Fully automated quantification of biventricular volumes and function in cardiovascular magnetic resonance: applicability to clinical routine settings

verfasst von: Sören J. Backhaus, Wieland Staab, Michael Steinmetz, Christian O. Ritter, Joachim Lotz, Gerd Hasenfuß, Andreas Schuster, Johannes T. Kowallick

Erschienen in: Journal of Cardiovascular Magnetic Resonance | Ausgabe 1/2019

Abstract

Background

Cardiovascular magnetic resonance (CMR) represents the clinical gold standard for the assessment of biventricular morphology and function. Since manual post-processing is time-consuming and prone to observer variability, efforts have been directed towards automated volumetric quantification. In this study, we sought to validate the accuracy of a novel approach providing fully automated quantification of biventricular volumes and function in a “real-world” clinical setting.

Methods

Three-hundred CMR examinations were randomly selected from the local data base. Fully automated quantification of left ventricular (LV) mass, LV and right ventricular (RV) end-diastolic and end-systolic volumes (EDV/ESV), stroke volume (SV) and ejection fraction (EF) were performed overnight using commercially available software (suiteHEART®, Neosoft, Pewaukee, Wisconsin, USA). Parameters were compared to manual assessments (QMass®, Medis Medical Imaging Systems, Leiden, Netherlands). Sub-group analyses were further performed according to image quality, scanner field strength, the presence of implanted aortic valves and repaired Tetralogy of Fallot (ToF).

Results

Biventricular automated segmentation was feasible in all 300 cases. Overall agreement between fully automated and manually derived LV parameters was good (LV-EF: intra-class correlation coefficient [ICC] 0.95; bias − 2.5% [SD 5.9%]), whilst RV agreement was lower (RV-EF: ICC 0.72; bias 5.8% [SD 9.6%]). Lowest agreement was observed in case of severely altered anatomy, e.g. marked RV dilation but normal LV dimensions in repaired ToF (LV parameters ICC 0.73–0.91; RV parameters ICC 0.41–0.94) and/or reduced image quality (LV parameters ICC 0.86–0.95; RV parameters ICC 0.56–0.91), which was more common on 3.0 T than on 1.5 T.

Conclusions

Fully automated assessments of biventricular morphology and function is robust and accurate in a clinical routine setting with good image quality and can be performed without any user interaction. However, in case of demanding anatomy (e.g. repaired ToF, severe LV hypertrophy) or reduced image quality, quality check and manual re-contouring are still required.
Hinweise

Electronic supplementary material

The online version of this article (https://​doi.​org/​10.​1186/​s12968-019-0532-9) contains supplementary material, which is available to authorized users.
Abkürzungen
AVR
Aortic valve replacement
bSSFP
Balanced steady state free precession
CI
Confidence interval
CMR
Cardiovascular magnetic resonance
CoV
Coefficient of variation
ECG
Electrocardiogram
EDV
End-diastolic volume
EF
Ejection fraction
ESV
End-systolic volume
ICC
Interclass correlation coefficient
IQR
Interquartile range
LV
Left ventricle/left ventricular
PAH
Pulmonary artery hypertension
RV
Right ventricle/right ventricular
SAX
Short axis
SCMR
Society for Cardiovascular magnetic Resonance
SV
Stroke volume
ToF
Tetralogy of Fallot

Introduction

Cardiovascular magnetic resonance (CMR) imaging is the gold standard for the assessment of cardiac function and morphology [1, 2]. Left ventricular (LV) ejection fraction (EF) is the most established parameter for cardiac functional assessments in clinical routine and is used for the evaluation of disease severity, treatment follow-up and risk assessment for adverse events [3, 4].
To extract clinically relevant information such as LV mass, LV and right ventricular (RV) end-diastolic and end-systolic volume (EDV/ESV), stroke volume (SV) and EF, accurate post-processing of the cine CMR images is required. In daily clinical routine, post-processing is typically performed manually by delineating endocardial and epicardial LV borders as well as endocardial RV borders in all short-axis (SAX) slices covering the ventricles from atrioventricular ring to apex, in both end-diastolic and end-systolic phases. This task is time-consuming, tedious and subject to observer-variability [2, 57]. Emerging post-processing software based on deep-learning algorithms using convolutional neural networks now offer a fully automated approach for LV and RV volume assessments and have recently become commercially available [8]. Initial evaluations of these automated approaches are promising [9], however often based on pre-selected cases with excellent image quality or ‘cropped’ data [9, 10], i.e. SAX stacks are manually triaged to only include end-diastolic and end-systolic images effectively covering the ventricles before applying the automatic algorithm. Importantly, most of the observer-variability results from discrepancies in defining the most apical and basal short-axis SAX slices [11], which is whitewashed by previous manual ‘cropping’ of SAX stacks.
At the current time, it remains unknown whether fully automated quantification of biventricular morphology and function is feasible and accurate in clinical CMR routine. Accordingly, the aim of the present study was to evaluate the feasibility and accuracy of fully automated biventricular assessment of morphology and function in a variety of CMR data (neither pre-selected nor pre-processed) taken from a real-world data base of a tertiary care CMR unit.

Methods

Study design

The study population consisted of 300 randomly selected patients referred to CMR within clinical routine care between 2016 and 2018. The CMR imaging protocol was employed on clinical 1.5 or 3 Tesla (Magnetom Symphony or Magnetom Skyra, Siemens Healthineers, Erlangen, Germany) CMR scanners. Protocols were employed as appropriate for clinical routine, all of which including electrocardiogram (ECG)-gated balanced steady-state free precession (bSSFP) cine sequences for a SAX stack. Typical imaging parameters were as follows: 25 frames/cardiac cycle, pixel spacing 0.8 mm × 0.8 mm, 8 mm slice thickness as well as inter-slice gap, TE 1.5 ms, TR 3 ms. The study was approved by the Ethics Committee of the University Hospital Goettingen and complied with the Declaration of Helsinki. The Ethics Committee gave permission to waive informed consent for this retrospective analysis. Furthermore, agreement was assessed between the fully automated algorithm and expert consensus contours based on the Society for Cardiovascular Magnetic Resonance (SCMR) consensus data consisting of 15 cases with different pathologies [12].

