Ultrashort echo time MRI of the lung in children and adolescents: comparison with non-enhanced computed tomography and standard post-contrast T1w MRI sequences

verfasst von: Diane M. Renz, Karl-Heinz Herrmann, Martin Kraemer, Joachim Boettcher, Matthias Waginger, Paul-Christian Krueger, Alexander Pfeil, Florian Streitparth, Karim Kentouche, Bernd Gruhn, Jochen G. Mainz, Martin Stenzel, Ulf K. Teichgraeber, Juergen R. Reichenbach, Hans-Joachim Mentzel

Erschienen in: European Radiology | Ausgabe 3/2022

Abstract

Objectives

To compare the diagnostic value of ultrashort echo time (UTE) magnetic resonance imaging (MRI) for the lung versus the gold standard computed tomography (CT) and two T1-weighted MRI sequences in children.

Methods

Twenty-three patients with proven oncologic disease (14 male, 9 female; mean age 9.0 + / − 5.4 years) received 35 low-dose CT and MRI examinations of the lung. The MRI protocol (1.5-T) included the following post-contrast sequences: two-dimensional (2D) incoherent gradient echo (GRE; acquisition with breath-hold), 3D volume interpolated GRE (breath-hold), and 3D high-resolution radial UTE sequences (performed during free-breathing). Images were evaluated by considering image quality as well as distinct diagnosis of pulmonary nodules and parenchymal areal opacities with consideration of sizes and characterisations.

Results

The UTE technique showed significantly higher overall image quality, better sharpness, and fewer artefacts than both other sequences. On CT, 110 pulmonary nodules with a mean diameter of 4.9 + / − 2.9 mm were detected. UTE imaging resulted in a significantly higher detection rate compared to both other sequences (p < 0.01): 76.4% (84 of 110 nodules) for UTE versus 60.9% (67 of 110) for incoherent GRE and 62.7% (69 of 110) for volume interpolated GRE sequences. The detection of parenchymal areal opacities by the UTE technique was also significantly higher with a rate of 93.3% (42 of 45 opacities) versus 77.8% (35 of 45) for 2D GRE and 80.0% (36 of 45) for 3D GRE sequences (p < 0.05).

Conclusion

The UTE technique for lung MRI is favourable in children with generally high diagnostic performance compared to standard T1-weighted sequences as well as CT.
Key Points
Due to the possible acquisition during free-breathing of the patients, the UTE MRI sequence for the lung is favourable in children.
The UTE technique reaches higher overall image quality, better sharpness, and lower artefacts, but not higher contrast compared to standard post-contrast T1-weighted sequences.
In comparison to the gold standard chest CT, the detection rate of small pulmonary nodules small nodules ≤ 4 mm and subtle parenchymal areal opacities is higher with the UTE imaging than standard T1-weighted sequences.
Begleitmaterial
Hinweise

Supplementary Information

The online version contains supplementary material available at https://​doi.​org/​10.​1007/​s00330-021-08236-7.

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
2D
Two-dimensional
3D
Three-dimensional
CNR
Contrast-to-noise ratio
CT
Computed tomography
GRE
Gradient echo
MRI
Magnetic resonance imaging
SD
Standard deviation
SI
Signal intensity
TE
Echo time
TR
Repetition time
UTE
Ultrashort echo time

Introduction

Magnetic resonance imaging (MRI) of the lung is increasingly established in children and adolescents [17]. According to the recently published systematic review of Hirsch et al. [2], lung MRI enables to replace up to 90% of paediatric chest computed tomography (CT) examinations. In contrast to adults, paediatric primary lung tumours are rare with the majority of pulmonary malignancies being of metastatic origin [2, 4, 8, 9]. However, the detection of subtle pulmonary pathologies, particularly tiny nodules, by lung MRI is still challenging [1013]. Sub-centimetre sized pulmonary nodules are mostly benign in general paediatric populations [1416]. Otherwise, even small pulmonary nodules can be malignant in children with proven malignancies, as many paediatric tumours can develop metastases in the lungs [4, 17].
Because children are more sensitive to negative effects of X-ray exposure than adults, one main benefit of lung MRI is the absence of any radiation exposure [25]. Further advantages are the possibility to visualise pathologies by using differently weighted sequences and to provide morphologic as well as functional information [25, 18, 19]. The main issues of lung MRI are low proton density of lung parenchyma and strong magnetic susceptibility effects resulting in short T2 and T2* relaxation times [10, 1822]. Further challenges are motion artefacts due to patient movement, cardiac pulsations or respiration, which are particularly problematic for younger children who have high cardiac pulse and respiratory rates as well as low capabilities to cooperate during breath-hold manoeuvres [2, 4, 23]. T2-weighted lung MRI sequences can be acquired by breath-hold manoeuvres and/ or by respiratory navigator gating; this respiratory navigator gating technique cannot be sufficiently used in T1-weighted sequences due to their short echo (TE) and repetition times (TR) [18, 19].
Standard T1-weighted lung  MRI sequences, currently mainly used in clinical routine, are acquired during breath-hold technique with common slice thicknesses of ≥ 2.5 mm [18, 24, 25]. The recently developed ultrashort echo time (UTE) sequences have very short TE on the order of < 0.1 ms combined with rapid acquisition due to short TR, which makes them suitable to acquire signals from tissues with short T2/T2* relaxation times [18, 2629]. Three-dimensional (3D) UTE techniques allow isotropic resolutions with slice thicknesses around 1 mm and, due to their numerous excitations, can be acquired during free-breathing of the patients [2631]. As far as we know, this is the first study, which has performed a detailed analysis of a 3D UTE sequence of the lung with isotropic voxels in a paediatric oncologic population, including an objective and subjective imaging assessment. The aim of this prospective study was to compare the UTE technique to chest CT, which is still the gold standard, and to two T1-weighted breath-hold sequences (2D incoherent and 3D volume interpolated gradient echo (GRE)), which are currently the most performed standard T1-weighted sequences in clinical routine. The focus of this investigation was set on the diagnosis of challenging, subtle, pulmonary pathologies, including small nodules and ground-glass opacities.

