Erschienen in: BMC Ophthalmology 1/2018

Open Access 01.12.2018 | Case report

Impending extrusion of Ex-PRESS shunt treated by shunt-position adjustment: a case report

verfasst von: Yong Ju Song, Sumi Kim, Gil Joong Yoon

Erschienen in: BMC Ophthalmology | Ausgabe 1/2018

Abstract

Background

To report a case of impending extrusion of Ex-PRESS shunt treated by shunt-position adjustment.

Case presentation

A 56-year-old Asian woman presented with impending extrusion and malposition of Ex-PRESS shunt in her left eye. The bleb of the left eye was shallow and diffuse. In the past, the patient was treated by Ex-PRESS shunt implantation under the scleral flap in both eyes. There had been no Ex-PRESS shunt-related complication in her right eye, and she reported no history of left-eye trauma. Based on these findings, we hypothesized that the source of the left-eye problem was a loosely fixed Ex-PRESS shunt spur. It was thought, furthermore, that this inadequate scleral resistance during the Ex-PRESS shunt implantation was due to the low scleral rigidity resulting from high myopia and insufficient maintenance of the anterior chamber. We proceeded to make an incision in the area adjacent to the Ex-PRESS shunt using a super sharp blade. The shunt was then pushed into the anterior chamber with forceps, and the spur was fixed firmly. Pushing the shunt to the anterior chamber was found to have been sufficient to fix it firmly. In fact, when the sclera was palpated with a sponge, aqueous outflow was observed with no shunt displacement. Postoperative intraocular pressure (IOP) was managed well, and the bleb had formed with diffuse, prominent shapes. The Ex-PRESS shunt was well sustained with good positioning.

Conclusions

When an Ex-PRESS shunt operation is performed on a patient who shows a tendency for low scleral rigidity, shunt implantation should be accomplished carefully and with force adequate for firm spur fixation.
Abkürzungen
AXL
axial length
BCVA
best-corrected visual acuity
D
diopters
IOP
intraocular pressure
MMC
mitomycin C

Background

The Ex-PRESS shunt (Alcon Laboratories Inc., Fort Worth, TX) is a stainless steel non-valve tube. It has a spur designed for fixation in the trabecular meshwork without suturing. Originally implanted under the conjunctiva to connect the anterior chamber to the subconjunctival space [1], it is now positioned under the scleral flap, due to complications such as hypotony, conjunctival erosion, shunt extrusion, and exposure resulting from the earlier methodology [24].
In order to regulate intraocular pressure (IOP) adequately and prevent complications including shunt extrusion or exposure, the Ex-PRESS shunt must be properly positioned and securely fixed. There have been reports of Ex-PRESS shunt removal necessitated by shunt extrusion or exposure resulting from improper fixation or trauma, or simply by spontaneous extrusion or exposure [47]. We herein report a case of impending improper-fixation-caused extrusion of Ex-PRESS shunt treated by shunt-position adjustment without shunt removal.

