Erschienen in: Clinical Oral Investigations 2/2020

08.07.2019 | Original Article

Fracture load of three-unit full-contour fixed dental prostheses fabricated with subtractive and additive CAD/CAM technology

verfasst von: Moritz Zimmermann, Andreas Ender, Thomas Attin, Albert Mehl

Erschienen in: Clinical Oral Investigations | Ausgabe 2/2020

Einloggen, um Zugang zu erhalten

Abstract

Objectives

The aim of this study was to test the fracture load of ceramic and composite three-unit full-contour fixed dental prostheses (FDPs) fabricated with additive and subtractive computer-aided design (CAD)/computer-aided manufacturing (CAM) technology.

Materials and methods

A newly developed alveolar socket replica model for a three-unit FDP replacing one molar was used in this study. Five CAD/CAM materials were used for fabrication of three-unit FDPs (each n = 12). The subtractive CAD/CAM fabrication method was used for groups BC (BRILLIANT Crios), TC (Telio CAD), EX (e.max CAD), and TZ (inCoris TZI C), and the additive method was used for group 3D (els 3D resin even stronger). FDPs were adhesively seated to the abutment dies (PANAVIA V5 system). Thermomechanical loading was performed prior to fracture testing with a universal testing machine. The data for maximum fracture load values was analyzed with one-way ANOVA and post hoc Scheffé test (α = 0.05).

Results

All FDPs survived the thermomechanical loading test. Statistically significant differences were found for the fracture load of three-unit FDPs fabricated from different CAD/CAM materials (p < 0.05). The highest mean fracture load was found for group TZ (2099.5 ± 382.1 N). Group 3D showed the lowest mean fracture load (928.9 ± 193.8 N). Group BC performed statistically significantly differently from group 3D with a mean fracture load of 1494.8 ± 214.5 N (p < 0.05).

Conclusions

Particle-filled composite resin CAD/CAM materials showed fracture load values within the range of ceramic materials with a specific indication of use for three-unit FDPs.

