Novel low-dose protocols using cone beam computed tomography in dental medicine: a review focusing on indications, limitations, and future possibilities

verfasst von: Andy W. K. Yeung, Reinhilde Jacobs, Michael M. Bornstein

Erschienen in: Clinical Oral Investigations | Ausgabe 6/2019

Einloggen, um Zugang zu erhalten

Abstract

Objectives

A narrative review on the potential use of low-dose protocols for cone beam computed tomography (CBCT) was conducted to identify indications and their relevance for various dental disciplines.

Materials and methods

Google Scholar was searched using the words “low-dose CBCT”. Reviews, consensus papers, clinical studies, and experimental studies were eligible for the initial screening process, but for data extraction only original articles were selected. Similar search procedures were then performed with the additional search words “pedo,” “ortho,” “endo,” “implant,” “perio,” and “oral surgery.” Furthermore, references of included articles were examined to identify further relevant articles.

Results

After screening, 27 publications remained for the data extraction process. Low-dose protocols have been reported for specialties such as pediatric dentistry (evaluating orofacial clefts, periapical lesions, impacted teeth, and autotransplantation), orthodontics (cephalometric analysis and interim assessment of treatment results), endodontics (detecting root fractures, resorptions and periapical bone loss), implant dentistry (planning implant insertion, evaluating peri-implant fenestration and dehiscence), periodontology (assessing periodontal structures), and oral and maxillofacial surgery (assessing mandibular third molars and TMJs). Nevertheless, most of the literature available is related to non-clinical studies. Furthermore, there is a lack of position statements or guidelines from authoritative bodies regarding the use of low-dose protocols in dental medicine.

Conclusions

Low-dose protocols for CBCT imaging seem to have potential in various disciplines in dental medicine ranging from pediatric dentistry to oral and maxillofacial surgery. Dose reduction is usually achieved by mAs reduction, use of partial rotations, reduced number of projections, and larger voxel sizes, but seldom by kV reduction.

