The second ESGAR consensus statement on CT colonography

verfasst von: Emanuele Neri, Steve Halligan, Mikael Hellström, Philippe Lefere, Thomas Mang, Daniele Regge, Jaap Stoker, Stuart Taylor, Andrea Laghi, ESGAR CT Colonography Working Group

Erschienen in: European Radiology | Ausgabe 3/2013

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Abstract

Objective

To update quality standards for CT colonography based on consensus among opinion leaders within the European Society of Gastrointestinal and Abdominal Radiology (ESGAR).

Material and methods

A multinational European panel of nine members of the ESGAR CT colonography Working Group (representing six EU countries) used a modified Delphi process to rate their level of agreement on a variety of statements pertaining to the acquisition, interpretation and implementation of CT colonography. Four Delphi rounds were conducted, each at 2 months interval.

Results

The panel elaborated 86 statements.
In the final round the panelists achieved complete consensus in 71 of 86 statements (82 %). Categories including the highest proportion of statements with excellent Cronbach's internal reliability were colon distension, scan parameters, use of intravenous contrast agents, general guidelines on patient preparation, role of CAD and lesion measurement.
Lower internal reliability was achieved for the use of a rectal tube, spasmolytics, decubitus positioning and number of CT data acquisitions, faecal tagging, 2D vs. 3D reading, and reporting.

Conclusion

The recommendations of the consensus should be useful for both the radiologist who is starting a CTC service and for those who have already implemented the technique but whose practice may need updating.

Key Points

Computed tomographic colonography is the optimal radiological method of assessing the colon
This article reviews ESGAR quality standards for CT colonography
This article is aimed to provide CT-colonography guidelines for practising radiologists
The recommendations should help radiologists who are starting/updating their CTC services

Introduction

Since its introduction (in 1994) [1], clinical implementation of computed tomography (CT) colonography has been governed by advances in CT technology, improvements in dedicated analysis software, development of patient preparation regimens and local diagnostic policies.
In 2007 the European Society of Gastrointestinal and Abdominal Radiology (ESGAR) consensus statement on CT colonography was published, detailing how best to conduct and interpret the examination [2]. That document was based on collective experience up to the beginning of 2006, and the authors represented the EU countries in which CTC underwent consistent clinical implementation (UK, Italy, Belgium and The Netherlands). Over the last 5 years expansion of the CT colonography literature has continued and several important studies, including multicentre studies, have been published [35]. These new data have provided further insight regarding optimisation of the CT colonography technique, interpretation and diagnostic capabilities. Indeed CT colonography is now recommended for colorectal cancer screening by several international groupings and is widely used to investigate patients with symptoms suggestive of colorectal cancer [6, 7]. Although recent review articles provide some guidance regarding the optimal CT colonography technique, given the evolving data [811] there is a current need to update the ESGAR consensus document.
The purpose of this article is therefore to update quality standards for CT colonography based on examination of the existing literature and expert opinion from key opinion-leaders within the European Society of Gastrointestinal and Abdominal Radiology.

Materials and methods

Consensus panel

A multinational European panel of nine members of the ESGAR CTC Working Group (comprising J.S., S.H., S.T., P.L., T.M., D.R., M.H., A.L., E.N., and representing six EU countries: Austria, Belgium, Italy, The Netherlands, Sweden and the UK) used a modified Delphi process [12, 13]. The Delphi process consists of a survey conducted in two or more rounds; the answers (or statements) collected in the first survey are modified in the second, the third, etc., to reach the maximum consensus among the experts. We rated the level of agreement among the experts on a variety of statements pertaining to the acquisition, interpretation and implementation of CT colonography. Four Delphi rounds were conducted, each at 2 months interval.
One of the panellists was chosen as the facilitator (E.N.).
In the first round the facilitator emailed a questionnaire with 22 items pertaining to panel members’ personal approaches to CTC, including items on patient preparation, data acquisition technique, image interpretation and clinical implementation (Table 1). Responses collected from all panellists were merged into a unique datasheet that served to identify areas of agreement and conflict in panellist opinion.
Table 1
Second ESGAR CT colonography consensus. Survey of the first Delphi round
https://static-content.springer.com/image/art%3A10.1007%2Fs00330-012-2632-x/MediaObjects/330_2012_2632_Tab1_HTML.gif
In the second round, the panellists attended a 1-day, face-to-face meeting, and, on the basis of their main areas of research and expertise, were divided into four working groups (WG) as follows: bowel preparation and tagging (WG 1), insufflation and scanning protocols (WG 2), reading paradigm (WG 3) and reporting (WG 4). Each WG independently drafted a cluster of statements pertaining to their allocated subject (Table 2). Each statement was built on the basis of panelists’ expertise and available indexed literature. Each WG then presented their proposed statements to the whole panel for consideration and subsequent discussion, during which time the content and wording of statements were modified until a general consensus emerged.
Table 2
Statements elaborated by the panellists in the second Delphi round, and discussed in the third and fourth to reach the maximum consensus and Cronbach's internal reliability. Statements with score between 4 and 5 are highlighted to show the situations in which all panellists agreed on the statement but the level of support differed (i.e. “agree somewhat” versus “agree strongly”)
https://static-content.springer.com/image/art%3A10.1007%2Fs00330-012-2632-x/MediaObjects/330_2012_2632_Tab2a_HTML.gif https://static-content.springer.com/image/art%3A10.1007%2Fs00330-012-2632-x/MediaObjects/330_2012_2632_Tab2b_HTML.gif https://static-content.springer.com/image/art%3A10.1007%2Fs00330-012-2632-x/MediaObjects/330_2012_2632_Tab2c_HTML.gif https://static-content.springer.com/image/art%3A10.1007%2Fs00330-012-2632-x/MediaObjects/330_2012_2632_Tab2d_HTML.gif
In the third and fourth rounds, copies of the latest statements were sent by email to panellists, who then indicated independently their level of agreement with each individual statement using a 5-point scale, as follows: 1, strongly disagree with the statement; 2, disagree somewhat with the statement; 3, undecided; 4, agree somewhat with the statement; 5, strongly agree with the statement.
After the third round the facilitator collected panellists’ ratings and determined the agreement score for each statement. If the mean score for an individual item was lower than four (maximum possible = five) the facilitator asked panelists to review the statement and attempt to reach a consensus in the fourth round.