Clinical relevance

Particle filled composite CAD/CAM materials may offer new treatment possibilities for the CAD/CAM workflow.
Literatur
1.
Zurück zum Zitat Yamashita J, Wang Q, Dechow PC (2006) Biomechanical effects of fixed partial denture therapy on strain patterns of the mandible. J Prosthet Dent 95:55–62PubMed Yamashita J, Wang Q, Dechow PC (2006) Biomechanical effects of fixed partial denture therapy on strain patterns of the mandible. J Prosthet Dent 95:55–62PubMed
2.
Zurück zum Zitat Walton TR (2015) An up-to-15-year comparison of the survival and complication burden of three-unit tooth-supported fixed dental prostheses and implant-supported single crowns. Int J Oral Maxillofac Implants 30:851–861PubMed Walton TR (2015) An up-to-15-year comparison of the survival and complication burden of three-unit tooth-supported fixed dental prostheses and implant-supported single crowns. Int J Oral Maxillofac Implants 30:851–861PubMed
3.
Zurück zum Zitat Poggio CE, Ercoli C, Rispoli K, Maiorana C, Esposito M (2017) Metal-free materials for fixed prosthodontic restorations. Cochrane Database Syst Rev (12):Cd009606 Poggio CE, Ercoli C, Rispoli K, Maiorana C, Esposito M (2017) Metal-free materials for fixed prosthodontic restorations. Cochrane Database Syst Rev (12):Cd009606
4.
Zurück zum Zitat Kerschbaum T, Haastert B, Marinello CP (1996) Risk of debonding in three-unit resin-bonded fixed partial dentures. J Prosthet Dent 75:248–253PubMed Kerschbaum T, Haastert B, Marinello CP (1996) Risk of debonding in three-unit resin-bonded fixed partial dentures. J Prosthet Dent 75:248–253PubMed
5.
Zurück zum Zitat De Backer H, Van Maele G, De Moor N, Van den Berghe L (2006) Single-tooth replacement: is a 3-unit fixed partial denture still an option? A 20-year retrospective study. Int J Prosthodont 19:567–573PubMed De Backer H, Van Maele G, De Moor N, Van den Berghe L (2006) Single-tooth replacement: is a 3-unit fixed partial denture still an option? A 20-year retrospective study. Int J Prosthodont 19:567–573PubMed
6.
Zurück zum Zitat Kern M, Sasse M, Wolfart S (2012) Ten-year outcome of three-unit fixed dental prostheses made from monolithic lithium disilicate ceramic. J Am Dent Assoc 143:234–240PubMed Kern M, Sasse M, Wolfart S (2012) Ten-year outcome of three-unit fixed dental prostheses made from monolithic lithium disilicate ceramic. J Am Dent Assoc 143:234–240PubMed
7.
Zurück zum Zitat Pelaez J, Cogolludo PG, Serrano B, Lozano JF, Suarez MJ (2012) A prospective evaluation of zirconia posterior fixed dental prostheses: three-year clinical results. J Prosthet Dent 107:373–379PubMed Pelaez J, Cogolludo PG, Serrano B, Lozano JF, Suarez MJ (2012) A prospective evaluation of zirconia posterior fixed dental prostheses: three-year clinical results. J Prosthet Dent 107:373–379PubMed
8.
Zurück zum Zitat Reich S, Endres L, Weber C, Wiedhahn K, Neumann P, Schneider O, Rafai N, Wolfart S (2014) Three-unit CAD/CAM-generated lithium disilicate FDPs after a mean observation time of 46 months. Clin Oral Investig 18:2171–2178PubMed Reich S, Endres L, Weber C, Wiedhahn K, Neumann P, Schneider O, Rafai N, Wolfart S (2014) Three-unit CAD/CAM-generated lithium disilicate FDPs after a mean observation time of 46 months. Clin Oral Investig 18:2171–2178PubMed
9.
Zurück zum Zitat Yang Y, Yu J, Gao J, Guo J, Li L, Zhao Y, Zhan S (2016) Clinical outcomes of different types of tooth-supported bilayer lithium disilicate all-ceramic restorations after functioning up to 5 years: a retrospective study. J Dent 51:56–61PubMed Yang Y, Yu J, Gao J, Guo J, Li L, Zhao Y, Zhan S (2016) Clinical outcomes of different types of tooth-supported bilayer lithium disilicate all-ceramic restorations after functioning up to 5 years: a retrospective study. J Dent 51:56–61PubMed
