Open Access 24.10.2021 | COVID-19 | Original Paper

Analytical performance of the point-of-care BIOSYNEX COVID-19 Ag BSS for the detection of SARS‐CoV‐2 nucleocapsid protein in nasopharyngeal swabs: a prospective field evaluation during the COVID-19 third wave in France

verfasst von: Frédéric Fitoussi, Serge Tonen-Wolyec, Natalio Awaida, Raphaël Dupont, Laurent Bélec

Erschienen in: Infection | Ausgabe 3/2022

Abstract

Background

The accuracy and reliability of rapid diagnostic tests are critical for monitoring and diagnosing SARS-CoV-2 infection in the general population. This study aimed to evaluate the analytical performance of the BIOSYNEX COVID-19 Ag BSS (Biosynex Swiss SA, Fribourg, Switzerland) antigen rapid diagnostic test (BIOSYNEX Ag-RDT), which targets the SARS-CoV-2 N-nucleocapsid protein for the diagnosis of COVID-19. The Ag-RDT was compared with a real-time RT-PCR (rtRT-PCR) as gold standard for performance measurement.

Methods

Two nasopharyngeal flocked swabs were prospectively collected simultaneously in March and April 2021 from 967 individuals aged ≥ 18 years tested for SARS-CoV-2 in two private laboratories, Paris, France.

Results

Overall, the Ag-RDT demonstrated high sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) of 81.8%, 99.6%, 96.6%, and 97.5%, respectively. The agreement (97.0%), reliability assessed using Cohen’s κ-coefficient (0.87), and accuracy evaluated using Youden index (J) (81.6%) in detecting SARS-CoV-2 were high. The analytical performance of the Ag-RDT remained high when there was significant viral shedding (i.e., N gene Ct values ≤ 33 on reference RT-PCR). The sensitivity was only 55.2% in case of low or very low viral excretion (Ct > 33).

Conclusions

The BIOSYNEX Ag-RDT is a promising, potentially simple diagnostic tool, especially in symptomatic COVID-19 patients with substantial viral excretion in the nasopharynx.
Hinweise

Supplementary Information

The online version contains supplementary material available at https://​doi.​org/​10.​1007/​s15010-021-01723-5.
Abkürzungen
Ag-RDT
Antigen-detecting rapid diagnostic test
BIOSYNEX  Ag-RDT
BIOSYNEX COVID-19 Ag BSS
CDC
Centers for Disease Control and Prevention
COVID-19
Coronavirus disease 2019
NAAT
Nucleic acid amplification test
rtRT‐PCR
real-time reverse transcription-polymerase chain reaction
SARS-CoV-2
Severe acute respiratory syndrome coronavirus 2
WHO
World Health Organization

Introduction

The 2019 coronavirus pandemic (COVID-19) continues to spread worldwide. The effective isolation and early treatment of patients infected by the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) require rapid, accurate, and straightforward diagnostic tools.
While currently recommended nucleic acid amplification tests (NAAT), such as real-time reverse transcription-polymerase chain reaction (rtRT-PCR) assays, remain the gold standard cornerstone for the diagnosis of SARS‐CoV‐2 infection [1, 2], viral antigens can be detected using immunological methods [24]. Indeed, conducting rtRT-PCR is expensive, time-consuming, and requires special equipment and qualified operators. Point-of-care antigen-detecting rapid diagnostic tests (Ag-RDT) constitute simple and less expensive alternative tests [3]. Ag-RDT relies on direct detection of SARS-CoV-2 viral proteins in nasal swabs and other respiratory secretions. The N-nucleocapsid protein is frequently targeted because of its relative abundance and conserved structure, or other viral proteins such as the spike protein [4]. Most Ag-RDTs rely on sandwich catching using anti-SARS-CoV-2 monoclonal antibodies to detect viral antigens in the simple-to-use lateral flow immunoassay format allowing results in < 30 min. However, significant variability has been reported about their diagnostic performance and a lack of external validation for many available tests, which still require clinical validation [59].
Our study aimed to evaluate the qualitative membrane-based immunochromatographic BIOSYNEX COVID-19 Ag BSS Ag-RDT (Biosynex Swiss SA, Freiburg, Switzerland; reference SW40006; abbreviated by BIOSYNEX Ag-RDT) using monoclonal antibodies detecting SARS-CoV-2 N-nucleocapsid protein to diagnose COVID‐19 from prospectively collected nasopharyngeal secretion samples in adults living in the Paris region throughout the third wave of the COVID-19 epidemic in France.

Materials and methods

Rapid antigen test

The BIOSYNEX Ag-RDT consists of a reaction membrane and three buffers (sample, reagent, and absorbent). The reagent buffer contains colloidal gold particles conjugated with monoclonal antibodies directed against the N protein of SARS-CoV-2. Secondary antibodies against the N protein are fixed on the reaction membrane. The manufacturer’s instructions were followed to conduct the test by mixing nasopharyngeal secretions with 300 µl of dilution buffer in a tube. After 1 min, four drops were added to the well on the cassette.
If SARS-CoV-2 antigens are present in the sample, the complexes between the anti-SARS-CoV-2 conjugate and the virus are captured by anti-SARS-CoV-2 monoclonal antibodies specific to the test line area (T). The lack of the T line indicates that the result is negative. A red line appears in the control line area (C) to serve as a procedural control, indicating that the correct sample volume has been added and the membrane has played its role. Reading is carried out after 15 min.

Study population and procedures

During the third wave of the COVID-19 epidemic (March and April 2021), two sites had been used to consecutively collect paired nasopharyngeal swabs. Site A was the Centre Cardiologique du Nord, Saint-Denis, France. Site B was the Laboratoire Paris XV, Paris, France. Participants aged ≥ 18 years and those consenting to undergo two nasopharyngeal swabs for rtRT-PCR and Ag-RDT were included. All participants were given a questionnaire that recorded demographic information (sex and age), reasons for testing, and current and past 14-day symptoms in symptomatic patients. Suggestive symptoms of COVID-19 were headache, fatigue, fever, or upper or lower respiratory symptoms. Asymptomatic individuals were defined as those not reporting any of these symptoms. At both sites, a health care professional first collected nasopharyngeal secretions in one nostril, using the swab provided in the BIOSYNEX Ag-RDT. A second nasopharyngeal swab in the other nostril served as specimen for the rtRT-PCR. The COVID-19 antigen rapid test was performed immediately on-site using the Ag-RDT following the manufacturer’s instructions. The other nasopharyngeal swabs were stored in physiological saline (NaCl 0.9%) (1000 μL) at + 4 °C and analyzed within 24–48 h by the reference rtRT-PCR.

Molecular detection of SARS-CoV-2

The multiplex real-time PCR Novel Coronavirus (2019-nCoV) Real-Time Multiplex RT-PCR Kit (Detection for 3 Genes) (Liferiver & Shanghai ZJ Bio-Tech Co., Ltd, Shanghai, China) was the reference multiplex molecular detection of SARS-CoV-2 RNA. Individual cycle threshold (Ct) values for each target gene (E, N and RdRP). According to manufacturer’s recommendations, samples with Ct values ≤ 41 for three or two gene targets were considered as positive; those with Ct values ≤ 41 for only one gene target were possibly positive; samples with Ct value > 41 for the three gene targets were negative.
The Ct values of the N gene in the RT-PCR reference were chosen for stratification of viral load in clinical samples because the Ag-RDT detects the SARS-CoV-2 N-nucleocapsid protein.