Materials and methods

Study design and patients

The study was planned and performed in accordance with the ethical guidelines of the Declaration of Helsinki and was approved by the local ethics committee (institutional review board number 4562–10/15). The custodians and, in the case of competence (comprehension and intelligence), also the children and adolescents gave consent before participating in the study. This prospective investigation was initiated by the investigators and designed as an intraindividual comparison between three T1-weighted post-contrast MRI sequences and non-enhanced chest CT examinations at one single study centre.
Children and adolescents with proven oncologic disease, who had been referred to clinically indicated MRI examinations with an intravenous application of MR contrast medium as well as to an unenhanced low-dose chest CT, were included. Main clinical indications for the MRI examinations were the diagnosis of lymph nodes and bone metastases. Further inclusion criteria were the performance of the MRI and the CT scans within a maximum of 10 days to provide a reliable comparison, a health status of the patients, which enables to slightly prolong the MRI examinations for study purposes and the possibility to acquire images during breath-hold of the patients. Exclusion criteria were contraindications against MRI and/or MR contrast media, chronic kidney disease ≥ stage 3 (glomerular filtration rate < 60 mL min−1 (1.73 m2)−1), and application of any contrast medium within 24 h before the MRI and/or CT examinations.
Twenty-three patients (14 male, 9 female) with 35 CT and MRI examinations fulfilled the inclusion criteria. The patients suffered from the following oncologic diseases: nephroblastoma (n = 8), acute lymphoblastic leukaemia (n = 4), Ewing’s sarcoma (n = 3), osteosarcoma (n = 2), Hodgkin’s lymphoma (n = 2), rhabdomyosarcoma (n = 2), thyroid carcinoma (n = 1), and mediastinal seminoma (n = 1). The mean age of the patients was 9.0 years (standard deviation, SD, 5.4 years; range 1–17 years). Fourteen of the 35 CT and MRI examinations (40.0%) were acquired under general anaesthesia of the patients, the remaining 21 CT and MRI scans (60.0%) without any sedation or anaesthesia.

CT and MRI protocol

All patients underwent unenhanced low-dose chest CT by using either a 64-slice or a 256-slice multi-detector system (GE LightSpeed VCT 64™ and GE Revolution™; General Electric Healthcare); for CT protocol see electronic supplementary material. Lung MRI was performed on a 1.5 Tesla system (Magnetom Symphony fit™; Siemens Healthineers). All patients underwent the following three T1-weighted MRI sequences after application of gadobutrol (Gadovist®, Bayer Healthcare) with a dose of 0.1 mmol kg−1 body weight: 2D incoherent (spoiled) GRE, 3D volume interpolated GRE, and 3D UTE sequences. The GRE sequences (both with the Cartesian acquisition of the k-space) were performed during breath-hold. The UTE sequence was in-house developed and adapted to lung imaging [28, 29, 32, 33]. The very short echo times of the UTE sequence result in numerous excitations, which makes the sequence robust against motion artefacts [28, 29, 32, 33]. The UTE sequence used 3D radial, centre-out k-space trajectories, making it even more robust against motion, flow, and susceptibility artefacts and allowing an acquisition during free-breathing [28, 29, 32, 33]. Image reconstruction for the UTE images was performed in MATLAB (The MathWorks Inc.) [34]. During the lung MRI examinations of the patients, two T2-weighted sequences were also acquired (see electronic supplementary material). The T1-weighted post-contrast sequences were performed in the same order: 2D incoherent (spoiled) GRE, 3D volume interpolated GRE, and 3D UTE sequences. Table 1 shows the standardised technical parameters of the three sequences.
Table 1
Technical parameters of the T1-weighted magnetic resonance imaging (MRI) sequences
Parameters
Incoherent GRE
Volume interpolated GRE
UTE
Repetition time
(TR) in ms
207.00
3.22
1.10
Echo time (TE)
in ms
2.24
1.06
0.08
Flip angle in °
90
10
6
Acquisition matrix
320
320
240
Field of view in mm
350
350
280
Spatial resolution in mm3
1.50 × 1.10 × 4.00
1.20 × 1.10 × 4.00
1.17 × 1.17 × 1.17
Slice orientation
Axial
Axial
Axial
Fat saturation
Yes
Yes
Yes
Breath-hold
Yes; acquisition by using four concatenations
Yes, acquisition within one breath-hold
No, acquired during free-breathing with two signal averages
Acquisition time
1.02 min
18.0 s
6.20 min
GRE, gradient echo; UTE, ultrashort echo time

Subjective image analysis

All CT and MR images were visually evaluated in consensus by two well-trained radiologists (one general and one paediatric radiologist), each with experience of more than 8 years in chest imaging. The T1-weighted post-contrast MRI sequences and the CT examinations were each separately assessed in an individual blinded session. Subsequently, the reviewers performed a second-look session and compared each of the three MRI sequences with the CT examinations. In this fifth session, the two radiologists assessed whether the pulmonary pathologies could be detected when the CT scans served as a direct comparison. In the case of detection of the pulmonary pathologies in the second-look session, the observers also rated the characteristics of these pathologies.
A structured report was provided for each evaluation session (the imaging evaluation is described in detail in the electronic supplementary material). The image quality of the CT examinations and the three MRI sequences were assessed by considering four criteria: overall image quality, contrast, sharpness, and the presence of artefacts [19, 35, 36]. These imaging parameters were scored on a visual 5-point ordinal Likert scale [19, 35, 36]. The detected pulmonary pathologies, the nodules, and parenchymal areal opacities were described according to current guidelines [3739]; for details, see the electronic supplementary material.

Objective image analysis

For quantitative assessment, regions of interests (ROIs) were drawn on the images of the three T1-weighted MRI sequences within the pulmonary lesions (nodules and areal opacities). ROIs with the same size were placed in the lung parenchyma adjacent to the lesions on the same slice. The ROI sizes were chosen as large as possible, while matching the lesions best avoiding vessels, bronchi as well as hilar structures. For each ROI, the mean and the SD of the signal intensity (SI) were recorded. The CNR of the lesions was computed as follows [40, 41]:
$$\mathrm{CNR}=\left({\mathrm{meanSI}}_{\mathrm{lesion}}-{\mathrm{meanSI}}_{\mathrm{parenchyma}}\right)/{\mathrm{SDSI}}_{\mathrm{parenchyma}}$$

Statistical evaluation

Statistical analysis was performed by using SPSS version 26.0 for Windows (IBM SPSS Statistics). Differences between the image quality parameters of the CT examinations and the MRI sequences were analysed by using Wilcoxon rank sum tests, as the values were not normally distributed according to the results of Shapiro–Wilk tests. McNemar tests evaluated differences in the detection rates of pulmonary nodules and opacities between the MRI sequences. In the case of parameters with normal distribution and interval or ratio scale, paired Student’s t-tests were performed to analyse the differences between the CT and the MRI examinations. p values < 0.05 (two-sided) were considered in all statistical tests to indicate significance.

