Abstract

The ABG I uncemented hip prosthesis has demonstrated an unacceptably high failure rate and poor functional outcome in our patients. A prospective review was performed of 38 implanted hips with a mean follow-up of 11 years. Sixteen hips have been revised or are candidates for revision hip surgery, due principally to polyethylene wear and pelvic osteolysis. Contributing factors were likely to be the poor wear characteristics of the polyethylene liners—which were gamma irradiated in air—and the thinness of these liners. Of further concern were the poor pain and function scores in the remaining hips [mean Harris Hip Scores (HHS) of 70 (SD 22) at 11-year follow-up]. The influence of the surgical learning curve on clinical results may have been an important factor. Undersized femoral components were associated with poor pain and functional scores in five patients with HHS < 60. All ABG I hip implants must remain under careful, long-term review.
Literatur
1.
Zurück zum Zitat Blacha J (2004) High osteolysis and revision rate with the hydroxy-apatite-coated ABG hip prostheses. Acta Orthop Scand 75(3):276–282PubMedCrossRef Blacha J (2004) High osteolysis and revision rate with the hydroxy-apatite-coated ABG hip prostheses. Acta Orthop Scand 75(3):276–282PubMedCrossRef
2.
Zurück zum Zitat Duffy P, Sher JL, Partington PF (2004) Premature wear and osteolysis in an HA-coated uncemented total hip arthroplasty. J Bone Joint Surg Br 86(1):34–38PubMed Duffy P, Sher JL, Partington PF (2004) Premature wear and osteolysis in an HA-coated uncemented total hip arthroplasty. J Bone Joint Surg Br 86(1):34–38PubMed
3.
Zurück zum Zitat Devane PA, Bourne RB, Rorabeck CH, MacDonald S, Robinson EJ (1995) Measurement of polyethylene wear in metal-backed acetabular cups II: clinical application. Clin Orthop 319:317–326PubMed Devane PA, Bourne RB, Rorabeck CH, MacDonald S, Robinson EJ (1995) Measurement of polyethylene wear in metal-backed acetabular cups II: clinical application. Clin Orthop 319:317–326PubMed
5.
Zurück zum Zitat Giannikas KA, Din R, Sadiq S, Dunningham TH (2002) Medium-term results of the ABG total hip arthroplasty in young patients. J Arthroplasty 17:184–188CrossRefPubMed Giannikas KA, Din R, Sadiq S, Dunningham TH (2002) Medium-term results of the ABG total hip arthroplasty in young patients. J Arthroplasty 17:184–188CrossRefPubMed
6.
Zurück zum Zitat Huk O, Bansal M, Betts F (1994) Polyethylene and metal debris generated by non articulating surfaces of modular acetabular components. J Bone Joint Surg Br 76-B:568–574 Huk O, Bansal M, Betts F (1994) Polyethylene and metal debris generated by non articulating surfaces of modular acetabular components. J Bone Joint Surg Br 76-B:568–574
7.
Zurück zum Zitat Harris H (1969) Traumatic arthritis of the hip after dislocation and acetabular fracture: treatment by mould arthroplasty: an end result study using a new method of result evaluation. J Bone Joint Surg Am 51:737–755PubMed Harris H (1969) Traumatic arthritis of the hip after dislocation and acetabular fracture: treatment by mould arthroplasty: an end result study using a new method of result evaluation. J Bone Joint Surg Am 51:737–755PubMed
8.
Zurück zum Zitat Katz JN, Losina E, Barrett J et al (2001) Association between hospital and surgeon procedure volume and outcomes of total hip replacement in the United States Medicare population. J Bone Joint Surg Am 83:1622–1629CrossRefPubMed Katz JN, Losina E, Barrett J et al (2001) Association between hospital and surgeon procedure volume and outcomes of total hip replacement in the United States Medicare population. J Bone Joint Surg Am 83:1622–1629CrossRefPubMed
11.
Zurück zum Zitat Sutula LC, Collier JP, Saum KA et al (1995) Impact of gamma sterilization on clinical performance of polyethylene in the hip. Clin Orthop 319:28–40PubMed Sutula LC, Collier JP, Saum KA et al (1995) Impact of gamma sterilization on clinical performance of polyethylene in the hip. Clin Orthop 319:28–40PubMed
12.
Zurück zum Zitat Shih CH, Lee PC, Chen JH et al (1997) Measurement of polyethylene wear in cementless total hip arthroplasty. J Bone Joint Surg Br 79B:361–365CrossRef Shih CH, Lee PC, Chen JH et al (1997) Measurement of polyethylene wear in cementless total hip arthroplasty. J Bone Joint Surg Br 79B:361–365CrossRef
Metadaten
Titel
Eleven-year results of the ABG I hip replacement
verfasst von
W. Y. Kim
B. N. Muddu
Publikationsdatum
01.06.2006
Verlag
Springer-Verlag
Erschienen in
International Orthopaedics / Ausgabe 3/2006
Print ISSN: 0341-2695
Elektronische ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-006-0075-1

Weitere Artikel der Ausgabe 3/2006

International Orthopaedics 3/2006 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

TEP mit Roboterhilfe führt nicht zu größerer Zufriedenheit

15.05.2024 Knie-TEP Nachrichten

Der Einsatz von Operationsrobotern für den Einbau von Totalendoprothesen des Kniegelenks hat die Präzision der Eingriffe erhöht. Für die postoperative Zufriedenheit der Patienten scheint das aber unerheblich zu sein, wie eine Studie zeigt.

Lever-Sign-Test hilft beim Verdacht auf Kreuzbandriss

15.05.2024 Vordere Kreuzbandruptur Nachrichten

Mit dem Hebelzeichen-Test lässt sich offenbar recht zuverlässig feststellen, ob ein vorderes Kreuzband gerissen ist. In einer Metaanalyse war die Vorhersagekraft vor allem bei positivem Testergebnis hoch.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.