Statistical analysis

To measure the internal consistency of panellist’s ratings for each statement, a quality analysis was performed using Cronbach's α correlation coefficient and SPSS (SPSS, Chicago, Ill.) [14]. Cronbach's α was determined after each round.
Cronbach’s α reliability coefficient normally ranges between 0 and 1. The closer the Cronbach’s α coefficient is to 1.0, the greater the internal consistency of the item. An α coefficient > 0.9 was considered excellent, α > 0.8 good, α > 0.7 acceptable, α > 0.6 questionable, α > 0.5 poor and α < 0.5 unacceptable. For the iterations, an α of 0.8 was considered a reasonable goal for internal reliability. All panellist ratings for each statement were also analysed with descriptive statistics, estimating the mean, maximum and minimum score, and their standard deviation.
A mean score of 4 was considered to represent “good” agreement between panellists, a score of 5 “complete” agreement.

Results

Based on the questionnaire provided by the facilitator, the panel elaborated 86 statements that were collected by the facilitator and organised into nine groups, as follows: (1) rectal tube, (2) spasmolytics, (3) colon distension, (4) image acquisition, (5) patient preparation, (6) faecal tagging, (7) reading paradigm, (8) lesion measurement and (9) reporting (Table 2).
In the third round the panelists achieved complete consensus (i.e. mean score 5) in 64 of 86 statements (75 %), which improved to 71 (82 %) in the fourth round (Table 2).
Categories including the highest proportion of statements achieving excellent internal reliability (i.e. Cronbach's α value >0.7) in the final round were colon distension, scan parameters, use of intravenous contrast medium, general guidelines on patient preparation, role of CAD and lesion measurement.
Lower internal reliability was achieved for statements regarding the use of a rectal tube, spasmolytics, decubitus positioning and number of CT data acquisitions, faecal tagging, 2D vs. 3D reading and reporting. However, in the last round, no panellist scored their individual statements as less than 4 on the 5-point rating scale. This indicates that all panellists agreed on the statement but the level of support differed (i.e. “agree somewhat” versus “agree strongly”).

Discussion

Full consensus was reached by our expert panel in 82 % of the statements. In the remaining statements, full consensus was not reached but all panellists achieved a “good” level of agreement. In total, the panellists completed fours rounds; the first and second rounds served to elaborate the basic statements. The third and fourth rounds contained the core of the discussion and were necessary to reach the maximum consensus possible, so creating an optimised, homogeneous opinion for each statement.
All panellists exhibited a high level of agreement for the technical performance of CTC, with clear recommendations regarding colon distension, CT parameters, use of intravenous contrast agents and patient preparation. Full agreement was also reached regarding the role of CAD and lesion measurement. These data reflect a general homogeneity of approach between panel members despite their wide geographical spread. All panel members are regular tutors on the ESGAR CTC course, which may have increased their level of agreement; there is a tendency to promote a common message during panel discussions occurring during the ESGAR CTC courses [15, 16]. Furthermore, in these areas the indexed literature is relatively mature and stable; for example available data supporting the use of automated CO2 for optimal colonic distension is relatively consistent [1720].
However, certain aspects of practice achieved less than “full” agreement. In particular, a digital rectal examination, before insertion of the rectal tube (if rectal examination had not been performed previously), was not standard practice in many centres, but was nevertheless recommended by some panellists (with a mean score 4.56). This difference could be explained by the practice to perform a digital rectal examination before CTC amongst a few of the experts involved in the consensus. Similarly, practice differed regarding the use of intravenous spasmolytics, with many administering such agents to all patients, whereas some (in Italy) only used it in selected individuals [21, 22]. Accordingly, use of spasmolytics is recommended by the majority but is not considered mandatory.
There were minor variations in recommended CT parameters between panellists but all recommended data acquistion in at least two patient positions, without any overall preference regarding the order of acquisitions (i.e. supine or prone first). The differences in CT protocols included the need for additional CT data acquisition and insufflation in cases of poor colonic distension; a minority of experts did not consider this mandatory although they agreed it should be recommended. An additional decubitus acquisition was recommended, if required, to improve the diagnostic quality of the examination [23, 24].
Although available CT technology differed among panellists, all agreed that 2.5-mm collimation was the maximum permissible (although thinner collimation is recommended when available) and use of low radiation dose protocols is to be employed when the overriding purpose of the study is the evaluation of the colonic lumen, for example as in screening [25, 26]. A low radiation dose should be considered a study in which the median effective dose is lower than 5.7 mSv, according to the results of the survey by Leidenbaum et al. [26]. For the staging of patients with known malignancy all the panellists agreed upon the use of standard-dose protocols and intravenous contrast medium [27, 28].
Substantial agreement was reached between panelists regarding the reading methods for interpretation of CT colonography. A combination of 2D and 3D reading was emphasised. Most of the panel were primary 2D readers but all recognised the importance of 3D integration, noting the range of different three-dimensional approaches available. The need for the reader to be adequately trained before interpreting CT colonography was emphasised and is strongly supported by the indexed literature [2933].
Computer-aided diagnosis was acknowledged by all panellists as a potentially useful tool for CTC interpretation, if employed in a second reader paradigm. Accordingly, the use of CAD was recommended provided that readers have already undergone adequate training in general CT colonography interpretation so that they can discriminate between true- and false-positive CAD marks appropriately [3442].
Panellists acknowledged that accurate polyp measurement is problematic for both CTC and endoscopy, with some evidence that CTC may be the superior technique [43, 44]. Despite this advantage, it is still uncertain whether a 2D or a 3D measurement should be made from CT. Moreover, the accuracy of such measurements has important clinical implications for the correct classification and risk stratification of lesions, influencing subsequent recommendations for patient management [4550]. The panel concluded that the maximal diameter of lesions should be primarily estimated using axial and MPR 2D views (which were considered to be the most reliable), avoiding a narrow CT window. Some caution should be exercised when measurements are taken using 3D perspectives given the potential for distortion generated by the three-dimensional endoluminal rendering [5155].
All panellists agreed that CTC should only be reported by a radiologist, and then only after adequate training [5659]. Motivations behind this recommendation are mainly the medico-legal implications of non-radiologists reporting CTC in EU countries. In all EU countries the radiological report is definitively validated by the radiologist despite, in a few centres, a preliminary reading being performed by a radiographer. Adequate training means having interpreted a minimum amount of colonoscopy-verified cases. Although the precise number has not yet been clearly defined, the literature shows that 175 is even not sufficient for several individuals [60, 61].
It was acknowledged that diagnostic accuracy is lower for polyps with a maximal diameter less than 6 mm [3, 4] but if detected with high confidence, and particularly if more than three in number, such polyps should still be reported. This contrasts with recommendations from the CT Colonography Reporting and Data System (C-RADS), authored by Zalis et al., where lesions less than 6 mm are considered diminutive and the recommendation is that they should not be reported [45]. The panel agreed that the patient’s risk (age, family history of colorectal cancer, previous polypectomy, etc.), as well as the number of diminutive lesions detected, should be considered in the decision to report them or not.
There was little disagreement between panellists regarding the need to calibrate the laxative effect of bowel preparation/purgation to the individual patient and potential target lesion. All panellists agreed that faecal tagging should be used routinely. Different preferences for specific laxative and tagging agents were expressed (for example sodium phosphate, magnesium citrate, polyethylene glycol for cleansing, and barium, iodine or a combination of both agents for tagging), reflecting local practice [6275].
In summary, the panel covered all important aspects regarding the practice of CTC and reached full agreement on most statements. The Consensus has been structured to give clear guidelines for the practice of CT colonography. The recommendations should be useful for both the radiologist who is starting a CTC service and for those who have already implemented the technique but whose practice may need updating in the light of recent developments.

