Skip to main content
Erschienen in: Documenta Ophthalmologica 1/2019

Open Access 23.11.2018 | Original Research Article

Flash visual evoked potentials (FVEP) in various stimulation conditions

verfasst von: Dorota Pojda-Wilczek, Wojciech Maruszczyk, Sebastian Sirek

Erschienen in: Documenta Ophthalmologica | Ausgabe 1/2019

Abstract

Aim

To compare flash visual evoked potentials (FVEP) elicited using a Ganzfeld bowl (G), Mini Ganzfeld (MG) and Flash Goggles (GG) with eyes open and closed.

Patients and method

The study group comprised 17 volunteers with mean age of 30 years; all of them were examined with the Roland Consult electrophysiological diagnostic system. Active electrodes were placed at O1 and O2. With the G and MG stimulators, the flash generated by white-light-emitting diodes (LEDs) presented standard flash of 3 cd s m−2. The GG used red LED flash of 3 cd s m−2. Stimulus frequency of 1.0 Hz, low-pass filter of 1.0 Hz and high-pass filters of 100 Hz (G); 50 Hz (MG); 30 Hz (GG) were used. P2 amplitude and latency were compared by the means of the Wilcoxon matched-pairs signed-rank test.

Results

After right eye stimulation (from O1; n = 17), the mean amplitudes of P2, elicited with the G, MG and GG, were 13, 7 and 10 µV, respectively. The respective latencies were 129, 114 and 110 ms. Hence, the difference between the results obtained with these stimulators was statistically significant (p < 0.05). The mean P2 amplitudes, acquired by the means of the G, MG and GG, were 13 µV, 7 µV and 10 µV for open eyes, and 11 µV, 8 µV and 8 µV for closed eyes. The respective latencies were 129, 114 and 110 ms for eyes open, and 127, 125 and 121 ms for eyes closed. These results of the MG (latency only) and GG (latency and amplitude) stimulation differed significantly (p < 0.05).

Conclusion

The amplitudes and latencies of the FVEP P2 elicited with different stimulators are not suitable for comparison. Closing the eye during the examination had a significant effect on the components of FVEP waveform elicited with the Flash Goggle and on the latency of P2 elicited with the MG.

Introduction

Flash visual evoked potentials (FVEP) have been used in ophthalmology and neurology since 1960s. Nowadays, the flash stimulus may not be considered to be the first choice in obtaining VEP, since pattern visual evoked potentials (PVEP) are much more reproducible and consistent among patients. However, PVEP require stable fixation, good cooperation and correction of refractive errors. Thus, this type of stimulation is sometimes difficult to perform in babies and young children, especially in the presence of a developmental disability or severe vision impairment. In the presence of developmental disability compliance and understanding are limited and therefore standard vision assessment techniques are not possible regardless of age. Therefore, for many patients flash stimulation may be the optimum tool for visual evoked potential assessment.
Apart from the standard Ganzfeld stimulator, other stimulators, such as the Mini Ganzfeld, Flash Baby, Flash Goggles, are used, especially for the examination of infants and children. Also a variety of different devices in use to assess FVEP in different attentional states of the patient are used. Intraoperative monitoring using VEP has been described. Benedičič and Bošnjak [1] used the Flash Goggles, while Luo et al. the light-stimulating device [2] for intraoperative assessment of the integrity of intracranial visual pathways. The conditions of stimulation vary from one laboratory to another one. Chayasirisobhon et al. [3] used LED goggles to assess awake neonates, while Klebermass-Scherehof et al. [4] and Feng et al. [5] examined infants during sleep. Jethani and Jethani used standard handheld stimulation and examination under sedation [6], and Shepherd et al. examined infants in different states of sleep using stroboscopic flash [7].
When interpreting FVEP, the type of device needs to be considered and also the use of anesthesia, the state of patient alertness (asleep, alert) and whether eyes are open or closed, e.g., Tartaglione, Bandini et al. 1995 “Eye Closure affects flash VEP latency in dementia” [8]. This study demonstrated particular latency delays of the P2 (eyes closed 3 ms delayed compared with eyes open) versus other components with the delay being significant in those with dementia versus control. All these factors need to be considered in the interpretation of flash VEP.
There is a lack of studies juxtaposing the results of examinations using various stimulators, and only some studies concentrate on the differences between examinations with open and closed eyes [9]. The aim of this study is to clarify whether results from FVEP stimulators differ and the effect of examination in the waking (open eye) versus closed eye state.

