Echocardiographic parameters to predict inadequate defibrillation safety margin in patients receiving implantable cardioverter defibrillators for primary prevention

verfasst von: Sachin Kumar Amruthlal Jain, Hamid Ghanbari, Rayan Hourani, Timothy R. Larsen, Marcos Daccarett, Christian Machado

Erschienen in: Journal of Interventional Cardiac Electrophysiology | Ausgabe 1/2013

Einloggen, um Zugang zu erhalten

Abstract

Background

Implantable cardioverter defibrillators (ICDs) have become an important part of the management of patients with congestive heart failure. At the time of ICD implantation, ventricular fibrillation (VF) is induced to assess adequate energy required for defibrillation. There are multiple parameters which influence the defibrillation safety margin (DSM); however, these factors are not well-established when ICDs are implanted for the primary prevention of sudden cardiac death (SCD) in patients with severe systolic dysfunction. We evaluated multiple clinical and echocardiographic parameters as predictors of adequate DSM in patients referred for ICD implantation for the primary prevention of SCD.

Methods

We prospectively enrolled 41 patients for ICD implantation with clinical indications for the primary prevention of SCD. Two blinded independent readers evaluated the prespecified echocardiographic parameters. These included left ventricular (LV) mass, indices of right ventricular and LV systolic and diastolic functions, and LV geometric dimensions. Basic clinical demographics, including age, gender, comorbidities, and etiology of cardiomyopathy, were also evaluated. DSM was established using our standard protocol for defibrillation testing which includes VF with successful first shock terminating VF at a value at least 10 J below the maximum output of the implanted device. High defibrillation thresholds (DFT) were defined as >21 J.

Results

The mean age is 61.8 ± 14.7 years, with men comprising the majority of the patients (73 %). The only clinical variables which predicted the high DFT were age (in years) (54.5 ± 17.5 vs. 65.7 ± 11.3, p = 0.044), QRS duration (in milliseconds) (116.0 ± 29.5 vs. 110.5 ± 21.8, p = 0.03), LV mass (in grams) (241.0 ± 77.9 vs. 181.9 ± 52.3, p = 0.006), and LV mass index (in grams per square meter) (111.1 ± 38.2 vs. 86.4 ± 21.1, p = 0.02). On multivariate logistic regression analysis, LV mass was the only independent predictor of low DFT (≤22 J) in patients with ICD implanted for the primary prevention of SCD.