Open Access

This article is distributed under the terms of the Creative Commons Attribution Noncommercial License which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.
Open AccessThis is an open access article distributed under the terms of the Creative Commons Attribution Noncommercial License (https://​creativecommons.​org/​licenses/​by-nc/​2.​0), which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Vining DJ, Celfand DW, Bechtold RE, Scharling ES, Grishaw EK, Shifrin RY (1994) Technical feasibility of colon imaging with helical CT and virtual reality. AJR 162:104 Vining DJ, Celfand DW, Bechtold RE, Scharling ES, Grishaw EK, Shifrin RY (1994) Technical feasibility of colon imaging with helical CT and virtual reality. AJR 162:104
2.
Zurück zum Zitat Taylor SA, Laghi A, Lefere P, Halligan S, Stoker J (2007) European Society of Gastrointestinal and Abdominal Radiology (ESGAR): consensus statement on CT colonography. Eur Radiol 17:575–579PubMedCrossRef Taylor SA, Laghi A, Lefere P, Halligan S, Stoker J (2007) European Society of Gastrointestinal and Abdominal Radiology (ESGAR): consensus statement on CT colonography. Eur Radiol 17:575–579PubMedCrossRef
3.
Zurück zum Zitat Johnson CD, Chen MH, Toledano AY et al (2008) Accuracy of CT colonography for detection of large adenomas and cancers. N Engl J Med 359:1207–1217PubMedCrossRef Johnson CD, Chen MH, Toledano AY et al (2008) Accuracy of CT colonography for detection of large adenomas and cancers. N Engl J Med 359:1207–1217PubMedCrossRef
4.
Zurück zum Zitat Regge D, Laudi C, Galatola G et al (2009) Diagnostic accuracy of computed tomographic colonography for the detection of advanced neoplasia in individuals at increased risk of colorectal cancer. JAMA 17:2453–2461CrossRef Regge D, Laudi C, Galatola G et al (2009) Diagnostic accuracy of computed tomographic colonography for the detection of advanced neoplasia in individuals at increased risk of colorectal cancer. JAMA 17:2453–2461CrossRef
5.
Zurück zum Zitat Graser A, Stieber P, Nagel D et al (2009) Comparison of CT colonography, colonoscopy, sigmoidoscopy and faecal occult blood tests for the detection of advanced adenoma in an average risk population. Gut 58:241–248PubMedCrossRef Graser A, Stieber P, Nagel D et al (2009) Comparison of CT colonography, colonoscopy, sigmoidoscopy and faecal occult blood tests for the detection of advanced adenoma in an average risk population. Gut 58:241–248PubMedCrossRef
6.
Zurück zum Zitat Levin B, Lieberman DA, McFarland B et al (2008) American Cancer Society Colorectal Cancer Advisory Group; US Multi-Society Task Force; American College of Radiology Colon Cancer Committee. Screening and surveillance for the early detection of colorectal cancer and adenomatous polyps, 2008: a joint guideline from the American Cancer Society, the US Multi-Society Task Force on Colorectal Cancer, and the American College of Radiology. CA Cancer J Clin 58:130–160PubMedCrossRef Levin B, Lieberman DA, McFarland B et al (2008) American Cancer Society Colorectal Cancer Advisory Group; US Multi-Society Task Force; American College of Radiology Colon Cancer Committee. Screening and surveillance for the early detection of colorectal cancer and adenomatous polyps, 2008: a joint guideline from the American Cancer Society, the US Multi-Society Task Force on Colorectal Cancer, and the American College of Radiology. CA Cancer J Clin 58:130–160PubMedCrossRef
7.
Zurück zum Zitat McFarland EG, Fletcher JG, Pickhardt P et al (2009) American College of Radiology. ACR Colon Cancer Committee white paper: status of CT colonography 2009. J Am Coll Radiol 6:756–772PubMedCrossRef McFarland EG, Fletcher JG, Pickhardt P et al (2009) American College of Radiology. ACR Colon Cancer Committee white paper: status of CT colonography 2009. J Am Coll Radiol 6:756–772PubMedCrossRef
8.
Zurück zum Zitat Mang T, Graser A, Schima W, Maier A (2007) CT colonography: techniques, indications, findings. Eur J Radiol 61:388-399PubMedCrossRef Mang T, Graser A, Schima W, Maier A (2007) CT colonography: techniques, indications, findings. Eur J Radiol 61:388-399PubMedCrossRef
9.
Zurück zum Zitat Yee J, Rosen MP, Blake MA (2010) ACR Appropriateness Criteria on colorectal cancer screening. J Am Coll Radiol 7:670–678PubMedCrossRef Yee J, Rosen MP, Blake MA (2010) ACR Appropriateness Criteria on colorectal cancer screening. J Am Coll Radiol 7:670–678PubMedCrossRef
10.
Zurück zum Zitat Cash BD (2010) Establishing a CT colonography service. Gastrointest Endosc Clin N Am 20:379–398PubMedCrossRef Cash BD (2010) Establishing a CT colonography service. Gastrointest Endosc Clin N Am 20:379–398PubMedCrossRef
11.
Zurück zum Zitat Mang T, Schima W, Brownstone E et al (2011) Consensus statement of the Austrian Society of Radiology, the Austrian Society of Gastroenterology and Hepatology and the Austrian Society of Surgery on CT colonography (Virtual Colonoscopy). Rofo 183:177–184PubMedCrossRef Mang T, Schima W, Brownstone E et al (2011) Consensus statement of the Austrian Society of Radiology, the Austrian Society of Gastroenterology and Hepatology and the Austrian Society of Surgery on CT colonography (Virtual Colonoscopy). Rofo 183:177–184PubMedCrossRef
12.
