Introduction

Placement of tibial component is expected to fulfill both maximum surface coverage and recommended anterior–posterior (AP) alignment in total knee arthroplasty (TKA). The purpose of this study is to evaluate the effect of the tibial baseplate shape on AP axis.

Materials and methods

Virtual surgery of TKA was performed with three-dimensional bone models reconstructed from 77 osteoarthritis varus knees. Two differently designed tibial baseplates, symmetrically and anatomically, were set to the cut surface under posterior slopes of 0°, 3°, and 7°. The AP axes were defined by connecting the geometrical center of the cut surface with the medial edge (axis MED) and medial 1/3 (axis 1/3MED) of patella tendon attachment. We evaluated the overhang rates as well as the most fitting AP axis which passes through the geometric center.

Results

Overhang rates when aligned to axis MED were 12–25% for the symmetrical-type group and 13–22% for the anatomical-type group. Overhang rates when aligned to axis 1/3MED were 42–48% for the symmetrical-type group and 3–7% for the anatomical-type group. The most fitting AP axis of tibial baseplate was located 2.5° external to axis MED for the symmetrical-type group and around 3.3° internal to axis 1/3MED for the anatomical-type group.

Conclusions

Symmetrically or anatomically designed tibial baseplates have their own favored AP axis and specific performance on coverage. When aligned to axis 1/3MED, anatomically designed tibial baseplates will effectively lower the mismatch rates compared to a symmetrically designed tibial baseplate. Orthopaedic surgeons are expected to place the tibial components to the cut surface during TKA with full understanding of the features between different baseplate designs, AP axes, and posterior slopes for an ideal tibial rotational position.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
2.
Zurück zum Zitat Sharkey PF, Lichstein PM, Shen C, Tokarski AT, Parvizi J (2014) Why are total knee arthroplasties failing today—has anything changed after 10 years? J Arthroplasty 29:1774–1778CrossRefPubMed Sharkey PF, Lichstein PM, Shen C, Tokarski AT, Parvizi J (2014) Why are total knee arthroplasties failing today—has anything changed after 10 years? J Arthroplasty 29:1774–1778CrossRefPubMed
4.
Zurück zum Zitat Nishikawa K, Okazaki K, Matsuda S, Tashiro Y, Kawahara S, Nakahara H et al (2013) Improved design decreases wear in total knee arthroplasty with varus malalignment. Knee Surg Sports Traumatol Arthrosc 22:2635–2640CrossRefPubMed Nishikawa K, Okazaki K, Matsuda S, Tashiro Y, Kawahara S, Nakahara H et al (2013) Improved design decreases wear in total knee arthroplasty with varus malalignment. Knee Surg Sports Traumatol Arthrosc 22:2635–2640CrossRefPubMed
5.
Zurück zum Zitat Indelli PF, Marcucci M, Cariello D, Poli P, Innocenti M (2011) Contemporary femoral designs in total knee arthroplasty: effects on the patello-femoral congruence. Int Orthop 36:1167–1173CrossRefPubMedPubMedCentral Indelli PF, Marcucci M, Cariello D, Poli P, Innocenti M (2011) Contemporary femoral designs in total knee arthroplasty: effects on the patello-femoral congruence. Int Orthop 36:1167–1173CrossRefPubMedPubMedCentral
6.
Zurück zum Zitat Kuriyama S, Ishikawa M, Furu M, Ito H, Matsuda S (2014) Malrotated tibial component increases medial collateral ligament tension in total knee arthroplasty. J Orthop Res 32:1658–1666CrossRefPubMed Kuriyama S, Ishikawa M, Furu M, Ito H, Matsuda S (2014) Malrotated tibial component increases medial collateral ligament tension in total knee arthroplasty. J Orthop Res 32:1658–1666CrossRefPubMed