Statistical analyses

Collected data were analyzed using IBM® SPSS® Statistics 20 software (IBM, SPSS Inc, Armonk, New York, USA). Results of quantitative variables were expressed as medians; however, the proportion with their 95% confidence interval (CI) assessed according to the Wilson score bounds were estimated for categorical variables [10]. Comparisons were carried out using Pearson’s Chi square test or Fisher’s exact test based on validity conditions. The PPV and NPV were calculated according to Bayes’ formulas, taking into account the officially reported prevalence of SARS-CoV-2 RNA positivity in symptomatic patients in the Paris region on 12th April 2021, e.g., around the peak of the third wave epidemic in France (Santé publique France 2021; https://​www.​santepubliquefra​nce.​fr/​).

Ethics statement

The purpose of the study was to clinically evaluate the continuous quality improvement program and performance evaluation of COVID-19 management measures following the National Medical-Biological Laboratory Accreditation [11]. The data set was anonymous and contained no identifiable personal health information.

Results

Paired swab samples were obtained from 967 participants, including 741 from site A and 226 from site B (Table 1). Participants ranged in age from 18 to 95 (median = 34 years). The main reasons for testing were air travel (35.6%), contact-case exposure of an individual infected with SARS-CoV-2 (35.1%), suspected COVID-19 (n = 212, 21.9%), preoperative assessment (4.4%), and control of SARS-CoV-2 infection in the previous 30 days (3.0%). The majority (722/967, 74.7%) of included persons were asymptomatic, while a minority (245/967, 25.3%) reported at least one COVID-19-related symptom [including 212 suspected COVID-19 cases, 29 (8.5%) contact cases, 3 (0.9%) travelers, and 1 (3.0%) patient with a recent history of COVID-19]. The median symptom duration before sampling was four days (range, 0–20 days). All comparisons between positive and negative Ag-RDT and rtRT-PCR testing results for both sites and all other variables did not achieve statistical significance (not shown).
Table 1
Characteristics of persons providing paired upper respiratory swab (n = 967) for real-time reverse transcription-polymerase chain reaction (rtRT-PCR) testing and BIOSYNEX COVID-19 Ag BSS rapid diagnostic testing for SARS-CoV-2 at two private laboratory sites, by test results, Paris, France, spring 2021
Characteristics
To number of persons (column %)
Number of persons (row%)
rtRT-PCR negative
rtRT-PCR positive
Antigen test negative
Antigen test positive
Total
967 (100)
819 (84.7)
148 (15.3)
844 (87.3)
123 (12.7)
Testing site
 A
741 (74.6)
633 (85.4)
108 (14.6)
651 (87.9)
90 (12.1)
 B
226 (25.4)
186 (82.4)
40 (17.6)
193 (85.4)
33 (14.6)
Sex
 Female
498 (51.5)
429 (86.2)
69 (13.8)
438 (88.0)
60 (12.0)
 Male
469 (48.5)
390 (83.2)
79 (16.8)
406 (86.6)
63 (13.4)
Age group, years
 18–49
740 (76.4)
629 (85.0)
111 (15.0)
647 (87.4)
93 (12.6)
 50–64
157 (16.3)
129 (82.2)
28 (17.8)
133 (84.7)
24 (15.3)
  ≥ 65
70 (7.3)
61 (87.1)
9 (12.9)
64 (91.4)
6 (8.6)
 Median age (range)
34 (18–83)
34 (18–83)
32 (18–82)
34 (18–83)
37 (18–82)
Any symptoms
  ≥ 1
245 (25.3)
165 (67.4)
80 (32.6)
168 (68.6)
77 (31.4)
 None
722 (74.7)
654 (90.6)
68 (9.4)
676 (93.6)
46 (6.4)
Days from onset of symptoms
 0–3
107 (43.3)
38 (35.5)
69 (64.5)
41(38.3)
66 (61.7)
 4–7
122 (49.8)
48 (39.4)
74 (60.6)
69 (56.6)
53 (43.4)
  > 7
16 (6.9)
11 (68.7)
5 (31.3)
12 (75.0)
4 (25.0)
 Median (range)
4 (0–20)
4 (0–20)
3 (0–15)
4 (0–20)
3 (0–10)
Air travel intention
 Yes
344 (35.6)
330 (95.9)
14 (4.1)
337 (98.0)
7 (2.0)
 No
623 (64.4)
489 (78.5)
134 (21.5)
507 (81.4)
116 (18.6)
Contact case (exposure to a diagnosed COVID-19 case)
 Yes
340 (35.1)
282 (82.9)
58 (17.1)
293 (86.2)
47 (13.8)
 No/unknown
627 (64.9)
537 (85.7)
90 (14.3)
551 (87.9)
76 (12.1)
Suspected COVID-19
 Yes (≥ 1 COVID-19 symptoms
212 (21.9)
149 (70.3)
63 (29.7)
151 (71.2)
61 (28.8)
 No
755 (78.1)
670 (88.8)
85 (11.2)
693 (91.8)
62 (8.2)
Preoperative assessment
 Yes
42 (4.4)
38 (90.5)
4 (9.5)
40 (95.2)
2 (2.8)
 No
925 (95.6)
781 (84.4)
144 (15.6)
804 (86.9)
121 (13.1)
Control of SARS-CoV-2 positive test results in past 30 days
 Yes
29 (3.0)
20 (68.9)
9 (31.1)
21 (72.4)
8 (27.6)
 No/unknown
938 (97.0)
799 (85.2)
139 (14.8)
823 (87.7)
115 (12.3)
Among the 148 positive samples using the gold standard rtRT-PCR, 146 were positive for the three gene targets, and two were positive for only E and N genes. The mean ± SD of the Ct values were 26.1 ± 4.4 arbitrary units (a.u.) for the E gene, 26.5 ± 5.0 a.u. for the RdRP gene, and 26.9 ± 5.1 a.u. for the N gene.
The vast majority (114/123, 92.7%) of positive results were visible in the window of the cassette of the Ag-RDT within the first 5 min. Table 2 shows the test results and primary performance characteristics of the BIOSYNEX Ag-RDT compared with the reference rtRT‐PCR in the study population according to COVID-19-compatible symptoms. Using rtRT-PCR as the standard, three false-positive BIOSYNEX Ag-RDT results occurred among specimens from asymptomatic individuals (n = 2) or symptomatic patients (n = 1). Of the 148 rtRT-PCR positive results, 27 (18.2%) were false-negative BIOSYNEX Ag-RDT (23 specimens from asymptomatic persons and 4 specimens from symptomatic patients). Overall, the BIOSYNEX Ag-RDT showed high sensitivity (81.8%), specificity (99.6%), PPV (96.6%), and NPV (97.5%). Among symptomatic patients, sensitivity was 95.0%, specificity was 99.4%, PPV was 95.6%, and NPV was 96.3% (Table 2). Within 7 days from symptom beginning, the BIOSYNEX Ag-RDT showed a sensitivity of 96.6%, a specificity of 99.4%, whereas the PPV and NPV were 95.7% and 99.4%, respectively.
Table 2
Test results and performances characteristics of the BIOSYNEX COVID-19 Ag BSS rapid diagnostic test compared with real-time reverse transcription-polymerase chain reaction (rtRT-PCR) for SARS-CoV-2 testing among asymptomatic and symptomatic persons at two private laboratory sites, by test results, Paris, France, spring 2021
Results and performances
rtRT-PCR (number of test, %)
Total
Positive
Negative
BIOSYNEX COVID-19 Ag BSS results
 All participants (n = 967)
  Positive
121 (12.5)
3 (0.3)
124 (12.8)
  Negative
27 (2.8)
816 (84.4)
843 (87.2)
  Total
148 (21.2)
819 (78.8)
967 (100)
 Asymptomatic (n = 722)
  Positive
45 (4.2)
2 (0.3)
47 (6.5)
  Negative
23 (5.2)
652 (90.3)
675 (95.5)
  Total
68 (9.4)
654 (90.6)
722 (100)
 Symptomatic (≥ 1 symptom) (n = 245)
  Positive
76 (31.0)
1 (0.4)
77 (31.4)
  Negative
4 (1.6)
164 (67.0)
168 (68.6)
  Total
80 (32.6)
165 (67.4)
245 (100)
 Symptomatic (≤ 7 days from symptom onset) (n = 229)
  Positive
72 (31.4)
1 (0.5)
73 (31.9)
  Negative
3 (1.3)
153 (66.8)
156 (68.1)
  Total