CMR analyses

Volumetric analyses were performed manually in short-axis orientations by an experienced investigator according to standardized recommendations [11] using commercially available post-processing software (QMass®, Version 3.1.16.0, Medis Medical Imaging Systems, Leiden, Netherlands). Fully automated segmentation was performed employing dedicated commercially available software (suiteHEART®, Version 4.0.6, Neosoft, Pewaukee, Wisconsin, USA). The papillary muscles were included within the myocardium, trabecular tissue was excluded from the myocardial mass using both, the manual (QMass®) and the automated (suiteHEART®) software. Manual segmentation was performed by simple delineation of the LV endocardial- and epicardial borders and the RV endocardial border with Bézier smoothing at end-diastole and end-systole. No thresholding or edge detection was applied. Cross-referencing from long-axis locations was used to adjust for systolic atrioventricular ring descent. Fully automated segmentation was done overnight without any user-interaction neither by pre-processing the acquired short axis stack nor by post-processing automatically traced borders. Analyses included LV mass, and biventricular EDV, ESV, SV and EF. Agreement was tested between manual and fully automated analyses. Reproducibility was tested by reapplying the fully automated tracking algorithm on 20 randomly selected patients and by manual volumetric analyses by two experienced investigators including intra- and inter-observer reproducibility. All operators were blinded to each other’s results. Furthermore, the analysis time needed to perform manual segmentations was measured in the subset of 20 patients. The presence and relevance of artefacts impacting image quality was graded adopting the criteria proposed by Klinke et al. [13] taking wrap around, respiratory ghost, cardiac ghost, image blurring, metal and shimming artefacts into account (Table 1). One point was given if the artefact impeded the visualization of > 1/3 of the ventricular endocardial border at end-systole and/or end-diastole on a single SAX slice. If such artefact involved 2 slices or ≥ 3 slices, 2 and 3 points were given, respectively. Furthermore, accurate short-axis orientation was evaluated, resulting in an image quality score between 0 (= excellent quality) and 6 (= poor quality). Image quality scores were separately assessed for the LV and RV myocardium.
Table 1
Quality assessment of cine short-axis (SAX) images. The image quality score corresponds to the sum of qualitative scoring based on 6 criteria (range of score: 0–5). One point was given if an artefact impeded the visualization of > 1/3 of the ventricular endocardial border at end-systole and/or end-diastole on a single SAX slice. If such artefact involved 2 slices or ≥ 3 slices, 2 and 3 points were given, respectively. Incorrect short-axis orientation was graded with 2 points
 
0
1
2
3
Maximum Score
1. Wrap around
No
1 slice
2 slices
≥3 slices
3
2. Respiratory ghost
No
1 slice
2 slices
≥3 slices
3. Cardiac ghost
No
1 slice
2 slices
≥3 slices
4. Image blurring / mis-triggering
No
1 slice
2 slices
≥3 slices
5. Metallic artefacts
No
1 slice
2 slices
≥3 slices
6. Orientation of stack
Correct
Incorrect
2
For the SCMR consensus data, only LV parameters were compared between automated analyses and manual expert consensus parameters, since RV parameters were not provided. According to the method described by Suinesiaputra et al. [12], papillary muscles and trabecular tissue were excluded from the myocardial mass.

Statistics

Continuous variables were checked for normal distribution using the Shapiro-Wilk test and are presented as median with interquartile range (IQR). Biventricular volumes and LV mass were indexed to body surface area. Dependent variables were tested using the Wilcoxon signed-rank test. Agreement of manual and automated analyses as well as intra- and inter-observer variability was assessed first using Bland-Altman analysis [mean difference between measurements with 95% confidence interval (CI)] [14], second intra-class correlation coefficients (ICC) based on a model of absolute agreement, considered excellent if ICC > 0.74, good between 0.60 and 0.74, fair between 0.4 and 0.59 and poor below 0.4 [15], and third the coefficient of variation (CoV, = standard deviation [SD] of the differences divided by the mean) [16, 17]. P-values provided are two-sided, an alpha level below 0.05 was considered statistically significant. Statistical analyses were performed using IBM SPSS Statistic Software Version 24 (International Business Machines, Armonk, New York, USA) and Microsoft Excel (Microsoft, Redmond, Washington, USA).

Results

Study population

Patient characteristics and cardiac volumes for both manual and automated assessments are presented in Table 2. Biventricular automatic segmentation was feasible in all 300 cases. In comparison with manual evaluations, automatic assessments depicted higher LV volumes, lower LVEF, higher LV mass as well as higher RV EDV, RV SV and RVEF (p < 0.001 for all). The study population consisted of 100 referrals to evaluate ischemic heart disease, 120 patients with myocardial disease, 70 patients with congenital heart disease and 10 others. Table 3 provides an overview of clinical indications. There were 31 patients imaged after aortic valve replacement (AVR) of whom 18 received transcatheter aortic valve replacement (Edwards SAPIEN 3™, Edwards Lifesciences, Irvine, California, USA), 7 patients after open-surgery AVR using a bioprosthesis (Carpentier-Edwards Perimount™, Edwards Lifesciences) and 6 patients after open-surgery AVR with a mechanical aortic valve (SJM Regent™, St. Jude Medical Inc., St Paul, Minnesota, USA).
Table 2
Demographics and biventricular volumes
Parameter
Study population
 
Gender (f/m)
118/182
 
Age
52 (30, 71)
 
BSA
1.92 (1.72, 2.05)
 
 
Automated
Manual
p
LV Mass, g/m2
59.8 (48.8, 74.1)
58.0 (46.0, 73.0)
< 0.001
LV EDV, ml/m2
88.5 (77.4, 105.8)
83.0 (71.0, 100.0)
< 0.001
LV ESV, ml/m2
37.7 (29.4, 51.1)
33.0 (25.0, 47.0)
< 0.001
LV SV, ml/m2
48.8 (41.2, 55.8)
48.0 (40.0, 56.0)
0.133
LV EF
58.0 (48.0, 63.0)
60.0 (51.0, 66.0)
< 0.001
RV EDV, ml/m2
87.0 (71.1, 107.4)
79.0 (65.0, 97.0)
< 0.001
RV ESV, ml/m2
38.4 (26.9, 48.0)
39.0 (28.0, 51.0)
0.125
RV SV, ml/m2
47.8 (40.3, 55.9)
40.5 (32.0, 48.0)
< 0.001
RV EF
56.0 (50.0, 63.0)
51.0 (44.0, 58.0)
< 0.001
Continuous variables are expressed as median and interquartile range (IQR) and were tested using the Wilcoxon signed-rank test. LV/RV left/right ventricle, EDV/ESV end-diastolic/systolic volume, SV stroke volume, EF ejection fraction
Numbers in bold type indicate a significant difference
Table 3
Clinical CMR indications
 