Case presentation

A 56-year-old Asian woman visited our clinic. Her best-corrected visual acuity (BCVA) was 20/20, and the IOP in the problem eye (the left) was 13 mmHg. Slit-lamp examination showed impending Ex-PRESS shunt extrusion or exposure. The internal opening was tilted to the corneal endothelium obliquely in the anterior chamber, and the external plate was prominent in the subconjunctival space (Fig. 1). The color of the bleb was pale, and its shape was diffuse and shallow, though there was no microcyst.
The patient had been followed up on for senile cataract and open-angle glaucoma, with maximal medical therapy. Seven (7) months previously, IOP in both eyes was uncontrolled, ranging from 25 to 28 mmHg. The spherical error was −9.5 diopters(D) and axial length (AXL) was 26.98 mm in the right eye. The spherical error was −5.0D and AXL was 27.06 mm in the left eye. BCVA was 20/200 in the right eye and 20/25 in the left eye, due to nuclear cataracts. This had necessitated combined surgery (cataract + Ex-PRESS shunt implantation) in our hospital. A 3 × 3 mm scleral flap of half thickness was made at superotemporal area. Then, phacoemulsification and lens implantation was performed through a 2.2 mm corneal incision site located in the temporal area. The tract was made with a 25-gauge needle in one-third of the medial area of the blue-gray transition zone, under a preformed scleral flap. The Ex-PRESS shunt (P200; Alcon Laboratories Inc., Fort Worth, TX) was then pushed into the anterior chamber. After surgery, right-eye visual acuity, IOP, bleb shape and Ex-PRESS shunt positioning were all satisfactory. In the left eye, however, the spur of the Ex-PRESS shunt was not firmly fixed in the trabecular meshwork, and the shunt had been implanted obliquely. These left-eye problems were thought to have been caused by low scleral rigidity resulting from high myopia and insufficient maintenance of the anterior chamber, which presumably had rendered scleral resistance inadequate during Ex-PRESS shunt implantation. We decided to observe carefully without surgical intervention, because the external plate of the shunt was covered well by scleral flap and there was no extrusion sign.
Seven (7) months postoperatively, we observed an impending extrusion sign that the external plate was prominent in the subconjunctival space, though it was covered by the scleral flap. Because Ex-PRESS shunt exposure can necessitate shunt removal or additional operation for IOP control, surgery was performed prior to any exposure. A limbus-based conjunctival incision was made followed by careful dissection of the adhesions between the subconjunctival and Tenon tissue. A subsequent incision was made along the previous scleral-flap wound with a super sharp blade, and then the scleral flap was opened carefully until the Ex-PRESS shunt was visible. After dissection of adhesions between the external plate of the EX-PRESS shunt and the surrounding tissue was performed, the shunt was grasped with forceps. However, as the shunt was fixed stronger than expected and, thus, was not easily removable by forceps, we decided against removal and opted for adjustment of its position instead. To make sufficient space for easy adjustment, an incision with the super sharp blade was made from the external-plate-inserted site, in both lateral directions, into the anterior chamber parallel to the iris plane. Then, the shunt was inserted perpendicularly to the limbus, parallel to the iris plane, and well pushed in anterior chamber by forceps for firm fixation of the spur. The internal opening was parallel to the iris, and the external plate was tight on the sclera (Fig. 2). When sclera was palpated with a sponge, aqueous outflow was observed without shunt displacement. After checking repeatedly, the scleral flap was closed with 4 simple sutures. Mitomycin C (MMC) 0.04% was applied for 4 min, followed by irrigation. The MMC was applied as far as possible from the distal margin of the scleral flap, so as to minimize the possibility of MMC penetration into the anterior chamber. Finally, the conjunctival incision was sutured. Postoperatively, IOP was maintained between 10 and 12 mmHg and the bleb was limited to the diffuse, prominent form. The Ex-PRESS shunt was sustained with good positioning (Fig. 3).

Discussion

Glaucoma is a leading cause of irreversible blindness worldwide [8]. If IOP cannot be controlled with medication, surgery is necessary. The standard procedure has been trabeculectomy. For better trabeculectomy results, the Ex-PRESS shunt (Optonol Ltd., Neve Ilan, Israel) was introduced by Belkin and Glovinsky in 1998 [9]. It is a stainless steel, non-valve tube composed of a body, spur and external plate. Currently, a device of 2.64 mm length and 50 or 200 um internal diameter (P50 or P200; Alcon Laboratories Inc., Fort Worth, TX) is commonly used [10]. The Ex-PRESS shunt connects the anterior chamber and subconjunctival space. Unlike conventional trabeculectomy, Ex-PRESS shunt surgery can be formulated. In trabeculectomy, the sizes of the fistula and corresponding iridectomy vary by operator’s preference. However, the size of lumen of the Ex-PRESS shunt is standardized and as such, it is possible to estimate the filtration amount. And Ex-PRESS shunt surgery does not require sclerectomy or iridectomy. Compared with trabeculectomy in theologically, it can decrease inflammation and hyphema [11]. Moreover, a meta-analysis reported that Ex-PRESS shunt surgery showed similar efficacy to conventional trabeculectomy in terms of IOP lowering, vision recovery and operative success rate [12]. Given these advantages, Ex-PRESS shunt surgery has been utilized widely in the treatment of glaucoma.
In the early days of its application, the Ex-PRESS shunt was implanted under the conjunctiva; however, complications such as hypotony, conjunctival erosion, shunt extrusion or exposure were encountered. To prevent such complications, Dahan and Carmichael performed implantation under the scleral flap [13]. In the decade since this innovation, complications have been rare. Among Stein JD et al.’s 8 documented cases (8 eyes) of Ex-PRESS shunt exposure, only 2 had been implanted under the scleral flap (the 6 others, under the conjunctiva) [6]. In each case, the shunt was removed and a scleral patch graft was performed to close the residual 400 um diameter wound. Kourin AS et al. reported one case of Ex-PRESS shunt scleral-flap implantation necessitating shunt removal [7]. Performing a 2-mm incision adjacent to the shunt and penetrating the anterior chamber parallel to iris, they grasped the shunt with a needle holder and rotated it counterclockwise in order to align the spur with the incision, thereby easily removing the shunt thorough the incision site.
Ex-PRESS shunt scleral-flap implantation has, in most cases, provided resistance adequate for prevent shunt extrusion. The potential for spontaneous post-operative displacement, accordingly, is low. This means that Ex-PRESS shunt extrusion occurs more readily as a consequence of intraoperative improper fixation. In our case, a patient with senile cataract and open-angle glaucoma underwent combined surgery (cataract + Ex-PRESS shunt implantation) in both eyes. Considering that the right-eye shunt sustained good function without displacement, and keeping in mind all of the above-noted findings, we supposed that the left-eye shunt extrusion had occurred by intraoperative improper spur fixation. The determined causes of the improper spur fixation are as follows: first, the patient had low scleral rigidity originating from high myopia; second, in the left-eye operation, unlike the right-eye case, the viscoelastic material utilized in the anterior chamber was insufficient. These factors combined to compromise scleral resistance to the implanted Ex-PRESS shunt.
According to the standard procedure, when malposition or exposure on conjunctiva of the Ex-PRESS shunt occurs, the shunt is removed and secondary glaucoma surgery is performed. However, we treated a patient with impending extrusion of Ex-PRESS shunt by shunt-position adjustment. To our best knowledge, this is first case of the treatment of malpositioned EX-PRESS shunt by shunt-position adjustment. This treatment is considered to be a particularly effective method, because it renders shunt removal and secondary glaucoma surgery unnecessary.
Ex-PRESS shunt removal is always quite traumatic and in cases of impending erosion it impossible to try to reposition it instead of straight forward removal. So to prevent this situation, Ex-PRESS shunt should be inserted perpendicularly to the limbus, parallel to the iris plane and well pushed in anterior chamber for firm fixation of the spur.