75 (32.7)
154 (67.3)
229 (100)
BIOSYNEX COVID-19 Ag BSS performances (%, 95%CI)
 All participants
  Sensitivity
81.8 (79.2–84.1)
  Specificity
99.6 (98.9–99.8)
  PPV£
96.6 (95.3–97.6)
  NPV£
97.5 (96.3–98.3)
 Asymptomatic
  Sensitivity
79.4 (76.3–82.2)
  Specificity
99.7 (98.9–99.9)
  PPV
97.3 (95.8–98.2)
  NPV
97.2 (95.7–98.2)
 Symptomatic
  Sensitivity
95.6 (92.2–97.5)
  Specificity
99.3 (97.2–99.8)
  PPV
95.6 (92.2–97.5)
  NPV
99.3 (97.2–99.8)
 Symptomatic (≤ 7 days from onset)
  Sensitivity
96.0 (92.6–97.9)
  Specificity
99.4 (97.3–99.9)
  PPV
95.7 (92.2–97.7)
  NPV
99.4 (97.3–99.9)
CI confidence interval; NVP negative predictive value; PPV positive predictive value
£PPV and NPV were calculated according to the Bayes’s formulae, by taking into account the official reported prevalence of SARS-CoV-2-RNA positivity in COVID-19-suspected patients in Paris’s area, France, of 12.2% on 12th April 2021 [Santé publique France 2021; https://​www.​santepubliquefra​nce.​fr/​]
The Table 3 shows the analytical results based on the level of viral excretion assessed by the N gene Ct values using the reference rtRT‐PCR. Overall, the BIOSYNEX Ag-RDT showed high agreement (97.0%), reliability using Cohen’s κ coefficient (0.87), and accuracy using Youden’s J index (81.6%) to detect SARS-CoV-2.
Table 3
Analytical performances of the BIOSYNEX COVID-19 Ag BSS rapid diagnostic test for the qualitative detection of the N protein of SARS-CoV-2 using 967 prospectively collected nasopharyngeal swab samples by reference rtRT-PCR#, according to their N gene Ct values
N gene Ct (median; range)
N
BIOSYNEX COVID-19 Ag BSS§
FN (n)
TP (n)
Sensitivitya (% [95% CI])µ
Specificitya (% [95% CI])
Agreementb
Concordancec
Youden’s J indexd
PPVe% [95% CI])
NPVe (% [95% CI])
Detectable N gene Ct£ by rtRT-PCR#
 ≤ 20
17.9 (13.9–20.0)
35
0
35
100 (99.6–100)
99.6 (98.9–99.8)
99.6 (98.9–99.8)
0.95 (0.93–0.96)
99.6 (98.9–99.8))
97.2 (95.9–98.1)
100 (99.6–100)
21–33
27.2 (20.1–33.0)
84
14
70
83.3 (80.7–85.6)
99.6 (98.9–99.8)
98.1 (97.0–98.8)
0.65 (0.62–0.68)
82.9 (80.3–85.2)
96.7 (95.3–97.7)
97.7 (96.5–98.5)
 > 33–41
35.9 (34.0–39.2)
29
13
16
55.2 (51.8–58.5)
99.6 (98.9–99.8)
98.1 (96.9–98.8)
0.65 (0.62–0.68)
54.8 (51.4–58.1)
95.0 (93.3–96.3)
94.1 (92.3–95.5)
All positive Ct values
26.9 (13.9–39.2)
148
27
121
81.8 (79.2–84.1)
99.6 (98.9–99.8)
96.9 (95.9–97.8)
0.87 (0.85–0.89)
81.4 (78.8–83.7)
96.6 (95.3–97.6)
97.5 (96.3–98.3)
Ct cycle threshold; FN false negative; NPV negative predictive value; PPV positive predictive value; rtRT-PCR real-time reverse transcription-polymerase chain reaction; TP True positive
§Paired nasopharyngeal samples in each nostril were collected with a flocked swab for each volunteer patients by trained healthcare personnel (nurses, doctors or biologists). The collection of the two simultaneous samples was always carried out by the same operator. Molecular testing as well COVID-19 antigen detection was carried out on fresh samples.
aThe results of SARS-CoV-2 RNA detection using the multiplex rtRT-PCR were used as the reference standard to estimate the sensitivity and specificity of the study Ag-RDT, with corresponding 95% CI.
bAgreement = TP + TN/TP + FP + TN + FN, expressed in percentage.
cThe Cohen’s κ coefficient calculation was used to estimate the concordance [43] and interpreted according the Landis and Koch scale [44], as follows: < 0 as indicating no agreement, 0–0.20 as slight, 0.21–0.40 as fair, 0.41–0.60 as moderate, 0.61–0.80 as substantial, and 0.81–1 as almost perfect concordance.
dThe accuracy of the test BIOSYNEX COVID-19 Ag BSS to correctly diagnose SARS-CoV-2 infection was estimated by Youden’s J index (J = sensitivity + specificity − 1) [45].
ePPV and NPV were calculated according to the Bayes’s formulae, by taking into account the official reported prevalence of SARS-CoV-2-RNA positivity in COVID-19-suspected patients in Paris’s area, France, of 12.2% on 12th April 2021 [Santé publique France 2021; https://​www.​santepubliquefra​nce.​fr/​].
µ95% confidence intervals in brackets were calculated using the Wilson score bounds.
£The Ct values of N gene detection by the reference Liferiver rtRT-PCR were used to classify nasopharyngeal samples according to their level of SARS-CoV-2 RNA excretion; Ct of 20 and 33 were taken as thresholds of very high and high SARS-CoV-2 RNA excretion, respectively, as previously stated [36, 39, 4041] ; viral loads with Ct > 33 are considered low and correspond to moderate or very low viral excretion [36, 3941]. Conversely, samples with Ct ≤ 33 have a significant SARS-CoV-2 viral load, as in individuals symptomatic for COVID-19 or contagious. Ct values ≤ 20 indicate very high viral shedding [3941]
#The CE IVD-marked Novel Coronavirus (2019-nCoV) Real-Time Multiplex RT-PCR Kit (Detection for 3 Genes) (Liferiver & Shanghai ZJ Bio-Tech Co., Ltd, Shanghai, China) constituted the reference multiplex rtRT-PCR for SARS-CoV-2 RNA detection. This assay can detect three coronavirus target genes simultaneously, including the SARS-like (SARS-CoV-2, SARS-CoV, bat SARS-like coronavirus) conserved region of envelope protein gene (E), RNA-dependent RNA polymerase gene (ORF1ab of RdRP gene), and nucleocapsid protein gene (N), using reverse transcription. Nucleic acid extraction was performed from 300 μL elution volume of a nasopharyngeal flocked swab sample, using an EX3600 extractor (Liferiver & Shanghai ZJ Bio-Tech Co.), according to the manufacturer’s instructions, and finally eluted in 50 μL (final volume). SARS-CoV-2 was detected in 5 μL of extracted RNA. Real-time PCR was conducted using CFX96™ Real-Time PCR Detection System (Bio-Rad Laboratories, Hercules, CA, USA) according to the manufacturer’s instructions. The experiment and analysis of the results were performed according to the manufacturer’s protocol.
In case of high or very high viral loads (Ct ≤ 33), the BIOSYNEX Ag-RDT had a good analytical performance (sensitivities between 83.3% and 100.0%, specificities of 99.8%, PPV between 98.3% and 98.6%, and NPV between 97.7% and 100.0%). In case of low or very low viral loads (Ct > 33), the sensitivity of the BIOSYNEX Ag-RDT had reduced analytical performance (sensitivity of only 55.2%), while its specificity remained high (98.8%). Similar observations were made when the Ct values of the E or ORF1ab gene targets were chosen for stratification of viral load in clinical samples (data not shown).
Finally, the sensitivity of the BIOSYNEX Ag-RDT varied among the five participant groups as follows: (i) travel: 50.0% (7/14), (ii) contact-case exposure: 81.0% (47/58), (iii) preoperative assessment: 50.0% (2/4); (iv) suspected COVID − 19: 96.8% (61/63), and (v) control of SARS-CoV-2 positive test results in the last 30 days: 88.9% (8/9).