Results

CT and MR image quality

The results of the image quality assessment are summarised in Table 2. The UTE images scored significantly higher regarding better overall image quality, better sharpness, and fewer artefacts compared to the other T1-weighted sequences. However, the contrast score of UTE imaging was lower with the difference to the incoherent GRE sequence being statistically significant (p = 0.002). Comparing image quality parameters of the three T1-weighted MRI sequences between examinations, which were performed with versus without sedation/ anaesthesia and patients aged < 7 years versus ≥ 7 years old, the values were not significantly different (Wilcoxon rank sum tests, p > 0.05).
Table 2
Image quality parameters of 35 computed tomography (CT) examinations in comparison to the three T1-weighted MRI sequences: incoherent GRE (gradient echo), volume interpolated GRE, and UTE (ultrashort echo time)
Image quality parameter
Chest CT
Incoherent GRE
Volume interpolated GRE
UTE
Mean values of the image quality scores (in parentheses standard deviations)
  Overall image quality
4.4 (0.7)
3.5 (0.8)†,ǂ
3.6 (0.9),ǂ
4.1 (0.8)
  Contrast
4.7 (0.4)
4.3 (0.5),ǂ
4.1 (0.6)
3.9 (0.6)
  Sharpness
4.4 (0.7)
3.5 (0.7),ǂ
3.5 (0.8),ǂ
4.1 (0.9)
  Presence of artefacts
4.1 (0.8)
3.1 (0.8),ǂ
3.4 (0.8),ǂ
3.9 (0.8)
Wilcoxon rank sum tests evaluated differences between each of the MRI sequences in comparison to CT examinations as well as between the three MRI sequences
p < 0.05 indicated statistical significance between the MRI sequence and the CT examinations
ǂ p < 0.05 indicated statistical significance between each of the GRE sequences in comparison to UTE images

Pulmonary nodules

In the 35 chest CT examinations, 110 pulmonary nodules were detected with a mean of the axial average diameter of 4.9 mm (SD 2.9 mm, range 1–21 mm). The range of the distribution was between 0 (in 15 CT examinations), 1 (in 4), and 31 nodules (in 1) with a mean of 3.1 + / − 6.1 nodules per CT examination. Thus, in 16 CT examinations, at least 2 nodules were detected. A table and a histogram showing the distribution of the 110 nodules within the 35 CT examinations are provided in the electronic supplementary material. The overall detection rate of the 110 nodules, found by CT, was 60.9% (67 of 110 nodules) for incoherent GRE and 62.7% (69 of 110) for volume interpolated GRE sequences; significantly more nodules (84 of 110; 76.4%) were observed in UTE images (McNemar tests, p < 0.01). Out of the 4 CT examinations with solely 1 nodule, the UTE technique detected all of the 4 nodules (versus only 1 of 4 nodules was also observed in the incoherent and volume interpolated GRE sequences; axial average diameter 10 mm). The axial average diameters of the 3 only with UTE detected nodules were 4 mm, 5 mm, and 5 mm (Fig. 1). Regarding examinations with ≥ 2 nodules, at least 1 nodule was found in UTE images. Thus, there were no false-negative MRI examinations regarding the UTE technique.
The characteristics of the 110 nodules are described in Table 3. Considering the average diameter of the nodules, more lesions with a size up to 10 mm were found with UTE sequences, whereas the detection rates for nodules with sizes of 1–4 mm differed significantly (see Table 4). Furthermore, nodules, located in the lung periphery, presenting no calcifications, with solid and subsolid appearances as well as with smooth and irregular margins were significantly more often found by UTE imaging (Table 4).
Table 3
Characteristics of 110 pulmonary nodules, detected on computed tomography (CT)
Characteristics
Size group based on the axial average diameter
  1–4 mm
63 (57.3%)
  5–7 mm
32 (29.1%)
  8–10 mm
11 (10.0%)
  > 10 mm
4 (3.6%)
Mediolateral location
  Central
31 (28.2%)
  Peripheral
79 (71.8%)
Appearance
  Solid
85 (77.3%)
  Subsolid
25 (22.7%)
Margin
  Smooth
60 (54.5%)
  Irregular
50 (45.5%)
Presence of calcifications
  Yes
23 (20.9%)
  No
87 (79.1%)
Table 4
Nodule detection rates in the initial analysis with the MRI sequences incoherent GRE (gradient echo), volume interpolated GRE, and UTE (ultrashort echo time) based on the size and the characteristics of the pulmonary nodules
Detection rates
Incoherent GRE
Volume interpolated GRE
UTE
Size group based on the axial average diameter
  1–4 mm
46.0% (29 of 63)*
49.2% (31 of 63)*
68.3% (43 of 63)*
  5–7 mm
78.1% (25 of 32)
78.1% (25 of 32)
84.4% (27 of 32)
  8–10 mm
81.8% (9 of 11)
81.8% (9 of 11)
90.9% (10 of 11)
  > 10 mm
100.0% (4 of 4)
100.0% (4 of 4)
100.0% (4 of 4)
Mediolateral location
  Central
41.9% (13 of 31)
51.6% (16 of 31)
61.3% (19 of 31)
  Peripheral
68.4% (54 of 79)*
67.1% (53 of 79)*
82.3% (65 of 79)*
Appearance
  Solid
67.1% (57 of 85)*
69.4% (59 of 85)*
78.8% (67 of 85)*
  Subsolid
40.0% (10 of 25)*
40.0% (10 of 25)*
68.0% (17 of 25)*
Margin
  Smooth
60.0% (36 of 60)*
61.7% (37 of 60)*
75.0% (45 of 60)*
  Irregular
62.0% (31 of 50)*
64.0% (32 of 50)*
78.0% (39 of 50)*
Presence of calcifications
  Yes
52.2% (12 of 23)
47.8% (11 of 23)
56.5% (13 of 23)
  No
63.2% (55 of 87)*
66.7% (58 of 87)*
81.6% (71 of 87)*
McNemar tests evaluated differences between each of the GRE sequences in comparison to UTE images
*p < 0.05 indicated statistical significance
In the second-look session, additional nodules were diagnosed on lung MRI: 15 in incoherent GRE, 12 in volume interpolated GRE, and 11 in UTE images. Seventy-nine pulmonary nodules were seen on chest CT as well as in all three MRI sequences. The average axial diameter of these 79 lesions was as follows: mean size of 5.5 (SD 3.0) mm on CT, 5.2 (SD 3.0) mm in incoherent GRE, 5.3 (SD 3.0) mm in volume interpolated GRE, and 5.3 (SD 3.0) mm in UTE sequences. The size difference between CT and incoherent GRE sequences was statistically significant (paired Student’s t-test, p = 0.017), between CT and volume interpolated GRE images nearly significant (paired Student’s t-test, p = 0.052).
In the objective image analysis, the CNR values of the 79 nodules were highest in incoherent GRE sequences with a mean value of 23.7 (SD 13.5), which was significantly higher compared to both, volume interpolated GRE (mean 20.7, SD 13.1; paired Student’s t-test, p = 0.009) and UTE sequences (mean 19.7, SD 10.9; paired Student’s t-test, p = 0.013). The mean difference of CNR between volume interpolated GRE and UTE imaging was not significant (paired Student’s t-test, p = 0.532).

