Skip to main content
Erschienen in: Strahlentherapie und Onkologie 4/2021

Open Access 17.10.2020 | Original Article

Hybrid volumetric modulated arc therapy for hypofractionated radiotherapy of breast cancer: a treatment planning study

verfasst von: Alexander Venjakob, Michael Oertel, Dominik Alexander Hering, Christos Moustakis, Uwe Haverkamp, Hans Theodor Eich

Erschienen in: Strahlentherapie und Onkologie | Ausgabe 4/2021

Abstract

Purpose

This study aims to evaluate the best possible practice using hybrid volumetric modulated arc therapy (H-VMAT) for hypofractionated radiation therapy of breast cancer. Different combinations of H‑VMAT—a combination of three-dimensional radiotherapy (3D-CRT) and VMAT—were analyzed regarding planning target volume (PTV), dose coverage, and exposure to organs at risk (OAR).

Methods

Planning computed tomography scans were acquired in deep-inspiration breath-hold. A total of 520 treatment plans were calculated and evaluated for 40 patients, comprising six different H‑VMAT plans and a 3D-CRT plan as reference. H‑VMAT plans consisted of two treatment plans including 3D-CRT and VMAT. During H‑VMAT planning, the use of hard wedge filters (HWF) and beam energies were varied. The reference plans were planned with mixed beam energies and the inclusion/omission of HWF.

Results

Compared to the reference treatment plans, all H‑VMAT plans showed consistently better PTV dose coverage, conformity, and homogeneity. Additionally, OAR protection was significantly improved with several H‑VMAT combinations (p < 0.05). The comparison of different H‑VMAT combinations showed that inclusion of HWF in the base plan had a negative impact on PTV dose coverage, conformity, and OAR exposure. It also increased the planned monitor units and beam-on time. Advantages of using lower beam energies (6-MV photons) in both the base plan and in the VMAT supplementary dose were observed.

Conclusion

The H‑VMAT technique is an effective possibility for generating homogenous and conformal dose distributions. With the right choice of H‑VMAT combination, superior OAR protection is achieved compared to 3D-CRT.

Introduction

Radiotherapy (RT) is pivotal in the adjuvant treatment of breast cancer, improving both local control and overall survival [1, 2]. Various radiation concepts and techniques have been established over time. Static three-dimensional radiotherapy (3D-CRT) represents the conventional radiation technique, whereby tangential opposing fields with hard wedge filters (HWF) are used. Modern dynamic irradiation techniques, such as intensity-modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT), attempt to generate more homogeneous and conformal dose distributions for the planning target volume (PTV). Furthermore, better protection of organs at risk (OAR) may be achieved [3, 4]. Nevertheless, dynamic radiation techniques bear the risk of increased induction of secondary tumors attributed to larger areas of low-dose exposure and increased monitor units (MU) [5]. To balance the respective advantages of static and dynamic radiation techniques, Mayo et al. [6] have developed a composite approach combining 3D-CRT and IMRT named hybrid intensity-modulated radiation therapy (H‑IMRT). In our study, we examined the impact of different variations with respect to HWF in the 3D-CRT base plan and beam energies for the application of hybrid volumetric arc therapy (H‑VMAT). Patients were evaluated in terms of PTV dose coverage and OAR exposure. Concerning OAR, dose comparisons were established and differences between left- and right-sided breast therapy were analyzed.

Materials and methods

Patient selection, positioning, and computed tomography

The patients included in this study were selected by defining a volume-based standardized breast size (n = 200), whereby the PTV sizes were evaluated as a measure of breast size. The computed tomography (CT) scans (Canon Aquilion LB, Canon Medical Systems Europe B.V., Zoetermeer, the Netherlands) were acquired with a slice thickness of 3.0 mm. Patients were positioned in a headfirst supine position. An arm board helped to immobilize the chest and thorax with the arms positioned overhead. To minimize intrafractional movement, the CT scans were carried out in deep-inspiration breath-hold (DIBH). This method bears dosimetric advantages, especially for cardiopulmonary OAR exposure and secondary lung cancer risk [710]. Of the 200 patients analyzed, 110 patients met the inclusion criteria of the study regarding PTV size and no exclusion criteria were included, such as funnel chests, pathological enlarged hearts, or anatomic variations. Out of these patients, 40 (20 per breast side) were selected randomly.

