Skip to main content
Erschienen in: European Journal of Nuclear Medicine and Molecular Imaging 11/2003

01.11.2003 | Original Article

Immunoscintigraphy of septic loosening of knee endoprosthesis: a retrospective evaluation of the antigranulocyte antibody BW 250/183

verfasst von: Rigobert Klett, Jens Kordelle, Ulrich Stahl, Alexander Khalisi, Maximillian Puille, Dagmar Steiner, Richard Bauer

Erschienen in: European Journal of Nuclear Medicine and Molecular Imaging | Ausgabe 11/2003

Einloggen, um Zugang zu erhalten

Abstract

Immunoscintigraphy with the use of the antigranulocyte antibody BW 250/183 is a reliable method for detecting infection, especially in septic loosening of hip prostheses, for which purpose quantitative interpretation of the time-activity course is employed. Therefore, we retrospectively studied whether similar results could be achieved in knee prostheses. We verified 28 scintigraphic examinations in 26 patients by histology. Scintigraphy was performed during an early (4–6 h post injection) and a late phase (23–25 h post injection). Infection was diagnosed when activity around the knee prosthesis increased by more than 10% compared with bone marrow. We found a sensitivity and a negative predictive value of 100%, a specificity of 80%, a positive predictive value of 81% and an accuracy of 89%. Specificity and accuracy are lower than in the evaluation of hip prostheses; however, in comparison to other scintigraphic modalities, scintigraphy with the antigranulocyte antibody BW 250/183 is superior in excluding infection and has better specificity and accuracy. Therefore, as in the case of hip prostheses, the described methodology appears to be the scintigraphic modality of choice for knee prostheses.
Literatur
1.
Zurück zum Zitat Berberich R, Hennes P, Alexander C. Entzündungsnachweis und HAMA-Bildung nach Applikation des monoklonalen Antikörpers BW 250/183. Nuklearmedizin 1992; 31:70–73.PubMed Berberich R, Hennes P, Alexander C. Entzündungsnachweis und HAMA-Bildung nach Applikation des monoklonalen Antikörpers BW 250/183. Nuklearmedizin 1992; 31:70–73.PubMed
2.
Zurück zum Zitat Kaim A, Maurer T, Ochsner P, et al. Chronic complicated osteomyelitis of the appendicular skeleton: diagnosis with technetium-99m-labelled antigranulocyte antibody-immunoscintigraphy. Eur J Nucl Med 1997; 24:732–738.CrossRefPubMed Kaim A, Maurer T, Ochsner P, et al. Chronic complicated osteomyelitis of the appendicular skeleton: diagnosis with technetium-99m-labelled antigranulocyte antibody-immunoscintigraphy. Eur J Nucl Med 1997; 24:732–738.CrossRefPubMed
3.
Zurück zum Zitat Steinstraesser A, Oberhausen E. Granulocyte labelling kit BW 250/183. Results of the European Multicenter Trial. Nuklearmedizin 1996; 35:1–11.PubMed Steinstraesser A, Oberhausen E. Granulocyte labelling kit BW 250/183. Results of the European Multicenter Trial. Nuklearmedizin 1996; 35:1–11.PubMed
4.
Zurück zum Zitat Boubaker A, Bischof Delaloye A, Blanc CH, et al. Immunoscintigraphy with antigranulocyte monoclonal antibodies for the diagnosis of septic loosening of hip prostheses. Eur J Nucl Med 1995; 22:139–147.PubMed Boubaker A, Bischof Delaloye A, Blanc CH, et al. Immunoscintigraphy with antigranulocyte monoclonal antibodies for the diagnosis of septic loosening of hip prostheses. Eur J Nucl Med 1995; 22:139–147.PubMed