Parenchymal areal opacities

Forty-five parenchymal areal opacities were diagnosed on chest CT: 16 consolidations, 10 ground-glass opacities, and 19 parenchymal bands. The detection rates in CT and the three MRI sequences are shown in Table 5 (see also Fig. 2). In the second-look session, five additional parenchymal areal opacities were observed with both, incoherent GRE and volume interpolated GRE, and two additional lesions with UTE imaging. Forty pathologies were detected on chest CT as well as with all three MRI sequences in the blinded and/or second-look sessions. The 40 opacities had the following axial average diameters: mean size of 24.1 (SD 17.0) mm on CT, 23.0 (SD 16.5) mm in incoherent GRE, 23.1 (SD 17.4) mm in volume interpolated GRE, and 23.4 (SD 16.8) mm in UTE sequences. The size differences between CT and incoherent GRE (paired Student’s t-test, p = 0.005) as well as between CT and volume interpolated GRE images (p = 0.025) were statistically significant. The size difference between CT and UTE imaging reached a p value of 0.088 (paired Student’s t-test).
Table 5
Detection rates of parenchymal areal opacities in the initial analysis with the MRI sequences incoherent GRE (gradient echo), volume interpolated GRE, and UTE (ultrashort echo time) based on the types of the opacities
Detection rates
Incoherent GRE
Volume interpolated GRE
UTE
Overall parenchymal areal opacities
 
77.8% (35 of 45)*
80.0% (36 of 45)*
93.3% (42 of 45)*
Consolidations
 
100.0% (16 of 16)
100.0% (16 of 16)
100.0% (16 of 16)
Ground-glass opacities
 
50.0% (5 of 10)
60.0% (6 of 10)
90.0% (9 of 10)
Parenchymal bands
 
73.7% (14 of 19)
73.7% (14 of 19)
89.5% (17 of 19)
McNemar tests evaluated differences between each of the GRE sequences in comparison to UTE images
*p < 0.05 indicated statistical significance
As in the case of pulmonary nodules, the CNR of the 40 opacities, visible in all three MRI sequences, was highest in incoherent GRE sequence with a value of 28.3 (SD 19.0) followed by UTE (mean 25.8, SD 16.3) and volume interpolated GRE sequences (mean 24.4, SD 15.7) in the objective image analysis. The mean difference between incoherent GRE and UTE images was not significant (paired Student’s t-test, p = 0.247), the comparison between the GRE sequences was nearly significant (paired Student’s t-test, p = 0.067). The mean difference of CNR between volume interpolated GRE and UTE imaging was not significant (paired Student’s t-test, p = 0.480).

Discussion

Thin-section UTE imaging is a sensitive technique to detect pulmonary nodules and parenchymal areal opacities in children and adolescents with malignancies. To our knowledge, this is the first study, which compared the UTE sequence with two T1-weighted standard sequences, mostly performed in clinical routine, and to the still gold standard chest CT in a paediatric oncologic population. The image quality parameters of the UTE sequence are higher compared to both GRE sequences, except for the contrast.
The high contrast of the incoherent GRE sequence is a direct consequence of the scan parameters, resulting in high signal even from tissues with longer T1-relaxation times. The ultrashort echo time of the UTE sequence leads to substantial signal gain in the lung parenchyma [27, 42]. Nevertheless, despite this reduction of contrast, which could be mitigated by increasing the TR (albeit at the cost of scan time), the nodular detection rate of UTE was superior compared to the other two MRI sequences, mainly due to the reduced artefacts and the improved lesion sharpness (due to the high isotropic resolution). The mean overall image quality of 4.1 for the UTE sequence in our study is comparable with published findings in adults. Ohno et al. [43] reported a mean overall image quality of 4.4 with 1 mm3 isotropic UTE imaging, using a similar visual 5-point Likert scale. In the study of Cha et al. [36], the UTE sequence resulted in a mean overall image quality of 3.9–4.0 and a mean value of 3.7 for the presence of image artefacts, scored by two readers with a similar 5-point Likert scale.
The significantly higher detection of pulmonary nodules with UTE sequence in our study is also similar to published reports in adults, which compared MRI sequences to chest CT. In the study of Burris et al. [12], the overall detection rate was 73% (60 of 82 nodules) for UTE and 30% (25 of 82 nodules) for dual-echo GRE sequences, whereby the nodules had a mean short-axis diameter of 6.2 mm. Wielpütz et al. [39] reported a detection rate of the UTE sequence of 92.5% for nodules with a relatively high mean maximum long-axis diameter of 17.4 mm. In our investigation, the UTE sequence was most beneficial, compared to GRE images, in diagnosing challenging nodules with sizes up to 4 mm. In the study of Burris et al. [12], the higher detection of the UTE sequence was statistically significant for nodules with diameters ≥ 4 mm, but not for sizes ≥ 8 mm. We detected significantly more solid and subsolid nodules with the UTE sequence compared to the GRE images. Ohno et al. [44] evaluated in adults the performance of UTE imaging compared to chest CT depending on the nodular appearance and found that 35.3% ground-glass, 33.3% partly solid, and 94.7% solid nodules (long-axis diameters of 4–6 mm) were detected with the UTE technique in consensus reading. The UTE sequence is also suited in diagnosing subsolid nodules, which are diagnostically challenging. This higher detection in our study is most likely due to the combination of higher spatial resolution and higher lung signal of the UTE technique. However, the UTE sequence showed no significant benefit in detecting calcified nodules in our study. Calcified nodules remain a diagnostic challenge for MRI due to their low signal intensity [4, 17, 45]. Although most calcified pulmonary nodules are likely to be benign, some bone-forming paediatric tumours, e.g., osteosarcomas, can show metastases with at least some calcifications in the lungs [4, 17, 45].
As described, the UTE sequence missed no patients with pulmonary nodules; out of the 4 CT examinations with solely 1 nodule, the UTE technique detected all of the 4 nodules in the initial reading. Thus, the missing of some nodules by UTE did not have a significant impact on the treatment of the patients. Furthermore, 7 of the 15 in total missed nodules could be visualised on MRI by considering all sequences (with also T2-weighted images). Although the radiation doses of chest CT examinations are quite low by using new technologies, any radiation exposure should be avoided in children if radiation-sensitive organs, e.g., thyroid gland, breasts, red bone marrow, are located in the exposure fields and in the case of tumour predisposition syndromes [2]. The application of contrast media may be favourable for diagnosing mediastinal and bihilar lymph node metastases on MR and CT imaging, whereas enhanced CT examinations result in higher radiation exposures than unenhanced CT imaging. Another advantage of MRI is the distinct diagnosis of bone metastases [3, 5, 45]. As no breath-hold manoeuvres are necessary for the performance of the UTE sequence, the higher detection rates and the possible acquisition without intubation anaesthesia compensate for the higher acquisition time of the UTE sequence compared to the common standard T1-weighted GRE sequences.
Lung MRI is increasingly performed in children for diagnosing pulmonary infections [46, 47]. All consolidations were detected with all three MRI sequences in our study, in contrast to the more subtle ground-glass opacities and parenchymal bands. Due to the low number of these pathologies, the higher detection rate with UTE compared to the GRE images was statistically not significant. The diagnostic challenge of detecting ground-glass opacities by MRI has been reported. Sodhi et al. [46] evaluated an MRI protocol with T2-weighted and volume interpolated GRE sequences and found a high correlation with chest CT for the diagnosis of consolidations, but not for ground-glass opacities in children (age range 5–15 years). Roach et al. [48] compared UTE imaging (acquisition time 5–10 min) with chest CT in patients with early cystic fibrosis lung disease (mean age 31.8 months) and reported Pearson correlation coefficients of 0.45 for diagnosing consolidations and of 0.21 for diagnosing ground-glass opacities. In older patients with cystic fibrosis (mean age 22.6 years), Dournes et al. [49] found that their evaluated UTE sequence showed good to very good agreement with CT regarding all morphologic pulmonary manifestations, such as bronchiectasis, peribronchial thickening, and mosaic pattern. Thus, the UTE technique becomes increasingly important for the diagnosis of subtle parenchymal pathologies.
Since the inclusion criteria required a homogeneous study design, only a limited number of patients could be included in this prospective study. All patients had clinical indications for the administration of MR contrast media; no agent was applied only for the study purpose. The use of contrast medium might have had an impact on the diagnostic performance, as the detection of pulmonary nodules can be improved by possible enhancement [10, 40]. However, MR contrast media must be in general administered judiciously in children and adolescents, due to possible gadolinium deposition in different tissues, such as brain, bone, or liver, and the brain development during childhood and adolescence [50]. The evaluated sequences were acquired at the end of the MRI examinations; thus, image artefacts due to patient movements or less deep anaesthesia may have occurred. Furthermore, a wash-out of the contrast media might have developed. The diagnostic performance of the T1-weighted sequences, particularly the UTE technique, which was performed as the last sequence, could be even higher if they were acquired earlier in the MRI examinations.
In conclusion, the findings of our study indicate that UTE imaging of the lung leads to improved diagnostic performance compared to conventional T1-weighted sequences in children and adolescents, if chest CT served as the gold standard. In particular, the detection rate of small pulmonary nodules ≤ 4 mm is superior with UTE imaging, presumably due to the higher spatial resolution and reduced artefacts. The visibility of subtle, diffuse parenchymal lesions, e.g., ground-glass opacities, in UTE images is promising. Lung MRI may be used instead of chest CT examinations during the staging of children with oncologic diseases, particularly if the diagnosis of lymph node and bone metastases are also important issues.

