Evaluation of the usage-induced degradation of Genius and Reciproc nickel–titanium reciprocating instruments

verfasst von: Luigi Generali, Alessia Borghi, Luca Lusvarghi, Giovanni Bolelli, Paolo Veronesi, Andrea Vecchi, Ugo Consolo, Carlo Becce, Carlo Bertoldi, Paolo Sassatelli

Erschienen in: Odontology | Ausgabe 4/2019

Einloggen, um Zugang zu erhalten

Abstract

The aim of this study was to characterize the main features and the usage-induced degradation of the Genius file after four severely curved root canal instrumentations and to compare their properties to the Reciproc files. Brand new and ex vivo used files were analysed by scanning electron microscopy (SEM) with energy-dispersive X-ray spectrometry (EDS), differential scanning calorimetry (DSC), X-ray diffraction (XRD), optical metallography, and nano-indentation to disclose their morphological, chemical, mechanical, thermal, and phase composition features. Nano-indentation data were statistically analysed using the Student’s t test for normal distribution or the Kolmogorov–Smirnov test for not-normal distributions. SEM analysis showed the presence of micro-cracks near the tip on both files after ex vivo usage test. EDS analysis confirmed that both files are manufactured from an almost equiatomic NiTi alloy. DSC analysis revealed that the transition temperature of the Genius is below 20 °C, while that of the Reciproc is above 20 °C. XRD analysis of Genius files identified cubic B2 austenite with minor peaks of residual monoclinic B19 martensite, while the contemporaneous presence of martensite, austenite and hexagonal R-phase was observed in the Reciproc files. Significant differences in nanohardness and modulus of elasticity (P < .05) were observed in both Genius and Reciproc files before and after use. The collected results showed that both instruments can be safely used as single-use files.
Literatur
1.
Zurück zum Zitat Shen Y, Zhou HM, Zheng YF, Peng B, Haapasalo M. Current challenges and concepts of the thermomechanical treatment of nickel–titanium instruments. J Endod. 2013;39:163–72.CrossRefPubMed Shen Y, Zhou HM, Zheng YF, Peng B, Haapasalo M. Current challenges and concepts of the thermomechanical treatment of nickel–titanium instruments. J Endod. 2013;39:163–72.CrossRefPubMed
2.
Zurück zum Zitat Walia H, Brantley WA, Gerstein H. An initial Investigation of the bending and torsional properties of nitinol root canal files. J Endod. 1988;14:346–51.CrossRefPubMed Walia H, Brantley WA, Gerstein H. An initial Investigation of the bending and torsional properties of nitinol root canal files. J Endod. 1988;14:346–51.CrossRefPubMed
3.
Zurück zum Zitat Pedullà E, Benites A, La Rosa GM, Plotino G, Grande NM, Rapisarda E, Generali L. Cyclic fatigue resistance of heat-treated nikel–titanium instruments after immersion in sodium hypochlorite and/or sterilization. J Endod. 2018;44:648–53.CrossRefPubMed Pedullà E, Benites A, La Rosa GM, Plotino G, Grande NM, Rapisarda E, Generali L. Cyclic fatigue resistance of heat-treated nikel–titanium instruments after immersion in sodium hypochlorite and/or sterilization. J Endod. 2018;44:648–53.CrossRefPubMed
4.
Zurück zum Zitat Schäfer E. Root canal instruments for manual use: a review. Endod Dental Traumatol. 1997;13:51–64.CrossRef Schäfer E. Root canal instruments for manual use: a review. Endod Dental Traumatol. 1997;13:51–64.CrossRef
5.
Zurück zum Zitat Schäfer E, Erler M, Dammaschke T. Comparative study on the shaping ability and cleaning efficiency of rotary Mtwo instruments. Part 1. Shaping ability in simulated curved canals. Int Endod J. 2006;39:196–202.CrossRefPubMed Schäfer E, Erler M, Dammaschke T. Comparative study on the shaping ability and cleaning efficiency of rotary Mtwo instruments. Part 1. Shaping ability in simulated curved canals. Int Endod J. 2006;39:196–202.CrossRefPubMed