Conclusion

LV mass may help predict an adequate DSM in patients who are referred for ICD implantation for the primary prevention of SCD. These results may help distinguish the patients who may require high-energy devices prior to the implantation procedure. These results may help distinguish patients requiring high-energy devices, coils, or advanced programming prior to implantation and appropriate referral to electrophysiologists.
Literatur
1.
Zurück zum Zitat Epstein, A. E., DiMarco, J. P., Ellenbogen, K. A., Estes, N. A., 3rd, Freedman, R. A., Gettes, L. S., Gillinov, A. M., Gregoratos, G., Hammill, S. C., Hayes, D. L., Hlatky, M. A., Newby, L. K., Page, R. L., Schoenfeld, M. H., Silka, M. J., Stevenson, L. W., Sweeney, M. O., Smith, S. C., Jr., Jacobs, A. K., Adams, C. D., Anderson, J. L., Buller, C. E., Creager, M. A., Ettinger, S. M., Faxon, D. P., Halperin, J. L., Hiratzka, L. F., Hunt, S. A., Krumholz, H. M., Kushner, F. G., Lytle, B. W., Nishimura, R. A., Ornato, J. P., Page, R. L., Riegel, B., Tarkington, L. G., & Yancy, C. W. (2008). ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: A report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices) developed in collaboration with the American Association for Thoracic Surgery and Society of Thoracic Surgeons. Journal of the American College of Cardiology, 51, e1–e62.PubMedCrossRef Epstein, A. E., DiMarco, J. P., Ellenbogen, K. A., Estes, N. A., 3rd, Freedman, R. A., Gettes, L. S., Gillinov, A. M., Gregoratos, G., Hammill, S. C., Hayes, D. L., Hlatky, M. A., Newby, L. K., Page, R. L., Schoenfeld, M. H., Silka, M. J., Stevenson, L. W., Sweeney, M. O., Smith, S. C., Jr., Jacobs, A. K., Adams, C. D., Anderson, J. L., Buller, C. E., Creager, M. A., Ettinger, S. M., Faxon, D. P., Halperin, J. L., Hiratzka, L. F., Hunt, S. A., Krumholz, H. M., Kushner, F. G., Lytle, B. W., Nishimura, R. A., Ornato, J. P., Page, R. L., Riegel, B., Tarkington, L. G., & Yancy, C. W. (2008). ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: A report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices) developed in collaboration with the American Association for Thoracic Surgery and Society of Thoracic Surgeons. Journal of the American College of Cardiology, 51, e1–e62.PubMedCrossRef
2.
Zurück zum Zitat Mark, D. B., Nelson, C. L., Anstrom, K. J., Al-Khatib, S. M., Tsiatis, A. A., Cowper, P. A., Clapp-Channing, N. E., Davidson-Ray, L., Poole, J. E., Johnson, G., Anderson, J., Lee, K. L., & Bardy, G. H. (2006). Cost-effectiveness of defibrillator therapy or amiodarone in chronic stable heart failure: Results from the Sudden Cardiac Death in Heart Failure Trial (SCD-HeFT). Circulation, 114, 135–142.PubMedCrossRef Mark, D. B., Nelson, C. L., Anstrom, K. J., Al-Khatib, S. M., Tsiatis, A. A., Cowper, P. A., Clapp-Channing, N. E., Davidson-Ray, L., Poole, J. E., Johnson, G., Anderson, J., Lee, K. L., & Bardy, G. H. (2006). Cost-effectiveness of defibrillator therapy or amiodarone in chronic stable heart failure: Results from the Sudden Cardiac Death in Heart Failure Trial (SCD-HeFT). Circulation, 114, 135–142.PubMedCrossRef
3.
Zurück zum Zitat Curtis, A. B., Ellenbogen, K. A., Hammill, S. C., Hayes, D. L., Reynolds, D. W., Wilber, D. J., & Cain, M. E. (2004). Clinical competency statement: Training pathways for implantation of cardioverter defibrillators and cardiac resynchronization devices. Heart Rhythm, 3, 371–375.CrossRef Curtis, A. B., Ellenbogen, K. A., Hammill, S. C., Hayes, D. L., Reynolds, D. W., Wilber, D. J., & Cain, M. E. (2004). Clinical competency statement: Training pathways for implantation of cardioverter defibrillators and cardiac resynchronization devices. Heart Rhythm, 3, 371–375.CrossRef
4.
Zurück zum Zitat Gerstenfeld, E. P. (2005). Defibrillation threshold testing: Is one shock enough? Heart Rhythm, 2, 123–124.PubMedCrossRef Gerstenfeld, E. P. (2005). Defibrillation threshold testing: Is one shock enough? Heart Rhythm, 2, 123–124.PubMedCrossRef