Aim

The aim of the study was to compare the flash visual evoked potentials (FVEP) elicited using the Ganzfeld bowl, Mini Ganzfeld and Flash Goggles with eyes open and closed.

Patients and methods

Seventeen volunteers (34 eyes), including 11 females, aged 23–56 (mean age: 30 years) were examined by the means of the Reti-port (Roland Consult, Germany) electrophysiological equipment. The examination was performed in a dimly illuminated room. The flash visual evoked potentials (FVEP) were examined using the following stimulators: the Ganzfeld bowl, Mini Ganzfeld and Flash Goggles. Each stimulator was used to conduct the following examinations: right eye open, left eye open and then eyes closed starting from the right one.
During the examination using the Ganzfeld bowl and the Mini Ganzfeld, the fellow eye was completely patched by a black, carton obturator with a gauze tampon on the closed eyelid. In the case of the Flash Goggles examination, the eyes were not covered.
The pupils were not dilated. According to the International 10/20 system, two active gold cup electrodes were placed in O1 and O2 points in relation to the electrode located in Fz. A ground electrode (gold cup) was located on the forehead. The examination conditions were as follows: stimulus frequency 1.0 Hz, cycle time 1 s, background light off, automatic artifact rejection of signals exceeding ± 50–100 µV in amplitude, low-pass filter 1.0 Hz and high-pass filters: 100 Hz (Ganzfeld bowl) or 50 Hz (Mini Ganzfeld) or 30 Hz (Flash Goggles), average of 100 sweeps. With the Ganzfeld bowl and Mini Ganzfeld stimulators, the flash generated by white-light-emitting diodes (LEDs) presented standard flash of 3 cd s m−2. The Flash Goggles used red LED flash of 3 cd s m−2. Flash duration in Ganzfeld bowl was 0.91 s; in Mini Ganzfeld and Flash Goggles, it was 1 s. Amplifier V2.0 with original settings for each stimulator (done by Roland Consult) was used.
Measurements of the P2 amplitude was taken from the preceding N2 negative peak to positive P2 peak at around 120 ms (from 100 to 140 ms). The P2 wave amplitude after initial Ganzfeld stimulation with eye open was higher than 5 µV at least from one of two electrodes (O1 and O2). The P2 wave was identified as the largest peak between 100 and 140 ms after stimulation.
Using the Wilcoxon signed-rank test, the amplitude and the latency of the P2 wave, obtained from electrode locations O1 and O2 separately after right eye (RE) and left eye (LE) stimulation (n = 17), were compared. The first comparison concerned the results acquired with each of the three stimulators and the second the results acquired with open and closed eyes.

