Skip to main content
Erschienen in: Die Anaesthesiologie 8/2019

22.07.2019 | Notfallmedizin

Defibrillator charging before rhythm analysis causes peri-shock pauses exceeding guideline recommended maximum 5 s

A randomized simulation trial

verfasst von: M. Kemper, A. Zech, M. Lazarovici, B. Zwissler, S. Prückner, O. Meyer

Erschienen in: Die Anaesthesiologie | Ausgabe 8/2019

Einloggen, um Zugang zu erhalten

Abstract

Introduction

Charging defibrillators prior to analyzing heart rhythms may decrease the no-flow time during rhythm check pauses while resuscitating in cardiac arrest. Although this anticipatory method is already used in some centers little is known about its safety. This study was carried out to confirm the safety and feasibility of the anticipatory method. It was hypothesized that this anticipatory method results in shorter total no-flow times, while other parameters of defibrillation efficacy including defibrillator safety and minimization of peri-shock pauses are unchanged.

Methods

This manikin study assigned 243 medical students randomly to study groups, 121 to the anticipatory method and 122 to the recommended European Resuscitation Council (ERC) algorithm. Of these 237 students ultimately underwent training (112 anticipatory method vs. 125 ERC algorithm). Participants were assessed and video recorded during a simulated cardiac arrest scenario which included three different heart rhythms (ventricular fibrillation [VF], pulseless ventricular tachycardia [pVT], asystole) in randomized order. Video and software analyses were performed. Defibrillation safety was assessed using a 17-item checklist defined beforehand.

Results

A total of 203 simulated cardiac arrests (75 anticipatory method and 128 ERC 2010 algorithm) were analyzed. The anticipatory method did not significantly reduce no-flow time (25.8 s, standard deviation, SD 7.4 s vs. 27.4 s SD 8.4 s, p = 0.19); however, peri-shock pauses were significantly longer in the anticipatory group compared to the ERC 2010 group (9.5 s SD 2.8 s vs. 3.3 s SD 1.9 s, p < 0.001). No significant difference concerning defibrillation safety between the groups was observed according to the 17-item checklist (14.6 SD 1.6 vs. 15.0 SD 1.4, p = 0.07).

Conclusion

Charging defibrillators before rhythm analysis did not decrease total no-flow time in simulated cardiac arrests but resulted in significantly longer peri-shock pauses exceeding 5 s. No significant differences in defibrillation safety were observed between the groups.
Fußnoten
1
Ethics committee: LMU Munich, Pettenkoferstr. 8a, 80336 Munich, Germany, 15.10.2012, project number: 427-12.
 