10.
Zurück zum Zitat Wolfart S, Eschbach S, Scherrer S, Kern M (2009) Clinical outcome of three-unit lithium-disilicate glass-ceramic fixed dental prostheses: up to 8 years results. Dent Mater 25:e63–e71PubMed Wolfart S, Eschbach S, Scherrer S, Kern M (2009) Clinical outcome of three-unit lithium-disilicate glass-ceramic fixed dental prostheses: up to 8 years results. Dent Mater 25:e63–e71PubMed
11.
Zurück zum Zitat Land MF, Hopp CD (2010) Survival rates of all-ceramic systems differ by clinical indication and fabrication method. J Evid Based Dent Pract 10:37–38PubMed Land MF, Hopp CD (2010) Survival rates of all-ceramic systems differ by clinical indication and fabrication method. J Evid Based Dent Pract 10:37–38PubMed
12.
Zurück zum Zitat Beuer F, Steff B, Naumann M, Sorensen JA (2008) Load-bearing capacity of all-ceramic three-unit fixed partial dentures with different computer-aided design (CAD)/computer-aided manufacturing (CAM) fabricated framework materials. Eur J Oral Sci 116:381–386PubMed Beuer F, Steff B, Naumann M, Sorensen JA (2008) Load-bearing capacity of all-ceramic three-unit fixed partial dentures with different computer-aided design (CAD)/computer-aided manufacturing (CAM) fabricated framework materials. Eur J Oral Sci 116:381–386PubMed
13.
14.
Zurück zum Zitat Wiegand A, Stucki L, Hoffmann R, Attin T, Stawarczyk B (2015) Repairability of CAD/CAM high-density PMMA- and composite-based polymers. Clin Oral Investig 19:2007–2013PubMed Wiegand A, Stucki L, Hoffmann R, Attin T, Stawarczyk B (2015) Repairability of CAD/CAM high-density PMMA- and composite-based polymers. Clin Oral Investig 19:2007–2013PubMed
15.
Zurück zum Zitat Pfeilschifter M, Preis V, Behr M, Rosentritt M (2018) Edge strength of CAD/CAM materials. J Dent 74:95–100PubMed Pfeilschifter M, Preis V, Behr M, Rosentritt M (2018) Edge strength of CAD/CAM materials. J Dent 74:95–100PubMed
16.
Zurück zum Zitat Zimmermann M, Mehl A, Reich S (2013) New CAD/CAM materials and blocks for chairside procedures. Int J Comput Dent 16:173–181PubMed Zimmermann M, Mehl A, Reich S (2013) New CAD/CAM materials and blocks for chairside procedures. Int J Comput Dent 16:173–181PubMed
17.
Zurück zum Zitat Giordano R (2006) Materials for chairside CAD/CAM-produced restorations. J Am Dent Assoc 137 Suppl:14s–21sPubMed Giordano R (2006) Materials for chairside CAD/CAM-produced restorations. J Am Dent Assoc 137 Suppl:14s–21sPubMed
19.
Zurück zum Zitat Rosentritt M, Behr M, Lang R, Handel G (2004) Flexural properties of prosthetic provisional polymers. Eur J Prosthodont Restor Dent 12:75–79PubMed Rosentritt M, Behr M, Lang R, Handel G (2004) Flexural properties of prosthetic provisional polymers. Eur J Prosthodont Restor Dent 12:75–79PubMed
20.
Zurück zum Zitat Edelhoff D, Schraml D, Eichberger M, Stawarczyk B (2016) Comparison of fracture loads of CAD/CAM and conventionally fabricated temporary fixed dental prostheses after different aging regimens. Int J Comput Dent 19:101–112PubMed Edelhoff D, Schraml D, Eichberger M, Stawarczyk B (2016) Comparison of fracture loads of CAD/CAM and conventionally fabricated temporary fixed dental prostheses after different aging regimens. Int J Comput Dent 19:101–112PubMed
21.
Zurück zum Zitat Behr M, Rosentritt M, Handel G (2003) Fiber-reinforced composite crowns and FPDs: a clinical report. Int J Prosthodont 16:239–243PubMed Behr M, Rosentritt M, Handel G (2003) Fiber-reinforced composite crowns and FPDs: a clinical report. Int J Prosthodont 16:239–243PubMed
22.
