Skip to main content
Erschienen in: European Spine Journal 7/2015

01.11.2015 | Original Article

Pedicle screw placement accuracy in thoracic and lumbar spinal surgery with a patient-matched targeting guide: a cadaveric study

verfasst von: Claudio Lamartina, Riccardo Cecchinato, Zsolt Fekete, Alberto Lipari, Meinrad Fiechter, P. Berjano

Erschienen in: European Spine Journal | Sonderheft 7/2015

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Pedicle screw placement is an increasingly common procedure for the correction of spine degenerative disease, deformity and trauma. However, screw placement is demanding, with complications resulting from inaccurate screw placement. While several different techniques have been developed to improve accuracy, they all have their limitations.

Methods

We examined the MySpine (Medacta International SA, Castel San Pietro, CH) patient-matched pedicle targeting guide in three cadaveric spine specimens operated on by three surgeons. A three-dimensional (3D) preoperative plan was constructed from spinal computed tomography scans, from which individualised guides were developed for the placement of Medacta Unconstrained Screw Technology pedicle screws. Following screw placement, the 3D positioning of the screws was compared to the preoperative plan against a series of pre-defined criteria.

Results

Of 46 inserted screws eligible for assessment, 91.3 % were fully inside the pedicle. There were no cases of Grade B (2–4 mm) or C (>4 mm) pedicle perforation. The mean deviation between the planned and actual screw position at the midpoint of the pedicle was 0.70 mm, the mean horizontal deviation was 0.60 mm and the mean vertical deviation was 0.77 mm. The mean angular deviation in the sagittal plane was 1.74°, versus 1.32° in the transverse plane. The mean deviation in screw depth was 1.55 mm. On all measures, the accuracy of screw placement was within the predefined criteria.