Acknowledgements

The study was initiated by the investigators and was not supported by any research grants. Thus, there was no conflict of interest regarding the study.

Declarations

Guarantor

The scientific guarantor of this publication is Prof. Dr. Diane M. Renz, M.D.

Conflict of interest

The authors of this manuscript declare no relationships with any companies whose products or services may be related to the subject matter of the article.

Statistics and biometry

Some of the authors have significant statistical expertise. No complex statistical methods were necessary for this paper.
The custodians and, in the case of competence (comprehension and intelligence), also the children and adolescents gave informed consent before participating in the study.

Ethical approval

The study was planned and performed in accordance with the ethical guidelines of the Declaration of Helsinki and was approved by the local ethics committee (institutional review board number 4562–10/15).

Methodology

• prospective
• cross-sectional study
• performed at one institution
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Anhänge

Supplementary Material

Literatur
1.
Zurück zum Zitat Portelli JL, McNulty JP, Bezzina P, Rainford L (2016) Frequency of paediatric medical imaging examinations performed at a European teaching hospital over a 7-year period. Eur Radiol 26:4221–4230PubMedCrossRef Portelli JL, McNulty JP, Bezzina P, Rainford L (2016) Frequency of paediatric medical imaging examinations performed at a European teaching hospital over a 7-year period. Eur Radiol 26:4221–4230PubMedCrossRef
2.
Zurück zum Zitat Hirsch FW, Sorge I, Vogel-Claussen J et al (2020) The current status and further prospects for lung magnetic resonance imaging in pediatric radiology. Pediatr Radiol 50:734–749PubMedPubMedCentralCrossRef Hirsch FW, Sorge I, Vogel-Claussen J et al (2020) The current status and further prospects for lung magnetic resonance imaging in pediatric radiology. Pediatr Radiol 50:734–749PubMedPubMedCentralCrossRef
3.
Zurück zum Zitat Tiddens HAWM, Kuo W, van Straten M, Ciet P (2018) Paediatric lung imaging: the times they are a-changin’. Eur Respir Rev 27:170097 Tiddens HAWM, Kuo W, van Straten M, Ciet P (2018) Paediatric lung imaging: the times they are a-changin’. Eur Respir Rev 27:170097
4.
Zurück zum Zitat Hirsch W, Sorge I, Krohmer S, Weber D, Meier K, Till H (2008) MRI of the lungs in children. Eur J Radiol 68:278–288PubMedCrossRef Hirsch W, Sorge I, Krohmer S, Weber D, Meier K, Till H (2008) MRI of the lungs in children. Eur J Radiol 68:278–288PubMedCrossRef
6.
Zurück zum Zitat Schaefer JF, Hector A, Schmidt K et al (2018) A semiquantitative MRI-score can predict loss of lung function in patients with cystic fibrosis: preliminary results. Eur Radiol 28:74–84PubMedCrossRef Schaefer JF, Hector A, Schmidt K et al (2018) A semiquantitative MRI-score can predict loss of lung function in patients with cystic fibrosis: preliminary results. Eur Radiol 28:74–84PubMedCrossRef
7.
Zurück zum Zitat Zirpoli S, Munari AM, Primolevo A et al (2019) Agreement between magnetic resonance imaging and computed tomography in the postnatal evaluation of congenital lung malformations: a pilot study. Eur Radiol 29:4544–4554PubMedCrossRef Zirpoli S, Munari AM, Primolevo A et al (2019) Agreement between magnetic resonance imaging and computed tomography in the postnatal evaluation of congenital lung malformations: a pilot study. Eur Radiol 29:4544–4554PubMedCrossRef
8.
Zurück zum Zitat Meier-Schroers M, Homsi R, Gieseke J, Schild HH, Thomas D (2019) Lung cancer screening with MRI: evaluation of MRI for lung cancer screening by comparison of LDCT- and MRI-derived Lung-RADS categories in the first two screening rounds. Eur Radiol 29:898–905PubMedCrossRef Meier-Schroers M, Homsi R, Gieseke J, Schild HH, Thomas D (2019) Lung cancer screening with MRI: evaluation of MRI for lung cancer screening by comparison of LDCT- and MRI-derived Lung-RADS categories in the first two screening rounds. Eur Radiol 29:898–905PubMedCrossRef
9.
Zurück zum Zitat Wan Q, Deng YS, Lei Q et al (2019) Differentiating between malignant and benign solid solitary pulmonary lesions: are intravoxel incoherent motion and diffusion kurtosis imaging superior to conventional diffusion-weighted imaging? Eur Radiol 29:1607–1615PubMedCrossRef Wan Q, Deng YS, Lei Q et al (2019) Differentiating between malignant and benign solid solitary pulmonary lesions: are intravoxel incoherent motion and diffusion kurtosis imaging superior to conventional diffusion-weighted imaging? Eur Radiol 29:1607–1615PubMedCrossRef
10.
Zurück zum Zitat Kurihara Y, Matsuoka S, Yamashiro T et al (2014) MRI of pulmonary nodules. AJR Am J Roentgenol 202:W210–W216PubMedCrossRef Kurihara Y, Matsuoka S, Yamashiro T et al (2014) MRI of pulmonary nodules. AJR Am J Roentgenol 202:W210–W216PubMedCrossRef
12.
Zurück zum Zitat Burris NS, Johnson KM, Larson PE et al (2016) Detection of small pulmonary nodules with ultrashort echo time sequences in oncology patients by using a PET/MR system. Radiology 278:239–246PubMedCrossRef Burris NS, Johnson KM, Larson PE et al (2016) Detection of small pulmonary nodules with ultrashort echo time sequences in oncology patients by using a PET/MR system. Radiology 278:239–246PubMedCrossRef
13.
