Effect of closed endotracheal suction in high-frequency ventilated premature infants measured with electrical impedance tomography

verfasst von: Mariëtte B. van Veenendaal, Martijn Miedema, Frans H. C. de Jongh, Johanna H. van der Lee, Inez Frerichs, Anton H. van Kaam

Erschienen in: Intensive Care Medicine | Ausgabe 12/2009

Abstract

Objective

To determine the global and regional changes in lung volume during and after closed endotracheal tube (ETT) suction in high-frequency ventilated preterm infants with respiratory distress syndrome (RDS).

Design

Prospective observational clinical study.

Setting

Neonatal intensive care unit.

Patients

Eleven non-muscle relaxed preterm infants with RDS ventilated with open lung high-frequency ventilation (HFV).

Interventions

Closed ETT suction.

Measurements and results

Changes in global and regional lung volume were measured with electrical impedance tomography. ETT suction resulted in an acute loss of lung volume followed by spontaneous recovery with a median residual loss of 3.3% of the maximum volume loss. The median stabilization time was 8 s. At the regional level, the lung volume changes during and after ETT suction were heterogeneous in nature.

Conclusions

Closed ETT suction causes an acute, transient and heterogeneous loss of lung volume in premature infants with RDS treated with open lung HFV.

Introduction

Bronchopulmonary dysplasia (BPD) is a common complication of premature birth, and (regional) alveolar collapse and overdistension during mechanical ventilation are considered important risk factors [1]. In an attempt to reduce the incidence of BPD in preterm infants, high-frequency ventilation (HFV), using a so-called open lung ventilation strategy aiming to minimize alveolar collapse and overdistension, has been advocated [2, 3].
Endotracheal tube (ETT) suction is essential in ventilated patients to maintain airway patency. However, ETT suction may also lead to (transient) hypoxia and cardiovascular instability, side effects generally attributed to loss of lung volume due to atelectasis [46]. This increased risk of atelectasis during ETT suction may compromise the efficacy of open lung ventilation to attenuate lung injury.
Only two studies have explored the effect of ETT suction on lung volume in newborn infants on HFV, using respiratory inductive plethysmography (RIP) [7, 8]. However, these studies included relatively mature, and in some cases muscle relaxed, newborn infants ventilated for various causes of respiratory failure. The results may be different in non-muscle relaxed preterm infants with respiratory distress syndrome (RDS). Furthermore, RIP only measures changes in total lung volume, providing no information on regional atelectasis and overdistension.
Electrical impedance tomography (EIT) is a new non-invasive technique measuring both global and regional changes in lung impedance, which are highly correlated with changes in gas volume [9]. The aim of this study was to examine the effect of closed ETT suction on the relative changes in both global and regional lung impedance in preterm infants with RDS treated with open lung HFV.

Methods

Subjects

The study was performed in the neonatal intensive care unit of the Emma Children’s Hospital AMC in Amsterdam and approved by the Institutional Review Board. Premature infants with RDS treated with primary HFV were included after obtaining written informed consent from both parents.
HFV was delivered with a Sensormedics 3100A oscillator (Cardinal Health, Yorba Linda, CA) and at all times combined with an individualized open lung ventilation strategy aiming to recruit and stabilize collapsed alveoli with the lowest possible airway pressure [2].

Suction protocol

ETT suction, using a neonatal closed tracheal suction system (Trach Care, Ballard Medical Products, Draper, UT), was performed in supine position with the patient’s head in the midline. Catheter sizes were standardized according to ETT size. Following saline (0.3 ml) instillation, the catheter was inserted to the tip of the ETT, and suctioning was applied for approximately 3 s at −75 mmHg. Next, the catheter was withdrawn while continuing suctioning.

