Reliability of cone beam computed tomography in scalar localization of the electrode array: a radio histological study

verfasst von: Mathieu Marx, Frank Risi, Bernard Escudé, Irfan Durmo, Christopher James, Frédéric Lauwers, Olivier Deguine, Bernard Fraysse

Erschienen in: European Archives of Oto-Rhino-Laryngology | Ausgabe 4/2014

Einloggen, um Zugang zu erhalten

Abstract

Postoperative imaging plays a growing role in clinical studies concerning prognostic factors in cochlear implantation. Indeed, intracochlear position of the cochlear implant has recently been identified as a contributor in functional outcomes and radiological tools must be accurate enough to determine the final placement of the electrode array. The aim of our study was to validate cone beam computed tomography as a reliable technique for scalar localization of the electrode array. We performed therefore a temporal bone study on ten specimens that were implanted with a perimodiolar implant prototype. Cone beam reconstructions were performed and images were analyzed by two physicians both experienced in cochlear implant imaging, who determined the scalar localization of the implant. Temporal bones then underwent histological control to document this scalar localization and hypothetical intracochlear lesions. In four cases, a dislocation from scala tympani to scala vestibuli was suspected on cone beam reconstructions of the ascending part of the basal turn. In three of these four specimens, dislocation in pars ascendens was confirmed histologically. In the remaining temporal bone, histological analysis revealed an elevation with rupture of the basilar membrane. Histological assessment revealed spiral ligament tearing in another bone. We conclude that cone beam is a reliable tool to assess scalar localization of the selectrode array and may be used in future clinical studies.
Literatur
1.
Zurück zum Zitat Blamey P, Arndt P, Bergeron F, Bredberg G, Brimacombe J, Facer G, Larky J, Lindstrom B, Nedzelski J, Peterson A, Shipp D, Staller S, Whitford L (1996) Factors affecting auditory performance of postlinguistically deaf adults using cochlear implants. Audiol Neurootol 1:293–306PubMedCrossRef Blamey P, Arndt P, Bergeron F, Bredberg G, Brimacombe J, Facer G, Larky J, Lindstrom B, Nedzelski J, Peterson A, Shipp D, Staller S, Whitford L (1996) Factors affecting auditory performance of postlinguistically deaf adults using cochlear implants. Audiol Neurootol 1:293–306PubMedCrossRef
2.
Zurück zum Zitat UK Cochlear Implant Study Group (2004) Criteria of candidacy for unilateral cochlear implantation in postlingually deafened adults iii: prospective evaluation of an actuarial approach to defining a criterion. Ear Hear 25:361–374 UK Cochlear Implant Study Group (2004) Criteria of candidacy for unilateral cochlear implantation in postlingually deafened adults iii: prospective evaluation of an actuarial approach to defining a criterion. Ear Hear 25:361–374
3.
Zurück zum Zitat Blamey P, Artieres F, Baskent D, Bergeron F, Beynon A, Burke E, Dillier N, Dowell R, Fraysse B, Gallego S, Govaerts PJ, Green K, Huber AM, Kleine-Punte A, Maat B, Marx M, Mawman D, Mosnier I, O’Connor AF, O’Leary S, Rousset A, Schauwers K, Skarzynski H, Skarzynski PH, Sterkers O, Terranti A, Truy E, Van de Heyning P, Venail F, Vincent C, Lazard DS (2012) Factors affecting auditory performance of postlinguistically deaf adults using cochlear implants: an update with 2251 patients. Audiol Neurootol 18:36–47PubMedCrossRef Blamey P, Artieres F, Baskent D, Bergeron F, Beynon A, Burke E, Dillier N, Dowell R, Fraysse B, Gallego S, Govaerts PJ, Green K, Huber AM, Kleine-Punte A, Maat B, Marx M, Mawman D, Mosnier I, O’Connor AF, O’Leary S, Rousset A, Schauwers K, Skarzynski H, Skarzynski PH, Sterkers O, Terranti A, Truy E, Van de Heyning P, Venail F, Vincent C, Lazard DS (2012) Factors affecting auditory performance of postlinguistically deaf adults using cochlear implants: an update with 2251 patients. Audiol Neurootol 18:36–47PubMedCrossRef