6.
Zurück zum Zitat Setzer FC, Kwon TK, Karabucak B. Comparison of apical transportation between two rotary file systems and two hybrid rotary instrumentation sequences. J Endod. 2010;36:1226–9.CrossRefPubMed Setzer FC, Kwon TK, Karabucak B. Comparison of apical transportation between two rotary file systems and two hybrid rotary instrumentation sequences. J Endod. 2010;36:1226–9.CrossRefPubMed
7.
Zurück zum Zitat Shim KS, Oh S, Kum K, Kim YC, Jee KK, Chang SW. Mechanical and metallurgical properties of various nickel–titanium rotary instruments. Biomed Res Int. 2017;2017:4528601.CrossRefPubMedPubMedCentral Shim KS, Oh S, Kum K, Kim YC, Jee KK, Chang SW. Mechanical and metallurgical properties of various nickel–titanium rotary instruments. Biomed Res Int. 2017;2017:4528601.CrossRefPubMedPubMedCentral
8.
Zurück zum Zitat Otsuka K, Ren X. Physical metallurgy of Ti–Ni-based shape memory alloys. Prog Mater Sci. 2005;50:511–678.CrossRef Otsuka K, Ren X. Physical metallurgy of Ti–Ni-based shape memory alloys. Prog Mater Sci. 2005;50:511–678.CrossRef
9.
Zurück zum Zitat Jamleh A, Kobayashi C, Yahata Y, Ebihara A, Suda H. Deflecting load of nickel titanium rotary instruments during cyclic fatigue. Dent Mater J. 2012;31:389–93.CrossRefPubMed Jamleh A, Kobayashi C, Yahata Y, Ebihara A, Suda H. Deflecting load of nickel titanium rotary instruments during cyclic fatigue. Dent Mater J. 2012;31:389–93.CrossRefPubMed
10.
Zurück zum Zitat Cheung GSP, Peng B, Bian Z, Shen Y, Darvell BW. Defects in ProTaper S1 instruments after clinical use: fractographic examination. Int Endod J. 2005;38:802–9.CrossRefPubMed Cheung GSP, Peng B, Bian Z, Shen Y, Darvell BW. Defects in ProTaper S1 instruments after clinical use: fractographic examination. Int Endod J. 2005;38:802–9.CrossRefPubMed
11.
Zurück zum Zitat Kramkowski TR, Bahcall J. An in vitro comparison of torsional stress and cyclic fatigue resistance of ProFile GT and ProFile GT series X rotary nickel–titanium files. J Endod. 2009;35:404–7.CrossRefPubMed Kramkowski TR, Bahcall J. An in vitro comparison of torsional stress and cyclic fatigue resistance of ProFile GT and ProFile GT series X rotary nickel–titanium files. J Endod. 2009;35:404–7.CrossRefPubMed
12.
Zurück zum Zitat Blum JY, Machtou P, Ruddle C, Micallef JP. Analysis of mechanical preparations in extracted teeth using ProTaper rotary instruments: value of the safety quotient. J Endod. 2003;29:567–75.CrossRefPubMed Blum JY, Machtou P, Ruddle C, Micallef JP. Analysis of mechanical preparations in extracted teeth using ProTaper rotary instruments: value of the safety quotient. J Endod. 2003;29:567–75.CrossRefPubMed
13.
Zurück zum Zitat Generali L, Righi E, Todesca MV, Consolo U. Canal shaping with WaveOne reciprocating files: influence of operator experience on instrument breakage and canal preparation time. Odontology. 2014;102:217–22.CrossRefPubMed Generali L, Righi E, Todesca MV, Consolo U. Canal shaping with WaveOne reciprocating files: influence of operator experience on instrument breakage and canal preparation time. Odontology. 2014;102:217–22.CrossRefPubMed
14.
Zurück zum Zitat Martín B, Zelada G, Varela P, Bahillo JG, Magán F, Ahn S, Rodríguez C. Factors influencing the fracture of nickel–titanium rotary instruments. Int Endod J. 2003;36:262–6.CrossRefPubMed Martín B, Zelada G, Varela P, Bahillo JG, Magán F, Ahn S, Rodríguez C. Factors influencing the fracture of nickel–titanium rotary instruments. Int Endod J. 2003;36:262–6.CrossRefPubMed