9.
Zurück zum Zitat Kuriyama S, Hyakuna K, Inoue S, Tamaki Y, Ito H, Matsuda S (2014) Tibial rotational alignment was significantly improved by use of a CT-navigated control device in total knee arthroplasty. J Arthroplasty 29:2352–2356CrossRefPubMed Kuriyama S, Hyakuna K, Inoue S, Tamaki Y, Ito H, Matsuda S (2014) Tibial rotational alignment was significantly improved by use of a CT-navigated control device in total knee arthroplasty. J Arthroplasty 29:2352–2356CrossRefPubMed
10.
Zurück zum Zitat Mizu-uchi H, Matsuda S, Miura H, Okazaki K, Akasaki Y, Iwamoto Y (2008) The evaluation of post-operative alignment in total knee replacement using a CT-based navigation system. J Bone Joint Surg Br 90-B:1025–31 Mizu-uchi H, Matsuda S, Miura H, Okazaki K, Akasaki Y, Iwamoto Y (2008) The evaluation of post-operative alignment in total knee replacement using a CT-based navigation system. J Bone Joint Surg Br 90-B:1025–31
12.
Zurück zum Zitat Garg A, Walker PS (1986) The effect of the interface on the bone stresses beneath tibial components. J Biomech 19:957–967CrossRefPubMed Garg A, Walker PS (1986) The effect of the interface on the bone stresses beneath tibial components. J Biomech 19:957–967CrossRefPubMed
14.
Zurück zum Zitat Bozkurt M, Akkaya M, Tahta M, Gursoy S, Firat A (2017) Tibial base plate for total knee arthroplasty: symmetric or asymmetric? Clin Orthop Surg 9:e42CrossRef Bozkurt M, Akkaya M, Tahta M, Gursoy S, Firat A (2017) Tibial base plate for total knee arthroplasty: symmetric or asymmetric? Clin Orthop Surg 9:e42CrossRef
15.
Zurück zum Zitat Dai Y, Scuderi GR, Bischoff JE, Bertin K, Tarabichi S, Rajgopal A (2014) Anatomic tibial component design can increase tibial coverage and rotational alignment accuracy: a comparison of six contemporary designs. Knee Surg Sports Traumatol Arthrosc 22:2911–2923CrossRefPubMedPubMedCentral Dai Y, Scuderi GR, Bischoff JE, Bertin K, Tarabichi S, Rajgopal A (2014) Anatomic tibial component design can increase tibial coverage and rotational alignment accuracy: a comparison of six contemporary designs. Knee Surg Sports Traumatol Arthrosc 22:2911–2923CrossRefPubMedPubMedCentral
16.
Zurück zum Zitat Martin S, Saurez A, Ismaily S, Ashfaq K, Noble P, Incavo SJ (2013) Maximizing tibial coverage is detrimental to proper rotational alignment. Clin Orthop Relat Res 472:121–125CrossRefPubMedCentral Martin S, Saurez A, Ismaily S, Ashfaq K, Noble P, Incavo SJ (2013) Maximizing tibial coverage is detrimental to proper rotational alignment. Clin Orthop Relat Res 472:121–125CrossRefPubMedCentral
17.
Zurück zum Zitat Stulberg SD, Goyal N (2015) Which tibial tray design achieves maximum coverage and ideal rotation: anatomic, symmetric, or asymmetric? An mri-based study. J Arthroplasty 30:1839–1841CrossRefPubMed Stulberg SD, Goyal N (2015) Which tibial tray design achieves maximum coverage and ideal rotation: anatomic, symmetric, or asymmetric? An mri-based study. J Arthroplasty 30:1839–1841CrossRefPubMed
18.
Zurück zum Zitat Cobb JP, Dixon H, Dandachli W, Iranpour F (2008) The anatomical tibial axis: reliable rotational orientation in knee replacement. J Bone Joint Surg Br 90-B:1032–8 Cobb JP, Dixon H, Dandachli W, Iranpour F (2008) The anatomical tibial axis: reliable rotational orientation in knee replacement. J Bone Joint Surg Br 90-B:1032–8
19.