Conclusions

In conclusion, when Ex-PRESS shunt implantation is performed on a patient who shows a tendency to low scleral rigidity, care should be taken to ensure that the shunt spur is firmly fixed.

Acknowledgements

None.

Funding

None.

Availability of data and materials

All the data supporting the conclusions of this article is included in the present article.
Ethical approval is not required because this is a single case report that does not include data that can identify the patient.
Written informed consent was obtained from the patient for publication of this case report and any accompanying images. A copy of the consent is available for review by the Editor of this journal.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
1.
Zurück zum Zitat Gandolfi S, Traverso CF, Bron A, Sellem E, Kaplan-Messas A, Belkin M. Short-term results of a miniature draining implant for glaucoma in combined surgery with phacoemulsification. Acta Ophthalmol Scand Suppl. 2002;236:66.CrossRefPubMed Gandolfi S, Traverso CF, Bron A, Sellem E, Kaplan-Messas A, Belkin M. Short-term results of a miniature draining implant for glaucoma in combined surgery with phacoemulsification. Acta Ophthalmol Scand Suppl. 2002;236:66.CrossRefPubMed
2.
Zurück zum Zitat Stewart RM, Diamond JG, Ashmore ED, Ayyala RS. Complications following ex-press glaucoma shunt implantation. Am J Ophthalmol. 2005;140(2):340–1.CrossRefPubMed Stewart RM, Diamond JG, Ashmore ED, Ayyala RS. Complications following ex-press glaucoma shunt implantation. Am J Ophthalmol. 2005;140(2):340–1.CrossRefPubMed
3.
Zurück zum Zitat Dahan E. Subconjunctival insertion of ex-PRESS R-50 miniature glaucoma implant. J Cataract Refract Surg. 2008;34(5):716. author reply 716-717CrossRefPubMed Dahan E. Subconjunctival insertion of ex-PRESS R-50 miniature glaucoma implant. J Cataract Refract Surg. 2008;34(5):716. author reply 716-717CrossRefPubMed
4.
Zurück zum Zitat Tavolato M, Babighian S, Galan A. Spontaneous extrusion of a stainless steel glaucoma drainage implant (ex-PRESS). Eur J Ophthalmol. 2006;16(5):753–5.PubMed Tavolato M, Babighian S, Galan A. Spontaneous extrusion of a stainless steel glaucoma drainage implant (ex-PRESS). Eur J Ophthalmol. 2006;16(5):753–5.PubMed
5.
Zurück zum Zitat Garg SJ, Kanitkar K, Weichel E, Fischer D: Trauma-induced extrusion of an Ex-PRESS glaucoma shunt presenting as an intraocular foreign body. Archives of ophthalmology (Chicago, Ill : 1960) 2005, 123(9):1270–1272. Garg SJ, Kanitkar K, Weichel E, Fischer D: Trauma-induced extrusion of an Ex-PRESS glaucoma shunt presenting as an intraocular foreign body. Archives of ophthalmology (Chicago, Ill : 1960) 2005, 123(9):1270–1272.
6.
Zurück zum Zitat Stein JD, Herndon LW, Brent Bond J, Challa P. Exposure of ex-PRESS miniature glaucoma devices: case series and technique for tube shunt removal. J Glaucoma. 2007;16(8):704–6.CrossRefPubMed Stein JD, Herndon LW, Brent Bond J, Challa P. Exposure of ex-PRESS miniature glaucoma devices: case series and technique for tube shunt removal. J Glaucoma. 2007;16(8):704–6.CrossRefPubMed
7.
Zurück zum Zitat Khouri AS, Khan MN, Fechtner RD, Vold SD. Technique for removal of malpositioned ex-PRESS glaucoma device. J Glaucoma. 2014;23(7):435–6.CrossRefPubMed Khouri AS, Khan MN, Fechtner RD, Vold SD. Technique for removal of malpositioned ex-PRESS glaucoma device. J Glaucoma. 2014;23(7):435–6.CrossRefPubMed