Zurück zum Zitat Renz DM, Scholz O, Böttcher J et al (2015) Comparison between magnetic resonance imaging and computed tomography of the lung in patients with cystic fibrosis with regard to clinical, laboratory, and pulmonary functional parameters. Invest Radiol 50:733–742PubMedCrossRef Renz DM, Scholz O, Böttcher J et al (2015) Comparison between magnetic resonance imaging and computed tomography of the lung in patients with cystic fibrosis with regard to clinical, laboratory, and pulmonary functional parameters. Invest Radiol 50:733–742PubMedCrossRef
14.
Zurück zum Zitat van Riel SJ, Ciompi F, Jacobs C et al (2017) Malignancy risk estimation of screen-detected nodules at baseline CT: comparison of the PanCan model, Lung-RADS and NCCN guidelines. Eur Radiol 27:4019–4029PubMedPubMedCentralCrossRef van Riel SJ, Ciompi F, Jacobs C et al (2017) Malignancy risk estimation of screen-detected nodules at baseline CT: comparison of the PanCan model, Lung-RADS and NCCN guidelines. Eur Radiol 27:4019–4029PubMedPubMedCentralCrossRef
15.
Zurück zum Zitat Samim A, Littooij AS, van den Heuvel-Eibrink MM, Wessels FJ, Nievelstein RAJ, de Jong PA (2017) Frequency and characteristics of pulmonary nodules in children at computed tomography. Pediatr Radiol 47:1751–1758PubMedPubMedCentralCrossRef Samim A, Littooij AS, van den Heuvel-Eibrink MM, Wessels FJ, Nievelstein RAJ, de Jong PA (2017) Frequency and characteristics of pulmonary nodules in children at computed tomography. Pediatr Radiol 47:1751–1758PubMedPubMedCentralCrossRef
16.
Zurück zum Zitat Renne J, Linderkamp C, Wacker F, Berthold LD, Weidemann J (2015) Prevalence and configuration of pulmonary nodules on multi-row CT in children without malignant diseases. Eur Radiol 25:2651–2656PubMedCrossRef Renne J, Linderkamp C, Wacker F, Berthold LD, Weidemann J (2015) Prevalence and configuration of pulmonary nodules on multi-row CT in children without malignant diseases. Eur Radiol 25:2651–2656PubMedCrossRef
17.
Zurück zum Zitat Kusma J, Young C, Yin H, Stanek JR, Yeager N, Aldrink JH (2017) Pulmonary nodule size <5 mm still warrants investigation in patients with osteosarcoma and Ewing sarcoma. J Pediatr Hematol Oncol 39:184–187PubMedCrossRef Kusma J, Young C, Yin H, Stanek JR, Yeager N, Aldrink JH (2017) Pulmonary nodule size <5 mm still warrants investigation in patients with osteosarcoma and Ewing sarcoma. J Pediatr Hematol Oncol 39:184–187PubMedCrossRef
19.
Zurück zum Zitat Scholz O, Denecke T, Böttcher J et al (2017) MRI of cystic fibrosis lung manifestations: sequence evaluation and clinical outcome analysis. Clin Radiol 72:754–763PubMedCrossRef Scholz O, Denecke T, Böttcher J et al (2017) MRI of cystic fibrosis lung manifestations: sequence evaluation and clinical outcome analysis. Clin Radiol 72:754–763PubMedCrossRef
20.
Zurück zum Zitat Behrendt L, Voskrebenzev A, Klimeš F et al (2020) Validation of automated perfusion-weighted phase-resolved functional lung (PREFUL)-MRI in patients with pulmonary diseases. J Magn Reson Imaging 52:103–114PubMed Behrendt L, Voskrebenzev A, Klimeš F et al (2020) Validation of automated perfusion-weighted phase-resolved functional lung (PREFUL)-MRI in patients with pulmonary diseases. J Magn Reson Imaging  52:103–114PubMed
21.
Zurück zum Zitat Martini K, Gygax CM, Benden C, Morgan AR, Parker GJM, Frauenfelder T (2018) Volumetric dynamic oxygen-enhanced MRI (OE-MRI): comparison with CT Brody score and lung function in cystic fibrosis patients. Eur Radiol 28:4037–4047PubMedCrossRef Martini K, Gygax CM, Benden C, Morgan AR, Parker GJM, Frauenfelder T (2018) Volumetric dynamic oxygen-enhanced MRI (OE-MRI): comparison with CT Brody score and lung function in cystic fibrosis patients. Eur Radiol 28:4037–4047PubMedCrossRef
22.
Zurück zum Zitat Nguyen AH, Perez-Rovira A, Wielopolski PA et al (2019) Technical challenges of quantitative chest MRI data analysis in a large cohort pediatric study. Eur Radiol 29:2770–2782PubMedCrossRef Nguyen AH, Perez-Rovira A, Wielopolski PA et al (2019) Technical challenges of quantitative chest MRI data analysis in a large cohort pediatric study. Eur Radiol 29:2770–2782PubMedCrossRef
23.
Zurück zum Zitat Ciet P, Tiddens HAWM, Wielopolski PA et al (2015) Magnetic resonance imaging in children: common problems and possible solutions for lung and airways imaging. Pediatr Radiol 45:1901–1915PubMedPubMedCentralCrossRef Ciet P, Tiddens HAWM, Wielopolski PA et al (2015) Magnetic resonance imaging in children: common problems and possible solutions for lung and airways imaging. Pediatr Radiol 45:1901–1915PubMedPubMedCentralCrossRef
24.
Zurück zum Zitat Fink C, Puderbach M, Biederer J et al (2007) Lung MRI at 1.5 and 3 Tesla: observer preference study and lesion contrast using five different pulse sequences. Invest Radiol 42:377–383PubMedCrossRef Fink C, Puderbach M, Biederer J et al (2007) Lung MRI at 1.5 and 3 Tesla: observer preference study and lesion contrast using five different pulse sequences. Invest Radiol 42:377–383PubMedCrossRef
25.
Zurück zum Zitat Frericks BB, Meyer BC, Martus P, Wendt M, Wolf KJ, Wacker F (2008) MRI of the thorax during whole-body MRI: evaluation of different MR sequences and comparison to thoracic multidetector computed tomography (MDCT). J Magn Reson Imaging 27:538–545PubMedCrossRef Frericks BB, Meyer BC, Martus P, Wendt M, Wolf KJ, Wacker F (2008) MRI of the thorax during whole-body MRI: evaluation of different MR sequences and comparison to thoracic multidetector computed tomography (MDCT). J Magn Reson Imaging 27:538–545PubMedCrossRef
26.
Zurück zum Zitat Wielpütz MO, Triphan SMF, Ohno Y, Jobst BJ, Kauczor HU (2019) Outracing lung signal decay – potential of ultrashort echo time MRI. Rofo 191:415–423PubMedCrossRef Wielpütz MO, Triphan SMF, Ohno Y, Jobst BJ, Kauczor HU (2019) Outracing lung signal decay – potential of ultrashort echo time MRI. Rofo 191:415–423PubMedCrossRef