EIT measurements and analysis

Changes in lung impedance and airway pressures were continuously recorded from 60 s before until 120 s after ETT suction with the “Goettingen Goe-MF II” EIT system (Cardinal Health, Yorba Linda, CA) using an electrical current of 5 mArms (50 kHz) and a scan rate of 44 Hz [10].
EIT data were analyzed off-line. Each recording was divided into three phases: (1) pre-suction (baseline), defined as −40 to −10 s before starting ETT suction; (2) suction, defined as the period in which negative pressure was applied; (3) post-suction, defined as +5 to +30 s after stabilization of the EIT signal. If the EIT signal was unstable, a stable time period was selected as close as possible to the predefined time windows. The post-suction EIT signal was considered stabilized when no visual increases were seen at the peaks and troughs of the signal.
Using the pre-suction phase as the reference recording, the mean global change in lung impedance after ETT suction was calculated by averaging all unfiltered impedance values. We also determined the lowest impedance level in the suction phase. With these values we could determine the residual global loss of impedance after suction as a percentage of the maximum global loss during suction. We also calculated the maximum and residual impedance loss in four regions of interest (ROI), i.e., ventral, dorsal, left and right. The time necessary to recover 80% of the maximum loss from the time point of maximum impedance loss and the time necessary for the EIT signal to stabilize after ETT suction was stopped were calculated.

Data collection

We collected the following parameters: transcutaneous oxygen saturation (SpO2) before, during and after ETT suction, episodes of bradycardia (heart rate below 100 bpm) during ETT suction and ventilator settings at the start of ETT suction. Adjustments in ventilator settings during the study were recorded.

Statistical analysis

Data are presented as median and interquartile range (IQR). Differences in SpO2 before and during ETT suction were analyzed with the Wilcoxon signed ranks test. A P value less than 0.05 is considered significant.

Results

Patient characteristics

Twelve patients were studied. One patient had to be excluded because the EIT signal before suction was not stable. Table 1 shows the characteristics of the remaining 11 patients. The indication for ETT suction was either routine suctioning before surfactant administration (n = 10) or clinical signs of secretion in the upper airways and/or the ETT (n = 1).
Table 1
Patient characteristics
Infant no.
Diagnosis
GA (weeks)
Weight (kg)
Age (h)
ETT size (mm)
Cat size (Fr)
Areaa (mm2)
CDP (cmH2O)
Amplitude (cmH2O)
FiO2
1
RDS
25.3
0.83
14
3.0
6
4.0
14
25
0.25
2
RDS
27.3
0.98
3
3.0
6
4.0
14
18
0.21
3
RDS + PPHN
29.3
1.37
45
3.5
7
5.3
12
24
0.80
4
RDS
31.1
1.44
48
3.0
6
4.0
9
19
0.22
5
RDS
33.6
2.12
71
3.0
6
4.0
8
24
0.25
6
RDS
28.3
1.30
9
3.0
6
4.0
18
29
0.24
7
RDS
31.4
1.39
9
3.5
7
5.3
18
22
0.30
8
RDS + PPHN
31.4
1.83
33
3.5
7
5.3
14
28
0.55
9
RDS
25.6
0.79
10
2.5
5
2.7
16
22
0.30
10
RDS + PPHN
31.0
1.60
36
3.0
6
4.0
13
21
0.60
11
RDS
28.4
1.12
7
3.0
6
4.0
18
25
0.35
Median
 
29.3
1.37
14
3.0
6
4.2
14
24
0.30
GA Gestational age, ETT endotracheal tube, Cat suction catheter, CDP continuous distending pressure, FiO 2 fraction of inspired oxygen, RDS respiratory distress syndrome, PPHN persistent pulmonary hypertensions of the newborn
aUnoccluded cross-sectional area of the ETT during suctioning

