Evaluation of vessel injury after simulated catheter use in an endothelialized silicone model of the intracranial arteries

verfasst von: Alyssa McCulloch, Brianna Yang, Sergey Frenklakh, Pratika Sah, Kristen O’Halloran Cardinal

Erschienen in: Neuroradiology | Ausgabe 10/2023

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Neurothrombectomy catheters can disrupt or injure the vessel wall. This potential injury is often studied in animal or cadaver models, but prior work suggests that endothelialized silicone models may be an option for early in vitro assessment. The purpose of this work was to create a complex, clinically-relevant endothelialized neurovascular silicone model, and to determine the utility of the model for evaluating vessel injury due to catheter simulated use.

Methods

Models of the ICA and MCA were fabricated out of silicone, sterilized, coated with fibronectin, placed in bioreactors, and endothelialized with HUVECs. These silicone vessels were maintained under flow for 3 and 7 days, and cellular linings were assessed. Subsequently, 24 silicone vessels were created and treated with neurovascular catheters. Vessels were accessed with a guidewire, microcatheter, and/or aspiration catheter, either once (1-pass) or three times (3-pass). Vessels were then fixed, and injury was evaluated through quantitative image analysis and a visual scoring system.

Results

Complex silicone models were successfully endothelialized and maintained with consistent cell linings. The transparent silicone permitted catheter simulated use without fluoroscopy, and injury to the vessel wall was observed and successfully imaged and characterized. Vessels subjected to 3-passes exhibited more injury than 1-pass, and injury increased with the number and size of devices. These results illustrated expected trends and support use of these models for early assessment of vessel injury.