Literatur
1.
Zurück zum Zitat Atwood C, Eisenberg MS, Herlitz J et al (2005) Incidence of EMS-treated out-of-hospital cardiac arrest in Europe. Resuscitation 67:75–80CrossRefPubMed Atwood C, Eisenberg MS, Herlitz J et al (2005) Incidence of EMS-treated out-of-hospital cardiac arrest in Europe. Resuscitation 67:75–80CrossRefPubMed
2.
Zurück zum Zitat Auble TE, Menegazzi JJ, Paris PM (1995) Effect of out-of-hospital defibrillation by basic life support providers on cardiac arrest mortality: A metaanalysis. Ann Emerg Med 25:642–648CrossRefPubMed Auble TE, Menegazzi JJ, Paris PM (1995) Effect of out-of-hospital defibrillation by basic life support providers on cardiac arrest mortality: A metaanalysis. Ann Emerg Med 25:642–648CrossRefPubMed
3.
Zurück zum Zitat Cheskes S, Schmicker RH, Christenson J et al (2011) Perishock pause: An independent predictor of survival from out-of-hospital shockable cardiac arrest. Circulation 124:58–66CrossRefPubMedPubMedCentral Cheskes S, Schmicker RH, Christenson J et al (2011) Perishock pause: An independent predictor of survival from out-of-hospital shockable cardiac arrest. Circulation 124:58–66CrossRefPubMedPubMedCentral
4.
Zurück zum Zitat Cheskes S, Schmicker RH, Verbeek PR et al (2014) The impact of peri-shock pause on survival from out-of-hospital shockable cardiac arrest during the Resuscitation Outcomes Consortium PRIMED trial. Resuscitation 85:336–342CrossRefPubMed Cheskes S, Schmicker RH, Verbeek PR et al (2014) The impact of peri-shock pause on survival from out-of-hospital shockable cardiac arrest during the Resuscitation Outcomes Consortium PRIMED trial. Resuscitation 85:336–342CrossRefPubMed
5.
Zurück zum Zitat Deakin CD, Nolan JP, Soar J et al (2010) European Resuscitation Council Guidelines for Resuscitation 2010 Section 4. Adult advanced life support. Resuscitation 81:1305–1352CrossRefPubMed Deakin CD, Nolan JP, Soar J et al (2010) European Resuscitation Council Guidelines for Resuscitation 2010 Section 4. Adult advanced life support. Resuscitation 81:1305–1352CrossRefPubMed
6.
Zurück zum Zitat Deakin CD, Nolan JP, Sunde K et al (2010) European Resuscitation Council Guidelines for Resuscitation 2010 Section 3. Electrical therapies: Automated external defibrillators, defibrillation, cardioversion and pacing. Resuscitation 81:1293–1304CrossRefPubMed Deakin CD, Nolan JP, Sunde K et al (2010) European Resuscitation Council Guidelines for Resuscitation 2010 Section 3. Electrical therapies: Automated external defibrillators, defibrillation, cardioversion and pacing. Resuscitation 81:1293–1304CrossRefPubMed
7.
Zurück zum Zitat Deakin CD, Thomsen JE, Lofgren B et al (2015) Achieving safe hands-on defibrillation using electrical safety gloves—A clinical evaluation. Resuscitation 90:163–167CrossRefPubMed Deakin CD, Thomsen JE, Lofgren B et al (2015) Achieving safe hands-on defibrillation using electrical safety gloves—A clinical evaluation. Resuscitation 90:163–167CrossRefPubMed
8.
Zurück zum Zitat Dickinson CL, Hall CR, Soar J (2008) Accidental shock to rescuer during successful defibrillation of ventricular fibrillation—A case of human involuntary automaticity. Resuscitation 76:489CrossRefPubMed Dickinson CL, Hall CR, Soar J (2008) Accidental shock to rescuer during successful defibrillation of ventricular fibrillation—A case of human involuntary automaticity. Resuscitation 76:489CrossRefPubMed
9.
Zurück zum Zitat Edelson DP, Abella BS, Kramer-Johansen J et al (2006) Effects of compression depth and pre-shock pauses predict defibrillation failure during cardiac arrest. Resuscitation 71:137–145CrossRefPubMed Edelson DP, Abella BS, Kramer-Johansen J et al (2006) Effects of compression depth and pre-shock pauses predict defibrillation failure during cardiac arrest. Resuscitation 71:137–145CrossRefPubMed
10.
Zurück zum Zitat Edelson DP, Robertson-Dick BJ, Yuen TC et al (2010) Safety and efficacy of defibrillator charging during ongoing chest compressions: A multi-center study. Resuscitation 81:1521–1526CrossRefPubMedPubMedCentral Edelson DP, Robertson-Dick BJ, Yuen TC et al (2010) Safety and efficacy of defibrillator charging during ongoing chest compressions: A multi-center study. Resuscitation 81:1521–1526CrossRefPubMedPubMedCentral
11.
