Dosing of the second-generation cryoballoon using acute time-to-pulmonary vein isolation as an indicator of durable ablation in a canine model

verfasst von: Wilber Su, Nicolas Coulombe, Nicole Kirchhof, Erin Grassl, Dan Wittenberger

Erschienen in: Journal of Interventional Cardiac Electrophysiology | Ausgabe 3/2018

Einloggen, um Zugang zu erhalten

Abstract

Background

Rigid time-based dosing protocol(s) currently used in the clinic for cryoballoon ablation of atrial fibrillation may be inadequate to guide the circumferential and transmural cryothermal energy transfer across the pulmonary vein (PV) and may result in injury to collateral tissues or electrical gaps between the PV and left atrium (LA).

Objective

A physiologic endpoint (e.g., acute time-to-PV isolation a.k.a. time-to-effect; TTE) may be effective in the determination of a transmural lesion formation and may allow for individualized ablation dosing across each PV.

Methods

Thirty PVs from 15 dogs were randomized into five dosing protocols, including (1) TTE + 60 s, (2) TTE + 90 s, (3) TTE + 120 s, (4) TTE + 150 s, and (5) 2 × 180 s. Ablations were conducted with a 23-mm second-generation cryoballoon, and TTE was assessed during a freeze by pacing from an inner balloon-lumen circular diagnostic catheter to a quadripolar diagnostic catheter in the coronary sinus. After ablation, animals were survived for 30 to 34 days, and repeat electrophysiology assessment of PV isolation was conducted after which animals were euthanized for gross anatomy and histological examination.

Results

At study termination, efficacy endpoint evaluations were based on maintenance of PV electrical isolation, gross anatomy assessment of PV lesions, and histological examination of PVs. Five efficacy endpoint failures were noted, including the following: 1 PV in the TTE + 90 sec group; 2 PVs in the TTE + 120 sec group; 1 PV in the TTE + 150 s group; and 1 PV in the 2 × 180 s group. Regarding safety, one phrenic nerve injury was observed in the 2 × 180 s cohort. No other complications were observed.

