Comparison of Sterilization of Reusable Endoscopic Biopsy Forceps by Autoclaving and Ethylene Oxide Gas

verfasst von: Jai Hoon Yoon, Byung Chul Yoon, Hang Lak Lee, Jun Kyu Lee, Yong-Tae Kim, Dong Ho Lee, Il Ju Choi, Don Haeng Lee, Dong Hee Kim

Erschienen in: Digestive Diseases and Sciences | Ausgabe 2/2012

Einloggen, um Zugang zu erhalten

Abstract

Background and Aims

Every country has standardized reprocessing guidelines for reducing the risk of microorganism transmission via reusable biopsy forceps. Sterilization is performed either by autoclaving or with the use of ethylene oxide (EO) gas. However, there are no clear standard global recommendations. The aim of this study was to determine whether EO gas or autoclaving is a safer and more effective method for the sterilization of reusable forceps.

Methods

This was a prospective study conducted at multiple tertiary referral centers. Seventy reusable biopsy forceps that had been reused at least 20 times each were collected from six endoscopy centers. In all, 61 forceps from five centers were sterilized using EO gas, and the nine forceps from the remaining center were placed in an autoclave. We performed real-time polymerase chain reaction (RT-PCR) for Mycobacterium tuberculosis and hepatitis B virus and performed bacterial cultures on the reusable forceps, which were cut into 2- to 3-cm sections. The forceps were also scanned with an electron microscope (EM) to detect surface damage and contamination.

Results

Escherichia coli bacteria were cultured from 2 of the 61 (3.3%) reusable biopsy forceps sterilized with EO gas. On EM scanning, abundant debris and tissue materials remained on the cup surfaces of the reused biopsy forceps and on their inner wires. No microorganisms were found on the autoclaved forceps.