15.
Zurück zum Zitat Zhou HM, Shen Y, Zheng W, Li L, Zheng YF, Haapasalo M. Mechanical properties of controlled memory and superelastic nickel–titanium wires used in the manufacture of rotary endodontic instruments. J Endod. 2012;38:1535–40.CrossRefPubMed Zhou HM, Shen Y, Zheng W, Li L, Zheng YF, Haapasalo M. Mechanical properties of controlled memory and superelastic nickel–titanium wires used in the manufacture of rotary endodontic instruments. J Endod. 2012;38:1535–40.CrossRefPubMed
16.
Zurück zum Zitat Panitvisai P, Parunnit P, Sathorn C, Messer HH. Impact of a retained instrument on treatment outcome: a systematic review and meta-analysis. J Endod. 2010;36:775–80.CrossRefPubMed Panitvisai P, Parunnit P, Sathorn C, Messer HH. Impact of a retained instrument on treatment outcome: a systematic review and meta-analysis. J Endod. 2010;36:775–80.CrossRefPubMed
17.
Zurück zum Zitat McGuigan MB, Louca C, Duncan HF. The impact of fractured endodontic instruments on treatment outcome. Br Dent J. 2013;214:285–9.CrossRefPubMed McGuigan MB, Louca C, Duncan HF. The impact of fractured endodontic instruments on treatment outcome. Br Dent J. 2013;214:285–9.CrossRefPubMed
18.
Zurück zum Zitat Kiefner P, Ban M, De-Deus G. Is the reciprocating movement per se able to improve the cyclic fatigue resistance of instruments? Int Endod J. 2014;47:430–6.CrossRefPubMed Kiefner P, Ban M, De-Deus G. Is the reciprocating movement per se able to improve the cyclic fatigue resistance of instruments? Int Endod J. 2014;47:430–6.CrossRefPubMed
19.
Zurück zum Zitat De-Deus G, Moreira EJ, Lopes HP, Elias CN. Extended cyclic fatigue life of F2 ProTaper instruments used in reciprocating movement. Int Endod J. 2010;43:1063–8.CrossRefPubMed De-Deus G, Moreira EJ, Lopes HP, Elias CN. Extended cyclic fatigue life of F2 ProTaper instruments used in reciprocating movement. Int Endod J. 2010;43:1063–8.CrossRefPubMed
20.
Zurück zum Zitat van der Vyver PJ, Jonker C. Reciprocating instruments in endodontics: a review of the literature. SADJ. 2014;64:404–9. van der Vyver PJ, Jonker C. Reciprocating instruments in endodontics: a review of the literature. SADJ. 2014;64:404–9.
21.
Zurück zum Zitat De-Deus G, Silva EJ, Vieira VT, Belladonna FG, Elias CN, Plotino G, Grande NM. Blue thermomechanical treatment optimizes fatigue resistance and flexibility of the Reciproc files. J Endod. 2017;43:462–6.CrossRefPubMed De-Deus G, Silva EJ, Vieira VT, Belladonna FG, Elias CN, Plotino G, Grande NM. Blue thermomechanical treatment optimizes fatigue resistance and flexibility of the Reciproc files. J Endod. 2017;43:462–6.CrossRefPubMed
23.
Zurück zum Zitat Iacono F, Pirani C, Generali L, Bolelli G, Sassatelli P, Lusvarghi L, Gandolfi MG, Giorgini L, Prati C. Structural analysis of HyFlex EDM instruments. Int Endod J. 2017;50:303–13.CrossRefPubMed Iacono F, Pirani C, Generali L, Bolelli G, Sassatelli P, Lusvarghi L, Gandolfi MG, Giorgini L, Prati C. Structural analysis of HyFlex EDM instruments. Int Endod J. 2017;50:303–13.CrossRefPubMed
24.
Zurück zum Zitat Özyürek T, Gündogår M, Yilmaz K, Uslu G. Bending resistance and cyclic fatigue life of Reciproc Blue, WaveOne Gold, and Genius files in a double (S-shaped) curved canal. J Dent Res Dent Clin Dent Prospects. 2017;11:241–6.PubMedPubMedCentral Özyürek T, Gündogår M, Yilmaz K, Uslu G. Bending resistance and cyclic fatigue life of Reciproc Blue, WaveOne Gold, and Genius files in a double (S-shaped) curved canal. J Dent Res Dent Clin Dent Prospects. 2017;11:241–6.PubMedPubMedCentral
26.