Conclusions

In a canine model, effective PV isolation could be found even in the shortest duration dosing cohort (TTE + 60 s). One complication (phrenic nerve injury) was observed in the longest duration dosing group (2 × 180 s). Further studies will be required to correlate these results to a 28-mm cryoballoon (more commonly used in the cryoablation of a human LA); however, to date, this is the first reporting of a successful cryoablation using TTE + 60 s dosing (approximately 90 s total duration of freezing).
Literatur
1.
Zurück zum Zitat Packer DL, Kowal RC, Wheelan KR, Irwin JM, Champagne J, Guerra PG, et al. Cryoballoon ablation of pulmonary veins for paroxysmal atrial fibrillation: first results of the North American Arctic Front (STOP AF) pivotal trial. J Am Coll Cardiol. 2013;61(16):1713–23.CrossRef Packer DL, Kowal RC, Wheelan KR, Irwin JM, Champagne J, Guerra PG, et al. Cryoballoon ablation of pulmonary veins for paroxysmal atrial fibrillation: first results of the North American Arctic Front (STOP AF) pivotal trial. J Am Coll Cardiol. 2013;61(16):1713–23.CrossRef
2.
Zurück zum Zitat Cardoso R, Mendirichaga R, Fernandes G, Healy C, Lambrakos LK, Viles-Gonzalez JF, et al. Cryoballoon versus radiofrequency catheter ablation in atrial fibrillation: a meta-analysis. J Cardiovasc Electrophysiol. 2016;27(10):1151–9.CrossRef Cardoso R, Mendirichaga R, Fernandes G, Healy C, Lambrakos LK, Viles-Gonzalez JF, et al. Cryoballoon versus radiofrequency catheter ablation in atrial fibrillation: a meta-analysis. J Cardiovasc Electrophysiol. 2016;27(10):1151–9.CrossRef
3.
Zurück zum Zitat Garg J, Chaudhary R, Palaniswamy C, Shah N, Krishnamoorthy P, Bozorgnia B, et al. Cryoballoon versus radiofrequency ablation for atrial fibrillation: a meta-analysis of 16 clinical trials. J Atr Fibrillation. 2016;9(3):1429–38.CrossRef Garg J, Chaudhary R, Palaniswamy C, Shah N, Krishnamoorthy P, Bozorgnia B, et al. Cryoballoon versus radiofrequency ablation for atrial fibrillation: a meta-analysis of 16 clinical trials. J Atr Fibrillation. 2016;9(3):1429–38.CrossRef
4.
Zurück zum Zitat Mugnai G, de Asmundis C, Velagic V, Hünük B, Ströker E, Wauters K, et al. Phrenic nerve injury during ablation with the second-generation cryoballoon: analysis of the temperature drop behaviour in a large cohort of patients. Europace. 2016;18(5):702–9.CrossRef Mugnai G, de Asmundis C, Velagic V, Hünük B, Ströker E, Wauters K, et al. Phrenic nerve injury during ablation with the second-generation cryoballoon: analysis of the temperature drop behaviour in a large cohort of patients. Europace. 2016;18(5):702–9.CrossRef
5.
Zurück zum Zitat Andrade JG, Dubuc M, Ferreira J, Guerra PG, Landry E, Coulombe N, et al. Histopathology of cryoballoon ablation-induced phrenic nerve injury. J Cardiovasc Electrophysiol. 2014;25(2):187–94.CrossRef Andrade JG, Dubuc M, Ferreira J, Guerra PG, Landry E, Coulombe N, et al. Histopathology of cryoballoon ablation-induced phrenic nerve injury. J Cardiovasc Electrophysiol. 2014;25(2):187–94.CrossRef
6.
Zurück zum Zitat Straube F, Dorwarth U, Hartl S, Bunz B, Wankerl M, Ebersberger U, et al. Outcome of paroxysmal atrial fibrillation ablation with the cryoballoon using two different application times: the 4- versus 3-min protocol. J Interv Card Electrophysiol. 2016;45(2):169–77.CrossRef Straube F, Dorwarth U, Hartl S, Bunz B, Wankerl M, Ebersberger U, et al. Outcome of paroxysmal atrial fibrillation ablation with the cryoballoon using two different application times: the 4- versus 3-min protocol. J Interv Card Electrophysiol. 2016;45(2):169–77.CrossRef
7.