Treatment planning and techniques

The Eclipse software (version 15.6, Varian Medical Systems, Palo Alto, CA, USA) was used for treatment planning. Treatment plans were created for a Varian TrueBeam linear accelerator, the calculation model was based on the anisotropic analytical algorithm (AAA, version 15.5.12) and a calculation grid size of 2.5 mm. Each PTV included the whole breast and was cropped 5 mm from the skin surface, resulting in a VMAT optimization where the dose is not forcibly modulated to the surface. The thoracic wall was defined as part of the PTV. Individual treatment plans were created using the H‑VMAT technique and the 3D-CRT technique as reference (Ref). The H‑VMAT plans represented sum plans: each consisted of two treatment plans including 3D-CRT and VMAT. The prescribed dose was 40.05 Gy in 15 fractions (2.67 Gy per fraction) based on the START B trail, which demonstrated the non-inferiority of this hypofractionated regime regarding locoregional control and toxicity in comparison to normofractionated treatment [11]. The reference treatment plans were planned as tangential opposing fields with mixed beam energies (6 MV and 15 MV) and HWF. All planning parameters such as gantry and wedge filter angles, collimator position, and field weightings were individually optimized. H‑VMAT plans were weighted 80/20 between 3D-CRT and VMAT, which included 3D-CRT as a base plan for VMAT optimization. The supplementary dose consisted of a single VMAT field with an identical isocenter position. Its range of rotation angle corresponded to the tangential angles from the base plan. The collimator rotation angle was set to 5° and the maximum dose rate to 600 MU/min. A standard optimization protocol was used for the inverse planning process of VMAT, which was adapted individually. Flattening-filter beams were used in all treatment plans. Within the hybrid approach, six different H‑VMAT combinations (HV1–HV6, Table 1) were used, varying beam energy and the application or omission of HWF in the 3D-CRT plans. The selection of these combinations was based on a previous evaluation by our group. This previous study was an initial assessment of 32 different combinations, 16 of which were within the weightings 60/40 and 80/20, which were evaluated based on a case number of 5 patients. To assess the optimal hybrid combination, superior combinations of this evaluation with respect to quality indices (PTV coverage, conformity, homogeneity) and OAR doses were analyzed in detail (Table 1).
Table 1
Definition of irradiation techniques using different H‑VMAT constellations and conventional 3D-CRT as a reference treatment plan
IT/H-VMAT constellation
3D-CRT base plan energy (MV)
Individual hard wedge filters (base plan)
VMAT energy (MV)
Ref
6/15 (mixed on both sides)
Yes
HV1
6/15 (mixed on both sides)
No
6
HV2
6
No
6
HV3
6
No
15
HV4
6
Yes
6
HV5
6
Yes
15
HV6
6 (posterior lateral), 15 (anterior medial)
No
6
IT irradiation techniques, H‑VMAT hybrid volumetric modulated arc therapy, HV1–HV6 H‑VMAT constellations, 3D-CRT conventional three-dimensional radiotherapy, Ref reference treatment plan
PTV dose coverage had the first priority in treatment planning. The PTV dose requirements in our study were based on the recommendations of ICRU Report 83 [12]:
$$\mathrm{PTV} \text{D}_{\text{mean}}=40.05\,\mathrm{Gy}(100 \% )$$
$$38.05\,\mathrm{Gy}(95\mathrm{\% })\leq \text{PTV D}_{95\mathrm{\%}}\leq 38.85\,\mathrm{Gy}(97\mathrm{\% })$$
$$\mathrm{PTV} \text{D}_{2\mathrm{\% }}\leq 42.85\,\mathrm{Gy}(107\mathrm{\% })$$
In order to ensure an appropriate plan comparison with regard to OAR doses, these PTV requirements should be kept as consistent as possible. For this purpose, an upper limit for PTV D95% was set. Minimizing OAR doses was a secondary priority. The protection was implemented so that the PTV dose requirements could be guaranteed. Each of the prepared 520 treatment plans was separately optimized for individual patients.