5.
Zurück zum Zitat Sciuk J, Puskás C, Greitemann B, et al. White blood cell scintigraphy with monoclonal antibodies in the study of the infected endoprosthesis. Eur J Nucl Med 1992; 19:497–502. Sciuk J, Puskás C, Greitemann B, et al. White blood cell scintigraphy with monoclonal antibodies in the study of the infected endoprosthesis. Eur J Nucl Med 1992; 19:497–502.
6.
Zurück zum Zitat Klett R, Khalisi A, Haas H, Puille M, Steiner D, Stürz H, Bauer R. Stufendiagnostik septischer Hüftendoprothesenlockerungen mittels Laborparametern und Antigranulozytenszintigraphie. Z Orthop 2001; 139:415–419.CrossRefPubMed Klett R, Khalisi A, Haas H, Puille M, Steiner D, Stürz H, Bauer R. Stufendiagnostik septischer Hüftendoprothesenlockerungen mittels Laborparametern und Antigranulozytenszintigraphie. Z Orthop 2001; 139:415–419.CrossRefPubMed
7.
Zurück zum Zitat Klett R, Steiner D, Puille M, Khalisi A, Matter HP, Stürz H, Bauer R. Antigranulocyte scintigraphy of septic loosening of hip prosthesis: influence of different analyzing methods. Nuklearmedizin 2001; 40:75–79.PubMed Klett R, Steiner D, Puille M, Khalisi A, Matter HP, Stürz H, Bauer R. Antigranulocyte scintigraphy of septic loosening of hip prosthesis: influence of different analyzing methods. Nuklearmedizin 2001; 40:75–79.PubMed
8.
Zurück zum Zitat Mirra JM, Marder MD, Amstutz HC. The pathology of failed total joint arthroplasty. Clin Orthop 1982; 170:175–183.PubMed Mirra JM, Marder MD, Amstutz HC. The pathology of failed total joint arthroplasty. Clin Orthop 1982; 170:175–183.PubMed
9.
Zurück zum Zitat Love C, Tomas MB, Marwin SE, Pugliese PV, Palestro CJ. Role of nuclear medicine in diagnosis of the infected joint replacement. Radiographics 2001; 21:1229–1238.PubMed Love C, Tomas MB, Marwin SE, Pugliese PV, Palestro CJ. Role of nuclear medicine in diagnosis of the infected joint replacement. Radiographics 2001; 21:1229–1238.PubMed
10.
Zurück zum Zitat Steinstraesser A, Berberich R, Kuhlmann L, Zabori S, Schwarz A. Binding of the monoclonal antibody BW 250/183 to human granulocytes. Nuklearmedizin 1992; 31:57–63.PubMed Steinstraesser A, Berberich R, Kuhlmann L, Zabori S, Schwarz A. Binding of the monoclonal antibody BW 250/183 to human granulocytes. Nuklearmedizin 1992; 31:57–63.PubMed
11.
Zurück zum Zitat Larikka MJ, Ahonen AK, Niemelä O, Puronto O, Junila JA, Hämäläinen MM, Britton K, Syrjälä HP.99mTc-ciprofloxacin (Infecton) imaging in the diagnosis of knee prosthesis infections. Nucl Med Commun 2002; 23:167–170.CrossRefPubMed Larikka MJ, Ahonen AK, Niemelä O, Puronto O, Junila JA, Hämäläinen MM, Britton K, Syrjälä HP.99mTc-ciprofloxacin (Infecton) imaging in the diagnosis of knee prosthesis infections. Nucl Med Commun 2002; 23:167–170.CrossRefPubMed
12.
Zurück zum Zitat Ivanéevic V, Perka C, Hasart O, Sandrock D, Munz DL. Imaging of low-grade bone infection with a technetium-99m labelled monoclonal anti-NCA-90 Fab' fragment in patients with previous joint surgery. Eur J Nucl Med 2002; 29:547–551.CrossRef Ivanéevic V, Perka C, Hasart O, Sandrock D, Munz DL. Imaging of low-grade bone infection with a technetium-99m labelled monoclonal anti-NCA-90 Fab' fragment in patients with previous joint surgery. Eur J Nucl Med 2002; 29:547–551.CrossRef