Zurück zum Zitat Ho KCT, Saevarsson SK, Ramm H, Lieck R, Zachow S, Sharma GB et al (2012) Computed tomography analysis of knee pose and geometry before and after total knee arthroplasty. J Biomech 45:2215–2221CrossRefPubMed Ho KCT, Saevarsson SK, Ramm H, Lieck R, Zachow S, Sharma GB et al (2012) Computed tomography analysis of knee pose and geometry before and after total knee arthroplasty. J Biomech 45:2215–2221CrossRefPubMed
20.
Zurück zum Zitat Forster-Horvath C, Kremo V, Müller-Gerbl M, Nowakowski AM (2015) Using the anatomical tibial axis for total knee arthroplasty alignment may lead to an internal rotation error. Int Orthop 39:2347–2353CrossRefPubMed Forster-Horvath C, Kremo V, Müller-Gerbl M, Nowakowski AM (2015) Using the anatomical tibial axis for total knee arthroplasty alignment may lead to an internal rotation error. Int Orthop 39:2347–2353CrossRefPubMed
21.
Zurück zum Zitat Nowakowski AM, Kamphausen M, Pagenstert G, Valderrabano V, Müller-Gerbl M (2014) Influence of tibial slope on extension and flexion gaps in total knee arthroplasty: increasing the tibial slope affects both gaps. Int Orthop 38:2071–2077CrossRefPubMed Nowakowski AM, Kamphausen M, Pagenstert G, Valderrabano V, Müller-Gerbl M (2014) Influence of tibial slope on extension and flexion gaps in total knee arthroplasty: increasing the tibial slope affects both gaps. Int Orthop 38:2071–2077CrossRefPubMed
22.
Zurück zum Zitat Mizu-uchi H, Colwell CW Jr, Matsuda S, Flores-Hernandez C, Iwamoto Y, D’Lima DD (2011) Effect of total knee arthroplasty implant position on flexion angle before implant-bone impingement. Arthroplasty 26:721–727CrossRef Mizu-uchi H, Colwell CW Jr, Matsuda S, Flores-Hernandez C, Iwamoto Y, D’Lima DD (2011) Effect of total knee arthroplasty implant position on flexion angle before implant-bone impingement. Arthroplasty 26:721–727CrossRef
23.
Zurück zum Zitat Okamoto S, Mizu-uchi H, Okazaki K, Hamai S, Tashiro Y, Nakahara H et al (2016) Two-dimensional planning can result in internal rotation of the femoral component in total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 24:229–235CrossRefPubMed Okamoto S, Mizu-uchi H, Okazaki K, Hamai S, Tashiro Y, Nakahara H et al (2016) Two-dimensional planning can result in internal rotation of the femoral component in total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 24:229–235CrossRefPubMed
25.
Zurück zum Zitat Nishikawa K, Mizu-uchi H, Okazaki K, Matsuda S, Tashiro Y, Iwamoto Y (2015) Accuracy of proximal tibial bone cut using anterior border of tibia as bony landmark in total knee arthroplasty. J Arthroplasty 30:2121–2124CrossRefPubMed Nishikawa K, Mizu-uchi H, Okazaki K, Matsuda S, Tashiro Y, Iwamoto Y (2015) Accuracy of proximal tibial bone cut using anterior border of tibia as bony landmark in total knee arthroplasty. J Arthroplasty 30:2121–2124CrossRefPubMed
27.
Zurück zum Zitat Akagi M, Oh M, Nonaka T, Tsujimoto H, Asano T, Hamanishi C (2004) An anteroposterior axis of the tibia for total knee arthroplasty. Clin Orthop Relat Res 213–9 Akagi M, Oh M, Nonaka T, Tsujimoto H, Asano T, Hamanishi C (2004) An anteroposterior axis of the tibia for total knee arthroplasty. Clin Orthop Relat Res 213–9
28.
Zurück zum Zitat Kawahara S, Okazaki K, Matsuda S, Mitsuyasu H, Nakahara H, Okamoto S et al (2013) Medial sixth of the patellar tendon at the tibial attachment is useful for the anterior reference in rotational alignment of the tibial component. Knee Surg Sports Traumatol Arthrosc 22:1070–1075CrossRefPubMed Kawahara S, Okazaki K, Matsuda S, Mitsuyasu H, Nakahara H, Okamoto S et al (2013) Medial sixth of the patellar tendon at the tibial attachment is useful for the anterior reference in rotational alignment of the tibial component. Knee Surg Sports Traumatol Arthrosc 22:1070–1075CrossRefPubMed