Dosimetric parameters

Four different indices were used to assess PTV dose conformity and homogeneity:
1. Coverage index [13]
$$\mathrm{COV}=\frac{\mathrm{PTV}_{\mathrm{PI}}}{\mathrm{PTV}}(\text{ideal value}\,1.000)$$
2. External index [14]
$$\mathrm{EI}=\frac{\mathrm{TV}_{\mathrm{PI}}-\mathrm{PTV}_{\mathrm{PI}}}{\mathrm{PTV}}(\text{ideal value}\,0.000)$$
3. Conformation number [15]
$$\text{CN}=\frac{\mathrm{PTV}_{\mathrm{PI}}}{\mathrm{PTV}}\mathrm{*}\frac{\mathrm{PTV}_{\mathrm{PI}}}{\mathrm{TV}_{\mathrm{PI}}}(\text{ideal value}\,1.000)$$
4. Homogeneity index [12]
$$\mathrm{HI}=\frac{\text{D}_{2}-\text{D}_{98}}{\text{D}_{\text{mean}}}(\text{ideal value}\ 0.000)$$
PTV =
planning target volume (cm3)
PTVPI =
partial volume in the PTV with at least the prescribed dose (cm3)
TVPI =
partial volume in the whole body with at least the prescribed dose (cm3)
D2 =
near maximum dose: dose in 2% of PTV (Gy)
D98 =
near minimum dose: dose in 98% of PTV (Gy)
Dmean =
mean dose of the PTV (Gy) [16].
The Radiation Therapy Oncology Group (RTOG) Report 1005 [17] served as a guide for dosimetric evaluations of the OAR doses (for detailed OAR constraints, see Table 2). Our previous evaluation showed clinically negligible doses to the contralateral lung and spinal cord, so that these OAR were not considered in this study.
Table 2
Dose constraints of OAR defined in RTOG Report 1005 [17] for hypofractionated breast radiation therapy
OAR
Dose specifications (per protocol, variation acceptable)
Ipsilateral lung
V16 = 15–20%
V8 = 35–40%
V4 = 50–55%
Heart
Dmean = 320–400 cGy
Contralateral breast
D5 = 144–240 cGy
OAR organ at risk, RTOG Radiation Therapy Oncology Group, VX volume (%) of the OAR exposed to at least X Gy, D5 minimal dose (Gy) in 5% of the OAR volume with the highest exposure, Dmean mean dose
Further evaluation criteria were the planned monitor units (MU) and beam-on time (BOT) that were observed over the total number of patients.
SPSS software (for Windows, version 26, IBM, Armonk, NY, USA) was used for the statistical analysis. A two-sided Wilcoxon rank-sum test for paired samples was used to evaluate dosimetric parameters. A p-value <0.05 was considered statistically significant.

Results

Planning target volume sizes

A volume-specific standard breast of 961 ± 346 cm3 was determined from the data of 200 patients. A Kolmogorov–Smirnov test confirmed the presence of a normal sample distribution. The analysis included only patients with a PTV or breast size within the range of the standard deviation, integrating 68% of all women, whereas anatomical variations such as funnel chests and pathological enlarged hearts were not considered.