13.
Zurück zum Zitat Niijhof MW, Oyen WJG, van Kampen A, Claessens RAMJ, van der Meer JWM, Corstens FHM. Hip and knee arthroplasty infection. In-111-IgG scintigraphy in 102 cases. Acta Orthop Scand 1997; 68:332–336.PubMed Niijhof MW, Oyen WJG, van Kampen A, Claessens RAMJ, van der Meer JWM, Corstens FHM. Hip and knee arthroplasty infection. In-111-IgG scintigraphy in 102 cases. Acta Orthop Scand 1997; 68:332–336.PubMed
14.
Zurück zum Zitat van Acker F, Nuyts J, Maes A, Vanquickenborne B, Stuyck J, Bellemans J, Vleugels S, Bormans G, Mortelmans L. FDG-PET,99mTc-HMPAO white blood cell SPET and bone scintigraphy in the evaluation of painful total knee arthroplasties. Eur J Nucl Med 2001; 28:1496–1504.CrossRefPubMed van Acker F, Nuyts J, Maes A, Vanquickenborne B, Stuyck J, Bellemans J, Vleugels S, Bormans G, Mortelmans L. FDG-PET,99mTc-HMPAO white blood cell SPET and bone scintigraphy in the evaluation of painful total knee arthroplasties. Eur J Nucl Med 2001; 28:1496–1504.CrossRefPubMed
15.
Zurück zum Zitat Zhuang H, Duarte PS, Pourdehnad M, Maes A, van Acker F, Shnier D, Garino JP, Fitzgerald RH, Alavi A. The promising role of18F-FDG PET in detecting infected lower limb prosthesis implants. J Nucl Med 2001; 42:44–48.PubMed Zhuang H, Duarte PS, Pourdehnad M, Maes A, van Acker F, Shnier D, Garino JP, Fitzgerald RH, Alavi A. The promising role of18F-FDG PET in detecting infected lower limb prosthesis implants. J Nucl Med 2001; 42:44–48.PubMed
16.
Zurück zum Zitat Joseph TN, Mujtaba M, Chen AL, Maurer SL, Zuckerman JD, Maldjian C, Di Cesare PE. Efficacy of combined technetium-99m sulfur colloid/indium-111 leukocyte scans to detect infected total hip and knee arthroplasties. J Arthroplasty 2001; 16:753–758.CrossRefPubMed Joseph TN, Mujtaba M, Chen AL, Maurer SL, Zuckerman JD, Maldjian C, Di Cesare PE. Efficacy of combined technetium-99m sulfur colloid/indium-111 leukocyte scans to detect infected total hip and knee arthroplasties. J Arthroplasty 2001; 16:753–758.CrossRefPubMed
17.
Zurück zum Zitat Scher DM, Pak K, Lonner JH, Finkel JE, Zuckerman JD, Cesare PE. The predictive value of indium-111 leukocyte scans in the diagnosis of infected total hip, knee, or resection arthroplasties. J Arthroplasty 2000; 15:295–300.PubMed Scher DM, Pak K, Lonner JH, Finkel JE, Zuckerman JD, Cesare PE. The predictive value of indium-111 leukocyte scans in the diagnosis of infected total hip, knee, or resection arthroplasties. J Arthroplasty 2000; 15:295–300.PubMed
Metadaten
Titel
Immunoscintigraphy of septic loosening of knee endoprosthesis: a retrospective evaluation of the antigranulocyte antibody BW 250/183
verfasst von
Rigobert Klett
Jens Kordelle
Ulrich Stahl
Alexander Khalisi
Maximillian Puille
Dagmar Steiner
Richard Bauer
Publikationsdatum
01.11.2003
Verlag
Springer-Verlag
Erschienen in
European Journal of Nuclear Medicine and Molecular Imaging / Ausgabe 11/2003
Print ISSN: 1619-7070
Elektronische ISSN: 1619-7089
DOI
https://doi.org/10.1007/s00259-003-1275-1

Weitere Artikel der Ausgabe 11/2003

European Journal of Nuclear Medicine and Molecular Imaging 11/2003 Zur Ausgabe