Conclusions

Our cadaver study indicates that pedicle screw placement with the system is accurate and should be investigated in larger in vitro and in vivo studies.
Literatur
1.
Zurück zum Zitat Metz LN, Burch S (2008) Computer-assisted surgical planning and image-guided surgical navigation in refractory adult scoliosis surgery: case report and review of the literature. Spine (Phila Pa 1976) 33:E287–E292. doi:10.1097/BRS.0b013e31816d256e CrossRef Metz LN, Burch S (2008) Computer-assisted surgical planning and image-guided surgical navigation in refractory adult scoliosis surgery: case report and review of the literature. Spine (Phila Pa 1976) 33:E287–E292. doi:10.​1097/​BRS.​0b013e31816d256e​ CrossRef
2.
Zurück zum Zitat Gaines RW Jr (2000) The use of pedicle-screw internal fixation for the operative treatment of spinal disorders. J Bone Joint Surg Am 82-A:1458–1476PubMed Gaines RW Jr (2000) The use of pedicle-screw internal fixation for the operative treatment of spinal disorders. J Bone Joint Surg Am 82-A:1458–1476PubMed
3.
Zurück zum Zitat Richter M, Wilke HJ, Kluger P, Neller S, Claes L, Puhl W (2000) Biomechanical evaluation of a new modular rod-screw implant system for posterior instrumentation of the occipito-cervical spine: in vitro comparison with two established implant systems. Eur Spine J 9:417–425PubMedCentralCrossRefPubMed Richter M, Wilke HJ, Kluger P, Neller S, Claes L, Puhl W (2000) Biomechanical evaluation of a new modular rod-screw implant system for posterior instrumentation of the occipito-cervical spine: in vitro comparison with two established implant systems. Eur Spine J 9:417–425PubMedCentralCrossRefPubMed
4.
Zurück zum Zitat Schmidt R, Wilke HJ, Claes L, Puhl W, Richter M (2003) Pedicle screws enhance primary stability in multilevel cervical corpectomies: biomechanical in vitro comparison of different implants including constrained and nonconstrained posterior instumentations. Spine (Phila Pa 1976) 28:1821–1828. doi:10.1097/01.BRS.0000083287.23521.48 CrossRef Schmidt R, Wilke HJ, Claes L, Puhl W, Richter M (2003) Pedicle screws enhance primary stability in multilevel cervical corpectomies: biomechanical in vitro comparison of different implants including constrained and nonconstrained posterior instumentations. Spine (Phila Pa 1976) 28:1821–1828. doi:10.​1097/​01.​BRS.​0000083287.​23521.​48 CrossRef
5.
Zurück zum Zitat Weinstein JN, Rydevik BL, Rauschning W (1992) Anatomic and technical considerations of pedicle screw fixation. Clin Orthop Relat Res 284:34–46PubMed Weinstein JN, Rydevik BL, Rauschning W (1992) Anatomic and technical considerations of pedicle screw fixation. Clin Orthop Relat Res 284:34–46PubMed
6.
Zurück zum Zitat Merloz P, Troccaz J, Vouaillat H, Vasile C, Tonetti J, Eid A, Plaweski S (2007) Fluoroscopy-based navigation system in spine surgery. Proc Inst Mech Eng H 221:813–820CrossRefPubMed Merloz P, Troccaz J, Vouaillat H, Vasile C, Tonetti J, Eid A, Plaweski S (2007) Fluoroscopy-based navigation system in spine surgery. Proc Inst Mech Eng H 221:813–820CrossRefPubMed
8.
Zurück zum Zitat Brasiliense LB, Theodore N, Lazaro BC, Sayed ZA, Deniz FE, Sonntag VK, Crawford NR (2010) Quantitative analysis of misplaced pedicle screws in the thoracic spine: how much pullout strength is lost?: presented at the 2009 Joint Spine Section Meeting. J Neurosurg Spine 12:503–508. doi:10.3171/2009.11.SPINE09408 CrossRefPubMed Brasiliense LB, Theodore N, Lazaro BC, Sayed ZA, Deniz FE, Sonntag VK, Crawford NR (2010) Quantitative analysis of misplaced pedicle screws in the thoracic spine: how much pullout strength is lost?: presented at the 2009 Joint Spine Section Meeting. J Neurosurg Spine 12:503–508. doi:10.​3171/​2009.​11.​SPINE09408 CrossRefPubMed
9.
Zurück zum Zitat Kim YJ, Lenke LG (2005) Thoracic pedicle screw placement: free-hand technique. Neurol India 53:512–519CrossRefPubMed Kim YJ, Lenke LG (2005) Thoracic pedicle screw placement: free-hand technique. Neurol India 53:512–519CrossRefPubMed
10.
Zurück zum Zitat Lonstein JE, Denis F, Perra JH, Pinto MR, Smith MD, Winter RB (1999) Complications associated with pedicle screws. J Bone Joint Surg Am 81:1519–1528PubMed Lonstein JE, Denis F, Perra JH, Pinto MR, Smith MD, Winter RB (1999) Complications associated with pedicle screws. J Bone Joint Surg Am 81:1519–1528PubMed
11.
Zurück zum Zitat Chen L, Xu L, Qiu Y et al (2015) The risks of aorta impingement from pedicle screw may increase due to aorta movement during posterior instrumentation in Lenke 5C curve: a computed tomography study. Eur Spine J 24:1481–1489CrossRefPubMed Chen L, Xu L, Qiu Y et al (2015) The risks of aorta impingement from pedicle screw may increase due to aorta movement during posterior instrumentation in Lenke 5C curve: a computed tomography study. Eur Spine J 24:1481–1489CrossRefPubMed
13.
14.
Zurück zum Zitat Yang L, Wang X, Zhou L, Fu Q (2013) A systematic review and meta-analysis of randomized controlled trials of unilateral versus bilateral Kyphoplasty for osteoporotic vertebral compression fractures. Pain Physician 16:277–290PubMed Yang L, Wang X, Zhou L, Fu Q (2013) A systematic review and meta-analysis of randomized controlled trials of unilateral versus bilateral Kyphoplasty for osteoporotic vertebral compression fractures. Pain Physician 16:277–290PubMed
16.