Planning target volume dose coverage

Compared to the reference treatment plan, all analyzed H‑VMAT combinations consistently delivered significantly better values for PTV dose coverage (Fig. 1, Table 3). With the conventional radiation technique (Ref), dose coverage was consistently less successful than with H‑VMAT. Compared to Ref, all H‑VMAT combinations achieved significantly better EI values. The use of HWF in H‑VMAT was disadvantageous for the EI in left-sided breast therapy, whereby HV4 (EI = 0.012 ± 0.008) and HV5 (EI = 0.013 ± 0.009) showed worse values compared to other combinations without HWF use such as HV1 (EI = 0.008 ± 0.003). Compared to Ref (left CN = 0.391 ± 0.050, right CN = 0.395 ± 0.046), the evaluation of conformity showed better values in all H‑VMAT combinations. HV4 (left CN = 0.611 ± 0.016, right CN = 0.622 ± 0.021) and HV5 (left CN = 0.598 ± 0.025, right CN = 0.617 ± 0.020) revealed inferior results compared to other H‑VMAT combinations for both breast sides. Better dose homogeneity within the PTV (HI) was generally achieved in the H‑VMAT plans.
Table 3
Representation of the considered dosimetric parameters using H‑VMAT combinations and the reference treatment plan
Breast side
Dosimetric parameter
Radiation technique/H-VMAT combinationa
Ref
HV1
HV2
HV3
HV4
HV5
HV6
Left
COV
0.523 ± 0.042
0.645 ± 0.029
(p < 0.0001)
0.647 ± 0.017
(p < 0.0001)
0.653 ± 0.019
(p < 0.0001)
0.622 ± 0.018
(p < 0.0001)
0.610 ± 0.028
(p < 0.0001)
0.639 ± 0.014
(p < 0.0001)
EI
0.181 ± 0.041
0.008 ± 0.003
(p < 0.0001)
0.011 ± 0.006
(p < 0.0001)
0.010 ± 0.007
(p < 0.0001)
0.012 ± 0.008
(p < 0.0001)
0.013 ± 0.009
(p < 0.0001)
0.009 ± 0.005
(p < 0.0001)
CN
0.391 ± 0.050
0.637 ± 0.029
(p < 0.0001)
0.637 ± 0.018
(p < 0.0001)
0.642 ± 0.020
(p < 0.0001)
0.611 ± 0.016
(p < 0.0001)
0.598 ± 0.025
(p < 0.0001)
0.630 ± 0.015
(p < 0.0001)
HI
0.143 ± 0.018
0.098 ± 0.005
(p < 0.0001)
0.095 ± 0.004
(p < 0.0001)
0.095 ± 0.004
(p < 0.0001)
0.095 ± 0.005
(p < 0.0001)
0.097 ± 0.005
(p < 0.00001)
0.100 ± 0.004
(p < 0.00001)
IL V16 (%)
14.9 ± 5.2
14.3 ± 5.0
(p < 0.0001)
13.9 ± 4.9
(p < 0.0001)
14.0 ± 4.9
(p < 0.0001)
14.2 ± 4.9
(p < 0.0001)
14.3 ± 4.9
(p < 0.0001)
14.4 ± 4.9
(p = 0.0005)
IL V8 (%)
22.7 ± 6.4
23.1 ± 6.8
(p = 0.5437)
21.7 ± 6.5
(p = 0.0249)
21.9 ± 6.6
(p = 0.0971)
23.0 ± 6.1
(p = 0.5913)
23.3 ± 6.5
(p = 0.2942)
22.9 ± 6.8
(p = 0.7953)
IL V4 (%)
44.4 ± 9.4
43.6 ± 10.8
(p = 0.5530)
41.1 ± 9.9
(p = 0.0028)
42.0 ± 10.3
(p = 0.0348)
46.4 ± 9.4
(p = 0.0259)
47.3 ± 10.4
(p = 0.0084)
43.0 ± 10.7
(p = 0.2730)
IL Dmean (Gy)
8.5 ± 2.0
7.8 ± 1.9
(p < 0.0001)
7.4 ± 1.8
(p < 0.0001)
7.5 ± 1.8
(p < 0.0001)
8.1 ± 1.6
(p = 0.0023)
8.2 ± 1.7
(p = 0.0181)
7.7 ± 1.9
(p < 0.0001)
Heart Dmean (cGy)
216 ± 42
128 ± 37
(p < 0.0001)
121 ± 34
(p < 0.0001)
123 ± 35
(p < 0.0001)
204 ± 33
(p = 0.0003)
205 ± 33
(p = 0.0005)
137 ± 39
(p < 0.0001)
CB D5 (cGy)
200 ± 33
88 ± 26
(p < 0.0001)
70 ± 23
(p < 0.0001)
84 ± 29
(p < 0.0001)
176 ± 29
(p < 0.0001)
183 ± 32
(p < 0.0001)
90 ± 36
(p < 0.0001)
Right
COV
0.535 ± 0.028
0.644 ± 0.018
(p < 0.0001)
0.655 ± 0.019
(p < 0.0001)
0.655 ± 0.016
(p < 0.0001)
0.632 ± 0.022
(p < 0.0001)
0.627 ± 0.021
(p < 0.0001)
0.637 ± 0.015
(p < 0.0001)
EI
0.197 ± 0.072
0.008 ± 0.003
(p < 0.0001)
0.010 ± 0.004
(p < 0.0001)
0.010 ± 0.004
(p < 0.0001)
0.010 ± 0.005
(p < 0.0001)
0.010 ± 0.005
(p < 0.0001)
0.009 ± 0.004
(p < 0.0001)
CN
0.395 ± 0.046
0.636 ± 0.018
(p < 0.0001)
0.644 ± 0.019
(p < 0.0001)
0.645 ± 0.016
(p < 0.0001)
0.622 ± 0.021
(p < 0.0001)
0.617 ± 0.020
(p < 0.0001)
0.628 ± 0.015
(p < 0.0001)
HI
0.140 ± 0.014
0.097 ± 0.006
(p < 0.0001)
0.097 ± 0.004
(p < 0.0001)
0.096 ± 0.005
(p < 0.0001)
0.096 ± 0.006
(p < 0.0001)
0.097 ± 0.005
(p < 0.0001)
0.099 ± 0.004
(p < 0.0001)
IL V16 (%)
15.3 ± 3.7
15.3 ± 3.6
(p = 0.7427)
14.7 ± 3.6
(p < 0.0001)
14.8 ± 3.7
(p = 0.0001)
15.0 ± 3.6
(p = 0.0166)
15.0 ± 3.7
(p = 0.0129)
15.1 ± 3.6
(p = 0.1621)
IL V8 (%)
23.4 ± 4.1