EIT results

In all patients ETT suction resulted in a steep reduction of global lung impedance followed by a more gradual recovery once suctioning was stopped (Fig. 1, Table 2). Comparing the ROI showed that the distribution of the maximum impedance loss was heterogeneous in nature in almost all patients and that there was no consistent pattern favoring one of these regions.
Table 2
Changes in impedance during and after closed endotracheal tube suction compared to the baseline (zero) values
Infant no.
Maximum impedance loss during suction
Residual impedance loss after suction
Global ΔZ
Region of interest (% of maximum global loss)
Global ΔZ (% of max loss)
Region of interest (% of residual global loss)
V
D
R
L
V
D
R
L
1
−51
−31 (59)
−20 (41)
−24 (48)
−26 (52)
−5.1 (10)
−4.5 (88)
−0.6 (12)
−3.7 (72)
−1.5 (28)
2
−68
−39 (57)
−29 (43)
−33 (48)
−35 (52)
−2.3 (3)
−2.3a
0.07a
−1.2 (51)
−1.1 (49)
3
−43
−15 (35)
−28 (65)
−24 (55)
−19 (45)
6.2b
3.7 (60)
2.5 (40)
2.1 (34)
4.1 (66)
4
−62
−31 (50)
−31 (50)
−35 (57)
−27 (43)
−1.8 (3)
1.9a
−3.7a
−1.4 (76)
−0.4 (24)
5
−58
−41 (70)
−17 (30)
−27 (46)
−31 (54)
0.02 (0)
0.5a
−0.4a
−1.3a
1.3a
6
−67
−31 (46)
−36 (54)
−46 (69)
−21 (31)
−0.1 (0.2)
−0.3a
0.2a
−1.2a
1.1a
7
−34
−17 (50)
−17 (50)
−19 (57)
−15 (43)
−3.7 (11)
−0.3 (7)
−3.4 (92)
−2.4 (65)
−1.3 (35)
8
−40
−22 (55)
−18 (45)
−21 (52)
−19 (48)
−3.5 (9)
−2.8 (81)
−0.7 (19)
−0.8 (23)
−2.7 (77)
9
−59
−35 (60)
−24 (40)
−31 (52)
−28 (48)
−8.7 (15)
−3.4 (39)
−5.3 (61)
−4.1 (47)
−4.6 (53)
10
−49
−25 (51)
−24 (49)
−24 (49)
−25 (51)
−6.6 (13)
−4.2 (63)
−2.4 (37)
−4.0 (61)
−2.6 (39)
11
−48
−32 (66)
−16 (34)
−27 (56)
−21 (44)
9.3b
2.3 (25)
7.0 (75)
3.5 (37)
5.8 (63)
Median
−51
−31 (55)
−24 (45)
−27 (52)
−25 (48)
−2.3 (3)
−0.3
−0.6
−1.3
−1.1
V Ventral, D dorsal, R right, L left
aRegions with volume loss and increase after endotracheal suction (percentages not shown because of the opposite regional values)
bImpedance stabilized at higher level after suction compared to baseline
In all patients the global impedance stabilized within 60 s after suctioning, with a median stabilization time of 8 s (IQR 3–29 s). The median time needed to recover 80% of the maximum loss from the time at which this maximum volume loss occurred was 8 s (IQR 4–14 s).
In 7 of the 11 patients (64%) global impedance stabilized below pre-suction values, and 3 patients failed to ever recover 90% of the impedance lost during suctioning (Table 2). In the remaining patients global impedance returned to baseline (n = 2) or stabilized at higher levels (n = 2). The median residual loss in global impedance after stabilization for all 11 patients was 3.3% of the maximum loss during suction (IQR 0–11%). Again, the loss of impedance after stabilization showed an inconsistent and heterogeneous pattern across the different ROIs.

Physiological data

ETT suctioning resulted in a transient decrease in the median SpO2 from 95% (IQR 94–98%) to 90% (IQR 89–94%) (P < 0.01). Bradycardia was not observed in any of the patients. No adjustments were made in the ventilator settings. In the majority of infants spontaneous breathing was absent or limited to just a few breaths per minute.