Results

The mean P2 amplitude (latency) acquired with the Ganzfeld bowl stimulation with open eyes (right eye, electrode at O1) amounted to 13 µV (129 ms), that acquired with the Mini Ganzfeld stimulation to 7 µV (114 ms), and that acquired with the Flash Goggles to 10 µV (110 ms) (Tables 1, 2; Figs. 1, 2). Statistically, the differences of P2 amplitude between Ganzfeld bowl and Mini Ganzfeld and between Mini Ganzfeld and Flash Goggles were significant (p = 0.005 and p = 0.008, respectively). The difference between Ganzfeld bowl and Flash Goggles was not significant (p = 0.1). The differences of P2 latency between Ganzfeld bowl and Mini Ganzfeld and between Ganzfeld bowl and Flash Goggles were significant (p = 0.003 and p = 0.002, respectively). The difference between Mini Ganzfeld and Flash Goggles was not significant (p = 0.3).
Table 1
P2 amplitude in examinations with the Ganzfeld bowl, Mini Ganzfeld and Flash Goggles with eyes open (n = 17)
Amplitude P2 [μV] RE O1
Average
Median
Min.
Max.
SE
Amplitude P2 [μV] LE O1
Average
Median
Min.
Max.
SE
Ganzfeld
13.5
8.9
4.8
38.2
2.44
Ganzfeld
12.4
12.4
5.8
20.8
1.06
Mini Ganzfeld
7.3
6.6
2.4
14.3
0.89
Mini Ganzfeld
7.2
6.7
1.3
17.8
0.98
Flash Goggles
10.2
9.8
4.1
20.6
1.08
Flash Goggles
9.1
8.3
3.7
15.4
0.83
Amplitude P2 [μV] RE O2
Average
Median
Min.
Max.
SE
Amplitude P2 [μV] LE O2
Average
Median
Min.
Max.
SE
Ganzfeld
14.9
12.8
2.3
40.3
2.49
Ganzfeld
13.9
12.8
4.9
27.7
1.43
Mini Ganzfeld
8.1
8.3
0.9
12.7
0.71
Mini Ganzfeld
8.8
9.1
0.4
18
1.02
Flash Goggles
11.5
10.3
5.7
19.8
0.95
Flash Goggles
10.9
10.5
4.6
21.9
1.16
Table 2
P2 latency in examinations with the Ganzfeld bowl, Mini Ganzfeld and Flash Goggles with eyes open (n = 17)
Latency P2 [ms] RE O1
Average
Median
Min.
Max.
SE
Latency P2 [ms] LE O1
Average
Median
Min.
Max.
SE
Ganzfeld
129
132
100
147
3.03
Ganzfeld
126
126
104
148
3.37
Mini Ganzfeld
114
117
81
141
4.02
Mini Ganzfeld
122
127
81
142
3.68
Flash Goggles
110
114
86
149
3.96
Flash Goggles
112
113
88
149
3.77
Latency P2 [ms] RE O2
Average
Median
Min.
Max.
SE
Latency P2 [ms] LE O2
Average
Median
Min.
Max.
SE
Ganzfeld
129
132
100
147
3.06
Ganzfeld
125
123
103
148
3.56
Mini Ganzfeld
119
124
79
142
3.89
Mini Ganzfeld
120
127
81
136
3.71
Flash Goggles
112
115
81
146
4.3
Flash Goggles
112
112
91
144
3.29
The mean P2 amplitude (latency) acquired with the Ganzfeld bowl stimulation with eyes closed amounted to 11 µV (127 ms), that acquired with the Mini Ganzfeld stimulation to 8 µV (125 ms), and that acquired with the Flash Goggles to 8 µV (121 ms) (Tables 3, 4). The results for open and closed eyes were, from a statistical point of view, significantly different in the case of both the Mini Ganzfeld (P2 latency: p = 0.003; P2 amplitude: p = 0.3) and the Flash Goggles (P2 latency: p = 0.006; P2 amplitude: p = 0.06) stimulators (Figs. 3, 4, 5). No significant differences between the results for open and closed eyes acquired with the Ganzfeld bowl were found (P2 latency: p = 0.8; P2 amplitude p = 0.4).
Table 3
P2 amplitude in examinations with the Ganzfeld bowl, Mini Ganzfeld and Flash Goggles with eyes closed (n = 17)
Amplitude P2 [μV] RE O1
Average
Median
Min.
Max.
SE
Amplitude P2 [μV] LE O1
Average
Median
Min.
Max.
SE
Ganzfeld
11.6
10.4
2.9
21.3
1.34
Ganzfeld
11.3
11.3
2.26
21.4
1.32
Mini Ganzfeld
8.2
6.5
2.8
17.5
1.04
Mini Ganzfeld
9.4
9.2
1.3
18.2
1.11
Flash Goggles
8.4
6.9
3.5
27.6
1.38
Flash Goggles
7.1
5.9
2.3
21.1
1.13
Amplitude P2 [μV] RE O2
Average
Median
Min.
Max.
SE
Amplitude P2 [μV] LE O2
Average
Median
Min.
Max.
SE
Ganzfeld
11.6
11.5
2.7
19.9
1.5
Ganzfeld
13.3
12.2
0.64
24.9
1.87
Mini Ganzfeld
9.8
11
1.8
19.2
1.17
Mini Ganzfeld
10.6
10.7
0.72
22.7
1.42
Flash Goggles
8.1
7.7
1.7
25.4
1.39
Flash Goggles
8.2
6
1.5
22.8
1.31
Table 4
P2 latency in examinations with the Ganzfeld bowl, Mini Ganzfeld and Flash Goggles with eyes closed (n = 17)
Latency P2 [ms] RE O1
Average
Median
Min.
Max.
SE
Latency P2 [ms] LE O1
Average
Median
Min.
Max.
SE
Ganzfeld
127
130
90
150
4.1
Ganzfeld
130
139
100
155
4.23
Mini Ganzfeld
125
120
100
150
4.1
Mini Ganzfeld
130
137
105
150
4.26
Flash Goggles
121
122
78
148
5.65
Flash Goggles
119
120
92
156
5.52
Latency P2 [ms] RE O2
Average
Median
Min.
Max.
SE
Latency P2 [ms] LE O2
Average
Median
Min.
Max.
SE
Ganzfeld
129
132
90
151
4.24
Ganzfeld
128
131
100
145
3.66
Mini Ganzfeld
124
124
101
149
3.53
Mini Ganzfeld
131
139
106
151
4.11
Flash Goggles
121
125
83
146
5.15
Flash Goggles
118
120
88
156
5.11