Zurück zum Zitat Eftestol T (2002) Effects of interrupting precordial compressions on the calculated probability of defibrillation success during out-of-hospital cardiac arrest. Circulation 105:2270–2273CrossRefPubMed Eftestol T (2002) Effects of interrupting precordial compressions on the calculated probability of defibrillation success during out-of-hospital cardiac arrest. Circulation 105:2270–2273CrossRefPubMed
12.
Zurück zum Zitat Hansen LK, Folkestad L, Brabrand M (2013) Defibrillator charging before rhythm analysis significantly reduces hands-off time during resuscitation: A simulation study. Am J Emerg Med 31:395–400CrossRefPubMed Hansen LK, Folkestad L, Brabrand M (2013) Defibrillator charging before rhythm analysis significantly reduces hands-off time during resuscitation: A simulation study. Am J Emerg Med 31:395–400CrossRefPubMed
13.
Zurück zum Zitat Hoke RS, Heinroth K, Trappe HJ et al (2009) Is external defibrillation an electric threat for bystanders? Resuscitation 80:395–401CrossRefPubMed Hoke RS, Heinroth K, Trappe HJ et al (2009) Is external defibrillation an electric threat for bystanders? Resuscitation 80:395–401CrossRefPubMed
14.
Zurück zum Zitat Iirola T, Lund VE, Katila AJ et al (2002) Teaching hospital physicians’ skills and knowledge of resuscitation algorithms are deficient. Acta Anaesthesiol Scand 46:1150–1154CrossRefPubMed Iirola T, Lund VE, Katila AJ et al (2002) Teaching hospital physicians’ skills and knowledge of resuscitation algorithms are deficient. Acta Anaesthesiol Scand 46:1150–1154CrossRefPubMed
15.
Zurück zum Zitat Koch Hansen L, Mohammed A, Pedersen M et al (2015) The Stop-Only-While-Shocking algorithm reduces hands-off time by 17 % during cardiopulmonary resuscitation—A simulation study. Eur J Emerg Med 23(6):413–417CrossRef Koch Hansen L, Mohammed A, Pedersen M et al (2015) The Stop-Only-While-Shocking algorithm reduces hands-off time by 17 % during cardiopulmonary resuscitation—A simulation study. Eur J Emerg Med 23(6):413–417CrossRef
16.
Zurück zum Zitat Kramer-Johansen J, Edelson DP, Abella BS et al (2007) Pauses in chest compression and inappropriate shocks: A comparison of manual and semi-automatic defibrillation attempts. Resuscitation 73:212–220CrossRefPubMed Kramer-Johansen J, Edelson DP, Abella BS et al (2007) Pauses in chest compression and inappropriate shocks: A comparison of manual and semi-automatic defibrillation attempts. Resuscitation 73:212–220CrossRefPubMed
17.
Zurück zum Zitat Krautter M, Weyrich P, Schultz JH et al (2011) Effects of Peyton’s four-step approach on objective performance measures in technical skills training: A controlled trial. Teach Learn Med 23:244–250CrossRefPubMed Krautter M, Weyrich P, Schultz JH et al (2011) Effects of Peyton’s four-step approach on objective performance measures in technical skills training: A controlled trial. Teach Learn Med 23:244–250CrossRefPubMed
18.
Zurück zum Zitat Lindekaer AL, Jacobsen J, Andersen G et al (1997) Treatment of ventricular fibrillation during anaesthesia in an anaesthesia simulator. Acta Anaesthesiol Scand 41:1280–1284CrossRefPubMed Lindekaer AL, Jacobsen J, Andersen G et al (1997) Treatment of ventricular fibrillation during anaesthesia in an anaesthesia simulator. Acta Anaesthesiol Scand 41:1280–1284CrossRefPubMed
19.
Zurück zum Zitat Pollack RA, Brown SP, Rea T et al (2018) Impact of bystander automated external defibrillator use on survival and functional outcomes in shockable observed public cardiac arrests. Circulation 137:2104–2113CrossRefPubMedPubMedCentral Pollack RA, Brown SP, Rea T et al (2018) Impact of bystander automated external defibrillator use on survival and functional outcomes in shockable observed public cardiac arrests. Circulation 137:2104–2113CrossRefPubMedPubMedCentral
20.
Zurück zum Zitat Pytte M, Pedersen TE, Ottem J et al (2007) Comparison of hands-off time during CPR with manual and semi-automatic defibrillation in a manikin model. Resuscitation 73:131–136CrossRefPubMed Pytte M, Pedersen TE, Ottem J et al (2007) Comparison of hands-off time during CPR with manual and semi-automatic defibrillation in a manikin model. Resuscitation 73:131–136CrossRefPubMed
21.
Zurück zum Zitat Sans S, Kesteloot H, Kromhout D (1997) The burden of cardiovascular diseases mortality in Europe. Task Force of the European Society of Cardiology on Cardiovascular Mortality and Morbidity Statistics in Europe. Eur Heart J 18:1231–1248CrossRefPubMed Sans S, Kesteloot H, Kromhout D (1997) The burden of cardiovascular diseases mortality in Europe. Task Force of the European Society of Cardiology on Cardiovascular Mortality and Morbidity Statistics in Europe. Eur Heart J 18:1231–1248CrossRefPubMed