Conclusions

Sterilization with EO gas may be inadequate because the complicated structure of the forceps may interfere with sterilization. Therefore, for optimum safety, reusable biopsy forceps should be sterilized by autoclaving.
Literatur
1.
Zurück zum Zitat Kimmey MB, Burnett DA, Carr-Locke DL, et al. Transmission of infection by gastrointestinal endoscopy. Gastrointest Endosc. 1993;39:885–888.CrossRef Kimmey MB, Burnett DA, Carr-Locke DL, et al. Transmission of infection by gastrointestinal endoscopy. Gastrointest Endosc. 1993;39:885–888.CrossRef
2.
Zurück zum Zitat Spach DH, Silverstein FE, Stamm WE. Transmission of infection by gastrointestinal endoscopy and bronchoscopy. Ann Intern Med. 1993;118:117–128.PubMed Spach DH, Silverstein FE, Stamm WE. Transmission of infection by gastrointestinal endoscopy and bronchoscopy. Ann Intern Med. 1993;118:117–128.PubMed
4.
Zurück zum Zitat Jung M, Beilenhoff U, Pietsch M, Kraft B, Rippin G. Standardized reprocessing of reusable colonoscopy biopsy forceps is effective: results of a German multicenter study. Endoscopy. 2003;35:197–202. doi:10.1055/s-2003-37270.PubMedCrossRef Jung M, Beilenhoff U, Pietsch M, Kraft B, Rippin G. Standardized reprocessing of reusable colonoscopy biopsy forceps is effective: results of a German multicenter study. Endoscopy. 2003;35:197–202. doi:10.​1055/​s-2003-37270.PubMedCrossRef
5.
Zurück zum Zitat Favero MS, Bond WW. Disinfection of medical and surgical materials. In: Block SS, ed. Disinfection, Sterilization, and Preservation. 5th ed. Philadelphia, PA: Lippincott Williams & Wilkins; 2001:881–917. Favero MS, Bond WW. Disinfection of medical and surgical materials. In: Block SS, ed. Disinfection, Sterilization, and Preservation. 5th ed. Philadelphia, PA: Lippincott Williams & Wilkins; 2001:881–917.
6.
Zurück zum Zitat Kim Y-T. Introduction of proposed guideline for endoscopy disinfection. Korean J Gastrointest Endosc. 2010;40:49. Kim Y-T. Introduction of proposed guideline for endoscopy disinfection. Korean J Gastrointest Endosc. 2010;40:49.
7.
Zurück zum Zitat Wilkinson M, Simmons N, Bramble M, et al. Report of the Working Party of the Endoscopy Committee of the British Society of Gastroenterology on the reuse of endoscopic accessories. Gut. 1998;42:304–306.PubMedCrossRef Wilkinson M, Simmons N, Bramble M, et al. Report of the Working Party of the Endoscopy Committee of the British Society of Gastroenterology on the reuse of endoscopic accessories. Gut. 1998;42:304–306.PubMedCrossRef
8.
Zurück zum Zitat Tandon RK. Disinfection of gastrointestinal endoscopes and accessories. J Gastroenterol Hepatol. 2000;15:G69–G72.PubMedCrossRef Tandon RK. Disinfection of gastrointestinal endoscopes and accessories. J Gastroenterol Hepatol. 2000;15:G69–G72.PubMedCrossRef
9.
Zurück zum Zitat Systchenko R, Marchetti B, Canard JN, et al. Guidelines of the French Society of Digestive Endoscopy: recommendations for setting up cleaning and disinfection procedures in gastrointestinal endoscopy. Endoscopy. 2000;32:807–818. doi:10.1055/s-2000-7710.PubMedCrossRef Systchenko R, Marchetti B, Canard JN, et al. Guidelines of the French Society of Digestive Endoscopy: recommendations for setting up cleaning and disinfection procedures in gastrointestinal endoscopy. Endoscopy. 2000;32:807–818. doi:10.​1055/​s-2000-7710.PubMedCrossRef
10.
Zurück zum Zitat Alvarado CJ, Reichelderfer M. APIC guideline for infection prevention and control in flexible endoscopy. Association for Professionals in Infection Control. Am J Infect Control. 2000;28:138–155. doi:S0196655300780515.PubMedCrossRef Alvarado CJ, Reichelderfer M. APIC guideline for infection prevention and control in flexible endoscopy. Association for Professionals in Infection Control. Am J Infect Control. 2000;28:138–155. doi:S019665530078051​5.PubMedCrossRef
11.
Zurück zum Zitat Nelson DB, Jarvis WR, Rutala WA, et al. Multi-society guideline for reprocessing flexible gastrointestinal endoscopes Society for Healthcare Epidemiology of America. Infect Control Hosp Epidemiol. 2003;24:532–537. doi:10.1086/502237.PubMedCrossRef Nelson DB, Jarvis WR, Rutala WA, et al. Multi-society guideline for reprocessing flexible gastrointestinal endoscopes Society for Healthcare Epidemiology of America. Infect Control Hosp Epidemiol. 2003;24:532–537. doi:10.​1086/​502237.PubMedCrossRef
13.
Zurück zum Zitat Ogoshi K. Reprocessing of gastrointestinal endoscopic accessories. J Gastroenterol Hepatol. 2000;15:G82–G85.PubMedCrossRef Ogoshi K. Reprocessing of gastrointestinal endoscopic accessories. J Gastroenterol Hepatol. 2000;15:G82–G85.PubMedCrossRef
14.
15.
16.
Zurück zum Zitat Shim TS, Chi HS, Lee SD, et al. Adequately washed bronchoscope does not induce false-positive amplification tests on bronchial aspirates in the diagnosis of pulmonary tuberculosis. Chest. 2002;121:774–781.PubMedCrossRef Shim TS, Chi HS, Lee SD, et al. Adequately washed bronchoscope does not induce false-positive amplification tests on bronchial aspirates in the diagnosis of pulmonary tuberculosis. Chest. 2002;121:774–781.PubMedCrossRef
17.
Zurück zum Zitat Abe A, Inoue K, Tanaka T, et al. Quantitation of hepatitis B virus genomic DNA by real-time detection PCR. J Clin Microbiol. 1999;37:2899–2903.PubMed Abe A, Inoue K, Tanaka T, et al. Quantitation of hepatitis B virus genomic DNA by real-time detection PCR. J Clin Microbiol. 1999;37:2899–2903.PubMed
18.
Zurück zum Zitat O’Connor HJ, Axon AT. Gastrointestinal endoscopy: infection and disinfection. Gut. 1983;24:1067–1077.PubMedCrossRef O’Connor HJ, Axon AT. Gastrointestinal endoscopy: infection and disinfection. Gut. 1983;24:1067–1077.PubMedCrossRef
19.
Zurück zum Zitat Langenberg W, Rauws EA, Oudbier JH, Tytgat GN. Patient-to-patient transmission of Campylobacter pylori infection by fiberoptic gastroduodenoscopy and biopsy. J Infect Dis. 1990;161:507–511.PubMedCrossRef Langenberg W, Rauws EA, Oudbier JH, Tytgat GN. Patient-to-patient transmission of Campylobacter pylori infection by fiberoptic gastroduodenoscopy and biopsy. J Infect Dis. 1990;161:507–511.PubMedCrossRef
20.
Zurück zum Zitat Dwyer DM, Klein EG, Istre GR, Robinson MG, Neumann DA, McCoy GA. Salmonella newport infections transmitted by fiberoptic colonoscopy. Gastrointest Endosc. 1987;33:84–87.PubMedCrossRef Dwyer DM, Klein EG, Istre GR, Robinson MG, Neumann DA, McCoy GA. Salmonella newport infections transmitted by fiberoptic colonoscopy. Gastrointest Endosc. 1987;33:84–87.PubMedCrossRef
24.
Zurück zum Zitat Hambrick D 3rd. Reprocessing of single-use endoscopic biopsy forceps and snares. One hospital’s study. Gastroenterol Nurs. 2001;24:112–115.PubMedCrossRef Hambrick D 3rd. Reprocessing of single-use endoscopic biopsy forceps and snares. One hospital’s study. Gastroenterol Nurs. 2001;24:112–115.PubMedCrossRef
26.
Zurück zum Zitat Deprez PH, Horsmans Y, Van Hassel M, Hoang P, Piessevaux H, Geubel A. Disposable versus reusable biopsy forceps: a prospective cost evaluation. Gastrointest Endosc. 2000;51:262–265. doi:S0016510700262603.PubMedCrossRef Deprez PH, Horsmans Y, Van Hassel M, Hoang P, Piessevaux H, Geubel A. Disposable versus reusable biopsy forceps: a prospective cost evaluation. Gastrointest Endosc. 2000;51:262–265. doi:S001651070026260​3.PubMedCrossRef
27.
Zurück zum Zitat Rizzo J, Bernstein D, Gress F. A performance, safety and cost comparison of reusable and disposable endoscopic biopsy forceps: a prospective, randomized trial. Gastrointest Endosc. 2000;51:257–261. doi:S0016510700514291.PubMedCrossRef Rizzo J, Bernstein D, Gress F. A performance, safety and cost comparison of reusable and disposable endoscopic biopsy forceps: a prospective, randomized trial. Gastrointest Endosc. 2000;51:257–261. doi:S001651070051429​1.PubMedCrossRef
28.
Zurück zum Zitat Wolfsen HC. Advantages of reusable accessories. Gastrointest Endosc Clin N Am. 2000;10:349–359.PubMed Wolfsen HC. Advantages of reusable accessories. Gastrointest Endosc Clin N Am. 2000;10:349–359.PubMed
29.
Zurück zum Zitat Yang R, Ng S, Nichol M, Laine L. A cost and performance evaluation of disposable and reusable biopsy forceps in GI endoscopy. Gastrointest Endosc. 2000;51:266–270. doi:S0016510700499414.PubMedCrossRef Yang R, Ng S, Nichol M, Laine L. A cost and performance evaluation of disposable and reusable biopsy forceps in GI endoscopy. Gastrointest Endosc. 2000;51:266–270. doi:S001651070049941​4.PubMedCrossRef
Metadaten
Titel
Comparison of Sterilization of Reusable Endoscopic Biopsy Forceps by Autoclaving and Ethylene Oxide Gas
verfasst von
Jai Hoon Yoon
Byung Chul Yoon
Hang Lak Lee
Jun Kyu Lee
Yong-Tae Kim
Dong Ho Lee
Il Ju Choi
Don Haeng Lee
Dong Hee Kim
Publikationsdatum
01.02.2012
Verlag
Springer US
Erschienen in
Digestive Diseases and Sciences / Ausgabe 2/2012
Print ISSN: 0163-2116
Elektronische ISSN: 1573-2568
DOI
https://doi.org/10.1007/s10620-011-1884-7