Zurück zum Zitat Shi L, Fok AS (2009) Structural optimization of the fibre-reinforced composite substructure in a three-unit dental bridge. Dent Mater 25:791–801PubMed Shi L, Fok AS (2009) Structural optimization of the fibre-reinforced composite substructure in a three-unit dental bridge. Dent Mater 25:791–801PubMed
23.
Zurück zum Zitat Erkmen E, Meric G, Kurt A, Tunc Y, Eser A (2011) Biomechanical comparison of implant retained fixed partial dentures with fiber reinforced composite versus conventional metal frameworks: a 3D FEA study. J Mech Behav Biomed Mater 4:107–116PubMed Erkmen E, Meric G, Kurt A, Tunc Y, Eser A (2011) Biomechanical comparison of implant retained fixed partial dentures with fiber reinforced composite versus conventional metal frameworks: a 3D FEA study. J Mech Behav Biomed Mater 4:107–116PubMed
24.
Zurück zum Zitat Keilig L, Stark H, Bourauel C (2016) Does the material stiffness of novel high-performance polymers for fixed partial dentures influence their biomechanical behavior? Int J Prosthodont 30:595–597PubMed Keilig L, Stark H, Bourauel C (2016) Does the material stiffness of novel high-performance polymers for fixed partial dentures influence their biomechanical behavior? Int J Prosthodont 30:595–597PubMed
25.
Zurück zum Zitat Krejci I, Reich T, Lutz F, Albertoni M (1990) An in vitro test procedure for evaluating dental restoration systems. 1. A computer-controlled mastication simulator. Schweiz Monatsschr Zahnmed 100:953–960PubMed Krejci I, Reich T, Lutz F, Albertoni M (1990) An in vitro test procedure for evaluating dental restoration systems. 1. A computer-controlled mastication simulator. Schweiz Monatsschr Zahnmed 100:953–960PubMed
26.
Zurück zum Zitat Tortopidis D, Lyons MF, Baxendale RH, Gilmour WH (1998) The variability of bite force measurement between sessions, in different positions within the dental arch. J Oral Rehabil 25:681–686PubMed Tortopidis D, Lyons MF, Baxendale RH, Gilmour WH (1998) The variability of bite force measurement between sessions, in different positions within the dental arch. J Oral Rehabil 25:681–686PubMed
27.
Zurück zum Zitat Lyons MF, Cadden SW, Baxendale RH, Yemm R (1996) Twitch interpolation in the assessment of the maximum force-generating capacity of the jaw-closing muscles in man. Arch Oral Biol 41:1161–1168PubMed Lyons MF, Cadden SW, Baxendale RH, Yemm R (1996) Twitch interpolation in the assessment of the maximum force-generating capacity of the jaw-closing muscles in man. Arch Oral Biol 41:1161–1168PubMed
28.
Zurück zum Zitat Özcan M, Jonasch M (2018) Effect of cyclic fatigue tests on aging and their translational implications for survival of all-ceramic tooth-borne single crowns and fixed dental prostheses. J Prosthodont 27:364–375PubMed Özcan M, Jonasch M (2018) Effect of cyclic fatigue tests on aging and their translational implications for survival of all-ceramic tooth-borne single crowns and fixed dental prostheses. J Prosthodont 27:364–375PubMed
29.
Zurück zum Zitat Stawarczyk B, Ender A, Trottmann A, Özcan M, Fischer J, Hämmerle CH (2012) Load-bearing capacity of CAD/CAM milled polymeric three-unit fixed dental prostheses: effect of aging regimens. Clin Oral Investig 16:1669–1677 Stawarczyk B, Ender A, Trottmann A, Özcan M, Fischer J, Hämmerle CH (2012) Load-bearing capacity of CAD/CAM milled polymeric three-unit fixed dental prostheses: effect of aging regimens. Clin Oral Investig 16:1669–1677
30.
Zurück zum Zitat Blackburn C, Rask H, Awada A (2018) Mechanical properties of resin-ceramic CAD-CAM materials after accelerated aging. J Prosthet Dent 119:954–958PubMed Blackburn C, Rask H, Awada A (2018) Mechanical properties of resin-ceramic CAD-CAM materials after accelerated aging. J Prosthet Dent 119:954–958PubMed