27.
Zurück zum Zitat Tyler DJ, Robson MD, Henkelman RM, Young IR, Bydder GM (2007) Magnetic resonance imaging with ultrashort TE (UTE) PULSE sequences: technical considerations. J Magn Reson Imaging 25:279–289PubMedCrossRef Tyler DJ, Robson MD, Henkelman RM, Young IR, Bydder GM (2007) Magnetic resonance imaging with ultrashort TE (UTE) PULSE sequences: technical considerations. J Magn Reson Imaging 25:279–289PubMedCrossRef
28.
Zurück zum Zitat Herrmann KH, Krämer M, Reichenbach JR (2016) Time efficient 3D radial UTE sampling with fully automatic delay compensation on a clinical 3T MR scanner. PLoS One 11:e0150371PubMedPubMedCentralCrossRef Herrmann KH, Krämer M, Reichenbach JR (2016) Time efficient 3D radial UTE sampling with fully automatic delay compensation on a clinical 3T MR scanner. PLoS One 11:e0150371PubMedPubMedCentralCrossRef
29.
Zurück zum Zitat Krämer M, Kollert MR, Brisson NM, Maggioni MB, Duda GN, Reichenbach JR (2020) Immersion of Achilles tendon in phosphate-buffered saline influences T1 and T2* relaxation times: an ex vivo study. NMR Biomed 33:e4288PubMedCrossRef Krämer M, Kollert MR, Brisson NM, Maggioni MB, Duda GN, Reichenbach JR (2020) Immersion of Achilles tendon in phosphate-buffered saline influences T1 and T2* relaxation times: an ex vivo study. NMR Biomed 33:e4288PubMedCrossRef
30.
Zurück zum Zitat Benlala I, Point S, Leung C et al (2020) Volumetric quantification of lung MR signal intensities using ultrashort TE as an automated score in cystic fibrosis. Eur Radiol 30:5479–5488PubMedCrossRef Benlala I, Point S, Leung C et al (2020) Volumetric quantification of lung MR signal intensities using ultrashort TE as an automated score in cystic fibrosis. Eur Radiol 30:5479–5488PubMedCrossRef
31.
Zurück zum Zitat Johnson KM, Fain SB, Schiebler ML, Nagle S (2013) Optimized 3D ultrashort echo time pulmonary MRI. Magn Reson Med 70:1241–1250PubMedCrossRef Johnson KM, Fain SB, Schiebler ML, Nagle S (2013) Optimized 3D ultrashort echo time pulmonary MRI. Magn Reson Med 70:1241–1250PubMedCrossRef
32.
Zurück zum Zitat Krämer M, Herzau B, Reichenbach JR (2020) Segmentation and visualization of the human cranial bone by T2* approximation using ultra-short echo time (UTE) magnetic resonance imaging. Z Med Phys 30:51–59PubMedCrossRef Krämer M, Herzau B, Reichenbach JR (2020) Segmentation and visualization of the human cranial bone by T2* approximation using ultra-short echo time (UTE) magnetic resonance imaging. Z Med Phys 30:51–59PubMedCrossRef
33.
Zurück zum Zitat Krämer M, Maggioni MB, Brisson NM et al (2019) T1 and T2* mapping of the human quadriceps and patellar tendons using ultra-short echo-time (UTE) imaging and bivariate relaxation parameter-based volumetric visualization. Magn Reson Imaging 63:29–36PubMedCrossRef Krämer M, Maggioni MB, Brisson NM et al (2019) T1 and T2* mapping of the human quadriceps and patellar tendons using ultra-short echo-time (UTE) imaging and bivariate relaxation parameter-based volumetric visualization. Magn Reson Imaging 63:29–36PubMedCrossRef
34.
Zurück zum Zitat Zwart NR, Johnson KO, Pipe JG (2012) Efficient sample density estimation by combining gridding and an optimized kernel. Magn Reson Med 67:701–710PubMedCrossRef Zwart NR, Johnson KO, Pipe JG (2012) Efficient sample density estimation by combining gridding and an optimized kernel. Magn Reson Med 67:701–710PubMedCrossRef
35.
Zurück zum Zitat Pierre EY, Grodzki D, Aandal G et al (2014) Parallel imaging-based reduction of acoustic noise for clinical magnetic resonance imaging. Invest Radiol 49:620–626PubMedCrossRef Pierre EY, Grodzki D, Aandal G et al (2014) Parallel imaging-based reduction of acoustic noise for clinical magnetic resonance imaging. Invest Radiol 49:620–626PubMedCrossRef
36.
Zurück zum Zitat Cha MJ, Park HJ, Paek MY et al (2018) Free-breathing ultrashort echo time lung magnetic resonance imaging using stack-of-spirals acquisition: a feasibility study in oncology patients. Magn Reson Imaging 51:137–143PubMedCrossRef Cha MJ, Park HJ, Paek MY et al (2018) Free-breathing ultrashort echo time lung magnetic resonance imaging using stack-of-spirals acquisition: a feasibility study in oncology patients. Magn Reson Imaging 51:137–143PubMedCrossRef
37.
Zurück zum Zitat Hansell DM, Bankier AA, MacMahon H, McLoud TC, Müller NL, Remy J (2008) Fleischner Society: glossary of terms for thoracic imaging. Radiology 246:697–722PubMedCrossRef Hansell DM, Bankier AA, MacMahon H, McLoud TC, Müller NL, Remy J (2008) Fleischner Society: glossary of terms for thoracic imaging. Radiology 246:697–722PubMedCrossRef
38.
Zurück zum Zitat Truong MT, Ko JP, Rossi SE et al (2014) Update in the evaluation of the solitary pulmonary nodule. Radiographics 34:1658–1679PubMedCrossRef Truong MT, Ko JP, Rossi SE et al (2014) Update in the evaluation of the solitary pulmonary nodule. Radiographics 34:1658–1679PubMedCrossRef
39.
Zurück zum Zitat Wielpütz MO, Lee HY, Koyama H et al (2018) Morphologic characterization of pulmonary nodules with ultrashort TE MRI at 3T. AJR Am J Roentgenol 210:1216–1225PubMedCrossRef Wielpütz MO, Lee HY, Koyama H et al (2018) Morphologic characterization of pulmonary nodules with ultrashort TE MRI at 3T. AJR Am J Roentgenol 210:1216–1225PubMedCrossRef
40.