Zurück zum Zitat Graham B, Regehr G, Wright JG (2003) Delphi as a method to establish consensus for diagnostic criteria. J Clin Epidemiol 56:1150–1156PubMedCrossRef Graham B, Regehr G, Wright JG (2003) Delphi as a method to establish consensus for diagnostic criteria. J Clin Epidemiol 56:1150–1156PubMedCrossRef
13.
Zurück zum Zitat Vakil N (2011) Editorial: consensus guidelines: method or madness? Am J Gastroenterol 106:225–227PubMedCrossRef Vakil N (2011) Editorial: consensus guidelines: method or madness? Am J Gastroenterol 106:225–227PubMedCrossRef
14.
Zurück zum Zitat Cronbach LJ (1951) Coefficient alpha and the internal structure of tests. Psychometrika 16:3CrossRef Cronbach LJ (1951) Coefficient alpha and the internal structure of tests. Psychometrika 16:3CrossRef
15.
Zurück zum Zitat Burling D (2010) International Collaboration for CT Colonography Standards. CT colonography standards. Clin Radiol 65:474–480PubMedCrossRef Burling D (2010) International Collaboration for CT Colonography Standards. CT colonography standards. Clin Radiol 65:474–480PubMedCrossRef
16.
Zurück zum Zitat Boone D, Halligan S, Frost R et al (2011) CT colonography: who attends training? A survey of participants at educational workshops. Clin Radiol 66:510–516PubMedCrossRef Boone D, Halligan S, Frost R et al (2011) CT colonography: who attends training? A survey of participants at educational workshops. Clin Radiol 66:510–516PubMedCrossRef
17.
Zurück zum Zitat Shinners TJ, Pickhardt PJ, Taylor AJ, Jones DA, Olsen CH (2006) Patient-controlled room air insufflation versus automated carbon dioxide delivery for CT colonography. AJR Am J Roentgenol 186:1491–1496PubMedCrossRef Shinners TJ, Pickhardt PJ, Taylor AJ, Jones DA, Olsen CH (2006) Patient-controlled room air insufflation versus automated carbon dioxide delivery for CT colonography. AJR Am J Roentgenol 186:1491–1496PubMedCrossRef
18.
Zurück zum Zitat Burling D, Taylor SA, Halligan S et al (2006) Automated insufflation of carbon dioxide for MDCT colonography: distension and patient experience compared with manual insufflation. AJR Am J Roentgenol 186:96–103PubMedCrossRef Burling D, Taylor SA, Halligan S et al (2006) Automated insufflation of carbon dioxide for MDCT colonography: distension and patient experience compared with manual insufflation. AJR Am J Roentgenol 186:96–103PubMedCrossRef
19.
Zurück zum Zitat Kim SY, Park SH, Choi EK et al (2008) Automated carbon dioxide insufflation for CT colonography: effectiveness of colonic distention in cancer patients with severe luminal narrowing. AJR Am J Roentgenol 190:698–706PubMedCrossRef Kim SY, Park SH, Choi EK et al (2008) Automated carbon dioxide insufflation for CT colonography: effectiveness of colonic distention in cancer patients with severe luminal narrowing. AJR Am J Roentgenol 190:698–706PubMedCrossRef
20.
Zurück zum Zitat Neri E, Laghi A, Regge D (2008) Re: Colonic perforation during screening CT colonography using automated CO2 insufflation in an asymptomatic adult. Abdom Imaging 33:748–749PubMedCrossRef Neri E, Laghi A, Regge D (2008) Re: Colonic perforation during screening CT colonography using automated CO2 insufflation in an asymptomatic adult. Abdom Imaging 33:748–749PubMedCrossRef
21.
Zurück zum Zitat Taylor SA, Halligan S, Goh V et al (2003) Optimizing colonic distention for multi-detector row CT colonography: effect of hyoscine butylbromide and rectal balloon catheter. Radiology 229:99–108PubMedCrossRef Taylor SA, Halligan S, Goh V et al (2003) Optimizing colonic distention for multi-detector row CT colonography: effect of hyoscine butylbromide and rectal balloon catheter. Radiology 229:99–108PubMedCrossRef
22.
Zurück zum Zitat Rogalla P, Lembcke A, Rückert JC et al (2005) Spasmolysis at CT colonography: butyl scopolamine versus glucagon. Radiology 236:184–188PubMedCrossRef Rogalla P, Lembcke A, Rückert JC et al (2005) Spasmolysis at CT colonography: butyl scopolamine versus glucagon. Radiology 236:184–188PubMedCrossRef
23.
Zurück zum Zitat Gryspeerdt SS, Herman MJ, Baekelandt MA, van Holsbeeck BG, Lefere PA (2004) Supine/left decubitus scanning: a valuable alternative to supine/prone scanning in CT colonography. Eur Radiol 14:768–777PubMedCrossRef Gryspeerdt SS, Herman MJ, Baekelandt MA, van Holsbeeck BG, Lefere PA (2004) Supine/left decubitus scanning: a valuable alternative to supine/prone scanning in CT colonography. Eur Radiol 14:768–777PubMedCrossRef
24.
Zurück zum Zitat Buchach CM, Kim DH, Pickhardt PJ (2011) Performing an additional decubitus series at CT colonography. Abdom Imaging 36:538–544PubMedCrossRef Buchach CM, Kim DH, Pickhardt PJ (2011) Performing an additional decubitus series at CT colonography. Abdom Imaging 36:538–544PubMedCrossRef
25.
Zurück zum Zitat Graser A, Wintersperger BJ, Suess C, Reiser MF, Becker CR (2006) Dose reduction and image quality in MDCT colonography using tube current modulation. AJR Am J Roentgenol 187:695–701PubMedCrossRef Graser A, Wintersperger BJ, Suess C, Reiser MF, Becker CR (2006) Dose reduction and image quality in MDCT colonography using tube current modulation. AJR Am J Roentgenol 187:695–701PubMedCrossRef
26.
Zurück zum Zitat Liedenbaum MH, Venema HW, Stoker J (2008) Radiation dose in CT colonography–trends in time and differences between daily practice and screening protocols. Eur Radiol 18:2222–2230PubMedCrossRef Liedenbaum MH, Venema HW, Stoker J (2008) Radiation dose in CT colonography–trends in time and differences between daily practice and screening protocols. Eur Radiol 18:2222–2230PubMedCrossRef
