Skip to main content
Erschienen in: European Journal of Nuclear Medicine and Molecular Imaging 10/2007

01.10.2007 | Original Article

Prospective comparison of two gamma probes for intraoperative detection of 18F-FDG: in vitro assessment and clinical evaluation in differentiated thyroid cancer patients with iodine-negative recurrence

verfasst von: Chantal Curtet, Thomas Carlier, Eric Mirallié, Caroline Bodet-Milin, Caroline Rousseau, Jacques Barbet, Francoise Kraeber-Bodéré

Erschienen in: European Journal of Nuclear Medicine and Molecular Imaging | Ausgabe 10/2007

Einloggen, um Zugang zu erhalten

Abstract

Purpose

This study was aimed at evaluating the spatial resolution and sensitivity of two hand-held gamma probes. Radioguided surgery was tested in seven patients with iodine-negative differentiated thyroid cancer (DTC) recurrence using 18F-FDG PET.

Methods

Two gamma probes were evaluated: Clerad’s GammaSup with a collimated CsI(Tl) scintillator and Novelec’s Modelo2 with a BGO scintillator. Five measurement tests were performed following the NEMA guidelines (NU3-2004). Radioguided surgery was performed in patients with recurrent DTC and abnormal 18F-FDG uptake on preoperative 18F-FDG PET images. Patients were injected with rTSH 2 days before surgery. A mean activity of 211 MBq of 18F-FDG was injected 60 min before surgery. In vivo and ex vivo counts were recorded for suspected tumours and normal tissue.

Results

Spatial resolution was higher with the CsI(Tl) than with the BGO detector: 20.2–40.6 mm vs 20.6–55.3 mm from 0 to 20 mm depth. Sensitivity in air and water and through side shielding was higher for BGO but the signal-to-noise ratio was 88 and 22 with the BGO compared to 131 and 76 with the CsI(Tl) at 10 and 30 mm depth. Median in vivo SNR (tumour/non-tumour ratio) was 1.8 with both the BGO and the CsI(Tl) detector, while ex vivo ratios of 2.3 and 2.1, respectively, were obtained. Radioguided surgery allowed detection of all of the tumours identified by 18F-FDG PET images.