Zurück zum Zitat Miyazaki S, Hachiya H, Nakamura H, Taniguchi H, Takagi T, Hirao K, et al. Pulmonary vein isolation using a second-generation cryoballoon in patients with paroxysmal atrial fibrillation: one-year outcome using a single big-balloon 3-minute freeze technique. J Cardiovasc Electrophysiol. 2016;27(12):1375–80.CrossRef Miyazaki S, Hachiya H, Nakamura H, Taniguchi H, Takagi T, Hirao K, et al. Pulmonary vein isolation using a second-generation cryoballoon in patients with paroxysmal atrial fibrillation: one-year outcome using a single big-balloon 3-minute freeze technique. J Cardiovasc Electrophysiol. 2016;27(12):1375–80.CrossRef
8.
Zurück zum Zitat Chun KR, Stich M, Fürnkranz A, Bordignon S, Perrotta L, Dugo D, et al. Individualized cryoballoon energy pulmonary vein isolation guided by real-time pulmonary vein recordings, the randomized ICE-T trial. Heart Rhythm. 2017;14(4):495–500.CrossRef Chun KR, Stich M, Fürnkranz A, Bordignon S, Perrotta L, Dugo D, et al. Individualized cryoballoon energy pulmonary vein isolation guided by real-time pulmonary vein recordings, the randomized ICE-T trial. Heart Rhythm. 2017;14(4):495–500.CrossRef
9.
Zurück zum Zitat Aryana A, Kenigsberg DN, Kowalski M, Koo CH, Lim HW, O'Neill PG, et al. Verification of a novel atrial fibrillation cryoablation dosing algorithm guided by time-to-pulmonary vein isolation: results from the Cryo-DOSING study (Cryoballoon-ablation DOSING based on the assessment of time-to-effect and pulmonary vein isolation guidance). Heart Rhythm. 2017;14(9):1319–25.CrossRef Aryana A, Kenigsberg DN, Kowalski M, Koo CH, Lim HW, O'Neill PG, et al. Verification of a novel atrial fibrillation cryoablation dosing algorithm guided by time-to-pulmonary vein isolation: results from the Cryo-DOSING study (Cryoballoon-ablation DOSING based on the assessment of time-to-effect and pulmonary vein isolation guidance). Heart Rhythm. 2017;14(9):1319–25.CrossRef
10.
Zurück zum Zitat Su W, Kowal R, Kowalski M, Metzner A, Svinarich JT, Wheelan K, et al. Best practice guide for cryoballoon ablation in atrial fibrillation: the compilation experience of more than 3000 procedures. Heart Rhythm. 2015;12(7):1658–66.CrossRef Su W, Kowal R, Kowalski M, Metzner A, Svinarich JT, Wheelan K, et al. Best practice guide for cryoballoon ablation in atrial fibrillation: the compilation experience of more than 3000 procedures. Heart Rhythm. 2015;12(7):1658–66.CrossRef
11.
Zurück zum Zitat Heeger CH, Wissner E, Mathew S, Hayashi K, Sohns C, Reißmann B, et al. Short tip-big difference? First-in-man experience and procedural efficacy of pulmonary vein isolation using the third-generation cryoballoon. Clin Res Cardiol. 2016;105(6):482–8.CrossRef Heeger CH, Wissner E, Mathew S, Hayashi K, Sohns C, Reißmann B, et al. Short tip-big difference? First-in-man experience and procedural efficacy of pulmonary vein isolation using the third-generation cryoballoon. Clin Res Cardiol. 2016;105(6):482–8.CrossRef
12.
Zurück zum Zitat Aryana A, Mugnai G, Singh SM, Pujara DK, de Asmundis C, Singh SK, et al. Procedural and biophysical indicators of durable pulmonary vein isolation during cryoballoon ablation of atrial fibrillation. Heart Rhythm. 2016;13(2):424–32.CrossRef Aryana A, Mugnai G, Singh SM, Pujara DK, de Asmundis C, Singh SK, et al. Procedural and biophysical indicators of durable pulmonary vein isolation during cryoballoon ablation of atrial fibrillation. Heart Rhythm. 2016;13(2):424–32.CrossRef
13.
Zurück zum Zitat Coulombe N, Paulin J, Su W. Improved in vivo performance of second-generation cryoballoon for pulmonary vein isolation. J Cardiovasc Electrophysiol. 2013;24(8):919–25.CrossRef Coulombe N, Paulin J, Su W. Improved in vivo performance of second-generation cryoballoon for pulmonary vein isolation. J Cardiovasc Electrophysiol. 2013;24(8):919–25.CrossRef
14.
Zurück zum Zitat Neumann T, Vogt J, Schumacher B, Dorszewski A, Kuniss M, Neuser H, et al. Circumferential pulmonary vein isolation with the cryoballoon technique results from a prospective 3-center study. J Am Coll Cardiol. 2008;52(4):273–8.CrossRef Neumann T, Vogt J, Schumacher B, Dorszewski A, Kuniss M, Neuser H, et al. Circumferential pulmonary vein isolation with the cryoballoon technique results from a prospective 3-center study. J Am Coll Cardiol. 2008;52(4):273–8.CrossRef