22.
Zurück zum Zitat Savastano S, Vanni V, Burkart R et al (2017) Comparative performance assessment of commercially available automatic external defibrillators: A simulation and real-life measurement study of hands-off time. Resuscitation 110:12–17CrossRefPubMed Savastano S, Vanni V, Burkart R et al (2017) Comparative performance assessment of commercially available automatic external defibrillators: A simulation and real-life measurement study of hands-off time. Resuscitation 110:12–17CrossRefPubMed
23.
Zurück zum Zitat Sell RE, Sarno R, Lawrence B et al (2010) Minimizing pre- and post-defibrillation pauses increases the likelihood of return of spontaneous circulation (ROSC). Resuscitation 81:822–825CrossRefPubMed Sell RE, Sarno R, Lawrence B et al (2010) Minimizing pre- and post-defibrillation pauses increases the likelihood of return of spontaneous circulation (ROSC). Resuscitation 81:822–825CrossRefPubMed
24.
Zurück zum Zitat Soar J, Nolan JP, Bottiger BW et al (2015) European Resuscitation Council Guidelines for Resuscitation 2015: Section 3. Adult advanced life support. Resuscitation 95:100–147CrossRefPubMed Soar J, Nolan JP, Bottiger BW et al (2015) European Resuscitation Council Guidelines for Resuscitation 2015: Section 3. Adult advanced life support. Resuscitation 95:100–147CrossRefPubMed
25.
Zurück zum Zitat Steinmetz J, Barnung S, Nielsen SL et al (2008) Improved survival after an out-of-hospital cardiac arrest using new guidelines. Acta Anaesthesiol Scand 52:908–913CrossRefPubMed Steinmetz J, Barnung S, Nielsen SL et al (2008) Improved survival after an out-of-hospital cardiac arrest using new guidelines. Acta Anaesthesiol Scand 52:908–913CrossRefPubMed
26.
Zurück zum Zitat Stiell IG, Wells GA, Demaio VJ et al (1999) Modifiable factors associated with improved cardiac arrest survival in a multicenter basic life support/defibrillation system: OPALS study phase I results. Ontario Prehospital Advanced Life Support. Ann Emerg Med 33:44–50CrossRefPubMed Stiell IG, Wells GA, Demaio VJ et al (1999) Modifiable factors associated with improved cardiac arrest survival in a multicenter basic life support/defibrillation system: OPALS study phase I results. Ontario Prehospital Advanced Life Support. Ann Emerg Med 33:44–50CrossRefPubMed
27.
Zurück zum Zitat Stiell IG, Wells GA, Field BJ et al (1999) Improved out-of-hospital cardiac arrest survival through the inexpensive optimization of an existing defibrillation program: OPALS study phase II. Ontario Prehospital Advanced Life Support. JAMA 281:1175–1181CrossRefPubMed Stiell IG, Wells GA, Field BJ et al (1999) Improved out-of-hospital cardiac arrest survival through the inexpensive optimization of an existing defibrillation program: OPALS study phase II. Ontario Prehospital Advanced Life Support. JAMA 281:1175–1181CrossRefPubMed
28.
Zurück zum Zitat Thim T, Grove EL, Lofgren B (2012) Charging the defibrillator before rhythm check reduces hands-off time during CPR: A randomised simulation study. Resuscitation 83:e210–e211CrossRefPubMed Thim T, Grove EL, Lofgren B (2012) Charging the defibrillator before rhythm check reduces hands-off time during CPR: A randomised simulation study. Resuscitation 83:e210–e211CrossRefPubMed
29.
Zurück zum Zitat Tomkins WG, Swain AH, Bailey M et al (2013) Beyond the pre-shock pause: The effect of prehospital defibrillation mode on CPR interruptions and return of spontaneous circulation. Resuscitation 84:575–579CrossRefPubMed Tomkins WG, Swain AH, Bailey M et al (2013) Beyond the pre-shock pause: The effect of prehospital defibrillation mode on CPR interruptions and return of spontaneous circulation. Resuscitation 84:575–579CrossRefPubMed
Metadaten
Titel
Defibrillator charging before rhythm analysis causes peri-shock pauses exceeding guideline recommended maximum 5 s
A randomized simulation trial
verfasst von
M. Kemper
A. Zech
M. Lazarovici
B. Zwissler
S. Prückner
O. Meyer
Publikationsdatum
22.07.2019
Verlag
Springer Medizin
Erschienen in
Die Anaesthesiologie / Ausgabe 8/2019
Print ISSN: 2731-6858
Elektronische ISSN: 2731-6866
DOI
https://doi.org/10.1007/s00101-019-0623-x

Weitere Artikel der Ausgabe 8/2019

Die Anaesthesiologie 8/2019 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.