5.
Zurück zum Zitat Birnie, D., Tung, S., Simpson, C., Crystal, E., Exner, D., Ayala Paredes, F. A., Krahn, A., Parkash, R., Khaykin, Y., Philippon, F., Guerra, P., Kimber, S., Cameron, D., & Healey, J. S. (2008). Complications associated with defibrillation threshold testing: The Canadian experience. Heart Rhythm, 5, 387–390.PubMedCrossRef Birnie, D., Tung, S., Simpson, C., Crystal, E., Exner, D., Ayala Paredes, F. A., Krahn, A., Parkash, R., Khaykin, Y., Philippon, F., Guerra, P., Kimber, S., Cameron, D., & Healey, J. S. (2008). Complications associated with defibrillation threshold testing: The Canadian experience. Heart Rhythm, 5, 387–390.PubMedCrossRef
6.
Zurück zum Zitat Marchlinski, F. E., Flores, B., Miller, J. M., Gottlieb, C. D., & Hargrove, W. C., 3rd. (1988). Relation of the intraoperative defibrillation threshold to successful postoperative defibrillation with an automatic implantable cardioverter defibrillator. The American Journal of Cardiology, 62, 393–398.PubMedCrossRef Marchlinski, F. E., Flores, B., Miller, J. M., Gottlieb, C. D., & Hargrove, W. C., 3rd. (1988). Relation of the intraoperative defibrillation threshold to successful postoperative defibrillation with an automatic implantable cardioverter defibrillator. The American Journal of Cardiology, 62, 393–398.PubMedCrossRef
7.
Zurück zum Zitat Russo, A. M., Sauer, W., Gerstenfeld, E. P., Hsia, H. H., Lin, D., Cooper, J. M., Dixit, S., Verdino, R. J., Nayak, H. M., Callans, D. J., Patel, V., & Marchlinski, F. E. (2005). Defibrillation threshold testing: Is it really necessary at the time of implantable cardioverter-defibrillator insertion? Heart Rhythm, 2, 456–461.PubMedCrossRef Russo, A. M., Sauer, W., Gerstenfeld, E. P., Hsia, H. H., Lin, D., Cooper, J. M., Dixit, S., Verdino, R. J., Nayak, H. M., Callans, D. J., Patel, V., & Marchlinski, F. E. (2005). Defibrillation threshold testing: Is it really necessary at the time of implantable cardioverter-defibrillator insertion? Heart Rhythm, 2, 456–461.PubMedCrossRef
8.
Zurück zum Zitat Mainigi, S. K., Cooper, J. M., Russo, A. M., Nayak, H. M., Lin, D., Dixit, S., Gerstenfeld, E. P., Hsia, H. H., Callans, D. J., Marchlinski, F. E., & Verdino, R. J. (2006). Elevated defibrillation thresholds in patients undergoing biventricular defibrillator implantation: Incidence and predictors. Heart Rhythm, 3, 1010–1016.PubMedCrossRef Mainigi, S. K., Cooper, J. M., Russo, A. M., Nayak, H. M., Lin, D., Dixit, S., Gerstenfeld, E. P., Hsia, H. H., Callans, D. J., Marchlinski, F. E., & Verdino, R. J. (2006). Elevated defibrillation thresholds in patients undergoing biventricular defibrillator implantation: Incidence and predictors. Heart Rhythm, 3, 1010–1016.PubMedCrossRef
9.
Zurück zum Zitat Raitt, M. H., Johnson, G., Dolack, G. L., Poole, J. E., Kudenchuk, P. J., & Bardy, G. H. (1995). Clinical predictors of the defibrillation threshold with the unipolar implantable defibrillation system. Journal of the American College of Cardiology, 25, 1576–1583.PubMedCrossRef Raitt, M. H., Johnson, G., Dolack, G. L., Poole, J. E., Kudenchuk, P. J., & Bardy, G. H. (1995). Clinical predictors of the defibrillation threshold with the unipolar implantable defibrillation system. Journal of the American College of Cardiology, 25, 1576–1583.PubMedCrossRef
10.
Zurück zum Zitat Kopp, D. E., Blakeman, B. P., Kall, J. G., Olshansky, B., Kinder, C. A., & Wilber, D. J. (1995). Predictors of defibrillation energy requirements with nonepicardial lead systems. Pacing and Clinical Electrophysiology, 18, 253–260.PubMedCrossRef Kopp, D. E., Blakeman, B. P., Kall, J. G., Olshansky, B., Kinder, C. A., & Wilber, D. J. (1995). Predictors of defibrillation energy requirements with nonepicardial lead systems. Pacing and Clinical Electrophysiology, 18, 253–260.PubMedCrossRef
11.