Zurück zum Zitat Pirani C, Iacono F, Generali L, Sassatelli P, Nucci C, Lusvarghi L, Gandolfi MG, Prati C. HyFlex EDM: superficial features, metallurgical analysis and fatigue resistance of innovative electro discharge machined NiTi rotary instruments. Int Endod J. 2016;49:483–93.CrossRefPubMed Pirani C, Iacono F, Generali L, Sassatelli P, Nucci C, Lusvarghi L, Gandolfi MG, Prati C. HyFlex EDM: superficial features, metallurgical analysis and fatigue resistance of innovative electro discharge machined NiTi rotary instruments. Int Endod J. 2016;49:483–93.CrossRefPubMed
27.
Zurück zum Zitat Hou X, Yahata Y, Hayashi Y, Ebihara A, Hanawa T, Suda H. Phase transformation behaviour and bending property of twisted nickel–titanium endodontic instruments. Int Endod J. 2011;44:253–8.CrossRefPubMed Hou X, Yahata Y, Hayashi Y, Ebihara A, Hanawa T, Suda H. Phase transformation behaviour and bending property of twisted nickel–titanium endodontic instruments. Int Endod J. 2011;44:253–8.CrossRefPubMed
28.
Zurück zum Zitat ISO 14577-1. Metallic materials—instrumented indentation tests for hardness and materials parameters—Part 1: test method. 2015. ISO 14577-1. Metallic materials—instrumented indentation tests for hardness and materials parameters—Part 1: test method. 2015.
29.
Zurück zum Zitat Oliver WC, Pharr GM. An improved technique for determining hardness and elastic modulus using load and displacement sensing indentation experiments. J Mater Res. 1992;7:1564–83.CrossRef Oliver WC, Pharr GM. An improved technique for determining hardness and elastic modulus using load and displacement sensing indentation experiments. J Mater Res. 1992;7:1564–83.CrossRef
30.
Zurück zum Zitat Wayman CM, Duerig TW. An introduction to martensite and shape memory. In: Duerig TW, Melton KN, Stöckel D, Wayman CM, editors. Engineering aspects of shape memory alloys. 1st ed. London: Buttheworth-Heinemann; 1990. pp. 3–9.CrossRef Wayman CM, Duerig TW. An introduction to martensite and shape memory. In: Duerig TW, Melton KN, Stöckel D, Wayman CM, editors. Engineering aspects of shape memory alloys. 1st ed. London: Buttheworth-Heinemann; 1990. pp. 3–9.CrossRef
31.
Zurück zum Zitat Arslan H, Doğanay E, Alsancak M, Çapar ID, Karataş E, Gündüz HA. Comparison of apically extruded debris after root canal instrumentation using Reciproc® instruments with various kinematics. Int Endod J. 2016;49:307–10.CrossRefPubMed Arslan H, Doğanay E, Alsancak M, Çapar ID, Karataş E, Gündüz HA. Comparison of apically extruded debris after root canal instrumentation using Reciproc® instruments with various kinematics. Int Endod J. 2016;49:307–10.CrossRefPubMed
32.
Zurück zum Zitat Bürklein S, Hinschitza K, Dammaschke T, Schäfer E. Shaping ability and cleaning effectiveness of two single-file systems in severely curved root canals of extracted teeth: Reciproc and WaveOne versus Mtwo and ProTaper. Int Endod J. 2012;45:449–61.CrossRefPubMed Bürklein S, Hinschitza K, Dammaschke T, Schäfer E. Shaping ability and cleaning effectiveness of two single-file systems in severely curved root canals of extracted teeth: Reciproc and WaveOne versus Mtwo and ProTaper. Int Endod J. 2012;45:449–61.CrossRefPubMed
33.
Zurück zum Zitat Pereira ES, Peixoto IF, Viana AC, Oliveira II, Gonzales BM, Buono VT, Bahia MG. Physical and mechanical properties of a thermomechanically treated NiTi wire used in the manufacture of rotary endodontic instruments. Int Endod J. 2012;45:469–74.CrossRefPubMed Pereira ES, Peixoto IF, Viana AC, Oliveira II, Gonzales BM, Buono VT, Bahia MG. Physical and mechanical properties of a thermomechanically treated NiTi wire used in the manufacture of rotary endodontic instruments. Int Endod J. 2012;45:469–74.CrossRefPubMed
34.