26.1 ± 4.8
(p = 0.0033)
23.2 ± 4.6
(p = 0.7689)
23.6 ± 4.7
(p = 0.6489)
24.6 ± 4.0
(p = 0.0240)
24.9 ± 4.9
(p = 0.0191)
25.1 ± 4.5
(p = 0.0178)
IL V4 (%)
44.8 ± 5.6
49.0 ± 7.9
(p = 0.0107)
43.3 ± 7.8
(p = 0.2904)
44.4 ± 7.8
(p = 0.8226)
48.7 ± 7.8
(p = 0.0164)
49.3 ± 8.7
(p = 0.0180)
46.9 ± 7.4
(p = 0.1834)
IL Dmean (Gy)
8.1 ± 1.2
8.2 ± 1.2
(p = 0.7931)
7.5 ± 1.2
(p = 0.0006)
7.6 ± 1.3
(p = 0.0059)
8.1 ± 1.1
(p = 0.8219)
8.1 ± 1.2
(p = 0.8098)
8.0 ± 1.1
(p = 0.4405)
Heart Dmean (cGy)
132 ± 19
71 ± 26
(p < 0.0001)
63 ± 17
(p < 0.0001)
65 ± 20
(p < 0.0001)
133 ± 16
(p = 0.6923)
132 ± 20
(p = 0.8997)
78 ± 27
(p < 0.0001)
CB D5 (cGy)
233 ± 45
116 ± 42
(p < 0.0001)
96 ± 35
(p < 0.0001)
111 ± 40
(p < 0.0001)
205 ± 38
(p < 0.0001)
213 ± 39
(p < 0.0001)
121 ± 51
(p < 0.0001)
Planned monitor units (MU)
490 ± 43
342 ± 27
365 ± 19
351 ± 40
540 ± 40
530 ± 40
342 ± 17
Planned beam-on time (min)
0.82 ± 0.07
0.91 ± 0.03
0.95 ± 0.03
0.94 ± 0.03
1.26 ± 0.06
1.25 ± 0.06
0.91 ± 0.03
H‑VMAT hybrid volumetric modulated arc therapy, HV1–HV6 H‑VMAT constellations, Ref reference treatment plan, COV planning target volume dose coverage, EI external index, CN conformation number, HI homogeneity index, IL ipsilateral, CB contralateral, VX volume (%) of the organ at risk (OAR) exposed to at least X Gy, D5 minimal dose (Gy) in 5% of the OAR volume with the highest exposure, Dmean mean dose
aHV1–HV6 are presented with p-values to demonstrate significant differences to Ref, using a two-sided Wilcoxon rank sum test for paired samples for the statistical analysis
The analysis within the hybrid combinations showed potential for improvement with regard to the quality indices (Table 4). The inclusion of HWF in the base plan (HV2/HV4 and HV3/HV5) resulted in significant disadvantages regarding COV and CN. The omission of HWF resulted in a significantly better homogeneity index in left-sided, but not in right-sided breast therapy (HV3/HV5). By using higher beam energies in the VMAT supplementary dose (HV2/HV3 and HV4/HV5), better CN values could be achieved. In addition, advantages in PTV dose coverage and conformity only in left-sided therapy (HV4/HV5) were observed.
Table 4
Representation of the considered dosimetric parameters using H‑VMAT combinations
Breast side
Dosimetric parameter
Supplementary dose energy 6 MV/15 MVa
Base plan without using HWF/using HWFa
HV2/HV3
HV4/HV5
HV2/HV4
HV3/HV5
Left
COV
0.1675
0.0055
<0.0001
<0.0001
EI
0.8252
0.6079
0.3626
0.1987
CN
0.1686
0.0038
<0.0001
<0.0001
HI
0.4521
0.0319
0.7829
0.0196
IL V16
0.1274
0.1139
<0.0001
<0.0001
IL V8
0.3137
0.1323
0.0001
0.0005
IL V4
0.1085
0.0356
<0.0001
<0.0001
IL Dmean
0.0696
0.0436
<0.0001
<0.0001
Heart Dmean
0.4134
0.4445
<0.0001
<0.0001
CB D5
<0.0001
<0.0001
<0.0001
<0.0001
Right
COV
0.9909
0.0556
0.0005
<0.0001
EI
0.5597
0.5217
0.8532
0.7558
CN
0.9339
0.0491
0.0005
<0.0001
HI
0.5291
0.4407
0.4650
0.8714
IL V16
0.1995
0.5058
<0.0001
<0.0001
IL V8
0.0646
0.5284
0.0007
0.0003
IL V4
0.0413
0.5445
<0.0001
<0.0001
IL Dmean
0.0006
0.9428
<0.0001
0.0001
Heart Dmean
0.1249
0.7775
<0.0001
<0.0001
CB D5
0.0001
0.0011
<0.0001
<0.0001
H‑VMAT hybrid volumetric modulated arc therapy, HV5–HV5 H‑VMAT constellations, COV planning target volume dose coverage, EI external index, CN conformation number, HI homogeneity index, IL ipsilateral, CB contralateral, VX volume (%) of the organ at risk (OAR) exposed to at least X Gy, D5 minimal dose (Gy) in 5% of the OAR volume with the highest exposure, Dmean mean dose
aP-values demonstrate the significant differences between the various H‑VMAT combinations, using a two-sided Wilcoxon rank sum test for paired samples for the statistical analysis. Values of the dosimetric parameters are listed in Table 3. HV2/HV3 and HV4/HV5 contain the same base plan, the supplementary dose energy was varied. HV2/HV4 and HV3/HV5 contain the same supplementary dose energy, the base plan was varied using HWF