Zurück zum Zitat Burke, P. T., Ghanbari, H., Alexander, P. B., Shaw, M. K., Daccarett, M., & Machado, C. (2010). A protocol for patients with cardiovascular implantable devices undergoing magnetic resonance imaging (MRI): Should defibrillation threshold testing be performed post-(MRI). Journal of Interventional Cardiac Electrophysiology, 28, 59–66.PubMedCrossRef Burke, P. T., Ghanbari, H., Alexander, P. B., Shaw, M. K., Daccarett, M., & Machado, C. (2010). A protocol for patients with cardiovascular implantable devices undergoing magnetic resonance imaging (MRI): Should defibrillation threshold testing be performed post-(MRI). Journal of Interventional Cardiac Electrophysiology, 28, 59–66.PubMedCrossRef
12.
Zurück zum Zitat Lang, R. M., Bierig, M., Devereux, R. B., Flachskampf, F. A., Foster, E., Pellikka, P. A., Picard, M. H., Roman, M. J., Seward, J., Shanewise, J. S., Solomon, S. D., Spencer, K. T., Sutton, M. S., & Stewart, W. J. (2005). Recommendations for chamber quantification: A report from the American Society of Echocardiography’s Guidelines and Standards Committee and the Chamber Quantification Writing Group, developed in conjunction with the European Association of Echocardiography, a branch of the European Society of Cardiology. Journal of the American Society of Echocardiography, 18, 1440–1463.PubMedCrossRef Lang, R. M., Bierig, M., Devereux, R. B., Flachskampf, F. A., Foster, E., Pellikka, P. A., Picard, M. H., Roman, M. J., Seward, J., Shanewise, J. S., Solomon, S. D., Spencer, K. T., Sutton, M. S., & Stewart, W. J. (2005). Recommendations for chamber quantification: A report from the American Society of Echocardiography’s Guidelines and Standards Committee and the Chamber Quantification Writing Group, developed in conjunction with the European Association of Echocardiography, a branch of the European Society of Cardiology. Journal of the American Society of Echocardiography, 18, 1440–1463.PubMedCrossRef
13.
Zurück zum Zitat Uyguanco, E. R., Berger, A., Budzikowski, A. S., Gunsburg, M., & Kassotis, J. (2008). Management of high defibrillation threshold. Expert Rev Cadiovasc Ther, 6, 1237–1248.CrossRef Uyguanco, E. R., Berger, A., Budzikowski, A. S., Gunsburg, M., & Kassotis, J. (2008). Management of high defibrillation threshold. Expert Rev Cadiovasc Ther, 6, 1237–1248.CrossRef
14.
Zurück zum Zitat Chen, P. S., Wolf, P. D., & Ideker, R. E. (1991). Mechanism of cardiac defibrillation: A different point of view. Circulation, 84, 913–919.PubMedCrossRef Chen, P. S., Wolf, P. D., & Ideker, R. E. (1991). Mechanism of cardiac defibrillation: A different point of view. Circulation, 84, 913–919.PubMedCrossRef
15.
Zurück zum Zitat Rodriguez, B., Eason, J. C., & Trayanova, N. A. (2006). Differences between left and right ventricular anatomy determine the type of reentrant circuits induced by an external electric shock. A rabbit heart simulation study. Progress in Biophysics and Molecular Biology, 90, 399–413.PubMedCrossRef Rodriguez, B., Eason, J. C., & Trayanova, N. A. (2006). Differences between left and right ventricular anatomy determine the type of reentrant circuits induced by an external electric shock. A rabbit heart simulation study. Progress in Biophysics and Molecular Biology, 90, 399–413.PubMedCrossRef
16.
Zurück zum Zitat Chapman, P. D., Sagar, K. B., Wetherbee, J. N., & Troup, P. J. (1987). Relationship of left ventricular mass to defibrillation threshold for the implantable defibrillator: A combined clinical and animal study. American Heart Journal, 114, 274–278.PubMedCrossRef Chapman, P. D., Sagar, K. B., Wetherbee, J. N., & Troup, P. J. (1987). Relationship of left ventricular mass to defibrillation threshold for the implantable defibrillator: A combined clinical and animal study. American Heart Journal, 114, 274–278.PubMedCrossRef
17.
