of electrophysiological steady-state responses. Doc Ophthalmol 98(3):207–232PubMedCrossRef Meigen T, Bach M (1999) On the statistical significance of electrophysiological steady-state responses. Doc Ophthalmol 98(3):207–232PubMedCrossRef
37.
Zurück zum Zitat Preiser D et al (2013) Photopic negative response versus pattern electroretinogram in early glaucoma. Invest Ophthalmol Vis Sci 54(2):1182–1191PubMedCrossRef Preiser D et al (2013) Photopic negative response versus pattern electroretinogram in early glaucoma. Invest Ophthalmol Vis Sci 54(2):1182–1191PubMedCrossRef
38.
Zurück zum Zitat Miyata K et al (2007) Reduction of oscillatory potentials and photopic negative response in patients with autosomal dominant optic atrophy with OPA1 mutations. Invest Ophthalmol Vis Sci 48(2):820–824PubMedCrossRef Miyata K et al (2007) Reduction of oscillatory potentials and photopic negative response in patients with autosomal dominant optic atrophy with OPA1 mutations. Invest Ophthalmol Vis Sci 48(2):820–824PubMedCrossRef
39.
Zurück zum Zitat Gotoh Y, Machida S, Tazawa Y (2004) Selective loss of the photopic negative response in patients with optic nerve atrophy. Arch Ophthalmol 122(3):341–346PubMedCrossRef Gotoh Y, Machida S, Tazawa Y (2004) Selective loss of the photopic negative response in patients with optic nerve atrophy. Arch Ophthalmol 122(3):341–346PubMedCrossRef
Metadaten
Titel
Comparison of human expert and computer-automated systems using magnitude-squared coherence (MSC) and bootstrap distribution statistics for the interpretation of pattern electroretinograms (PERGs) in infants with optic nerve hypoplasia (ONH)
verfasst von
Anthony C. Fisher
Daphne L. McCulloch
Mark S. Borchert
Pamela Garcia-Filion
Cassandra Fink
Antonio Eleuteri
David M. Simpson
Publikationsdatum
01.08.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
Documenta Ophthalmologica / Ausgabe 1/2015
Print ISSN: 0012-4486
Elektronische ISSN: 1573-2622
DOI
https://doi.org/10.1007/s10633-015-9493-y

Weitere Artikel der Ausgabe 1/2015

Documenta Ophthalmologica 1/2015 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Ophthalmika in der Schwangerschaft

Die Verwendung von Ophthalmika in der Schwangerschaft und Stillzeit stellt immer eine Off-label-Anwendung dar. Ein Einsatz von Arzneimitteln muss daher besonders sorgfältig auf sein Risiko-Nutzen-Verhältnis bewertet werden. In der vorliegenden …

Operative Therapie und Keimnachweis bei endogener Endophthalmitis

Vitrektomie Originalie

Die endogene Endophthalmitis ist eine hämatogen fortgeleitete, bakterielle oder fungale Infektion, die über choroidale oder retinale Gefäße in den Augapfel eingeschwemmt wird [ 1 – 3 ]. Von dort infiltrieren die Keime in die Netzhaut, den …

Bakterielle endogene Endophthalmitis

Vitrektomie Leitthema

Eine endogene Endophthalmitis stellt einen ophthalmologischen Notfall dar, der umgehender Diagnostik und Therapie bedarf. Es sollte mit geeigneten Methoden, wie beispielsweise dem Freiburger Endophthalmitis-Set, ein Keimnachweis erfolgen. Bei der …

So erreichen Sie eine bestmögliche Wundheilung der Kornea

Die bestmögliche Wundheilung der Kornea, insbesondere ohne die Ausbildung von lichtstreuenden Narben, ist oberstes Gebot, um einer dauerhaften Schädigung der Hornhaut frühzeitig entgegenzuwirken und die Funktion des Auges zu erhalten.   

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.