Number of patients
Ischemic Heart Disease
100
 Coronary Heart Disease
97
 Acute Myocardial Infarction
3
Myocardial disease
120
 Myocarditis
64
 Arrhythmogenic RV Cardiomyopathy
19
 Dilated Cardiomyopathy
14
 Hypertrophic Cardiomyopathy
11
 Sarcoidosis
7
 Iron Overload Cardiomyopathy
3
 Non-Compaction Cardiomyopathy
1
 Anderson Fabry Disease
1
Congenital Heart Disease
70
 Repaired Tetralogy of Fallot
47
 Aortic Coarctation
12
 Atrial Septal Defect
6
 Aortic Dilatation in Bicuspid Aortic Valve
5
Others
10
 Cardiac Mass
5
 Rheumatic Disease
2
 Pericarditis Constrictiva
1
 Pericardial Effusion
1
 Pulmonary Arterial Hypertension
1
RV right ventricular

Image quality and post-processing

LV-image quality was graded with 1.0 (SD 1.3) (Score 0 n = 168, Score 1 n = 19, Score 2 n = 46, and Score 3 n = 67 points. RV-image quality was graded with 1.1 (SD 1.3) (Score 0 n = 151, Score 1 n = 37, Score 2 n = 39, and Score 3 n = 73 points. Appropriate short-axis orientation was fulfilled in 298 case, the highest image quality score assigned was 3. Manual post-processing took on average 11.3 ± 1.5 min as opposed to automated pre-processing with < 1 min/SAX stack. Representative examples of high and low segmentation accuracy are given in Fig. 1. Corresponding videos including automatic segmentation of all phases and SAX slices can be found in Additional file 1.

Agreement of manual and automated analyses

Results comparing automated and manual volume assessments including mean differences with corresponding SD, ICC and CoV are presented in Tables 4 and 5. Corresponding Bland-Altman plots are presented in Figs. 2, 3 and 4. Agreement of manual and automated assessments in the overall cohort of 300 patients was excellent for all LV parameters (ICC ≥0.91), best for EDV (ICC 0.98) closely followed by ESV (ICC 0.96) as well as mass and EF (both ICC 0.95). The automated algorithm slightly overestimated LV mass, EDV and ESV while underestimating LV EF (mean difference − 2.5%, limits of agreement [LOA] -14.6 to 9.1%), p < 0.001). Agreement for RV volumes was excellent for RV EDV and ESV (both ICC 0.92) and good for RV SV (ICC 0.73) and EF (ICC 0.72). Similar to LV measurement, the automatic algorithm overestimated RV EDV, and also RV EF (mean difference 5.8%, LOA -13.0 to 24.6%, p < 0.001). Higher field strength (3 vs 1.5 Tesla) was associated with reduced agreement in biventricular volumes, though it was also associated with a decrease in image quality (1.5 T: LV image quality score 0.8 (SD 1.2), RV image quality score 0.7 (SD 1.1); 3.0 T: LV image quality score 1.4 (SD 1.3), RV image quality score 1.1 (SD 1.3); p < 0.001 for all). Similarly, aortic valve replacement resulted in lower agreement but was also accompanied by lower image quality (LV image quality score 1.9 (SD 1.2); RV image quality score 2.0 (SD 1.2). Repaired ToF was associated with decreased RV image quality (RV image quality score 1.8 [SD 1.1]) but preserved LV image quality (LV image quality score 0.62 [0.99]). Despite preserved LV image quality, agreement was reduced for both LV and RV volumes (Table 3.).
Table 4
Agreement between manual and automated segmentations. Agreement was analysed in the entire study group (n = 300) as well as in subgroups according to field strength, aortic valve replacement and repaired Tetralogy of Fallot
 
Parameter
Mean Difference (SD of the Diff.)
ICC (95% CI)
CoV (%)
All
LV Mass
2.4 (9.3)
0.95 (0.94–0.97)
14.6
(n = 300)
LV EDV
5.0 (7.9)
0.98 (0.94–0.99)
8.5
LV ESV
4.4 (11.1)
0.96 (0.94–0.97)
25.0
LV SV
0.3 (7.3)
0.91 (0.89–0.93)
15.1
LV EF
−2.5 (5.9)
0.95 (0.92–0.97)
10.6
RV EDV
7.4 (12.0)
0.92 (0.81–0.96)
14.0
RV ESV
−1.6 (9.8)
0.92 (0.89–0.93)
24.0
RV SV
9.0 (10.6)
0.73 (0.26–0.87)
23.5
RV EF
5.8 (9.6)
0.72 (0.47–0.83)
17.8
1.5 T
LV Mass
4.0 (7.2)
0.96 (0.91–0.98)
11.7
(n = 132)
LV EDV
4.4 (8.4)
0.98 (0.96–0.99)
8.8
LV ESV
2.9 (6.1)
0.99 (0.98–1.00)
12.2
LV SV
1.1 (6.4)
0.95 (0.92–0.96)
13.7
LV EF
−1.5 (4.9)
0.97 (0.96–0.98)
9.3
RV EDV
10.6 (9.6)
0.90 (0.33–0.97)
11.7
RV ESV
2.1 (8.0)
0.94 (0.91–0.96)
20.5
RV SV
8.5 (8.8)
0.72 (0.11–0.88)
20.6
RV EF
3.9 (8.8)
0.77 (0.62–0.86)
16.5
3 T
LV Mass
0.1 (9.7)
0.97 (0.95–0.98)
13.2
(n = 90)
LV EDV
5.5 (7.3)
0.98 (0.91–0.99)
7.8
LV ESV
6.9 (17.7)
0.88 (0.79–0.92)
40.5
LV SV
−1.4 (8.5)
0.84 (0.75–0.89)
17.1
LV EF
−3.8 (6.9)
0.92 (0.83–0.96)
12.3
RV EDV
5.8 (12.2)
0.89 (0.79–0.94)
15.0
RV ESV
−0.7 (7.9)
0.94 (0.92–0.96)
20.7
RV SV
6.2 (11.0)
0.64 (0.35–0.79)
25.3
RV EF
4.1 (8.4)
0.83 (0.68–0.90)
15.3
Aortic Valve replacement
LV Mass
1.3 (11.5)
0.89 (0.76–0.94)
16.7
(n = 31)
LV EDV
5.0 (7.7)
0.97 (0.91–0.99)
8.7
LV ESV
5.5 (5.6)
0.97 (0.76–0.99)
14.8
LV SV
−0.4 (7.6)
0.93 (0.85–0.96)
15.2
LV EF
−4.4 (6.4)
0.92 (0.73–0.97)
10.8
RV EDV
6.0 (13.9)
0.90 (0.78–0.95)
16.6
RV ESV
−6.9 (10.6)
0.86 (0.60–0.94)
28.9
RV SV
12.8 (15.0)
0.54 (0.00–0.79)
31.6
RV EF
10.3 (13.7)
0.54 (0.00–0.79)
23.7
Tetralogy of Fallot
LV Mass
2.9 (11.3)
0.75 (0.55–0.86)
23.0
(n = 47)
LV EDV
5.6 (7.8)
0.91 (0.69–0.96)
9.4
LV ESV
3.1 (6.4)
0.83 (0.66–0.91)
18.7
LV SV
1.9 (6.5)
0.87 (0.76–0.93)
13.1
LV EF
−1.5 (5.3)
0.73 (0.52–0.85)
8.9
RV EDV
2.6 (14.3)
0.94 (0.90–0.97)
13.5
RV ESV
− 10.1 (10.5)
0.82 (0.25–0.93)
20.2
RV SV
13.0 (9.4)
0.81 (0.00–0.94)
17.4
RV EF
11.3 (7.5)
0.41 (0.00–0.73)
14.8
Biventricular volumes and LV mass were indexed to body surface area. T: Tesla. SD standard deviation, ICC intraclass correlation coefficient, CoV coefficient of variation, LV left ventricular, RV right ventricular, EDV/ESV end-diastolic/systolic volume, SV stroke volume, EF ejection fraction
Table 5
Agreement between manual and automated analyses according to image quality
 