27.
Zurück zum Zitat Filippone A, Ambrosini R, Fuschi M, Marinelli T, Genovesi D, Bonomo L (2004) Preoperative T and N staging of colorectal cancer: accuracy of contrast-enhanced multi-detector row CT colonography–initial experience. Radiology 231:83–90PubMedCrossRef Filippone A, Ambrosini R, Fuschi M, Marinelli T, Genovesi D, Bonomo L (2004) Preoperative T and N staging of colorectal cancer: accuracy of contrast-enhanced multi-detector row CT colonography–initial experience. Radiology 231:83–90PubMedCrossRef
28.
Zurück zum Zitat Mainenti PP, Cirillo LC, Camera L et al (2006) Accuracy of single phase contrast enhanced multidetector CT colonography in the preoperative staging of colo-rectal cancer. Eur J Radiol 60:453–459PubMedCrossRef Mainenti PP, Cirillo LC, Camera L et al (2006) Accuracy of single phase contrast enhanced multidetector CT colonography in the preoperative staging of colo-rectal cancer. Eur J Radiol 60:453–459PubMedCrossRef
29.
Zurück zum Zitat Taylor SA, Halligan S, Slater A et al (2006) Polyp detection with CT colonography: primary 3D endoluminal analysis versus primary 2D transverse analysis with computer-assisted reader software. Radiology 239:759–767PubMedCrossRef Taylor SA, Halligan S, Slater A et al (2006) Polyp detection with CT colonography: primary 3D endoluminal analysis versus primary 2D transverse analysis with computer-assisted reader software. Radiology 239:759–767PubMedCrossRef
30.
Zurück zum Zitat Neri E, Vannozzi F, Vagli P, Bardine A, Bartolozzi C (2006) Time efficiency of CT colonography: 2D vs 3D visualization. Comput Med Imaging Graph 30:175–180PubMedCrossRef Neri E, Vannozzi F, Vagli P, Bardine A, Bartolozzi C (2006) Time efficiency of CT colonography: 2D vs 3D visualization. Comput Med Imaging Graph 30:175–180PubMedCrossRef
31.
Zurück zum Zitat Mang T, Schaefer-Prokop C, Schima W, Maier A et al (2009) Comparison of axial, coronal, and primary 3D review in MDCT colonography for the detection of small polyps: a phantom study. Eur J Radiol 70:86–93PubMedCrossRef Mang T, Schaefer-Prokop C, Schima W, Maier A et al (2009) Comparison of axial, coronal, and primary 3D review in MDCT colonography for the detection of small polyps: a phantom study. Eur J Radiol 70:86–93PubMedCrossRef
32.
Zurück zum Zitat Mang T, Kolligs FT, Schaefer C, Reiser MF, Graser A (2011) Comparison of diagnostic accuracy and interpretation times for a standard and an advanced 3D visualisation technique in CT colonography. Eur Radiol 21:653–662PubMedCrossRef Mang T, Kolligs FT, Schaefer C, Reiser MF, Graser A (2011) Comparison of diagnostic accuracy and interpretation times for a standard and an advanced 3D visualisation technique in CT colonography. Eur Radiol 21:653–662PubMedCrossRef
33.
Zurück zum Zitat Lostumbo A, Wanamaker C, Tsai J, Suzuki K, Dachman AH (2010) Comparison of 2D and 3D views for evaluation of flat lesions in CT colonography. Acad Radiol 17:39–47PubMedCrossRef Lostumbo A, Wanamaker C, Tsai J, Suzuki K, Dachman AH (2010) Comparison of 2D and 3D views for evaluation of flat lesions in CT colonography. Acad Radiol 17:39–47PubMedCrossRef
34.
Zurück zum Zitat Baker ME, Bogoni L, Obuchowski NA et al (2007) Computer-aided detection of colorectal polyps: can it improve sensitivity of less-experienced readers? Preliminary findings. Radiology 245:140–149PubMedCrossRef Baker ME, Bogoni L, Obuchowski NA et al (2007) Computer-aided detection of colorectal polyps: can it improve sensitivity of less-experienced readers? Preliminary findings. Radiology 245:140–149PubMedCrossRef
35.
Zurück zum Zitat Taylor SA, Burling D, Roddie M et al (2008) Computer-aided detection for CT colonography: incremental benefit of observer training. Br J Radiol 81:180–186PubMedCrossRef Taylor SA, Burling D, Roddie M et al (2008) Computer-aided detection for CT colonography: incremental benefit of observer training. Br J Radiol 81:180–186PubMedCrossRef
36.
Zurück zum Zitat Petrick N, Haider M, Summers RM et al (2008) CT colonography with computer-aided detection as a second reader: observer performance study. Radiology 246:148–156PubMedCrossRef Petrick N, Haider M, Summers RM et al (2008) CT colonography with computer-aided detection as a second reader: observer performance study. Radiology 246:148–156PubMedCrossRef
37.
Zurück zum Zitat Taylor SA, Charman SC, Lefere P et al (2008) CT colonography: investigation of the optimum reader paradigm by using computer-aided detection software. Radiology 246:463–471PubMedCrossRef Taylor SA, Charman SC, Lefere P et al (2008) CT colonography: investigation of the optimum reader paradigm by using computer-aided detection software. Radiology 246:463–471PubMedCrossRef
38.
Zurück zum Zitat Regge D, Hassan C, Pickhardt PJ et al (2009) Impact of computer-aided detection on the cost-effectiveness of CT colonography. Radiology 250:488–497PubMedCrossRef Regge D, Hassan C, Pickhardt PJ et al (2009) Impact of computer-aided detection on the cost-effectiveness of CT colonography. Radiology 250:488–497PubMedCrossRef
39.