Conclusion

Complex silicone neurovascular models can be endothelialized and used in vitro to assess and compare injury due to the use of neurovascular catheters.
Literatur
1.
Zurück zum Zitat Lv X (2021) The history and development of endovascular neurosurgery. In: Frontiers in Clinical Neurosurgery. IntechOpenCrossRef Lv X (2021) The history and development of endovascular neurosurgery. In: Frontiers in Clinical Neurosurgery. IntechOpenCrossRef
3.
Zurück zum Zitat Nogueira RG, Lutsep HL, Gupta R et al (2012) Trevo versus Merci retrievers for thrombectomy revascularisation of large vessel occlusions in acute ischaemic stroke (TREVO 2): a randomised trial. Lancet 380:1231–1240CrossRefPubMedPubMedCentral Nogueira RG, Lutsep HL, Gupta R et al (2012) Trevo versus Merci retrievers for thrombectomy revascularisation of large vessel occlusions in acute ischaemic stroke (TREVO 2): a randomised trial. Lancet 380:1231–1240CrossRefPubMedPubMedCentral
5.
Zurück zum Zitat Hidaka M, Yamaguchi S, Koyanagi Y, Arakawa S (2019) Reocclusion of the treated vessel due to endothelial injury after mechanical thrombectomy in a patient with acute ischaemic stroke. BMJ Case Rep 12:e228937CrossRefPubMedPubMedCentral Hidaka M, Yamaguchi S, Koyanagi Y, Arakawa S (2019) Reocclusion of the treated vessel due to endothelial injury after mechanical thrombectomy in a patient with acute ischaemic stroke. BMJ Case Rep 12:e228937CrossRefPubMedPubMedCentral
6.
Zurück zum Zitat Liu Y, Abbasi M, Arturo Larco JL et al (2021) Preclinical testing platforms for mechanical thrombectomy in stroke: a review on phantoms, in-vivo animal, and cadaveric models. J Neurointerv Surg 13:816–822CrossRefPubMedPubMedCentral Liu Y, Abbasi M, Arturo Larco JL et al (2021) Preclinical testing platforms for mechanical thrombectomy in stroke: a review on phantoms, in-vivo animal, and cadaveric models. J Neurointerv Surg 13:816–822CrossRefPubMedPubMedCentral
7.
Zurück zum Zitat Chueh JY, Wakhloo AK, Gounis MJ (2009) Neurovascular Modeling: Small-Batch Manufacturing of Silicone Vascular Replicas. Am J Neuroradiol 30:1159–1164CrossRefPubMedPubMedCentral Chueh JY, Wakhloo AK, Gounis MJ (2009) Neurovascular Modeling: Small-Batch Manufacturing of Silicone Vascular Replicas. Am J Neuroradiol 30:1159–1164CrossRefPubMedPubMedCentral
8.
Zurück zum Zitat Sommer KN, Bhurwani MMS, Tutino V et al (2021) Use of patient specific 3D printed neurovascular phantoms to simulate mechanical thrombectomy. 3D Print Med 7:32CrossRefPubMedPubMedCentral Sommer KN, Bhurwani MMS, Tutino V et al (2021) Use of patient specific 3D printed neurovascular phantoms to simulate mechanical thrombectomy. 3D Print Med 7:32CrossRefPubMedPubMedCentral
9.
Zurück zum Zitat Reddy AS, Liu Y, Cockrum J et al (2021) Construction of a comprehensive endovascular test bed for research and device development in mechanical thrombectomy in stroke. J Neurosurg 134:1190–1197CrossRef Reddy AS, Liu Y, Cockrum J et al (2021) Construction of a comprehensive endovascular test bed for research and device development in mechanical thrombectomy in stroke. J Neurosurg 134:1190–1197CrossRef
10.
Zurück zum Zitat Teng D, Pannell JS, Rennert RC et al (2015) Endothelial Trauma From Mechanical Thrombectomy in Acute Stroke. Stroke 46:1099–1106CrossRefPubMed Teng D, Pannell JS, Rennert RC et al (2015) Endothelial Trauma From Mechanical Thrombectomy in Acute Stroke. Stroke 46:1099–1106CrossRefPubMed
11.
Zurück zum Zitat Carniato S, Mehra M, King RM et al (2013) Porcine Brachial Artery Tortuosity for In Vivo Evaluation of Neuroendovascular Devices. Am J Neuroradiol 34:E36–E38CrossRefPubMedPubMedCentral Carniato S, Mehra M, King RM et al (2013) Porcine Brachial Artery Tortuosity for In Vivo Evaluation of Neuroendovascular Devices. Am J Neuroradiol 34:E36–E38CrossRefPubMedPubMedCentral
12.
Zurück zum Zitat Georganos SA, Guilbert F, Salazkin I et al (2004) Surgical construction of an in vivo carotid siphon model to test neurovascular devices. Neurosurgery 54:1239–1243CrossRefPubMed Georganos SA, Guilbert F, Salazkin I et al (2004) Surgical construction of an in vivo carotid siphon model to test neurovascular devices. Neurosurgery 54:1239–1243CrossRefPubMed
13.
Zurück zum Zitat Gao B-L, Wang Y-L, Zhang X-J et al (2017) Construction of an in vivo carotid siphon model for testing endovascular devices for neuro-interventions. Interv Neuroradiol 23:325–329CrossRefPubMedPubMedCentral Gao B-L, Wang Y-L, Zhang X-J et al (2017) Construction of an in vivo carotid siphon model for testing endovascular devices for neuro-interventions. Interv Neuroradiol 23:325–329CrossRefPubMedPubMedCentral
14.
Zurück zum Zitat Gory B, Bresson D, Rouchaud A et al (2013) A novel swine model to evaluate arterial vessel injury after mechanical endovascular thrombectomy. Interv Neuroradiol 19:147–152CrossRefPubMedPubMedCentral Gory B, Bresson D, Rouchaud A et al (2013) A novel swine model to evaluate arterial vessel injury after mechanical endovascular thrombectomy. Interv Neuroradiol 19:147–152CrossRefPubMedPubMedCentral
15.
Zurück zum Zitat Peschillo S, Tomasello A, Diana F et al (2018) Comparison of subacute vascular damage caused by adapt versus stent retriever devices after thrombectomy in acute ischemic stroke: histological and ultrastructural study in an animal model. Interv Neurol 7:501–512CrossRefPubMedPubMedCentral Peschillo S, Tomasello A, Diana F et al (2018) Comparison of subacute vascular damage caused by adapt versus stent retriever devices after thrombectomy in acute ischemic stroke: histological and ultrastructural study in an animal model. Interv Neurol 7:501–512CrossRefPubMedPubMedCentral