Parameter
Mean Difference (SD of the Diff.)
ICC (95% CI)
CoV (%)
Good image quality (Score ≤ 1)
LV (n = 187)
LV Mass
3.0 (7.9)
0.96 (0.94–0.98)
12.8
LV EDV
3.4 (6.2)
0.99 (0.97–0.99)
6.5
LV ESV
1.7 (4.1)
0.99 (0.99–1.00)
9.0
LV SV
1.5 (5.2)
0.95 (0.93–0.96)
10.5
LV EF
−0.6 (3.5)
0.98 (0.98–0.99)
6.2
RV (n = 188)
RV EDV
7.8 (10.5)
0.93 (0.75–0.97)
12.2
RV ESV
1.0 (6.9)
0.96 (0.95–0.97)
16.9
RV SV
6.7 (8.5)
0.79 (0.40–0.90)
18.9
RV EF
3.0 (6.1)
0.88 (0.78–0.93)
11.5
Reduced image quality (Score ≥ 2)
LV (n = 113)
LV Mass
1.3 (11.2)
0.94 (0.91–0.96)
16.7
LV EDV
7.5 (9.7)
0.95 (0.82–0.98)
10.8
LV ESV
8.9 (16.3)
0.87 (0.74–0.93)
37.9
LV SV
−1.6 (9.6)
0.86 (0.80–0.90)
20.5
LV EF
−5.6 (7.5)
0.90 (0.67–0.96)
13.8
RV (n = 112)
RV EDV
6.7 (14.3)
0.91 (0.83–0.95)
16.8
RV ESV
−5.9 (12.1)
0.84 (0.70–0.90)
30.2
RV SV
12.7 (12.6)
0.67 (0.02–0.86)
27.9
RV EF
10.6 (12.1)
0.56 (0.03–0.77)
22.5
Biventricular volumes and LV mass were indexed to body surface area. SD standard deviation, ICC intraclass correlation coefficient, CoV coefficient of variation, LV left ventricular, RV right ventricular, EDV/ESV end-diastolic/systolic volume, SV stroke volume, EF ejection fraction
If classified according to image quality score, 0 to 1 point was associated with considerable better agreement than 2 to 3 points, both for LV and RV automated analyses. Considering an image quality score of ≤1, both LV and RV agreements were excellent for all variables with a bias of − 0.6% (LOA -7.6 to 6.4%) and 3.0% (LOA -9.2 to 15.2%) for LV EF and RV EF, respectively. However, large differences were observed in case of reduced image quality (score ≥ 2) with a bias of − 5.6% (LOA -20.6 to 9.4%) and 10.6% (LOA -13.6 to 34.8%) for LV EF and RV EF, respectively. LV and RV stroke volumes were very consistent in automated analyses, LV 48.1 ml/m2 compared to RV 47.6 ml/m2 in median, p = 0.435.
Results from the comparison between automatically and manually derived expert consensus LV parameters based on the SCMR consensus data are provided in the Additional file 1. In accordance with the study’s results, agreement was excellent in the majority of cases (ICC ≥ 0.95 for all LV parameters) (Additional file 2: Tables S1 and S2). There was one patient with a 20% difference in LV EF between automatic and manual expert results (Case # 15), which was a patient with severe LV hypertrophy (Additional file 2: Table S1, Figure S1), similar to the case shown in Fig. 1c.

Reproducibility

Reproducibility for manual segmentations was better for LV than for RV measurements. The automated algorithm yielded exactly the same results when being reapplied. Table 6 shows ICC, CoV and mean differences (SD) within and between observers.
Table 6
Reproducibility of manual and automated analyses
 