Zurück zum Zitat de Vries AH, Jensch S, Liedenbaum MH et al (2009) Does a computer-aided detection algorithm in a second read paradigm enhance the performance of experienced computed tomography colonography readers in a population of increased risk? Eur Radiol 19:941–950PubMedCrossRef de Vries AH, Jensch S, Liedenbaum MH et al (2009) Does a computer-aided detection algorithm in a second read paradigm enhance the performance of experienced computed tomography colonography readers in a population of increased risk? Eur Radiol 19:941–950PubMedCrossRef
40.
Zurück zum Zitat Fisichella VA, Jäderling F, Horvath S et al (2009) Computer-aided detection (CAD) as a second reader using perspective filet view at CT colonography: effect on performance of inexperienced readers. Clin Radiol 64:972–982PubMedCrossRef Fisichella VA, Jäderling F, Horvath S et al (2009) Computer-aided detection (CAD) as a second reader using perspective filet view at CT colonography: effect on performance of inexperienced readers. Clin Radiol 64:972–982PubMedCrossRef
41.
Zurück zum Zitat Halligan S, Mallett S, Altman DG et al (2011) Incremental benefit of computer-aided detection when used as a second and concurrent reader of CT colonographic data: multiobserver study. Radiology 258:469–476PubMedCrossRef Halligan S, Mallett S, Altman DG et al (2011) Incremental benefit of computer-aided detection when used as a second and concurrent reader of CT colonographic data: multiobserver study. Radiology 258:469–476PubMedCrossRef
42.
Zurück zum Zitat Dachman AH, Obuchowski NA, Hoffmeister JW et al (2010) Effect of computer-aided detection for CT colonography in a multireader, multicase trial. Radiology 256:827–835PubMedCrossRef Dachman AH, Obuchowski NA, Hoffmeister JW et al (2010) Effect of computer-aided detection for CT colonography in a multireader, multicase trial. Radiology 256:827–835PubMedCrossRef
43.
Zurück zum Zitat Punwani S, Halligan S, Irving P et al (2008) Measurement of colonic polyps by radiologists and endoscopists: who is most accurate? Eur Radiol 18:874–881PubMedCrossRef Punwani S, Halligan S, Irving P et al (2008) Measurement of colonic polyps by radiologists and endoscopists: who is most accurate? Eur Radiol 18:874–881PubMedCrossRef
44.
Zurück zum Zitat Jeong JY, Kim MJ, Kim SS (2008) Manual and automated polyp measurement comparison of CT colonography with optical colonoscopy. Acad Radiol 15:231–239PubMedCrossRef Jeong JY, Kim MJ, Kim SS (2008) Manual and automated polyp measurement comparison of CT colonography with optical colonoscopy. Acad Radiol 15:231–239PubMedCrossRef
45.
Zurück zum Zitat Zalis ME, Barish MA, Choi JR et al (2005) Working Group on Virtual Colonoscopy. CT colonography reporting and data system: a consensus proposal. Radiology 236:3–9PubMedCrossRef Zalis ME, Barish MA, Choi JR et al (2005) Working Group on Virtual Colonoscopy. CT colonography reporting and data system: a consensus proposal. Radiology 236:3–9PubMedCrossRef
46.
Zurück zum Zitat Kim DH, Pickhardt PJ, Taylor AJ (2007) Characteristics of advanced adenomas detected at CT colonographic screening: implications for appropriate polyp size thresholds for polypectomy versus surveillance. AJR Am J Roentgenol 188:940–944PubMedCrossRef Kim DH, Pickhardt PJ, Taylor AJ (2007) Characteristics of advanced adenomas detected at CT colonographic screening: implications for appropriate polyp size thresholds for polypectomy versus surveillance. AJR Am J Roentgenol 188:940–944PubMedCrossRef
47.
Zurück zum Zitat Pickhardt PJ, Hassan C, Laghi A et al (2008) Clinical management of small (6- to 9-mm) polyps detected at screening CT colonography: a cost-effectiveness analysis. AJR Am J Roentgenol 191:1509–1516PubMedCrossRef Pickhardt PJ, Hassan C, Laghi A et al (2008) Clinical management of small (6- to 9-mm) polyps detected at screening CT colonography: a cost-effectiveness analysis. AJR Am J Roentgenol 191:1509–1516PubMedCrossRef
48.
Zurück zum Zitat Shah JP, Hynan LS, Rockey DC (2009) Management of small polyps detected by screening CT colonography: patient and physician preferences. Am J Med 122:687–689PubMed Shah JP, Hynan LS, Rockey DC (2009) Management of small polyps detected by screening CT colonography: patient and physician preferences. Am J Med 122:687–689PubMed
49.
Zurück zum Zitat Heresbach D, Chauvin P, Hess-Migliorretti A, Riou F, Grolier J, Josselin JM (2010) Cost-effectiveness of colorectal cancer screening with computed tomography colonography according to a polyp size threshold for polypectomy. Eur J Gastroenterol Hepatol 22:716–723PubMedCrossRef Heresbach D, Chauvin P, Hess-Migliorretti A, Riou F, Grolier J, Josselin JM (2010) Cost-effectiveness of colorectal cancer screening with computed tomography colonography according to a polyp size threshold for polypectomy. Eur J Gastroenterol Hepatol 22:716–723PubMedCrossRef
50.
Zurück zum Zitat Neri E, Faggioni L, Vagli P et al (2011) Patients' preferences about follow-up of medium size polyps detected at screening CT colonography. Abdom Imaging 36:713–717PubMedCrossRef Neri E, Faggioni L, Vagli P et al (2011) Patients' preferences about follow-up of medium size polyps detected at screening CT colonography. Abdom Imaging 36:713–717PubMedCrossRef
51.
Zurück zum Zitat Pickhardt PJ, Lee AD, McFarland EG, Taylor AJ (2005) Linear polyp measurement at CT colonography: in vitro and in vivo comparison of two-dimensional and three-dimensional displays. Radiology 236:872–878PubMedCrossRef Pickhardt PJ, Lee AD, McFarland EG, Taylor AJ (2005) Linear polyp measurement at CT colonography: in vitro and in vivo comparison of two-dimensional and three-dimensional displays. Radiology 236:872–878PubMedCrossRef