Conclusion

This study demonstrated the feasibility of high-energy photon detection with a conventional scintillator equipped with a collimator. The CsI(TI) probe detected more true events from background than did the BGO detector during surgery.
Literatur
1.
Zurück zum Zitat Krag DN, Weaver DL, Alex JC, Fairbank JT. Surgical resection and radiolocalization of sentinel node in breast cancer using a gamma probe. Surg Oncol 1993;2:335–40.PubMedCrossRef Krag DN, Weaver DL, Alex JC, Fairbank JT. Surgical resection and radiolocalization of sentinel node in breast cancer using a gamma probe. Surg Oncol 1993;2:335–40.PubMedCrossRef
2.
Zurück zum Zitat Veronesi U, Paganelli G, Galimberti V, Viale G, Zurrida S, Bedoni M, et al. Sentinel-node biopsy to avoid axillary dissection in breast cancer with clinically negative lymph-nodes. Lancet 1997;349:1864–7.PubMedCrossRef Veronesi U, Paganelli G, Galimberti V, Viale G, Zurrida S, Bedoni M, et al. Sentinel-node biopsy to avoid axillary dissection in breast cancer with clinically negative lymph-nodes. Lancet 1997;349:1864–7.PubMedCrossRef
3.
Zurück zum Zitat Travagli JP, Cailleux AF, Ricard M, Baudin E, Caillou B, Parmentier C, et al. Combination of radioiodine (131I) and probe-guided surgery for persistent or recurrent thyroid carcinoma. J Clin Endocrinol Metab 1998;83:2675–80.PubMedCrossRef Travagli JP, Cailleux AF, Ricard M, Baudin E, Caillou B, Parmentier C, et al. Combination of radioiodine (131I) and probe-guided surgery for persistent or recurrent thyroid carcinoma. J Clin Endocrinol Metab 1998;83:2675–80.PubMedCrossRef
4.
Zurück zum Zitat Peltier P, Curtet C, Chatal JF, Le Doussal JM, Daniel G, Aillet G, et al. Radioimmunodetection of medullary thyroid cancer using a bispecific anti-CEA/anti-indium-DTPA antibody and an indium-111-labeled DTPA dimer. J Nucl Med 1993;34:1267–73.PubMed Peltier P, Curtet C, Chatal JF, Le Doussal JM, Daniel G, Aillet G, et al. Radioimmunodetection of medullary thyroid cancer using a bispecific anti-CEA/anti-indium-DTPA antibody and an indium-111-labeled DTPA dimer. J Nucl Med 1993;34:1267–73.PubMed
5.
Zurück zum Zitat Filippi L, Valentini FB, Gossetti B, Gossetti F, De Vincentis G, Scopinaro F, et al. Intraoperative gamma probe detection of head and neck paragangliomas with 111In-pentetreotide: a pilot study. Tumori 2005;91:173–6.PubMed Filippi L, Valentini FB, Gossetti B, Gossetti F, De Vincentis G, Scopinaro F, et al. Intraoperative gamma probe detection of head and neck paragangliomas with 111In-pentetreotide: a pilot study. Tumori 2005;91:173–6.PubMed
6.
Zurück zum Zitat Essner R, Hsueh EC, Haigh PI, Glass EC, Huynh Y, Daghighian F. Application of an [18F]fluorodeoxyglucose-sensitive probe for the intraoperative detection of malignancy. J Surg Res 2001;96:120–6.PubMedCrossRef Essner R, Hsueh EC, Haigh PI, Glass EC, Huynh Y, Daghighian F. Application of an [18F]fluorodeoxyglucose-sensitive probe for the intraoperative detection of malignancy. J Surg Res 2001;96:120–6.PubMedCrossRef
7.
Zurück zum Zitat Zervos EE, Desai DC, DePalatis LR, Soble D, Martin EW. 18F-labeled fluorodeoxyglucose positron emission tomography-guided surgery for recurrent colorectal cancer: a feasibility study. J Surg Res 2001;97:9–13.PubMedCrossRef Zervos EE, Desai DC, DePalatis LR, Soble D, Martin EW. 18F-labeled fluorodeoxyglucose positron emission tomography-guided surgery for recurrent colorectal cancer: a feasibility study. J Surg Res 2001;97:9–13.PubMedCrossRef
8.
Zurück zum Zitat Helal BO, Merlet P, Toubert ME, Franc B, Schvartz C, Gauthier-Koelesnikov H, et al. Clinical impact of 18F-FDG PET in thyroid carcinoma patients with elevated thyroglobulin levels and negative 131I scanning results after therapy. J Nucl Med 2001;42:1464–9.PubMed Helal BO, Merlet P, Toubert ME, Franc B, Schvartz C, Gauthier-Koelesnikov H, et al. Clinical impact of 18F-FDG PET in thyroid carcinoma patients with elevated thyroglobulin levels and negative 131I scanning results after therapy. J Nucl Med 2001;42:1464–9.PubMed
9.
Zurück zum Zitat Pineda JD, Lee T, Ain K, Reynolds JC, Robbins J. Iodine-131 therapy for thyroid cancer patients with elevated thyroglobulin and negative diagnostic scan. J Clin Endocrinol Metab 1995;80:1488–92.PubMedCrossRef Pineda JD, Lee T, Ain K, Reynolds JC, Robbins J. Iodine-131 therapy for thyroid cancer patients with elevated thyroglobulin and negative diagnostic scan. J Clin Endocrinol Metab 1995;80:1488–92.PubMedCrossRef
10.
Zurück zum Zitat Uematsu H, Sadato N, Ohtsubo T, Tsuchida T, Nakamura S, Sugimoto K, et al. Fluorine-18-fluorodeoxyglucose PET versus thallium-201 scintigraphy evaluation of thyroid tumors. J Nucl Med 1998;39:453–9.PubMed Uematsu H, Sadato N, Ohtsubo T, Tsuchida T, Nakamura S, Sugimoto K, et al. Fluorine-18-fluorodeoxyglucose PET versus thallium-201 scintigraphy evaluation of thyroid tumors. J Nucl Med 1998;39:453–9.PubMed
11.