Parameter
Mean Difference (SD of the Diff.)
ICC (95% CI)
CoV (%)
Manual
Intra-observer
LV Mass
−0.3 (3.4)
0.99 (0.97–1.00)
5.3
LV EDV
−3.3 (3.4)
0.99 (0.91–1.00)
4.0
LV ESV
−2.6 (3.0)
0.99 (0.95–1.00)
8.4
LV SV
−0.6 (3.9)
0.97 (0.93–0.99)
8.0
LV EF
1.7 (3.7)
0.98 (0.94–0.99)
6.3
RV EDV
−1.5 (6.7)
0.96 (0.90–0.98)
9.6
RV ESV
1.8 (4.3)
0.95 (0.87–0.98)
14.2
RV SV
−3.5 (6.3)
0.81 (0.49–0.93)
15.6
RV EF
−2.9 (5.1)
0.82 (0.49–0.93)
8.9
Inter-observer
LV Mass
−0.7 (3.3)
0.99 (0.97–1.00)
5.6
LV EDV
−5.6 (5.8)
0.97 (0.74–0.99)
6.7
LV ESV
−6.3 (4.8)
0.97 (0.46–0.99)
12.5
LV SV
0.8 (3.7)
0.97 (0.93–0.99)
7.7
LV EF
5.2 (57.9)
0.94 (0.44–0.98)
8.0
RV EDV
−15.5 (7.8)
0.82 (0.00–0.96)
10.0
RV ESV
−7.3 (5.7)
0.84 (0.00–0.96)
16.4
RV SV
−8.1 (6.9)
0.75 (0.00–0.93)
16.2
RV EF
1.3 (7.2)
0.62 (0.04–0.85)
12.9
Automatic
Intra-observer
LV Mass
0.0 (0.0)
1.00
0.0
LV EDV
0.0 (0.0)
1.00
0.0
LV ESV
0.0 (0.0)
1.00
0.0
LV SV
0.0 (0.0)
1.00
0.0
LV EF
0.0 (0.0)
1.00
0.0
RV EDV
0.0 (0.0)
1.00
0.0
RV ESV
0.0 (0.0)
1.00
0.0
RV SV
0.0 (0.0)
1.00
0.0
RV EF
0.0 (0.0)
1.00
0.0
Inter-observer
LV Mass
0.0 (0.0)
1.00
0.0
LV EDV
0.0 (0.0)
1.00
0.0
LV ESV
0.0 (0.0)
1.00
0.0
LV SV
0.0 (0.0)
1.00
0.0
LV EF
0.0 (0.0)
1.00
0.0
RV EDV
0.0 (0.0)
1.00
0.0
RV ESV
0.0 (0.0)
1.00
0.0
RV SV
0.0 (0.0)
1.00
0.0
RV EF
0.0 (0.0)
1.00
0.0
Biventricular volumes and LV mass were indexed to body surface area. SD: standard deviation. ICC: intraclass correlation coefficient. CoV: coefficient of variation. LV: left ventricular. RV: right ventricular. EDV/ESV end-diastolic/systolic volume. SV: stroke volume. EF: ejection fraction

Discussion

The present study demonstrates the feasibility of fully automated quantification of biventricular morphology and function and reveals its current pitfalls and limitations in a ‘real-world’ clinical setting. Several notable findings should be considered. First, automatically and manually derived volumes agree well in case of good image quality; however, severe differences occur in case of reduced image quality. Second, agreement is better for LV than for RV volumes. Third, demanding anatomical circumstances (e.g. in patients with repaired ToF) result in lower agreement. Forth, different field strengths or the presence of valve replacements do not impede automated assessments as long as image quality is preserved.

Agreement of automated and manual assessment

CMR represents the reference standard for cardiac volume assessment [1] with incremental accuracy and reproducibility as compared to echocardiography [18]. However, CMR acquisition time is long and further requires time-consuming post-processing to extract clinically relevant information. Thus, efforts have been directed towards automated post-processing analyses based on deep-learning algorithms within the last decade [8, 19, 20]. The current literature demonstrates excellent agreement for automated and manual LV volume assessments [6, 10]; however, studies concerned with automatic RV segmentation are scarce [21]. Noteworthy, the study by Queirόs et al. [10] applied an automatic algorithm on cropped data, that is after manually defining the most basal and apical slices with subsequent cropping the SAX stack to include images effectively covering the LV before applying the automatic algorithm. Furthermore, ED and ES time points were manually pre-selected. However, the correct definition of the most basal slice is amongst the most challenging steps in SAX volume assessments and one of the most important source of observer variability [22], therefore representing a clear bias in testing the reliability of an automated algorithm. In the present study, we sought to simulate a real-world clinical scenario by randomly selecting patients from clinical routine imaging. We applied a commercially available automatic algorithm on clinically acquired SAX stacks – occasionally comprising both atria and ventricles – without any manual pre- or post-processing. The final data was acquired on 1.5 and 3.0 T scanners. Our data elaborates on the excellent agreement between automatically and manually derived volumes in case of good image quality, with overall better agreement for LV than for RV measurements. Indeed, quantification of RV volumes is generally more challenging as opposed to LV volumes due to the complex RV anatomy [23, 24]. Nevertheless, LV and RV stroke volumes were consistent in automated analyses in this patient group without intra- or extracardiac shunt.
Manual post-processing time took on average more than 11 min as compared to fully automated assessments with < 1 min. Importantly, automatic analyses of several CMR examinations (in this case 300 scans) run completely user-independent and were performed overnight. Furthermore, automated analyses promise to overcome limitations in observer variability, since the algorithm yields exactly the same measures when being reapplied by different users. Thus, the automated frame-work provides a highly reproducible approach and is able to extremely shorten post-processing times of CMR examinations with subsequent potential to improve cost-effectiveness [25]. Furthermore, the framework may provide ‘on-the-fly’ post-processing parallel to finishing the CMR scan (e.g. during late gadolinium enhancement acquisitions).

Impact of image quality

Our data demonstrate that image quality is the leading determinant of accuracy for fully automatic volume assessment. In case of good image quality (image quality score of ≤1 adopted to the criteria proposed by Klinke et al. [13], Table 1), the bias of both LV and RV function was within acceptable limits. However, in case of reduced image quality (image quality score ≥ 2), a large bias of > 5% was observed for both LV and RV EF with wide LOA, particularly for RV EF. Importantly, the relevance of RV function and volumes is increasingly recognized in various diseases [26]. For example, the diagnosis of arrhythmogenic right ventricular cardiomyopathy is challenging and heavily relies on the assessment of RV EDV and RV EF [27]. If considered for clinical use and decision making, a precise volume assessment is of utmost importance, and cannot be achieved with the proposed fully automatic algorithm in case of impaired image quality yet.