Zurück zum Zitat Kosti E, Zinelis S, Molyvdas I, Lambrianidis T. Effect of root canal curvature on the failure incidence of ProFile royary Ni–Ti endodontic instruments. Int Endod J. 2011;44:917–25.CrossRefPubMed Kosti E, Zinelis S, Molyvdas I, Lambrianidis T. Effect of root canal curvature on the failure incidence of ProFile royary Ni–Ti endodontic instruments. Int Endod J. 2011;44:917–25.CrossRefPubMed
35.
Zurück zum Zitat Plotino G, Grande NM, Porciani PF. Deformation and fracture incidence of Reciproc instruments: a clinical evaluation. Int Endod J. 2015;48:199–205.CrossRefPubMed Plotino G, Grande NM, Porciani PF. Deformation and fracture incidence of Reciproc instruments: a clinical evaluation. Int Endod J. 2015;48:199–205.CrossRefPubMed
36.
Zurück zum Zitat Gu Y, Kum KY, Perinpanayagam H, Kim C, Kum DJ, Lim SM, Chang SW, Baek SH, Zhu Q, Yoo YJ. Various heat-treated nickel–titanium rotary instruments evaluated in S-shaped simulated resin canals. J Dent Sci. 2017;12:14–20.CrossRefPubMed Gu Y, Kum KY, Perinpanayagam H, Kim C, Kum DJ, Lim SM, Chang SW, Baek SH, Zhu Q, Yoo YJ. Various heat-treated nickel–titanium rotary instruments evaluated in S-shaped simulated resin canals. J Dent Sci. 2017;12:14–20.CrossRefPubMed
37.
Zurück zum Zitat Iacono F, Pirani C, Generali L, Sassatelli P, Nucci C, Gandolfi MG, Prati C. Wear analysis and cyclic fatigue resistance of electro discharge machined NiTi rotary instruments. G It Endo. 2016;30:64–8.CrossRef Iacono F, Pirani C, Generali L, Sassatelli P, Nucci C, Gandolfi MG, Prati C. Wear analysis and cyclic fatigue resistance of electro discharge machined NiTi rotary instruments. G It Endo. 2016;30:64–8.CrossRef
38.
Zurück zum Zitat Sangid MD. The physics of fatigue crack initiation. Int J Fatigue. 2013;57:58–72.CrossRef Sangid MD. The physics of fatigue crack initiation. Int J Fatigue. 2013;57:58–72.CrossRef
39.
Zurück zum Zitat Brantley WA, Svec TA, Iijima M, Powers JM, Grentzer TH. Differential scanning calorimetric studies of nickel titanium rotary endodontic instruments. J Endod. 2002;28:567–72.CrossRefPubMed Brantley WA, Svec TA, Iijima M, Powers JM, Grentzer TH. Differential scanning calorimetric studies of nickel titanium rotary endodontic instruments. J Endod. 2002;28:567–72.CrossRefPubMed
40.
Zurück zum Zitat Shagmir H, Nili-Ahmadabadi M, Huang Y, Langdon TG. Evolution of microstructure and hardness in NiTi shape memory alloys processed by high-pressure torsion. J Mater Sci. 2014;49:2998–3009.CrossRef Shagmir H, Nili-Ahmadabadi M, Huang Y, Langdon TG. Evolution of microstructure and hardness in NiTi shape memory alloys processed by high-pressure torsion. J Mater Sci. 2014;49:2998–3009.CrossRef
41.
Zurück zum Zitat Pirani C, Paolucci A, Ruggeri O, Bossù M, Polimeni A, Gatto MR, Gandolfi MG, Prati C. Wear and metallographic analysis of WaveOne and reciproc NiTi instruments before and after three uses in root canals. Scanning. 2014;36:517–25.CrossRefPubMed Pirani C, Paolucci A, Ruggeri O, Bossù M, Polimeni A, Gatto MR, Gandolfi MG, Prati C. Wear and metallographic analysis of WaveOne and reciproc NiTi instruments before and after three uses in root canals. Scanning. 2014;36:517–25.CrossRefPubMed
Metadaten
Titel
Evaluation of the usage-induced degradation of Genius and Reciproc nickel–titanium reciprocating instruments
verfasst von
Luigi Generali
Alessia Borghi
Luca Lusvarghi
Giovanni Bolelli
Paolo Veronesi
Andrea Vecchi
Ugo Consolo
Carlo Becce
Carlo Bertoldi
Paolo Sassatelli
Publikationsdatum
28.03.2019
Verlag
Springer Japan
Erschienen in
Odontology / Ausgabe 4/2019
Print ISSN: 1618-1247
Elektronische ISSN: 1618-1255
DOI
https://doi.org/10.1007/s10266-019-00423-9

