Skip to main content
Erschienen in: European Journal of Nuclear Medicine and Molecular Imaging 3/2011

01.03.2011 | Original Article

Wide beam reconstruction for half-dose or half-time cardiac gated SPECT acquisitions: optimization of resources and reduction in radiation exposure

verfasst von: Claudio Marcassa, Riccardo Campini, Orazio Zoccarato, Paolo Calza

Erschienen in: European Journal of Nuclear Medicine and Molecular Imaging | Ausgabe 3/2011

Einloggen, um Zugang zu erhalten

Abstract

Purpose

A new iterative reconstruction algorithm (WBR™) has been recently proposed for cardiac single photon emission computed tomography (SPECT). The WBR™ technology is designed to reduce noise, improving lesion identification without affecting the image resolution, allowing SPECT studies with reduced count statistic. This allows for either half-time (HT) or half-dose (HD) cardiac SPECT, with image quality and quantitative data comparable to standard-time (ST) or standard-dose (SD) SPECT. Few data exist on the comparison between conventional filtered backprojection (FBP) and this new algorithm in a clinical setting. The aim of this study was to compare the performance of FBP and WBR™.

Methods

Phantoms studies were performed to compare spatial resolution and contrast recovery with FBP, ordered subset expectation maximization (OSEM) and WBR™. A group of 92 patients, with different cardiac pathology, scheduled for a stress-rest SPECT were studied: 52 patients (group A) were injected with a SD of tracer and underwent both ST and HT SPECT; 40 patients (group B) were injected with a half dose of tracer and underwent ST SPECT and immediately after an additional SPECT at double time/projection (DT), to compensate for the low count statistic. A 2-day 99mTc-sestamibi protocol was used in all patients. SD/ST and HD/DT SPECT were reconstructed with a conventional FBP; SD/HT and HD/ST SPECT were reconstructed with WBR™. The summed stress score (SSS) and summed rest score (SRS) were calculated; the left ventricular ejection fraction (LVEF) was automatically derived.

Results

In group A (SD), no significant differences were observed between ST FBP SPECT and HT WBR™ in SSS (11.1 and 11.7, respectively) and SRS (9.4 and 10.3, respectively, NS). LVEF on rest acquisitions was also comparable (50% on ST SPECT and 49% on HT SPECT, NS); LVEF on post-stress studies in HT SPECT (46%) was lower than ST SPECT (50%), although not statistically significant. In group B (HD), SSS (6.2 in ST and 5.3 in DT) and SRS (4.0 in ST and 3.3 in DT) were also comparable. No differences were documented between ST and DT in rest (47 and 48%, respectively) and stress (48 and 50%, respectively) LVEF.