Technical and anatomical considerations

To further elucidate limitations of the commercially available software, we compared the agreement of automated and manually derived volumes for subgroups according to field strengths, the presence of aortic valve replacements as well as repaired ToF. Agreement was better at 1.5 T compared to 3 T scans; however, at 3 T considerably more artefacts (mainly due to inadequate breath-holding and shimming) were present. Reduced agreement at 3 T is therefore more likely a result of lower image quality. Due to the growing number of percutaneously implanted aortic valves [28] and increasing indications for CMR imaging [29] including aortic valve stenosis [30], the presence of valve replacement in CMR studies is likely to grow. As long as image quality was preserved in these patients, agreement of LV volumes remained acceptable, enabling the use of automated algorithms in this group of patients. In contrast, in patients with repaired ToF, both RV and LV agreement were considerably decreased, despite low image quality solely affecting the RV (metal artefacts resulting from sternal wires). Since LV image quality was good, reduced agreement is most likely due to the more demanding anatomy in these patients (distinctly larger RV than LV volumes), which points out the current limitations of fully automated analysis. Here, it remains to be investigated whether or not the proposed automatic deep-learning frame-work is able to further learn from these cases with subsequent improvement of accuracy.

Limitations

The study’s conclusions are derived from the comparison of 300 automatically and manually quantified clinical CMR examinations from a single centre. Although manual contouring was performed by experienced observers, intra- and inter-observer variability may limit its use as a reference standard. Details of the automatic algorithm are not disclosed by the software vendor and therefore cannot be reported. RV mass was not measured, since the automatic algorithm does not provide RV mass quantification.

Conclusion

Fully automated quantification of biventricular morphology and function is feasible and accurate in the majority of cases in a clinical routine setting and has the potential to extremely accelerate post-processing times and to improve reproducibility. However, in case of limited image quality or in patients with demanding anatomy (e.g. in patients with repaired ToF) the proposed fully-automatic frame-work does not yet provide satisfying results and still requires manual re-contouring.

Acknowledgements

We thank Marcus Billing and Nick Scholand for the support in IT applications. We also thank Professor Alistair A. Young and Dr. Avan Suinesiaputra for providing access to the SCMR consensus data.

Funding

None.