16.
Zurück zum Zitat Li C-H, Gao B-L, Wang J-W et al (2019) Construction of aneurysmal models on a curved vascular segment of a carotid siphon model for testing endovascular devices. World Neurosurg 123:e581–e587CrossRefPubMed Li C-H, Gao B-L, Wang J-W et al (2019) Construction of aneurysmal models on a curved vascular segment of a carotid siphon model for testing endovascular devices. World Neurosurg 123:e581–e587CrossRefPubMed
17.
Zurück zum Zitat Arai D, Ishii A, Chihara H et al (2016) Histological examination of vascular damage caused by stent retriever thrombectomy devices. J Neurointerv Surg 8:992–995CrossRefPubMed Arai D, Ishii A, Chihara H et al (2016) Histological examination of vascular damage caused by stent retriever thrombectomy devices. J Neurointerv Surg 8:992–995CrossRefPubMed
18.
Zurück zum Zitat Arturo Larco JL, Madhani SI, Liu Y et al (2023) Human “live cadaver” neurovascular model for proximal and distal mechanical thrombectomy in stroke. J Neurointerv Surg 15:465–472CrossRefPubMed Arturo Larco JL, Madhani SI, Liu Y et al (2023) Human “live cadaver” neurovascular model for proximal and distal mechanical thrombectomy in stroke. J Neurointerv Surg 15:465–472CrossRefPubMed
19.
Zurück zum Zitat Villadolid C, Puccini B, Dennis B et al (2020) Custom tissue engineered aneurysm models with varying neck size and height for early stage in vitro testing of flow diverters. J Mater Sci Mater Med 31:34CrossRefPubMedPubMedCentral Villadolid C, Puccini B, Dennis B et al (2020) Custom tissue engineered aneurysm models with varying neck size and height for early stage in vitro testing of flow diverters. J Mater Sci Mater Med 31:34CrossRefPubMedPubMedCentral
20.
Zurück zum Zitat McCulloch A, Turcott A, Graham G et al (2021) Endothelialized silicone aneurysm models for in vitro evaluation of flow diverters. J Neurointerv Surg 13:727–731CrossRefPubMed McCulloch A, Turcott A, Graham G et al (2021) Endothelialized silicone aneurysm models for in vitro evaluation of flow diverters. J Neurointerv Surg 13:727–731CrossRefPubMed
21.
Zurück zum Zitat Sugiu K, Martin J-B, Jean B et al (2003) Artificial cerebral aneurysm model for medical testing, training, and research. Neurol Med Chir 43:69–72 discussion 73CrossRef Sugiu K, Martin J-B, Jean B et al (2003) Artificial cerebral aneurysm model for medical testing, training, and research. Neurol Med Chir 43:69–72 discussion 73CrossRef
22.
Zurück zum Zitat Thompson JW, Elwardany O, McCarthy DJ et al (2019) In vivo cerebral aneurysm models. Neurosurg Focus 47:E20CrossRefPubMed Thompson JW, Elwardany O, McCarthy DJ et al (2019) In vivo cerebral aneurysm models. Neurosurg Focus 47:E20CrossRefPubMed
23.
Zurück zum Zitat Kent KC, Shindo S, Ikemoto T, Whittemore AD (1989) Species variation and the success of endothelial cell seeding. J Vasc Surg 9:271–276CrossRefPubMed Kent KC, Shindo S, Ikemoto T, Whittemore AD (1989) Species variation and the success of endothelial cell seeding. J Vasc Surg 9:271–276CrossRefPubMed
24.
Zurück zum Zitat Sharefkin JB, van Wart HE, Cruess DF et al (1986) Adult human endothelial cell enzymatic harvesting. Estimates of efficiency and comparison of crude and partially purified bacterial collagenase preparations by replicate microwell culture and fibronectin degradation measured by enzyme-linked immunosorbent assay. J Vasc Surg 4:567–577CrossRefPubMed Sharefkin JB, van Wart HE, Cruess DF et al (1986) Adult human endothelial cell enzymatic harvesting. Estimates of efficiency and comparison of crude and partially purified bacterial collagenase preparations by replicate microwell culture and fibronectin degradation measured by enzyme-linked immunosorbent assay. J Vasc Surg 4:567–577CrossRefPubMed
25.
Zurück zum Zitat Fitzgerald S, Ryan D, Thornton J, Nogueira RG (2021) Preclinical evaluation of Millipede 088 intracranial aspiration catheter in cadaver and in vitro thrombectomy models. J Neurointerv Surg 13:447–452CrossRefPubMed Fitzgerald S, Ryan D, Thornton J, Nogueira RG (2021) Preclinical evaluation of Millipede 088 intracranial aspiration catheter in cadaver and in vitro thrombectomy models. J Neurointerv Surg 13:447–452CrossRefPubMed
26.
Zurück zum Zitat Mejia J, Gutierrez J, Patino Hoyos M et al (2020) E-130 Step forward: early experience using OCT technology in neurovascular field. In: Electronic Poster Abstracts. BMJ Publishing Group Ltd., BMA House, Tavistock Square, London, WC1H 9JR, p A99.1-A99 Mejia J, Gutierrez J, Patino Hoyos M et al (2020) E-130 Step forward: early experience using OCT technology in neurovascular field. In: Electronic Poster Abstracts. BMJ Publishing Group Ltd., BMA House, Tavistock Square, London, WC1H 9JR, p A99.1-A99
Metadaten
Titel
Evaluation of vessel injury after simulated catheter use in an endothelialized silicone model of the intracranial arteries
verfasst von
Alyssa McCulloch
Brianna Yang
Sergey Frenklakh
Pratika Sah
Kristen O’Halloran Cardinal
Publikationsdatum
17.07.2023
Verlag
Springer Berlin Heidelberg
Erschienen in
Neuroradiology / Ausgabe 10/2023
Print ISSN: 0028-3940
Elektronische ISSN: 1432-1920
DOI
https://doi.org/10.1007/s00234-023-03197-8