31.
Zurück zum Zitat Plengsombut K, Brewer JD, Monaco EA Jr, Davis EL (2009) Effect of two connector designs on the fracture resistance of all-ceramic core materials for fixed dental prostheses. J Prosthet Dent 101:166–173PubMed Plengsombut K, Brewer JD, Monaco EA Jr, Davis EL (2009) Effect of two connector designs on the fracture resistance of all-ceramic core materials for fixed dental prostheses. J Prosthet Dent 101:166–173PubMed
32.
Zurück zum Zitat Oh W, Gotzen N, Anusavice KJ (2002) Influence of connector design on fracture probability of ceramic fixed-partial dentures. J Dent Res 81:623–627PubMed Oh W, Gotzen N, Anusavice KJ (2002) Influence of connector design on fracture probability of ceramic fixed-partial dentures. J Dent Res 81:623–627PubMed
33.
Zurück zum Zitat Vult von Steyern P, Kokubo Y, Nilner K (2005) Use of abutment-teeth vs. dental implants to support all-ceramic fixed partial dentures: an in-vitro study on fracture strength. Swed Dent J 29:53–60PubMed Vult von Steyern P, Kokubo Y, Nilner K (2005) Use of abutment-teeth vs. dental implants to support all-ceramic fixed partial dentures: an in-vitro study on fracture strength. Swed Dent J 29:53–60PubMed
34.
Zurück zum Zitat Wimmer T, Ender A, Roos M, Stawarczyk B (2013) Fracture load of milled polymeric fixed dental prostheses as a function of connector cross-sectional areas. J Prosthet Dent 110:288–295PubMed Wimmer T, Ender A, Roos M, Stawarczyk B (2013) Fracture load of milled polymeric fixed dental prostheses as a function of connector cross-sectional areas. J Prosthet Dent 110:288–295PubMed
35.
Zurück zum Zitat Choi JW, Kim SY, Bae JH, Bae EB, Huh JB (2017) In vitro study of the fracture resistance of monolithic lithium disilicate, monolithic zirconia, and lithium disilicate pressed on zirconia for three-unit fixed dental prostheses. J Adv Prosthodont 9:244–251PubMedPubMedCentral Choi JW, Kim SY, Bae JH, Bae EB, Huh JB (2017) In vitro study of the fracture resistance of monolithic lithium disilicate, monolithic zirconia, and lithium disilicate pressed on zirconia for three-unit fixed dental prostheses. J Adv Prosthodont 9:244–251PubMedPubMedCentral
36.
Zurück zum Zitat Rosentritt M, Plein T, Kolbeck C, Behr M, Handel G (2000) In vitro fracture force and marginal adaptation of ceramic crowns fixed on natural and artificial teeth. Int J Prosthodont 13:387–391PubMed Rosentritt M, Plein T, Kolbeck C, Behr M, Handel G (2000) In vitro fracture force and marginal adaptation of ceramic crowns fixed on natural and artificial teeth. Int J Prosthodont 13:387–391PubMed
37.
Zurück zum Zitat Guess PC, Schultheis S, Wolkewitz M, Zhang Y, Strub JR (2013) Influence of preparation design and ceramic thicknesses on fracture resistance and failure modes of premolar partial coverage restorations. J Prosthet Dent 110:264–273PubMedPubMedCentral Guess PC, Schultheis S, Wolkewitz M, Zhang Y, Strub JR (2013) Influence of preparation design and ceramic thicknesses on fracture resistance and failure modes of premolar partial coverage restorations. J Prosthet Dent 110:264–273PubMedPubMedCentral
38.
Zurück zum Zitat Wimmer T, Erdelt KJ, Eichberger M, Roos M, Edelhoff D, Stawarzak B (2014) Influence of abutment model materials on the fracture loads of three-unit fixed dental prostheses. Dent Mater J 33:717–724PubMed Wimmer T, Erdelt KJ, Eichberger M, Roos M, Edelhoff D, Stawarzak B (2014) Influence of abutment model materials on the fracture loads of three-unit fixed dental prostheses. Dent Mater J 33:717–724PubMed
39.