Discussion

We evaluated the analytical performance of the novel point-of-care BIOSYNEX Ag-RDT compared to multiplex rtRT-PCR as gold standard for detecting SARS-CoV-2 RNA in a real-life setting. In this study, the sensitivity of the BIOSYNEX Ag-RDT was lower among specimens from asymptomatic persons (79.4%) than among specimens from symptomatic patients (95.0%). It was high in patients with suspected COVID-19 (96.8%). Specificity (> 99.0%) was high in specimens from both asymptomatic individuals and symptomatic patients. The prevalence of SARS-CoV-2 RNA-positive rt-RT-PCR results in this population was 15.3% overall, 9.4% for asymptomatic individuals, and 32.6% for symptomatic patients. The estimated PPVs and NPVs of the BIOSYNEX Ag-RDT were elevated in all groups of participants. However, administering the Ag-RDT in low prevalence settings will likely result in lower predictive values. In the event of significant viral excretion (i.e., N gene Ct values below 33 based on reference rtRT-PCR), the BIOSYNEX Ag-RDT showed high sensitivity (from 83.3% to 100.0%) and specificity (> 99.0%) for SARS-CoV-2 RNA detection. Concordance, reliability, as well as accuracy were great with the reference assay and PPVs and NPVs above 97.0%. However, the sensitivity of the study Ag-RDT dropped to 55.2% with low or very low viral shedding (Ct > 33). Together, these observations demonstrated the high analytical performance of the BIOSYNEX Ag-RDT. This performance made it suitable for use as point-of-care Ag-RDT in various hospital and non-hospital settings where a rapid diagnosis of SARS-CoV-2 is necessary. Although less sensitive than RT-PCR, the BIOSYNEX Ag-RDT could be beneficial by obtaining quick results, ease of use, and independence from existing laboratory structures. Testing criteria focusing on patients during the early onset of symptoms could further increase its diagnostic value.
The sensitivity of the BIOSYNEX Ag-RDT was 81.8% overall, and the positive detection rate was comparable to the rtRT-PCR in the majority (88.2%) of subjects with Ct ≤ 33. False-negative test results of 12/14 (85.7%) subjects with significant viral excretion (Ct ≤ 33) were asymptomatic, although conflicting evidence exists about the relationship between symptom severity and viral shedding [12]. False-positive test results were rarely observed, providing 99.6%-specificity, exceeding the performance recommended by the World Health Organization (WHO) [13]. False-positive results have been reported as well in other antigen tests [1416]. False positivity could be associated with high viscosity of tested specimen samples as well as interference with mucosal antibodies [17].
Finally, the BIOSYNEX Ag-RDT meets the current WHO criteria which stipulate that Ag-RDTs for SARS-CoV-2 antigen detection must have a sensitivity greater than 80% and a specificity greater than 97% (97%–100%) [13]. Furthermore, analytical performances comparable to those in our study Ag-RDT were previously reported for some Ag-RDTs in lateral flow immunoassay format [7, 9, 14, 1828], while several studies have reported much lower sensitivity levels contrasting with consistently high specificity [3, 2934]. In addition, the BIOSYNEX Ag-RDT also fulfilled the current recommendations of the French High Authority of Health (Haute Autorité de santé, Saint-Denis, France) for a screening Ag-RTD stating that, at minimum, Ag-RDTs would need to correctly identify significant proportions of symptomatic patients (sensitivity ≥ 80%) as well as asymptomatic individuals (sensitivity ≥ 50%) and have high specificity (≥ 90%) [35].
We analyzed our results based on the estimated viral load in SARS-CoV-2 in the samples. There is an ongoing debate about the Ct value corresponding to the threshold of infectivity (i.e., patient considered as contagious) [7, 36, 37]. La Scola et al. found that patients with Ct values > 33 are not infectious because of the low number of positive cultures [38]. The Centers for Disease Control and Prevention (CDC), Atlanta, USA, propose a Ct cut-off value of 33 as a marker for contagiousness [39], and stress that Ct values ≤ 20 correspond to very high viral excretion [7, 36, 40, 41]. Our results confirm that the analytical performances of the BIOSYNEX Ag-RDT were much better in specimens with a high viral load. These observations demonstrate the capability of the BIOSYNEX Ag-RDT as a rapid rule-in test for COVID-19 with samples at high viral load in symptomatic patients, for example, and raise caution about its use as a singular rule-out test, especially in samples with lower viral loads.
Our study has several strengths. All samples were collected from one nasopharynx with flocked swabs, optimal for evaluating Ag-RDT clinical performances in our study. The Ag-RDT and reference rtRT-PCR were carried out in parallel. The study population included various situations outside the hospital setting, with mostly young adults without comorbidities who had typical and mild COVID-19 symptoms when being symptomatic.
The study presents also some limitations. Participants may have inadvertently reported general, non-specific symptoms as COVID-19 compatible symptoms. This investigation evaluated the BIOSYNEX Ag-RDT; the results presented here cannot be generalized to other agencies-authorized SARS-CoV-2 antigen tests. Otherwise, the CDC clarified that Ct values using the rtRT-PCR platform is not a quantitative measure of viral burden in clinical samples and cannot be used to assess whether a person is infectious [42]. Consequently, our stratification of samples according to Ct values of the N gene does not necessarily reflect the actual infectivity of the participants. Finally, higher rate of asymptomatic persons in the study (with lower virus level) could have resulted in decreased sensitivity of the BIOSYNEX Ag-RDT.