Organs at risk dose exposure

Evaluation of the ipsilateral lung (IL) dose value V16 displayed significantly better values in almost all H‑VMAT combinations on both breast sides (Fig. 2, Table 3). Only for HV1 (V16 = 15.3 ± 3.6%) and HV6 (V16 = 15.1 ± 3.6%) in the case of right-sided breast irradiation was there no significant difference to the reference treatment plan (Ref V16 = 15.3 ± 3.7%). The IL V8 doses in left-sided breast irradiation were equivalent across all RT combinations, with only HV2 (V8 = 21.7 ± 6.5%) showing statistical benefit over Ref (V8 = 22.7 ± 6.4%). In right-sided breast irradiation, most hybrid combinations, HV1, HV4, HV5, and HV6, provided disadvantages. Overall, the tolerable dose values of the RTOG Report 1005 (V8 = 35–40%) were met in most used techniques (except for one patient HV1, HV5, and HV6 for the right breast). When considering the IL dose V4, only HV2 (V4 = 41.1 ± 9.9%) and HV3 (V4 = 42.0 ± 10.3%) were superior to Ref (V4 = 44.4 ± 9.4%) in left-sided therapy. HV4 (V4 = 46.4 ± 9.4%) and HV5 (V4 = 47.3 ± 10.4%) delivered significantly worse values as well as HV1, HV4, and HV5 on right-sided therapy. All other H‑VMAT combinations were equivalent to Ref regarding IL V4. The analysis of the mean ipsilateral lung dose showed advantages of the hybrid technique on left breast therapy, with significantly better values for H‑VMAT. In right-sided therapy, only H‑VMAT combinations HV2 (Dmean = 7.5 ± 1.2 Gy) and HV3 (Dmean = 7.6 ± 1.3 Gy) offered statistically significant advantages over the reference (Dmean = 8.1 ± 1.2 Gy).
All dose values of the heart complied with the dose specification of RTOG Report 1005 regarding Dmean = 320–400 cGy (Fig. 3, Table 3). Especially in left-sided breast therapy, the heart was better protected with all H‑VMAT combinations considered. HV1 (Dmean = 128 ± 37 cGy), HV2 (Dmean = 121 ± 34 cGy), HV3 (Dmean = 123 ± 35 cGy), and HV6 (Dmean = 137 ± 39 cGy) showed the best values and thus the most pronounced advantages over Ref (Dmean = 216 ± 42 cGy). In right-sided breast irradiation, the same trend could be observed (HV1 Dmean = 71 ± 26 cGy, HV2 Dmean = 63 ± 17 cGy, HV3 Dmean = 65 ± 20 cGy, HV6 Dmean = 78 ± 27 cGy, and Ref Dmean = 132 ± 19 cGy). The combinations HV4 and HV5 consistently achieved the highest heart exposure in H‑VMAT. In general, higher heart dose values were evaluated in left-sided breast therapy.
Some treatment techniques could not fulfill the dose constraints D5 = 144–240 cGy of the contralateral breast according to the guidelines of RTOG Report 1005 (Fig. 3, Table 3). All considered hybrid combinations showed advantages over the 3D-CRT treatment plan (left D5 = 200 ± 33 cGy, right D5 = 233 ± 45 cGy). The lowest dose values with the greatest advantages were achieved with H‑VMAT combinations without the inclusion of HWF. HV2 (left D5 = 70 ± 23 cGy, right D5 = 96 ± 35 cGy) and HV3 (left D5 = 84 ± 29 cGy, right D5 = 111 ± 40 cGy) provided the best values on both breast sides.
The comparison of hybrid combinations showed a possibility to reduce the OAR doses (Table 4). Significant advantages of combinations without the use of HWF in the base plan were found for all OAR. Regarding the VMAT supplementary dose, there is a tendency that the choice of higher beam energy (15 MV instead of 6 MV) is disadvantageous for OAR exposure. For IL V4, IL Dmean, and CB D5, significant advantages could be observed using a lower VMAT energy.