Clinical relevance

Albeit low-dose protocols have potential to result in a reduction of dose exposure for 3D imaging due to dental indications, there is a need to more clearly specify indications and limitations to avoid indiscriminate use of standard and high-dose CBCT scans in the future on the lines of ALARA/ALADA principles.
Literatur
1.
Zurück zum Zitat Mozzo P, Procacci C, Tacconi A, Martini PT, Andreis IB (1998) A new volumetric CT machine for dental imaging based on the cone-beam technique: preliminary results. Eur Radiol 8:1558–1564CrossRefPubMed Mozzo P, Procacci C, Tacconi A, Martini PT, Andreis IB (1998) A new volumetric CT machine for dental imaging based on the cone-beam technique: preliminary results. Eur Radiol 8:1558–1564CrossRefPubMed
2.
Zurück zum Zitat Scarfe WC, Farman AG, Sukovic P (2006) Clinical applications of cone-beam computed tomography in dental practice. J Can Dent Assoc 72:75–80 Scarfe WC, Farman AG, Sukovic P (2006) Clinical applications of cone-beam computed tomography in dental practice. J Can Dent Assoc 72:75–80
4.
Zurück zum Zitat Horner K, Islam M, Flygare L, Tsiklakis K, Whaites E (2009) Basic principles for use of dental cone beam computed tomography: consensus guidelines of the European Academy of Dental and Maxillofacial Radiology. Dentomaxillofac Radiol 38:187–195CrossRef Horner K, Islam M, Flygare L, Tsiklakis K, Whaites E (2009) Basic principles for use of dental cone beam computed tomography: consensus guidelines of the European Academy of Dental and Maxillofacial Radiology. Dentomaxillofac Radiol 38:187–195CrossRef
5.
Zurück zum Zitat Dula K, Bornstein MM, Buser D, Dagassan-Berndt D, Ettlin DA, Filippi A, Gabioud F, Katsaros C, Krastl G, Lambrecht JT (2014) SADMFR guidelines for the use of cone-beam computed tomography/digital volume tomography. Swiss Dent J 124:1169–1183PubMed Dula K, Bornstein MM, Buser D, Dagassan-Berndt D, Ettlin DA, Filippi A, Gabioud F, Katsaros C, Krastl G, Lambrecht JT (2014) SADMFR guidelines for the use of cone-beam computed tomography/digital volume tomography. Swiss Dent J 124:1169–1183PubMed
6.
Zurück zum Zitat Dula K, Benic GI, Bornstein M, Dagassan-Berndt D, Filippi A, Hicklin S, Kissling-Jeger F, Luebbers H-T, Sculean A, Sequeira-Byron P (2015) SADMFR guidelines for the use of cone-beam computed tomography/digital volume tomography. Swiss Dent J 125:945–953PubMed Dula K, Benic GI, Bornstein M, Dagassan-Berndt D, Filippi A, Hicklin S, Kissling-Jeger F, Luebbers H-T, Sculean A, Sequeira-Byron P (2015) SADMFR guidelines for the use of cone-beam computed tomography/digital volume tomography. Swiss Dent J 125:945–953PubMed
8.
Zurück zum Zitat Oenning AC, Jacobs R, Pauwels R, Stratis A, Hedesiu M, Salmon B, Group DR (2018) Cone-beam CT in paediatric dentistry: DIMITRA project position statement. Pediatr Radiol 48:308–316CrossRefPubMed Oenning AC, Jacobs R, Pauwels R, Stratis A, Hedesiu M, Salmon B, Group DR (2018) Cone-beam CT in paediatric dentistry: DIMITRA project position statement. Pediatr Radiol 48:308–316CrossRefPubMed
9.
Zurück zum Zitat Evans CA, Scarfe WC, Ahmad M, Cevidanes LH, Ludlow JB, Palomo JM, Simmons KE, White SC (2013) Clinical recommendations regarding use of cone beam computed tomography in orthodontics. Position statement by the American Academy of Oral and Maxillofacial Radiology. Oral Surg Oral Med Oral Pathol Oral Radiol 116:238–257CrossRef Evans CA, Scarfe WC, Ahmad M, Cevidanes LH, Ludlow JB, Palomo JM, Simmons KE, White SC (2013) Clinical recommendations regarding use of cone beam computed tomography in orthodontics. Position statement by the American Academy of Oral and Maxillofacial Radiology. Oral Surg Oral Med Oral Pathol Oral Radiol 116:238–257CrossRef
10.
Zurück zum Zitat Patel S, Durack C, Abella F, Roig M, Shemesh H, Lambrechts P, Lemberg K (2014) European Society of Endodontology position statement: the use of CBCT in endodontics. Int Endod J 47:502–504CrossRefPubMed Patel S, Durack C, Abella F, Roig M, Shemesh H, Lambrechts P, Lemberg K (2014) European Society of Endodontology position statement: the use of CBCT in endodontics. Int Endod J 47:502–504CrossRefPubMed
11.