Discussion

This study shows that closed ETT suction during open lung HFV in non-muscle relaxed preterm infants with RDS results in an acute and heterogeneous loss of impedance followed by a gradual spontaneous recovery within the first minute after suctioning. We did not observe major hypoxia or cardiovascular instability.
Animal and human adult studies have shown that the changes in lung impedance during mechanical ventilation are highly correlated with changes in air content seen on computed tomography [9, 11, 12]. Although comparable studies in newborn infants are difficult to perform and therefore currently not available, it seems plausible that these observations also apply to ventilated newborn infants and that the observed changes in lung impedance as reported in the present study are indeed changes in lung aeration.
Our observed changes in global lung volume are strikingly similar to previous studies using RIP to measure changes in lung volume following ETT suction in more mature and, in some cases, muscle relaxed newborn infants ventilated for various causes of respiratory failure [7, 8]. At first glance, these consistent results seem to indicate that atelectasis caused by ETT suctioning during open lung HFV is only transient and that a recruitment maneuver to restore lung volume and attenuate possible lung injury is often not necessary. However, unlike RIP, EIT also provides important information on the regional distribution. To our knowledge, this is the first study that shows that the distribution of lung volume changes after ETT suction in ventilated preterm infants with RDS is heterogeneous in nature. Similar heterogeneity following ETT suction has also been described in children with acute respiratory distress syndrome [13].
Although our study cannot measure the exact volume states of different lung regions, our findings do suggest that recovery in global lung volume after ETT suctioning may still be accompanied by regional atelectasis and over-distension and thus ongoing lung injury. It might well be that these patients can indeed benefit from a recruitment maneuver aiming to reestablish the homogeneity in regional air distribution. Future studies will have to investigate the efficacy of such a maneuver.
It was interesting to observe that patients with the largest tube size showed the smallest decrease in lung volume during ETT suctioning. Although this could be explained by the larger unoccluded cross-sectional area of the ETT in these patients [14], the small sample size of this study precludes reliable analysis and thus firm conclusions.
This study has several limitations that need to be addressed. First of all, this study does not provide absolute lung volume changes in response to ETT suction. Because the main focus of this study was to determine the relative changes in lung volume following ETT suction, we decided not to subject the patients to a period of possibly injurious positive pressure ventilation with the sole purpose of calibrating the impedance changes. Secondly, this study mainly included surfactant-deficient preterm infants. The result may be different in other groups of patients, although the consistency of our findings with previous studies using RIP seems reassuring. Finally, the effect of ETT suction on lung volume may vary according to the suction protocol.

Conclusion

Closed ETT suction causes an acute and heterogeneous loss of lung volume in non-muscle relaxed premature infants with RDS treated with open lung HFV. Although there is an almost complete spontaneous recovery of the global volume loss within the first minute after ETT suction, at the regional level this recovery is heterogeneous in nature.

Acknowledgments

This work was financially supported by Cardinal Health Care and Nycomed, The Netherlands.

Conflict of interest statement

The authors have not disclosed any potential conflicts of interest.

Open Access

This article is distributed under the terms of the Creative Commons Attribution Noncommercial License which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.
Open AccessThis is an open access article distributed under the terms of the Creative Commons Attribution Noncommercial License (https://​creativecommons.​org/​licenses/​by-nc/​2.​0), which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Innere Medizin

Kombi-Abonnement

Mit e.Med Innere Medizin erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Innere Medizin, den Premium-Inhalten der internistischen Fachzeitschriften, inklusive einer gedruckten internistischen Zeitschrift Ihrer Wahl.