Zurück zum Zitat Silbermann J, Riese F, Allam Y, Reichert T, Koeppert H, Gutberlet M (2011) Computer tomography assessment of pedicle screw placement in lumbar and sacral spine: comparison between free-hand and O-arm based navigation techniques. Eur Spine J 20:875–881. doi:10.1007/s00586-010-1683-4 PubMedCentralCrossRefPubMed Silbermann J, Riese F, Allam Y, Reichert T, Koeppert H, Gutberlet M (2011) Computer tomography assessment of pedicle screw placement in lumbar and sacral spine: comparison between free-hand and O-arm based navigation techniques. Eur Spine J 20:875–881. doi:10.​1007/​s00586-010-1683-4 PubMedCentralCrossRefPubMed
19.
Zurück zum Zitat Lamartina C, Capuzzo A, Cecchinato R et al (2014) Adolescent idiopathic scoliosis surgery with patient-specific screw placement-guides. Eur Spine J 23(12):2765–2766CrossRefPubMed Lamartina C, Capuzzo A, Cecchinato R et al (2014) Adolescent idiopathic scoliosis surgery with patient-specific screw placement-guides. Eur Spine J 23(12):2765–2766CrossRefPubMed
20.
Zurück zum Zitat Landi A, Gregori F, Mancarella C, Delfini R (2015) 3D printed tubular guides for pedicle screw placement: the answer for the need of a greater accuracy in spinal stabilization. Orthop Muscular Syst 4:e116. doi:10.4127/2161-0533.1000e116 CrossRef Landi A, Gregori F, Mancarella C, Delfini R (2015) 3D printed tubular guides for pedicle screw placement: the answer for the need of a greater accuracy in spinal stabilization. Orthop Muscular Syst 4:e116. doi:10.​4127/​2161-0533.​1000e116 CrossRef
21.
Zurück zum Zitat Gertzbein SD, Robbins SE (1990) Accuracy of pedicular screw placement in vivo. Spine (Phila Pa 1976) 15:11–14CrossRef Gertzbein SD, Robbins SE (1990) Accuracy of pedicular screw placement in vivo. Spine (Phila Pa 1976) 15:11–14CrossRef
22.
Zurück zum Zitat Aoude AA, Fortin M, Figueiredo R et al (2015) Methods to determine pedicle screw placement accuracy in spine surgery: a systematic review. Eur Spine J (Germany) 24(5):p990–1004CrossRef Aoude AA, Fortin M, Figueiredo R et al (2015) Methods to determine pedicle screw placement accuracy in spine surgery: a systematic review. Eur Spine J (Germany) 24(5):p990–1004CrossRef
23.
Zurück zum Zitat Gelalis ID, Paschos NK, Pakos EE, Politis AN, Arnaoutoglou CM, Karageorgos AC, Ploumis A, Xenakis TA (2012) Accuracy of pedicle screw placement: a systematic review of prospective in vivo studies comparing free hand, fluoroscopy guidance and navigation techniques. Eur Spine J 21:247–255. doi:10.1007/s00586-011-2011-3 PubMedCentralCrossRefPubMed Gelalis ID, Paschos NK, Pakos EE, Politis AN, Arnaoutoglou CM, Karageorgos AC, Ploumis A, Xenakis TA (2012) Accuracy of pedicle screw placement: a systematic review of prospective in vivo studies comparing free hand, fluoroscopy guidance and navigation techniques. Eur Spine J 21:247–255. doi:10.​1007/​s00586-011-2011-3 PubMedCentralCrossRefPubMed
25.
Zurück zum Zitat Sugawara T, Higashiyama N, Kaneyama S, Takabatake M, Watanabe N, Uchida F, Sumi M, Mizoi K (2013) Multistep pedicle screw insertion procedure with patient-specific lamina fit-and-lock templates for the thoracic spine: clinical article. J Neurosurg Spine 19:185–190. doi:10.3171/2013.4.SPINE121059 CrossRefPubMed Sugawara T, Higashiyama N, Kaneyama S, Takabatake M, Watanabe N, Uchida F, Sumi M, Mizoi K (2013) Multistep pedicle screw insertion procedure with patient-specific lamina fit-and-lock templates for the thoracic spine: clinical article. J Neurosurg Spine 19:185–190. doi:10.​3171/​2013.​4.​SPINE121059 CrossRefPubMed
31.
Zurück zum Zitat Rampersaud YR, Pik JH, Salonen D, Farooq S (2005) Clinical accuracy of fluoroscopic computer-assisted pedicle screw fixation: a CT analysis. Spine (Phila Pa 1976) 30:E183–E190CrossRef Rampersaud YR, Pik JH, Salonen D, Farooq S (2005) Clinical accuracy of fluoroscopic computer-assisted pedicle screw fixation: a CT analysis. Spine (Phila Pa 1976) 30:E183–E190CrossRef
32.
Zurück zum Zitat Vr V, Kamath V, Shetty AP, Rajasekaran S (2010) Iso-C3D navigation assisted pedicle screw placement in deformities of the cervical and thoracic spine. Indian J Orthop 44:163–168. doi:10.4103/0019-5413.62083 CrossRef Vr V, Kamath V, Shetty AP, Rajasekaran S (2010) Iso-C3D navigation assisted pedicle screw placement in deformities of the cervical and thoracic spine. Indian J Orthop 44:163–168. doi:10.​4103/​0019-5413.​62083 CrossRef
Metadaten
Titel
Pedicle screw placement accuracy in thoracic and lumbar spinal surgery with a patient-matched targeting guide: a cadaveric study
verfasst von
Claudio Lamartina
Riccardo Cecchinato
Zsolt Fekete
Alberto Lipari
Meinrad Fiechter
P. Berjano
Publikationsdatum
01.11.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
European Spine Journal / Ausgabe Sonderheft 7/2015
Print ISSN: 0940-6719
Elektronische ISSN: 1432-0932
DOI
https://doi.org/10.1007/s00586-015-4261-y

Weitere Artikel der Sonderheft 7/2015

European Spine Journal 7/2015 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

TEP mit Roboterhilfe führt nicht zu größerer Zufriedenheit

15.05.2024 Knie-TEP Nachrichten

Der Einsatz von Operationsrobotern für den Einbau von Totalendoprothesen des Kniegelenks hat die Präzision der Eingriffe erhöht. Für die postoperative Zufriedenheit der Patienten scheint das aber unerheblich zu sein, wie eine Studie zeigt.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.