29.
Zurück zum Zitat Matsuda S, Miura H, Nagamine R, Mawatari T, Tokunaga M, Nabeyama R et al (2006) Anatomical analysis of the femoral condyle in normal and osteoarthritic knees. J Orthop Res 22:104–109CrossRef Matsuda S, Miura H, Nagamine R, Mawatari T, Tokunaga M, Nabeyama R et al (2006) Anatomical analysis of the femoral condyle in normal and osteoarthritic knees. J Orthop Res 22:104–109CrossRef
30.
Zurück zum Zitat Landis JR, Koch GG (1977) The Measurement of Observer Agreement for Categorical Data. Biometrics 33:159CrossRefPubMed Landis JR, Koch GG (1977) The Measurement of Observer Agreement for Categorical Data. Biometrics 33:159CrossRefPubMed
32.
Zurück zum Zitat Cheng T, Zhang G, Zhang X (2011) Imageless Navigation System Does Not Improve Component Rotational Alignment in Total Knee Arthroplasty. J Surg Res 171:590–600CrossRefPubMed Cheng T, Zhang G, Zhang X (2011) Imageless Navigation System Does Not Improve Component Rotational Alignment in Total Knee Arthroplasty. J Surg Res 171:590–600CrossRefPubMed
33.
Zurück zum Zitat Hernandez-Vaquero D, Noriega-Fernandez A, Fernandez-Carreira JM, Fernandez-Simon JM, de Los Rios JL (2014) Computer-assisted surgery improves rotational positioning of the femoral component but not the tibial component in total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 22:3127–3134CrossRefPubMed Hernandez-Vaquero D, Noriega-Fernandez A, Fernandez-Carreira JM, Fernandez-Simon JM, de Los Rios JL (2014) Computer-assisted surgery improves rotational positioning of the femoral component but not the tibial component in total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 22:3127–3134CrossRefPubMed
34.
Zurück zum Zitat Farronato G, Giannini L, Galbiati G, Pisani L, Mortellaro C, Maspero C (2014) Verification of the Reliability of the Three-dimensional Virtual Presurgical Orthodontic Diagnostic Protocol. J Craniofac Surg 25:2013–2016CrossRefPubMed Farronato G, Giannini L, Galbiati G, Pisani L, Mortellaro C, Maspero C (2014) Verification of the Reliability of the Three-dimensional Virtual Presurgical Orthodontic Diagnostic Protocol. J Craniofac Surg 25:2013–2016CrossRefPubMed
35.
Zurück zum Zitat Victor J, Van Doninck D, Labey L, Innocenti B, Parizel PM, Bellemans J (2009) How precise can bony landmarks be determined on a CT scan of the knee? Knee 16:358–365CrossRefPubMed Victor J, Van Doninck D, Labey L, Innocenti B, Parizel PM, Bellemans J (2009) How precise can bony landmarks be determined on a CT scan of the knee? Knee 16:358–365CrossRefPubMed
Metadaten
Titel
Effects of tibial baseplate shape on rotational alignment in total knee arthroplasty: three-dimensional surgical simulation using osteoarthritis knees
verfasst von
Yuan Ma
Hideki Mizu-uchi
Ken Okazaki
Tetsuro Ushio
Koji Murakami
Satoshi Hamai
Yukio Akasaki
Yasuharu Nakashima
Publikationsdatum
26.10.2017
Verlag
Springer Berlin Heidelberg
Erschienen in
Archives of Orthopaedic and Trauma Surgery / Ausgabe 1/2018
Print ISSN: 0936-8051
Elektronische ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-017-2828-2

Weitere Artikel der Ausgabe 1/2018

Archives of Orthopaedic and Trauma Surgery 1/2018 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

TEP mit Roboterhilfe führt nicht zu größerer Zufriedenheit

15.05.2024 Knie-TEP Nachrichten

Der Einsatz von Operationsrobotern für den Einbau von Totalendoprothesen des Kniegelenks hat die Präzision der Eingriffe erhöht. Für die postoperative Zufriedenheit der Patienten scheint das aber unerheblich zu sein, wie eine Studie zeigt.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.