Zurück zum Zitat Fayad MI, Nair M, Levin MD, Benavides E, Rubinstein RA, Barghan S, Hirschberg CS, Ruprecht A (2015) AAE and AAOMR joint position statement: use of cone beam computed tomography in endodontics 2015 update. Oral Surg Oral Med Oral Pathol Oral Radiol 120:508–512CrossRef Fayad MI, Nair M, Levin MD, Benavides E, Rubinstein RA, Barghan S, Hirschberg CS, Ruprecht A (2015) AAE and AAOMR joint position statement: use of cone beam computed tomography in endodontics 2015 update. Oral Surg Oral Med Oral Pathol Oral Radiol 120:508–512CrossRef
12.
Zurück zum Zitat Kim DM, Bassir SH (2017) When is cone-beam computed tomography imaging appropriate for diagnostic inquiry in the management of inflammatory periodontitis? An American Academy of Periodontology best evidence review. J Periodontol 88:978–998CrossRefPubMed Kim DM, Bassir SH (2017) When is cone-beam computed tomography imaging appropriate for diagnostic inquiry in the management of inflammatory periodontitis? An American Academy of Periodontology best evidence review. J Periodontol 88:978–998CrossRefPubMed
13.
Zurück zum Zitat Harris D, Horner K, Gröndahl K, Jacobs R, Helmrot E, Benic GI, Bornstein MM, Dawood A, Quirynen M (2012) E.A.O. guidelines for the use of diagnostic imaging in implant dentistry 2011. A consensus workshop organized by the European Association for Osseointegration at the Medical University of Warsaw. Clin Oral Implants Res 23:1243–1253CrossRefPubMed Harris D, Horner K, Gröndahl K, Jacobs R, Helmrot E, Benic GI, Bornstein MM, Dawood A, Quirynen M (2012) E.A.O. guidelines for the use of diagnostic imaging in implant dentistry 2011. A consensus workshop organized by the European Association for Osseointegration at the Medical University of Warsaw. Clin Oral Implants Res 23:1243–1253CrossRefPubMed
14.
Zurück zum Zitat Benavides E, Rios HF, Ganz SD, An C-H, Resnik R, Reardon GT, Feldman SJ, Mah JK, Hatcher D, Kim M-J (2012) Use of cone beam computed tomography in implant dentistry: the International Congress of Oral Implantologists consensus report. Implant Dent 21:78–86CrossRefPubMed Benavides E, Rios HF, Ganz SD, An C-H, Resnik R, Reardon GT, Feldman SJ, Mah JK, Hatcher D, Kim M-J (2012) Use of cone beam computed tomography in implant dentistry: the International Congress of Oral Implantologists consensus report. Implant Dent 21:78–86CrossRefPubMed
15.
Zurück zum Zitat Bornstein MM, Scarfe WC, Vaughn VM, Jacobs R (2014) Cone beam computed tomography in implant dentistry: a systematic review focusing on guidelines, indications, and radiation dose risks. Int J Oral Maxillofac Implants 29:55–77CrossRefPubMed Bornstein MM, Scarfe WC, Vaughn VM, Jacobs R (2014) Cone beam computed tomography in implant dentistry: a systematic review focusing on guidelines, indications, and radiation dose risks. Int J Oral Maxillofac Implants 29:55–77CrossRefPubMed
16.
Zurück zum Zitat White SC, Scarfe WC, Schulze RK, Lurie AG, Douglass JM, Farman AG, Law CS, Levin MD, Sauer RA, Valachovic RW (2014) The image gently in dentistry campaign: promotion of responsible use of maxillofacial radiology in dentistry for children. Oral Surg Oral Med Oral Pathol Oral Radiol 118:257–261CrossRefPubMed White SC, Scarfe WC, Schulze RK, Lurie AG, Douglass JM, Farman AG, Law CS, Levin MD, Sauer RA, Valachovic RW (2014) The image gently in dentistry campaign: promotion of responsible use of maxillofacial radiology in dentistry for children. Oral Surg Oral Med Oral Pathol Oral Radiol 118:257–261CrossRefPubMed
17.
Zurück zum Zitat Mettler FA Jr, Huda W, Yoshizumi TT, Mahesh M (2008) Effective doses in radiology and diagnostic nuclear medicine: a catalog. Radiology 248:254–263CrossRefPubMed Mettler FA Jr, Huda W, Yoshizumi TT, Mahesh M (2008) Effective doses in radiology and diagnostic nuclear medicine: a catalog. Radiology 248:254–263CrossRefPubMed
18.
Zurück zum Zitat Hendee WR, Edwards FM (1986) ALARA and an integrated approach to radiation protection. Semin Nucl Med 16:142–150CrossRefPubMed Hendee WR, Edwards FM (1986) ALARA and an integrated approach to radiation protection. Semin Nucl Med 16:142–150CrossRefPubMed
19.