Conclusion

The BIOSYNEX Ag-RDT demonstrated high specificity and sufficient sensitivity for detecting SARS-CoV-2. Given the simple procedures and short turnaround time for this test, it is a promising option as an alternative diagnostic modality, especially in symptomatic COVID-19 patients. The test may also be used to test asymptomatic individuals suspicious of exposure to SARS-CoV-2 and as part of a population-level mass screening.

Acknowledgements

The authors thank all nurses and staff at the Centre Cardiologique du Nord, Saint-Denis, and Laboratoire Paris XV, Paris, and the subjects for their willingness to participate in the study. We are grateful to Biosynex, Strasbourg, France, for providing the tests for the study.

Declarations

Conflict of interest

The authors report no conflicts of interest. The authors alone are responsible for the content and the writing of the paper.

Ethics approval

This study was approved by the local scientific committee of the Parc de l’Innovation, Strasbourg, France (03 March 2021).
All authors approved the submission of the manuscript for publication. All participants consented to take part in this study.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Unsere Produktempfehlungen

Neuer Inhalt

e.Med Interdisziplinär

Kombi-Abonnement

Jetzt e.Med zum Sonderpreis bestellen!

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Jetzt bestellen und 100 € sparen!

e.Med Innere Medizin

Kombi-Abonnement

Mit e.Med Innere Medizin erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Innere Medizin, den Premium-Inhalten der internistischen Fachzeitschriften, inklusive einer gedruckten internistischen Zeitschrift Ihrer Wahl.

Jetzt bestellen und 100 € sparen!

