How to Measure/Calculate Radiation Dose in Patients?

verfasst von: Reinhard Loose, Michael Wucherer

Erschienen in: CardioVascular and Interventional Radiology | Ausgabe 6/2021

Abstract

Patients in fluoroscopically guided interventions (FGI) may be exposed to substantial radiation dose levels (SRDL). The most commonly reported adverse reactions are skin injuries with erythema or necrosis. It is therefore important for the interventional radiologist to know deterministic effects with their threshold doses. If possible all relevant modality parameters should be displayed on the interventionalists screen. Dosimetric parameters should be displayed in digital imaging and communications in medicine (DICOM) units and stored as DICOM Radiation Dose Structured Report (RDSR). The peak skin dose (PSD) is the most relevant risk parameter for skin injuries. Dose management systems (DMS) help optimising radiation exposure of patients. However, their calculation of skin dose maps is only available after a FGI. Therefore, dose maps and PSD should preferably be calculated and displayed in real time by the modality.
Hinweise

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
CT
Computed tomography
DICOM
Digital Imaging and Communications in Medicine
DMS
Dose management system
DRL
Diagnostic reference level
ED
Effective dose
ESD
Entrance surface dose
EU-BSS
European Basic Safety Standards
FGI
Fluoroscopically Guided Intervention
FOV
Field of View
HIS
Hospital Information System
ICRP
International Commission on Radiological Protection
IEC
International Electrotechnical Commission,
KAP
Kerma-area product
Ka,r
Air kerma at the patient entrance reference point
MPE
Medical physics expert
NCRP
National Council on Radiation Protection and Measurements
OSL
Optically stimulated luminescence dosimeter
PACS
Picture archiving and communication system
PSD
Peak Skin Dose
RDSR
Radiation Dose Structured Report
RF
Radio Fluoroscopy
RIS
Radiology Information System
SRDL
Substantial Radiation Dose Level
TLD
Thermoluminescent Dosimeter
XA
X-ray angiography

Introduction

Patient exposure in fluoroscopically guided interventions (FGI) spans a wide dose range and can reach levels at which deterministic effects may occur. The reasons for this include the variety of modalities used and the wide range of interventional procedures. It is therefore important in complex interventions with substantial radiation dose levels (SRDL) to be able to measure or estimate the patient dose. This dose information should not only be available after a procedure by means of the recorded exposure parameters, but also to the interventionalist during a FGI.
The following article essentially refers to stationary angiography systems on which complex interventions e.g. in radiology, neuroradiology, cardiology or vascular surgery can be performed. Patient exposure in FGI depends on many patient, procedure and modality related parameters like kV, mAs, filtration, detector entrance dose rate, pulse rate, number of images, image processing, fluoroscopy time, geometric properties of the modality and field of view (FOV) [1, 2]. The DICOM radiation dose structured report (RDSR) is available in most new angiographic systems and enables a more detailed analysis of all exposure parameters from fluoroscopy and radiographic images or cine series [3].
For all aspects of quality assurance and dose management, the involvement of a medical physics expert (MPE) is therefore imperative in accordance with the European Directive 2013/59/EURATOM (EU-BSS) [4], especially for procedures with a higher dose such as computed tomography (CT) or FGI. Furthermore, FGI are the procedures with the highest risk of deterministic effects. If appropriate, these are to be reported to the competent authorities as "unintended exposures" in accordance with the national implementation of the EU-BSS.