5.
Zurück zum Zitat Aschendorff A, Kromeier J, Klenzner T, Laszig R (2007) Quality control after insertion of the nucleus contour and contour advance electrode in adults. Ear Hear 28:75S–79SPubMedCrossRef Aschendorff A, Kromeier J, Klenzner T, Laszig R (2007) Quality control after insertion of the nucleus contour and contour advance electrode in adults. Ear Hear 28:75S–79SPubMedCrossRef
6.
Zurück zum Zitat Finley CC, Holden TA, Holden LK, Whiting BR, Chole RA, Neely GJ, Hullar TE, Skinner MW (2008) Role of electrode placement as a contributor to variability in cochlear implant outcomes. Otol Neurotol 29:920–928PubMedCentralPubMedCrossRef Finley CC, Holden TA, Holden LK, Whiting BR, Chole RA, Neely GJ, Hullar TE, Skinner MW (2008) Role of electrode placement as a contributor to variability in cochlear implant outcomes. Otol Neurotol 29:920–928PubMedCentralPubMedCrossRef
7.
Zurück zum Zitat Shepherd RK, Hatsushika S, Clark GM (1993) Electrical stimulation of the auditory nerve: the effect of electrode position on neural excitation. Hear Res 66:108–120PubMedCrossRef Shepherd RK, Hatsushika S, Clark GM (1993) Electrical stimulation of the auditory nerve: the effect of electrode position on neural excitation. Hear Res 66:108–120PubMedCrossRef
8.
Zurück zum Zitat Lecerf P, Bakhos D, Cottier JP, Lescanne E, Trijolet JP, Robier A (2011) Midmodiolar reconstruction as a valuable tool to determine the exact position of the cochlear implant electrode array. Otol Neurotol 32:1075–1081PubMedCrossRef Lecerf P, Bakhos D, Cottier JP, Lescanne E, Trijolet JP, Robier A (2011) Midmodiolar reconstruction as a valuable tool to determine the exact position of the cochlear implant electrode array. Otol Neurotol 32:1075–1081PubMedCrossRef
9.
Zurück zum Zitat Dahmani-Causse M, Marx M, Deguine O, Fraysse B, Lepage B, Escude B (2011) Morphologic examination of the temporal bone by cone beam computed tomography: comparison with multislice helical computed tomography. Eur Ann Otorhinolaryngol Head Neck Dis 128:230–235PubMedCrossRef Dahmani-Causse M, Marx M, Deguine O, Fraysse B, Lepage B, Escude B (2011) Morphologic examination of the temporal bone by cone beam computed tomography: comparison with multislice helical computed tomography. Eur Ann Otorhinolaryngol Head Neck Dis 128:230–235PubMedCrossRef
10.
Zurück zum Zitat Cushing SL, Daly MJ, Treaba CG, Chan H, Irish JC, Blaser S, Gordon KA, Papsin BC (2012) High-resolution cone-beam computed tomography: a potential tool to improve atraumatic electrode design and position. Acta Otolaryngol 132:361–368PubMedCrossRef Cushing SL, Daly MJ, Treaba CG, Chan H, Irish JC, Blaser S, Gordon KA, Papsin BC (2012) High-resolution cone-beam computed tomography: a potential tool to improve atraumatic electrode design and position. Acta Otolaryngol 132:361–368PubMedCrossRef
11.
Zurück zum Zitat Briggs RJ, Tykocinski M, Lazsig R, Aschendorff A, Lenarz T, Stover T, Fraysse B, Marx M, Roland JT Jr, Roland PS, Wright CG, Gantz BJ, Patrick JF, Risi F (2011) Development and evaluation of the modiolar research array–multi-centre collaborative study in human temporal bones. Cochlear Implants Int 12:129–139PubMedCentralPubMedCrossRef Briggs RJ, Tykocinski M, Lazsig R, Aschendorff A, Lenarz T, Stover T, Fraysse B, Marx M, Roland JT Jr, Roland PS, Wright CG, Gantz BJ, Patrick JF, Risi F (2011) Development and evaluation of the modiolar research array–multi-centre collaborative study in human temporal bones. Cochlear Implants Int 12:129–139PubMedCentralPubMedCrossRef