Zurück zum Zitat Hodgson, D. M., Olsovsky, M. R., Shorofsky, S. R., Daly, B., & Gold, M. R. (2002). Clinical predictors of defibrillation thresholds with an active pectoral pulse generator lead system. PACE, 25, 408–413.PubMedCrossRef Hodgson, D. M., Olsovsky, M. R., Shorofsky, S. R., Daly, B., & Gold, M. R. (2002). Clinical predictors of defibrillation thresholds with an active pectoral pulse generator lead system. PACE, 25, 408–413.PubMedCrossRef
18.
Zurück zum Zitat Horton, R. P., Canby, R. C., Roman, C. A., Hull, M. L., Kaye, S. A., Jessen, M. E., & Page, R. L. (1997). Determinants of nonthoracotomy biphasic defibrillation. PACE, 20, 60–64.PubMedCrossRef Horton, R. P., Canby, R. C., Roman, C. A., Hull, M. L., Kaye, S. A., Jessen, M. E., & Page, R. L. (1997). Determinants of nonthoracotomy biphasic defibrillation. PACE, 20, 60–64.PubMedCrossRef
19.
Zurück zum Zitat Lubinski, A., Lewicka-Nowak, E., Zienciuk, A., Krolak, T., Kempa, M., Pazdyga, A., & Swiatecka, G. (2005). Clinical predictors of defibrillation threshold in patients with implantable cardioverter-defibrillators. Kardiologia Polska, 62, 317–328.PubMed Lubinski, A., Lewicka-Nowak, E., Zienciuk, A., Krolak, T., Kempa, M., Pazdyga, A., & Swiatecka, G. (2005). Clinical predictors of defibrillation threshold in patients with implantable cardioverter-defibrillators. Kardiologia Polska, 62, 317–328.PubMed
20.
Zurück zum Zitat Schuger, C., Ellenbogen, K. A., Faddis, M., Knight, B. P., Yong, P., & Sample, R. (2006). Defibrillation energy requirements in an ICD population receiving cardiac resynchronization therapy. Journal of Cardiovascular Electrophysiology, 17, 247–250.PubMedCrossRef Schuger, C., Ellenbogen, K. A., Faddis, M., Knight, B. P., Yong, P., & Sample, R. (2006). Defibrillation energy requirements in an ICD population receiving cardiac resynchronization therapy. Journal of Cardiovascular Electrophysiology, 17, 247–250.PubMedCrossRef
21.
Zurück zum Zitat Kolb, C., Tzeis, S., & Zrenner, B. (2009). Defibrillation threshold testing: Tradition or necessity? Pacing and Clinical Electrophysiology, 32, 570–572.PubMedCrossRef Kolb, C., Tzeis, S., & Zrenner, B. (2009). Defibrillation threshold testing: Tradition or necessity? Pacing and Clinical Electrophysiology, 32, 570–572.PubMedCrossRef
22.
Zurück zum Zitat Strickberger, S. A., & Klein, G. J. (2004). Is defibrillation testing required for defibrillator implantation? Journal of the American College of Cardiology, 44, 88–91.PubMedCrossRef Strickberger, S. A., & Klein, G. J. (2004). Is defibrillation testing required for defibrillator implantation? Journal of the American College of Cardiology, 44, 88–91.PubMedCrossRef
Metadaten
Titel
Echocardiographic parameters to predict inadequate defibrillation safety margin in patients receiving implantable cardioverter defibrillators for primary prevention
verfasst von
Sachin Kumar Amruthlal Jain
Hamid Ghanbari
Rayan Hourani
Timothy R. Larsen
Marcos Daccarett
Christian Machado
Publikationsdatum
01.06.2013
Verlag
Springer US
Erschienen in
Journal of Interventional Cardiac Electrophysiology / Ausgabe 1/2013
Print ISSN: 1383-875X
Elektronische ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-012-9764-4

Weitere Artikel der Ausgabe 1/2013

Journal of Interventional Cardiac Electrophysiology 1/2013 Zur Ausgabe

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

Niedriger diastolischer Blutdruck erhöht Risiko für schwere kardiovaskuläre Komplikationen

25.04.2024 Hypotonie Nachrichten

Wenn unter einer medikamentösen Hochdrucktherapie der diastolische Blutdruck in den Keller geht, steigt das Risiko für schwere kardiovaskuläre Ereignisse: Darauf deutet eine Sekundäranalyse der SPRINT-Studie hin.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Adipositas-Medikament auch gegen Schlafapnoe wirksam

24.04.2024 Adipositas Nachrichten

Der als Antidiabetikum sowie zum Gewichtsmanagement zugelassene Wirkstoff Tirzepatid hat in Studien bei adipösen Patienten auch schlafbezogene Atmungsstörungen deutlich reduziert, informiert der Hersteller in einer Vorab-Meldung zum Studienausgang.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.