How to Analyse DICOM Dose Reports in FGI

About 20 years ago, exposure parameters from FGI were usually recorded manually in a Radiology or Hospital Information System (RIS/HIS) or paper based. Later, storage was provided together with the angiographic images in a picture archiving and communication system (PACS) as bitmap report. All of these recording methods allow only difficult analysis of the patient exposure. Today DICOM RDSR is available in most new angiography systems and provides an easy solution to collect dose parameters. This includes all exposure parameters for each fluoroscopic scene, all radiographic images or cine series with kV, mAs, geometrical parameters of C-arm, detector and more. Table 1 shows an excerpt of an angiography RDSR. In radiography and CT, the exposure data can be extracted relatively reliable from the DICOM image data even without RDSR. This is not the case with fluoroscopic procedures, as fluoroscopy scenes are usually not stored in the PACS. The DICOM image data therefore lack the dose contribution from fluoroscopy, which can easily exceed 50% of the total dose depending on the type of intervention. Recording and processing of patient exposure was driven by the EU-BSS which requires member states of the European Union to ensure justification and optimisation of radiological procedures and store information on patient exposure for analysis and quality assurance [4, 5]. Various commercial dose management systems (DMSs) with varying characteristics are available today [6]. In contrast to radiography and CT, complex RDSR reports are not always correctly and completely saved as DICOM objects in FGI and are not always correctly and completely evaluated by DMS providers. This is particularly important because all contributions from fluoroscopy and radiography / cine series are required to determine the total exposure of a patient. Furthermore, a complete recording of all individual radiation events is required to calculate the dose distribution on the patient's surface and to identify locations where overlapping radiation fields can lead to a high peak skin dose (PSD) and thus to potential deterministic skin injuries.
Table 1
Excerpt from a DICOM RDSR with relevant modality and exposure parameters. Depending on the type of procedure a RDSR can consists of many kilobytes of exposure events
https://static-content.springer.com/image/art%3A10.1007%2Fs00270-021-02772-x/MediaObjects/270_2021_2772_Tab1_HTML.png
The most commonly used exposure parameters are Kerma-area product (KAP) and Air kerma at the patient entrance reference point (Ka,r). KAP is used for diagnostic reference levels (DRLs) in most countries and is also displayed or transmitted by the manufacturers of all angiography systems. KAP is a public tag in both the Radio Fluoroscopy (RF) and X-ray angiography (XA) DICOM Service Class objects. The second most important parameter is Ka,r, which correlates more than KAP with the skin dose, followed by the total fluoroscopy time in the number of cine series or images. KAP and Ka,r are usually transmitted cumulatively in RDSR for an entire examination, while the dose distribution on the body surface with PSD has to be calculated from all individual exposure events. Since dosimetric data are usually transferred to a PACS after an examination has been completed, the dose distribution is only available in a DMS after the procedure. An online display of the skin dose distribution on the modality screen during the intervention would be desirable in order to avoid high PSD by changing the projection direction and hence the skin entrance field from time to time.

Parameters of Patient Exposure

Patient exposure to radiation can be measured directly or indirectly [7]. Direct dosimetry requires the use of dosimeters. Real time measurements are performed with ionisation chambers, diodes, metal oxide semiconductor field-effect transistors (MOSFET) and other devices. Non-real time measurements use thermoluminescent dosimeters (TLD) or optically stimulated luminescence dosimeters (OSL), other devices or film dosimetry in earlier times. Currently, most of the modern interventional systems use an ionization transmission chamber to measure the KAP and estimate Ka,r. Some manufacturers calculate these dosimetric quantities from the modality parameters (e.g. tube output).
Direct dosimetry is time-consuming and therefore mostly limited to measurements on phantoms or patients in clinical studies. In clinical routine, indirect dosimetry with dose parameters derived from the modalities are used because of their easy availability [1].
Electronic real time dosimeters are also used for occupational dosimetry of staff members. It must be noted that these dosimeters are suitable for measuring scattered radiation, but not radiation in the primary beam, since in this case they will display incorrect dose data.
Figure 1 shows a simple illustration of a monoplane angiography system with the relevant dosimetric parameters.

Kerma-Area-Product (KAP)

KAP is the integral of air kerma across the entire x-ray beam emitted from the x- ray tube (also called Dose Area Product DAP) [8]. KAP is a surrogate measurement for the entire amount of energy delivered to the patient by the beam. KAP is measured in [Gy*cm2] and does not include scatter.
KAP is the most important and most frequently used exposure parameter in radiographic and fluoroscopic examinations. It is available on most modalities and display and recording is required by law in many countries. Most of the DRLs in radiography and fluoroscopy are based on KAP. However, KAP does not allow an estimate about the risk of deterministic injuries to the skin or organs, as KAP is a product of dose and area and therefore large areas with a low dose or small areas with a high dose can provide identical values of KAP. For measurement of KAP an ionization chamber is placed beyond the X-ray collimators and must intercept the entire radiation field for accurate dosimetric results. Instead of measurement with ionization chambers, some manufacturers calculate KAP from modality parameters, e.g. tube output. With this procedure, it must be taken into account that a missing or incorrect calibration leads to incorrect KAP values.