12.
Zurück zum Zitat Biedron S, Prescher A, Ilgner J, Westhofen M (2010) The internal dimensions of the cochlear scalae with special reference to cochlear electrode insertion trauma. Otol Neurotol 31:731–737PubMedCrossRef Biedron S, Prescher A, Ilgner J, Westhofen M (2010) The internal dimensions of the cochlear scalae with special reference to cochlear electrode insertion trauma. Otol Neurotol 31:731–737PubMedCrossRef
13.
Zurück zum Zitat Briggs RJ, Tykocinski M, Xu J, Risi F, Svehla M, Cowan R, Stover T, Erfurt P, Lenarz T (2006) Comparison of round window and cochleostomy approaches with a prototype hearing preservation electrode. Audiol Neurootol 11(Suppl 1):42–48PubMedCrossRef Briggs RJ, Tykocinski M, Xu J, Risi F, Svehla M, Cowan R, Stover T, Erfurt P, Lenarz T (2006) Comparison of round window and cochleostomy approaches with a prototype hearing preservation electrode. Audiol Neurootol 11(Suppl 1):42–48PubMedCrossRef
14.
Zurück zum Zitat Lenarz T, Stover T, Buechner A, Paasche G, Briggs R, Risi F, Pesch J, Battmer RD (2006) Temporal bone results and hearing preservation with a new straight electrode. Audiol Neurootol 11(Suppl 1):34–41PubMedCrossRef Lenarz T, Stover T, Buechner A, Paasche G, Briggs R, Risi F, Pesch J, Battmer RD (2006) Temporal bone results and hearing preservation with a new straight electrode. Audiol Neurootol 11(Suppl 1):34–41PubMedCrossRef
15.
Zurück zum Zitat Eshraghi AA, Yang NW, Balkany TJ (2003) Comparative study of cochlear damage with three perimodiolar electrode designs. Laryngoscope 113:415–419PubMedCrossRef Eshraghi AA, Yang NW, Balkany TJ (2003) Comparative study of cochlear damage with three perimodiolar electrode designs. Laryngoscope 113:415–419PubMedCrossRef
16.
Zurück zum Zitat Lane JI, Driscoll CL, Witte RJ, Primak A, Lindell EP (2007) Scalar localization of the electrode array after cochlear implantation: a cadaveric validation study comparing 64-slice multidetector computed tomography with microcomputed tomography. Otol Neurotol 28:191–194PubMedCrossRef Lane JI, Driscoll CL, Witte RJ, Primak A, Lindell EP (2007) Scalar localization of the electrode array after cochlear implantation: a cadaveric validation study comparing 64-slice multidetector computed tomography with microcomputed tomography. Otol Neurotol 28:191–194PubMedCrossRef
17.
Zurück zum Zitat Lane JI, Witte RJ, Driscoll CL, Shallop JK, Beatty CW, Primak AN (2007) Scalar localization of the electrode array after cochlear implantation: clinical experience using 64-slice multidetector computed tomography. Otol Neurotol 28:658–662PubMedCrossRef Lane JI, Witte RJ, Driscoll CL, Shallop JK, Beatty CW, Primak AN (2007) Scalar localization of the electrode array after cochlear implantation: clinical experience using 64-slice multidetector computed tomography. Otol Neurotol 28:658–662PubMedCrossRef
18.
Zurück zum Zitat Teymouri J, Hullar TE, Holden TA, Chole RA (2011) Verification of computed tomographic estimates of cochlear implant array position: a micro-ct and histologic analysis. Otol Neurotol 32:980–986PubMedCentralPubMedCrossRef Teymouri J, Hullar TE, Holden TA, Chole RA (2011) Verification of computed tomographic estimates of cochlear implant array position: a micro-ct and histologic analysis. Otol Neurotol 32:980–986PubMedCentralPubMedCrossRef
19.
Zurück zum Zitat Aschendorff A, Kubalek R, Turowski B, Zanella F, Hochmuth A, Schumacher M, Klenzner T, Laszig R (2005) Quality control after cochlear implant surgery by means of rotational tomography. Otol Neurotol 26:34–37PubMedCrossRef Aschendorff A, Kubalek R, Turowski B, Zanella F, Hochmuth A, Schumacher M, Klenzner T, Laszig R (2005) Quality control after cochlear implant surgery by means of rotational tomography. Otol Neurotol 26:34–37PubMedCrossRef