52.
Zurück zum Zitat Burling D, Halligan S, Altman DG et al (2006) Polyp measurement and size categorisation by CT colonography: effect of observer experience in a multi-centre setting. Eur Radiol 16:1737–1744PubMedCrossRef Burling D, Halligan S, Altman DG et al (2006) Polyp measurement and size categorisation by CT colonography: effect of observer experience in a multi-centre setting. Eur Radiol 16:1737–1744PubMedCrossRef
53.
Zurück zum Zitat Park SH, Choi EK, Lee SS et al (2007) Polyp measurement reliability, accuracy, and discrepancy: optical colonoscopy versus CT colonography with pig colonic specimens. Radiology 244:157–164PubMedCrossRef Park SH, Choi EK, Lee SS et al (2007) Polyp measurement reliability, accuracy, and discrepancy: optical colonoscopy versus CT colonography with pig colonic specimens. Radiology 244:157–164PubMedCrossRef
54.
Zurück zum Zitat Park SH, Choi EK, Lee SS et al (2008) Linear polyp measurement at CT colonography: 3D endoluminal measurement with optimized surface-rendering threshold value and automated measurement. Radiology 246:157–167PubMedCrossRef Park SH, Choi EK, Lee SS et al (2008) Linear polyp measurement at CT colonography: 3D endoluminal measurement with optimized surface-rendering threshold value and automated measurement. Radiology 246:157–167PubMedCrossRef
55.
Zurück zum Zitat Bethea E, Nwawka OK, Dachman AH (2009) Comparison of polyp size and volume at CT colonography: implications for follow-up CT colonography. AJR Am J Roentgenol 193:1561–1567PubMedCrossRef Bethea E, Nwawka OK, Dachman AH (2009) Comparison of polyp size and volume at CT colonography: implications for follow-up CT colonography. AJR Am J Roentgenol 193:1561–1567PubMedCrossRef
56.
Zurück zum Zitat Pickhardt PJ (2009) Editorial: CTC interpretation by gastroenterologists: feasible but largely impractical, undesirable, and misguided. Am J Gastroenterol 104:2932–2934PubMedCrossRef Pickhardt PJ (2009) Editorial: CTC interpretation by gastroenterologists: feasible but largely impractical, undesirable, and misguided. Am J Gastroenterol 104:2932–2934PubMedCrossRef
57.
Zurück zum Zitat Carpenter S (2010) Gastroenterologists should read CT colonography. Gastrointest Endosc Clin N Am 20:271–277PubMedCrossRef Carpenter S (2010) Gastroenterologists should read CT colonography. Gastrointest Endosc Clin N Am 20:271–277PubMedCrossRef
58.
Zurück zum Zitat Kim DH, Pickhardt PJ (2010) Radiologists should read CT colonography. Gastrointest Endosc Clin N Am 20:259–69PubMedCrossRef Kim DH, Pickhardt PJ (2010) Radiologists should read CT colonography. Gastrointest Endosc Clin N Am 20:259–69PubMedCrossRef
59.
Zurück zum Zitat Fletcher JG, Chen MH, Herman BA et al (2010) Can radiologist training and testing ensure high performance in CT colonography? Lessons From the National CT Colonography Trial. AJR Am J Roentgenol 195:117–125PubMedCrossRef Fletcher JG, Chen MH, Herman BA et al (2010) Can radiologist training and testing ensure high performance in CT colonography? Lessons From the National CT Colonography Trial. AJR Am J Roentgenol 195:117–125PubMedCrossRef
60.
Zurück zum Zitat Heresbach D, Djabbari M, Riou F et al (2011) Accuracy of computed tomographic colonography in a nationwide multicentre trial, and its relation to radiologist expertise. Gut 60:658–665PubMedCrossRef Heresbach D, Djabbari M, Riou F et al (2011) Accuracy of computed tomographic colonography in a nationwide multicentre trial, and its relation to radiologist expertise. Gut 60:658–665PubMedCrossRef
61.
Zurück zum Zitat Liedenbaum MH, Bipat S, Bossuyt PM et al (2011) Evaluation of a standardized CT colonography training program for novice readers. Radiology 258:477–487PubMedCrossRef Liedenbaum MH, Bipat S, Bossuyt PM et al (2011) Evaluation of a standardized CT colonography training program for novice readers. Radiology 258:477–487PubMedCrossRef
62.
Zurück zum Zitat Lefere PA, Gryspeerdt SS, Dewyspelaere J, Baekelandt M, Van Holsbeeck BG (2002) Dietary fecal tagging as a cleansing method before CT colonography: initial results polyp detection and patient acceptance. Radiology 224:393–403PubMedCrossRef Lefere PA, Gryspeerdt SS, Dewyspelaere J, Baekelandt M, Van Holsbeeck BG (2002) Dietary fecal tagging as a cleansing method before CT colonography: initial results polyp detection and patient acceptance. Radiology 224:393–403PubMedCrossRef
63.
Zurück zum Zitat Iannaccone R, Laghi A, Catalano C et al (2004) Computed tomographic colonography without cathartic preparation for the detection of colorectal polyps. Gastroenterology 127:1300–1311PubMedCrossRef Iannaccone R, Laghi A, Catalano C et al (2004) Computed tomographic colonography without cathartic preparation for the detection of colorectal polyps. Gastroenterology 127:1300–1311PubMedCrossRef
64.
Zurück zum Zitat Gryspeerdt S, Lefere P, Herman M et al (2005) CT colonography with fecal tagging after incomplete colonoscopy. Eur Radiol 15:1192–1202PubMedCrossRef Gryspeerdt S, Lefere P, Herman M et al (2005) CT colonography with fecal tagging after incomplete colonoscopy. Eur Radiol 15:1192–1202PubMedCrossRef
65.