Weitere Artikel der Ausgabe 4/2019

Odontology 4/2019 Zur Ausgabe

Parodontalbehandlung verbessert Prognose bei Katheterablation

19.04.2024 Vorhofflimmern Nachrichten

Werden Personen mit Vorhofflimmern in der Blanking-Periode nach einer Katheterablation gegen eine bestehende Parodontitis behandelt, verbessert dies die Erfolgsaussichten. Dafür sprechen die Resultate einer prospektiven Untersuchung.

Invasive Zahnbehandlung: Wann eine Antibiotikaprophylaxe vor infektiöser Endokarditis schützt

11.04.2024 Endokarditis Nachrichten

Bei welchen Personen eine Antibiotikaprophylaxe zur Prävention einer infektiösen Endokarditis nach invasiven zahnärztlichen Eingriffen sinnvoll ist, wird diskutiert. Neue Daten stehen im Einklang mit den europäischen Leitlinienempfehlungen.

Zell-Organisatoren unter Druck: Mechanismen des embryonalen Zahnwachstums aufgedeckt

08.04.2024 Zahnmedizin Nachrichten

Der Aufbau von Geweben und Organen während der Embryonalentwicklung wird von den Zellen bemerkenswert choreografiert. Für diesen Prozess braucht es spezielle sogenannte „Organisatoren“. In einer aktuellen Veröffentlichung im Fachjournal Nature Cell Biology berichten Forschende durch welchen Vorgang diese Organisatoren im Gewebe entstehen und wie sie dann die Bildung von Zähnen orchestrieren.

Die Oralprophylaxe & Kinderzahnheilkunde umbenannt

11.03.2024 Kinderzahnmedizin Nachrichten

Infolge der Umbenennung der Deutschen Gesellschaft für Kinderzahnheilkunde in Deutsche Gesellschaft für Kinderzahnmedizin (DGKiZ) wird deren Mitgliederzeitschrift Oralprophylaxe & Kinderzahnheilkunde in Oralprophylaxe & Kinderzahnmedizin umbenannt. Aus diesem Grunde trägt die erste Ausgabe in 2024 erstmalig den neuen Titel.

Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Zahnmedizin und bleiben Sie gut informiert – ganz bequem per eMail.