Concordance analysis of paired cancer antigen (CA) 15-3 and 27.29 testing
verfasst von:
David C. Lin, Jonathan R. Genzen
Erschienen in:
Breast Cancer Research and Treatment
|
Ausgabe 1/2018
Einloggen, um Zugang zu erhalten
Abstract
Purpose
Cancer antigens (CA) 15-3 and 27.29 are used in the clinical management of many breast cancer patients. Given that immunoassays for CA 15-3 and CA 27.29 target epitopes on the same glycoprotein—Mucin 1 (MUC1)—the present analysis was conducted to evaluate the potential concordance of tumor marker results when both tests were ordered by providers on the same specimens.
Methods
A retrospective limited dataset of paired CA 15-3 (Roche Diagnostics) and CA 27.29 (Siemens Diagnostics) test results was obtained from a national clinical reference laboratory. Concordance according to reference interval (RI) status and percent (%) change between consecutive test results was analyzed.
Results
37,652 paired results from 12,470 distinct patients were obtained. The correlation between CA 15-3 and CA 27.29 results was high (correlation coefficient: Pearson, 0.967), although across the dataset a significant difference between CA 15-3 and CA 27.29 results was observed (P < 0.05). RI concordance between CA 15-3 and CA 27.29 results was observed in 93.7% of pairs (35,280 of 37,652). Correlation was also observed in the % change of CA 15-3 and CA 27.29 results between consecutive specimens for individual patients. Using doubling or halving thresholds (i.e., 100% increase or 50% decrease), concordance in % change was observed between CA 15-3 and CA 27.29 in approximately 90% of cases. Individual patient results trended similarly across both markers over time.
Conclusion
While generally concordant, CA 15-3 and CA 27.29 results should not be used interchangeably. The present report provides no evidence for added value in performing both tests routinely for individual patients.