Discussion

The increasing number of patients with hypoxic brain damage can contribute to the fact that the FVEP will be useful as a prognostic value not only for children, but also for adult patients [10]. The conditions of the examinations have to be often adjusted to the patient; for example, the examination can be conducted in a supine position. When the improved health condition and cooperation with the examined patient make it possible to conduct a control examination in the standard conditions, it is important to bear in mind that even the same flash frequency, even though obtained by the means of a different type of stimulator, will trigger a response of a different value of latency and amplitude.
In the examination using the Mini Ganzfeld and Flash Goggles, the P2 amplitudes were lower than after the stimulation with the Ganzfeld bowl. P2 latency was longest after the Ganzfeld bowl stimulation and shortest after the Flash Goggles. The differences between particular stimuli stem mainly from the difference in the settings of amplifiers and stimulus wavelength characteristics.
It is easier to explain the differences between responses obtained with open and closed eyes. Decreased flash luminance, as a result of closed eyelids, was the reason for the extension of the wave latency. Such was also the conclusion following from the study conducted by Subramanian et al. [11]. At the same time, the researchers proved a higher FVEP reproducibility and result stabilization in the consecutive examinations with closed eyes. Therefore, they suggest conducting the FVEP with closed eyes.
In the present study after closing eyes, VEPs in the majority of cases became more regular, almost sinusoidal with not so sharp and clearly mark peaks. It cannot be ruled out that the decrease in the quality of the flash, in the case of closed eyes, caused a weaker response to stimulation, with a simultaneously recorded higher basic brain activity. Halliday recorded sinusoidal discharge after flash stimulation and described its potential connection with the resting rhythm alpha EEG (electroencephalography) [12]. The alpha rhythm is recorded, among others, in coma [13]. The lack of fixation in the case of closed eyes and reduced alertness could have caused the visual response to be significantly modified by the increase in a different activity of the brain. Opening eyes probably causes this activity to be suppressed by a dominant activity of the visual system. A lower P2 amplitude after stimulation through eyelid should be expected because of lower intensity of flash. On the other hand, superimposition of signals evoked by flash and other brain rhythms in some conditions may be an explanation of higher P2 amplitude when eyes are closed.
A different response after the Flash Goggles stimulation can also result from the application of a dozen or so many red light diodes in this stimulator. In the clinical practice, it is crucial to standardize the means of stimulation in order to enable the juxtaposition of results obtained from a patient in different age and clinical condition. It is particularly important in, for example, children and adults with hydrocephalus, craniostenosis or brain tumors, examined repeatedly in connection with the possibility of occurrence of intracranial hypertension.
Studies by Sjöstrom et al. [14] and Desch [15] constitute examples of applying the VEP to monitor the intracranial pressure. Applying the Ganzfeld bowl stimulation from the very beginning seems to be the best choice, even in young children. The Ganzfeld bowl examination gives the child the biggest freedom, contact with the stimulator is minimal, and the examination does not require sedation.
The results of this study should be of interest within the context of multicenter clinical trials and in clinical practice when a given patient’s condition is evaluated on the basis of the FVEP results obtained at different times and using different methods.
The current study had numerous limitations: The examination lasted about 30 min per patient, and stimulations were not repeated. Samples are small, and in some cases, the differences did not reach statistical significance.