Weitere Produktempfehlungen anzeigen
Anhänge

Supplementary Information

Below is the link to the electronic supplementary material.
Literatur
1.
Zurück zum Zitat Smithgall MC, Dowlatshahi M, Spitalnik SL, Hod EA, Rai AJ. Types of assays for SARS-CoV-2 testing: a review. Lab Med. 2020;51:e59–65.CrossRef Smithgall MC, Dowlatshahi M, Spitalnik SL, Hod EA, Rai AJ. Types of assays for SARS-CoV-2 testing: a review. Lab Med. 2020;51:e59–65.CrossRef
2.
Zurück zum Zitat Rai P, Kumar BK, Deekshit VK, Karunasagar I, Karunasagar I. Detection technologies and recent developments in the diagnosis of COVID-19 infection. Appl Microbiol Biotechnol. 2021;105(2):441–55.CrossRef Rai P, Kumar BK, Deekshit VK, Karunasagar I, Karunasagar I. Detection technologies and recent developments in the diagnosis of COVID-19 infection. Appl Microbiol Biotechnol. 2021;105(2):441–55.CrossRef
3.
Zurück zum Zitat Dinnes J, Deeks JJ, Berhane S, Taylor M, Adriano A, Davenport C, Dittrich S, Emperador D, Takwoingi Y, Cunningham J, Beese S, Domen J, Dretzke J, Ferrante di Ruffano L, Harris IM, Price MJ, Taylor-Phillips S, Hooft L, Leeflang MM, McInnes MD, Spijker R, Van den Bruel A; Cochrane COVID-19 Diagnostic Test Accuracy Group. Rapid, point-of-care antigen and molecular-based tests for diagnosis of SARS-CoV-2 infection. Cochrane Database Syst Rev. 2021;3:CD013705. https://doi.org/10.1002/14651858.CD013705.pub2.CrossRef Dinnes J, Deeks JJ, Berhane S, Taylor M, Adriano A, Davenport C, Dittrich S, Emperador D, Takwoingi Y, Cunningham J, Beese S, Domen J, Dretzke J, Ferrante di Ruffano L, Harris IM, Price MJ, Taylor-Phillips S, Hooft L, Leeflang MM, McInnes MD, Spijker R, Van den Bruel A; Cochrane COVID-19 Diagnostic Test Accuracy Group. Rapid, point-of-care antigen and molecular-based tests for diagnosis of SARS-CoV-2 infection. Cochrane Database Syst Rev. 2021;3:CD013705. https://​doi.​org/​10.​1002/​14651858.​CD013705.​pub2.CrossRef
4.
Zurück zum Zitat Li D, Li J. Immunologic testing for SARS-CoV-2 infection from the antigen perspective. J Clin Microbiol. 2020:JCM.02160–20. Li D, Li J. Immunologic testing for SARS-CoV-2 infection from the antigen perspective. J Clin Microbiol. 2020:JCM.02160–20.
7.
Zurück zum Zitat Favresse J, Gillot C, Oliveira M, Cadrobbi J, Elsen M, Eucher C, Laffineur K, Rosseels C, Van Eeckhoudt S, Nicolas JB, Morimont L, Dogné JM, Douxfils J. Head-to-head comparison of rapid and automated antigen detection tests for the diagnosis of SARS-CoV-2 infection. J Clin Med. 2021;10:E265.CrossRef Favresse J, Gillot C, Oliveira M, Cadrobbi J, Elsen M, Eucher C, Laffineur K, Rosseels C, Van Eeckhoudt S, Nicolas JB, Morimont L, Dogné JM, Douxfils J. Head-to-head comparison of rapid and automated antigen detection tests for the diagnosis of SARS-CoV-2 infection. J Clin Med. 2021;10:E265.CrossRef
8.
Zurück zum Zitat Fitzpatrick MC, Pandey A, Wells CR, Sah P, Galvani AP. Buyer beware: inflated claims of sensitivity for rapid COVID-19 tests. Lancet. 2021;397:24–5.CrossRef Fitzpatrick MC, Pandey A, Wells CR, Sah P, Galvani AP. Buyer beware: inflated claims of sensitivity for rapid COVID-19 tests. Lancet. 2021;397:24–5.CrossRef
9.
Zurück zum Zitat Schildgen V, Demuth S, Lüsebrink J, Schildgen O. Limits and opportunities of SARS-CoV-2 antigen rapid tests: an experienced-based perspective. Pathogens. 2021;10:E38.CrossRef Schildgen V, Demuth S, Lüsebrink J, Schildgen O. Limits and opportunities of SARS-CoV-2 antigen rapid tests: an experienced-based perspective. Pathogens. 2021;10:E38.CrossRef
10.
Zurück zum Zitat Newcombe RG. Two-sided confidence intervals for the single proportion: comparison of 362 seven methods. Stat Med. 1998;17:857–72.CrossRef Newcombe RG. Two-sided confidence intervals for the single proportion: comparison of 362 seven methods. Stat Med. 1998;17:857–72.CrossRef
12.
Zurück zum Zitat Magleby R, Westblade LF, Trzebucki A, Simon MS, Rajan M, Park J, Goyal P, Safford MM, Satlin MJ. Impact of SARS-CoV-2 viral load on risk of intubation and mortality among hospitalized patients with coronavirus disease 2019. Clin Infect Dis. 2020;ciaa851. Magleby R, Westblade LF, Trzebucki A, Simon MS, Rajan M, Park J, Goyal P, Safford MM, Satlin MJ. Impact of SARS-CoV-2 viral load on risk of intubation and mortality among hospitalized patients with coronavirus disease 2019. Clin Infect Dis. 2020;ciaa851.
13.
Zurück zum Zitat World Health Organization. Interim guidance. Antigen-detection in the diagnosis of SARS-CoV-2 infection using rapid immune-assays. 2020. (Accessed 04 May 2021). World Health Organization. Interim guidance. Antigen-detection in the diagnosis of SARS-CoV-2 infection using rapid immune-assays. 2020. (Accessed 04 May 2021).
14.
Zurück zum Zitat Chaimayo C, Kaewnaphan B, Tanlieng N, Athipanyasilp N, Sirijatuphat R, Chayakulkeeree M, Angkasekwinai N, Sutthent R, Puangpunngam N, Tharmviboonsri T, Pongraweewan O, Chuthapisith S, Sirivatanauksorn Y, Kantakamalakul W, Horthongkham N. Rapid SARS-CoV-2 antigen detection assay in comparison with real-time RT-PCR assay for laboratory diagnosis of COVID-19 in Thailand. Virol J. 2020;17:177.CrossRef Chaimayo C, Kaewnaphan B, Tanlieng N, Athipanyasilp N, Sirijatuphat R, Chayakulkeeree M, Angkasekwinai N, Sutthent R, Puangpunngam N, Tharmviboonsri T, Pongraweewan O, Chuthapisith S, Sirivatanauksorn Y, Kantakamalakul W, Horthongkham N. Rapid SARS-CoV-2 antigen detection assay in comparison with real-time RT-PCR assay for laboratory diagnosis of COVID-19 in Thailand. Virol J. 2020;17:177.CrossRef
15.
Zurück zum Zitat Tanimoto T, Matsumura M, Tada S, Fujita S, Ueno S, Hamai K, Omoto T, Maeda H, Nishisaka T, Ishikawa N. Need for a high-specificity test for confirming weakly positive result in an immunochromatographic SARS-CoV-2-specific antigen test: a case report. J Microbiol Immunol Infect. 2020;S1684–1182:30272–3. Tanimoto T, Matsumura M, Tada S, Fujita S, Ueno S, Hamai K, Omoto T, Maeda H, Nishisaka T, Ishikawa N. Need for a high-specificity test for confirming weakly positive result in an immunochromatographic SARS-CoV-2-specific antigen test: a case report. J Microbiol Immunol Infect. 2020;S1684–1182:30272–3.
18.
Zurück zum Zitat Cerutti F, Burdino E, Milia MG, Allice T, Gregori G, Bruzzone B, Ghisetti V. Urgent need of rapid tests for SARS CoV-2 antigen detection: evaluation of the SD-biosensor antigen test for SARS-CoV-2. J Clin Virol. 2020;132:104654. Cerutti F, Burdino E, Milia MG, Allice T, Gregori G, Bruzzone B, Ghisetti V. Urgent need of rapid tests for SARS CoV-2 antigen detection: evaluation of the SD-biosensor antigen test for SARS-CoV-2. J Clin Virol. 2020;132:104654.
19.
Zurück zum Zitat Diao B, Wen K, Zhang J, Chen J, Han C, Chen Y, Wang S, Deng G, Zhou H, Wu Y. Accuracy of a nucleocapsid protein antigen rapid test in the diagnosis of SARS-CoV-2 infection. Clin Microbiol Infect. 2020;S1198–743X(20)30611-X. Diao B, Wen K, Zhang J, Chen J, Han C, Chen Y, Wang S, Deng G, Zhou H, Wu Y. Accuracy of a nucleocapsid protein antigen rapid test in the diagnosis of SARS-CoV-2 infection. Clin Microbiol Infect. 2020;S1198–743X(20)30611-X.
20.
Zurück zum Zitat Linares M, Pérez-Tanoira R, Carrero A, Romanyk J, Pérez-García F, Gómez-Herruz P, Arroyo T, Cuadros J. Panbio antigen rapid test is reliable to diagnose SARS-CoV-2 infection in the first 7 days after the onset of symptoms. J Clin Virol. 2020;133:104659.CrossRef Linares M, Pérez-Tanoira R, Carrero A, Romanyk J, Pérez-García F, Gómez-Herruz P, Arroyo T, Cuadros J. Panbio antigen rapid test is reliable to diagnose SARS-CoV-2 infection in the first 7 days after the onset of symptoms. J Clin Virol. 2020;133:104659.CrossRef
21.
Zurück zum Zitat Toptan T, Eckermann L, Pfeiffer AE, Hoehl S, Ciesek S, Drosten C, Corman VM. Evaluation of a SARS-CoV-2 rapid antigen test: potential to help reduce community spread? J Clin Virol. 2020;135:104713.CrossRef Toptan T, Eckermann L, Pfeiffer AE, Hoehl S, Ciesek S, Drosten C, Corman VM. Evaluation of a SARS-CoV-2 rapid antigen test: potential to help reduce community spread? J Clin Virol. 2020;135:104713.CrossRef
22.
Zurück zum Zitat Weitzel T, Legarraga P, Iruretagoyena M, Pizarro G, Vollrath V, Araos R, Munita JM, Porte L. Comparative evaluation of four rapid SARS-CoV-2 antigen detection tests using universal transport medium. Travel Med Infect Dis. 2020;39:101942.CrossRef Weitzel T, Legarraga P, Iruretagoyena M, Pizarro G, Vollrath V, Araos R, Munita JM, Porte L. Comparative evaluation of four rapid SARS-CoV-2 antigen detection tests using universal transport medium. Travel Med Infect Dis. 2020;39:101942.CrossRef
23.
Zurück zum Zitat Berger A, Nsoga MTN, Perez-Rodriguez FJ, Aad YA, Sattonnet-Roche P, Gayet-Ageron A, Jaksic C, Torriani G, Boehm E, Kronig I, Sacks JA, de Vos M, Bausch FJ, Chappuis F, Renzoni A, Kaiser L, Schibler M, Eckerle I. Diagnostic accuracy of two commercial SARS-CoV-2 antigen-detecting rapid tests at the point of care in community-based testing centers. PLoS ONE. 2021;16:e0248921.CrossRef Berger A, Nsoga MTN, Perez-Rodriguez FJ, Aad YA, Sattonnet-Roche P, Gayet-Ageron A, Jaksic C, Torriani G, Boehm E, Kronig I, Sacks JA, de Vos M, Bausch FJ, Chappuis F, Renzoni A, Kaiser L, Schibler M, Eckerle I. Diagnostic accuracy of two commercial SARS-CoV-2 antigen-detecting rapid tests at the point of care in community-based testing centers. PLoS ONE. 2021;16:e0248921.CrossRef
