01.12.2017 | Research article | Ausgabe 1/2017 Open Access

BMC Pediatrics 1/2017

Limitations and opportunities of whole blood bilirubin measurements by GEM premier 4000®

Zeitschrift:
BMC Pediatrics > Ausgabe 1/2017
Autoren:
Li Wang, Arianne Y. K. Albert, Benjamin Jung, Keyvan Hadad, Martha E. Lyon, Melanie Basso

Abstract

Background

Neonatal hyperbilirubinemia has traditionally been screened by either total serum bilirubin or transcutaneous bilirubin. Whole blood bilirubin (TwB) by the GEM Premier 4000® blood gas analyzer (GEM) is a relatively new technology and it provides fast bilirubin results with a small sample volume and can measure co-oximetry and other analytes. Our clinical study was to evaluate the reliability of TwB measured by the GEM and identify analytical and clinical factors that may contribute to possible bias.

Methods

440 consecutive healthy newborn samples that had plasma bilirubin ordered for neonatal hyperbilirubinemia screening were included. TwB was first measured using the GEM, after which the remainder of the blood was spun and plasma neonatal bilirubin was measured using the VITROS 5600® (VITROS).

Results

62 samples (14%) were excluded from analysis due to failure in obtaining GEM results. Passing-Bablok regression suggested that the GEM results were negatively biased at low concentrations of bilirubin and positively biased at higher concentrations relative to the VITROS results (y = 1.43x-61.13). Bland-Altman plots showed an overall negative bias of the GEM bilirubin with a wide range of differences compared to VITROS. Both hemoglobin concentration and hemolysis affected the accuracy of the GEM results. Clinically, male infants had higher mean bilirubin levels, and infants delivered by caesarean section had lower hemoglobin levels. When comparing the number of results below the 40th percentile and above the 95th percentile cut-offs in the Bhutani nomogram which would trigger discharge or treatment, GEM bilirubin exhibited poor sensitivity and poor specificity in contrast to VITROS bilirubin.

Conclusions

An imperfect correlation was observed between whole blood bilirubin measured on the GEM4000® and plasma bilirubin on the VITROS 5600®. The contributors to the observed differences between the two instruments were specimen hemolysis and the accuracy of hemoglobin measurements, the latter of which affects the calculation of plasma-equivalent bilirubin. Additionally, the lack of standardization of total bilirubin calibration particularly in newborn specimens, may also account for some of the disagreement in results.
Literatur
Über diesen Artikel

Weitere Artikel der Ausgabe 1/2017

BMC Pediatrics 1/2017 Zur Ausgabe

Neu im Fachgebiet Pädiatrie

Meistgelesene Bücher aus dem Fachgebiet

2014 | Buch

Pädiatrische Notfall- und Intensivmedizin

Wenn es um pädiatrische Notfälle auf Station oder um schwerst kranke Kinder auf der Intensivstation geht, ist dieses Buch ein verlässlicher Begleiter. Übersichtlich nach Organen eingeteilt, enthält das Kitteltaschenbuch sämtliche essenziellen Themen der Kinderintensivmedizin, von kardiozirkulatorischen bis infektiologischen Erkrankungen.

Autor:
Prof. Dr. Thomas Nicolai

2015 | Buch

Häufige Hautkrankheiten im Kindesalter

Klinik - Diagnose - Therapie

Das Buch bietet für die 30 häufigsten Hauterkrankungen bei Kindern einen Diagnose- und Behandlungsfahrplan: Für jede Erkrankung gibt es präzise Texte, zahlreiche typische klinische Farbfotos sowie Tabellen zu Differentialdiagnosen und zum therapeutischen Prozedere. Für alle Ärzte, die Kinder mit Hauterkrankungen behandeln! 

Herausgeber:
Dietrich Abeck, Hansjörg Cremer

Mail Icon II Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Pädiatrie und bleiben Sie gut informiert – ganz bequem per eMail.

Bildnachweise