Zurück zum Zitat Pauwels R, Beinsberger J, Collaert B, Theodorakou C, Rogers J, Walker A, Cockmartin L, Bosmans H, Jacobs R, Bogaerts R (2012) Effective dose range for dental cone beam computed tomography scanners. Eur J Radiol 81:267–271CrossRef Pauwels R, Beinsberger J, Collaert B, Theodorakou C, Rogers J, Walker A, Cockmartin L, Bosmans H, Jacobs R, Bogaerts R (2012) Effective dose range for dental cone beam computed tomography scanners. Eur J Radiol 81:267–271CrossRef
20.
Zurück zum Zitat Jacobs R, Salmon B, Codari M, Hassan B, Bornstein MM (2018) Cone beam computed tomography in implant dentistry: recommendations for clinical use. BMC Oral Health 18:88CrossRefPubMedPubMedCentral Jacobs R, Salmon B, Codari M, Hassan B, Bornstein MM (2018) Cone beam computed tomography in implant dentistry: recommendations for clinical use. BMC Oral Health 18:88CrossRefPubMedPubMedCentral
21.
Zurück zum Zitat Guerrero ME, Jacobs R, Loubele M, Schutyser F, Suetens P, van Steenberghe D (2006) State-of-the-art on cone beam CT imaging for preoperative planning of implant placement. Clin Oral Investig 10:1–7CrossRefPubMed Guerrero ME, Jacobs R, Loubele M, Schutyser F, Suetens P, van Steenberghe D (2006) State-of-the-art on cone beam CT imaging for preoperative planning of implant placement. Clin Oral Investig 10:1–7CrossRefPubMed
22.
Zurück zum Zitat Van der Molen A, Schilham A, Stoop P, Prokop M, Geleijns J (2013) A national survey on radiation dose in CT in the Netherlands. Insights Imaging 4:383–390CrossRefPubMedPubMedCentral Van der Molen A, Schilham A, Stoop P, Prokop M, Geleijns J (2013) A national survey on radiation dose in CT in the Netherlands. Insights Imaging 4:383–390CrossRefPubMedPubMedCentral
23.
Zurück zum Zitat Qu X, Li G, Zhang Z, Ma X (2012) Thyroid shields for radiation dose reduction during cone beam computed tomography scanning for different oral and maxillofacial regions. Eur J Radiol 81:e376–e380CrossRefPubMed Qu X, Li G, Zhang Z, Ma X (2012) Thyroid shields for radiation dose reduction during cone beam computed tomography scanning for different oral and maxillofacial regions. Eur J Radiol 81:e376–e380CrossRefPubMed
24.
Zurück zum Zitat Brenner DJ, Elliston CD, Hall EJ, Berdon WE (2001) Estimated risks of radiation-induced fatal cancer from pediatric CT. AJR Am J Roentgenol 176:289–296CrossRefPubMed Brenner DJ, Elliston CD, Hall EJ, Berdon WE (2001) Estimated risks of radiation-induced fatal cancer from pediatric CT. AJR Am J Roentgenol 176:289–296CrossRefPubMed
25.
Zurück zum Zitat Jacobs R, Pauwels R, Scarfe WC, De Cock C, Dula K, Willems G, Verdonck A, Politis C (2018) Pediatric cleft palate patients show a 3-to 5-fold increase in cumulative radiation exposure from dental radiology compared with an age-and gender-matched population: a retrospective cohort study. Clin Oral Investig 22:1783–1793CrossRefPubMed Jacobs R, Pauwels R, Scarfe WC, De Cock C, Dula K, Willems G, Verdonck A, Politis C (2018) Pediatric cleft palate patients show a 3-to 5-fold increase in cumulative radiation exposure from dental radiology compared with an age-and gender-matched population: a retrospective cohort study. Clin Oral Investig 22:1783–1793CrossRefPubMed
26.
Zurück zum Zitat McCollough CH, Schueler BA, Atwell TD, Braun NN, Regner DM, Brown DL, LeRoy AJ (2007) Radiation exposure and pregnancy: when should we be concerned? Radiographics 27:909–917CrossRef McCollough CH, Schueler BA, Atwell TD, Braun NN, Regner DM, Brown DL, LeRoy AJ (2007) Radiation exposure and pregnancy: when should we be concerned? Radiographics 27:909–917CrossRef
27.
Zurück zum Zitat Kelaranta A, Ekholm M, Toroi P, Kortesniemi M (2015) Radiation exposure to foetus and breasts from dental X-ray examinations: effect of lead shields. Dentomaxillofac Radiol 45:20150095CrossRefPubMedPubMedCentral Kelaranta A, Ekholm M, Toroi P, Kortesniemi M (2015) Radiation exposure to foetus and breasts from dental X-ray examinations: effect of lead shields. Dentomaxillofac Radiol 45:20150095CrossRefPubMedPubMedCentral
28.
Zurück zum Zitat Marcu M, Hedesiu M, Salmon B, Pauwels R, Stratis A, Oenning ACC, Cohen ME, Jacobs R, Baciut M, Roman R (2018) Estimation of the radiation dose for pediatric CBCT indications: a prospective study on ProMax3D. Int J Paediatr Dent 28:300–309CrossRefPubMed Marcu M, Hedesiu M, Salmon B, Pauwels R, Stratis A, Oenning ACC, Cohen ME, Jacobs R, Baciut M, Roman R (2018) Estimation of the radiation dose for pediatric CBCT indications: a prospective study on ProMax3D. Int J Paediatr Dent 28:300–309CrossRefPubMed