Zurück zum Zitat Rosentritt M, Behr M, Gebhard R, Handel G (2006) Influence of stress simulation parameters on the fracture strength of all-ceramic fixed-partial dentures. Dent Mater 22:176–182PubMed Rosentritt M, Behr M, Gebhard R, Handel G (2006) Influence of stress simulation parameters on the fracture strength of all-ceramic fixed-partial dentures. Dent Mater 22:176–182PubMed
40.
Zurück zum Zitat Kinney JH, Marshall SJ, Marshall GW (2003) The mechanical properties of human dentin: a critical review and re-evaluation of the dental literature. Crit Rev Oral Biol Med 14:13–29PubMed Kinney JH, Marshall SJ, Marshall GW (2003) The mechanical properties of human dentin: a critical review and re-evaluation of the dental literature. Crit Rev Oral Biol Med 14:13–29PubMed
41.
Zurück zum Zitat Ho SP, Kurylo MP, Fong TK, Lee SS, Wagner HD, Ryder MI, Marshall GW (2010) The biomechanical characteristics of the bone-periodontal ligament-cementum complex. Biomaterials 31(25):6635–6646PubMedPubMedCentral Ho SP, Kurylo MP, Fong TK, Lee SS, Wagner HD, Ryder MI, Marshall GW (2010) The biomechanical characteristics of the bone-periodontal ligament-cementum complex. Biomaterials 31(25):6635–6646PubMedPubMedCentral
42.
Zurück zum Zitat Scherrer SS, de Rijk WG, Belser UC, Meyer JM (1994) Effect of cement film thickness on the fracture resistance of a machinable glass-ceramic. Dent Mater 10:172–177PubMed Scherrer SS, de Rijk WG, Belser UC, Meyer JM (1994) Effect of cement film thickness on the fracture resistance of a machinable glass-ceramic. Dent Mater 10:172–177PubMed
43.
Zurück zum Zitat Wimmer T, Erdelt KJ, Raith S, Schneider JM, Stawarczyk B, Beuer F (2014) Effects of differing thickness and mechanical properties of cement on the stress levels and distributions in a three-unit zirconia fixed prosthesis by FEA. J Prosthodont 23:358–366PubMed Wimmer T, Erdelt KJ, Raith S, Schneider JM, Stawarczyk B, Beuer F (2014) Effects of differing thickness and mechanical properties of cement on the stress levels and distributions in a three-unit zirconia fixed prosthesis by FEA. J Prosthodont 23:358–366PubMed
44.
Zurück zum Zitat Reymus M, Roos M, Eichberger M, Edelhoff D, Hickel R, Stawarczyk B (2019) Bonding to new CAD/CAM resin composites: influence of air abrasion and conditioning agents as pretreatment strategy. Clin Oral Investig 23:529–538PubMed Reymus M, Roos M, Eichberger M, Edelhoff D, Hickel R, Stawarczyk B (2019) Bonding to new CAD/CAM resin composites: influence of air abrasion and conditioning agents as pretreatment strategy. Clin Oral Investig 23:529–538PubMed
45.
Zurück zum Zitat Mormann WH, Stawarczyk B, Ender A, Sener B, Attin T, Mehl A (2013) Wear characteristics of current aesthetic dental restorative CAD/CAM materials: two-body wear, gloss retention, roughness and Martens hardness. J Mech Behav Biomed Mater 20:113–125PubMed Mormann WH, Stawarczyk B, Ender A, Sener B, Attin T, Mehl A (2013) Wear characteristics of current aesthetic dental restorative CAD/CAM materials: two-body wear, gloss retention, roughness and Martens hardness. J Mech Behav Biomed Mater 20:113–125PubMed
Metadaten
Titel
Fracture load of three-unit full-contour fixed dental prostheses fabricated with subtractive and additive CAD/CAM technology
verfasst von
Moritz Zimmermann
Andreas Ender
Thomas Attin
Albert Mehl
Publikationsdatum
08.07.2019
Verlag
Springer Berlin Heidelberg
Erschienen in
Clinical Oral Investigations / Ausgabe 2/2020
Print ISSN: 1432-6981
Elektronische ISSN: 1436-3771
DOI
https://doi.org/10.1007/s00784-019-03000-0

Weitere Artikel der Ausgabe 2/2020

Clinical Oral Investigations 2/2020 Zur Ausgabe

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Update Zahnmedizin

Bestellen Sie unseren kostenlosen Newsletter und bleiben Sie gut informiert – ganz bequem per eMail.