Dose at the Patient Entrance Reference Point (Ka,r)

Ka,r is a dose accumulated at a specific point in space relative to the fluoroscopic gantry during a procedure. The International Electrotechnical Commission (IEC) defines a patient entrance reference point as 15 cm from the isocentre of a C-arm x-ray unit on the central beam towards the focus (Fig. 1) [9]. Ka,r is measured in Gy. Ka,r is sometimes referred to as reference dose, cumulative dose, or cumulative air kerma. Ka,r is only a rough estimation of skin dose and not the equivalent to the skin dose. The patient entrance reference point may correspond to the skin level, a point within the patient, or a point outside of the patient. The patient entrance reference point is thus more a technical characteristic of the equipment than operational information of the exposure of a real patient. In addition, Ka,r does not include beam repositioning, backscatter, or the attenuation of the table. Since mid-2006, the U.S. Food and Drug Administration requires that all fluoroscopes sold in the U.S. be capable of displaying the total air kerma at the IRP.

Entrance Surface Dose (ESD)

The entrance surface dose or entrance skin dose (ESD) is the measure of the radiation dose [mGy] that is absorbed by the skin of a patient. Entrance skin dose includes backscatter, should include (or estimate) the attenuation of the table [10] and is either directly measurable using dosimeters or can be calculated using the modality related dose parameters [11, 12].
However ESD is a poor indicator of radiation risk as it does not account for tissue sensitivity, penetration and exposed field size. Since ESD does not include the exposed field size, ESD is a better surrogate parameter for estimating the risk of deterministic skin reactions, provided the position of the radiation field on the skin does not change.

Peak Skin Dose

PSD is defined as the highest dose at any portion of a patient’s skin during a procedure. PSD includes contributions from both the primary X-ray beam and from scattered radiation and is measured in Gy. The level and distribution of the skin dose in FGI can either be measured or calculated. In the past, measurements were often carried out with radiochromic films, the density of which is a measure of the dose and the distribution of the radiation entry fields. Figure 2 shows the dose distribution with radiochromic film of a cardiological intervention [2]. Since the introduction of the DICOM RDSR in angiography, PSD can also be calculated with certain errors. In addition to KAP and / or Ka,r all geometric parameters of the respective position of the C-arm, table and collimation are required in the RDSR for all radiation events (fluoroscopy and image / cine series) [13]. Figure 3 shows the relevant geometrical parameters, KAP and Ka,r at the modality screen. All changes in the position of the table or C-arm are recorded in individual RDSR objects together with the respective collimation and the current values of KAP and Ka,r. From these data, a dose distribution can be calculated, represented graphically and the PSD determined. Figure 4 shows a coloured dose map during a cardiological procedure [14]. Since all geometrical data are based on parameters of the modality, large errors or unusable results arise when the patient's position relative to the table changes. In the meantime, DMS from several manufacturers can calculate dose maps and PSD and display them graphically. This always happens after the end of an intervention. It would therefore be an advantage if dose maps and PSD were calculated online by the device manufacturer and displayed live on a modality screen in the intervention room.

Organ Dose

The organ dose of tissue located in the radiation field can be estimated via the ESD. This requires depth-dose curves in tissue for the respective radiation quality. The attenuation in a patient by means of half-value thicknesses of the tissue can also be used as a rough estimate. As a first approximation, the half-value thickness of 3 cm in tissue (water) can be assumed for typical radiation spectra in angiography systems. The organ dose also depends on the extent to which the organ has been fully or partial exposed and, accordingly, on how the dose is distributed in the organ. Organ doses can also be determined using Monte Carlo software or conversion factors from KAP. The latter can also be used to estimate the effective dose (ED) [15].
The dose estimation is subject to many uncertainties (metrological, patient-specific). The uncertainty for the KAP is in a range of 25% that of the ESD calculated from KAP in the range of 50%. This means that uncertainties of over 50% must be expected for individual organ dose estimates. Backscatter factors depend strongly on the field size and the filtration and tube voltage and are in case of FGI around 1.4 (1.3–1.5) for water [10].