Conclusion

The amplitudes and latencies of the FVEP P2 elicited with different stimulators are not suitable for comparison. Closing the eye during the examination had a significant effect on the components of FVEP waveform elicited with the Flash Goggle and on the latency of P2 elicited with the Mini Ganzfeld.

Acknowledgements

The study was supported by the Medical University of Silesia, Grant KNW-1-156/N/3/0. The sponsor had no role in the design or conduct of this research.

Compliance with ethical standards

Conflict of interest

All authors certify that they have no affiliations with or involvement in any organization or entity with any financial interest (such as honoraria; educational grants; participation in speakers’ bureaus; membership, employment, consultancies, stock ownership, or other equity interest; and expert testimony or patent-licensing arrangements), or non-financial interest (such as personal or professional relationships, affiliations, knowledge or beliefs) in the subject matter or materials discussed in this manuscript.

Ethical approval

The research was performed under the Medical University of Silesia in Katowice Bioethical Commission agreement—no. KNW/0022/KB1/141/11.

Statement of human rights

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee (Bioethical Commission of the Medical University of Silesia in Katowice) and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.

Statement on the welfare of animals

This article does not contain any studies with animals performed by any of the authors.
Informed consent was obtained from all individual participants included in the study.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Literatur
1.
Zurück zum Zitat Benedičič M, Bošnjak R (2011) Optic nerve potentials and cortical potentials after stimulation of the anterior visual pathway during neurosurgery. Doc Ophthalmol 122(2):115–125CrossRefPubMed Benedičič M, Bošnjak R (2011) Optic nerve potentials and cortical potentials after stimulation of the anterior visual pathway during neurosurgery. Doc Ophthalmol 122(2):115–125CrossRefPubMed
3.
Zurück zum Zitat Chayasirisobhon S, Gurbani S, Chai EE, Spurgeon B, Yu L, Bosu S, Gurbani S (2012) Evaluation of maturation and function of visual pathways in neonates: role of flash visual-evoked potentials revisited. Clin EEG Neurosci 43(1):18–22CrossRefPubMed Chayasirisobhon S, Gurbani S, Chai EE, Spurgeon B, Yu L, Bosu S, Gurbani S (2012) Evaluation of maturation and function of visual pathways in neonates: role of flash visual-evoked potentials revisited. Clin EEG Neurosci 43(1):18–22CrossRefPubMed
4.
Zurück zum Zitat Klebermass-Schrehof K, Rona Z, Waldhör T, Czaba C, Beke A, Weninger M, Olischar M (2013) Can neurophysiological assessment improve timing of intervention in posthaemorrhagic ventricular dilatation? Arch Dis Child Fetal Neonatal Ed 98:F291–F297CrossRefPubMed Klebermass-Schrehof K, Rona Z, Waldhör T, Czaba C, Beke A, Weninger M, Olischar M (2013) Can neurophysiological assessment improve timing of intervention in posthaemorrhagic ventricular dilatation? Arch Dis Child Fetal Neonatal Ed 98:F291–F297CrossRefPubMed
5.
Zurück zum Zitat Feng JJ, Wang WP, Guo SJ, Liu ZW, Xu X (2013) Flash visual evoked potentials in preterm infants. Ophthalmology 120(3):489–494CrossRefPubMed Feng JJ, Wang WP, Guo SJ, Liu ZW, Xu X (2013) Flash visual evoked potentials in preterm infants. Ophthalmology 120(3):489–494CrossRefPubMed
6.
Zurück zum Zitat Jethani J, Jethani M (2013) Flash visual evoked potentials in patients with periventricular leucomalacia in children less than 1 year of age. Indian J Ophthalmol 61(11):634–635CrossRefPubMedPubMedCentral Jethani J, Jethani M (2013) Flash visual evoked potentials in patients with periventricular leucomalacia in children less than 1 year of age. Indian J Ophthalmol 61(11):634–635CrossRefPubMedPubMedCentral