Availability of data and materials

Regarding data availability, we confirm that all relevant data are within the paper and all data underlying the findings are fully available without restriction and can be accessed at the University Medical Centre Goettingen by researchers who meet the criteria for access to confidential data. Anonymized DICOM data of this dataset are available from the corresponding author on reasonable request to test other software packages.
The study was approved by the Ethics Committee of the University Hospital Goettingen and complied with the Declaration of Helsinki. The Ethics Committee gave permission to waive informed consent for this retrospective analysis.
Not applicable.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
1.
Zurück zum Zitat Pennell DJ. Cardiovascular magnetic resonance. Circulation. 2010;121:692–705.CrossRef Pennell DJ. Cardiovascular magnetic resonance. Circulation. 2010;121:692–705.CrossRef
2.
Zurück zum Zitat Miller CA, Pearce K, Jordan P, Argyle R, Clark D, Stout M, et al. Comparison of real-time three-dimensional echocardiography with cardiovascular magnetic resonance for left ventricular volumetric assessment in unselected patients. Eur Heart J Cardiovasc Imaging. 2012;13:187–95.CrossRef Miller CA, Pearce K, Jordan P, Argyle R, Clark D, Stout M, et al. Comparison of real-time three-dimensional echocardiography with cardiovascular magnetic resonance for left ventricular volumetric assessment in unselected patients. Eur Heart J Cardiovasc Imaging. 2012;13:187–95.CrossRef
3.
Zurück zum Zitat White HD, Norris RM, Brown MA, Brandt PW, Whitlock RM, Wild CJ. Left ventricular end-systolic volume as the major determinant of survival after recovery from myocardial infarction. Circulation. 1987;76:44–51.CrossRef White HD, Norris RM, Brown MA, Brandt PW, Whitlock RM, Wild CJ. Left ventricular end-systolic volume as the major determinant of survival after recovery from myocardial infarction. Circulation. 1987;76:44–51.CrossRef
4.
Zurück zum Zitat Eitel I, de Waha S, Wöhrle J, Fuernau G, Lurz P, Pauschinger M, et al. Comprehensive prognosis assessment by CMR imaging after ST-segment elevation myocardial infarction. J Am Coll Cardiol. 2014;64:1217–26.CrossRef Eitel I, de Waha S, Wöhrle J, Fuernau G, Lurz P, Pauschinger M, et al. Comprehensive prognosis assessment by CMR imaging after ST-segment elevation myocardial infarction. J Am Coll Cardiol. 2014;64:1217–26.CrossRef
5.
Zurück zum Zitat Petitjean C, Dacher J-N. A review of segmentation methods in short axis cardiac MR images. Med Image Anal. 2011;15:169–84.CrossRef Petitjean C, Dacher J-N. A review of segmentation methods in short axis cardiac MR images. Med Image Anal. 2011;15:169–84.CrossRef
6.
Zurück zum Zitat Hautvast GLTF, Salton CJ, Chuang ML, Breeuwer M, O'Donnell CJ, Manning WJ. Accurate computer-aided quantification of left ventricular parameters: experience in 1555 cardiac magnetic resonance studies from the Framingham heart study. Magn Reson Med. 2012;67:1478–86.CrossRef Hautvast GLTF, Salton CJ, Chuang ML, Breeuwer M, O'Donnell CJ, Manning WJ. Accurate computer-aided quantification of left ventricular parameters: experience in 1555 cardiac magnetic resonance studies from the Framingham heart study. Magn Reson Med. 2012;67:1478–86.CrossRef
7.
Zurück zum Zitat Axel L, Sodickson DK. The need for speed: accelerating CMR imaging assessment of cardiac function. JACC Cardiovasc Imaging. 2014;7:893–5.CrossRef Axel L, Sodickson DK. The need for speed: accelerating CMR imaging assessment of cardiac function. JACC Cardiovasc Imaging. 2014;7:893–5.CrossRef
8.
Zurück zum Zitat Bernard O, Lalande A, Zotti C, Cervenansky F, Yang X, Heng P-A, et al. Deep Learning Techniques for Automatic MRI Cardiac Multi-Structures Segmentation and Diagnosis: Is the Problem Solved? IEEE Trans Med Imaging. 2018;37:2514–25. Bernard O, Lalande A, Zotti C, Cervenansky F, Yang X, Heng P-A, et al. Deep Learning Techniques for Automatic MRI Cardiac Multi-Structures Segmentation and Diagnosis: Is the Problem Solved? IEEE Trans Med Imaging. 2018;37:2514–25.
9.
Zurück zum Zitat van Geuns RJM, Baks T, Gronenschild EHBM, Aben J-PMM, Wielopolski PA, Cademartiri F, de Feyter PJ. Automatic quantitative left ventricular analysis of cine MR images by using three-dimensional information for contour detection. Radiology. 2006;240:215–21.CrossRef van Geuns RJM, Baks T, Gronenschild EHBM, Aben J-PMM, Wielopolski PA, Cademartiri F, de Feyter PJ. Automatic quantitative left ventricular analysis of cine MR images by using three-dimensional information for contour detection. Radiology. 2006;240:215–21.CrossRef
10.
Zurück zum Zitat Queirós S, Barbosa D, Engvall J, Ebbers T, Nagel E, Sarvari SI, et al. Multi-Centre validation of an automatic algorithm for fast 4D myocardial segmentation in cine CMR datasets. Eur Heart J Cardiovasc Imaging. 2016;17:1118–27.CrossRef Queirós S, Barbosa D, Engvall J, Ebbers T, Nagel E, Sarvari SI, et al. Multi-Centre validation of an automatic algorithm for fast 4D myocardial segmentation in cine CMR datasets. Eur Heart J Cardiovasc Imaging. 2016;17:1118–27.CrossRef
11.
Zurück zum Zitat Schulz-Menger J, Bluemke DA, Bremerich J, Flamm SD, Fogel MA, Friedrich MG, et al. Standardized image interpretation and post processing in cardiovascular magnetic resonance: Society for Cardiovascular Magnetic Resonance (SCMR) board of trustees task force on standardized post processing. J Cardiovasc Magn Reson. 2013;15:35.CrossRef Schulz-Menger J, Bluemke DA, Bremerich J, Flamm SD, Fogel MA, Friedrich MG, et al. Standardized image interpretation and post processing in cardiovascular magnetic resonance: Society for Cardiovascular Magnetic Resonance (SCMR) board of trustees task force on standardized post processing. J Cardiovasc Magn Reson. 2013;15:35.CrossRef
12.
Zurück zum Zitat Suinesiaputra A, Bluemke DA, Cowan BR, Friedrich MG, Kramer CM, Kwong R, et al. Quantification of LV function and mass by cardiovascular magnetic resonance: multi-center variability and consensus contours. J Cardiovasc Magn Reson. 2015;17:63.CrossRef Suinesiaputra A, Bluemke DA, Cowan BR, Friedrich MG, Kramer CM, Kwong R, et al. Quantification of LV function and mass by cardiovascular magnetic resonance: multi-center variability and consensus contours. J Cardiovasc Magn Reson. 2015;17:63.CrossRef
13.
Zurück zum Zitat Klinke V, Muzzarelli S, Lauriers N, Locca D, Vincenti G, Monney P, et al. Quality assessment of cardiovascular magnetic resonance in the setting of the European CMR registry: description and validation of standardized criteria. J Cardiovasc Magn Reson. 2013;15:55.CrossRef Klinke V, Muzzarelli S, Lauriers N, Locca D, Vincenti G, Monney P, et al. Quality assessment of cardiovascular magnetic resonance in the setting of the European CMR registry: description and validation of standardized criteria. J Cardiovasc Magn Reson. 2013;15:55.CrossRef
14.
Zurück zum Zitat Bland M, Altman D. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;327:307–10.CrossRef Bland M, Altman D. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;327:307–10.CrossRef
15.
Zurück zum Zitat Kowallick JT, Morton G, Lamata P, Jogiya R, Kutty S, Lotz J, et al. Inter-study reproducibility of left ventricular torsion and torsion rate quantification using MR myocardial feature tracking. J Magn Reson Imaging. 2016;43:128–37.CrossRef Kowallick JT, Morton G, Lamata P, Jogiya R, Kutty S, Lotz J, et al. Inter-study reproducibility of left ventricular torsion and torsion rate quantification using MR myocardial feature tracking. J Magn Reson Imaging. 2016;43:128–37.CrossRef