29.
Zurück zum Zitat Hidalgo Rivas JA, Horner K, Thiruvenkatachari B, Davies J, Theodorakou C (2015) Development of a low-dose protocol for cone beam CT examinations of the anterior maxilla in children. Br J Radiol 88:20150559CrossRefPubMedPubMedCentral Hidalgo Rivas JA, Horner K, Thiruvenkatachari B, Davies J, Theodorakou C (2015) Development of a low-dose protocol for cone beam CT examinations of the anterior maxilla in children. Br J Radiol 88:20150559CrossRefPubMedPubMedCentral
30.
Zurück zum Zitat Widmann G, Bischel A, Stratis A, Bosmans H, Jacobs R, Gassner E-M, Puelacher W, Pauwels R (2017) Spatial and contrast resolution of ultralow dose dentomaxillofacial CT imaging using iterative reconstruction technology. Dentomaxillofac Radiol 46:20160452CrossRefPubMedPubMedCentral Widmann G, Bischel A, Stratis A, Bosmans H, Jacobs R, Gassner E-M, Puelacher W, Pauwels R (2017) Spatial and contrast resolution of ultralow dose dentomaxillofacial CT imaging using iterative reconstruction technology. Dentomaxillofac Radiol 46:20160452CrossRefPubMedPubMedCentral
31.
Zurück zum Zitat McGuigan M, Duncan H, Horner K (2018) An analysis of effective dose optimization and its impact on image quality and diagnostic efficacy relating to dental cone beam computed tomography (CBCT). Swiss Dent J 128:297–316PubMed McGuigan M, Duncan H, Horner K (2018) An analysis of effective dose optimization and its impact on image quality and diagnostic efficacy relating to dental cone beam computed tomography (CBCT). Swiss Dent J 128:297–316PubMed
32.
Zurück zum Zitat Pauwels R, Zhang G, Theodorakou C, Walker A, Bosmans H, Jacobs R, Bogaerts R, Horner K, Consortium SP (2014) Effective radiation dose and eye lens dose in dental cone beam CT: effect of field of view and angle of rotation. Br J Radiol 87:20130654CrossRefPubMedPubMedCentral Pauwels R, Zhang G, Theodorakou C, Walker A, Bosmans H, Jacobs R, Bogaerts R, Horner K, Consortium SP (2014) Effective radiation dose and eye lens dose in dental cone beam CT: effect of field of view and angle of rotation. Br J Radiol 87:20130654CrossRefPubMedPubMedCentral
33.
Zurück zum Zitat Ludlow J, Timothy R, Walker C, Hunter R, Benavides E, Samuelson D, Scheske M (2015) Effective dose of dental CBCT—a meta analysis of published data and additional data for nine CBCT units. Dentomaxillofac Radiol 44:20140197CrossRef Ludlow J, Timothy R, Walker C, Hunter R, Benavides E, Samuelson D, Scheske M (2015) Effective dose of dental CBCT—a meta analysis of published data and additional data for nine CBCT units. Dentomaxillofac Radiol 44:20140197CrossRef
34.
Zurück zum Zitat Stratis A, Zhang G, Lopez-Rendon X, Politis C, Hermans R, Jacobs R, Bogaerts R, Shaheen E, Bosmans H (2017) Two examples of indication specific radiation dose calculations in dental CBCT and multidetector CT scanners. Phys Med 41:71–77CrossRefPubMed Stratis A, Zhang G, Lopez-Rendon X, Politis C, Hermans R, Jacobs R, Bogaerts R, Shaheen E, Bosmans H (2017) Two examples of indication specific radiation dose calculations in dental CBCT and multidetector CT scanners. Phys Med 41:71–77CrossRefPubMed
35.
Zurück zum Zitat Jones D, Mannocci F, Andiappan M, Brown J, Patel S (2015) The effect of alteration of the exposure parameters of a cone-beam computed tomographic scan on the diagnosis of simulated horizontal root fractures. J Endod 41:520–525CrossRefPubMed Jones D, Mannocci F, Andiappan M, Brown J, Patel S (2015) The effect of alteration of the exposure parameters of a cone-beam computed tomographic scan on the diagnosis of simulated horizontal root fractures. J Endod 41:520–525CrossRefPubMed
36.
Zurück zum Zitat Neves F, Vasconcelos T, Vaz S, Freitas D, Haiter-Neto F (2012) Evaluation of reconstructed images with different voxel sizes of acquisition in the diagnosis of simulated external root resorption using cone beam computed tomography. Int Endod J 45:234–239CrossRefPubMed Neves F, Vasconcelos T, Vaz S, Freitas D, Haiter-Neto F (2012) Evaluation of reconstructed images with different voxel sizes of acquisition in the diagnosis of simulated external root resorption using cone beam computed tomography. Int Endod J 45:234–239CrossRefPubMed