7.
Zurück zum Zitat Shepherd AJ, Saunders KJ, McCulloch DL, Dutton GN (1999) Prognostic value of flash visual evoked potentials in preterm infants. Dev Med Child Neurol 41:9–15CrossRefPubMed Shepherd AJ, Saunders KJ, McCulloch DL, Dutton GN (1999) Prognostic value of flash visual evoked potentials in preterm infants. Dev Med Child Neurol 41:9–15CrossRefPubMed
8.
Zurück zum Zitat Tartaglione A, Bandini F, Maculotti M, Marogna M, Spadavecchia L, Favale E (1995) Eye closure affects flash VEP latency in dementia. Electroencephalogr Clin Neurophysiol 96:197–205CrossRefPubMed Tartaglione A, Bandini F, Maculotti M, Marogna M, Spadavecchia L, Favale E (1995) Eye closure affects flash VEP latency in dementia. Electroencephalogr Clin Neurophysiol 96:197–205CrossRefPubMed
9.
Zurück zum Zitat Strigaro G, Prandi P, Varrasi C, Monaco F, Cantello R (2012) Defective visual inhibition in photosensitive idiopathic generalized epilepsy. Epilepsia 53(4):695–704CrossRefPubMed Strigaro G, Prandi P, Varrasi C, Monaco F, Cantello R (2012) Defective visual inhibition in photosensitive idiopathic generalized epilepsy. Epilepsia 53(4):695–704CrossRefPubMed
11.
Zurück zum Zitat Subramanian SK, Gaur GS, Narayan SK (2013) Low luminance/eyes closed and monochromatic stimulations reduce variability of flash visual evoked potential latency. Ann Indian Acad Neurol 16(4):614–618CrossRefPubMedPubMedCentral Subramanian SK, Gaur GS, Narayan SK (2013) Low luminance/eyes closed and monochromatic stimulations reduce variability of flash visual evoked potential latency. Ann Indian Acad Neurol 16(4):614–618CrossRefPubMedPubMedCentral
12.
Zurück zum Zitat Halliday AM (1993) The visual evoked potential in healthy subjects. In: Halliday AM (ed) Evoked potentials in clinical testing, 2nd edn. Churchill Livingstone, Edinburgh, pp 57–113 Halliday AM (1993) The visual evoked potential in healthy subjects. In: Halliday AM (ed) Evoked potentials in clinical testing, 2nd edn. Churchill Livingstone, Edinburgh, pp 57–113
13.
Zurück zum Zitat Bragatti JA, Mattos AM, Bastes H, Riesgo RS (2008) Alfa coma pattern in a child. Clin EEG Neurosci 39(4):206–209CrossRefPubMed Bragatti JA, Mattos AM, Bastes H, Riesgo RS (2008) Alfa coma pattern in a child. Clin EEG Neurosci 39(4):206–209CrossRefPubMed
14.
Zurück zum Zitat Sjöstrom A, Uvebrant P, Roos A (1995) The light flash-evoked response as a possible indicator of increased intracranial pressure in hydrocephalus. Childs Nerv Syst 11(7):381–387CrossRefPubMed Sjöstrom A, Uvebrant P, Roos A (1995) The light flash-evoked response as a possible indicator of increased intracranial pressure in hydrocephalus. Childs Nerv Syst 11(7):381–387CrossRefPubMed
15.
Zurück zum Zitat Desch LW (2001) Longitudinal stability of visual evoked potentials in children and adolescents with hydrocephalus. Dev Med Child Neurol 43(2):113–117CrossRefPubMed Desch LW (2001) Longitudinal stability of visual evoked potentials in children and adolescents with hydrocephalus. Dev Med Child Neurol 43(2):113–117CrossRefPubMed
Metadaten
Titel
Flash visual evoked potentials (FVEP) in various stimulation conditions
verfasst von
Dorota Pojda-Wilczek
Wojciech Maruszczyk
Sebastian Sirek
Publikationsdatum
23.11.2018
Verlag
Springer Berlin Heidelberg
Erschienen in
Documenta Ophthalmologica / Ausgabe 1/2019
Print ISSN: 0012-4486
Elektronische ISSN: 1573-2622
DOI
https://doi.org/10.1007/s10633-018-9663-9

Weitere Artikel der Ausgabe 1/2019

Documenta Ophthalmologica 1/2019 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.