16.
Zurück zum Zitat Morton G, Schuster A, Jogiya R, Kutty S, Beerbaum P, Nagel E. Inter-study reproducibility of cardiovascular magnetic resonance myocardial feature tracking. J Cardiovasc Magn Reson. 2012;14:43.CrossRef Morton G, Schuster A, Jogiya R, Kutty S, Beerbaum P, Nagel E. Inter-study reproducibility of cardiovascular magnetic resonance myocardial feature tracking. J Cardiovasc Magn Reson. 2012;14:43.CrossRef
17.
Zurück zum Zitat Gertz RJ, Lange T, Kowallick JT, Backhaus SJ, Steinmetz M, Staab W, et al. Inter-vendor reproducibility of left and right ventricular cardiovascular magnetic resonance myocardial feature-tracking. PLoS One. 2018;13:e0193746.CrossRef Gertz RJ, Lange T, Kowallick JT, Backhaus SJ, Steinmetz M, Staab W, et al. Inter-vendor reproducibility of left and right ventricular cardiovascular magnetic resonance myocardial feature-tracking. PLoS One. 2018;13:e0193746.CrossRef
18.
Zurück zum Zitat Grothues F, Smith GC, Moon JC, Bellenger NG, Collins P, Klein HU, Pennell DJ. Comparison of interstudy reproducibility of cardiovascular magnetic resonance with two-dimensional echocardiography in normal subjects and in patients with heart failure or left ventricular hypertrophy. Am J Cardiol. 2002;90:29–34.CrossRef Grothues F, Smith GC, Moon JC, Bellenger NG, Collins P, Klein HU, Pennell DJ. Comparison of interstudy reproducibility of cardiovascular magnetic resonance with two-dimensional echocardiography in normal subjects and in patients with heart failure or left ventricular hypertrophy. Am J Cardiol. 2002;90:29–34.CrossRef
19.
Zurück zum Zitat Shahzad R, Tao Q, Dzyubachyk O, Staring M, Lelieveldt BPF, van der Geest RJ. Fully-automatic left ventricular segmentation from long-axis cardiac cine MR scans. Med Image Anal. 2017;39:44–55.CrossRef Shahzad R, Tao Q, Dzyubachyk O, Staring M, Lelieveldt BPF, van der Geest RJ. Fully-automatic left ventricular segmentation from long-axis cardiac cine MR scans. Med Image Anal. 2017;39:44–55.CrossRef
20.
Zurück zum Zitat Avendi MR, Kheradvar A, Jafarkhani H. Fully automatic segmentation of heart chambers in cardiac MRI using deep learning. J Cardiovasc Magn Reson. 2016;18:351.CrossRef Avendi MR, Kheradvar A, Jafarkhani H. Fully automatic segmentation of heart chambers in cardiac MRI using deep learning. J Cardiovasc Magn Reson. 2016;18:351.CrossRef
21.
Zurück zum Zitat Avendi MR, Kheradvar A, Jafarkhani H. Automatic segmentation of the right ventricle from cardiac MRI using a learning-based approach. Magn Reson Med. 2017;78:2439–48.CrossRef Avendi MR, Kheradvar A, Jafarkhani H. Automatic segmentation of the right ventricle from cardiac MRI using a learning-based approach. Magn Reson Med. 2017;78:2439–48.CrossRef
22.
Zurück zum Zitat Simprini LA, Goyal P, Codella N, Fieno DS, Afroz A, Mullally J, et al. Geometry-independent inclusion of basal myocardium yields improved cardiac magnetic resonance agreement with echocardiography and necropsy quantified left-ventricular mass. J Hypertens. 2013;31:2069–76.CrossRef Simprini LA, Goyal P, Codella N, Fieno DS, Afroz A, Mullally J, et al. Geometry-independent inclusion of basal myocardium yields improved cardiac magnetic resonance agreement with echocardiography and necropsy quantified left-ventricular mass. J Hypertens. 2013;31:2069–76.CrossRef
23.
Zurück zum Zitat Beerbaum P, Barth P, Kropf S, Sarikouch S, Kelter-Kloepping A, Franke D, et al. Cardiac function by MRI in congenital heart disease: impact of consensus training on interinstitutional variance. J Magn Reson Imaging. 2009;30:956–66.CrossRef Beerbaum P, Barth P, Kropf S, Sarikouch S, Kelter-Kloepping A, Franke D, et al. Cardiac function by MRI in congenital heart disease: impact of consensus training on interinstitutional variance. J Magn Reson Imaging. 2009;30:956–66.CrossRef
24.
Zurück zum Zitat Hudsmith L, Petersen S, Francis J, Robson M, Neubauer S. Normal Human Left and Right Ventricular and Left Atrial Dimensions Using Steady State Free Precession Magnetic Resonance Imaging. J Cardiovasc Magn Reson. 2005;7:775–82.CrossRef Hudsmith L, Petersen S, Francis J, Robson M, Neubauer S. Normal Human Left and Right Ventricular and Left Atrial Dimensions Using Steady State Free Precession Magnetic Resonance Imaging. J Cardiovasc Magn Reson. 2005;7:775–82.CrossRef
25.
Zurück zum Zitat Hegde VA, Biederman RW, Mikolich JR. Cardiovascular magnetic resonance imaging-incremental value in a series of 361 patients demonstrating cost savings and clinical benefits: an outcome-based study. Clin Med Insights Cardiol. 2017;11:1179546817710026.CrossRef Hegde VA, Biederman RW, Mikolich JR. Cardiovascular magnetic resonance imaging-incremental value in a series of 361 patients demonstrating cost savings and clinical benefits: an outcome-based study. Clin Med Insights Cardiol. 2017;11:1179546817710026.CrossRef
26.
Zurück zum Zitat Tadic M. Multimodality evaluation of the right ventricle: an updated review. Clin Cardiol. 2015;38:770–6.CrossRef Tadic M. Multimodality evaluation of the right ventricle: an updated review. Clin Cardiol. 2015;38:770–6.CrossRef
27.
Zurück zum Zitat te Riele ASJM, Tandri H, Bluemke DA. Arrhythmogenic right ventricular cardiomyopathy (ARVC): cardiovascular magnetic resonance update. J Cardiovasc Magn Reson. 2014;16:50.CrossRef te Riele ASJM, Tandri H, Bluemke DA. Arrhythmogenic right ventricular cardiomyopathy (ARVC): cardiovascular magnetic resonance update. J Cardiovasc Magn Reson. 2014;16:50.CrossRef
28.
Zurück zum Zitat Haussig S, Schuler G, Linke A. Worldwide TAVI registries: what have we learned? Clin Res Cardiol. 2014;103:603–12.CrossRef Haussig S, Schuler G, Linke A. Worldwide TAVI registries: what have we learned? Clin Res Cardiol. 2014;103:603–12.CrossRef
29.
Zurück zum Zitat von Knobelsdorff-Brenkenhoff F, Schulz-Menger J. Role of cardiovascular magnetic resonance in the guidelines of the European Society of Cardiology. J Cardiovasc Magn Reson. 2016;18:6.CrossRef von Knobelsdorff-Brenkenhoff F, Schulz-Menger J. Role of cardiovascular magnetic resonance in the guidelines of the European Society of Cardiology. J Cardiovasc Magn Reson. 2016;18:6.CrossRef
30.
Zurück zum Zitat Musa TA, Plein S, Greenwood JP. The role of cardiovascular magnetic resonance in the assessment of severe aortic stenosis and in post-procedural evaluation following transcatheter aortic valve implantation and surgical aortic valve replacement. Quant Imaging Med Surg. 2016;6:259–73.CrossRef Musa TA, Plein S, Greenwood JP. The role of cardiovascular magnetic resonance in the assessment of severe aortic stenosis and in post-procedural evaluation following transcatheter aortic valve implantation and surgical aortic valve replacement. Quant Imaging Med Surg. 2016;6:259–73.CrossRef
Metadaten
Titel
Fully automated quantification of biventricular volumes and function in cardiovascular magnetic resonance: applicability to clinical routine settings
verfasst von
Sören J. Backhaus
Wieland Staab
Michael Steinmetz
Christian O. Ritter
Joachim Lotz
Gerd Hasenfuß
Andreas Schuster
Johannes T. Kowallick
Publikationsdatum
01.12.2019
Verlag
BioMed Central
Erschienen in
Journal of Cardiovascular Magnetic Resonance / Ausgabe 1/2019
Elektronische ISSN: 1532-429X
DOI
https://doi.org/10.1186/s12968-019-0532-9

Weitere Artikel der Ausgabe 1/2019

Journal of Cardiovascular Magnetic Resonance 1/2019 Zur Ausgabe

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.