24.
Zurück zum Zitat Courtellemont L, Guinard J, Guillaume C, Giaché S, Rzepecki V, Seve A, Gubavu C, Baud K, Le Helloco C, Cassuto GN, Pialoux G, Hocqueloux L, Prazuck T. High performance of a novel antigen detection test on nasopharyngeal specimens for diagnosing SARS-CoV-2 infection. J Med Virol. 2021;93:3152–7.CrossRef Courtellemont L, Guinard J, Guillaume C, Giaché S, Rzepecki V, Seve A, Gubavu C, Baud K, Le Helloco C, Cassuto GN, Pialoux G, Hocqueloux L, Prazuck T. High performance of a novel antigen detection test on nasopharyngeal specimens for diagnosing SARS-CoV-2 infection. J Med Virol. 2021;93:3152–7.CrossRef
25.
Zurück zum Zitat Mboumba Bouassa RS, Veyer D, Péré H, Bélec L. Analytical performances of the point-of-care SIENNATM COVID-19 antigen rapid test for the detection of SARS-CoV-2 nucleocapsid protein in nasopharyngeal swabs: a prospective evaluation during the COVID-19 second wave in France. Int J Infect Dis. 2021;18:8–12.CrossRef Mboumba Bouassa RS, Veyer D, Péré H, Bélec L. Analytical performances of the point-of-care SIENNATM COVID-19 antigen rapid test for the detection of SARS-CoV-2 nucleocapsid protein in nasopharyngeal swabs: a prospective evaluation during the COVID-19 second wave in France. Int J Infect Dis. 2021;18:8–12.CrossRef
26.
Zurück zum Zitat Landaas ET, Storm ML, Tollånes MC, Barlinn R, Kran AB, Bragstad K, Christensen A, Andreassen T. Diagnostic performance of a SARS-CoV-2 rapid antigen test in a large Norwegian cohort. J Clin Virol. 2021;137:104789.CrossRef Landaas ET, Storm ML, Tollånes MC, Barlinn R, Kran AB, Bragstad K, Christensen A, Andreassen T. Diagnostic performance of a SARS-CoV-2 rapid antigen test in a large Norwegian cohort. J Clin Virol. 2021;137:104789.CrossRef
28.
Zurück zum Zitat Takeuchi Y, Akashi Y, Kato D, Kuwahara M, Muramatsu S, Ueda A, Notake S, Nakamura K, Ishikawa H, Suzuki H. The evaluation of a newly developed antigen test (QuickNavi-COVID19 Ag) for SARS-CoV-2: a prospective observational study in Japan. J Infect Chemother. 2021;27:890–4.CrossRef Takeuchi Y, Akashi Y, Kato D, Kuwahara M, Muramatsu S, Ueda A, Notake S, Nakamura K, Ishikawa H, Suzuki H. The evaluation of a newly developed antigen test (QuickNavi-COVID19 Ag) for SARS-CoV-2: a prospective observational study in Japan. J Infect Chemother. 2021;27:890–4.CrossRef
29.
Zurück zum Zitat Albert E, Torres I, Bueno F, Huntley D, Molla E, Fernández-Fuentes MÁ, Martínez M, Poujois S, Forqué L, Valdivia A, Solano de la Asunción C, Ferrer J, Colomina J, Navarro D. Field evaluation of a rapid antigen test (Panbio COVID-19 Ag Rapid Test Device) for COVID-19 diagnosis in primary healthcare centres. Clin Microbiol Infect. 2020;S1198–743X(20)30697–2. Albert E, Torres I, Bueno F, Huntley D, Molla E, Fernández-Fuentes MÁ, Martínez M, Poujois S, Forqué L, Valdivia A, Solano de la Asunción C, Ferrer J, Colomina J, Navarro D. Field evaluation of a rapid antigen test (Panbio COVID-19 Ag Rapid Test Device) for COVID-19 diagnosis in primary healthcare centres. Clin Microbiol Infect. 2020;S1198–743X(20)30697–2.
30.
Zurück zum Zitat Scohy A, Anantharajah A, Bodéus M, Kabamba-Mukadi B, Verroken A, Rodriguez-Villalobos H. Low performance of rapid antigen detection test as frontline testing for COVID-19 diagnosis. J Clin Virol. 2020;129:104455.CrossRef Scohy A, Anantharajah A, Bodéus M, Kabamba-Mukadi B, Verroken A, Rodriguez-Villalobos H. Low performance of rapid antigen detection test as frontline testing for COVID-19 diagnosis. J Clin Virol. 2020;129:104455.CrossRef
31.
Zurück zum Zitat Yamayoshi S, Sakai-Tagawa Y, Koga M, Akasaka O, Nakachi I, Koh H, Maeda K, Adachi E, Saito M, Nagai H, Ikeuchi K, Ogura T, Baba R, Fujita K, Fukui T, Ito F, Hattori SI, Yamamoto K, Nakamoto T, Furusawa Y, Yasuhara A, Ujie M, Yamada S, Ito M, Mitsuya H, Omagari N, Yotsuyanagi H, Iwatsuki-Horimoto K, Imai M, Kawaoka Y. Comparison of rapid antigen tests for COVID-19. Viruses. 2020;12:1420.CrossRef Yamayoshi S, Sakai-Tagawa Y, Koga M, Akasaka O, Nakachi I, Koh H, Maeda K, Adachi E, Saito M, Nagai H, Ikeuchi K, Ogura T, Baba R, Fujita K, Fukui T, Ito F, Hattori SI, Yamamoto K, Nakamoto T, Furusawa Y, Yasuhara A, Ujie M, Yamada S, Ito M, Mitsuya H, Omagari N, Yotsuyanagi H, Iwatsuki-Horimoto K, Imai M, Kawaoka Y. Comparison of rapid antigen tests for COVID-19. Viruses. 2020;12:1420.CrossRef
33.
Zurück zum Zitat Torres I, Poujois S, Albert E, Colomina J, Navarro D. Evaluation of a rapid antigen test (Panbio COVID-19 Ag rapid test device) for SARS-CoV-2 detection in asymptomatic close contacts of COVID-19 patients. Clin Microbiol Infect. 2021;S1198–743X(20)30782–5. Torres I, Poujois S, Albert E, Colomina J, Navarro D. Evaluation of a rapid antigen test (Panbio COVID-19 Ag rapid test device) for SARS-CoV-2 detection in asymptomatic close contacts of COVID-19 patients. Clin Microbiol Infect. 2021;S1198–743X(20)30782–5.
34.
Zurück zum Zitat Torres I, Poujois S, Albert E, Álvarez G, Colomina J, Navarro D. Point-of-care evaluation of a rapid antigen test (CLINITESTⓇ Rapid COVID-19 Antigen Test) for diagnosis of SARS-CoV-2 infection in symptomatic and asymptomatic individuals. J Infect. 2021;82:e11–2.CrossRef Torres I, Poujois S, Albert E, Álvarez G, Colomina J, Navarro D. Point-of-care evaluation of a rapid antigen test (CLINITEST Rapid COVID-19 Antigen Test) for diagnosis of SARS-CoV-2 infection in symptomatic and asymptomatic individuals. J Infect. 2021;82:e11–2.CrossRef
35.
Zurück zum Zitat Haute Autorité de santé, Saint-Denis, France. Revue rapide sur les tests de détection antigénique du virus SARS-CoV-2. 2020 Haute Autorité de santé, Saint-Denis, France. Revue rapide sur les tests de détection antigénique du virus SARS-CoV-2. 2020
36.
Zurück zum Zitat Jefferson T, Spencer EA, Brassey J, Heneghan C. Viral cultures for COVID-19 infectious potential assessment - a systematic review. Clin Infect Dis. 2020;ciaa1764. Jefferson T, Spencer EA, Brassey J, Heneghan C. Viral cultures for COVID-19 infectious potential assessment - a systematic review. Clin Infect Dis. 2020;ciaa1764.
37.
Zurück zum Zitat Zou L, Ruan F, Huang M, Liang L, Huang H, Hong Z, Yu J, Kang M, Song Y, Xia J, Guo Q, Song T, He J, Yen HL, Peiris M, Wu J. SARS-CoV-2 viral load in upper respiratory specimens of infected patients. N Engl J Med. 2020;382:1177–9.CrossRef Zou L, Ruan F, Huang M, Liang L, Huang H, Hong Z, Yu J, Kang M, Song Y, Xia J, Guo Q, Song T, He J, Yen HL, Peiris M, Wu J. SARS-CoV-2 viral load in upper respiratory specimens of infected patients. N Engl J Med. 2020;382:1177–9.CrossRef
38.
Zurück zum Zitat La Scola B, Le Bideau M, Andreani J, Hoang VT, Grimaldier C, Colson P, Gautret P, Raoult D. Viral RNA load as determined by cell culture as a management tool for discharge of SARS-CoV-2 patients from infectious disease wards. Eur J Clin Microbiol Infect Dis. 2020;39:1059–61.CrossRef La Scola B, Le Bideau M, Andreani J, Hoang VT, Grimaldier C, Colson P, Gautret P, Raoult D. Viral RNA load as determined by cell culture as a management tool for discharge of SARS-CoV-2 patients from infectious disease wards. Eur J Clin Microbiol Infect Dis. 2020;39:1059–61.CrossRef
41.
Zurück zum Zitat Yu F, Yan L, Wang N, Yang S, Wang L, Tang Y, Gao G, Wang S, Ma C, Xie R, Wang F, Tan C, Zhu L, Guo Y, Zhang F. Quantitative detection and viral load analysis of SARS-CoV-2 in infected patients. Clin Infect Dis. 2020;71:793–8.CrossRef Yu F, Yan L, Wang N, Yang S, Wang L, Tang Y, Gao G, Wang S, Ma C, Xie R, Wang F, Tan C, Zhu L, Guo Y, Zhang F. Quantitative detection and viral load analysis of SARS-CoV-2 in infected patients. Clin Infect Dis. 2020;71:793–8.CrossRef
43.
Zurück zum Zitat Cohen J. A coefficient of agreement for nominal scales. Educ Psychol Meas. 1960;20:37–46.CrossRef Cohen J. A coefficient of agreement for nominal scales. Educ Psychol Meas. 1960;20:37–46.CrossRef
44.
Zurück zum Zitat Landlis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics. 1977;33:159–74.CrossRef Landlis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics. 1977;33:159–74.CrossRef
45.
Metadaten
Titel
Analytical performance of the point-of-care BIOSYNEX COVID-19 Ag BSS for the detection of SARS‐CoV‐2 nucleocapsid protein in nasopharyngeal swabs: a prospective field evaluation during the COVID-19 third wave in France
verfasst von
Frédéric Fitoussi
Serge Tonen-Wolyec
Natalio Awaida
Raphaël Dupont
Laurent Bélec
Publikationsdatum
24.10.2021
Verlag
Springer Berlin Heidelberg
Schlagwort
COVID-19
Erschienen in
Infection / Ausgabe 3/2022
Print ISSN: 0300-8126
Elektronische ISSN: 1439-0973
DOI
https://doi.org/10.1007/s15010-021-01723-5