15.
Zurück zum Zitat Ottaviano L, Chierchia GB, Bregasi A, Bruno N, Antonelli A, Alsheraei AT, et al. Cryoballoon ablation for atrial fibrillation guided by real-time three-dimensional transoesophageal echocardiography: a feasibility study. Europace. 2013;15(7):944–50.CrossRef Ottaviano L, Chierchia GB, Bregasi A, Bruno N, Antonelli A, Alsheraei AT, et al. Cryoballoon ablation for atrial fibrillation guided by real-time three-dimensional transoesophageal echocardiography: a feasibility study. Europace. 2013;15(7):944–50.CrossRef
16.
Zurück zum Zitat Takami M, Misiri J, Lehmann HI, Parker KD, Johnson SB, Sarmiento RI, et al. Spatial and time-course thermodynamics during pulmonary vein isolation using the second-generation cryoballoon in a canine in vivo model. Circ Arrhythm Electrophysiol. 2015;8(1):186–92.CrossRef Takami M, Misiri J, Lehmann HI, Parker KD, Johnson SB, Sarmiento RI, et al. Spatial and time-course thermodynamics during pulmonary vein isolation using the second-generation cryoballoon in a canine in vivo model. Circ Arrhythm Electrophysiol. 2015;8(1):186–92.CrossRef
17.
Zurück zum Zitat Ghosh J, Martin A, Keech AC, Chan KH, Gomes S, Singarayar S, et al. Balloon warming time is the strongest predictor of late pulmonary vein electrical reconnection following cryoballoon ablation for atrial fibrillation. Heart Rhythm. 2013;10(9):1311–7.CrossRef Ghosh J, Martin A, Keech AC, Chan KH, Gomes S, Singarayar S, et al. Balloon warming time is the strongest predictor of late pulmonary vein electrical reconnection following cryoballoon ablation for atrial fibrillation. Heart Rhythm. 2013;10(9):1311–7.CrossRef
18.
Zurück zum Zitat Goldberger JJ, Mitrani RD. Cryoballoon ablation for atrial fibrillation: double jeopardy? J Cardiovasc Electrophysiol. 2016;27(12):1381–3.CrossRef Goldberger JJ, Mitrani RD. Cryoballoon ablation for atrial fibrillation: double jeopardy? J Cardiovasc Electrophysiol. 2016;27(12):1381–3.CrossRef
19.
Zurück zum Zitat John RM, Kapur S, Ellenbogen KA, Koneru JN. Atrioesophageal fistula formation with cryoballoon ablation is most commonly related to the left inferior pulmonary vein. Heart Rhythm. 2017;14(2):184–9.CrossRef John RM, Kapur S, Ellenbogen KA, Koneru JN. Atrioesophageal fistula formation with cryoballoon ablation is most commonly related to the left inferior pulmonary vein. Heart Rhythm. 2017;14(2):184–9.CrossRef
Metadaten
Titel
Dosing of the second-generation cryoballoon using acute time-to-pulmonary vein isolation as an indicator of durable ablation in a canine model
verfasst von
Wilber Su
Nicolas Coulombe
Nicole Kirchhof
Erin Grassl
Dan Wittenberger
Publikationsdatum
14.04.2018
Verlag
Springer US
Erschienen in
Journal of Interventional Cardiac Electrophysiology / Ausgabe 3/2018
Print ISSN: 1383-875X
Elektronische ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-018-0346-y

Weitere Artikel der Ausgabe 3/2018

Journal of Interventional Cardiac Electrophysiology 3/2018 Zur Ausgabe

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

Niedriger diastolischer Blutdruck erhöht Risiko für schwere kardiovaskuläre Komplikationen

25.04.2024 Hypotonie Nachrichten

Wenn unter einer medikamentösen Hochdrucktherapie der diastolische Blutdruck in den Keller geht, steigt das Risiko für schwere kardiovaskuläre Ereignisse: Darauf deutet eine Sekundäranalyse der SPRINT-Studie hin.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Adipositas-Medikament auch gegen Schlafapnoe wirksam

24.04.2024 Adipositas Nachrichten

Der als Antidiabetikum sowie zum Gewichtsmanagement zugelassene Wirkstoff Tirzepatid hat in Studien bei adipösen Patienten auch schlafbezogene Atmungsstörungen deutlich reduziert, informiert der Hersteller in einer Vorab-Meldung zum Studienausgang.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.