01.12.2010 | Original Paper

Comparison of ocular response analyzer, dynamic contour tonometer and Goldmann applanation tonometer

verfasst von: Charlotte Renier, Thierry Zeyen, Steffen Fieuws, Sofie Vandenbroeck, Ingeborg Stalmans

Erschienen in: International Ophthalmology | Ausgabe 6/2010

Einloggen, um Zugang zu erhalten

Abstract

The aim of this study was to compare the intra-ocular pressure (IOP) obtained by ocular response analyzer (ORA), dynamic contour tonometer (DCT) and Goldmann applanation tonometer (GAT). In 102 patients (47 with primary open-angle glaucoma and 55 healthy controls) IOP was measured with GAT, ORA and DCT in one eye. The agreement between GAT, DCT and ORA values was assessed using Bland–Altman plots. The discrepancy between the methods was related to central corneal thickness (CCT), corneal hysteresis (CH) and corneal resistance factor (CRF) using linear regression models. Significant differences were observed amongst DCT, corneal compensated ORA (ORAcc) and GAT (P < 0.01). Only the ORAcc and DCT were comparable. ORAcc and DCT significantly over-estimated IOP compared to GAT and for ORAcc this difference depended on the height of IOP. A significant correlation was found between CCT and the deviation of DCT and ORAcc from corrected GAT (both P < 0.0001). Our study showed a low degree of agreement between IOP measured by ORA, DCT and GAT. DCT and ORAcc over-estimated the IOP compared to GAT.
Literatur
1.
Zurück zum Zitat Whitacre MM, Stein R (1993) Sources of error with use of Goldmann-type tonometers. Surv Ophthalmol 38:1–30CrossRefPubMed Whitacre MM, Stein R (1993) Sources of error with use of Goldmann-type tonometers. Surv Ophthalmol 38:1–30CrossRefPubMed
2.
Zurück zum Zitat Doughty MJ, Zaman ML (2000) Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach. Surv Ophthalmol 44:367–408CrossRefPubMed Doughty MJ, Zaman ML (2000) Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach. Surv Ophthalmol 44:367–408CrossRefPubMed
3.
Zurück zum Zitat Noecker RS, Dirks MS, Choplin NT, Bernstein P, Batoosingh AL, Whitcup SM (2003) A six-month randomized clinical trial comparing the intraocular pressure-lowering efficacy of bimatoprost and latanoprost in patients with ocular hypertension or glaucoma. Am J Ophthalmol 13:55–63CrossRef Noecker RS, Dirks MS, Choplin NT, Bernstein P, Batoosingh AL, Whitcup SM (2003) A six-month randomized clinical trial comparing the intraocular pressure-lowering efficacy of bimatoprost and latanoprost in patients with ocular hypertension or glaucoma. Am J Ophthalmol 13:55–63CrossRef
4.
Zurück zum Zitat Hager A, Schroeder B, Sadeghi M, Grosserr M, Wiegand W (2007) The influence of corneal hysteresis and corneal resistance factor on the measurement of intraocular pressure. Ophthalmologe 104:484–489CrossRefPubMed Hager A, Schroeder B, Sadeghi M, Grosserr M, Wiegand W (2007) The influence of corneal hysteresis and corneal resistance factor on the measurement of intraocular pressure. Ophthalmologe 104:484–489CrossRefPubMed
5.
Zurück zum Zitat Kanngiesser HE, Kniestedt C, Robert YC (2005) Dynamic contour tonometry: presentation of a new tonometer. J Glaucoma 14:344–350CrossRefPubMed Kanngiesser HE, Kniestedt C, Robert YC (2005) Dynamic contour tonometry: presentation of a new tonometer. J Glaucoma 14:344–350CrossRefPubMed
6.
Zurück zum Zitat Quan H, Shih WJ (1996) Assessing reproducibility by the within-subject coefficient of variation with random effects models. Biometrics 52(4):1195–1203CrossRefPubMed Quan H, Shih WJ (1996) Assessing reproducibility by the within-subject coefficient of variation with random effects models. Biometrics 52(4):1195–1203CrossRefPubMed
7.
Zurück zum Zitat Bland JM, Altman BG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1(8476):307–310PubMed Bland JM, Altman BG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1(8476):307–310PubMed
8.
Zurück zum Zitat Bland JM, Altman DG (1999) Measuring agreement in method comparison studies. Stat Methods Med Res 8:135–160CrossRefPubMed Bland JM, Altman DG (1999) Measuring agreement in method comparison studies. Stat Methods Med Res 8:135–160CrossRefPubMed
9.
Zurück zum Zitat Ku JYF, Danesh-Meyer HV, Craig JP, Gamble GD, McGhee CNJ (2006) Comparison of intraocular pressure measured by Pascal dynamic contour tonometry and Goldmann applanation tonometry. Eye 20:191–198CrossRefPubMed Ku JYF, Danesh-Meyer HV, Craig JP, Gamble GD, McGhee CNJ (2006) Comparison of intraocular pressure measured by Pascal dynamic contour tonometry and Goldmann applanation tonometry. Eye 20:191–198CrossRefPubMed
10.
Zurück zum Zitat Schneider E, Grehn F (2006) Intraocular pressure measurement: comparison of dynamic contour tonometer and Goldmann applanation tonometer. J Glaucoma 15:2–6CrossRefPubMed Schneider E, Grehn F (2006) Intraocular pressure measurement: comparison of dynamic contour tonometer and Goldmann applanation tonometer. J Glaucoma 15:2–6CrossRefPubMed
11.
Zurück zum Zitat Doyle A, Lackhar Y (2005) Comparison of dynamic contour tonometry with Goldmann applanation tonometry over a wide range of central corneal thickness. J Glaucoma 14:288–292CrossRefPubMed Doyle A, Lackhar Y (2005) Comparison of dynamic contour tonometry with Goldmann applanation tonometry over a wide range of central corneal thickness. J Glaucoma 14:288–292CrossRefPubMed
12.
Zurück zum Zitat Salvetat ML, Zeppieri M, Tosoni C, Brushini P (2007) Comparisons between Pascal dynamic contour tonometry, the TonoPen, and Goldmann applanation tonometry in patients with glaucoma. Acta Ophthalmol Scand 85:272–279CrossRefPubMed Salvetat ML, Zeppieri M, Tosoni C, Brushini P (2007) Comparisons between Pascal dynamic contour tonometry, the TonoPen, and Goldmann applanation tonometry in patients with glaucoma. Acta Ophthalmol Scand 85:272–279CrossRefPubMed
13.
Zurück zum Zitat Kaufmann C, Bachmann LM, Thiel MA (2003) Intraocular pressure measurements using dynamic contour tonometry after laser in situ keratomileusis. Invest Ophthalmol Vis Sci 45:3118–3121CrossRef Kaufmann C, Bachmann LM, Thiel MA (2003) Intraocular pressure measurements using dynamic contour tonometry after laser in situ keratomileusis. Invest Ophthalmol Vis Sci 45:3118–3121CrossRef
14.
Zurück zum Zitat Pache M, Wilmsmeyer S, Lautebach S, Funk J (2005) Dynamic contour tonometry versus Goldmann applanation tonometry: a comparative study. Graefes Arch Clin Exp Ophthalmol 243:763–767CrossRefPubMed Pache M, Wilmsmeyer S, Lautebach S, Funk J (2005) Dynamic contour tonometry versus Goldmann applanation tonometry: a comparative study. Graefes Arch Clin Exp Ophthalmol 243:763–767CrossRefPubMed
15.
Zurück zum Zitat Martinez-de-la-Casa JM, Garcia-Feijoo J (2006) Ocular response analyzer versus Goldmann applanation tonometry for intraocular pressure measurements. Invest Ophthalmol Vis Sci 47:4410–4414CrossRefPubMed Martinez-de-la-Casa JM, Garcia-Feijoo J (2006) Ocular response analyzer versus Goldmann applanation tonometry for intraocular pressure measurements. Invest Ophthalmol Vis Sci 47:4410–4414CrossRefPubMed
16.
Zurück zum Zitat Johannesson G, Hallberg P, Eklund A, Linden C (2007) Pascal, ICare and Goldmann applanation tonometry: a comparative study. Acta Ophthalmol 86(6):614–621 Johannesson G, Hallberg P, Eklund A, Linden C (2007) Pascal, ICare and Goldmann applanation tonometry: a comparative study. Acta Ophthalmol 86(6):614–621
17.
Zurück zum Zitat Siganos D, Papastergiou G, Moedas C (2004) Assessment of the Pascal dynamic contour tonometer in monitoring intraocular pressure in unoperated eyes and eyes after LASIK. J Cataract Refract Surg 30:746–751CrossRefPubMed Siganos D, Papastergiou G, Moedas C (2004) Assessment of the Pascal dynamic contour tonometer in monitoring intraocular pressure in unoperated eyes and eyes after LASIK. J Cataract Refract Surg 30:746–751CrossRefPubMed
Metadaten
Titel
Comparison of ocular response analyzer, dynamic contour tonometer and Goldmann applanation tonometer
verfasst von
Charlotte Renier
Thierry Zeyen
Steffen Fieuws
Sofie Vandenbroeck
Ingeborg Stalmans
Publikationsdatum
01.12.2010
Verlag
Springer Netherlands
Erschienen in
International Ophthalmology / Ausgabe 6/2010
Print ISSN: 0165-5701
Elektronische ISSN: 1573-2630
DOI
https://doi.org/10.1007/s10792-010-9377-9