Zurück zum Zitat Lefere P, Gryspeerdt S, Marrannes J, Baekelandt M, Van Holsbeeck B (2005) CT colonography after fecal tagging with a reduced cathartic cleansing and a reduced volume of barium. AJR Am J Roentgenol 184:1836–1842PubMed Lefere P, Gryspeerdt S, Marrannes J, Baekelandt M, Van Holsbeeck B (2005) CT colonography after fecal tagging with a reduced cathartic cleansing and a reduced volume of barium. AJR Am J Roentgenol 184:1836–1842PubMed
66.
Zurück zum Zitat Zalis ME, Perumpillichira JJ, Magee C, Kohlberg G, Hahn PF (2006) Tagging-based, electronically cleansed CT colonography: evaluation of patient comfort and image readability. Radiology 239:149–159PubMedCrossRef Zalis ME, Perumpillichira JJ, Magee C, Kohlberg G, Hahn PF (2006) Tagging-based, electronically cleansed CT colonography: evaluation of patient comfort and image readability. Radiology 239:149–159PubMedCrossRef
67.
Zurück zum Zitat Taylor SA, Slater A, Burling DN et al (2008) CT colonography: optimisation, diagnostic performance and patient acceptability of reduced-laxative regimens using barium-based faecal tagging. Eur Radiol 18:32–42PubMedCrossRef Taylor SA, Slater A, Burling DN et al (2008) CT colonography: optimisation, diagnostic performance and patient acceptability of reduced-laxative regimens using barium-based faecal tagging. Eur Radiol 18:32–42PubMedCrossRef
68.
Zurück zum Zitat Slater A, Planner A, Bungay HK, Bose P, Milburn S (2009) Three-day regimen improves faecal tagging for minimal preparation CT examination of the colon. Br J Radiol 82:545–548PubMedCrossRef Slater A, Planner A, Bungay HK, Bose P, Milburn S (2009) Three-day regimen improves faecal tagging for minimal preparation CT examination of the colon. Br J Radiol 82:545–548PubMedCrossRef
69.
Zurück zum Zitat Neri E, Turini F, Cerri F, Vagli P, Bartolozzi C (2009) CT colonography: same-day tagging regimen with iodixanol and reduced cathartic preparation. Abdom Imaging 34:642–647PubMedCrossRef Neri E, Turini F, Cerri F, Vagli P, Bartolozzi C (2009) CT colonography: same-day tagging regimen with iodixanol and reduced cathartic preparation. Abdom Imaging 34:642–647PubMedCrossRef
70.
Zurück zum Zitat Liedenbaum MH, de Vries AH, Gouw CI et al (2010) CT colonography with minimal bowel preparation: evaluation of tagging quality, patient acceptance and diagnostic accuracy in two iodine-based preparation schemes. Eur Radiol 20:367–376PubMedCrossRef Liedenbaum MH, de Vries AH, Gouw CI et al (2010) CT colonography with minimal bowel preparation: evaluation of tagging quality, patient acceptance and diagnostic accuracy in two iodine-based preparation schemes. Eur Radiol 20:367–376PubMedCrossRef
71.
Zurück zum Zitat Campanella D, Morra L, Delsanto S et al (2010) Comparison of three different iodine-based bowel regimens for CT colonography. Eur Radiol 20:348–358PubMedCrossRef Campanella D, Morra L, Delsanto S et al (2010) Comparison of three different iodine-based bowel regimens for CT colonography. Eur Radiol 20:348–358PubMedCrossRef
72.
Zurück zum Zitat Zueco Zueco C, Sobrido Sampedro C, Corroto JD, Rodriguez Fernández P, Fontanillo Fontanillo M (2012) CT colonography without cathartic preparation: positive predictive value and patient experience in clinical practice. Eur Radiol 22:1195–204PubMedCrossRef Zueco Zueco C, Sobrido Sampedro C, Corroto JD, Rodriguez Fernández P, Fontanillo Fontanillo M (2012) CT colonography without cathartic preparation: positive predictive value and patient experience in clinical practice. Eur Radiol 22:1195–204PubMedCrossRef
73.
Zurück zum Zitat Liedenbaum MH, Denters MJ, de Vries AH et al (2010) Low-fiber diet in limited bowel preparation for CT colonography: Influence on image quality and patient acceptance. AJR Am J Roentgenol 195:31–37CrossRef Liedenbaum MH, Denters MJ, de Vries AH et al (2010) Low-fiber diet in limited bowel preparation for CT colonography: Influence on image quality and patient acceptance. AJR Am J Roentgenol 195:31–37CrossRef
74.
Zurück zum Zitat Davis W, Nisbet P, Hare C, Cooke P, Taylor SA (2011) Non-laxative CT colonography with barium-based faecal tagging: is additional phosphate enema beneficial and well tolerated? Br J Radiol 84:120–125PubMedCrossRef Davis W, Nisbet P, Hare C, Cooke P, Taylor SA (2011) Non-laxative CT colonography with barium-based faecal tagging: is additional phosphate enema beneficial and well tolerated? Br J Radiol 84:120–125PubMedCrossRef
75.
Zurück zum Zitat Liedenbaum MH, Denters MJ, Zijta FM et al (2011) Reducing the oral contrast dose in CT colonography: evaluation of faecal tagging quality and patient acceptance. Clin Radiol 66:30–37PubMedCrossRef Liedenbaum MH, Denters MJ, Zijta FM et al (2011) Reducing the oral contrast dose in CT colonography: evaluation of faecal tagging quality and patient acceptance. Clin Radiol 66:30–37PubMedCrossRef
Metadaten
Titel
The second ESGAR consensus statement on CT colonography
verfasst von
Emanuele Neri
Steve Halligan
Mikael Hellström
Philippe Lefere
Thomas Mang
Daniele Regge
Jaap Stoker
Stuart Taylor
Andrea Laghi
ESGAR CT Colonography Working Group
Publikationsdatum
01.03.2013
Verlag
Springer-Verlag
Erschienen in
European Radiology / Ausgabe 3/2013
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-012-2632-x

Weitere Artikel der Ausgabe 3/2013

European Radiology 3/2013 Zur Ausgabe

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.