e.Med Anästhesiologie

Kombi-Abonnement

Mit e.Med Anästhesiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes AINS, den Premium-Inhalten der AINS-Fachzeitschriften, inklusive einer gedruckten AINS-Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Clark RH, Slutsky AS, Gerstmann DR (2000) Lung protective strategies of ventilation in the neonate: what are they? Pediatrics 105:112–114CrossRefPubMed Clark RH, Slutsky AS, Gerstmann DR (2000) Lung protective strategies of ventilation in the neonate: what are they? Pediatrics 105:112–114CrossRefPubMed
2.
Zurück zum Zitat de Jaegere A, van Veenendaal MB, Michiels A, van Kaam AH (2006) Lung recruitment using oxygenation during open lung high-frequency ventilation in preterm infants. Am J Respir Crit Care Med 174:639–645CrossRefPubMed de Jaegere A, van Veenendaal MB, Michiels A, van Kaam AH (2006) Lung recruitment using oxygenation during open lung high-frequency ventilation in preterm infants. Am J Respir Crit Care Med 174:639–645CrossRefPubMed
3.
Zurück zum Zitat Tingay DG, Mills JF, Morley CJ, Pellicano A, Dargaville PA (2006) The deflation limb of the pressure–volume relationship in infants during high-frequency ventilation. Am J Respir Crit Care Med 173:414–420CrossRefPubMed Tingay DG, Mills JF, Morley CJ, Pellicano A, Dargaville PA (2006) The deflation limb of the pressure–volume relationship in infants during high-frequency ventilation. Am J Respir Crit Care Med 173:414–420CrossRefPubMed
4.
Zurück zum Zitat Choong K, Chatrkaw P, Frndova H, Cox PN (2003) Comparison of loss in lung volume with open versus in-line catheter endotracheal suctioning. Pediatr Crit Care Med 4:69–73CrossRefPubMed Choong K, Chatrkaw P, Frndova H, Cox PN (2003) Comparison of loss in lung volume with open versus in-line catheter endotracheal suctioning. Pediatr Crit Care Med 4:69–73CrossRefPubMed
5.
Zurück zum Zitat Maggiore SM, Lellouche F, Pigeot J, Taille S, Deye N, Durrmeyer X, Richard JC, Mancebo J, Lemaire F, Brochard L (2003) Prevention of endotracheal suctioning-induced alveolar derecruitment in acute lung injury. Am J Respir Crit Care Med 167:1215–1224CrossRefPubMed Maggiore SM, Lellouche F, Pigeot J, Taille S, Deye N, Durrmeyer X, Richard JC, Mancebo J, Lemaire F, Brochard L (2003) Prevention of endotracheal suctioning-induced alveolar derecruitment in acute lung injury. Am J Respir Crit Care Med 167:1215–1224CrossRefPubMed
6.
Zurück zum Zitat Cereda M, Villa F, Colombo E, Greco G, Nacoti M, Pesenti A (2001) Closed system endotracheal suctioning maintains lung volume during volume-controlled mechanical ventilation. Intensive Care Med 27:648–654CrossRefPubMed Cereda M, Villa F, Colombo E, Greco G, Nacoti M, Pesenti A (2001) Closed system endotracheal suctioning maintains lung volume during volume-controlled mechanical ventilation. Intensive Care Med 27:648–654CrossRefPubMed
7.
Zurück zum Zitat Hoellering AB, Copnell B, Dargaville PA, Mills JF, Morley CJ, Tingay DG (2008) Lung volume and cardiorespiratory changes during open and closed endotracheal suction in ventilated newborn infants. Arch Dis Child Fetal Neonatal Ed 93:F436–F441CrossRefPubMed Hoellering AB, Copnell B, Dargaville PA, Mills JF, Morley CJ, Tingay DG (2008) Lung volume and cardiorespiratory changes during open and closed endotracheal suction in ventilated newborn infants. Arch Dis Child Fetal Neonatal Ed 93:F436–F441CrossRefPubMed
8.