Weitere Artikel der Ausgabe 6/2010

International Ophthalmology 6/2010 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Ophthalmika in der Schwangerschaft

Die Verwendung von Ophthalmika in der Schwangerschaft und Stillzeit stellt immer eine Off-label-Anwendung dar. Ein Einsatz von Arzneimitteln muss daher besonders sorgfältig auf sein Risiko-Nutzen-Verhältnis bewertet werden. In der vorliegenden …

Operative Therapie und Keimnachweis bei endogener Endophthalmitis

Vitrektomie Originalie

Die endogene Endophthalmitis ist eine hämatogen fortgeleitete, bakterielle oder fungale Infektion, die über choroidale oder retinale Gefäße in den Augapfel eingeschwemmt wird [ 1 – 3 ]. Von dort infiltrieren die Keime in die Netzhaut, den …

Bakterielle endogene Endophthalmitis

Vitrektomie Leitthema

Eine endogene Endophthalmitis stellt einen ophthalmologischen Notfall dar, der umgehender Diagnostik und Therapie bedarf. Es sollte mit geeigneten Methoden, wie beispielsweise dem Freiburger Endophthalmitis-Set, ein Keimnachweis erfolgen. Bei der …

So erreichen Sie eine bestmögliche Wundheilung der Kornea

Die bestmögliche Wundheilung der Kornea, insbesondere ohne die Ausbildung von lichtstreuenden Narben, ist oberstes Gebot, um einer dauerhaften Schädigung der Hornhaut frühzeitig entgegenzuwirken und die Funktion des Auges zu erhalten.   

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.