Zurück zum Zitat Frilling A, Tecklenborg K, Gorges R, Weber F, Clausen M, Broelsch EC. Preoperative diagnostic value of [18F] fluorodeoxyglucose positron emission tomography in patients with radioiodine-negative recurrent well-differentiated thyroid carcinoma. Ann Surg 2001;234:804–11.PubMedCrossRef Frilling A, Tecklenborg K, Gorges R, Weber F, Clausen M, Broelsch EC. Preoperative diagnostic value of [18F] fluorodeoxyglucose positron emission tomography in patients with radioiodine-negative recurrent well-differentiated thyroid carcinoma. Ann Surg 2001;234:804–11.PubMedCrossRef
12.
Zurück zum Zitat Grunwald F, Kalicke T, Feine U, Lietzenmayer R, Scheidhauer K, Dietlein M, et al. Fluorine-18 fluorodeoxyglucose positron emission tomography in thyroid cancer: results of a multicentre study. Eur J Nucl Med 1999;26:1547–52.PubMedCrossRef Grunwald F, Kalicke T, Feine U, Lietzenmayer R, Scheidhauer K, Dietlein M, et al. Fluorine-18 fluorodeoxyglucose positron emission tomography in thyroid cancer: results of a multicentre study. Eur J Nucl Med 1999;26:1547–52.PubMedCrossRef
13.
Zurück zum Zitat Hooft L, Hoekstra OS, Deville W, Lips P, Teule GJ, Boers M, et al. Diagnostic accuracy of 18F-fluorodeoxyglucose positron emission tomography in the follow-up of papillary or follicular thyroid cancer. J Clin Endocrinol Metab 2001;86:3779–86.PubMedCrossRef Hooft L, Hoekstra OS, Deville W, Lips P, Teule GJ, Boers M, et al. Diagnostic accuracy of 18F-fluorodeoxyglucose positron emission tomography in the follow-up of papillary or follicular thyroid cancer. J Clin Endocrinol Metab 2001;86:3779–86.PubMedCrossRef
14.
Zurück zum Zitat Gulec SA, Lee J, Morton DL. GLUT-1 expression correlates with overall and post-resection survival in patients with metastatic pulmonary melanoma. Eur J Nucl Med 2005;32 Suppl 1:S113. Gulec SA, Lee J, Morton DL. GLUT-1 expression correlates with overall and post-resection survival in patients with metastatic pulmonary melanoma. Eur J Nucl Med 2005;32 Suppl 1:S113.
15.
Zurück zum Zitat Hooft L, van der Veldt AA, van Diest PJ, Hoekstra OS, Berkhof J, Teule GJ, et al. [18F]fluorodeoxyglucose uptake in recurrent thyroid cancer is related to hexokinase i expression in the primary tumor. J Clin Endocrinol Metab 2005;90:328–34.PubMedCrossRef Hooft L, van der Veldt AA, van Diest PJ, Hoekstra OS, Berkhof J, Teule GJ, et al. [18F]fluorodeoxyglucose uptake in recurrent thyroid cancer is related to hexokinase i expression in the primary tumor. J Clin Endocrinol Metab 2005;90:328–34.PubMedCrossRef
16.
Zurück zum Zitat Chin BB, Patel P, Cohade C, Ewertz M, Wahl R, Ladenson P. Recombinant human thyrotropin stimulation of fluoro-d-glucose positron emission tomography uptake in well-differentiated thyroid carcinoma. J Clin Endocrinol Metab 2004;89:91–5.PubMedCrossRef Chin BB, Patel P, Cohade C, Ewertz M, Wahl R, Ladenson P. Recombinant human thyrotropin stimulation of fluoro-d-glucose positron emission tomography uptake in well-differentiated thyroid carcinoma. J Clin Endocrinol Metab 2004;89:91–5.PubMedCrossRef
17.
Zurück zum Zitat National Electrical Manufacturers Association. NEMA Standards Publication NU 3-2004. Performance measurements and quality control guidelines for non-imaging intraoperative gamma probes. Rosslyn, VA: National Electrical Manufacturers Association, 2004. National Electrical Manufacturers Association. NEMA Standards Publication NU 3-2004. Performance measurements and quality control guidelines for non-imaging intraoperative gamma probes. Rosslyn, VA: National Electrical Manufacturers Association, 2004.
18.
Zurück zum Zitat Zanzonico P, Heller S. The intraoperative gamma probe: basic principles and choices available. Semin Nucl Med 2000;30:33–4.PubMedCrossRef Zanzonico P, Heller S. The intraoperative gamma probe: basic principles and choices available. Semin Nucl Med 2000;30:33–4.PubMedCrossRef
19.
Zurück zum Zitat Daghighian F, Mazziotta JC, Hoffman EJ, Shenderov P, Eshaghian B, Siegel S, et al. Intraoperative beta probe: a device for detecting tissue labeled with positron or electron emitting isotopes during surger. Med Phys 1994;21:153–7.PubMedCrossRef Daghighian F, Mazziotta JC, Hoffman EJ, Shenderov P, Eshaghian B, Siegel S, et al. Intraoperative beta probe: a device for detecting tissue labeled with positron or electron emitting isotopes during surger. Med Phys 1994;21:153–7.PubMedCrossRef
20.
Zurück zum Zitat Hickernell TS, Barber HB, Barrett HH, Woolfenden JM. Dual-detector probe for surgical tumor staging. J Nucl Med 1988;29:1101–6.PubMed Hickernell TS, Barber HB, Barrett HH, Woolfenden JM. Dual-detector probe for surgical tumor staging. J Nucl Med 1988;29:1101–6.PubMed
Metadaten
Titel
Prospective comparison of two gamma probes for intraoperative detection of 18F-FDG: in vitro assessment and clinical evaluation in differentiated thyroid cancer patients with iodine-negative recurrence
verfasst von
Chantal Curtet
Thomas Carlier
Eric Mirallié
Caroline Bodet-Milin
Caroline Rousseau
Jacques Barbet
Francoise Kraeber-Bodéré
Publikationsdatum
01.10.2007
Verlag
Springer-Verlag
Erschienen in
European Journal of Nuclear Medicine and Molecular Imaging / Ausgabe 10/2007
Print ISSN: 1619-7070
Elektronische ISSN: 1619-7089
DOI
https://doi.org/10.1007/s00259-007-0393-6

Weitere Artikel der Ausgabe 10/2007

European Journal of Nuclear Medicine and Molecular Imaging 10/2007 Zur Ausgabe