Zurück zum Zitat Tingay DG, Copnell B, Mills JF, Morley CJ, Dargaville PA (2007) Effects of open endotracheal suction on lung volume in infants receiving HFOV. Intensive Care Med 33:689–693CrossRefPubMed Tingay DG, Copnell B, Mills JF, Morley CJ, Dargaville PA (2007) Effects of open endotracheal suction on lung volume in infants receiving HFOV. Intensive Care Med 33:689–693CrossRefPubMed
9.
Zurück zum Zitat Victorino JA, Borges JB, Okamoto VN, Matos GF, Tucci MR, Caramez MP, Tanaka H, Sipmann FS, Santos DC, Barbas CS, Carvalho CR, Amato MB (2004) Imbalances in regional lung ventilation: a validation study on electrical impedance tomography. Am J Respir Crit Care Med 169:791–800CrossRefPubMed Victorino JA, Borges JB, Okamoto VN, Matos GF, Tucci MR, Caramez MP, Tanaka H, Sipmann FS, Santos DC, Barbas CS, Carvalho CR, Amato MB (2004) Imbalances in regional lung ventilation: a validation study on electrical impedance tomography. Am J Respir Crit Care Med 169:791–800CrossRefPubMed
10.
Zurück zum Zitat Frerichs I, Schiffmann H, Hahn G, Hellige G (2001) Non-invasive radiation-free monitoring of regional lung ventilation in critically ill infants. Intensive Care Med 27:1385–1394CrossRefPubMed Frerichs I, Schiffmann H, Hahn G, Hellige G (2001) Non-invasive radiation-free monitoring of regional lung ventilation in critically ill infants. Intensive Care Med 27:1385–1394CrossRefPubMed
11.
Zurück zum Zitat Frerichs I, Hinz J, Herrmann P, Weisser G, Hahn G, Dudykevych T, Quintel M, Hellige G (2002) Detection of local lung air content by electrical impedance tomography compared with electron beam CT. J Appl Physiol 93:660–666PubMed Frerichs I, Hinz J, Herrmann P, Weisser G, Hahn G, Dudykevych T, Quintel M, Hellige G (2002) Detection of local lung air content by electrical impedance tomography compared with electron beam CT. J Appl Physiol 93:660–666PubMed
12.
Zurück zum Zitat Meier T, Luepschen H, Karsten J, Leibecke T, Grossherr M, Gehring H, Leonhardt S (2008) Assessment of regional lung recruitment and derecruitment during a PEEP trial based on electrical impedance tomography. Intensive Care Med 34:543–550CrossRefPubMed Meier T, Luepschen H, Karsten J, Leibecke T, Grossherr M, Gehring H, Leonhardt S (2008) Assessment of regional lung recruitment and derecruitment during a PEEP trial based on electrical impedance tomography. Intensive Care Med 34:543–550CrossRefPubMed
13.
Zurück zum Zitat Wolf GK, Grychtol B, Frerichs I, van Genderingen HR, Zurakowski D, Thompson JE, Arnold JH (2007) Regional lung volume changes in children with acute respiratory distress syndrome during a derecruitment maneuver. Crit Care Med 35:1972–1978CrossRefPubMed Wolf GK, Grychtol B, Frerichs I, van Genderingen HR, Zurakowski D, Thompson JE, Arnold JH (2007) Regional lung volume changes in children with acute respiratory distress syndrome during a derecruitment maneuver. Crit Care Med 35:1972–1978CrossRefPubMed
14.
Zurück zum Zitat Kiraly NJ, Tingay DG, Mills JF, Morley CJ, Copnell B (2008) Negative tracheal pressure during neonatal endotracheal suction. Pediatr Res 64:29–33CrossRefPubMed Kiraly NJ, Tingay DG, Mills JF, Morley CJ, Copnell B (2008) Negative tracheal pressure during neonatal endotracheal suction. Pediatr Res 64:29–33CrossRefPubMed
Metadaten
Titel
Effect of closed endotracheal suction in high-frequency ventilated premature infants measured with electrical impedance tomography
verfasst von
Mariëtte B. van Veenendaal
Martijn Miedema
Frans H. C. de Jongh
Johanna H. van der Lee
Inez Frerichs
Anton H. van Kaam
Publikationsdatum
01.12.2009
Verlag
Springer-Verlag