Effective Dose

Different tissues have varying degrees of radiosensitivity. For example, breast, bone marrow and colon are much more radiosensitive than bone surface, brain and skin. To account for this, tissue weighing factors have been developed [16, 17]. Mathematically, ED is the sum over irradiated tissues of the product of the equivalent dose and the tissue weighting factor for those tissues [18]. The unit of ED is the Sievert (Sv). It is important to note that tissue weighting factors are based on population age- and sex-specific averages, which contribute significantly to differences in individual risk. The chain below shows the relationship between absorbed dose, equivalent dose and effective dose [18].
It should be noted that ED has been introduced to assess the risk of occupational exposure. The ICRP does not recommend using ED for risk calculations of individual patients. Due to the uncertainty of all parameters involved in the chain of an ED calculation, the error of the ED estimate for individual patients is significantly above 50%.

Summary

The risk of skin radiation injuries, rarely of organ injuries, has long been known in interventional radiology. However, calculating or measuring patient doses and their distribution in organs or on the surface of the body is not an easy task. In the past, thresholds were suggested for easily available parameters such as KAP, Ka,r or fluoroscopy time above which deterministic reactions can be expected [7]. Once the threshold dose is exceeded, the injury becomes progressively more severe with increasing dose, although the true severity of major injuries will only become apparent weeks to months after the procedure. These thresholds are also called SRDL. Stecker et al. suggested parameters for a first SRDL notification [8]: PSD 2 Gy, Ka,r 3 Gy, KAP 300 Gy*cm2, fluoroscopy time 30 min.
The National Council on Radiation Protection and Measurements (NCRP) suggests a SRDL be defined at a PSD of 3 Gy or a Ka,r of 5 Gy [19].
In summary:
  • Interventionalist should know SRDL levels with high risk of skin injuries [20]
  • KAP, Ka,r, fluoroscopy time and if possible dose maps should be displayed online on the interventionalists screen
  • Dose maps and PSD should preferably be displayed in real time and not only in DMS
  • The radiation entrance field should be changed from time to time in complex interventions to avoid overlapping fields with high PSD
  • DMS is a helpful tool to analyse and optimise radiation protection of patients
  • All measured or calculated dose parameters should be stored for analysis