Weitere Artikel der Ausgabe 10/2023

Neuroradiology 10/2023 Zur Ausgabe

Leitlinien kompakt für die Neurologie

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Wartezeit nicht kürzer, aber Arbeit flexibler

Psychotherapie Medizin aktuell

Fünf Jahren nach der Neugestaltung der Psychotherapie-Richtlinie wurden jetzt die Effekte der vorgenommenen Änderungen ausgewertet. Das Hauptziel der Novellierung war eine kürzere Wartezeit auf Therapieplätze. Dieses Ziel wurde nicht erreicht, es gab jedoch positive Auswirkungen auf andere Bereiche.

„Restriktion auf vier Wochen Therapie bei Schlaflosigkeit ist absurd!“

06.05.2024 Insomnie Nachrichten

Chronische Insomnie als eigenständiges Krankheitsbild ernst nehmen und adäquat nach dem aktuellen Forschungsstand behandeln: Das forderte der Schlafmediziner Dr. Dieter Kunz von der Berliner Charité beim Praxis Update.

Stuhltransfusion könnte Fortschreiten von Parkinson-Symptomen bremsen

03.05.2024 Parkinson-Krankheit Nachrichten

Kann eine frühzeitige Stuhltransplantation das Fortschreiten von Parkinson-Symptomen verlangsamen? Die Ergebnisse einer randomisierten Phase-2-Studie scheinen dafür zu sprechen.

Frühe Tranexamsäure-Therapie nützt wenig bei Hirnblutungen

02.05.2024 Hirnblutung Nachrichten

Erhalten Personen mit einer spontanen Hirnblutung innerhalb von zwei Stunden nach Symptombeginn eine Tranexamsäure-Therapie, kann dies weder die Hämatomexpansion eindämmen noch die Mortalität senken.

Update Neurologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.