Weitere Artikel der Ausgabe 2/2012

Digestive Diseases and Sciences 2/2012 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Mehr Lebenszeit mit Abemaciclib bei fortgeschrittenem Brustkrebs?

24.05.2024 Mammakarzinom Nachrichten

In der MONARCHE-3-Studie lebten Frauen mit fortgeschrittenem Hormonrezeptor-positivem, HER2-negativem Brustkrebs länger, wenn sie zusätzlich zu einem nicht steroidalen Aromatasehemmer mit Abemaciclib behandelt wurden; allerdings verfehlte der numerische Zugewinn die statistische Signifikanz.

ADT zur Radiatio nach Prostatektomie: Wenn, dann wohl länger

24.05.2024 Prostatakarzinom Nachrichten

Welchen Nutzen es trägt, wenn die Strahlentherapie nach radikaler Prostatektomie um eine Androgendeprivation ergänzt wird, hat die RADICALS-HD-Studie untersucht. Nun liegen die Ergebnisse vor. Sie sprechen für länger dauernden Hormonentzug.

„Überwältigende“ Evidenz für Tripeltherapie beim metastasierten Prostata-Ca.

22.05.2024 Prostatakarzinom Nachrichten

Patienten mit metastasiertem hormonsensitivem Prostatakarzinom sollten nicht mehr mit einer alleinigen Androgendeprivationstherapie (ADT) behandelt werden, mahnt ein US-Team nach Sichtung der aktuellen Datenlage. Mit einer Tripeltherapie haben die Betroffenen offenbar die besten Überlebenschancen.

So sicher sind Tattoos: Neue Daten zur Risikobewertung

22.05.2024 Melanom Nachrichten

Das größte medizinische Problem bei Tattoos bleiben allergische Reaktionen. Melanome werden dadurch offensichtlich nicht gefördert, die Farbpigmente könnten aber andere Tumoren begünstigen.

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.