Compliance with Ethical Standards

Conflict of interest

none.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Innere Medizin

Kombi-Abonnement

Mit e.Med Innere Medizin erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Innere Medizin, den Premium-Inhalten der internistischen Fachzeitschriften, inklusive einer gedruckten internistischen Zeitschrift Ihrer Wahl.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Literatur
2.
Zurück zum Zitat ICRP Publication 120 (2013) International Commission on Radiological Protection. Radiological protection in cardiology. Ann ICRP. 42 (1): 1–125 ICRP Publication 120 (2013) International Commission on Radiological Protection. Radiological protection in cardiology. Ann ICRP. 42 (1): 1–125
4.
Zurück zum Zitat Council Directive 2013/59/Euratom of 5 December 2013 laying down basic safety standards for protection against the dangers arising from exposure to ionising radiation and repealing Directives 89/618/Euratom, 90/641/Euratom, 96/29/Euratom, 97/43/Euratom and 2003/122/Euratom. http://data.europa.eu/eli/dir/2013/59/oj (last access 06 January 2021) Council Directive 2013/59/Euratom of 5 December 2013 laying down basic safety standards for protection against the dangers arising from exposure to ionising radiation and repealing Directives 89/618/Euratom, 90/641/Euratom, 96/29/Euratom, 97/43/Euratom and 2003/122/Euratom. http://​data.​europa.​eu/​eli/​dir/​2013/​59/​oj (last access 06 January 2021)
5.
Zurück zum Zitat European Society of Radiology (ESR) (2015) Summary of the European Directive 2013/59/Euratom: Essentials for Health Professionals in Radiology. Insights Imaging. 6: 411–417 European Society of Radiology (ESR) (2015) Summary of the European Directive 2013/59/Euratom: Essentials for Health Professionals in Radiology. Insights Imaging. 6: 411–417
8.
Zurück zum Zitat Stecker M, Balter S, Towbin R, et al. Guidelines for Patient Radiation Dose Management. J Vasc Interv Radiol. 2009;20:263–73.CrossRef Stecker M, Balter S, Towbin R, et al. Guidelines for Patient Radiation Dose Management. J Vasc Interv Radiol. 2009;20:263–73.CrossRef
9.
Zurück zum Zitat International Electrotechnical Commission. Report 60601-2-43:2010 Medical electrical equipment – part 2–43: Particular requirements for the safety and essential performance of X-ray equipment for interventional procedures. Geneva, Switzerland: IEC; 2010. International Electrotechnical Commission. Report 60601-2-43:2010 Medical electrical equipment – part 2–43: Particular requirements for the safety and essential performance of X-ray equipment for interventional procedures. Geneva, Switzerland: IEC; 2010.
10.
Zurück zum Zitat Petoussi-Henss N, Zankl M, Drexler G, et al. Calculation of backscatter factors for diagnostic radiology using Monte Carlo methods. Phys Med Biol. 1998;43:2237–50.CrossRef Petoussi-Henss N, Zankl M, Drexler G, et al. Calculation of backscatter factors for diagnostic radiology using Monte Carlo methods. Phys Med Biol. 1998;43:2237–50.CrossRef
11.
Zurück zum Zitat Tsapaki V, Tsalafoutas IA, Chinofoti I, et al. Radiation doses to patients undergoing standard radiographic examinations: a comparison between two methods. The British Journal of Radiology. 2007;80:107–12.CrossRef Tsapaki V, Tsalafoutas IA, Chinofoti I, et al. Radiation doses to patients undergoing standard radiographic examinations: a comparison between two methods. The British Journal of Radiology. 2007;80:107–12.CrossRef
12.
Zurück zum Zitat Sharma R, Sharma SD, Pawar S, et al. Radiation dose to patients from X-ray radiographic examinations using computed radiography imaging system. J Med Phys. 2015;40(1):29–37.CrossRef Sharma R, Sharma SD, Pawar S, et al. Radiation dose to patients from X-ray radiographic examinations using computed radiography imaging system. J Med Phys. 2015;40(1):29–37.CrossRef
13.
Zurück zum Zitat Jones AK, Pasciak AS (2011) Calculating the peak skin dose resulting from fluoroscopically guided interventions. Part I: Methods. Journal of applied clinical medical physics. 12 (4): 231–244 Jones AK, Pasciak AS (2011) Calculating the peak skin dose resulting from fluoroscopically guided interventions. Part I: Methods. Journal of applied clinical medical physics. 12 (4): 231–244
14.
Zurück zum Zitat Malchair F, Dabin J, Deleu M, et al. Review of skin dose calculation software in interventional cardiology. Physica Med. 2020;80:75–83.CrossRef Malchair F, Dabin J, Deleu M, et al. Review of skin dose calculation software in interventional cardiology. Physica Med. 2020;80:75–83.CrossRef
16.
Zurück zum Zitat ICRP, 1991. 1990 Recommendations of the International Commission on Radiological Protection. ICRP Publication 60. Ann. ICRP 21 (1–3) ICRP, 1991. 1990 Recommendations of the International Commission on Radiological Protection. ICRP Publication 60. Ann. ICRP 21 (1–3)
17.
Zurück zum Zitat ICRP, 2007. The 2007 Recommendations of the International Commission on Radiological Protection. ICRP Publication 103. Ann. ICRP 37 (2–4) ICRP, 2007. The 2007 Recommendations of the International Commission on Radiological Protection. ICRP Publication 103. Ann. ICRP 37 (2–4)
18.
Zurück zum Zitat ICRP Publication 121 (2013) Radiological protection in paediatric diagnostic and interventional radiology. Ann ICRP. 42 (2): 1–63. ICRP Publication 121 (2013) Radiological protection in paediatric diagnostic and interventional radiology. Ann ICRP. 42 (2): 1–63.
Metadaten
Titel
How to Measure/Calculate Radiation Dose in Patients?
verfasst von
Reinhard Loose
Michael Wucherer
Publikationsdatum
03.03.2021
Verlag
Springer US
Erschienen in
CardioVascular and Interventional Radiology / Ausgabe 6/2021
Print ISSN: 0174-1551
Elektronische ISSN: 1432-086X
DOI
https://doi.org/10.1007/s00270-021-02772-x

Weitere Artikel der Ausgabe 6/2021

CardioVascular and Interventional Radiology 6/2021 Zur Ausgabe

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Endlich: Zi zeigt, mit welchen PVS Praxen zufrieden sind

IT für Ärzte Nachrichten

Darauf haben viele Praxen gewartet: Das Zi hat eine Liste von Praxisverwaltungssystemen veröffentlicht, die von Nutzern positiv bewertet werden. Eine gute Grundlage für wechselwillige Ärztinnen und Psychotherapeuten.

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.