9.
Zurück zum Zitat Nyska A, Glovinsky Y, Belkin M, Epstein Y. Biocompatibility of the ex-PRESS miniature glaucoma drainage implant. J Glaucoma. 2003;12(3):275–80.CrossRefPubMed Nyska A, Glovinsky Y, Belkin M, Epstein Y. Biocompatibility of the ex-PRESS miniature glaucoma drainage implant. J Glaucoma. 2003;12(3):275–80.CrossRefPubMed
10.
Zurück zum Zitat Chan JE, Netland PA: EX-PRESS Glaucoma Filtration Device: efficacy, safety, and predictability. Medical devices (Auckland, NZ) 2015, 8:381–388. Chan JE, Netland PA: EX-PRESS Glaucoma Filtration Device: efficacy, safety, and predictability. Medical devices (Auckland, NZ) 2015, 8:381–388.
11.
Zurück zum Zitat Sarkisian SR Jr. The ex-PRESS miniature glaucoma device and the glaucoma surgery treatment algorithm. Clin Exp Ophthalmol. 2011;39(5):381–2.CrossRefPubMed Sarkisian SR Jr. The ex-PRESS miniature glaucoma device and the glaucoma surgery treatment algorithm. Clin Exp Ophthalmol. 2011;39(5):381–2.CrossRefPubMed
12.
Zurück zum Zitat Wang W, Zhang X. Meta-analysis of randomized controlled trials comparing EX-PRESS implantation with trabeculectomy for open-angle glaucoma. PLoS One. 2014;9(6):e100578.CrossRefPubMedPubMedCentral Wang W, Zhang X. Meta-analysis of randomized controlled trials comparing EX-PRESS implantation with trabeculectomy for open-angle glaucoma. PLoS One. 2014;9(6):e100578.CrossRefPubMedPubMedCentral
13.
Zurück zum Zitat Dahan E, Carmichael TR. Implantation of a miniature glaucoma device under a scleral flap. J Glaucoma. 2005;14(2):98–102.CrossRefPubMed Dahan E, Carmichael TR. Implantation of a miniature glaucoma device under a scleral flap. J Glaucoma. 2005;14(2):98–102.CrossRefPubMed
Metadaten
Titel
Impending extrusion of Ex-PRESS shunt treated by shunt-position adjustment: a case report
verfasst von
Yong Ju Song
Sumi Kim
Gil Joong Yoon
Publikationsdatum
01.12.2018
Verlag
BioMed Central
Erschienen in
BMC Ophthalmology / Ausgabe 1/2018
Elektronische ISSN: 1471-2415
DOI
https://doi.org/10.1186/s12886-017-0665-2

Weitere Artikel der Ausgabe 1/2018

BMC Ophthalmology 1/2018 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Ophthalmika in der Schwangerschaft

Die Verwendung von Ophthalmika in der Schwangerschaft und Stillzeit stellt immer eine Off-label-Anwendung dar. Ein Einsatz von Arzneimitteln muss daher besonders sorgfältig auf sein Risiko-Nutzen-Verhältnis bewertet werden. In der vorliegenden …

Operative Therapie und Keimnachweis bei endogener Endophthalmitis

Vitrektomie Originalie

Die endogene Endophthalmitis ist eine hämatogen fortgeleitete, bakterielle oder fungale Infektion, die über choroidale oder retinale Gefäße in den Augapfel eingeschwemmt wird [ 1 – 3 ]. Von dort infiltrieren die Keime in die Netzhaut, den …

Bakterielle endogene Endophthalmitis

Vitrektomie Leitthema

Eine endogene Endophthalmitis stellt einen ophthalmologischen Notfall dar, der umgehender Diagnostik und Therapie bedarf. Es sollte mit geeigneten Methoden, wie beispielsweise dem Freiburger Endophthalmitis-Set, ein Keimnachweis erfolgen. Bei der …

So erreichen Sie eine bestmögliche Wundheilung der Kornea

Die bestmögliche Wundheilung der Kornea, insbesondere ohne die Ausbildung von lichtstreuenden Narben, ist oberstes Gebot, um einer dauerhaften Schädigung der Hornhaut frühzeitig entgegenzuwirken und die Funktion des Auges zu erhalten.   

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.