20.
Zurück zum Zitat Wardrop P, Whinney D, Rebscher SJ, Roland JT Jr, Luxford W, Leake PA (2005) A temporal bone study of insertion trauma and intracochlear position of cochlear implant electrodes. I: comparison of nucleus banded and nucleus contour electrodes. Hear Res 203:54–67PubMedCrossRef Wardrop P, Whinney D, Rebscher SJ, Roland JT Jr, Luxford W, Leake PA (2005) A temporal bone study of insertion trauma and intracochlear position of cochlear implant electrodes. I: comparison of nucleus banded and nucleus contour electrodes. Hear Res 203:54–67PubMedCrossRef
21.
Zurück zum Zitat Stover T, Issing P, Graurock G, Erfurt P, ElBeltagy Y, Paasche G, Lenarz T (2005) Evaluation of the advance off-stylet insertion technique and the cochlear insertion tool in temporal bones. Otol Neurotol 26:1161–1170PubMedCrossRef Stover T, Issing P, Graurock G, Erfurt P, ElBeltagy Y, Paasche G, Lenarz T (2005) Evaluation of the advance off-stylet insertion technique and the cochlear insertion tool in temporal bones. Otol Neurotol 26:1161–1170PubMedCrossRef
22.
Zurück zum Zitat Handzel O, Burgess BJ, Nadol JB Jr (2006) Histopathology of the peripheral vestibular system after cochlear implantation in the human. Otol Neurotol 27:57–64PubMedCrossRef Handzel O, Burgess BJ, Nadol JB Jr (2006) Histopathology of the peripheral vestibular system after cochlear implantation in the human. Otol Neurotol 27:57–64PubMedCrossRef
23.
Zurück zum Zitat Tien HC, Linthicum FH Jr (2002) Histopathologic changes in the vestibule after cochlear implantation. Otolaryngol Head Neck Surg 127:260–264PubMedCrossRef Tien HC, Linthicum FH Jr (2002) Histopathologic changes in the vestibule after cochlear implantation. Otolaryngol Head Neck Surg 127:260–264PubMedCrossRef
24.
Zurück zum Zitat Todt I, Basta D, Ernst A (2008) Does the surgical approach in cochlear implantation influence the occurrence of postoperative vertigo? Otolaryngol Head Neck Surg 138:8–12PubMedCrossRef Todt I, Basta D, Ernst A (2008) Does the surgical approach in cochlear implantation influence the occurrence of postoperative vertigo? Otolaryngol Head Neck Surg 138:8–12PubMedCrossRef
Metadaten
Titel
Reliability of cone beam computed tomography in scalar localization of the electrode array: a radio histological study
verfasst von
Mathieu Marx
Frank Risi
Bernard Escudé
Irfan Durmo
Christopher James
Frédéric Lauwers
Olivier Deguine
Bernard Fraysse
Publikationsdatum
01.04.2014
Verlag
Springer Berlin Heidelberg
Erschienen in
European Archives of Oto-Rhino-Laryngology / Ausgabe 4/2014
Print ISSN: 0937-4477
Elektronische ISSN: 1434-4726
DOI
https://doi.org/10.1007/s00405-013-2448-6

Weitere Artikel der Ausgabe 4/2014

European Archives of Oto-Rhino-Laryngology 4/2014 Zur Ausgabe

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Betalaktam-Allergie: praxisnahes Vorgehen beim Delabeling

16.05.2024 Pädiatrische Allergologie Nachrichten

Die große Mehrheit der vermeintlichen Penicillinallergien sind keine. Da das „Etikett“ Betalaktam-Allergie oft schon in der Kindheit erworben wird, kann ein frühzeitiges Delabeling lebenslange Vorteile bringen. Ein Team von Pädiaterinnen und Pädiatern aus Kanada stellt vor, wie sie dabei vorgehen.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Eingreifen von Umstehenden rettet vor Erstickungstod

15.05.2024 Fremdkörperaspiration Nachrichten

Wer sich an einem Essensrest verschluckt und um Luft ringt, benötigt vor allem rasche Hilfe. Dass Umstehende nur in jedem zweiten Erstickungsnotfall bereit waren, diese zu leisten, ist das ernüchternde Ergebnis einer Beobachtungsstudie aus Japan. Doch es gibt auch eine gute Nachricht.

Update HNO

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert – ganz bequem per eMail.