Weitere Artikel der Ausgabe 3/2022

Infection 3/2022 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

NSCLC: Progressionsfreies Überleben unter Osimertinib fast versiebenfacht

06.06.2024 ASCO 2024 Kongressbericht

Erste Ergebnisse der Phase-III-Studie LAURA etablieren Osimertinib als neuen Therapiestandard für Menschen mit nicht-resezierbarem, EGFR-mutiertem, nicht-kleinzelligem Lungenkarzinom im Stadium III, die nach definitiver Radiochemotherapie progressionsfrei sind. Auf der ASCO-Tagung wurden diese beeindruckenden Ergebnisse besprochen.

Hodgkin Lymphom: BrECADD-Regime übertrifft die Erwartungen

05.06.2024 ASCO 2024 Kongressbericht

Das Kombinationsregime BrECADD mit Brentuximab vedotin ermöglichte in der Studie HD21 beim fortgeschrittenen klassischen Hodgkin-Lymphom eine unerwartet hohe progressionsfreie Überlebensrate von 94,3% nach vier Jahren. Gleichzeitig war das Regime besser tolerabel als der bisherige Standard eBEACOPP.

Antikörper-Drug-Konjugat verdoppelt PFS bei Multiplem Myelom

05.06.2024 ASCO 2024 Nachrichten

Zwei Phase-3-Studien deuten auf erhebliche Vorteile des Antikörper-Wirkstoff-Konjugats Belantamab-Mafodotin bei vorbehandelten Personen mit Multiplem Myelom: Im Vergleich mit einer Standard-Tripeltherapie wurde das progressionsfreie Überleben teilweise mehr als verdoppelt.

Neuer TKI gegen CML: Höhere Wirksamkeit, seltener Nebenwirkungen

05.06.2024 Chronische myeloische Leukämie Nachrichten

Der Tyrosinkinasehemmer (TKI) Asciminib ist älteren Vertretern dieser Gruppe bei CML offenbar überlegen: Personen mit frisch diagnostizierter CML entwickelten damit in einer Phase-3-Studie häufiger eine gute molekulare Response, aber seltener ernste Nebenwirkungen.

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.