Zurück zum Zitat Bruegel M, Gaa J, Woertler K et al (2007) MRI of the lung: value of different turbo spin-echo, single-shot turbo spin-echo, and 3D gradient-echo pulse sequences for the detection of pulmonary metastases. J Magn Reson Imaging 25:73–81PubMedCrossRef Bruegel M, Gaa J, Woertler K et al (2007) MRI of the lung: value of different turbo spin-echo, single-shot turbo spin-echo, and 3D gradient-echo pulse sequences for the detection of pulmonary metastases. J Magn Reson Imaging 25:73–81PubMedCrossRef
41.
Zurück zum Zitat Ida M, Wakayama T, Nielsen ML, Abe T, Grodzki DM (2015) Quiet T1-weighted imaging using PETRA: initial clinical evaluation in intracranial tumor patients. J Magn Reson Imaging 41:447–453PubMedCrossRef Ida M, Wakayama T, Nielsen ML, Abe T, Grodzki DM (2015) Quiet T1-weighted imaging using PETRA: initial clinical evaluation in intracranial tumor patients. J Magn Reson Imaging 41:447–453PubMedCrossRef
42.
Zurück zum Zitat Higano NS, Fleck RJ, Spielberg DR et al (2017) Quantification of neonatal lung parenchymal density via ultrashort echo time MRI with comparison to CT. J Magn Reson Imaging 46:992–1000PubMedPubMedCentralCrossRef Higano NS, Fleck RJ, Spielberg DR et al (2017) Quantification of neonatal lung parenchymal density via ultrashort echo time MRI with comparison to CT. J Magn Reson Imaging 46:992–1000PubMedPubMedCentralCrossRef
43.
Zurück zum Zitat Ohno Y, Koyama H, Yoshikawa T et al (2016) Pulmonary high-resolution ultrashort TE MR imaging: comparison with thin-section standard- and low-dose computed tomography for the assessment of pulmonary parenchyma diseases. J Magn Reson Imaging 43:512–532PubMedCrossRef Ohno Y, Koyama H, Yoshikawa T et al (2016) Pulmonary high-resolution ultrashort TE MR imaging: comparison with thin-section standard- and low-dose computed tomography for the assessment of pulmonary parenchyma diseases. J Magn Reson Imaging 43:512–532PubMedCrossRef
44.
Zurück zum Zitat Ohno Y, Koyama H, Yoshikawa T et al (2017) Standard-, reduced-, and no-dose thin-section radiologic examinations: comparison of capability for nodule detection and nodule type assessment in patients suspected of having pulmonary nodules. Radiology 284:562–573PubMedCrossRef Ohno Y, Koyama H, Yoshikawa T et al (2017) Standard-, reduced-, and no-dose thin-section radiologic examinations: comparison of capability for nodule detection and nodule type assessment in patients suspected of having pulmonary nodules. Radiology 284:562–573PubMedCrossRef
45.
Zurück zum Zitat Biederer J, Ohno Y, Hatabu H et al (2017) Screening for lung cancer: does MRI have a role? Eur J Radiol 86:353–360PubMedCrossRef Biederer J, Ohno Y, Hatabu H et al (2017) Screening for lung cancer: does MRI have a role? Eur J Radiol 86:353–360PubMedCrossRef
46.
Zurück zum Zitat Sodhi KS, Khandelwal N, Saxena AK et al (2016) Rapid lung MRI in children with pulmonary infections: time to change our diagnostic algorithms. J Magn Reson Imaging 43:1196–1206PubMedCrossRef Sodhi KS, Khandelwal N, Saxena AK et al (2016) Rapid lung MRI in children with pulmonary infections: time to change our diagnostic algorithms. J Magn Reson Imaging 43:1196–1206PubMedCrossRef
47.
Zurück zum Zitat Ley-Zaporozhan J, Ley S, Sommerburg O, Komm N, Müller FM, Schenk JP (2009) Clinical application of MRI in children for the assessment of pulmonary diseases. Rofo 181:419–432PubMedCrossRef Ley-Zaporozhan J, Ley S, Sommerburg O, Komm N, Müller FM, Schenk JP (2009) Clinical application of MRI in children for the assessment of pulmonary diseases. Rofo 181:419–432PubMedCrossRef
48.
Zurück zum Zitat Roach DJ, Crémillieux Y, Fleck RJ et al (2016) Ultrashort echo-time magnetic resonance imaging is a sensitive method for the evaluation of early cystic fibrosis lung disease. Ann Am Thorac Soc 13:1923–1931PubMedPubMedCentralCrossRef Roach DJ, Crémillieux Y, Fleck RJ et al (2016) Ultrashort echo-time magnetic resonance imaging is a sensitive method for the evaluation of early cystic fibrosis lung disease. Ann Am Thorac Soc 13:1923–1931PubMedPubMedCentralCrossRef
49.
Zurück zum Zitat Dournes G, Menut F, Macey J et al (2016) Lung morphology assessment of cystic fibrosis using MRI with ultra-short echo time at submillimeter spatial resolution. Eur Radiol 26:3811–3820PubMedCrossRef Dournes G, Menut F, Macey J et al (2016) Lung morphology assessment of cystic fibrosis using MRI with ultra-short echo time at submillimeter spatial resolution. Eur Radiol 26:3811–3820PubMedCrossRef
50.
Zurück zum Zitat Blumfield E, Moore MM, Drake MK et al (2017) Survey of gadolinium-based contrast agent utilization among the members of the Society for Pediatric Radiology: a quality and safety committee report. Pediatr Radiol 47:665–673PubMedCrossRef Blumfield E, Moore MM, Drake MK et al (2017) Survey of gadolinium-based contrast agent utilization among the members of the Society for Pediatric Radiology: a quality and safety committee report. Pediatr Radiol 47:665–673PubMedCrossRef
Metadaten
Titel
Ultrashort echo time MRI of the lung in children and adolescents: comparison with non-enhanced computed tomography and standard post-contrast T1w MRI sequences
verfasst von
Diane M. Renz
Karl-Heinz Herrmann
Martin Kraemer
Joachim Boettcher
Matthias Waginger
Paul-Christian Krueger
Alexander Pfeil
Florian Streitparth
Karim Kentouche
Bernd Gruhn
Jochen G. Mainz
Martin Stenzel
Ulf K. Teichgraeber
Juergen R. Reichenbach
Hans-Joachim Mentzel
Publikationsdatum
20.10.2021
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 3/2022
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-021-08236-7

Weitere Artikel der Ausgabe 3/2022

European Radiology 3/2022 Zur Ausgabe

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.