37.
Zurück zum Zitat EzEldeen M, Stratis A, Coucke W, Codari M, Politis C, Jacobs R (2017) As low dose as sufficient quality: optimization of cone-beam computed tomographic scanning protocol for tooth autotransplantation planning and follow-up in children. J Endod 43:210–217CrossRefPubMed EzEldeen M, Stratis A, Coucke W, Codari M, Politis C, Jacobs R (2017) As low dose as sufficient quality: optimization of cone-beam computed tomographic scanning protocol for tooth autotransplantation planning and follow-up in children. J Endod 43:210–217CrossRefPubMed
38.
Zurück zum Zitat Brown AA, Scarfe WC, Scheetz JP, Silveira AM, Farman AG (2009) Linear accuracy of cone beam CT derived 3D images. Angle Orthod 79:150–157CrossRefPubMed Brown AA, Scarfe WC, Scheetz JP, Silveira AM, Farman AG (2009) Linear accuracy of cone beam CT derived 3D images. Angle Orthod 79:150–157CrossRefPubMed
39.
Zurück zum Zitat Cook V, Timock A, Crowe J, Wang M, Covell D (2015) Accuracy of alveolar bone measurements from cone beam computed tomography acquired using varying settings. Orthod Craniofacial Res 18:127–136CrossRef Cook V, Timock A, Crowe J, Wang M, Covell D (2015) Accuracy of alveolar bone measurements from cone beam computed tomography acquired using varying settings. Orthod Craniofacial Res 18:127–136CrossRef
40.
Zurück zum Zitat Kusnoto B, Kaur P, Salem A, Zhang Z, Galang-Boquiren MT, Viana G, Evans CA, Manasse R, Monahan R, BeGole E (2015) Implementation of ultra-low-dose CBCT for routine 2D orthodontic diagnostic radiographs: Cephalometric landmark identification and image quality assessment. Semin Orthod 21:233–247CrossRef Kusnoto B, Kaur P, Salem A, Zhang Z, Galang-Boquiren MT, Viana G, Evans CA, Manasse R, Monahan R, BeGole E (2015) Implementation of ultra-low-dose CBCT for routine 2D orthodontic diagnostic radiographs: Cephalometric landmark identification and image quality assessment. Semin Orthod 21:233–247CrossRef
41.
Zurück zum Zitat Ludlow JB, Walker C (2013) Assessment of phantom dosimetry and image quality of i-CAT FLX cone-beam computed tomography. Am J Orthod Dentofac Orthop 144:802–817CrossRef Ludlow JB, Walker C (2013) Assessment of phantom dosimetry and image quality of i-CAT FLX cone-beam computed tomography. Am J Orthod Dentofac Orthop 144:802–817CrossRef
42.
Zurück zum Zitat Bechara B, McMahan CA, Nasseh I, Geha H, Hayek E, Khawam G, Raad M, Noujeim M (2013) Number of basis images effect on detection of root fractures in endodontically treated teeth using a cone beam computed tomography machine: an in vitro study. Oral Surg Oral Med Oral Pathol Oral Radiol 115:676–681CrossRefPubMed Bechara B, McMahan CA, Nasseh I, Geha H, Hayek E, Khawam G, Raad M, Noujeim M (2013) Number of basis images effect on detection of root fractures in endodontically treated teeth using a cone beam computed tomography machine: an in vitro study. Oral Surg Oral Med Oral Pathol Oral Radiol 115:676–681CrossRefPubMed
43.
Zurück zum Zitat Durack C, Patel S, Davies J, Wilson R, Mannocci F (2011) Diagnostic accuracy of small volume cone beam computed tomography and intraoral periapical radiography for the detection of simulated external inflammatory root resorption. Int Endod J 44:136–147CrossRefPubMed Durack C, Patel S, Davies J, Wilson R, Mannocci F (2011) Diagnostic accuracy of small volume cone beam computed tomography and intraoral periapical radiography for the detection of simulated external inflammatory root resorption. Int Endod J 44:136–147CrossRefPubMed
44.
Zurück zum Zitat Hashem D, Brown JE, Patel S, Mannocci F, Donaldson AN, Watson TF, Banerjee A (2013) An in vitro comparison of the accuracy of measurements obtained from high-and low-resolution cone-beam computed tomography scans. J Endod 39:394–397CrossRefPubMed Hashem D, Brown JE, Patel S, Mannocci F, Donaldson AN, Watson TF, Banerjee A (2013) An in vitro comparison of the accuracy of measurements obtained from high-and low-resolution cone-beam computed tomography scans. J Endod 39:394–397CrossRefPubMed
45.