Planned monitor units and beam-on times

Increased MU were required when using HWF (Table 3). Therefore, Ref and the combinations HV4 and HV5 in hybrid planning showed the highest values. Compared to Ref, the H‑VMAT plans showed consistently higher values for beam-on times. The longest BOT were observed using HWF (HV4 and HV5).

Discussion

To our knowledge, this is the first study to analyze different base plan designs using 3D-CRT, in which the use of HWF and beam energies were varied. Variations in beam energy for the VMAT supplementary dose were also performed. The H‑VMAT combinations HV2 and HV3 (Table 1) uniformly achieved the best values regarding all dosimetric parameters in comparison to the reference treatment plan (Fig. 4; Table 3). Both favored combinations contained a 6-MV photon base plan without the use of HWF, which are conventionally used for shape compensation and dose homogenization. The comparison of different hybrid combinations showed significant disadvantages for quality indices and OAR doses when using HWF in the base plan (Table 4). In a tangential field arrangement using wedge orientation for breast surface compensation, the total body and OAR doses were increased. Advantages of virtual wedges over hard wedge filters are reported in the literature [1821]. Less scattering effects led to lower average doses outside the irradiation field. Consequently, smaller low-dose areas in the body and lower OAR exposure, e.g., of the contralateral breast in breast radiotherapy, were observed. Furthermore, a reduction in MU was achieved. Nevertheless, virtual wedges were omitted in our study because the collimator would have to be rotated by 90 ° for technical realization (Varian TrueBeam linear accelerator). This would have complicated the exact fitting of the MLC to the PTV contour and would have increased the exposure of the OAR, especially the ipsilateral lung. The use of HWF is not necessary in H‑VMAT for reasons of dynamic supplementary dose and may even be harmful due to the increased MU and beam-on time. Furthermore, higher photon energies in VMAT showed an unfavorable tendency with regard to OAR exposure (Table 3). The maximum of depth dose curves shifts the maximum dose deeper into the tissue and a longer dose extension is created, which also increases the total body and OAR doses.
In the present study, the weighting between the radiation techniques 3D-CRT and VMAT was uniform for all combinations (80/20). Other studies confirmed the benefit of this ratio when using the H‑VMAT technique. Balaji et al. [22, 23] analyzed different ratios of chest wall radiation and identified 80/20 and 70/30 as optimal weighting concerning dose coverage of the PTV and OAR exposure. Nevertheless, the choice of weighting should be adapted individually, depending on the clinical scenario. Patient age plays an important role, as late side effects are negligible in older patients, but both acute and late effects should be considered in younger patients. With the increase of the VMAT component weighting, the low-dose area expands. In younger patients, these areas can be minimized by choosing 80/20.
The study of Lin et al. [24] showed benefits of the H‑VMAT technique over pure-VMAT and a fixed-field IMRT technique in radiotherapy of left breast cancer. They used a base plan consisting of tangential IMRT fields (T-IMRT). In our study, 3D-CRT was used in the base plan, which enables a reduction in MU and delivery time compared to dynamic techniques like IMRT. Less MLC radiation leakage and internal body scattering can be the benefit of decreased MU, which leads to a reduction in the total body dose [25]. With a reduction of the delivery time, patients have to complete fewer breathing cycles per fraction, which makes the treatment more feasible when using DIBH.
For the supplementary dose, various concepts may be used. Balaji et al. [26] reported clinically similar results of H‑IMRT and H‑VMAT using 6‑MV flattening filter-free (FFF) photon beams. The researchers recommend H‑VMAT for hypofractionated breast RT in DIBH due to reduced MU and treatment delivery time. Another study by Chen et al. [27] consistently reported advantages of H‑VMAT over H‑IMRT with regard to dose conformity, heart dose, and delivery time. They compared H‑VMAT with two different H‑IMRT designs in the treatment of early-stage left-sided breast cancer. Two tangential beams were used as a base plan and two coplanar 90-degree arcs (H-VMAT) or four IMRT fields (H-IMRT) for the supplementary dose. Ramasubramanian et al. [28] analyzed the influence of different arc designs for left breast therapy, comparing two partial arcs (2A), four partial arcs (4A), and four tangential arcs (TA). The dosimetric superiority of PTV and OAR doses could be analyzed in 2A and 4A, with the 2A method additionally allowing a reduction of MU and beam-on time. We only used flattening-filter beams and a single VMAT field over the entire tangential angular range for the supplementary dose, such that our approach resembles the 2A method. This approach may realize a further shortening of treatment time due to the uninterrupted continuous arc.
The basis for proper dose application with H‑VMAT to the patient is the existence of good reproducibility of the patient positioning. Dynamic techniques like H‑VMAT may be prone to intra- and interfractional positioning uncertainties, which demand reproducibility of patient positioning both in treatment planning and execution [27, 2931]. In order to minimize positioning uncertainties and deviations between planned and delivered dose distributions, CT acquisition and radiotherapy should be carried out in DIBH. Besides, studies have shown benefits of DIBH concerning irradiated cardiac volume and secondary lung cancer risk in left-sided breast radiotherapy [710, 32]. Efficient therapy in DIBH, however, requires special measures and skills. Employee training, patient selection criteria, patient coaching, and protocols for verifying the treatment are essential requirements for efficient therapy [33].
Another key limitation of our study is the restricted patient selection. We evaluated and analyzed standardized breasts or PTV sizes only. Anatomical deviations in size or non-conventional shapes may change the findings.