Conclusion

WBR™ performance and image quality were comparable to those of conventional FBP, allowing for either HT or HD studies. The former allows for an increased patient throughput and optimization of resources. The latter modalities would allow for a significant reduction in both patients’ and operators’ exposure. Further studies are needed to validate the clinical use of this method.
Literatur
1.
Zurück zum Zitat Borges-Neto S, Pagnanelli RA, Shaw LK, Honeycutt E, Shwartz SC, Adams GL, et al. Clinical results of a novel wide beam reconstruction method for shortening scan time of Tc-99m cardiac SPECT perfusion studies. J Nucl Cardiol 2007;14:555–65.CrossRefPubMed Borges-Neto S, Pagnanelli RA, Shaw LK, Honeycutt E, Shwartz SC, Adams GL, et al. Clinical results of a novel wide beam reconstruction method for shortening scan time of Tc-99m cardiac SPECT perfusion studies. J Nucl Cardiol 2007;14:555–65.CrossRefPubMed
2.
Zurück zum Zitat Venero CV, Heller GV, Bateman TM, McGhie AI, Ahlberg AW, Courter D, et al. A multicenter evaluation of a new post-processing method with depth-dependent collimator resolution applied to full-time and half-time acquisitions without and with simultaneously acquired attenuation correction. J Nucl Cardiol 2009;16:714–25.CrossRefPubMed Venero CV, Heller GV, Bateman TM, McGhie AI, Ahlberg AW, Courter D, et al. A multicenter evaluation of a new post-processing method with depth-dependent collimator resolution applied to full-time and half-time acquisitions without and with simultaneously acquired attenuation correction. J Nucl Cardiol 2009;16:714–25.CrossRefPubMed
3.
Zurück zum Zitat DePuey EG, Bommireddipalli S, Clark J, Thompson L, Srour Y. Wide beam reconstruction “quarter-time” gated myocardial perfusion SPECT functional imaging: a comparison to “full-time” ordered subset expectation maximum. J Nucl Cardiol 2009;16:736–52.CrossRefPubMed DePuey EG, Bommireddipalli S, Clark J, Thompson L, Srour Y. Wide beam reconstruction “quarter-time” gated myocardial perfusion SPECT functional imaging: a comparison to “full-time” ordered subset expectation maximum. J Nucl Cardiol 2009;16:736–52.CrossRefPubMed
4.
Zurück zum Zitat Marcassa C, Campini R, Zoccarato O, Calza P. Wide beam reconstruction for gated-SPECT: going toward optimization of resources and reduction in radiation exposure. Eur J Nucl Med Mol Imaging 2008;35(Suppl 2):S251. Marcassa C, Campini R, Zoccarato O, Calza P. Wide beam reconstruction for gated-SPECT: going toward optimization of resources and reduction in radiation exposure. Eur J Nucl Med Mol Imaging 2008;35(Suppl 2):S251.
5.
Zurück zum Zitat DePuey G, Bommireddipalli S. Half-dose myocardial perfusion SPECT with wide beam reconstruction. Circulation 2009;120:S334CrossRef DePuey G, Bommireddipalli S. Half-dose myocardial perfusion SPECT with wide beam reconstruction. Circulation 2009;120:S334CrossRef
6.
Zurück zum Zitat Hesse B, Tägil K, Cuocolo A, Anagnostopoulos C, Bardiés M, Bax J, et al. EANM/ESC procedural guidelines for myocardial perfusion imaging in nuclear cardiology. Eur J Nucl Med Mol Imaging 2005;32(7):855–97.CrossRefPubMed Hesse B, Tägil K, Cuocolo A, Anagnostopoulos C, Bardiés M, Bax J, et al. EANM/ESC procedural guidelines for myocardial perfusion imaging in nuclear cardiology. Eur J Nucl Med Mol Imaging 2005;32(7):855–97.CrossRefPubMed
7.
Zurück zum Zitat Germano G, Kiat H, Kavanagh PB, Moriel M, Mazzanti M, Su HT, et al. Automatic quantification of ejection fraction from gated myocardial perfusion SPECT. J Nucl Med 1995;36:2138–47.PubMed Germano G, Kiat H, Kavanagh PB, Moriel M, Mazzanti M, Su HT, et al. Automatic quantification of ejection fraction from gated myocardial perfusion SPECT. J Nucl Med 1995;36:2138–47.PubMed
8.
Zurück zum Zitat Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurements. Lancet 1986;1(8476):307–10.PubMed Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurements. Lancet 1986;1(8476):307–10.PubMed
9.
Zurück zum Zitat Zoccarato O, Campini R, Marcassa C, Calza P. Comparison between filtered back projection SPECT reconstruction and a new iterative reconstruction algorithm: a phantom study. Eur J Nucl Med Mol Imaging 2007;34(Suppl 2):S237. Zoccarato O, Campini R, Marcassa C, Calza P. Comparison between filtered back projection SPECT reconstruction and a new iterative reconstruction algorithm: a phantom study. Eur J Nucl Med Mol Imaging 2007;34(Suppl 2):S237.
10.
Zurück zum Zitat Zoccarato O, Campini R, Marcassa C, Calza P. Comparison between filtered back projection SPECT reconstruction and a new iterative reconstruction algorithm: an anthropomorphic cardiac phantom study. Eur J Nucl Med Mol Imaging 2007;34(Suppl 2):S280. Zoccarato O, Campini R, Marcassa C, Calza P. Comparison between filtered back projection SPECT reconstruction and a new iterative reconstruction algorithm: an anthropomorphic cardiac phantom study. Eur J Nucl Med Mol Imaging 2007;34(Suppl 2):S280.
11.
Zurück zum Zitat Zoccarato O, Campini R, Marcassa C, Calza P. Performance of a new iterative reconstruction algorithm for cardiac short-time SPECT: preliminary results in an anthropomorphic cardiac phantom study. Comput Cardiol 2008;35:329–332. Zoccarato O, Campini R, Marcassa C, Calza P. Performance of a new iterative reconstruction algorithm for cardiac short-time SPECT: preliminary results in an anthropomorphic cardiac phantom study. Comput Cardiol 2008;35:329–332.
12.
Zurück zum Zitat Marcassa C, Bax JJ, Bengel F, Hesse B, Petersen L, Reyes E, et al. Clinical value, cost-effectiveness, and safety of myocardial perfusion scintigraphy: a position statement. Eur Heart J 2008;29:557–63.CrossRefPubMed Marcassa C, Bax JJ, Bengel F, Hesse B, Petersen L, Reyes E, et al. Clinical value, cost-effectiveness, and safety of myocardial perfusion scintigraphy: a position statement. Eur Heart J 2008;29:557–63.CrossRefPubMed
Metadaten
Titel
Wide beam reconstruction for half-dose or half-time cardiac gated SPECT acquisitions: optimization of resources and reduction in radiation exposure
verfasst von
Claudio Marcassa
Riccardo Campini
Orazio Zoccarato
Paolo Calza
Publikationsdatum
01.03.2011
Verlag
Springer-Verlag
Erschienen in
European Journal of Nuclear Medicine and Molecular Imaging / Ausgabe 3/2011
Print ISSN: 1619-7070
Elektronische ISSN: 1619-7089
DOI
https://doi.org/10.1007/s00259-010-1654-3

Weitere Artikel der Ausgabe 3/2011

European Journal of Nuclear Medicine and Molecular Imaging 3/2011 Zur Ausgabe