Zurück zum Zitat Lennon S, Patel S, Foschi F, Wilson R, Davies J, Mannocci F (2011) Diagnostic accuracy of limited-volume cone-beam computed tomography in the detection of periapical bone loss: 360° scans versus 180° scans. Int Endod J 44:1118–1127CrossRefPubMed Lennon S, Patel S, Foschi F, Wilson R, Davies J, Mannocci F (2011) Diagnostic accuracy of limited-volume cone-beam computed tomography in the detection of periapical bone loss: 360° scans versus 180° scans. Int Endod J 44:1118–1127CrossRefPubMed
46.
Zurück zum Zitat Al-Ekrish AA (2012) Effect of exposure time on the accuracy and reliability of cone beam computed tomography in the assessment of dental implant site dimensions in dry skulls. Saudi Dent J 24:127–134CrossRefPubMedPubMedCentral Al-Ekrish AA (2012) Effect of exposure time on the accuracy and reliability of cone beam computed tomography in the assessment of dental implant site dimensions in dry skulls. Saudi Dent J 24:127–134CrossRefPubMedPubMedCentral
47.
Zurück zum Zitat Dawood A, Brown J, Sauret-Jackson V, Purkayastha S (2012) Optimization of cone beam CT exposure for pre-surgical evaluation of the implant site. Dentomaxillofac Radiol 41:70–74CrossRefPubMedPubMedCentral Dawood A, Brown J, Sauret-Jackson V, Purkayastha S (2012) Optimization of cone beam CT exposure for pre-surgical evaluation of the implant site. Dentomaxillofac Radiol 41:70–74CrossRefPubMedPubMedCentral
48.
Zurück zum Zitat de-Azevedo-Vaz SL, de Faria Vasconcelos K, Neves FS, Melo SLS, Campos PSF, Haiter-Neto F (2013) Detection of periimplant fenestration and dehiscence with the use of two scan modes and the smallest voxel sizes of a cone-beam computed tomography device. Oral Surg Oral Med Oral Pathol Oral Radiol 115:121–127CrossRefPubMed de-Azevedo-Vaz SL, de Faria Vasconcelos K, Neves FS, Melo SLS, Campos PSF, Haiter-Neto F (2013) Detection of periimplant fenestration and dehiscence with the use of two scan modes and the smallest voxel sizes of a cone-beam computed tomography device. Oral Surg Oral Med Oral Pathol Oral Radiol 115:121–127CrossRefPubMed
49.
Zurück zum Zitat El Sahili N, David-Tchouda S, Thoret S, Nasseh I, Berberi A, Fortin T (2018) Effect of milliamperage reduction on pre-surgical implant planning using cone beam computed tomography by surgeons of varying experience. J Maxillofac Oral Surg 17:520–530CrossRefPubMed El Sahili N, David-Tchouda S, Thoret S, Nasseh I, Berberi A, Fortin T (2018) Effect of milliamperage reduction on pre-surgical implant planning using cone beam computed tomography by surgeons of varying experience. J Maxillofac Oral Surg 17:520–530CrossRefPubMed
50.
Zurück zum Zitat Liljeholm R, Kadesjö N, Benchimol D, Hellén-Halme K, Shi X-Q (2017) Cone-beam computed tomography with ultra-low dose protocols for pre-implant radiographic assessment: an in vitro study. Eur J Oral Implantol 10:351–359PubMed Liljeholm R, Kadesjö N, Benchimol D, Hellén-Halme K, Shi X-Q (2017) Cone-beam computed tomography with ultra-low dose protocols for pre-implant radiographic assessment: an in vitro study. Eur J Oral Implantol 10:351–359PubMed
51.
Zurück zum Zitat Neves FS, Vasconcelos TV, Campos PS, Haiter-Neto F, Freitas DQ (2014) Influence of scan mode (180/360) of the cone beam computed tomography for preoperative dental implant measurements. Clin Oral Implants Res 25:e155–e158CrossRefPubMed Neves FS, Vasconcelos TV, Campos PS, Haiter-Neto F, Freitas DQ (2014) Influence of scan mode (180/360) of the cone beam computed tomography for preoperative dental implant measurements. Clin Oral Implants Res 25:e155–e158CrossRefPubMed
52.
Zurück zum Zitat Parsa A, Ibrahim N, Hassan B, Motroni A, Van der Stelt P, Wismeijer D (2013) Influence of cone beam CT scanning parameters on grey value measurements at an implant site. Dentomaxillofac Radiol 42:79884780CrossRefPubMedPubMedCentral Parsa A, Ibrahim N, Hassan B, Motroni A, Van der Stelt P, Wismeijer D (2013) Influence of cone beam CT scanning parameters on grey value measurements at an implant site. Dentomaxillofac Radiol 42:79884780CrossRefPubMedPubMedCentral
53.
Zurück zum Zitat Sur J, Seki K, Koizumi H, Nakajima K, Okano T (2010) Effects of tube current on cone-beam computerized tomography image quality for presurgical implant planning in vitro. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 110:e29–e33CrossRefPubMed Sur J, Seki K, Koizumi H, Nakajima K, Okano T (2010) Effects of tube current on cone-beam computerized tomography image quality for presurgical implant planning in vitro. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 110:e29–e33CrossRefPubMed