Conclusion

Our study showed the advantages of the H‑VMAT technique with a weighting of 80% 3D-CRT/20% VMAT compared to the conventional static tangential radiation technique. The combination of static and dynamic radiation techniques improved PTV dose coverage, conformity, and homogeneity. The doses of the OAR ipsilateral lung, heart, and contralateral breast were significantly reduced (p < 0.05) by a suitable choice of the plan combination of H‑VMAT. In treatment planning, a photon energy of 6 MV should be preferred for both the tangential radiation fields of the base plan and the VMAT supplementary dose. The use of radiation wedges in the base plan was found to be disadvantageous for PTV dose coverage and conformity, OAR doses, MU number, and delivery time.

Compliance with ethical guidelines

Conflict of interest

A. Venjakob, M. Oertel, D.A. Hering, C. Moustakis, U. Haverkamp, and H. T. Eich declare that they have no competing interests.

Ethical standards

For this article no studies with human participants or animals were performed by any of the authors. All studies performed were in accordance with the ethical standards indicated in each case.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Strahlentherapie und Onkologie

Print-Titel

•Übersichten, Originalien, Kasuistiken

•Kommentierte Literatur aus der Radioonkologie, Strahlenbiologie und -physik

Literatur
5.
Zurück zum Zitat Hall EJ (2006) Intensity-modulated radiation therapy, protons, and the risk of second cancers. Int J Radiat Oncol Biol Phys 65:1–7CrossRef Hall EJ (2006) Intensity-modulated radiation therapy, protons, and the risk of second cancers. Int J Radiat Oncol Biol Phys 65:1–7CrossRef
Metadaten
Titel
Hybrid volumetric modulated arc therapy for hypofractionated radiotherapy of breast cancer: a treatment planning study
verfasst von
Alexander Venjakob
Michael Oertel
Dominik Alexander Hering
Christos Moustakis
Uwe Haverkamp
Hans Theodor Eich
Publikationsdatum
17.10.2020
Verlag
Springer Berlin Heidelberg
Erschienen in
Strahlentherapie und Onkologie / Ausgabe 4/2021
Print ISSN: 0179-7158
Elektronische ISSN: 1439-099X
DOI
https://doi.org/10.1007/s00066-020-01696-8

Weitere Artikel der Ausgabe 4/2021

Strahlentherapie und Onkologie 4/2021 Zur Ausgabe

Update Onkologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.