54.
Zurück zum Zitat Vasconcelos TV, Neves FS, de Freitas DQ, Campos PSF, Watanabe PCA (2014) Influence of the milliamperage settings on cone beam computed tomography imaging for implant planning. Int J Oral Maxillofac Implants 29:1364–1368CrossRefPubMed Vasconcelos TV, Neves FS, de Freitas DQ, Campos PSF, Watanabe PCA (2014) Influence of the milliamperage settings on cone beam computed tomography imaging for implant planning. Int J Oral Maxillofac Implants 29:1364–1368CrossRefPubMed
55.
Zurück zum Zitat Waltrick KB, de Abreu Junior MJN, Corrêa M, Zastrow MD, D'Avila Dutra V (2013) Accuracy of linear measurements and visibility of the mandibular canal of cone-beam computed tomography images with different voxel sizes: an in vitro study. J Periodontol 84:68–77CrossRefPubMed Waltrick KB, de Abreu Junior MJN, Corrêa M, Zastrow MD, D'Avila Dutra V (2013) Accuracy of linear measurements and visibility of the mandibular canal of cone-beam computed tomography images with different voxel sizes: an in vitro study. J Periodontol 84:68–77CrossRefPubMed
56.
Zurück zum Zitat Al-Okshi A, Theodorakou C, Lindh C (2017) Dose optimization for assessment of periodontal structures in cone beam CT examinations. Dentomaxillofac Radiol 46:20160311CrossRefPubMedPubMedCentral Al-Okshi A, Theodorakou C, Lindh C (2017) Dose optimization for assessment of periodontal structures in cone beam CT examinations. Dentomaxillofac Radiol 46:20160311CrossRefPubMedPubMedCentral
57.
Zurück zum Zitat de Melo LPL, Oenning ACC, Nadaes MR, Nejaim Y, Neves FS, Oliveira ML, Freitas DQ (2017) Influence of acquisition parameters on the evaluation of mandibular third molars through cone beam computed tomography. Oral Surg Oral Med Oral Pathol Oral Radiol 124:183–190CrossRef de Melo LPL, Oenning ACC, Nadaes MR, Nejaim Y, Neves FS, Oliveira ML, Freitas DQ (2017) Influence of acquisition parameters on the evaluation of mandibular third molars through cone beam computed tomography. Oral Surg Oral Med Oral Pathol Oral Radiol 124:183–190CrossRef
58.
Zurück zum Zitat Neves FS, de Camargo Souza T, de-Azevedo-Vaz SL, Campos PSF, Bóscolo FN (2014) Influence of cone-beam computed tomography milliamperage settings on image quality of the mandibular third molar region. Oral Radiol 30:27–31CrossRef Neves FS, de Camargo Souza T, de-Azevedo-Vaz SL, Campos PSF, Bóscolo FN (2014) Influence of cone-beam computed tomography milliamperage settings on image quality of the mandibular third molar region. Oral Radiol 30:27–31CrossRef
59.
Zurück zum Zitat Yadav S, Palo L, Mahdian M, Upadhyay M, Tadinada A (2015) Diagnostic accuracy of 2 cone-beam computed tomography protocols for detecting arthritic changes in temporomandibular joints. Am J Orthod Dentofac Orthop 147:339–344CrossRef Yadav S, Palo L, Mahdian M, Upadhyay M, Tadinada A (2015) Diagnostic accuracy of 2 cone-beam computed tomography protocols for detecting arthritic changes in temporomandibular joints. Am J Orthod Dentofac Orthop 147:339–344CrossRef
60.
Zurück zum Zitat da Silva Moura W, Chiqueto K, Pithon GM, Neves LS, Castro R, Henriques JFC (2018) Factors influencing the effective dose associated with CBCT: a systematic review. Clin Oral Investig 23:1319–1330 da Silva Moura W, Chiqueto K, Pithon GM, Neves LS, Castro R, Henriques JFC (2018) Factors influencing the effective dose associated with CBCT: a systematic review. Clin Oral Investig 23:1319–1330
Metadaten
Titel
Novel low-dose protocols using cone beam computed tomography in dental medicine: a review focusing on indications, limitations, and future possibilities
verfasst von
Andy W. K. Yeung
Reinhilde Jacobs
Michael M. Bornstein
Publikationsdatum
25.04.2019
Verlag
Springer Berlin Heidelberg
Erschienen in
Clinical Oral Investigations / Ausgabe 6/2019
Print ISSN: 1432-6981
Elektronische ISSN: 1436-3771
DOI
https://doi.org/10.1007/s00784-019-02907-y

Weitere Artikel der Ausgabe 6/2019

Clinical Oral Investigations 6/2019 Zur Ausgabe

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Update Zahnmedizin

Bestellen Sie unseren kostenlosen Newsletter und bleiben Sie gut informiert – ganz bequem per eMail.